seasafe CERTIFICATE OF CALIBRATION FOR PORTABLE GAS DETECTOR Gertincate No.: 2212/28-001 () Name of VesseV/ Company MIURA; (IMO NO.: 8910902) Unit Under Test: Portable Multi Gas Detector Date of Test 28 December 2022 Particulars of Unit under Test Description ‘Multi Gas detector ‘Manufacturer/Mark RIKEN KEIKI Co., Ltd. Model/type No. GX-2009 ‘Identity/ SL No. 943110711RN 0 to 100% LEL, 0 to 40% VOL O2, | Rangefreeokition: 0-100 ppm VOL HS, 0-500ppm CO Particulars of Calibration Apparatus Used Test instrument PORTAGAS Calibration gas cylinder with regular flow indicator ‘SLJ/LOT No. 988250 Composition ‘CH4 = 50% LEL; O2: 18%; HaS : 25ppm, and CO : 100ppm_ Treconbiy “The calibration gas is traceably Calibrated to NIST weights N.LS.T#532662-1/049-509 Calibration Results Test Gas Oz CHa Recommendation ‘Measuring Range 78.0 vol%_| 60.0 %LEL Passed Alarm Point 48.0 vol% | 23.0 %LEL eee Indication Value before Passed collbreton 18.1 vol% | 49.6 %LEL Indication Value after Passed Catoration 18.0 vol% | 50.0 %LEL Note(s 1. “Passed” means that the equipment has been successfully calibrated to meet the specification of the equipment under test, as set by the equipment manufacturer. 2. Recommended next calibration date will be on 28 December 2023. Serviced by Seasafe Portiand Sattar Tower (6% floor) 176 Strand Road, Agrabad C/A Chittagong - 4100, Bangladesh Tel: +9638011002 Email: [email protected] seasafe CERTIFICATE OF CALIBRATION FOR PORTABLE GAS DETECTOR Certificate No.: 2212/29-002 (C) Name of Vessel/ Company MIURA; (IMO NO.: 8910902) Unit Under Test Portable Multi Gas Detector Date of Tes 28 December 2022 Particulars of Unit under Test Description Muli Gas detector ManufactureriNark RIKEN KEIKI Co,, Lid Modeltype No. GX-2009) Identty/ SL No. 943110714RN - Oto 100% LEL, 016 40% VOL Os, | Rangefresciutoty (0-100 ppm VOL HsS, 0-500ppm GO rHiculars of Calibration Apparatus Used Test instument PORTAGAS Calibration gas oyinder with regular flow indicator 'SLILOT No. 968250 ‘Composition CHA: 60% LEL; Ov: 18%, HaS_26ppm, and CO; 100ppm “The calibration gas is traceably calibrated to NIST weights Traceability NisaneaseseTONoSOS Calibration Results Test Gas, ‘02 Hai Recommendation. Measuring Range 78.0 volt 25 pom_| 100 pom Passed ‘Alarm Point 48.0vol% | 23.0%LEL | 10 ppm | 50 ppm Passed Fi See een 48.4 vol% | 49.6 %LEL | 25 ppm | 99.5 ppm eee Indication Value after Passed Calibration 18.0.vol% | 50.0%LEL | 25 ppm | 100 ppm Note(s): 1. “Passed” means that the equipment has been successfully calibrated to meet the specification of the ‘equipment under test, as set by the equipment manufacturer. 2, Recommended next calibration date will be on 28 December 2023. Calibrated By |: Misbah Mohammed Signature with | : Company Seal *) ‘Serviced by Seasafe Portland Sattar Tower (6" floor) 1776 Strand Road, Agrabad C/A Chittagong - 4100, Bangladesh Tel: +963801 1002 Email: [email protected] seasafe CERTIFICATE OF CALIBRATION FOR PORTABLE GAS DETECTOR Certificate No.: 2212/20-003 (C) Name of Vessel’ Company MIURA; (IMO NO.: 8910902) Unit Under Test: Portable Multi Gas Detector Date of Test: 28 December 2022 Particulars of Unit under Test Description ‘Multi Gas detector Manufacturer/Mark RIKEN KEIKI Co., Ltd Modeltype No. ‘GX-2008 Identity/ SL No. (943110713RN 0 to 100% LEL, 0 to 40% VOL Oz, | neces 0-100 ppm VOL HS, 0-500ppm CO Particulars of Calibration Apparatus Used Test instrument PORTAGAS Calibration gas cylinder with regular flow indicator 'SL/LOT No. ‘986250 ‘Composition ‘CHG : 50% LEL; O2: 18%; HaS : 25ppm, and CO: 100ppm “The calibration gas is traceably calibrated to NIST weights Traceability N1.8.7#532662-1/049-509 Calibration Results Test Gas Oz CHA ‘CO__| Recommendation Measuring Range 70 vol% | 6010 %LEL 700 pom | Passed Alarm Point 18.0 vol% | 23.0 %LEL 50 ppm — indication Value before eased calibration 18.4 vol% | 49.6 %LEL 99.5 ppm Indication Value after 500 %LEL Pena 18.0 vol% 100 ppm Calibration Note(s): 4. “Passed” means that the equipment has been successfully calibrated to meet the specification of the ‘equipment under test, as sot by the equipment manufacturer. 2. Recommended next calibration date will be on 28 December 2023. Calibrated By Signature with Company Seal Serviced by Seasafe Portland Sattar Tower (8" floor) 1776 Strand Road, Agrabad C/A Chittagong - 4100, Bangladesh Tel: +9638011002 Email: [email protected] seasafe CERTIFICATE OF CALIBRATION FOR PORTABLE GAS DETECTOR Certificate No.: 2212/28-004 (C) Name of Vessel/ Company MIURA; (IMO NO.: 8910902) Unit Under Test Portable Multi Gas Detector Date of Test 29 December 2022 Particulars of Unit under Test Description Multi Gas detector Manutacturer/Mark RIKEN KEIKI Co., Ltd. Modeltype No. (Gx-2008, Identity/ SL_No. '9X4051559RN : Oto 100% LEL, Oto 40% VOL On, 1 ee 0-100 ppm VOI. HsS, 0-600ppm CO Particulars of Calibration Apparatus Used Test instrument PORTAGAS Calibration gas cylinder with regular flow indicator SL/JLOT No. 988250 ‘Composition CHA: 50% LEL; Os: 18%, HS 26ppm, and CO; 100pem os “The calibration gas is traceably calibrated to NIST weights races, N.1.$.T#532662-1/049-509 Calibration Results Test Gas 02 CH4 Measuring Range 780 vo% | 50.0 %LEL Alarm Point 18.0 vol% | 23.0 %LEL indication Value before ‘colkeation 18.1 vol% | 49.6 %LEL Indication Value after Calibration Note(s): 1. “Passed” means that the equipment has been successfully calibrated to meet the specification of the ‘equipment under test, as set by the equipment manufacturer. 2. Recommended next calibration date will be on 28 December 2023. 18.0 vol% | 50.0 %LEL Calibrated By |: Misbah Mohammed Signature with Company Seal ‘Serviced by Seasafe Portland Sattar Tower (8* floor) 1776 Strand Road, Agrabad C/A Chittagong - 4100, Bangladesh Tel: +9638011002 Email: [email protected] Certificate No.: Name of Vessel/ Company Unit Under Test: Date of Test Particulars of Unit under Test. seasafe CERTIFICATE OF CALIBRATION FOR PORTABLE GAS DETECTOR 2242/29-005 (C), MIURA; (IMO NO.: 8910902) Portable Multi Gas Detector 29 December 2022 Description ‘Multi Gas detector Manufacturer/Mark RIKEN KEIKI Co., Ltd. Modelitype No, ‘GX-8000 Identity/ SL No. 499025055. Range/resolution: 0 to 100% LEL, 0 to 40% VOL Oz, 1 0-100 ppm VOL H2S, 0-500ppm CO Particulars of Calibration Apparatus Used Test instrument PORTAGAS Calibration gas cylinder with regular flow indicator ‘SL/LOT No. (988250 ‘Composition ‘CHA: 50% LEL; Os: 18%; HaS : 25ppm, and CO: 100ppm Traceability The calibration gas is traceably calibrated to NIST weights N.L.S.7#532662-1/049-508 Calibration Results Test Gas. ee ‘Oz CHa ‘CO__| Recommendation ‘Measuring Range 78.0 vol% | 60.0 %LEL Passed Alarm Point 18.0 vol% | 23.0 %LEL Passed Indication Value before calibration Passed 18.1 vol% | 49.6 %LEL Indication Value after Passed 18.0 vol% | 60.0 %LEL Calibration Note(s): 1, “Passed” means that the equi ‘equipment under test, as set by the equipment manufacturer. Recommended next calibration date will be on 28 December 2023. ipment has been successfully calibrated to meet the specification of the Calibrated By Signature with Company Seal ‘Serviced by Seasafo Portland Sattar Tower (6* floor) 1776 Strand Road, Agrabad C/A Chittagong — 4100, Bangladesh Tel: +9638011002 Email: [email protected] seasafe CERTIFICATE OF CALIBRATION FOR PORTABLE GAS DETECTOR Certificate No.: 2242/29-006 (C) Name of Vessel/ Company MIURA; (IMO NO.: 8910902) Unit Under Test: Portable Multi Gas Detector Date of Te: 28 December 2022 Particulars of Unit under Test Description Muli Gas detector ManufacturerIMark RIKEN KEIKI Co., Lid. Modeltype No. ‘GX-8000, Identity’ SL No, 489026056 ‘Oto 100% LEL, 0t6 40% VOL On, | enaeliescceers 0-100 ppm VOL. HaS, 0-500ppm CO Particulars of Calibration Apparatus Used Test instrument PORTAGAS Calibration gas oyinder witt regular flow indicator SL/LOT No. ‘988250 Composition CHA : 50% LEL, Or: 1896, HS: 26ppm, and CO 10pm Traceabity “The calibration gas is traceably calibrated to NIST weights NAL8.7#532662~1/049-609 Calibration Results Tost Gas, 02 CHa ‘Measuring Range 78.0 vol% | 50.0 %LEL Alarm Point 18.0 vol% | 23.0 *LEL Indication Value before caltration 18.1 vol% | 49.6 %LEL Indication Value after Calibration Note(s): 18.0 vol% | 50.0 %LEL 4. “Passed” means that the equipment has been successfully calibrated to meet the specification of the ‘equipment under test, as set by the equipment manufacturer. 2. Recommended next calibration date will be on 28 December 2023. Calibrated By + Misbah Mohammed Pee Signature with Company Seal ‘Serviced by Seasafe Portland Sattar Tower (6" floor) 1776 Strand Road, Agrabad C/A Chittagong - 4100, Bangladesh Tel: +9638011002 Email: [email protected]

You might also like