Defib Zoll M2
Defib Zoll M2
BEMS/UT/
TYPE CODE : 17882-047,
Defibrillator/Pacemaker, External
(ZOLL, M2)
Prepared by :
IZHAR IZZUDDIN BIN ABDUL AIR
BEMS ENGINEER
1. EQUIPMENT OVERVIEW
CONTENT
2. BASIC FUNCTION (not to include clinical application)
8. POTENTIAL HAZARD
EXTERIOR VIEW
EXTERIOR VIEW
13. CHARGER ON
1. SELECTOR SWITCH 14. Paper Tray
2. DEFIB ENERGY SELECT 15. SUMMARY
BUTTONS 16. CODE MARKER
3. DEFIB CHARGE 17. PCMCIA Data Card Slot
4. SHOCK 18. PC Card Modem Slot
5. ANALYZE (12-Lead Option Only)
6. SOFTKEYS 19. PACER OUTPUT mA
7. LEAD (Pacer Version Only)
8. SIZE 20. . 4:1 BUTTON (Pacer
9. ALARM SUSPEND Version Only)
10. RECORDER 21. PACER RATE ppm
11. BEEPER VOLUME (ECG) (Pacer Version Only)
12. BRIGHTNESS/CONTRAS 22. Systole and Alarm
T ADJUSTMENT Speaker
23. Microphone (Optional)
1. EQUIPMENT EXTERIOR VIEW
OVERVIEW
1. PEDIATRIC ELECTRODE
PLATE
2. ADULT ELECTRODE
PLATE
3. BUTTON
MANUAL DEFIBRILLATION
2. BASIC OPERATIONAL 1. Turn the SELECTOR SWITCH to DEFIB. The unit automatically defaults to 200 Joules
2. Energy Select - Observe the display and verify the selected energy is appropriate. To change
PROCEDURE
the energy setting use either pair of up/down arrow buttons. One pair is located on the front
panel of the unit, the other pair is located on the sternum paddle
3. Prepare Paddles - Remove the paddles from their holders by grasping the handles and
pressing down on the paddle release latch located above each paddle
4. Apply Paddles to Chest - Apply the paddles firmly to the anterior wall of the chest.
MANUAL DEFIBRILLATION (Cont)
2. BASIC OPERATIONAL 5. If an ECG cable and ECG electrodes are in use, press the LEAD button to select the desired ECG
PROCEDURE lead configuration
6. Press the CHARGE button on the front panel or on the apex paddle handle
7. when the CHARGE button is activated, the device will not charge and a “RELEASE SHOCK BUTTON”
or other message will appear on the display.
8. Using your thumbs, simultaneously press and hold both SHOCK buttons (one on each paddle) until
energy is delivered to the patient.
9. Once energy is delivered, the display will simultaneouslyshow “XXXJ DELIVERED” and “DEFIB XXXJ
SEL.” After approximately 5 seconds the “XXXJ DELIVERED” message will disappear and the
“DEFIB XXXJ SEL” message remains to indicate the selected energy level
10. Paddle plates and handles must be thoroughly cleanedafter each use
AUTOMATED EXTERNAL DEFIBRILLATOR (AED)
2. BASIC OPERATIONAL OPERATION
PROCEDURE 1. The unit will beep 4 times to indicate that it has passedthe power-on self-test. If
the audio recorder is present,the unit will begin recording audio data immediately
2. Energy Select- For non-Biphasic units shock number 1 is set at 200joules, shock 2
is set at 200 joules and shock 3 and up is set at 360 joules (default setting). For
Biphasic units shock number 1 is set at 120 joules, shock 2 is setat 120 joules and
shock 3 and up is set at 200 joules
3. Press the ANALYZE button to begin analysis of the patient’s ECG rhythm. The
device announces and displays a “STAND CLEAR” message
4. An “ANALYZING ECG” message is then displayed for upto 12 seconds while the
patient’s ECG is analyzed
5. Once the analysis is completed, the unit indicates whether or not a shock is
advised, when a non-shockable rhythm is detected the unit displays a “NO SHOCK
ADV.” message.
AUTOMATED EXTERNAL DEFIBRILLATOR
(AED) OPERATION (Cont)
2. BASIC OPERATIONAL
6. Immediately check pulse and breathing and resume other treatment per protocol. If the patient’s
PROCEDURE rhythm is shockable the unit will display a “SHOCK ADVISED” message.
7. The defibrillator will begin charging automatically to the pre-configured energy setting and
display a “CHARGING” message
8. When charging is completed the monitor displays the energy level to which the defibrillator has
been charged,“XXXJ READY
9. Once the unit has charged to the selected energy, the SHOCK button will illuminate and the
“PRESS SHOCK”message will be announced and displayed.
10. A continuous tone will sound for 10 seconds, followed byan intermittent beeping for 5 seconds.
The shock must be delivered within this 15 second interval or the defibrillator will disarm itself.
11. Press and hold the illuminated SHOCK button on the front panel until energy is delivered to the
patient.
12. After the energy is delivered to the patient, the displayreturns to XXX J SEL. SHOCKS: 1,
indicating the number of shocks administered to the patient.
13. Press the ANALYZE button to restart an ECG analysisand determine if additional shocks are
required.
ECG MONITORING
1. Depending upon local usage, the ECG leads are markedeither LA,RA, LL, RL, and V or L, R, F,
2. BASIC OPERATIONAL N and C
IEC Color AHA Color Placement of Electrodes
PROCEDURE Coding Coding
R/Red RA/White Place near right mid- clavicular
Electrode Electrode line, directlybelow clavicle.
L/Yellow LA/Black Place near left mid- clavicular
Electrode Electrode line, directlybelow clavicle. RA LA
clavicular line.
C/White V/Brown Single movable chestelectrode.
Electrode Electrode
2. Peel the protective backing from the ECG electrode. Becareful to keep adhesive surface free of
electrolyte gel.
3. Attach snap-on leads and check for good contact between the electrode and the lead
termination.
4. Plug the patient cable connector into the ECG input connector
ECG MONITORING (Cont)
5. Set SELECTOR SWITCH to the MONITOR position
2. BASIC OPERATIONAL 6. Press the LEAD button until the desired lead configuration is selected
7. If the “ECG LEAD OFF” message appears on the display, inspect the ECG electrodes, lead wires,
PROCEDURE and associated connections. If a “CHECK PADS” or “POOR PAD CONTACT” message is displayed
inspect the MFE Pads, cable, and associated connections
8. If heart rate alarms are enabled with paddles selected, the unit displays the message “SELECT
LIMB LEADS”.If you see this message, select limb or precordial leads.
9. Press the SIZE button until the desired waveform size isdisplayed.
10. Adjust QRS beeper volume to suitable level using the beeper volume button
To suspend the alarm tone for 90 seconds, press and release the ALARM SUSPEND button in less
2. BASIC OPERATIONAL than 1 second. The alarm tone stops
PROCEDURE
After 90 seconds, if the physiological parameter remains at a value that triggers the alarm, the unit
sounds the alarm tone again
If the alarm condition “clears” (the physiologicalparameter returns to a value within range) after you
suspend the alarm tone, the unit resets the alarm and displays the bell icon (no flashing, no “X”). The
alarm parameter displays normally (no highlighting).
Adding Patient Name
1. Press ID # softkey.
2. BASIC OPERATIONAL 2. Press Prev. Digit or Next Digit softkey to select letter for patient name.
PROCEDURE 3. Press Inc Digit or Dec. Digit softkey to change valueof letter.
Repeat steps 2 and 3 until you have entered the patient’s entire name.
4. Press Enter Name softkey.
5. Press Prev. Digit or Next Digit softkey to select digit or letter for identification
number.
6. Press Inc Digit or Dec. Digit softkey to change valueof digit.
Repeat steps 5 and 6 until you have entered the patient’s entire identification
number.
Press Enter ID and Return softkey.
Printing a Report
2. BASIC OPERATIONAL 1. Press the SUMMARY button.
PROCEDURE 2. Press the Print Chart softkey.
3. Press the Print Range softkey.
4. Press Prev. Event or Next Event softkey to scroll through
the events.
5. Press Print softkey.
Cleaning the Recorder Printhead
1. Press down and pull the Paper Compartment drawerwhere the RELEASE label is located.
3. DAILY AND ROUTINE 2. Remove the paper (if necessary).
MAINTENANCE 3. Pull drawer out all the way.
4. Tip unit backwards so that the bottom of the drawer iseasily visible.
5. Locate plastic tab in the back of the drawer.
6. Press tab (disengaging plastic ridge) and pull draweruntil removed.
7. Locate the row of soft, thin bristles.
8. Locate a thin black line (Printhead) adjacent andparallel to the bristles.
9. Gently wipe the thin black line with an alcohol (Isopropyl) moistened Q-tip. Dry any residual
alcoholwith a new Q-tip.
10. Place drawer and paper back into the unit.
Changing Paper
1. Press down and pull the paper tray drawer where the “RELEASE” label is located. The drawer
slides open.
2. Check for adequate paper supply. If paper supply is low,remove paper in the tray. Place a
new pad of thermal paper in the drawer with the paper coming off the top of the pad and the
grid facing up.
3. Pull enough paper off the pad so that the paper extendsout of the strip recorder when the
paper compartment drawer is closed.
4. Close paper compartment drawer by pushing the drawerin and pressing down lightly where
the “RELEASE” labelis located until the drawer is flush with the front of the device.
MANUAL DEFIBRILLATOR TESTING
3. DAILY AND ROUTINE If a “LOW BATTERY” message appears during testing at the beginning of a shift, the battery currently
in use is close to depletion and should be replaced and charged. The device does not test the battery
MAINTENANCE
for adequate charge to support extended use of the unit, capacity can only bedetermined by testing
the battery in an appropriate ZOLLBattery Charger.
CORRECTIVE 10. Unit displays the CABLE FAULT or PADDLEFAULT • Check the connection between the Multifunction cable
message. andthe MFE pads.
ACTION • Remove the CPRD-to-MFC connector, if in use, and plug
themultifunction cable directly into the CPR-D-padz.
26.No apparent energy delivery to patient. • Under certain circumstances, some patients will not “twitch”
when energy is delivered.
• Perform defibrillator self test as described in GENERAL
MAINTENANCE Section.
• Check for “CHECK PADS” and ”POOR PAD CONTACT”
messages alternating on the monitor.
• If Multi-Function Electrodes are used, ensure proper
placement and contact.
5. CAUSES AND Symptom • Recommended Action
27.“CHECK PADS” message. • Verify proper Multi-Function Cable / MFE Pad connection by
CORRECTIVE disconnecting and reconnecting the Mult-Function cable and
MFE Pads.
ACTION • Ensure proper contact of Multi-Function Pads and that the
patient does not have excessive hair beneath the electrodes.
• If message persists, disconnect Multi-Function cable from
MFE Pads and plug cable into test connector. “CHECK
PADS” should change to “DEFIB PAD SHORT” (ManualMode
Only).
• If test fails, try using paddles to defibrillate.
28.“USE PADS” message. • ECGAnalysis will only operate when MFE pads are attachedto
the patient.
• Disconnect paddle and connect MFE Pads for use in Semi-
automatic defibrillation.
• Activate manual mode to use Paddles.
DEFIBRILLATOR (Cont)
Symptom Recommended action
5. CAUSES AND 29.“NOISY ECG” • Check for proper application and adhesion of Multi-Function
“RETRYANALYSIS” message. Electrodes.
CORRECTIVE • Check to make sure that nobody is touching the patient
andthat the patient is motionless.
ACTION 30.“ECGTOO LARGE” • Press ANALYZE button again to begin analysis.
“RETRYANALYSIS” message.
31. No “TEST OK” message when performing a • Check to make sure unit is set to 30 joules.
defibrillator self-test. • If testing with Multi-Function Cable, make sure that cable is
firmly inserted into test connector.
• If testing with paddles, make sure to press the paddles
firmlyagainst the sides of the unit while discharging.
32.“DEFIB MAINT. REQUIRED” message. • Contact ZOLLTechnical Service Department.
AC Charger
34. “LOW BATTERY” message appears on monitorwhen • Replace battery pack with a fully charged battery pack.
unit is plugged intoAC mains. • Unplug device from AC mains and plug device back into AC
mains.
• Verify AC mains is working properly.
35. None of the CHARGER ON indicators are illuminated • Unplug device from AC mains and plug device back into AC
when the device is plugged into ACmains. mains.
• Verify AC mains is working properly.
Perkara ini adalah merujuk kepada:
6. CLEANLINESS
AND “MASTER AGREED PROCEDURE : BEM-TS-004 (INCIDENTS, HAZARDS
DECONTIMINATIO AND HANDLING HAZARDOUS AND CONTAMINATED BIOMEDICAL
EQUIPMENT ) PARA 6.3 (Handling Contaminated Biomedical
N AWARENESS equipment )”
6. CLEANLINESS Cleaning and care
1.Wipe the defibrillator monitor screen, Paddle pad and ECG Probe.
AND
DECONTIMINATIO 2.Wipe ECG cable, Paddle cable and Power cable. Avoid excessive use of
liquid.
N AWARENESS
3.Change or sterilize all defibrillator pad and invasive patient accessories.
Points to note:
− Do not use hypochlorite, acetone-, phenol- or ammonia based
cleaners.
Perform Corrective
7. MAINTENANCE User identify faults during use Maintenance
REPORT
Decontaminate faulty Perform performance
PROCEDURE equipment test/electrical safety test
2. BURNS - Do not use the gel in a rate that is too little because it can
cause burns to the patient skin and can affect for long term patient
lives.