G3300-90016 TOF Quickstart
G3300-90016 TOF Quickstart
Use this guide for your first steps with the Agilent 6210 LC/MS TOF Sys-
tem, and as a roadmap for your user information.
Agilent Technologies
You use the Agilent MassHunter Software for setting up and running data
acquisition. For data analysis, you have two choices. You can use the PE Sciex
Analyst QS 1.1 software package which Agilent provides which is especially
modified for the Agilent 6210 LC/MS TOF system.
You can also use the Agilent MassHunter Workstation Qualitative Analysis
program and the Agilent MassHunter Workstation Quantitative Analysis program.
Both of these programs are included. Before using either of these programs, you
need to translate your data files from the WIFF format to the format used by these
Agilent programs using the Translator program. You can run this program
interactively or directly from a worklist. The translator program is included in the
Agilent MassHunter Workstation Qualitative Analysis program’s installation.
Online Help
Press F1 To get more information about a pane or dialog box, place the cursor on
the part of the pane or dialog box of interest and press F1.
Help menu From the Help menu, access “How-to” help and reference help.
PE-Sciex Analyst online help Refer to Analyst online help to learn how to view,
quantitate and report on Agilent 6210 LC/MS TOF results.
Documents
You can find these manuals delivered with the TOF hardware or software. You can
also find a PDF version on the installation CD-ROM, in the Manuals folder.
TOF User’s Guide Use this guide to install and set up the TOF hardware. This
guide also contains background information to help you operate, maintain and
troubleshoot the TOF.
6210 LC/MS TOF System Installation Guide This guide is used by the Agilent
customer engineer to install the 6210 LC/MS TOF hardware and MassHunter
Software, configure the instrument, and verify performance.
You can find these manuals on the installation CD-ROM, in the Manuals folder.
Concepts Guide - The Big Picture Learn the background information to help you
make selections in the software.
Training
Familiarization Guide Use this guide as a training lab.
3 Set up an
Set up and acquisition
run samples method
1 Double-click the
Agilent TOF group
on your desktop.
2 Double-click the
TOF icon to start
the software
engines.
5 Double-click the
TOF icon in the
Agilent TOF
group (Figure 1).
The main window appears. See Figure 2 on page 7. 6 Make sure that the LC modules are the
The top pane of this window is the Instrument Status ones that you want configured with
pane. (Figure below) the instrument. (See below.)
• Check the
headers of the LC
modules labeled
in the Instrument
Status pane.
If LC modules other than those you intend to use appear in the
Instrument Status pane or the Configuration report, use the Online
Help to access instructions to reconfigure the instrument.
Click a button to
Instrument Status pane Real-time Plot pane Method pane Worklist pane
see the pane you
want to use.
Worklist pane
With this pane you enter sample information for individual samples and
information for batch samples. When you run the worklist, the samples and
batches are automatically run in the order listed in the worklist.
3 Set up DAD
Prepare the real-time
instrument plot displays
(optional)
5 View and
4 Prepare the
work with
TOF
the system
logbook
Purge the binary pump You purge the binary pump manually.
1 Right-click the
binary pump
status box in the
Instrument
Status pane.
2 Select Standby
in the shortcut
menu.
5 Enter a Flow of
5 ml/min, and
click Apply.
6 Select On in the
binary pump
shortcut menu.
7 After purging,
enter the flow 8 Close the black Purge for about 9 minutes to pass about
you use to valve. 45 mL or 3X the volume for the binary
equilibrate the pump.
column, and
click Apply.
1 Right-click
pump device in
the Instrument
Status pane to
bring up pump
device menu. 2 Select Purge to
bring up the Pump
Purge System
dialog box.
4 Monitor the
purge with this
dialog box and
click Cancel
after the purge
is complete.
2 Click Apply to
send the
parameters to the
LC.
Right-click the LC
module in the
Instrument Status
pane to bring up
the control menu
for that module.
1 Right-click the
Actuals box to
bring up Setup
item.
2 Click Setup to
bring up the tool
to select
instrument
actuals.
3 Select actuals
to set up to view
the actual
conditions in
the Instrument
Status pane.
1 Right-click the
signal plot to
bring up the
signal shortcut
menu.
2 Select Change to
bring up the tool
for selecting the The default display is of a TIC chromatogram.
signal and its plot
parameters.
3 Select a DAD
signal that you
intend to monitor,
and click Add.
4 Highlight a
Selected Signal.
1 Right-click the
line or profile
spectral plot to
bring up the
spectra shortcut
menu.
2 If you clicked the
line spectral plot,
select the DAD
spectrum as the
spectrum to Or, if you right-clicked the
view. profile spectral plot, select
Line first, then select the
DAD spectrum.
Polarity Switching
If you are using Polarity Switching, you need to use a different autotune method.
For each source, there is a positive and a negative method for Polarity Switching.
The name of the autotune method has either “PolaritySWPos” or “PolaritySWNeg”
appended to it.
• ESIautotunePolaritySWPos.m for ESI
• ESIautotunePolaritySWNeg.m for ESI
• nanoESIautotunePolaritySWPos.m for nanospray or dual nanospray
• nanoESIautotunePolaritySWNeg.m for nanospray or dual nanospray
• APPIautotunePolaritySWPos.m for APPI
• APPIautotunePolaritySWNeg.m for APPI
• APCIautotunePolaritySWPos.m for APCI
• APCIautotunePolaritySWNeg.m for APCI
• MMIautotunePolaritySWPos.m for MMI
• MMIautotunePolaritySWNeg.m for MMI
You will need to perform four autotunes to correctly tune the TOF system when
using Polarity Switching. First, you need to tune in both positive and negative
modes. Then, you need to tune using the Polarity Switching methods in both
positive and negative modes.
5 If you want to
select a different
set of masses or
you want to use a
different
calibration
standard, click
Show Extended.
6 Select masses to
use for
calibration, OR
load another
mass list.
7 Then calibrate
again.
2 Check to see if
the TOF needs
tuning.
4 Or click Show
Extended to do a
3 Click Autotune manual tune on the
for a tune that parameters that need
takes 15-20 adjusting.
minutes or
Quicktune for a
tune that takes
1-3 minutes.
1 Select the
masses for the
ramp.
2 Select the
parameter to
vary.
3 Select the
variable to
optimize, and
click Ramp.
If you do a manual
tune, you must also
do a calibration.
Autotune and
Quicktune include
mass calibration
with one mass.
After you do a tune,
the optimized
parameters appear
in the Parameters
panel.
Select System
OR, click the Log Select Columns
Logbook Viewer from
icon.
the Tools menu. Find
Filter Events
3 Set up to
Set up change TOF
acquisition parameters
methods during run
1 Select ion
polarity.
2 Enter TOF
parameters in
the Data, and
Acquisition tabs.
You enter these values for the initial Right-click each field to All entries in the Tune,
time 0.0 min and the whole run and for find the maximum and Calibration and Parameters tabs
one scan, unless you add other time minimum values. are not saved with the method.
segments and scans. See the next page.
You can also change Note that Ion Polarity can Note the values you Note the values you
3 Add up to four scans values on the Data be changed for each time can change with can change with
for each time tab for each time segment. The label for that each time segment. each scan.
segment, including segment. section will be “Ion
the initial time. Polarity (Seg.)” if there are
4 For each scan, more than one time
change a value segments.
or values.
1 Enable
reference mass
correction.
2 Mark Bottle A
to use the
Agilent
reference std.
1 Select the
default mass
list for your ion
source and
polarity.
2 Enter a new
name into the
Name field.
4 Add or delete
masses to the
new list.
3 Print a
Acquire real-time
data plot report
5 Print a 4 Modify a
worklist worklist in
report progress
4 Select a project
and enter the data
file name.
You can only create projects in Even though Both Acquisition and DA is a selection
Analyst. See “Step 5—Analyze data” in the “Part of method to run” list, it is not available
on page 35. for single samples in this version of software.
1 Right-click here to
bring up the worklist
shortcut menu.
3 Select Add
column(s) to add a
column for the You must select
empirical formula. default.anm to
produce an empirical
formula confirmation
4 Enter the generic report.
name (Formula_x)
and formula for an
EFC column.
If Track Worklist is
On (Worklist menu),
the row that is
running is
highlighted blue.
• To print a
real-time plot
report during the
run, select Print
> Real-Time Plot.
3 Review
Analyze spectra and
data sample
information
4 Set up and
5 Customize the generate a
data analysis
parameter
method,
optimization
default.anm
report
1 Double-click the
Analyst QS icon
in the Agilent TOF
Software group
window.
1 Right-click on
chromatogram to See the next page to learn about
bring up the spectral operations with Analyst.
shortcut menu.
1 Right-click on
spectrum to bring
up the shortcut
menu.
• Select View
Actuals from the
shortcut menu to
see the real-time
TOF parameter
values.
• Select
Re-Calibrate TOF
to recalibrate the
TOF from Analyst.
2 Select the
samples for
the report.
3 Continue
through the
wizard, then
click Finish.
The EFC report can now also include a backward database search
(called an EFC Database Screening Report). Based upon a formula, a
mass is determined and then XICs are extracted for that mass to see
if the compound can be found. You can limit the search of the
database to formulas with a certain retention time tolerance. A value
of -1 in the Retention time tolerance field indicates to not limit the
search based upon retention time.
In this book
This book contains brief
instructions to help you get
started with your Agilent 6210
Time-of-Flight TOF LC/MS
system. This books shows you
how to:
• Prepare the instrument for a
run
• Set up acquisition methods
• Set up and run an
interactive sample and
worklists
• Review data
To submit comments about this
guide, send an e-mail to
[email protected].
*G3300-90016*
G3300-90016
Agilent Technologies