IMX-3B - Operator Manual

1. The document is an operator manual for the IMX-3B Antarix Troco x-ray equipment that provides instructions for proper use and safety warnings. 2. Several warnings are given, including that only qualified personnel should operate the equipment, proper shielding and power supply are required, and precautions must be taken to prevent exposure to radiation. 3. The manual instructs users to carefully read it before operating the equipment and not to use the device for any purpose other than what is described. Technical works and interventions should only be done by personnel trained by the manufacturer.

Uploaded by

nik
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
110 views30 pages

IMX-3B - Operator Manual

1. The document is an operator manual for the IMX-3B Antarix Troco x-ray equipment that provides instructions for proper use and safety warnings. 2. Several warnings are given, including that only qualified personnel should operate the equipment, proper shielding and power supply are required, and precautions must be taken to prevent exposure to radiation. 3. The manual instructs users to carefully read it before operating the equipment and not to use the device for any purpose other than what is described. Technical works and interventions should only be done by personnel trained by the manufacturer.

Uploaded by

nik
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 30

IMX-3B

ANTARIX TROCO

Operator Manual
Code M70A Rev. 11

IMAGO RADIOLOGY S.r.l.


20081 ABBIATEGRASSO (MI) - Offices: Viale G.G. Sforza, 95 – Tel.+39029464120 / +39029465996 - Fax +39029450043
Partita IVA 09519270152 - C.C.I.A.A. Milano n. 1295725 - Trib. Milano Registro Società 288164 n° 2273 vol. 7344
Website: www.imagoradiology.it e-mail: [email protected]
Imago Radiology S.r.l. IMX-3B Antarix Troco

MANUAL
PREPARED ANDREA TROMBONI
CHECKED MORENO BELTRAMINI
APPROVED FABIO CAMPI
DATE OF ISSUE 29/03/2016

REVISION INDEX
M70A Manual drawing up
Modified paragraphs :
10.10.2002 M70A _01
3 (3.4) _ 6 _ 9 .
Modified paragraphs :
03.07.2003 M70A _02
3 (3.3-3.4) _ 4 (4.1-4.2) _ 9 .
03.01.2004 M70A _03 New Release
Modified paragraphs :
29.09.2004 M70A _04
1(1.2-1.3) _4 (4.1-4.2) _ 9 .
Modified paragraphs :
01.06.2005 M70A _05
1 _ 2 _3 _ 6 _ 9 .
17.03.2006 M70A _06 New Release
Modified paragraphs :
10.10.2006 M70A _07 VALIDITY of S/N 702057
0_1_2 _3_4_5_6_9.
Modified paragraphs :
20.11.2007 M70A _08 VALIDITY of S/N 702521
6_9.
Rdm
11.05.2011 M70A _09 New Release VALIDITY of S/N 702961
N° 384
Modified paragraphs : 2
Rdm
01.02.2016 M70A _10 Modified paragraphs on Technical manual : VALIDITY of S/N 703713
N° 417
7_8.
Rdm Modified paragraphs on Technical manual :
29.03.2016 M70A _11 VALIDITY of S/N 703684
N° 419 7 (K101100) _ 8 .

The original version of this manual is in ITALIAN language .

Operator’s Manual (M70A_11) index Page I


Imago Radiology S.r.l. IMX-3B Antarix Troco

INDEX
1. INTRODUCTION (Rev.09) ................................................................................................1
1.1. IMPORTANT NOTICE ............................................................................................................................ 2
1.2. WARRANTY............................................................................................................................................ 6

2. TECHNICAL FEATURES (Rev.10) .................................................................................1


2.1. DESCRIPTION........................................................................................................................................ 2
2.2. TECHNICAL DATA ................................................................................................................................. 2
2.2.1. Dimensions....................................................................................................................................... 4
2.3. INDENTIFICATION ................................................................................................................................ 5
2.4. DISPOSAL OF MATERIALS................................................................................................................... 6
2.5. TECHNICAL REGULATIONS and CLASSIFICATIONS ....................................................................... 7
2.6. SYMBOLS .............................................................................................................................................. 7

3. ISTRUCTIONS FOR USE (Rev.09) ..................................................................................1


3.1. GENERAL GLOBAL VIEW .................................................................................................................... 2
3.2. CONTROL PANEL ................................................................................................................................. 4
3.3. FUNCTIONAL DESCRIPTION .............................................................................................................. 5
3.3.1. Equipment Alignment ...................................................................................................................... 5
3.3.2. Tube Stand Rotation ....................................................................................................................... 8

4. CLEANING AND MAINTENANCE (Rev.09) ....................................................................1


4.1. CLEANING ............................................................................................................................................. 2
4.2. MAINTENANCE ..................................................................................................................................... 2

B. OPTIONALS ..................................................................................................................1
A.1. STANDARD ACCESSORIES ................................................................................................................ 2
A.2. OPTIONAL ACCESSORIES ................................................................................................................. 2

Operator’s Manual (M70A_11) index Page II


Imago Radiology S.r.l. IMX-3B Antarix Troco

1. INTRODUCTION

Operator’s-Technical Manual (M70A_9) introduction Page 1 of 6


Imago Radiology S.r.l. IMX-3B Antarix Troco

1.1. IMPORTANT NOTICE

WARNING
GENERAL
This user manual provides instructions for proper use of equipment and should be
considered as its integral and permanent part and must remain with the unit.

Carefully read the content of this manual before you start working on the equipment.

The working environment for this equipment is the "professional X-ray room" properly
prepared with adequate shielding for X-rays and a suitable power supply line
"dedicated" You can NOT use this equipment in buildings that have a domestic type
of mains power supply .

You can not install and use this equipment in operating theatres

The use of this equipment by unqualified and not authorized personnel is strictly
prohibited. Even if this equipment is designed and manufactured in compliance with
the most current safety standards, the X-ray emission is always a danger when the
operator is not qualified or duly informed, exposure to X-rays might cause injuries to
people.

All precautions must be taken to prevent unauthorized and unqualified personnel to


operate on the equipment as this might cause danger to himself and others.

Do not use the equipment in presence of vapors or flammable mixtures.

Avoid decanting and penetration of water and various fluids inside the equipment to
prevent the corrosion or short circuits.

Avoid, in the vicinity of the equipment, the use of other equipments not conform to the
EMC regulation.

Before activating any movement of the equipment, make sure that the patient is lying
in a correct position with legs and arms inside table-top.

During activation of the movements, the operator must pay close attention to the
patient and to other people who may be in the room. The operator must make sure
that there are no objects near the equipment that may interfere with its movements.

Operator’s-Technical Manual (M70A_9) introduction Page 2 of 6


Imago Radiology S.r.l. IMX-3B Antarix Troco

In the case of patient with thorn skin and / or injury a Biocompatible towel must be
placed between the patient and table top.

Avoid exposure to radiation in pregnant women and children under 6 years.

The equipment should not be used for purposes other than those mentioned.

Technical works on the equipment and opening of protection covers or any


intervention on any component of the equipment, should only be performed by
competent personnel, trained directly or indirectly by the manufacturer.

In case of faults or malfunctions, that can not be solved as indicated in this manual,
the authorized person must consult the manufacturer of the equipment for the
necessary and appropriate instructions.

The manufacturer disclaims any liability if the authorized personnel, during the course
of his duties, does not adhere to the requirements contained in this manual.

The manufacturer disclaims any liability arising from repairs performed by


unauthorized personnel.

It is advisable to turn off the equipment during periods of inactivity.

ALWAYS use protective gloves when handling lead counterweights.

RADIOPROTECTION

Before using the equipment, the authorized personnel must be informed about
security measures to be taken, as established by the "International Commission on
Radiological Protection" and its "National Standard".

Protect the patient from scattered radiation in the surrounding areas by means of
anti-X aprons, where necessary.

To protect the patient during radiographic exposure, always use the largest possible
focus –skin distances.

Before exposure remove any object, metallic or not unnecessary to radiographic


examination (eg, necklaces or rings worn by the patient).

The control point of the X-ray emission must be in an area protected and shielded
from radiation with an anti-X window on the equipment and loudspeakers to
communicate with the patient.

During radiographic exposures, only the patient can be present in the room. In case
you need the presence of medical staff or trained and authorized paramedics must
be present (eg, with non-cooperative patients) , they must either be provided with
protective anti X-ray clothing and must stay in Designed Zone of Occupancy as
outlined below.
Operator’s-Technical Manual (M70A_9) introduction Page 3 of 6
Imago Radiology S.r.l. IMX-3B Antarix Troco

DESIGNED ZONE OF OCCUPANCY

Fig. 1) Frontal view

Note:

"Fig 1) shows the frontal area of possible occupation with a dimension of 600x600 mm
(23.6" x23.6 ") x 2000 mm (78.7") height as required by applicable law.

Operator’s-Technical Manual (M70A_9) introduction Page 4 of 6


Imago Radiology S.r.l. IMX-3B Antarix Troco

Since the Column/ table-top/ bucky group can move longitudinally to explore the whole
body, the occupation zone is extended longitudinally for all the length of table top (2200
mm / 86.6”) as showed in the following “Fig-2)”:

Fig. 2) Top view

The maximum values of scattered radiation at the heights indicated in the above areas of
occupation with table in horizontal position are shown in the following "Table 1”).

H dal suolo Dose


400 mm (15,7") 0.04 mGy/h
600 mm (23,6") 0.12 mGy/h
1000 mm (39,4") 0.66 mGy/h
1200 mm (47,2") 0.62 mGy/h
1400 mm (55,1") 0.39 mGy/h
1600 mm (63,0") 0.10 mGy/h
1800 mm (70,9") 0.03 mGy/h
2000 mm (78,7") < 0.01 mGy/h

Table 1) Scattered radiations maximum values

The maximum values of scattered radiation were measured as described in the reference
standard IEC60601-1-3 with radiological factors of 100 KV, 100 mAs and Duty-cycle 1:60.

Operator’s-Technical Manual (M70A_9) introduction Page 5 of 6


Imago Radiology S.r.l. IMX-3B Antarix Troco

1.2. WARRANTY

The warranty validity of the products is 24 months from invoice date for any fault of
construction or employed material, except consumables. Warranty does neither include
transport nor travel expenses of our technical staff.

12 months Pro Rata Temporis gradual warranty is valuable for: generators, monoblocs,
X-ray tubes and Image Intensifiers.

Warranty is valuable as long as the equipment is used properly according to user’s


instructions. Any request of warranty replacement must indicate the purchase date and
the serial number of the unit.

Warranty does not cover:

• Damages caused by faults of the operator or lack of maintenance.


• Technical interventions made by unauthorized personnel.
• Replacements of not original spare parts.
• Inappropriate or negligent use
• Wrong installation
• Model, code or S.N° of the unit have been modified, erased, removed or made
unreadable.
• Damages caused by reasons beyond manufacturer’s responsibility, such as (but not
only) : natural events, storms, catastrophes, riots etc…

Before returning the goods to the supplier, the customer must send a technical report
describing the problem in detail and also working conditions. Any return of goods to our
Company must be on carriage paid basis and agreed in advance.

The request of free spare parts can be accepted only after having received the defective
parts (with transport to customer’s charge ) in order to be able to check the causes.
The manufacturer reserves the right to decide if the parts must be repaired or replaced.

The defective products must be sent to the following address:

Made by:
IMAGO RADIOLOGY S.r.l.
Via G.G.Sforza 95
20081 Abbiategrasso (MI) ITALY
Tel. +39 029464120 / +39 029465996 Fax: +39 029450043
Website : www.imagoradiology.it

Operator’s-Technical Manual (M70A_9) introduction Page 6 of 6


Imago Radiology S.r.l. IMX-3B Antarix Troco

2. TECHNICAL FEATURES

Operator’s-Technical Manual (M70A_10) technical features Page 1 of 9


Imago Radiology S.r.l. IMX-3B Antarix Troco

2.1. DESCRIPTION
ANTARIX TROCO is a radiographic table with table-top at fixed height, potter
bucky with vibrating grid, manual light beam collimator.
It is possible to horizontally move the table top in all directions by means of foot
pedal.
The column, sliding on floor rails, with the facility of x-ray tube arm rotation,
allows you to operate in combination with Vertical Bucky for upstanding
radiographies.

2.2. TECHNICAL DATA

NAME OF THE UNIT


ANTARIX TROCO IMX-3B
Power Supply 230Vac +/- 10% 50 Hz 5A
Weight of the equipment 434 Kg

TABLE
Table top dimensions 2200 x747
Table Top Height from the ground 730 mm
Longitudinal run ± 407 mm
Transversal run ± 120 mm
Table top to film distance 89 mm
Potter bucky longitudinal run 490 mm
Max patient’s weight 200 Kg
Absorption coefficient ~/< 0,8 mm aluminium

COLUMN STAND
Dimensions 3006 x h2154 x 1405 mm.
Tube arm vertical movement Min. 520 mm - Max. 1950 mm
Tube arm transversal movement 160 mm ( OPTIONAL )
Tube arm rotation around horizontal axis ± 110°
Stand maximum rotation on its vertical axis
with mechanical bloc every 90°
± 180°
Focus film distance Min. 1000 mm - Max. 1200 mm

TUBE FEATURE
Max. weight of X-ray tube 24 kG

Operator’s-Technical Manual (M70A_10) technical features Page 2 of 9


Imago Radiology S.r.l. IMX-3B Antarix Troco

POTTER BUCKY 35x43 cm


Technical data See enclosed manual

COLLIMATOR
Technical data See enclosed manual

STORAGE AND TRANSPORT CONDITIONS


Temperature -10° to +40° C
Humidity 20 a 80% without dew

WORKING CONDITIONS
Temperature +10° to +35° C
Humidity 35 to 75% without dew
Maximum Altitude Utilization 3500 meters

Continuous functioning with


Usage Conditions
intermittent charge

Note : Do not use the equipment in flammable or explosive environments.


Equipment not protected against water.

Operator’s-Technical Manual (M70A_10) technical features Page 3 of 9


Imago Radiology S.r.l. IMX-3B Antarix Troco

2.2.1. Dimensions

ATTENZIONE:
NON ruotare lo Stativo
oltre la Quota di ±90°

WARNING :
Do NOT rotate the stand
for more than ±90°

Fig. 2a

Operator’s-Technical Manual (M70A_10) technical features Page 4 of 9


Imago Radiology S.r.l. IMX-3B Antarix Troco

2.3. IDENTIFICATION

An identification data label is located above table’s left side and on the back side of the
column. It shows: equipment’s code, serial number and other needed information.

Vista Frontale Vista Posteriore


Front View Back View

Fig. 2b

You always need to give this information when requesting maintenance


or parts in replacement.

Operator’s-Technical Manual (M70A_10) technical features Page 5 of 9


Imago Radiology S.r.l. IMX-3B Antarix Troco

2.4. DISPOSAL OF MATERIALS

This equipment contains materials that can be recycled or reused.

• Packing :

¾ Wood and plastic sheet 100% recyclable

• Materials composing the unit:

¾ Carbon steel 100% recyclable


¾ Elastomers and plastic materials 100% recyclable
¾ Cables and other electrical materials
¾ Lead 100% recyclable

The equipment must not be disposed as a normal urban solid waste, but it must be taken
to an expert waste disposal centre who will dismantle it in conformity with the local laws in
force.

For all the equipments manufactured after August the 13th 2005, disposal must be made
as per n° 22/99 RAEE European decree.

Operator’s-Technical Manual (M70A_10) technical features Page 6 of 9


Imago Radiology S.r.l. IMX-3B Antarix Troco

2.5. TECHNICAL REGULATIONS AND CLASSIFICATIONS

The unit complies with the following directives :

♦ 93/42/EEC Directive (Class II) and subsequent amendments made 2007/47/CE.

Anf to the following technical standards:

♦ CEI EN 60601-1 (Class 1 type B) and further corrections.


♦ CEI EN 60601-1-2 (Electro Magnetic Compatibility)

Test Report IMQ - Annex A:


The ANTARIX is suitable for use in the specified electromagnetic envoronment.
The purchaser or user of the ANTARIX should assure that is used in an electromagnetic
environment as described below:
Emission test Compliance Electromagnetic Environment
RF emissions Group 1 This ANTARIX uses RF energy only for internal function.
Therefore, the RF emission is very low and not likely to
CISPR 11 cause any interference in nearby electronic equipment
Class B This ANTARIX is suitable for use in domestic
establishments and is establishments directly connected to
the low voltage power supply network which supplies
buildings used for domestic purposes
Harmonic Complies This ANTARIX is suitable for use in establishments
emissions directly connected to a public low voltage power supply
network
EN 61000-3-2 Complies This ANTARIX is suitable for use in establishments
Voltage directly connected to a public low voltage power supply
fluctuations/ network
Fliicker emissions
EN 61000-3-3

Test Report IMQ - Annex B:


The ANTARIX is suitable for use in the specified electromagnetic envoronment.
The purchaser or user of the ANTARIX should assure that is used in an electromagnetic
environment as described below:
EN 60601-1-2 Compliance
Immunity test Electromagnetic Environment
Test level level
Electrostatic 6 KV contac EN 60601-1-2
Discharge 8 KV air Test level Residential/Hospital
(ESD)
EN 61000-4-2
Electrical fast 2 KV for power EN 60601-1-2
Transient/burst supply lines Test level Residential/Hospital
EN 61000-4-4
1 KV for
input/output lines >
3m
Surge 1 KV differential EN 60601-1-2
EN 61000-4-5 mode Test level Residential/Hospital

2 KV common
Operator’s-Technical Manual (M70A_10) technical features Page 7 of 9
Imago Radiology S.r.l. IMX-3B Antarix Troco

mode
Viltage dips, 0% Un for 0.5 cycles EN 60601-1-2
short 40% Un for 5 cycles Test level Residential/Hospital
interruptions 70% Un for 25
and voltage cycles
variations on 0% Un for 5 cycles
power supply
input lines
EN 61000-4-11
Power 3 A/m EN 60601-1-2
frequency Test level Residential/Hospital
(50/60 Hz)
magnetic field
EN 61000-4-8

Test Report IMQ - Annex C: non-LIFE SUPPORTING EQUIPMENT


The ANTARIX is suitable for use in the specified electromagnetic envoronment.
The customer or the user of the ANTARIX should assure that it is used in such an environment
EN 60601-1-2 Compliance
Immunity test Electromagnetic Environment
Test level level
Portable and mobile RF
communications equipment should be
used no closer to any part of the
ANTARIX including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter
Recommended separation distance
Radiated RF 3 V/m 3 V/m d = 1.2x√P 80 MHz to 800 MHz
EN 61000-4-3 80 MHz to 2.5 GHz
d = 2.3x√P 808 MHz to 2.5 GHz
Conducted RF 3V 3V d = 1.2x√P
EN 61000-4-6 150 KHz to 80 MHz
Where P is maximum output rating of the
transmitter in watt (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).

Field strengths for fixed RF transmitter, as


determined by an electromagnetic site
survey, should be less then the
compliance level in each frequency range

Interference may occur in the vicinity of


equipment marked with the following
symbol:

Operator’s-Technical Manual (M70A_10) technical features Page 8 of 9


Imago Radiology S.r.l. IMX-3B Antarix Troco

2.6. SYMBOLS

Simboli Descrizione Description

Dir. 93/42/CEE e s.m.i Dir. 93/42/CEE e s.m.i


Marcatura CE di conformità EC mark

IEC 60417-5840 IEC 60417-5840


Parte applicata tipo B Applied part type B

ISO 7010-M002 ISO 7010-M002


Fare riferimento al libretto di Refer to the instruction manual
istruzioni

IEC 60417-5019 IEC 60417-5019


Terra di protezione Grounding protection

IEC 60417-5036 IEC 60417-5036


Pericolo alta tensione Danger High Tension

IEC 60417-5032 IEC 60417-5032


Corrente alternata A/C

UNI EN 980:2004 UNI EN 980:2004


Data di produzione ( aaaa-mm ) Manufactured ( yyyy-mm )

UNI EN 980:2004 UNI EN 980:2004


Prodotto da xxx Made in xxx

UNI EN 980:2004 UNI EN 980:2004


Serial Number Serial Number

UNI EN 980:2004 UNI EN 980:2004


Codice commerciale apparecchio Equipment commercial code

22/99 RAEE 22/99 RAEE


Smaltimento Disposal

Operator’s-Technical Manual (M70A_10) technical features Page 9 of 9


Imago Radiology S.r.l. IMX-3B Antarix Troco

3. ISTRUCTIONS FOR USE

Operator’s-Technical Manual (M70A_9) instructions for use Page 1 o f 8


Imago Radiology S.r.l. IMX-3B Antarix Troco

3.1. GENERAL GLOBAL VIEW

IMPORTANT

NOTA : Pay attention when the equipment is moving in general and especially if
there is a patient near it.
Avoid to hurt the patient or the equipment with any accessory or any other
rigid material.

O I

N K

B
M

G
F E

Fig. 3a

Operator’s-Technical Manual (M70A_9) instructions for use Page 2 o f 8


Imago Radiology S.r.l. IMX-3B Antarix Troco

IDENTIFICAZIONE DELLE PARTI / PARTS IDENTIFICATION (fig.3a) :

A. Elevating table.
A. Tavolo elevabile.
B. Potter bucky .
B. Potter.
C. Table top .
C. Piano Porta-Pazienti (P.P.P.).
D. Handle for table top .
D. Maniglia Piano Porta-Pazienti.
E. Switch for table top transversal movement release.
E. Interruttore di sblocco movimento trasversale P.P.P.
F. Pilot light for power supply.
F. Spia luminosa di presenza alimentazione.
G. Foot pedal for table top movements release.
G. Leva a Pedale di sblocco movimenti P.P.P.
H. Switch for potter bucky longitudinal movement
H. Interruttore di sblocco movimento longitudinale release.
Potter.
I. Stand.
I. Stativo.
J. X-ray tube holder.
J. Supporto Guaina.
K. Focal height indicator .
K. Indicatore Altezza Focale.
L. Collimator.
L. Collimatore.
M. Stand rails group.
M. Gruppo Rotaie Stativo.
N. Control Panel .
N. Pannello Comandi.
O. Rx tube .
O. Tubo RX .

Operator’s-Technical Manual (M70A_9) instructions for use Page 3 o f 8


Imago Radiology S.r.l. IMX-3B Antarix Troco

3.2. CONTROL PANEL

Fig. 3b

DESCRIZIONE PANNELLO / CONTROL PANEL DESCRIPTION (fig.3b) :

1. Comando sblocco del freno movimento verticale del 1. Control key for brake release of x-ray tube arm
Supporto Guaina (Premendo il pulsante il movimento vertical movement (the movement is released when
si libera). the key is pushed).
2. Comando sblocco/blocco freno rotazione Supporto 2. Control key for brake release and blocking of X-ray
Guaina (Premendo il pulsante il freno si libera). tube arm rotation. (The movement is released when
3. Comando sblocco freno movimento trasversale the key is pushed).
Supporto Guaina (Premendo il pulsante il movimento 3. Control key for brake release of X-ray tube arm
si libera). transversal movement (The movement is released
4. Comando sblocco del freno movimento longitudinale when the key is pushed)
Supporto Guaina (Funzionamento in modo 4. Control key for brake release of x-ray tube arm
“interruttore”: premere una prima volta per liberare il longitudinal movement. (It’s a release button: if you
movimento, premere una seconda volta per bloccare il push it, the movement is released , if you push it
movimento). again, the movement is blocked).
5. Indicatore angolo (gradi) rotazione Supporto Guaina. 5. Rotation arm angle indicator gauge (degrees).

Operator’s-Technical Manual (M70A_9) instructions for use Page 4 o f 8


Imago Radiology S.r.l. IMX-3B Antarix Troco

3.3. FUNCTIONAL DESCRIPTION

IMPORTANT
When you press control key for brake release of vertical movement of tube arm ( Control
key 1 – Fig. 3b), avoid to bump the table top; this might damage the system.

3.3.1. EQUIPMENT ALIGNMENT

Transversal centring of bucky table top (Fig.3c):

• Push the lock/unlock switch of table top transversal movement (E) and move
manually the table until the automatic centring is activated.
• Push the foot pedal (G) to release the longitudinal brake of the table top (C).
Remove your feet from the pedal to activate again the brake.
• Push again the switch (E) if you want to unlock the function of transversal centring.
In that way, when the foot pedal is pressed, you can unlock both longitudinal and
transversal movements.

Posizione di Sblocco
Release Position
Posizione di Blocco
Block Position Fig. 3c

Operator’s-Technical Manual (M70A_9) instructions for use Page 5 o f 8


Imago Radiology S.r.l. IMX-3B Antarix Troco

Tube arm and table centring (Fig.3d):

• Keep pushed KEY 3 (Fig.3b)


and move tube arm (O)
transversally towards the
centre of the table (A); the O
release of the internal I
mechanical positioner will
confirm that you have K
reached the table central
position.
• Keep pushed KEY 1 (Fig.3b)
and vertically move tube arm
(O) to reach a focal distance
(K) by taking as a reference
the cm indicator on the
column (I).
A

Fig. 3d

Loading Patient

• Loading patient on table-top.


• Make sure that the patient is lying in a correct position with legs and arms inside
table-top.

Patient – Potter Bucky – Tube arm Centring (Fig.3e):

• Turn on collimator centring light beam (P).


• Keep KEY 4 (Fig.3b) pressed and move tube arm longitudinally in comparison to
table (A).
• In case, move table top (C) longitudinally by means of foot pedal (H) in order to
align the anatomic part of the patient to be examined with collimator’s light beam.
• Remove your foot from foot pedal to block table top in the desired position.
• Remove cassette tray (3h-1) from potter bucky (B).
• Insert cassette between its centring clamps.
• Align cassette centre with the notch clamp (3h-2).
• Rotate handle (3h-3) to block cassette.
• Push switch (I) to release potter bucky longitudinal movement.
• Align potter bucky with collimator light beam.
• Re-push switch (I) to block potter bucky.
• Reinsert cassette tray in potter bucky.

Operator’s-Technical Manual (M70A_9) instructions for use Page 6 o f 8


Imago Radiology S.r.l. IMX-3B Antarix Troco

3h-1

3h-3

3h-2 I
Fig. 3e

NOW YOU CAN TAKE A RADIOGRAPHY PAYING ATTENTION TO STAY IN THE


POSITION OF X-RAY CONTROL.

WARNING
Avoid to expose to radiations children under 6.
Avoid to expose to radiations pregnant women.
Make sure that security rules are respected.
It is important that also operators take the necessary precautions to avoid
unnecessary exposures to X-rays.
Even the patient must not be exposed to excessive radiations.

Operator’s-Technical Manual (M70A_9) instructions for use Page 7 o f 8


Imago Radiology S.r.l. IMX-3B Antarix Troco

3.3.2 TUBE STAND ROTATION (Fig.3f):

• The column can rotate around its vertical axis to be placed in parking position, or in
order to enable you to make X-rays on lateral cassette holders and vertical buckies.

Push the foot pedal (1) located at the base of the stand to release its vertical
rotation.
Slightly rotate in the desired direction.
Release small foot pedal.
Go on with the rotation, the stand will automatically block every 90°.

Fig. 3f

WARNING:
In order to avoid to rotate the stand for more than 90°, in both directions :
DO NOT KEEP THE FOOT PEDAL PUSHED DURING STAND ROTATION.

Operator’s-Technical Manual (M70A_9) instructions for use Page 8 o f 8


Imago Radiology S.r.l. IMX-3B Antarix Troco

4. CLEANING AND MAINTENANCE

Operator’s Manual (M70A_9) cleaning and maintenance Page 1 o f 2


Imago Radiology S.r.l. IMX-3B Antarix Troco

4.1. CLEANING

• Cleaning must be made at regular intervals with the equipment turned OFF:

After every examination, clean all parts that have been in contact with the patient
with a disinfectant product approved by the Ministry of Health .
Every week clean the painted metallic and plastic parts, minding to use a proper
cleaning solution especially for the plastic parts.
Use cleaning products properly and avoid to let them penetrate inside the unit.

4.2. MAINTENANCE

• Maintenance must be make after the first 6 months since installation, after wards
regularly every 12 months.

• The ordinary and preventive maintenance must be carried out only by the
manufacturer's staff or by authorized personnel.

• The company refuses all liabilities for damages caused by interventions made by other
persons than those described above.

In case you find damages or malfunctioning that are not stated on this manual,
please call the TECHNICAL ASSISTENCE SERVICE of the manufacturer.
(See paragraph 1.2 )

Operator’s Manual (M70A_9) cleaning and maintenance Page 2 o f 2


Imago Radiology S.r.l. IMX-3B - Antarix Troco
IMX-3B - Antarix Troco

A. ACCESSORI /
ACCESSORIES

Manuale Operatore (M70A_9) accessori Page 1 of 2


Operator’s Manual (M70A_9) accessories
Imago Radiology S.r.l. IMX-3B - Antarix Troco
IMX-3B - Antarix Troco

A.1. ACCESSORI STANDARD / STANDARD ACCESSORIES

Potter IMAGO 35x43 cm. (24V-230V)


IMAGO Potter 35x43 cm. (24V-230V)
IMA-1 Completo di Griglia Foc. 100 cm. (IMA-
with of foc. Grid 100 cm. (IMA-252)
252)

A.2. ACCESSORI A RICHIESTA / OPTIONAL ACCESSORIES

Tendi-fascia con fascia di


IMA-13 Compression band with band
compressione
IMA-20 Coppia di reggi-cosce per ginecologia Gynaecological leg support
IMA-112 Porta cassette laterale snodabile Lateral cassette holder
Griglia come da specifiche del cliente Special grids as per customer’s specs
Kit per Camera di ionizzazione per Kit for measuring chamber for
l’esposizione automatica (come da automatic exposure. ( As per
specifiche del cliente) customer’s request)

Manuale Operatore (M70A_9) accessori Page 2 of 2


Operator’s Manual (M70A_9) accessories