FAQ Error 70
FAQ Error 70
Department: I IA SC PS3
Issued: 22.09.2010
QUESTION
Why does Error Code 70 randomly appear on the display of my SITRANS F M Transmitter?
ANSWER
All sensors are calibrated at a certain coil excitation frequency depending on type and
size. This frequency is stored in the SENSORPROM memory unit.
Error code 70 can randomly appear on the display, if the factory set coil excitation
frequency has for unknown reason changed to a higher value. The changed frequency
may also cause the measured flow values to be lower than the actual flow.
Further Information
www.siemens.com/flow