M1242e Multisampler 3861
M1242e Multisampler 3861
Manual
Nothing in this publication may be reproduced and/or made public by means of print, photocopy, microfilm or any other means without
previous written permission from Eijkelkamp Soil & Water. Technical data can be amended without prior notification.
Eijkelkamp Soil & Water is not responsible for (personal damage due to (improper) use of the product.
Eijkelkamp Soil & Water is interested in your reactions and remarks about its products and operating instructions.
2
On these operating instructions
1. Short description
The Multisampler has been developed to take samples from
layered strata of sludge and sediment as well as from layered
liquids.
The stratification of the sample taken remains reasonably intact.
In liquids column- as well as point samples can be taken.The
apparatus can be deployed up to a depth of several meters by
applying extension rods with bayonet connection (1).
The standard model of the Multisampler is supplied with two
cutting heads.
One open sharp cutting head (2) for taking column samples in
more or less tough material “without” agitation, and a cutting
head with ball valve seating (4) and ball (3) to be able to take
semi-agitated samples from fluid material. The large passage of
the ball valve enables sampling of extremely viscous- and/or
heterogeneous material.
The metal parts of the Multisampler have been made of stainless steel. The sampling tube is made of PMMA
acrylic high transparent plastic. A NBR-rubber piston (6) is applied. Under the condition that all precautions
(cleaning, controlling) have been taken, the samples taken can be examined for all kinds of pollution.
For taking layered samples from unknown chemicals we refer to the rod- or cable operated liquid layer samplers
(art. no.: 1240 and 1241). These have been fitted with a chemically inert teflon sampling tube.
3
2. Introduction and decontamination
Every Multisampler leaving our plant is subjected to a final quality inspection. If nevertheless you would find
any imperfections in the apparatus, then we request you to contact us.
On reception you should check whether all required and ordered parts are present and whether they are
undamaged. The minimum you need would be:
1. The Multisampler (124201B) with two cutting heads (one of which has been mounted).
2. A handle to be able to operate the Multisampler (011001B or 011004B).
3. Possible one or more extension rods (011007B).
4. A cord to operate the piston rod (01090002).
In combination with tissues, isopropyl alcohol can be used to quickly clean the apparatus in the field.
Volatile substances that entered the plastic through permeation will disappear from the apparatus
by placing it in a ventilated warm place after carefully cleaning inside- and outside. In extreme cases
the sampling tube can not be used again.
Note: In the event of affect (by certain chemicals) of the PMMA or corrosion of the stainless steel, iron, nickel
and chrome (from the stainless steel) and tin (from the PMMA) may be released. In such an event this should
be taken into account on interpreting the results of the analysis.
4
3. Application.
3.1 Taking sludge- and sediment samples
Principle of operation:
The apparatus is pushed into the sludge. To avoid compaction of the layers a piston is held at a stationary
position relative to the sludge.
Working procedure:
Assemble the Multisampler, fitted with the open cutting head (2) with centring pens (5) to avoid slipping off).
See to it that the wires (8) that are holding the cutting ring are not pulled too tight. The adjustable hook (9)
that pulls the wires can be unscrewed or screwed in. Mount no more than two extension rods and the handle
and fit the piston rod (11) with a cord (you could use a plasticized cord).
Lower the apparatus until any resistance is felt (in case of application at greater depths additional extension
rods are mounted during the lowering of the apparatus).
After feeling the sludge to be cut, the apparatus is lifted several decimeters.
Then the piston is blocked at a stationary level, for instance by pulling the cord tight and by standing on it.
Next the sampler is pushed into the sludge until it is filled-up completely or until too much resistance is felt
(in case of the application of several extension rods it is important to cut fast and with short brisk movements
to avoid bending).
After this the apparatus is pulled-up by the piston cord. Through an air valve in the top section of the
Multisampler, air can leave, but not enter, the space over the piston which causes the piston to remain in place
and the sample is kept inside by the slight underpressure.
- Often vertical lines can be seen in addition to the horizontal lines caused by the different sediment layers
in the sample. These lines are the result of the cutting of small pieces of sludge with a deviant color by the
cutting edge of the cutting head. It is a typical wall-effect and is called smearing.
- In case of a well cut sample the top section of the sample should be very watery (clear water). If this is not
the case then, the next time you take a sample, you should pull the apparatus further up after feeling the
bottom before you commence cutting. It is also possible to apply an electronic sludge layer detector to
determine the transition between water and sludge.
- Water on top of the piston has no significance. It leaked inside by passing alongside the piston rod.
5
To be able to execute procedure 1 the following is done:
r Examine or photograph the sample as it can be seen through the transparent sampling tube. If this method
of visual examination is sufficient the apparatus can be placed in the wide-mouth flask and emptied. To
maintain the anoxic character of the sample the flask will be filled to the rim. The rim should be cleaned
properly in order to allow proper sealing.
Examine the sample after it has been removed from the sampling tube. This can be executed as follows:
- Place a half-round plastic gutter (art. no.: 990812) with a length of at least 1 meter, or a sheet of plastic
on the ground.
- Hold the Multisampler almost horizontal and screw handle (10) on the piston rod (11). Empty the sampler
by keeping the piston rod stationary while pulling the sampler. In this way a ‘roll’ is created in which
the layer thickness are maintained.
- The circumference of the sample has been disturbed most due to the wall-affects. For this reason the
sample should be examined by opening it using a spatula. This avoids incorrect interpretation.
- The less the binding of the sludge, the easier it falls apart.
- Sub-samples can be taken using a spatula or a spoon.
r Immediately press the cap firmly in the lower section of the transparent tube. If necessary the sample can
be pulled up slightly by means of the piston rod.
r Unscrew the piston rod from the piston and remove the rod.
r Transport the sampling tube with the sample in a (semi)vertical position (not horizontal). This avoids mixing.
r A new sample can be taken after mounting another tube and piston.
Principle of operation:
1. Column sampling:
The apparatus is pressed into the fluid. In order to maintain the layered character of the liquids the piston
is kept at a stationary position in relation to the fluid. The stainless steel ball valve in the bottom section
prevents loss of the sample taken.
2. Spot sampling:
The apparatus is lowered to the proper depth. By pulling up the piston completely the apparatus is filled
with liquid from that particular depth (for instance the bottom of a reservoir).
3. Mix-sampling:
As under 2, but now the piston is raised a little (for instance 10 centimeters = 125 ml) at every sampling spot.
In this way a mix-sample of ten spots, divided over the depth or over the surface, can be obtained in one
action.
Application:
Assemble the Multisampler, fitted with an open cutting head. See to it that the wire is not pulled too tight. The
hook that pulls the wires can be unscrewed or screwed in. Mount no more than two extension rods and the
top piece and fit the piston rod with a cord (you could use a plasticized cord).
r Lower the apparatus until the right depth is reached (in case of application at greater depths the extension
rods are mounted during the lowering of the apparatus).
6
If a column sample has to be taken:
r Then the piston is blocked at a stationary level, for instance by pulling the cord tight and by standing on
it.
r Next the sampler is pushed down until it is filled-up completely or you reach the bottom of the reservoir.
In case a spot- or mix-sample has to be taken, action is taken as previously described. However, now the
apparatus is kept stationary and the piston is pulled up.
r Examine or photograph the sample as it can be seen through the transparent sampling tube. If this method
of visual examination is sufficient the apparatus can be placed in a wide-mouth flask and emptied. Hold
the Multisampler almost horizontal so that the ball comes out of the seat of the ball valve. The contents
of the tube will run into the wide-mouth flask.
It is also possible to transport the fluid sample in the sampling tube. To this purpose you should remove the
piston rod. Turn the Multisampler upside down. Disconnect the tension rope and remove the ball valve. The
ball will remain inside the sampling tube; it cannot be removed. Mount the cutting head on the tube. The
sample is now ready for transportation.
7
8