seasafe
CERTIFICATE OF CALIBRATION FOR PORTABLE GAS DETECTOR
Gertincate No.: 2212/28-001 ()
Name of VesseV/ Company MIURA; (IMO NO.: 8910902)
Unit Under Test: Portable Multi Gas Detector
Date of Test 28 December 2022
Particulars of Unit under Test
Description ‘Multi Gas detector
‘Manufacturer/Mark RIKEN KEIKI Co., Ltd.
Model/type No. GX-2009
‘Identity/ SL No. 943110711RN
0 to 100% LEL, 0 to 40% VOL O2, |
Rangefreeokition: 0-100 ppm VOL HS, 0-500ppm CO
Particulars of Calibration Apparatus Used
Test instrument PORTAGAS Calibration gas cylinder with regular flow indicator
‘SLJ/LOT No. 988250
Composition ‘CH4 = 50% LEL; O2: 18%; HaS : 25ppm, and CO : 100ppm_
Treconbiy “The calibration gas is traceably Calibrated to NIST weights
N.LS.T#532662-1/049-509
Calibration Results
Test Gas Oz CHa Recommendation
‘Measuring Range 78.0 vol%_| 60.0 %LEL Passed
Alarm Point 48.0 vol% | 23.0 %LEL eee
Indication Value before Passed
collbreton 18.1 vol% | 49.6 %LEL
Indication Value after Passed
Catoration 18.0 vol% | 50.0 %LEL
Note(s
1. “Passed” means that the equipment has been successfully calibrated to meet the specification of the
equipment under test, as set by the equipment manufacturer.
2. Recommended next calibration date will be on 28 December 2023.
Serviced by
Seasafe
Portiand Sattar Tower (6% floor)
176 Strand Road, Agrabad C/A
Chittagong - 4100, Bangladesh
Tel: +9638011002
Email: [email protected]seasafe
CERTIFICATE OF CALIBRATION FOR PORTABLE GAS DETECTOR
Certificate No.: 2212/29-002 (C)
Name of Vessel/ Company MIURA; (IMO NO.: 8910902)
Unit Under Test Portable Multi Gas Detector
Date of Tes 28 December 2022
Particulars of Unit under Test
Description Muli Gas detector
ManufactureriNark RIKEN KEIKI Co,, Lid
Modeltype No. GX-2009)
Identty/ SL No. 943110714RN
- Oto 100% LEL, 016 40% VOL Os, |
Rangefresciutoty (0-100 ppm VOL HsS, 0-500ppm GO
rHiculars of Calibration Apparatus Used
Test instument PORTAGAS Calibration gas oyinder with regular flow indicator
'SLILOT No. 968250
‘Composition CHA: 60% LEL; Ov: 18%, HaS_26ppm, and CO; 100ppm
“The calibration gas is traceably calibrated to NIST weights
Traceability NisaneaseseTONoSOS
Calibration Results
Test Gas, ‘02 Hai Recommendation.
Measuring Range 78.0 volt 25 pom_| 100 pom Passed
‘Alarm Point 48.0vol% | 23.0%LEL | 10 ppm | 50 ppm Passed
Fi
See een 48.4 vol% | 49.6 %LEL | 25 ppm | 99.5 ppm eee
Indication Value after Passed
Calibration 18.0.vol% | 50.0%LEL | 25 ppm | 100 ppm
Note(s):
1. “Passed” means that the equipment has been successfully calibrated to meet the specification of the
‘equipment under test, as set by the equipment manufacturer.
2, Recommended next calibration date will be on 28 December 2023.
Calibrated By |: Misbah Mohammed
Signature with | :
Company Seal *)
‘Serviced by
Seasafe
Portland Sattar Tower (6" floor)
1776 Strand Road, Agrabad C/A
Chittagong - 4100, Bangladesh
Tel: +963801 1002
Email: [email protected]seasafe
CERTIFICATE OF CALIBRATION FOR PORTABLE GAS DETECTOR
Certificate No.: 2212/20-003 (C)
Name of Vessel’ Company MIURA; (IMO NO.: 8910902)
Unit Under Test: Portable Multi Gas Detector
Date of Test: 28 December 2022
Particulars of Unit under Test
Description ‘Multi Gas detector
Manufacturer/Mark RIKEN KEIKI Co., Ltd
Modeltype No. ‘GX-2008
Identity/ SL No. (943110713RN
0 to 100% LEL, 0 to 40% VOL Oz, |
neces 0-100 ppm VOL HS, 0-500ppm CO
Particulars of Calibration Apparatus Used
Test instrument PORTAGAS Calibration gas cylinder with regular flow indicator
'SL/LOT No. ‘986250
‘Composition ‘CHG : 50% LEL; O2: 18%; HaS : 25ppm, and CO: 100ppm
“The calibration gas is traceably calibrated to NIST weights
Traceability N1.8.7#532662-1/049-509
Calibration Results
Test Gas Oz CHA ‘CO__| Recommendation
Measuring Range 70 vol% | 6010 %LEL 700 pom | Passed
Alarm Point 18.0 vol% | 23.0 %LEL 50 ppm —
indication Value before eased
calibration 18.4 vol% | 49.6 %LEL 99.5 ppm
Indication Value after 500 %LEL Pena
18.0 vol% 100 ppm
Calibration
Note(s):
4. “Passed” means that the equipment has been successfully calibrated to meet the specification of the
‘equipment under test, as sot by the equipment manufacturer.
2. Recommended next calibration date will be on 28 December 2023.
Calibrated By
Signature with
Company Seal
Serviced by
Seasafe
Portland Sattar Tower (8" floor)
1776 Strand Road, Agrabad C/A
Chittagong - 4100, Bangladesh
Tel: +9638011002
Email: [email protected]seasafe
CERTIFICATE OF CALIBRATION FOR PORTABLE GAS DETECTOR
Certificate No.: 2212/28-004 (C)
Name of Vessel/ Company MIURA; (IMO NO.: 8910902)
Unit Under Test Portable Multi Gas Detector
Date of Test 29 December 2022
Particulars of Unit under Test
Description Multi Gas detector
Manutacturer/Mark RIKEN KEIKI Co., Ltd.
Modeltype No. (Gx-2008,
Identity/ SL_No. '9X4051559RN
: Oto 100% LEL, Oto 40% VOL On, 1
ee 0-100 ppm VOI. HsS, 0-600ppm CO
Particulars of Calibration Apparatus Used
Test instrument PORTAGAS Calibration gas cylinder with regular flow indicator
SL/JLOT No. 988250
‘Composition CHA: 50% LEL; Os: 18%, HS 26ppm, and CO; 100pem
os “The calibration gas is traceably calibrated to NIST weights
races, N.1.$.T#532662-1/049-509
Calibration Results
Test Gas 02 CH4
Measuring Range 780 vo% | 50.0 %LEL
Alarm Point 18.0 vol% | 23.0 %LEL
indication Value before
‘colkeation 18.1 vol% | 49.6 %LEL
Indication Value after
Calibration
Note(s):
1. “Passed” means that the equipment has been successfully calibrated to meet the specification of the
‘equipment under test, as set by the equipment manufacturer.
2. Recommended next calibration date will be on 28 December 2023.
18.0 vol% | 50.0 %LEL
Calibrated By |: Misbah Mohammed
Signature with
Company Seal
‘Serviced by
Seasafe
Portland Sattar Tower (8* floor)
1776 Strand Road, Agrabad C/A
Chittagong - 4100, Bangladesh
Tel: +9638011002
Email: [email protected]Certificate No.:
Name of Vessel/ Company
Unit Under Test:
Date of Test
Particulars of Unit under Test.
seasafe
CERTIFICATE OF CALIBRATION FOR PORTABLE GAS DETECTOR
2242/29-005 (C),
MIURA; (IMO NO.: 8910902)
Portable Multi Gas Detector
29 December 2022
Description
‘Multi Gas detector
Manufacturer/Mark
RIKEN KEIKI Co., Ltd.
Modelitype No,
‘GX-8000
Identity/ SL No.
499025055.
Range/resolution:
0 to 100% LEL, 0 to 40% VOL Oz, 1
0-100 ppm VOL H2S, 0-500ppm CO
Particulars of Calibration Apparatus Used
Test instrument
PORTAGAS Calibration gas cylinder with regular flow indicator
‘SL/LOT No.
(988250
‘Composition
‘CHA: 50% LEL; Os: 18%; HaS : 25ppm, and CO: 100ppm
Traceability
The calibration gas is traceably calibrated to NIST weights
N.L.S.7#532662-1/049-508
Calibration Results
Test Gas.
ee
‘Oz CHa ‘CO__| Recommendation
‘Measuring Range
78.0 vol% | 60.0 %LEL Passed
Alarm Point
18.0 vol% | 23.0 %LEL Passed
Indication Value before
calibration
Passed
18.1 vol% | 49.6 %LEL
Indication Value after
Passed
18.0 vol% | 60.0 %LEL
Calibration
Note(s):
1, “Passed” means that the equi
‘equipment under test, as set by the equipment manufacturer.
Recommended next calibration date will be on 28 December 2023.
ipment has been successfully calibrated to meet the specification of the
Calibrated By
Signature with
Company Seal
‘Serviced by
Seasafo
Portland Sattar Tower (6* floor)
1776 Strand Road, Agrabad C/A
Chittagong — 4100, Bangladesh
Tel: +9638011002
Email: [email protected]seasafe
CERTIFICATE OF CALIBRATION FOR PORTABLE GAS DETECTOR
Certificate No.: 2242/29-006 (C)
Name of Vessel/ Company MIURA; (IMO NO.: 8910902)
Unit Under Test: Portable Multi Gas Detector
Date of Te: 28 December 2022
Particulars of Unit under Test
Description Muli Gas detector
ManufacturerIMark RIKEN KEIKI Co., Lid.
Modeltype No. ‘GX-8000,
Identity’ SL No, 489026056
‘Oto 100% LEL, 0t6 40% VOL On, |
enaeliescceers 0-100 ppm VOL. HaS, 0-500ppm CO
Particulars of Calibration Apparatus Used
Test instrument PORTAGAS Calibration gas oyinder witt regular flow indicator
SL/LOT No. ‘988250
Composition CHA : 50% LEL, Or: 1896, HS: 26ppm, and CO 10pm
Traceabity “The calibration gas is traceably calibrated to NIST weights
NAL8.7#532662~1/049-609
Calibration Results
Tost Gas, 02 CHa
‘Measuring Range 78.0 vol% | 50.0 %LEL
Alarm Point 18.0 vol% | 23.0 *LEL
Indication Value before
caltration 18.1 vol% | 49.6 %LEL
Indication Value after
Calibration
Note(s):
18.0 vol% | 50.0 %LEL
4. “Passed” means that the equipment has been successfully calibrated to meet the specification of the
‘equipment under test, as set by the equipment manufacturer.
2. Recommended next calibration date will be on 28 December 2023.
Calibrated By + Misbah Mohammed
Pee
Signature with
Company Seal
‘Serviced by
Seasafe
Portland Sattar Tower (6" floor)
1776 Strand Road, Agrabad C/A
Chittagong - 4100, Bangladesh
Tel: +9638011002
Email: [email protected]