Ap034 Inst Manual
Ap034 Inst Manual
Manual
AP034 Active
Differential Probe
AP034 Active Differential Probe
Operator’s Manual
May 2013
© 2013 Teledyne LeCroy, Inc. All rights reserved.
Unauthorized duplication of Teledyne LeCroy documentation materials other than for internal sales
and distribution purposes is strictly prohibited. However, clients are encouraged to distribute and
duplicate Teledyne LeCroy documentation for their own internal educational purposes.
WaveSurfer, WaveRunner, and Teledyne LeCroy are registered trademarks of Teledyne LeCroy, Inc.
Windows is a registered trademark of Microsoft Corporation. Other product or brand names are
trademarks or requested trademarks of their respective holders. Information in this publication
supersedes all earlier versions. Specifications are subject to change without notice.
Warranty
Teledyne LeCroy warrants this oscilloscope accessory for normal use and operation within
specification for a period of one year from the date of shipment. Spare parts, replacement parts and
repairs are warranted for 90 days.
In exercising its warranty, Teledyne LeCroy, at its option, will either repair or replace any assembly
returned within its warranty period to the Customer Service Department or an authorized service
center. However, this will be done only if the product is determined by Teledyne LeCroy’s
examination to be defective due to workmanship or materials, and the defect is not caused by
misuse, neglect, accident, abnormal conditions of operation, or damage resulting from attempted
repair or modifications by a non-authorized service facility.
The customer will be responsible for the transportation and insurance charges for the return of
products to the service facility. Teledyne LeCroy will return all products under warranty with
transportation charges prepaid.
This warranty replaces all other warranties, expressed or implied, including but not limited to any
implied warranty of merchantability, fitness or adequacy for any particular purposes or use.
Teledyne LeCroy shall not be liable for any special, incidental, or consequential damages, whether in
contract or otherwise.
922261-00 Rev A
May 2013
Operator’s Manual
Table of Contents
Safety Instructions............................................................................................... 1
Symbols .................................................................................................................. 1
Precautions............................................................................................................. 1
Operating Environment.......................................................................................... 2
Overview ............................................................................................................ 3
Description ............................................................................................................. 3
Applications ............................................................................................................ 3
Standard Accessories ............................................................................................. 4
Optional Accessories .............................................................................................. 4
Oscilloscope Software Compatibility...................................................................... 4
Specifications ...................................................................................................... 5
Operation ........................................................................................................... 8
Connecting the Probe to the Test Instrument ....................................................... 8
Connecting the Probe to the Test Circuit ............................................................... 8
Probe Input Loading ............................................................................................. 10
Grounding the Probe ............................................................................................ 10
Selecting the Proper Range .................................................................................. 11
Operation with Teledyne LeCroy Oscilloscopes ................................................... 12
Using AP034 with the APPPS Power Supply ......................................................... 12
Adding Offset........................................................................................................ 14
Autobalance ......................................................................................................... 15
Designing Test Fixtures for the AP034 Probe ....................................................... 15
Maintenance ..................................................................................................... 16
Cleaning ................................................................................................................ 16
Calibration Interval ............................................................................................... 16
Service Strategy .................................................................................................... 16
Troubleshooting ................................................................................................ 17
Returning a Probe ............................................................................................. 19
Replacement Parts ............................................................................................ 20
922261-00 Rev A i
AP034 Active Differential Probe
Matching Procedure for ÷10 Plugs ...................................................................... 22
Equipment Required.............................................................................................22
Procedure .............................................................................................................23
Performance Verification ................................................................................... 25
Test Equipment Required .....................................................................................25
Preliminary Procedure ..........................................................................................27
Procedure .............................................................................................................27
Adjustment Procedure ....................................................................................... 34
Introduction ..........................................................................................................34
Test Equipment Required .....................................................................................35
Preliminary Procedure ..........................................................................................37
Procedure .............................................................................................................38
Reference Information ....................................................................................... 47
Differential Mode and Common Mode ................................................................47
Differential Mode Range and Common Mode Range ..........................................47
Common Mode Rejection Ratio ...........................................................................48
Certifications ........................................................................................................49
Contact Teledyne LeCroy .................................................................................... 52
ii 922261-00 Rev A
Operator’s Manual
Safety Instructions
This section contains instructions that must be observed to keep this oscilloscope accessory
operating in a correct and safe condition. You are required to follow generally accepted safety
procedures in addition to the precautions specified in this section. The overall safety of any system
incorporating this accessory is the responsibility of the assembler of the system.
Symbols
These symbols may appear on the probe body or in this manual to alert you to important safety
considerations.
HIGH VOLTAGE, risk of electric shock.
DOUBLE INSULATION
Precautions
To avoid personal injury, or damage to the probe or test instrument, comply with the following
safety precautions.
Use product only as specified.
Connect and disconnect properly. Connect probe to the measurement instrument before
connecting the test leads to a circuit/signal being tested.
Use only accessories compatible with the probe. Use only accessories that are rated for the
application. Ensure connections between probe input leads and probe accessories are secure
before connecting them to a voltage source.
922261-00 Rev A 1
AP034 Active Differential Probe
Do not overload. To avoid electric shock, do not apply any potential that exceeds the maximum
rating of the probe and/or the probe accessory, whichever is less. Observe all terminal ratings of
the instrument before connecting the probe.
Be careful not to damage the insulation surface when making measurements.
Use only within operational environment listed. Do not use in wet or explosive atmospheres.
Keep product surfaces clean and dry. Use indoors only.
Handle with care. Probe accessory tips are sharp. They can puncture skin or cause other bodily
injury if not handled properly.
Keep fingers behind the finger guard of the probe accessories.
Do not operate with suspected failures. Before each use, inspect the probe and accessories for any
damage such as tears or other defects in the probe body, cable jacket, accessories, etc. If any part
is damaged, cease operation immediately and sequester the probe from inadvertent use.
Operating Environment
Only use the product within this operating environment:
Temperature: 0° to 50° C
Humidity: Maximum relative humidity 90 % for temperatures up to 31° C decreasing linearly to
50% relative humidity at 40° C
Altitude: Up to 10,000 ft (3,048 m)
2 922261-00 Rev A
Operator’s Manual
Overview
Description
The AP034 is a wide band differential active probe. The probe features low noise, low input
capacitance, high common mode rejection, and FET-buffered inputs in the probe head. User
selectable attenuation and offset give the probe flexibility to measure a large range of signal
amplitudes. Plug-on attenuator and AC coupling accessories further extend the application range.
Interconnect accessories included allow connection to surface mount and through-hole
components with minimal signal degradation. The input receptacles in the probe head are
compatible with standard 0.025 in. (0.635 mm) square pins. This provides a convenient low cost
method of creating device characterization test fixtures.
The probe is powered directly from a Teledyne LeCroy oscilloscope through the ProBus® interface.
The ProBus interface also allows local control of the probe through the oscilloscope user interface
and remote control through the interface buses, (GPIB, RS-232). The optional ADPPS power supply
allows the AP034 to be used with other instruments such as spectrum analyzers, network
analyzers, and oscilloscopes without ProBus interface.
Applications
The AP034 is ideal for acquiring high speed differential signals such as those found in disk drive
read channels, differential LAN, video, etc. It can also be used with spectrum analyzers to acquire
signals in some RF systems (for example, balanced IF mixers in hand held cellular telephones). The
high impedance characteristics of both inputs allow the probe to be used as a FET probe to make
single-ended measurements in digital systems without introducing a ground loop, as a conventional
FET probe would.
922261-00 Rev A 3
AP034 Active Differential Probe
Standard Accessories
Hard Case
÷10 Plug-on Attenuator
÷20 Plug-on Attenuator
Plug-on AC Coupler
Probe Connection Accessory Kit:
Flex Lead Set (1)
Mini Clip, 0.8 mm (3)
Mini Clip, 0.5 mm (2)
Ground Lead (1)
Offset Pins, Round (4)
Square Pin Header Strip (1)
Manual, AP034 Active Differential Probe Operator’s Manual
Optional Accessories
ADPPS Power Supply
4 922261-00 Rev A
Operator’s Manual
Specifications
Nominal Characteristics
Nominal characteristics describe parameters and attributes that have are guaranteed by design,
but do not have associated tolerances.
Input Configuration: True Differential (+ and – Inputs); with shield Ground connector.
922261-00 Rev A 5
AP034 Active Differential Probe
probe tip depends on proximity to probe cable ground, and is therefore non-reproducible. Teledyne LeCroy
has chosen to use a reproducible method of measurement, rather than obtain a more optimistic
measurement.
Typical Electrical Characteristics
Typical characteristics are parameters with no guaranteed performance. Tests for typical
characteristics are not provided in the Performance Verification Procedure.
Input Capacitance (each side to Ground): < 1.5 pF (See Figure 2.)
Harmonic Distortion (3rdorder distortion): –60 dB below fundamental (200 mVp-p output at 100 MHz)
(3rd order intercept): +20 dBm (at 100 MHz at output)
6 922261-00 Rev A
Operator’s Manual
General Characteristics
Temperature: 0° to 50° C Operating
-40° to 75° C Storage
Input Connectors: Compatible with 0.025” (0.635 mm) square pins.
0.036” (0.91 mm) maximum diameter (for round pins)
Power Requirements: Powered from oscilloscope through ProBus interface
or with ADPPS power supply.
Dimensions
Control Housing: Length: 3.625 in (9.2 cm)
Width: 1.50 in (3.8 cm)
Height: 1.00 in (2.5 cm)
Head: Length: 4.0 in (10.1 cm) w/o Attenuator or AC Coupler
Width: 2.25 in (2.25 cm)
Height: 0.625 in (1.6 cm)
Cable: Length: 42 in (106 cm)
Diameter: 0.275 in (7.0 mm)
Weight: Probe only: 6.4 oz (0.18 kg)
Shipping: 2 lbs, 8.4 oz (1.15 kg)
922261-00 Rev A 7
AP034 Active Differential Probe
Operation
CAUTION: The input circuits in the AP034 incorporate components that protect the probe
from damage resulting from electrostatic discharge (ESD). Keep in mind that this is an active probe,
and it should be handled carefully to avoid damage. When using the AP034, you are advised to take
precautions against potential instrument damage due to ESD.
8 922261-00 Rev A
Operator’s Manual
High common mode rejection requires precise matching of the relative gain or attenuation in the +
and – input signal paths. Mismatches in additional parasitic capacitance, inductance, delay, and a
source impedance difference between the + and – signal paths will lower the common mode
rejection ratio. Therefore, it is desirable to use the same length and type of wire and connectors for
both input connections. When possible, try to connect the inputs to points in the circuit with
approximately the same source impedance.
If AC coupling is desired, install the AC coupling accessory on the probe tip before connecting it to
the test circuit. The low-frequency cutoff (–3 dB point) of the AC coupler is approximately 16 Hz.
If the voltage in the test circuit exceeds the probe’s capability, add the external ÷10 or ÷20
attenuator* to the probe tip. If both the external attenuator and AC coupler are used, install the
attenuator on the probe tip first, then install the AC coupler on the attenuator input.
CAUTION: The external attenuators are precisely adjusted during manufacturing to match the
characteristics of the input of the probe with which they were shipped. The input characteristics of
the external attenuator itself do not exactly match those of the probe. Therefore, using the ÷10
and ÷20 attenuators at the same time is not recommended. The scale factor encoding system will
not operate correctly with both attenuators installed simultaneously.
The input characteristics of the AP034 1 GHz Active Differential Probe are significantly different
than those of the AP033 500 MHz Active Differential Probe. The external attenuators are not
interchangeable. The external attenuators currently supplied with the probes are labeled with the
appropriate model number. The attenuators supplied with model AP033 probes prior to the
introduction of the AP034 did not include the model number on the label. When using an external
attenuator with the AP034 Active Differential Probe, make sure it is labeled “AP034.”
Interchanging non-compatible attenuators will not damage the probes; however, the transient
response of the measured signals will be significantly in error.
In addition to being compatible with the included lead set, the probe input connectors will mate
with standard 0.025 in. (0.635 mm) square pins in any rotational orientation. To avoid damaging
the input connectors, do not attempt to insert connectors or wire larger than 0.036 in. (0.91 mm) in
diameter. Avoid rotating square pins after they are inserted into the input connectors.
The included accessories simplify the task of connecting the probe to the test circuit:
Use the small (0.5 mm) mini clips with the flexible lead set when connecting to fine-pitch
surface mount IC leads.
Use the larger (0.8 mm) mini clips to connect to through-hole leaded components.
Use the offset round pins for hand-held probing applications. Reposition the pins by
rotating them to obtain the required spacing.
922261-00 Rev A 9
AP034 Active Differential Probe
CAUTION: Do not use the attenuator encoding receptacle (unmarked socket near the – input)
to ground the probe. Connection to the encoding receptacle will not provide adequate grounding
and may result in an incorrect effective gain indication.
In many cases it is not necessary to ground the probe to the circuit under test. However, if the test
circuit is isolated from earth ground, it is usually necessary to connect the probe ground to a point
in the circuit. Grounding test circuits that are referenced to earth ground may improve the fidelity
of high frequency components in the waveforms. The potential for improvement with grounding
will vary depending on the common mode source impedance. However, connecting the probe
ground to a circuit that is referenced to earth ground can create a ground loop that may add noise
to low amplitude signals. The rejection of high frequency common mode signals is improved when
the probe head is ungrounded.
The best recommendation for connecting or not connecting the probe ground is to try both
configurations and select the one that performs the best.
10 922261-00 Rev A
Operator’s Manual
NOTE The AP034 Active Differential Probe transmits the measured signal differentially through the
probe cable. This essentially eliminates signal degradation from ground loop effects within the
probe. However, creating a ground loop may introduce signal distortions in the test circuit itself, or
in any coaxial cable between the ADPPS power supply and the test instrument.
Table 1. AP034 Dynamic ranges and input capacitance at different attenuator settings
922261-00 Rev A 11
AP034 Active Differential Probe
When using a differential probe or amplifier, be careful to avoid exceeding the common mode
range. Because the common mode signal is rejected by the differential probe, and is not displayed,
changes in the amplitude of the common mode component are not apparent to the user.
Exceeding the common mode range may introduce distortion into the AP034 output.
Circuitry in the probe detects the presence of either attenuator, and displays the effective gain of
the probe on the probe front panel.
12 922261-00 Rev A
Operator’s Manual
The AP034 offset allows you to measure signals up to 2.0 V (1.6 V of offset plus 400 mV of output
signal). With accessory plug-on attenuators, the effective offset is increased, and input signal 0f
20V (÷10) or 40 V (÷20) may be view ed. The peak signal that can be viewed at any equivalent scale
factor is the Input Differential Mode Range plus the offset used. See Adding Offset on the following
page.
When using the AP034 with the ADPPS power supply on oscilloscopes without ProBus interface,
use the following table to select the desired effective volts per division and determine the offset
available.
Table 2, Recommended settings to obtain desired equivalent Volts/Division
Input Maximum Input
Common
Equivalent Oscilloscope AP034 Differential Max Observable Noise
Mode
Scale Factor Scale Factor Attenuation Mode Offset Signal with (nV/√Hz)
Range
Range Offset Typical
2
2 mV/Div 2 mV/Div ÷1 ± 16V ± 8 mV ± 1.6 V ± 1.608 V 35
2
5 mV/Div 5 mV/Div ÷1 ± 16 V ± 20 mV ± 1.6 V ± 1.620 V 35
2
10 mV/Div 10 mV/Div ÷1 ± 16 V ± 40 mV ± 1.6 V ± 1.640 V 35
2
20 mV/Div 20 mV/Div ÷1 ± 16 V ± 80 mV ± 1.6 V ± 1.680 V 35
2
50 mV/Div 50 mV/Div ÷1 ± 16 V ± 200 mV ± 1.6 V ± 1.800 V 35
3
100 mV/Div 100 mV/Div ÷1 ± 16 V ± 400 mV ± 1.6 V ± 2.0 V 35
2
200 mV/Div 20 mV/Div ÷101 ± 42 V ± 800 mV ± 16 V ± 16.8 V 350
2
500 mV/Div 50 mV/Div ÷101 ± 42 V ±2V ± 16 V ± 18 V 350
3
1 V/Div 100 mV/Div ÷101 ± 42 V ±4V ± 16 V ± 20 V 350
3
2 V/Div 100 mV/Div ÷201 ± 42 V ±8V ± 32 V ± 40 V 700
922261-00 Rev A 13
AP034 Active Differential Probe
Adding Offset
When the AP034 Active Differential Probe is used with a Teledyne LeCroy oscilloscope equipped
with ProBus interface, the probe offset can be controlled with the channel OFFSET knob. The
buttons on the probe housing will be disabled.
When used with the ADPPS Power Supply on instruments that lack ProBus support, offset can be
controlled with the buttons on the probe’s front panel. The offset can be returned to zero at any
time by briefly pressing both the and OFFSET buttons at the same time. Pressing both the and
OFFSET buttons for more than 2 seconds will initiate an Autobalance cycle.
14 922261-00 Rev A
Operator’s Manual
NOTE: Probe offset is controlled with the channel OFFSET knob in oscilloscope firmware versions
8.1.0 or higher (including X-Stream firmware). The current offset value is displayed above the
graticule for a few seconds after a change is made by turning the knob. In prior software versions,
the channel OFFSET knob adjusted the offset of the oscilloscope input rather than the probe offset.
When the AP034 Active Differential Probe is used with Teledyne LeCroy oscilloscopes with older
software versions, it is important that the channel offset (controlled by the channel OFFSET knob)
remain set to 0.0000 V. Moving the displayed waveform with the channel offset rather than probe
offset will not re-center the dynamic operating range of the probe. Offsetting the trace with the
oscilloscope channel offset introduces the possibility of erroneous waveforms that result from
overloading the probe.
Autobalance
The AP034 Active Differential Probe incorporates an Autobalance function to remove the DC offset
from the output. Autobalance must be invoked by you. When power is first applied, the probe will
return to the internal values resulting from the last Autobalance cycle. For Autobalance to function
properly, all signals must be removed from the input.
After several minutes of warmup, or when the probe is exposed to a large shift in ambient
temperature, some DC offset drift may occur from thermal effects in the amplifier circuitry. To
initiate an Autobalance cycle, remove the probe from the test circuit and select the AUTOBALANCE
menu selection in the oscilloscope “COUPLING” menu for the channel the probe is connected to. If
the probe is being used with the ADPPS power supply, remove the input signal, then push and hold
both OFFSET buttons for two seconds. The AP034 must be terminated into 50Ω for successful
Autobalance.
Upon successful completion of the Autobalance cycle, all three of the EFFECTIVE GAIN indicators
will be briefly illuminated. If an input signal is present during auto balance and the routine fails, the
EFFECTIVE GAIN indicators will not illuminate. The probe will then revert to the offset values
resulting from the last. successful completion of the Autobalance cycle. In many situations, this will
be adequate to make routine measurements.
922261-00 Rev A 15
AP034 Active Differential Probe
Maintenance
This probe is a high quality, precision instrument. To maintain accuracy and signal fidelity,
mechanical shock should be avoided, as well as damage to the cable through excessive bending.
To achieve the small 2.5 mm tip size, the input tip diameter is narrower than those in larger probes.
Avoid placing excessive force sideways on the tip. Should the tip become damaged, it may be
replaced. Other maintenance and component replacement should be referred to qualified
personnel.
Cleaning
The outside of the probe should cleaned with a soft cloth dampened with either deionized /
distilled water or isopropyl alcohol. Allow the surface to dry completely before returning the probe
to service. Never immerse the probe in any liquid.
Calibration Interval
To guarantee accurate performance, you should have the probe calibrated every 12 months. Avoid
exposing the probe to extreme mechanical shock and excessive bending of the cable because these
may alter the calibration.
Service Strategy
The AP034 circuits utilize fine pitch surface mount devices; it is, therefore, impractical to attempt
component-level repair in the field. Defective probes must be returned to a Teledyne LeCroy
16 922261-00 Rev A
Operator’s Manual
service facility for diagnosis and exchange. A defective probe under warranty will be replaced with
a factory refurbished probe. A probe that is not under warranty can be exchanged for a factory
refurbished probe. A modest fee is charged for this service. The defective probe must be returned
in order to receive credit for the probe core.
Troubleshooting
If the probe is not operating properly the problem may be the way in which it is used. Before
assuming the probe is defective, perform the following troubleshooting procedures.
922261-00 Rev A 17
AP034 Active Differential Probe
C. DC Errors
Incorrect DC gain requires recalibration or factory repair. This can be determined by completing the
gain checks in the Performance Verification Procedure.
1. Extremely high source resistance will result in DC gain errors. Check the probe accuracy with
the oscilloscope calibrator signal.
2. Verify that the probe is not being overdriven into clipping for its current gain setting.
3. Excessive offset can result from large changes in ambient temperature. Remove the input
signal from the probe and repeat the Autobalance cycle. With the Offset set to zero, did the
trace return to the center of the graticule?
Returning a Probe
Return a probe for calibration or service by contacting your local Teledyne LeCroy sales
representative. They tell you where to return the product. All returned products should be
identified by both model and serial number. Provide your name and contact number, and a
description of the defect or failure (if possible).
Products returned to the factory require a Return Material Authorization (RMA) acquired by
contacting your nearest Teledyne LeCroy sales office, representative or the North America
Customer Care Center.
Return shipment should be prepaid. Teledyne LeCroy cannot accept COD or Collect Return
shipments. We recommend air-freighting.
1. Contact your local Teledyne LeCroy sales or service representative to obtain a Return Material
Authorization.
2. Remove all accessories from the probe. Do not include the manual.
3. Pack the probe in its case, surrounded by the original packing material (or equivalent) and box.
4. Label the case with a tag containing
The RMA
Name and address of the owner
Probe model and serial number
Description of failure
5. Package the probe case in a cardboard shipping box with adequate padding to avoid damage
in transit.
6. Mark the outside of the box with the shipping address given to you by the Teledyne LeCroy
representative; be sure to add the following:
ATTN: <RMA assigned by the Teledyne LeCroy representative>
FRAGILE
7. Insure the item for the replacement cost of the probe.
922261-00 Rev A 19
AP034 Active Differential Probe
8. If returning a probe to a different country, also:
Mark shipments returned for service as a “Return of US manufactured goods for
warranty repair/recalibration.”
If there is a cost involved in the service, put the service cost in the value column and
the replacement value of the probe in the body of the invoice marked “For insurance
purposes only.”
Be very specific as to the reason for shipment. Duties may have to be paid on the value
of the service.
Replacement Parts
The probe connection accessories, and other common parts can be ordered through the regional
customer care centers. Refer to Table 3 for Teledyne LeCroy part numbers. Defective probes can be
replaced on an exchange basis. The replacement exchange probe will have been factory repaired,
inspected and calibrated to the same standards as a new product. In order to obtain an exchange
probe, you must return the defective probe. The returned probe should be sent back to the
regional customer care center without any accessories, manual, or case.
922261-00 Rev A 21
AP034 Active Differential Probe
Equipment Required
Test Oscilloscope
The oscilloscope must support ProBus. Otherwise use a non- ProBus oscilloscope and ADPPS
power supply.
Signal Source
Low frequency square wave: Frequency 50 Hz to 5 kHz, Amplitude 1 V to 10 V. The output
waveform must have a square corner and flat top with minimum overshoot suitable for adjusting
compensation. The generator should have trigger output, or use a BNC Tee connector and separate
BNC cable from the output to provide the trigger signal for the test oscilloscope.
Interconnect Cable
This is for connecting the output of the signal source to the probe. A BNC cable and a pair of small
alligator clips or “lead grabber” adapter (Pomona #3788) may be used. 0.025 in. (0.635 mm)
Square Pins (3 required). The pins from the header supplied in the probe accessory kit are suitable.
Tools
Flat bladed screwdriver, 0.040 in. (1 mm) wide
Adjustment Tool: 0.025 in. (0.635 mm) square head
22 922261-00 Rev A
Operator’s Manual
Note: You can fabricate the Adjustment Tool by flattening the end of a 0.025 in. (0.635 mm)
square pin with a file. Insert the pin into a short length of rigid plastic tubing to serve as a handle.
Procedure
1. Attach the AP034 to the test oscilloscope. If the test oscilloscope is not equipped with ProBus,
use the ADPPS to provide power for the AP034.
2. Attach the ÷or ÷Attenuator Adapter to the AP034 probe tip.
3. Insert 0.025 in. (0.635 mm) square pins into the +, –, and input connectors of the Plug-on
Attenuator.
4. Attach the interconnect cable to the output of the signal source.
5. Attach the Trigger Out signal from the signal source to the External Trigger Input of the test
oscilloscope. If the signal source does not have a separate Trigger Out signal, use a BNC Tee
connector in the output. Run one cable to the External Trigger Input of the test oscilloscope.
Connect the other to the probe inputs.
6. Using the alligator clips on the end of the interconnect cable, connect the signal source ground
to the square pin on the Attenuator Adapter’s “–“ input. Attach the signal output to the square
pin on the Attenuator Adapter’s “+” input.
7. Turn on the test oscilloscope. The EFFECTIVE GAIN indicator for ÷10 or ÷20 should be lit. (NOTE:
If the X1 EFFECTIVE GAIN indicator is lit instead of the ÷10 or ÷20 indicator, make sure that the
plug-on adapter installed on the probe tip is the Attenuator, and not the AC Coupler.)
8. Set the test oscilloscope Volts/Div to 200 mV (for Teledyne LeCroy oscilloscopes with ProBus)
or 20 mV/Div when using the ADPPS with an oscilloscope without scale factor correction; 5
μs/Div; AUTO trigger mode; Trigger source: External. Set the Bandwidth Limiting to 20–30 MHz
and Average the waveform 1:31 to reduce noise.
9. Turn on the signal source. Set the output frequency to approximately 5 kHz. Set the output
amplitude to approximately 1 V for matching the ÷10 Attenuator Adapter, or 2 V for matching
the ÷20 Attenuator Adapter.
10. Adjust the test oscilloscope trigger level for a stable trace. If necessary, use the probe offset to
position the waveform to show the square corner of the test signal.
NOTE: Do not use the oscilloscope offset or position controls to reposition the trace. Many of the
signal generators used for compensation calibration only have square corners on one of the two
edges of the output waveform. (Rising or falling edge, but not both.) Be sure to display the correct
edge for this step.
922261-00 Rev A 23
AP034 Active Differential Probe
11. Using the square adjustment tool, adjust the +LF Comp (See Figure 7) to achieve the best
square corner and flat top of the displayed waveform. Note that the added capacitance of the
adjustment tool may change the compensation of the waveform when it is inserted. The
correct adjustment is achieved when the best corner is displayed with the adjustment tool
removed
.
Figure 7, ÷10 Attenuator Adjustment Locations
12. Move the connections on the interconnect cable so the signal generator ground is connected to
the square pin inserted in the ground connector of the Attenuator Adapter. Connect the signal
generator output to BOTH the + input and – input square pins in the Attenuator Adapter.
13. Set the signal generator frequency to about 50 Hz, and the output amplitude to about 10 V.
14. Set the test oscilloscope Volts/Div to 20 mV (for Teledyne LeCroy oscilloscopes with ProBus) or
2 mV/Div when using the ADPPS with an oscilloscope without scale factor correction; and set
the time scale to 2 ms/Div. It may be necessary to increase the averaging to 1:127 to remove
noise.
15. The displayed waveform is the common mode response. Using the adjustment screwdriver
with the flat blade, adjust the DC Atten. Balance (Figure 7) to minimize the amplitude of the flat
portions of the displayed waveform. This adjustment only affects the flat portions of the square
wave. Do not be concerned with any overshoot at the transitions.
16. Reduce the Test Oscilloscope Averaging weighting to 1:31. Return the Volts/Div to 200 mV (for
Teledyne LeCroy oscilloscopes with ProBus) or 20 mV/Div when using the ADPPS with an
oscilloscope without scale factor correction, and the Time/Div to 5 s.
17. Return the signal source output frequency to approximately 5 kHz. Set the output amplitude to
approximately 1 V for the ÷10 Attenuator Adapter, 2 V for the ÷20 Attenuator Adapter.
18. Using the 0.025 in. square adjustment tool, adjust the –LF Comp (Figure 7) to minimize the
amplitude of overshoot possible to completely eliminate the overshoot. As with the +LF Comp
adjustment, the added capacitance of the adjustment tool may change the amplitude of the
waveform when it is inserted. The correct adjustment is achieved when the overshoot is
minimized with the adjustment tool removed.
24 922261-00 Rev A
Operator’s Manual
Performance Verification
This procedure can be used to verify the warranted characteristics of the AP034 Active Differential
Probe. You can do the performance verification without removing the instrument covers and
exposing yourself to hazardous voltages.
Adjustment should only be attempted if a parameter measured in the Performance Verification
Procedure is outside of the specified limits. Adjustment should only be performed by qualified
personnel.
The recommended calibration interval for the model AP034 Active Differential Probe is one year.
The complete performance verification procedure should be performed as the first step of annual
calibration. You can record test results on a photocopy of the Test Record at the end of this
section.
922261-00 Rev A 25
AP034 Active Differential Probe
Table 4, Performance Verification Test Equipment
Description Minimum Requirements Test Equipment Examples
1
Wide Band Oscilloscope Minimum 1 GHz bandwidth Teledyne LeCroy LC584A
2 mV to 5 V scale factors ProBus
interface equipped
2% vertical accuracy
Digital Multimeter DC: 0.1% accuracy Agilent Technologies 34401A
AC: 0.2% accuracy to measure 200 Fluke 8842A-09
mV and 2 V rms @ 1 kHz 5½ digit
Keithley 2001
resolution
Function Generator Sine Wave and Square Wave output Agilent Technologies 33120A
waveforms
Stanford Research Model DS340
20 Vp-p into 1 MΩ
Leader LAG-120B
70 Hz to 10 MHz frequency range
High Frequency Sine Wave 50 to 500 MHz and 50 kHz Agilent Technologies 8648C
Generator Output adjustable from 300 mV to2
Vp-p
Terminator, in-line, BNC 50 Ω ±2% coaxial termination Pomona 4119-50
Terminator, precision, BNC 50 Ω ±0.2% Teledyne LeCroy TERM-CF01
Attenuator, BNC 50 Ω ±2%, ÷10 (20 dB) Pomona 4108-20dB
BNC coaxial cable (2 required) male-male BNC, 50 Ω, 36 in. Pomona 5697-36
Calibration Fixture ProBus Extension Cable Teledyne LeCroy PROBUS-CF01
Calibration Fixture AP033/AP034 Calibration Fixture Teledyne LeCroy AP03x-CF01
Banana Plug adapter BNC female-to-banana plug Pomona 1269
NOTE:1 If a Teledyne LeCroy ProBus equipped oscilloscope is not available, you may use an
alternate oscilloscope that meets the other minimum specifications listed, and the model ADPPS
power supply, to perform the performance verification procedure. The input termination of the
alternate oscilloscope must be set to 50 Ω and the offset or position must remain at center screen.
26 922261-00 Rev A
Operator’s Manual
Preliminary Procedure
1. Connect the AP034 Active Differential Probe to the female end of the ProBus Extension Cable.
Connect the male end of the ProBus extension cable to Channel 1 of the oscilloscope.
2. Turn the oscilloscope on and allow at least a 30- minutes warm-up time for the AP034 and test
equipment before performing the Verification Procedure.
3. Turn on the other test equipment and allow these to warm up for the time recommended by
the manufacturer.
4. While the instruments are reaching operating temperature, make a photocopy of the
Performance Verification Test Record (located at the end of this section), and fill in the
necessary information.
The warranted characteristics of the AP034 Active Differential Probe are valid at any temperature
within the Environmental Characteristics listed in the Specifications. However, some of the other
test equipment used to verify the performance may have environmental limitations required to
meet the accuracy needed for the procedure. Make sure that the ambient conditions meet the
requirements of all the test instruments used in this procedure.
Procedure
NOTE: Correct operation of the AP034 controls requires oscilloscope software version 8.1.0 or
higher. The use of earlier versions is not recommended. You can verify the software version in the
test oscilloscope by pressing the SHOW STATUS button, then selecting the System menu option.
Contact your local Teledyne LeCroy representative if the software in your oscilloscope requires
updating.
922261-00 Rev A 27
AP034 Active Differential Probe
4. Connect one end of a second BNC cable to the output of the Function Generator. Attach the
BNC-to-dual-male banana plug adapter to the free end of the second BNC cable. Connect the
banana plugs to the Digital Multimeter (DMM).
5. Set the DMM to measure AC Volts.
6. Set the mode of the Function Generator to Sine Wave; the frequency to approximately 1 kHz;
and the output amplitude to approximately 200 mV rms, as measured by the DMM.
7. Record the measured amplitude to 100 μV resolution in the Test Record.
8. Unplug the output cable from the DMM. Remove the BNC-to banana plug adapter from the
end of the cable.
9. Autobalance the AP034 by pressing the AUTOBALANCE menu button located in the “Coupling”
menu.
10. Carefully align the four pins that correspond to the Differential Drive No Termination portion
of the AP03x- CF01 Calibration Fixture with the input receptacles in the AP034 probe head.
Press the probe into the fixture to fully engage the pins.
11. Connect the banana plugs of the precision terminator to the digital multimeter (DMM)
12. Connect the free end of the output cable from the Function Generator to the Differential Drive
No Termination connector of the AP033/AP034 Calibration Fixture.
13. After the DMM reading has stabilized, record the measured output amplitude to 100 μV
resolution in the Test Record.
14. Divide the measured output voltage (recorded in step A-13) by the sine wave generator output
voltage (probe input voltage) from step A-7. Subtract 1.0 from the ratio and multiply the result
by 100 to get the error in percent.
15. Record the answer to two significant places (±x.xx%) on line A-15 in the Test Record.
16. Verify that the X1 gain error is less than ±2%.
17. Disconnect both BNC cables from the test setup.
28 922261-00 Rev A
Operator’s Manual
922261-00 Rev A 29
AP034 Active Differential Probe
10. Move the high frequency sine wave generator output cable from the oscilloscope channel 2
input to the Common Mode Drive 50 ohm Termination connector of the AP033/AP034
Calibration Fixture.
11. Set the high-frequency sine wave generator RF Output to ON.
12. Set the oscilloscope to display channel 1 only, set the trigger source to channel 1, and set the
channel 1 vertical scale factor as needed to measure the amplitude of the displayed waveform.
Adjust the trigger level as necessary for a stable display. This is the common mode signal. Turn
on Math Channel A. Press MATH SETUP, then REDEFINE A. Set the A Math type to Average, Avg
Type to Continuous, with 1:15 weighting, of channel 1. Turn off the trace 1 display.
NOTE: The amplitude of the Common Mode signal should be relatively small. If the output
waveform appears to be a 1 Volt square wave, check that the Common Mode Drive 50 ohm
Termination connector of the AP033/AP034 Calibration Fixture is being used and not the
Differential Drive 50 ohm Termination connector.
13. Measure the peak-to-peak output amplitude of the AP034. Record the reading to two-digit
resolution (xx mV) as “Common Mode Signal at 100 MHz” in the Test Record.
14. Calculate the Common Mode Rejection Ratio (CMRR) at 100 MHz by dividing 2,000 mV by the
measured Common Mode Signal recorded in step B-14 (direct reciprocal of the Common Mode
Gain). Record the result to two-digit resolution (xx:1) in the Test Record.
15. Verify that the CMRR at 100 MHz is greater than 18:1 (25 dB).
16. Turn off the high frequency sine wave generator RF output, and disconnect the output cables.
30 922261-00 Rev A
Operator’s Manual
second BNC cable from the remaining end of the BNC Tee adapter to the Common Mode Drive
No Termination input connector of the AP033/AP034 Calibration Fixture.
CAUTION: Make sure that the Common Mode Drive No Termination connection is used.
Prolonged application of the power levels used in the low frequency common mode test may
damage the termination resistance in either 50 ohm Termination input of the AP033/AP034
calibration fixture.
4. Set the sine wave generator frequency to 70 Hz, and the output amplitude to 20 Vp-p, (4
divisions on the oscilloscope). If necessary, adjust the trigger level for a stable display.
5. Set the oscilloscope to display channel 1, but leave the trigger source set to channel 2. Set the
vertical scale of channel 1 to 2 mV/div. Create a math waveform on channel A defined as the
Average of channel 1. Set the average factor as necessary to reduce noise. Increase the zoom
to a factor of four (0.5 mV/division). Turn off the waveform display of all channels except the A
Math waveform.
6. Increase the zoom of the A Math waveform as needed to measure the peak-to-peak amplitude.
This is the common mode signal.
7. Record the displayed “Common Mode Signal at 70 Hz” to two-digit resolution (x.xx mV) in the
Test Record.
8. Calculate the Common Mode Rejection Ratio (CMRR) at 70 Hz by dividing 20 000 mV by the
measured Common Mode Signal recorded in step C-7 (direct reciprocal of the Common Mode
Gain). Record the result to two-digit resolution (xx 000:1) in the Test Record.
9. Verify that the “Common Mode Rejection Ratio at 70 Hz” is greater than 10 000:1 (80 dB).
10. Change the output amplitude of the Function Generator to approximately 8.0 Vp-p, and the
frequency to 1 MHz.
11. Set the oscilloscope to display channel 2 only. Change the horizontal scale factor to 0.5 s/div,
and the channel 2 vertical scale to 2 V/div. Turn the BWL on channels 1 and 2 to 200 MHz.
12. Adjust oscilloscope trigger level as necessary for a stable display. Adjust the Function Generator
amplitude as needed to maintain 8.0 Vp-p as measured on channel 2.
13. Turn off the oscilloscope channel 2 display, turn on the display of channel 1 and the A Math
waveform (averaged channel 1). Adjust the scale factor of channel 1 and the Zoom of the A
Math waveform as necessary to accurately measure the amplitude of the averaged signal.
14. Record the displayed “Common Mode Signal at 1 MHz” to two-digit resolution in the Test
Record.
922261-00 Rev A 31
AP034 Active Differential Probe
15. Calculate the “Common Mode Rejection Ratio (CMRR) at 1 MHz” by dividing 8000 mV by the
measured Common Mode Signal recorded in step C-14. Record the result in the Test Record.
16. Verify that the Common Mode Rejection Ratio at 1 MHz is greater than 100:1 (40 dB).
17. Remove all cables and test fixtures from the AP034 probe. This completes the Performance
Verification of the AP034. Complete and file the results recorded in the AP034 Performance
Verification Test Record, as required by your quality procedures. Apply a suitable calibration
label to the AP034 housing as required.
32 922261-00 Rev A
Operator’s Manual
AP034
Model:
Serial Number:
Asset or Tracking Number:
Date:
Technician:
CALIBRATION
MODEL SERIAL NUMBER
DUE DATE
OSCILLOSCOPE
DIGITAL MULTIMETER
LEVELED SINE WAVE GENERATOR
1 The
function generator provides stimulus for making relative measurements. The output amplitude of the generator is
measured with the DMM or oscilloscope in the procedure. Thus, the generator is not required to be calibrated.
922261-00 Rev A 33
AP034 Active Differential Probe
Adjustment Procedure
Introduction
You can use this procedure to adjust the AP034 Active Differential Probe to meet the warranted
specifications. This procedure should only be performed if the instrument fails the Performance
Verification tests. If the probe cannot be adjusted to meet the Performance Verification limits,
repair may be necessary. To ensure instrument accuracy, check the calibration of the AP034 Active
Differential Probe every year. Before calibration, thoroughly clean and inspect this unit as discussed
in the “Cleaning” section.
Completion of each step in the Adjustment Procedure ensures that the differential probe meets
specifications. Some of the adjustments interact with other parts of the circuitry. Therefore, it is
necessary that all adjustments be performed in the order listed. For best overall instrument
performance, make each adjustment to the exact setting, even when adjustment is within the
limits stated in the procedure.
Adequate guard bands were designed into the AP034 Active Differential Probe to ensure that it will
meet or exceed published specifications over the entire operating temperature range. To continue
to meet the environmental specifications, all adjustments must be performed in a controlled
environment with an ambient temperature of 25 ±5 °C. The AP034 Active Differential Probe must
also be at stable operating temperature before performing adjustments.
CAUTION: The adjustment procedure will require removal of the probe covers. These covers
are part of the ESD protection system of the AP034 Active Differential Probe. To protect the probe,
you should perform the entire procedure on a static dissipating work surface Wear an antistatic
grounding wrist strap and follow standard static control procedures.
The probe tip housing provides physical rigidity to the input pins of the probe. When the covers are
removed, observe extra caution to avoid breaking the probe tip receptacles when mating the probe
to the calibration fixture.
34 922261-00 Rev A
Operator’s Manual
Digital Multimeter 0.1% DC volts accuracy at 0.4 V 0.2% Agilent Technologies 34401A
AC volts accuracy at 2 V and 1 kHz
Fluke 8842A-09
5½ digit resolution
Keithley 2001 with test leads
0.01 mV AC volts resolution
Function Generator Sine and square wave output Agilent Technologies 33120A
20 Vp-p into 1 MΩ Stanford Research Model DS340
50 Hz to 1 MHz frequency Range Leader LAG-120B
Sync. or Freq. Reference output
Fast Rise Pulse Generator Risetime <2 ns Calibrator signal from Teledyne
LeCroy oscilloscope
Amplitude >400 mV into 50 Ω
Over/Undershoot <6% Tegam / Tektronix PG506 with
power unit
Sine Wave Generator Output 250 MHz at 1 V RMS Agilent Technologies 8648C, with
male N to female BNC
adapter for the output connector.
Tegam SG504 with TM series
mainframe.
922261-00 Rev A 35
AP034 Active Differential Probe
Banana Plug Adapter BNC female to dual male banana plug Pomona 1269
(2 Required)
Shorting Plug Two 0.025 in. square pins on 0.100 in. Supplied with AP03X-CF01
centers
Low Capacitance
Screwdriver
NOTE: If a Teledyne LeCroy ProBus equipped oscilloscope is not available, you may perform the
adjustment procedure with an alternate oscilloscope that meets the other Minimum Specifications,
and the model ADPPS power supply. The input termination of the oscilloscope must be set at 50 Ω ,
the offset or position must remain at center screen, and the high-frequency adjustments must be
performed with the ADPPS connected directly to the input of the oscilloscope.
36 922261-00 Rev A
Operator’s Manual
Preliminary Procedure
1. Using pliers, carefully remove each of the push buttons from the ProBus interface housing by
gently pulling them away from the housing.
2. Remove the two screws that secure the plastic cover on the cable end of the ProBus interface
housing. Gently pull on the probe cable to slide the circuit board assembly from the metal
housing.
3. Remove the 5/64 in. (2 mm) Allen head cap screw from the small cover on the back of the
probe tip. Remove the cover along with the small cover on the opposite side of the probe. Hold
the shielded portion of the probe head in one hand and gently slide the larger cover off by
pulling it away from the probe tip end.
4. Connect the AP034 Active Differential Probe output to the female end of the ProBus Extension
Cable. Be careful to align the ProBus pins with the corresponding connector correctly. Connect
the male end of the ProBus Extension Cable to Channel 1 of the oscilloscope.
NOTE: The Logic board is connected to the Amplifier board by four small 8-pin connectors.
There are no additional mechanical fasteners holding the two boards together. A small foam
pad mounted on the circuit board ensures that the connectors remain engaged while the
assembly is mounted within the ProBus interface housing. Be sure that the connectors
between the boards are firmly engaged before applying power to the probe. Operating the
probe without the logic board will not damage it. However, to ensure reliable operation, the
logic board should only be mated with the Amplifier board with the power removed, and the
ProBus connector disengaged.
5. Apply power to the oscilloscope and to the other test instrumentation. Allow at least 30
minute’s warm-up time for the AP034 Active Differential Probe and test equipment before
continuing the calibration procedure.
NOTE: The probe tip cover also serves to thermally stabilize the input circuitry. The differential
input stage of the probe utilizes discrete transistors that need to maintain an approximate
match of junction temperatures for correct DC balance. With the covers removed, this circuitry
is susceptible to drift caused by air currents flowing over the components. Remove or redirect
any fans that may be blowing air currents over the probe tip during adjustment.
Correct operation of the AP034 Active Differential Probe controls requires software version
8.1.0 or higher. You can find out the software version loaded on your Teledyne LeCroy scope by
pressing SHOW STATUS, then selecting the System menu option. In version 8.1.0 and higher,
the OFFSET controls on the front panel of the probe are disabled. Offset is controlled by the
OFFSET knob in the oscilloscope CHANNEL section.
922261-00 Rev A 37
AP034 Active Differential Probe
Procedure
A. Adjust Coarse DC Balance (R226)
1. If necessary, set the probe offset to 0.000 V by rotating the OFFSET knob in the CHANNEL
section of the oscilloscope.
2. Insert the two-pin shorting plug into the socket on the Logic Board. Use the two holes
closest to the probe cable (Figure 9).
38 922261-00 Rev A
Operator’s Manual
922261-00 Rev A 39
AP034 Active Differential Probe
40 922261-00 Rev A
Operator’s Manual
10. Set the oscilloscope to display channel 3. The waveform is the common mode signal. Turn
the offset as necessary to keep the trace on screen. Turn on Math Channel A. Press MATH
SETUP, then REDEFINE A. Set the A Math type to Average, Avg Type to Continuous, with
1:15 weighting, of channel 3. Turn off the trace 3 display. Use the Math Zoom and Position
controls as necessary to view the waveform.
11. Adjust LF CMRR (R7) for minimum amplitude. (See Figure 11 for location.)
12. Remove the BNC cable from the output of the Function Generator and the calibration fixture.
Leave the BNC cable from the SYNC output to the oscilloscope external trigger input in
place.
922261-00 Rev A 41
AP034 Active Differential Probe
5. Connect the other end of the cable to the DMM input, making sure that the plug
corresponding to the BNC shield (marked “Ground”) is connected to the LO or COM input.
6. Set the DC Voltage Source to read as close as possible to +1.600 V on the DMM.
7. Autobalance the AP034 by pressing the AUTOBALANCE menu button located in the
“COUPLING” menu twice, or by pressing and holding both probe offset buttons for at least
two seconds.
8. Disconnect the BNC cable from the female BNC-to-dual male banana plug adapter on the
DMM.
9. Reconnect the BNC cable from the DC Voltage Source to the Differential Drive No
Termination connector on the AP033/AP034 Calibration Fixture.
10. Set the probe offset to –1.600 V by rotating the OFFSET knob in the channel section of the
oscilloscope.
11. Allow 10 seconds for the AP034 to stabilize.
12. Using a low capacitance screwdriver, adjust Offset (R330) to bring the trace back to exactly
center screen. (See Figure 9 for adjustment location.)
13. Remove the dual banana plug adapter from the output of the DC voltage source and
reconnect it, with the pins reversed. (The plug corresponding to the BNC shield (marked
“Ground”) should now be connected to the +Output connector.)
14. In the CHANNEL menu of the oscilloscope, set the probe offset to +1.600 V by rotating the
OFFSET knob.
15. Allow 10 seconds for the AP034 to stabilize.
16. Note the amplitude of the trace from center screen.
17. Adjust Offset Cal (R330) to position the trace to approximately ½ the amplitude from
center screen noted in the previous step. This amplitude should be within •}10 mV of
center screen. Note this value.
18. Again, remove the dual banana plug adapter from the output of the DC voltage source and
reconnect it, with the pins reversed. The plug corresponding to the BNC shield (marked
“Ground”) should now be connected to the – or Common Output connector.
19. In the CHANNEL menu of the oscilloscope, set the probe offset to –1.600 V by rotating the
OFFSET knob.
20. Allow 10 seconds for the AP034 to stabilize.
42 922261-00 Rev A
Operator’s Manual
21. The trace should be offset from the center line by the same amplitude but opposite
polarity of the value noted in step E-16.
22. Repeat steps E-13 through E-21 as necessary until the errors at +1.6 V and at –1.6 V are
approximately equal, and both are within 10 mV of the center scale. Reverse the polarity of
the dual banana plug adapter at the output of the DC voltage source, and the
corresponding probe offset value with each repetition.
23. Disconnect the cable from the DC Voltage Source. Keep the AP033/AP034 Calibration
Fixture connected to the AP034 for the next step.
922261-00 Rev A 43
AP034 Active Differential Probe
44 922261-00 Rev A
Operator’s Manual
3. Set the oscilloscope display to channel 4, coupling to DC1MΩ , vertical scale factor to 200
mV/div, and the trigger source to channel 4. Adjust the oscilloscope trigger level as
necessary for a stable display.
4. Set the oscilloscope to display channel 1, and vertical scale factor to 10 mV/div. Set the
OFFSET to 0.0 mv
5. 5. Adjust Attenuator CMRR (capacitor C18A) for minimum amplitude shift of the displayed
waveform at the trigger point in time. See Figure 9 for adjustment location.
6. Leave the setup connected for the next step.
922261-00 Rev A 45
AP034 Active Differential Probe
46 922261-00 Rev A
Operator’s Manual
Reference Information
Differential Mode and Common Mode
Differential probes amplify the voltage difference that appears between the + and – inputs. This
voltage is referred to as the Differential Mode or Normal Mode voltage. The voltage component
that is referenced to earth ground, and is identical on both inputs, is rejected by the amplifier. This
voltage is referred to as the Common Mode voltage, because it is common to both inputs. The
common mode voltage can be expressed as:
922261-00 Rev A 47
AP034 Active Differential Probe
48 922261-00 Rev A
Operator’s Manual
Certifications
This section certifies the probe’s Electromagnetic Compatibility (EMC), Safety and Environmental
compliance.
EMC Compliance
EC DECLARATION OF CONFORMITY - EMC
The probe meets intent of EC Directive 2004/108/EC for Electromagnetic Compatibility. Compliance
was demonstrated to the following specifications as listed in the Official Journal of the European
Communities:
EN 61326-1:2006, EN 61326-2-1:2006 EMC requirements for electrical equipment for
measurement, control, and laboratory use.
Electromagnetic Emissions:
CISPR 11:2003, Radiated and Conducted Emissions Group 1, Class A 1 2
Electromagnetic Immunity:
EN 61000-4-2:2001 Electrostatic Discharge, 4 kV contact, 8 kV air, 4 kV vertical/horizontal coupling
planes 3
EN 61000-4-3:2006 RF Radiated Electromagnetic Field, 3 V/m, 80-1000 MHz; 3 V/m, 1400 MHz - 2
GHz; 1 V/m, 2 GHz - 2.7 GHz 3
1 Emissions which exceed the levels required by this standard may occur when the probe is connected to a
test object.
2 This product is intended for use in nonresidential areas only. Use in residential areas may cause
electromagnetic interference.
3 Meets Performance Criteria “B” limits of the respective standard: during the disturbance, product
undergoes a temporary degradation or loss of function or performance which is self-recoverable.
European Contact:
Teledyne LeCroy Europe GmbH
Waldhofer Str 104
D-69123 Heidelberg
Germany
Tel: (49) 6221 82700
922261-00 Rev A 49
AP034 Active Differential Probe
Safety Compliance
EC DECLARATION OF CONFORMITY – LOW VOLTAGE
The probe meets intent of EC Directive 2006/95/EC for Product Safety. Compliance was
demonstrated to the following specifications as listed in the Official Journal of the European
Communities:
EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and
laboratory use – Part 1: General requirements
EN 61010-2:030:2010 Safety requirements for electrical equipment for measurement, control, and
laboratory use – Part 2-030: Particular requirements for testing and measuring circuits
EN 61010-031/A1:2008 Safety requirements for electrical equipment for measurement, control, and
laboratory use – Part 031: Safety requirements for hand-held probe assemblies for electrical measurement
and test.
50 922261-00 Rev A
Operator’s Manual
Environmental Compliance
END-OF-LIFE HANDLING
The probe is marked with this symbol to indicate that it complies with the applicable
European Union requirements to Directives 2002/96/EC and 2006/66/EC on Waste
Electrical and Electronic Equipment (WEEE) and Batteries.
The probe is subject to disposal and recycling regulations that vary by country and
region. Many countries prohibit the disposal of waste electronic equipment in
standard waste receptacles. For more information about proper disposal and
recycling of your Teledyne LeCroy product, please visit teledynelecroy.com/recycle.
ISO Certification
Manufactured under an ISO 9000 Registered Quality Management System. Visit
teledynelecroy.com to view the certificate.
922261-00 Rev A 51
AP034 Active Differential Probe
52 922261-00 Rev A