Flame Test Experiment
Flame Test Experiment
SAFETY FIRST!
Review the safety procedures In this lab, the solutions you will be using contain harmful materials.
and caution the students Avoid skin contact with these chemicals. Observe all precautions,
about the corrosive nature of especially the ones listed below. If you see a safety icon beside a step in
acids before they begin to the Procedure, refer to the list below for its meaning.
work.
Caution: Wear your safety goggles. (All steps.)
PROCEDURE
As you perform the experiment, record your observations in Data Table 1.
1. Place a test-tube rack on a paper towel. Write the chemical
name for each of the seven metal salts next to a position in the
rack where a test tube will be placed. Use scoopulas supplied
with each salt to place pea-sized samples of each metal salt
into a test tube. Place the tubes in the test-tube rack.
2. Pour about 15 mL of 6M HCl into a clean, labeled 50-mL
beaker. Dip the wire loop into the 6M HCl and then heat it in
the hot flame of a gas burner, as shown in Figure 6.1a. Continue
this procedure until no color comes from the wire when it is
put into the flame.
3. Dip the clean wire loop into a sample of metal salt and heat the
sample in the burner flame, as shown in Figure 6.1b. Record the
color of the flame in Data Table 1. Test the remaining samples,
cleaning the wire loop as described in Step 2, before each new
The solutions can be stored
sample is tested. Record your observations.
and reused or use the
© Pearson Education, Inc., publishing as Pearson Prentice Hall. All rights reserved.
following disposal methods 4. View the flame colors produced by NaNO3 and KNO3 through
for chemical waste. cobalt-blue glass. Record your observations.
Disposal 1: NaNO3, KNO3, 5. Perform a flame test on your unknown salt. Record your
Ca(NO3)2, Sr(NO3)2, LiNO3,
observations.
Cu(NO3)2.
Disposal 2: HCl(aq). 6. Dispose of the unused portions of your samples as directed by
your teacher.
Disposal 3: Ba(NO3)2.
OBSERVATIONS
DATA TABLE 1: FLAME TESTS
unknown
4. Would flame tests be useful for detecting metal ions present in a mixture of metal
ions? Explain.
The detection of metal ions in mixtures would be difficult. For example, Li and Sr2
5. The energy of colored light increases in the order red, yellow, green, blue, violet.
List the metallic elements used in the flame tests in increasing order of the energy
of the light emitted.
6. What is the purpose of using the cobalt glass in the identification of sodium and
potassium?
K nearly always gives a yellow flame because of the presence of trace amounts of Na.
Cobalt glass filters out the yellow, allowing the violet of K to be seen. The glass helps
to distinguish Na from K. If only Na is present, no flame is seen through the cobalt
GOING FURTHER
Do Research
In this lab, you observed that each element emits a unique color of light when heated
in a flame. If these light emissions were examined through a prism, you would observe
that the emitted light is actually composed of different wavelengths of light that may
lie in the violet region, the green region, or the red region of the visible spectrum. Each
element has a unique emission spectrum. Look up the emission spectra for the
elements tested in this lab. Do research on how scientists apply these emission
spectra to investigate the chemical composition of stars. For example, what is the
© Pearson Education, Inc., publishing as Pearson Prentice Hall. All rights reserved.
emission spectrum of the sun, and what does this spectrum reveal about the types of
elements in the sun?
Scientists compare the emission spectrum of the sun with the wavelengths of lines
emitted by elements in the laboratory to determine the kinds of elements in the sun. At
least 67 different elements have been detected in the sun’s emission spectrum. The sun
ENERGIES OF ELECTRONS
7
PURPOSE
Text Reference
To construct a simple flame spectrograph and measure a wavelength
Section 5.3 of light produced by the electronic excitations of sodium ions.
Time Required
40 minutes BACKGROUND
You should recall from Experiment 6 that flame tests are useful for
Objectives identifying metal ions that produce characteristic colors. Separating
• Construct a simple flame these characteristic colors into discrete wavelengths of light produces a
spectrograph.
pattern of individual lines that uniquely identifies the metal ion. This
• Measure the 589-nm pattern of lines is called an emission spectrum. With a reference source
flame omission line of the
of emission spectra, you would find it relatively easy to identify a
sodium ion.
particular metal ion.
You can separate the lines in the visible region of a flame emission
Advance Preparation spectrum by using an optical prism or a diffraction grating. A
6M hydrochloric acid
spectrograph is an instrument designed to produce electronic
Dilute concentrated HCl 1:1
with distilled water. excitations, separate the emitted light into its component wavelengths,
Approximately 200 mL will be and then record the wavelengths of emitted light. In this experiment,
required to meet the needs of you will construct a simple spectrograph and measure the wavelength of
15 student lab pairs. a strong excitation of sodium ions.
1M sodium chloride
© Pearson Education, Inc., publishing as Pearson Prentice Hall. All rights reserved.
SAFETY FIRST!
In this lab, observe all precautions, especially the ones listed below. If
you see a safety icon beside a step in the Procedure, refer to the list below
for its meaning.
Caution: Wear your safety goggles. (All steps.)
PROCEDURE
As you perform the experiment, record your data and calculation results
in Data Table 1.
1. Set up the apparatus shown in Figure 7.1.
© Pearson Education, Inc., publishing as Pearson Prentice Hall. All rights reserved.
Use the following disposal
6. Dispose of the solutions as directed by your teacher.
methods for chemical waste.
Disposal 2: NaCl(aq).
Disposal 3: HCl(aq).
Left
image
A
B
90 Y
Right
80 image
Z
70
60
5 0
40
20
30
30 Movable 20
slit
40 10
50
60
Figure 7.1
OBSERVATIONS
DATA TABLE 1: WAVELENGTH FOR THE SODIUM EMISSION LINE
distance Y 40.0 cm
distance Z 42.9 cm
sin 0.359
15.0 cm 15.8 cm
15.4 cm
2
2. Calculate the distance Z, using the Pythagorean theorem. Refer to Figure 7.1 for
relationship of distances.
© Pearson Education, Inc., publishing as Pearson Prentice Hall. All rights reserved.
Z 兹苶
X2 Y2苶
Record the value of Z in the data table.
Z 兹(15.4
苶苶 cm)2 苶
(40.0苶
cm)2
兹237.1
苶苶 6 cm2苶
16苶
00 cm2苶
42.9 cm
4. The wavelength () of the sodium flame emission line being investigated in this
experiment is given, in nanometers, by the Bragg equation:
1 107 nm
冢
d sin
1 cm 冣
1
d represents a diffraction grating constant: d where n is the number of lines,
n
per centimeter, scribed on the diffraction grating. Calculate the value of d for your
grating and enter it in Data Table 1.
The number of lines per inch is often 13 400; d may be calculated
as follows:
1 in. 2.54 cm
d 1.90 104 cm/line (constant)
13 400 lines 1 in.
5. Compute the wavelength of the bright line you viewed on the meter stick, using
the Bragg equation. Record this value in the data table.
1 107 nm
冢
(1.90 104 cm) (0.359) 682 nm
cm 冣
6. The accepted value of for the observed transition is 589.0 nm. Calculate the
percent error in your value.
682 nm 589 nm
percent error 100 15.8%
589 nm
© Pearson Education, Inc., publishing as Pearson Prentice Hall. All rights reserved.
The ions of each element give characteristic patterns of lines in the spectrograph.
Even elements that produce flame-test results that appear to the eye to be the same
GOING FURTHER
Do Research
Atomic absorption spectroscopy is one of the most sensitive methods available for the
detection of various metals. Do research to find out how this method is used to
quantify the amount of an element in a particular sample and compare the detection
limits of this method with other spectroscopic methods of analysis.
The detection limits of many modern atomic spectroscopic methods lie well below 1 ppm
for a number of metallic elements. The ultrasensitivity of this technique is a valuable asset
to scientists studying materials that contain only trace amounts of a given metal.
INTRODUCTION TO THE
8 SPECTROPHOTOMETER
PURPOSE
To determine the absorption spectrum of an aqueous solution of
Text Reference chromium(III) ions.
Section 5.3
SAFETY FIRST!
In this lab, observe all precautions, especially the following ones. If you
see a safety icon beside a step in the Procedure, refer to the following list
for its meaning.
PROCEDURE
As you perform the experiment, record your percent transmittance data
in Data Table 1.
1. Turn on the spectrophotometer and allow it to warm up for
about 20 minutes.
Step 2. 2. Set the wavelength control knob to 375 nanometers (375 nm).
To avoid damage, the Adjust the amplifier control knob to produce 0 percent
spectrophotometer knobs transmittance (0%T) at this wavelength.
must not be twisted past the
3. Add 3 mL of distilled water to a clean, small test tube. Wipe the
point at which resistance is
encountered. Twisting the
outside of the tube with a tissue to make certain that it is clean
amplifier control knob and dry. Avoid getting fingerprints on the tube. Dislodge any air
counterclockwise past the bubbles present in the water by gently tapping the tube with a
resistance point turns the finger.
spectrophotometer off.
4. Place the tube in the sample holder and close the cover. Adjust
Step 4. the light control knob until the spectrophotometer reads
Show students how to tilt the 100%T.
bottom of the test tube
slightly to the right (facing 5. Remove the first sample from the spectrophotometer. Add
the spectrophotometer) to fit 3 mL of 0.02M chromium(III) nitrate, Cr(NO3)3, to another
into the holder. clean test tube. Use a tissue to clean and dry the tube. Insert the
© Pearson Education, Inc., publishing as Pearson Prentice Hall. All rights reserved.
tube of chromium(III) nitrate into the sample holder. Close the
The energy of the radiation cover of the holder. Read the percent transmittance and record
striking the detector is not the reading in Data Table 1. Remove the sample from the
constant at all wavelengths, holder.
because the energy output of 6. Turn the wavelength dial to 400 nm. Use the amplifier control
the source is not constant.
knob to adjust the percent transmittance to 0%T. Place the
The amplifier control knob
opens or closes the slit to water sample in the holder. With the light control knob, adjust
bring the energy at the the meter to 100%T. Replace the water sample with the
detector to the constant chromium(III) nitrate sample. Measure and record the percent
value designated 100%T. transmittance at 400 nm.
Failure to adjust to 100%T at
7. For the remainder of the wavelengths listed in Data Table 1,
each wavelength will
produce an “absorption continue the procedure of setting 0%T, setting 100%T, and
spectrum” that includes the measuring the percent transmittance of the chromium(III)
energy profile of the source nitrate solution.
as well as the spectrum of 8. Unless directed otherwise by your teacher, return the aqueous
the absorbing species in the
chromium(III) nitrate to the dropper bottle.
sample solution.
OBSERVATIONS
DATA TABLE 1: PERCENT TRANSMITTANCE AND ABSORBANCE OF 0.02M
Cr(NO3)3 SOLUTION AT VARIOUS WAVELENGTHS
about 420 nm and the other is at about 580 nm. These wavelengths
100
80
Percent Transmittance (%T)
60
40
Sample Data:
Percent Transmittance (%T)
versus Wavelength (nm)
20
0
350 400 450 500 550 600 650
© Pearson Education, Inc., publishing as Pearson Prentice Hall. All rights reserved.
Wavelength (nm)
Sample Data:
Absorbance (A)
Absorbance (A)
0.2
0
350 400 450 500 550 600 650
Wavelength (nm)
3. Based on the answer to problem 2, would you expect a red solution to absorb or
transmit red light? Explain.
The previous answer suggests that colored solutions absorb light of a
complementary color. Blue Cr(NO3)3 solution absorbs orange light. (Orange is the
complement of blue.) A red solution absorbs green light (the complement of red)
and transmits red light. (A yellow solution would absorb its complement, yellow.)
the peaks in an absorbance spectrum, but not in a transmittance spectrum. For this
GOING FURTHER
Develop a Hypothesis
Based on the results of this lab, propose a hypothesis about how the absorption of
light by a solution of Cr(NO3)3 varies with the concentration of Cr3 ions in solution.
The amount of light absorbed is directly proportional to the concentration of the
Design an Experiment
Propose an experiment to test your hypothesis. In your protocol, include a method
for determining the concentration of Cr3 ions in an unknown solution of Cr(NO3)3.
If resources are available and you have your teacher’s permission, perform the
experiment.
Measure the absorption of a series of standard solutions of chromium(III) ions at the
of absorption versus concentration. Draw a best-fit line through these points. The
measuring the absorption and reading its concentration from the graph.
© Pearson Education, Inc., publishing as Pearson Prentice Hall. All rights reserved.