0270501299WA-'-25℃ ‘-30℃ '-40℃ Biomedical Freezer低温保存箱英文说明书 - 20181214 - 全部变更
0270501299WA-'-25℃ ‘-30℃ '-40℃ Biomedical Freezer低温保存箱英文说明书 - 20181214 - 全部变更
Operation Manual
Certificate of Quality
Model :
DW-25L92
checker:
DW-25L262
DW-25L300
DW-30L278
DW-30L420F
DW-30L508
DW-40L92
DW-40L262
Manufacturer: Qingdao Haier Biomedical Co., Ltd. DW-40L278
Address: Haier Industrial Park, Economic Technology Development DW-40L278J
Zone, Qingdao P.R. China DW-40L278(a)
DW-40L348
Web:www.haiermedical.com
DW-40L348J
Revision Date:12/2018 DW-40L508
Version:3rd,2018 DW-40L508J
Dedicated code:0270501299WA DW-40L508(a)
V13026 DW-40L420F
DW-40W100
DW-40W255
DW-40W380
57
Product Feature
Packing List
56 1
Content
Specification
Name Deep Freezer
Model DW-40W100 DW-40W255 DW-40W380
Exterior dimensions
685×650×810 1243×724×838 1554×724×838
(W × D × H) (mm)
Interior dimensions
500×385×630 1036×426×625 1375×453×640
(W × D × H) (mm)
Effective volume 100L 255L 380L
Product Feature ................................................................. 1 Door Solid door
Name of parts • Control panel .......................................... 13 Condenser/Evaporator Wire and tube type / coil tube type
Temperature control Microprocessor controlled
Application method ........................................................... 19
Refrigerant R404a 110g R404a 100g R404a 230g Mix refrigerant 162g
Alarm ................................................................................ 33 Weight 43kg 70kg 82kg
Cleaning maintenance ...................................................... 37 Temperature range -20℃ ~-40℃ -20℃ ~-40℃ -20℃ ~-40℃
Test holes 1 /
Alarms High/low temperature alarm, sensor error alarm and power failure alarm
Duration time of
18 hours (when battery is fully charged)
power outage alarm
Note: Climate Type 4 means the temperature is +30℃ ,the relative humidity is 55%.Technical
information may vary from the appliance that you have just purchased due to technical improvements.
2 55
Safety Precautions
Dear Customers,
Specification Thank you for choosing Haier deep freezer. Please make sure that you have carefully read the
Name Deep Freezer manual and follow the instruction as outlined. This is to prevent personal injury and damage to
products.
Model DW-30L420F/DW-40L420F
Exterior dimensions(W × D × H) (mm) 950×890×1900 Safety labels
4 53
Specification Disposal of discarded freezer shall be conducted by the relevant personnel. Dismantle the door
to prevent occurrence of accidents such as suffocation.
Name Deep Freezer Do not set off hazardous articles that are inflammable, explosive or volatilizable in the freezer,
Model DW-40L278/278J DW-40L278(a) DW-40L348/348J DW-40L508/508J DW-40L508(a) nor use the flammable spraying agent in the vicinity of the freezer, or otherwise, it may trigger
Exterior
an explosion or fire.
dimensions 770x710x1810 790x845x1860 940x845x1860 Do not store corrosive articles such as acid, alkali etc. in the freezer. Otherwise, it may spoil
(W × D × H) (mm) the internal units or electric parts of the freezer.
Interior dimensions
520×435×1230 535×610×1228 685×610×1228
(W × D × H) (mm) Do not place the packing plastic bags at the locations where children can easily take as plastic
Effective volume 278L 348L 490L bags may cause occurrence of suffocation accident.
Door Solid door
Do not climb up to the freezer or place articles on the freezer, otherwise it may cause personal
injury or damage of freezer due to overturn of the freezer.
Insulation CFC-Free foamed-in-place urethane
Do not insert any metal objects like iron nails, iron-wire into any orifice and gap or any air vent
Compressor High Quality Hermetically sealed compressor
used for internal air circulation of the freezer, otherwise, it may cause an electric shock or injury
Drawers 6 12 10 due to contact of the above objects with the moving parts.
Refrigeration Direct cooling
52 5
Application guidelines
During the normal process of use, please read the following precautions seriously before use, in order
to keep the low-temperature freezer in the best performance: Specification
• About the battery
Name Deep Freezer
Some types of this series of low-temperature freezer is equipped with built-in rechargeable batteries,
which could send alarm message about outage to you immediately when the power is off, so as to Model DW-25L262 DW-25L300 DW-40L262
protect your storage to the largest extent. In addition, you should understand and pay attention to
points as follows about the batteries: Exterior dimensions
700×705×1665
a.Type DW-25L92/262/300 and DW-40L92/262 are not equipped with rechargeable batteries, so they (W × D × H) (mm)
could not send any alarm message for outage; Interior dimensions
480×465×1430
b.The controllers of DW-30L278/508 and DW-40W100/255/380 are respectively equipped with a (W × D × H) (mm)
rechargeable battery, which is irreplaceable and could not be controlled by any switch. Under normal Effective volume 262L 300L 262L
condition, such battery could be fully recharged within about 24 hours, which could under normal
work for outage alarm and display for 18 hours. Therefore, if the power of this series of product is off Door Solid door
for more than 24 hours, the battery would be no power, and then the function of alarming for outage Insulation CFC-Free foamed-in-place urethane
would become invalid;
c. Type DW-40L278/278J/278(a)348/348J/508/508J/508(a)/420F/30L420F is respectively equipped Compressor High Quality Hermetically sealed compressor
with a replaceable rechargeable battery, which is located within the electric cabinet. In addition, a
Drawers 7
battery switch is equipped in it, too, so that when the power of product is scheduled to be off or would
be not used for a long time, you should unplug the power cable and turn the switch to "O" timely, in Refrigeration Direct cooling
order to prolong the service life; when it is planned to use the product under normal condition, you
Exterior / Interior Spraying steel plate / PS plate
should turn the switch to "I" in order to keep the alarm for outage normal.
• Storage and withdrawal of stored articles Condenser/Evaporator Wire and tube type
a. Before putting any article in the freezer, you should confirm that whether the storage temperature
Temperature control Microprocessor controlled
of article complies with the setting temperature of the freezer in advance;
b. Before putting any article in the freezer, you should confirm that whether the temperature within the Refrigerant R600a 90g R600a 90g R404a 220g R404a 200g
freezer has already reached to the setting temperature in advance;
Weight 88kg 88kg 88kg
c. When putting in the articles, you should put them successively in batches, and the volume of article
to be put each time could not be larger than 1/3 of volume of the freezer, in order not to make the Temperature range -10℃ ~-25℃ -10℃ ~-25℃ -20℃ ~-40℃
refrigerating system overloaded;
Voltage 220-240V~/50Hz 220-240V~/50Hz 220-240V~/50/60Hz 115V~/60Hz
d. When putting or taking the articles each time, please minimize the time for door closing, in order to
reduce the energy consumption and keep the temperature within the freezer relatively stable; Rated power/current 90W/0.7A 90W/0.7A 400W/4.8A 600W/7.5A
• Lock latch of vertical-type freezer
Noise Level 34dB(A) 34dB(A) 44dB(A)
The lock latch of vertical-type freezer should be fastened after the freezer is electrified for use,
otherwise the temperature would become higher and frosting would become serious, and the normal Foaming Agent CP/IP
use would be affected.
Climate Type 4
• Test hole of freezer
The test hole equipped on the wall of freezer is a through-hole, which is used to lead out the Type of protection against
I
measuring line within the freezer during test. After leading out the measuring line, please stuff the electric shock
test hole with heat insulating materials again, otherwise the temperature within the freezer would stay Test holes 1
high, and dew or frost would be condensed surrounding to the outside of through-hole.
Alarms High/low temperature alarm, sensor error alarm
Meaning of crossed –out wheeled dustbin: Note: under the ambient temperature of 25°C, the minimum temperature inside the empty equipment
of the product of DW-40L series could achieve -40°C; and under the ambient temperature of 32°C,
Do not dispose of electrical appliances as unsorted municipal waste, use separate collection facilities. the minimum temperature inside the empty equipment could achieve -35°C.
Contact you local government for information regarding the collection systems available.If electrical
Climate Type 4 means the temperature is +30°C,the relative humidity is 55%.
appliances are disposed of in landfills or dumps, hazardous substances can leak into the groundwater
Technical information may vary from the appliance that you have just purchased due to technical
and get into the food chain, damaging your health and well-being. improvements.
6 51
Specification Product installation
Door Solid door • As the freezer is sensitive to the environmental temperature. The freezer cannot
operate normally if it is installed in an environment other than the above. Please
Insulation CFC-Free foamed-in-place urethane use it after improving the environment.
Compressor High Quality Hermetically sealed compressor • It is prohibited to install the freezer in the open air environment. When the freezer
gets wet in the rainwater, it may cause leakage of electricity or electric shock
Drawers 3 accident.
Refrigeration Direct cooling
Installation place
Exterior / Interior Spraying steel plate / PS plate
For normal operation and optimum performance level of the freezer, its installation place shall
Condenser/Evaporator Wire and tube type conform to the following conditions:
Temperature control Microprocessor controlled • It can not be installed in a narrow, small and airtight space, the door frame of the room for placing
it shall not be less than and lower than this product, and shall at least guarantee its normal passing
Refrigerant R600a 52g R404a 150g in and out so as to avoid any maintenance difficulty and failure of in-time repair in occurrence of
Weight 46kg 46kg machine fault, causing spoilage to the stored articles.
• The ground at the installation place shall be solid and flat.
Temperature range -10℃ ~-25℃ -20℃ ~-40℃
• It shall be well-ventilated without direct sunshine.
Voltage 220-240V~/50Hz 220-240V~/50Hz • Do not share a socket with other equipment, the plug and socket shall be well connected.
Rated power/current 77W/0.35A 220W/2.1A • Do not distort or stress the power cord.
• If it is required to lengthen the power cord, the section area of the lengthened wire shall not be less
Noise Level 27dB(A) 40dB(A)
than 3mm2 and the length shall not be longer than 3m.
Foaming Agent CP/IP • Check the operating voltage before use, at the area where the voltage is not stable, use of a voltage
regulator suitable for the motor load shall be considered for voltage stabilization. So that the normal
Climate Type 4
input voltage is stabilized at the rated voltage ±10%, the power of the voltage regulator is greater than
Type of protection against electric shock I 4000W.
Test holes 1 • The freezer shall be guaranteed with reliable earthing.
• If the power socket has earth wire, check if the earthing is good before use.
Alarms High/low temperature alarm, sensor error alarm
• If the socket is not furnished with an earth wire, be sure to install one by the professional
Note: Climate Type 4 means the temperature is +30℃ ,the relative humidity is 55%. engineering and technical personnel.
Technical information may vary from the appliance that you have just purchased due to technical
• Do not make freezer earthing via gas piping, water supply pipe, telephone line or
improvements.
lightning rod that may easily cause an electric shock.
• After installation, the power plug must be accessible to facilitate unplugging of the
power cord in emergency. No articles shall block out the air vent of the freezer.
50 7
Preparation before Use for vertical type and DW-30L278/508)
1. Disassemble all packaging materials, including the base
Take out all packaging materials of product at suitable site, including the base, and dissemble all fixed
tapes or other fixtures on the freezer used for transportation safety.
8 49
4. Install side lock latch
DW-30/40L420F For Type DW-25L92 and DW-40L92, the attached lock latch kit should be taken out and then installed
on the left side of freezer according to the "Lock Catch Installation Instruction Leaflet" stipulated in the
Appendix.
• DW- 30L278/508
a. Place the side of handle with hole downwards;
1 2
b. Align the handle with its holder;
c. Fix the holder and the handle and then lift up the handle;
d. Fix the handle with screws from the bottom after the handle is
fastened.
3 4
48 9
7. Leveling of the unit
At the bottom of the upright freezer, there is a DW-40L278/278J/278(a)/348/348J/508/508J/508(a)
leveling leg that allows an adjustment of one of the
corners of the unit to secure its placement. Turn the
leveling leg accordingly to adjust the height.
Down Up
Leveling leg
10 47
Circuit Diagram Initial power-on
Please observe the following rules when conducting initial start-up and continuous running:
DW-25L92/262/300 and DW-40L92/262
Keep the unit empty; connect the unit to a dedicated power socket that meets the power
requirement of the freezer.
After power up, turn on the battery switch of the units to charge the battery. The audible
alarm may sound and this is normal. Press the Silence button (for DW- 25L92/262/300
and DW-40L92/262, press alarm button; for horizontal type and DW- 30L278/508, press
any button) to cancel the alarm on.
Power on, and set the temperature required without putting any article in the empty
freezer, and then observe that whether the freezer could be turned on or off under normal
condition for over 24 hours, in order to confirm that the performance of freezer is normal.
Start putting articles gradually into the freezer after completing a thorough inspection of the
freezer.
46 11
• A person shall be specially assigned for daily check and record of the product operating
conditions, in case the temperature inside the bin is on the high side or on the low side, just others are for horizontal type
transfer the articles to the other freezers till the fault is removed and then replace back the
articles.
• Before putting any article in the freezer, please confirm that the temperature required for
the article is consistent with the temperature range of this product in advance, in order not to
damage the article due to the failure of this product to reach to the temperature required for the
article. Capillary tube
• The indicated temperature of this product may have a certain difference with the actual
temperature inside the bins due to refrigeration inertia, this is normal.
• The temperature within the topside drawer of freezer would be relatively higher than that in
other drawers, so, for any article with strict requirements for temperature should be placed in Dryer filter
other drawers as possible.
For Type DW-40, if its temperature is set to be high (such as higher than
-25℃ ), the temperature deviation might become more serious due to the
refrigeration inertia, which would be more serious when the temperature is
Evaporator
set to be higher. Therefore, it is suggested that Type DW-40 should be used
under -30℃ , and for any service temperature that is higher than -25℃ /-
30℃ , please select Type DW-25/30.
• All the freezers are of storage equipment, and it is strictly forbidden to put excessive articles
that are relatively too hot in one go, this will cause long time of running of the compressor
without stop, slow drop of temperature, thus affecting service life of the compressor, the articles
must be put in batches so that the temperature drops in steps till the required temperature is
reached.
• Do not use any mechanic tool or other means without permit from manufacturer to accelerate
the process of defrosting.
• Do not spoil the refrigerating circuit. Compressor
• Inside the freezer, do not use the electric appliance without production permission.
DW-40W380
Name of parts
Dryer filter
Pad lock Control panel
T-branch
Test hole
Compressor
Door handle
Condenser
Solid door
44 13
DW-40L278/278J/278(a) and DW-30L278 DW-30/40L420F
Tpye DW-40L
A
A
Evaporator
Condenser
Compressor
B
Air return pipe
Solenoid valve
Condenser
anti-dew tube
B
Capillary tube
Dryer filter
Evaporator
Compressor
Caster and leveling leg
14 43
Refrigeration Schematic & Circuit Diagram
Auxiliary capillary
Lower door
Compressor
Exhaust pipe
Pad lock
42 15
DW-30/40L420F
Any of the following event will be normal, please do not worry about any of it:
Power switch •For initial use or start-up, the working time of refrigerating system will be longer and the noise will
(on the back) Door ajar alarm switch be more serious, and under the normal condition, you may find out sound like liquid flow, which is
caused by the circular flow of refrigerating fluid within the tube;
Control panel USB port Solid door
•There might be a difference between the temperature displayed on the freezer and that at the center
of freezer, that's because the sensor is not located at the center. However, the displayed temperature
will gradually be closed to that at the center after stable operation of the freezer;
Battery switch •For radiating tubes are equipped in the outer shell of freezer, the surface temperature of shell will be
(on the back) very high when the freezer is being operated.
Inner door
and handle
Door
Handle
(with lock)
Test hole
Storage drawer
(optional)
Temperature recorder
(optional)
16 41
FAQ
If you have any questions about the operation of the unit, please refer DW-40W100/255/380
to the following first. If an issue persists, contact your after sales office.
Do not maintain and dismantle the freezer by yourself!
40 17
Rechargeable battery recycling
The rechargeable battery equipped in the freezer could be recyclable. Therefore, when its service life
Control panel expires, please contact local organization for battery recycling to check or dump it in a correct way.
Rechargeable battery within the freezer is used for outage alarm, which is located in the electric
cabinet.
There are electric appliance components within the electric cabinet.In order to prevent electric
shock, the cover should be opened only by professional engineer or maintenance personnel.
1. Turn off the battery switch and then pull the plug out of the socket;
2. Screw out all screws on the cover of electric cabinet on top of the bin by screwdriver;
3. Pull out the wiring plug of battery;
4. Unfix the fixers of battery and take out the battery;
5. Recycle or correctly dispose of the battery based on relevant rules.
For replacement rechargeable battery, it should guarantee that the red wire should be connected
with the anode of battery while the black wire with the cathode without any error. Otherwise, the
charging circuit of computer board will be easily burnt, and the computer board will not be able to
recharge for the battery.
DW-30/40L420F The red line to connect with the positive battery end
Electronic temperature controller is adopted in horizontal type and Type DW-30L278/508, which
would not marked here.
18 39
Application method
• Temperature adjustment
• When the freezer is under electrified operation, please ensure that the battery switch keeps on, and
After pressing " " and " " for 3 seconds, the setting temperature would begin to flash, and then
the battery will be recharged. After the battery has been continuously recharged for a week, please
press the " " once, the temperature would be reduced by 1℃ until -40℃ /-25℃ , and for one more
test the quantity, for it should be full under normal condition. If the quantity is not full, please replace
press, it will be back to -10℃ ; and then press the " " once, 1℃ would be added until -10℃ , and
the rechargeable battery.
for one more press, it will back to -40℃ /-25℃ . it would exit from the setting status and then display
• The outage alarm battery is consumable, and the service life is 2 or 3 years generally. If any battery the normal temperature if no operation is performed for 5 seconds.
has been used for more than 3 years, the outage alarm function will become invalid. The battery
• Temperature setting of hi-low temperature alarm
should be replaced as soon as possible. For any replacement, please contact the after-sale staff of
Haier. The default of hi-low temperature alarm is ±5℃ of setting temperature, and the steps should be
followed when setting it:
After pressing "Alarm" + " " for 3 seconds, AH (high temperature)would begin to flash, and then
Discarding Disposal of Freezer press the " " once, the temperature would be reduced by 1℃ until 1, and for one more press, it will
be back to 10; and then press the " " once, 1℃ would be added until 10, and for one more press,
it will back to 1; it would exit from the setting status and save the setting, and then display the normal
• If the freezer is left unused for a long time in an unattended area, make sure
temperature if no operation is performed for 5 seconds.
that children will not approach the freezer and the freezer door will not be
closed completely. After pressing "Alarm" + " " for 3 seconds, AL (low temperature) would begin to flash, and then
press the " " once, the temperature would be reduced by 1℃ until 1, and for one more press, it will
• The scrap disposal should be carried out by corresponding staff, and the
be back to 10; and then press the " " once, 1℃ would be added until 10, and for one more press,
door should be demolished in order to prevent any accident such as asphyxia.
it will back to 1; it would exit from the setting status and save the setting, and then display the normal
temperature if no operation is performed for 5 seconds.
Purchasing of Recording Paper and Standby Battery of Recorder
DW-30L278/508
Temperature recorder is an optional device. The chart papers are custom-made. Additional chart
papers can be purchased through Haier.
The 9 V DC battery for the recorder can be acquired anywhere so long it matches the requirement.
Function button FN Up-regulated button
Customers can also buy high quality replacement battery from Haier.
38 19
Cleaning maintenance
The temperature controller will be locked automatically if there is no pressing Freezer cleaning
within 3 minutes. Then it should be unlocked when you need to operate it again.
• Change the setting temperature
• Clean the freezer once a month. Regular cleaning can keep the appearance of the freezer a brand
In general, the temperature will not be set again, if do, please refer to the steps as follows:
new look.
1. Please unlock the controller if the controller is locked;
• Wipe out the dust on shell, chamber and all the accessories of the freezer with a dry cloth. If the
2. Press "SET" once and then current setting temperature would be shown;
freezer is too dirty, remove the dirt with a clean cloth dipped with neutral detergent and wipe out the
3. Increase or decrease the setting value by " " or " ";
residues with a wet cloth. Then wipe with a dry cloth.
4. Press "SET" once again after you change the temperature to your expected value, and the
• Do not pour water on the shell or in the chamber of the freezer. Otherwise, it may damage the
controller will save the current setting and then exit to normal interface.
insulation of the electrical equipment and may lead to the occurrence of fault.
• Default setting temperature
• Mechanical compressors are hermetically sealed. They do not need lubrication.
The default setting temperature of this product is -25℃ .
• Conduct defrosting or deicing for the inner wall once every month.
• Query historical data
The product is equipped with simple function of recording historical data, which can save the
maximum value of temperature Ho1 and the minimum value Lo1 recorded within 24 hours of normal Defrosting for inner wall
operation. For any inquiry, it should be operated by the following steps:
1. Press "SET" for more than 3 seconds, then the digital tube will display the parameter code of "PS" After under operation for a certain time, the frost will be formed on the inner wall of freezer, which
(password); might affect the normal refrigerating effect if it is too thick, and might consume more energy.
2. Use " " and " " to roll the parameter Ho1 or Lo1; Therefore, after a certain time of operation, when the thickness of frost reaches to 5mm around, you
should use the attached defrosting shovel to clean it according to the following steps
3. Press "SET" to confirm the parameter and then display the value of parameter;
• Take out the articles kept in the bin and move to appropriate site;
4. Press "SET" to return to the displayed parameter;
• Switch off the power switch;
5. Use " " and " " again to roll for inquiring another parameter;
• Open the outer and inner doors for a while in order to defrost naturally;
6. Press "SET" to confirm the parameter and then display the value of parameter;
• After frost melts well, discard the water from the drawers or drain it from the drainage hole
7. Press "SET" for more than 3 seconds to return to normal.
• Dry the bottom of bin with a dry cloth;
• The extreme value available for inquiry should be data within 24 hours. If the • Restart the equipment after the bin is cleaned well;
operating time could not last for 24 hours after start, the extreme value for current • Put all articles back in the bin that has already been cooled down.
time will be displayed after start.
• When inquiry, if press Fn for 1 second, this record will be deleted (the value will
be changed to measured value by current sensor).
• Do not use knife or screwdriver or similar tools with sharp edges to defrost.
• For freezer with defrosting function (such as DW-40L278/278J/278(a)/348/348J/508/508J/508(a)/
420F and DW-30L420F), you could press "Defrost" on the computer board for 5s after keyboard is
unlocked, and the compressor will stop for operation. However, this key could not recover the status
automatically(DW-30/40L420F could be recovered automatically without pressing), so, you should
press this key for 5s again after defrosting is completed, and the compressor will be recovered for
normal operation.
• Defrosting should be carried out once per month, up to once per three months.
20 37
• Temperature setting of hi-low temperature alarm
Horizontal cabinet series When unlocking,
1. Press "SET" for more than 3s, then the screen will display the parameter code of "PS" (password);
• When the temperature captured by sensor is higher than+5℃ (or higher than the setting value of 2. Press "SET" button to enter password setting;
high-temperature alarm), "AH" and temperature will be alternately shown; when the temperature 3. Use " " and " " to roll the figure to "-15";
captured by sensor is lower than -5℃ (or lower than the setting value of low-temperature alarm), "AL"
4. Press "SET" to confirm the password;
and temperature will be alternately shown.
5. Use " " and " " to roll the parameter name (the parameter of high temperature alarm is AH,
• For any short circuit, open circuit, or fault, or incorrect connection happens on the temperature
and that of low temperature alarm is AL); 6. Press SET to display corresponding parameter value;
sensing probe, "Err" will be shown, and when the sensor's temperature is higher than 85℃ , "EHi"
will be shown, and when lower than -50℃ , "ELo" will be shown. At the same time, the beeper for 7. Use " " or " " to add or reduce the value of parameter;
temperature controller will be initiated, and corresponding alarm parameter (such as Err, Ehi, Elo) will 8. Press "SET" button to temporarily save the modified parameter value and return to the display
be shown on the screen correspondingly, as well as the alarm mark (without display of temperature). parameter;
• If the temperature sensing probe is judged for error, power should be cut off, and 9. When modifying other parameters, repeat steps 5~8;
the connection of sensing probe should be checked firstly. 10. Press "SET" for more than 3s to save the revised parameter and then exit the parameter setting
• If any internal code is shown when the user is adjusting the temperature procedure.
controller, please wait for 1 minute until the temperature is shown again, in order to Note: It will exit the setting procedure and then lock the controller if there is not any pressing within 3
prevent from any fault due to any change of control parameter. minutes, and it will perform based on the parameter set at last.
DW-40L278/278J/278(a)/348/348J/508/508J/508(a)
Automatic recovery of alarm
• Unlocking
• Under alarm, the beeper alarm could be ceased by pressing the cancellation key on board (for Type When the indicating light of "Lock" is turned on, all keys except "Digit Tran" will be locked, and it
DW-30L278/508, press any key, and the flashing alarm and remote alarm will not be ceased). should be unlocked before setting any parameter. For unlock, the following steps should be followed:
• If the alarm condition is still available, the beeper alarm will be recovered automatically 30 minutes 1. Press "Digit Tran" for more than 5 seconds, then "F0" will be shown, while the Digit "0" will flash;
after cease.
2. Press "Temp. Adjst" once and "F1" will be shown on the display area;
DW-25L92/262/300, DW-40L92/262 and DW-30L278/508 is not equipped with 3. Press "SET" once, the indicating light of "Locked" will be off and the keyboard will be unlocked.
such function. • User parameter setting
User could change the setting temperature of freezing chamber, alarm value, start delay, and other
Remote alarm terminal parameters via control panel.
1. Parameter inquiry and setting
Remote alarm terminal is respectively equipped in Type DW-40L278/278J/2 a. Under the unlock status, press "SET", "L0" will be shown on the screen, and Digit "0" will flash;
78(a)/348/348J/508/508J/508(a)/420F and DW-30L420F: b. Press "Digit Tran" could change "L0" to L1, L2, L3, L4, or L5;
• Remote alarm terminal is equipped on top of freezer, and the alarm signal COM c. After the designated parameter is set, press "SET" once again, the setting value will be shown on
of electric cabinet should be output via the terminal. The carrying capacity the screen, and the value of digit unit will flash. the list corresponding to parameter codes is as follow:
NO
of such terminal is 30V, 2A, DC. Parameters code Parameter meaning Parameters code Parameter meaning
NC
• Contact output: L0 Control temperature setting L3 Inquiry of current voltage
Remote alarm terminal could be classified into three types, including High-temperature alarm value Current ambient temperature
L1 L4
normally open, normally closed, and public. Users may select NO or NC setting query
according to their own requirements. Low-temperature alarm value
L2 L5 Delay of door ajar alarm
setting
The remote alarm function should be initiated after the user installs the alarm
device, which should be used cooperating with remote alarm interface. d. Press "Digit Tran" to change the flashing digit, and then press "Temp. Adjst" to change the value of
current digit;
e. Adjust to the required temperature value and then press "SET", the parameter will be changed and
saved.
36 21
• The default setting temperature is -36℃ ;
• Door ajar alarm delay will be only applicable to DW-40L278/278-ARQ/278J; DW-30/40L420F
• The voltage and ambient temperature value could only be inquired other than
After connecting with the power, the freezer start working, and the inner temperature will be shown
being set;
on screen.
• The range for setting hi-low temperature alarm is as follow:
• Historical alarm query
Maximum setting temperature + 5℃ ≥ high temperature alarm value ≥ set
When the inner temperature is shown normally, press the "Alarm Record" for 5s, and "HIS" will be
temperature + 5℃ .
shown on the panel (Inquiry of alarm history), and then press "Alarm Record" once again and enter
Minimum setting temperature - 5℃ ≥ low temperature alarm value ≥ set the inquiry list of over-temperature alarm history, including four parameters: AL1, AL2, AUL, and HAt.
temperature - 5℃ .
Parameters code Instruction
• For indication, the symbol "├" before the number refers to "+". AL1 The highest temperature in the last alarm process
2. Setting start delay AL2 The lowest temperature in the last alarm process
AUL Average temperature in the last alarm process
a. Under the unlock status, press "SET" and "Temp. Adjst" for 5 seconds at the same time, the current HAt Lasting time in last alarm (unit: minute)
start delay time will be shown on the screen;
Select parameter by " " and " ", and press "Alarm Record" to check the parameter value, then
b. Press "Temp. Adjst" could change the start delay time;
return by pressing "Alarm Record" again after completing checking, and then could select another
c.Adjust to the required value and then press "SET", the parameter will be changed and saved. parameter code for inquiry. After completion of parameter value inquiry, press "Alarm Record"
3. Defrosting twice to return to "HIS". You may exit the parameter inquiry interface if there is no operation on any
Under the unlock status, press "Defrost" for more than 5 seconds, the refrigerating system of freezing interface of inquiry above mentioned for 10s, and then current inner temperature will be shown.
chamber will cease for operation, and "dF" will be shown on the screen. This status could not be For inquiry of HIS parameter during over-temperature alarm, the parameter value for such alarm will
recovered automatically, but could only be exited after pressing "Defrost" for more than 5 seconds be displayed; otherwise, the parameter value for the last alarm will be displayed. The alarm history
again. While successful exit, "dF" will not be shown on the screen again. parameter should be saved in memory within 30min after completion of each over-temp. alarm.
For any adjustment of parameter, the system will save current setting and then • Cause of alarm
exit the setting mode if there is no operation within 10 seconds; Fault
Alarms Status Indication Beeper
code
For the first starting up, 3h after
DW-30/40L420F High When the inner temperature default setting, start buzzing
temperature is higher than the setting E9 Alarm indicator flashes alarm function. When in normal
alarm temperature of alarm operation, make the buzzing
• Temperature adjustment alarm after delay for 15 min.
Press "Set" and then "Ts" will be shown on the screen; press "Set" again, current setting temperature Low When the inner temperature
will be shown, the default is "-25.0", which could be increased or decreased via " " or " ". After temperature is lower than the setting E10 Alarm indicator flashes
entering the setting value, press "Set" for 5 seconds to save it and display current temperature. alarm temperature of alarm
Alarm indicator flashes,
During the process above mentioned, the setting of parameter will not be saved if there is no Power
the black screen and inner Beep alarm will be sent after
operation on any interface for 10 seconds, and then it will return to display the current temperature. failure Power failure of equipment /
temperature are on the 1min delay.
alarm
• DIS parameter inquiry display screen alternatively
Press " " and " " at the same time, "PS" will be shown on the control panel; then press "Set", Excessive
The ambient temperature is
ambient E14 Alarm indicator flashes The buzzer does not alarm
"00" will flash; press " " or " " to enter the correct password (default is 06); press "Set" to enter above 38 °C
temperature
the interface for adjusting system parameter: firstly, "DIS" parameter will be shown, and the first Fault occurs on the main Alarm indicator flashes and
"DIS1"will be shown after pressing "Set"; and the value of "DIS1" will be shown after pressing "Set" E2 Buzzer alarm
sensor in refrigerator E2 code is displayed
once again, and then it will return to show "DIS1" after pressing "SET" once again. By " " and " ", Sensor Fault occurs on the Alarm indicator flashes and
E1 Buzzer alarm
you may inquire the value of "DIS2", "DIS3", "DIS4", and then press "Set" to return after review. exception combinetemperature sensor E0 code is displayed
Fault occurs on the display Alarm indicator flashes and
Press "Set" again to return to show "DIS". You may exit the parameter inquiry interface if there is E0 Buzzer alarm
sensor E1 code is displayed
no operation on any interface of inquiry above mentioned for 10 seconds, and then current inner The battery is low or the
temperature will be shown. Low battery E5 Alarm indicator flashes The buzzer does not alarm
battery switch is not turned on
Door ajar The door is opened or not
E6 Alarm indicator flashes Buzzer alarm
alarm closed tightly
22 35
DIS display parameter:
DW-40L278/278J/278(a)/348/348J/508/508J/508(a) Parameters code Instruction
DIS1 Show the current temperature value of inner sensor (before adjustment)
After connecting with the power, the freezer start working, and the inner temperature will be shown DIS2 Display the current temperature value of the display sensor
on screen panel. DIS3 Show the current temperature value of ambient temperature sensor
• Working status of the display panel: • Restore factory setting
"Lock" pilot lamp: it is ON referring to that all settings are locked to avoid maloperation. Pressing "Set" + " " +" " 10 seconds at the same time, to restore factory setting.
"Alarm" pilot lamp: it is ON or flashing referring to the occurrence of fault.
• Function key Horizontal cabinet series
Stop beeping: under lock or unlock, the beep sound could be ceased by pressing "Silence" (recover
in 30 minutes), but flashing light alarm could not be ceased. • Electronic temperature controller
Alarm test: it is used to test that whether the alarm system is under normal status and whether the Electronic temperature controller is equipped on the control panel of freezer, and the actual
battery is low in power or not. Under unlock, press such key, all indicating lights will be on, and temperature within the bin will be shown on the digital screen after being electrified. Schematic
beeper will sound, and "8" will be shown on all digital tubes. Under lock, press such key when any diagram of electronic temperature controller.
alarm sounds, corresponding fault alarm code will be shown, and the reason could be inquired by the
list as follow: HG
Alarms Status Fault code Indication Beeper
Beep alarm will be
When the temperature
High initiated in 5h after initial
in the bin is higher than Alarm
temperature E9 start, which will send I F
the set value of high indicator flashes
alarm alarm for 15min delay
temperature alarm A E
under normal work.
When the temperature
Low
in the bin is lower than Alarm
temperature E10
the set value of low indicator flashes
alarm
temperature alarm B CD
Alarm indicator flashes,
Power the black screen and A Setting button F Add button
Power failure of Beep alarm will be sent
failure / inner temperature are B Locking indication G Refrigeration status
equipment after 30s delay
alarm on the display screen C Alarm indication H Power indication
alternatively D Defrosting indication I Function button
Excessive The ambient E Decrease button
Alarm indicator
ambient temperature is above E14 ------------
is always on • Under normal use, the setting temperature will not be adjusted generally. Any
temperature 38 °C
Alarm indicator adjustment might cause Hi (AH) or low (AL) temperature alarm, and at that time,
Fault of the main
E6 flashes and E6 code is Buzzer alarm please adjust the value of hi-low temperature alarm (AH, AL) at the same time.
Abnormal sensor in the bin
displayed • Generally, the value of hi-low temperature alarm should be set within the range of
sensor
Ambient temperature Alarm ±5℃ .
E0 ------------
sensor fault indicator flashes
The battery is low or • Temperature setting
Alarm indicator
Low battery the battery switch is not E5 ------------ Current temperature value will be shown on the screen after the freezer is electrified. If the setting
is always on
turned on temperature should be set again, for example: if the previous setting temperature is -20℃ , and now
Delay (10min by default) it should be adjusted to -25℃ , please follow the steps by numbers:
Door ajar Door is open or is not Alarm
E7 and then beeper will
alarm closed tightly. indicator flashes
sound
For more than one alarm is initiated, each code of alarm fault will be shown for 3s
respectively.
34 23
Alarm
24 33
setting. If during the setting process, you press "Set", "Digit Tran", and "Temp. Adjst" again for over
5s, it will save the setting and exit at once. Normal temperature will be shown on the screen. Test holes
• Setting the time for recording data by U disk interface (setting the system current time) (DW-
30/40L420F):
Press "Down" for 5s, then the left two digits on the screen (2 digits from left to right) would stably
show "1P" and the right two digits (2 digits from right to left) will show the Year by flashing (10 by
default); press "Up" once, 1 year will be added, while press "Down" once, 1 year will be reduced. It
could show in a cycle from 10 to 30; and then press "Set" to save after setting is completed. When
the left two digits on the screen (2 digits from left to right) stably show "2P" and the right two digits
(2 digits from right to left) show the Month by flashing (01 by default); press "Up" once, 1 month will
be added, while press "Down"once, 1 month will be reduced. It could show in a cycle from 01 to 12;
and then press "Set" to save after setting is completed. When the left two digits on the screen (2
digits from left to right) stably show "3P" and the right two digits (2 digits from right to left) show the
Date by flashing (01 by default); press "Up" once, 1 day will be added, while press "Down" once, 1
day will be reduced. It could show in a cycle from 01 to 31; and then press "Set" to save after setting
is completed. When the left two digits on the screen (2 digits from left to right) stably show "4P" and
the right two digits (2 digits from right to left) show the Hour by flashing (01 by default); press "Up"
once, 1 hour will be added, while press "Down" once, 1 hour will be reduced. It could show in a cycle
from 00 to 23; and then press "Set" to save after setting is completed. When the left two digits on the
screen (2 digits from left to right) stably show "5P" and the right two digits (2 digits from right to left)
show the Minute by flashing (01 by default); press "Up" once, 1 minute will be added, while press
"Down" once, 1 minute will be reduced. It could show in a cycle from 00 to 59; and then press "Set" to
save after setting is completed. If there is no pressing within 10 seconds during the setting process, it Test holes
will exit without saving the setting and then show the inner temperature. If "Set" is being pressed for 5
seconds during setting process, it will exit after saving and then show the inner temperature. • Purpose
1. Relevant test should be carried out by putting the test line in the freezer to test the articles or
environment within it.
2. Balance the pressure in and out of the bin in order to open the door cover easily.
• Use method
1. For any data test, you should only open the end caps of braiding at both terminal of the test hole,
and then put in the test line, and stuff the hole with the end caps, which are made of rubber so that
the test line will not be test line.
2. If the temperature within the bin is too low, the door cover will be difficult for open, that is because
the pressure within the bin is too small due to low temperature. At that time, you could open the end
cap at the outer side of test hole in order to reduce the internal pressure, so that the door cover could
be opened.
• The picture above is only a schematic diagram, and the test hole is located at
different site due to different model, so please carry out relevant operation based
on the model you have bought.
• After leading out the measuring line, please stuff the test hole with heat
insulating materials again, otherwise the temperature within the freezer would
stay high, and dew or frost would be condensed surrounding to the outside of
through-hole.
32 25
Temperature recorder USB-Connector
Inner temperature recorder with 6-inch standard panel could be optionally equipped with Type DW- • USB interface functions
40L278/278J/278(a)/348/348J/508/508J/508(a)/420F and DW-30L420F to record the change of inner
USB computer board could save 10 years of test data, which will collect data once every 6 minutes
temperature.
and then save it automatically; after memory is fully occupied, the latest data will automatically
• Installation and operation replace the earliest one.
In order to operate and use recorder correctly, please follow the following steps: 1. Format for saving exported data: the format of exported data will adopt *.CSV, each CSV file refers
1. Open the recorder door, and the recorder can be seen; one month of data with a file name of "Year+Month. CSV". One year of exported data will be saved in
2. Switch on the 9V DC battery on the upper right of recorder, which is a spare power; a folder automatically, which will named with the current year.
3. Check the paper of recorder, which should be replaced based on the operation steps stated in P27 2. Treatment of abnormal exported data: any abnormal status will be filled with corresponding
as needed. characters; for example, "NULL1" means outage fault; while "NULL2" means communication fault of
4. Take down the plastic cover of recorder pen and then close the door of recorder. USB board and main control panel.
The recorder will start working until the inner temperature falls in the recording 3. Export data: After the U disk is plugged, it will automatically recognize the U disk and import data
range. into it. Under locked status, you could rapidly press SET for 1 second (press "Up" for Type DW-
30/40L420F), and then "USb" will be shown on the display panel in a cycle of "1 second on and 1
second off", which means that data is being imported into U disk. If import is not completed, it will exit
and return to show current temperature after flashing for 5s. If "ALL" is shown steadily on the display
panel, it means that data has already been completed, and it will exit and return to show current
temperature after 5s. Take U disk from USB interface. Format of exported data is as follow:
序号 /No. 时间 /Time 设定温度 /Setting Temp. 箱内温度 /Inner Temp.
00000 20130425 11:40 -40.0 -40.1
00001 20130425 11:46 -40.0 -40.2
00002 20130425 11:52 -40.0 -40.0
... ... ... ...
• Setting the time for recording data by U disk interface (setting the system current time) (DW-40L278
/278J/278(a)/348/348J/508/ 508J/508(a)):
• Power Supply After it is unlocked, press "Set", "Digit Tran", and "Temp. Adjst" for over 5s, then the left two digits
AC power should be adopted for operation of freezer when the recorder is under normal condition. on the screen (2 digits from left to right) would stably show "1P" and the right two digits (2 digits
If AC power supply fails, the LED indicating light on the recorder will flash to indicate that the power from right to left) will show the Year by flashing (10 by default); press "Temp. Adjst" once, 1 year will
supply is abnormal. The recorder will continue to record the inner temperature by spare battery. Each be added, and it could show in a cycle from 10 to 30; and then press "Set" to save after setting is
spare battery could sustain for about 30 hours (note: such battery should be replaced timely when completed. When the left two digits on the screen (2 digits from left to right) stably show "2P" and the
the quantity is low, so as to prevent corrosion of battery button due to not-timely replacement) In right two digits (2 digits from right to left) show the Month by flashing (01 by default); press "Temp.
order to save the battery, please switch off the battery button when the freezer is not in use. In normal Adjst" once, 1 month will be added, and it could show in a cycle from 01 to 12; and then press "Set"
process, please install spare battery well in order to take records normally under outage condition. to save after setting is completed. When the left two digits on the screen (2 digits from left to right)
The LED indicating light of recorder would continue to flash until the main power (AC power) is stably show "3P" and the right two digits (2 digits from right to left) show the Date by flashing (01 by
connected and its spare battery is replaced. When the power of spare battery is low, LED indicating default); press "Temp. Adjst" once, 1 day will be added, and it could show in a cycle from 01 to 31;
light of recorder should be replaced. and then press "Set" to save after setting is completed. When the left two digits on the screen (2
• Repair and replacement of spare battery of recorder digits from left to right) stably show "4P" and the right two digits (2 digits from right to left) show the
Hour by flashing (01 by default); press "Temp. Adjst" once, 1 hour will be added, and it could show in
When the green LED light of recorder is flashing, there are two ways of checking, including:
a cycle from 00 to 23; and then press SET to save after setting is completed. When the left two digits
1. Remove the spare battery. If LED light is off and the recorder stops, it means that the main power on the screen (2 digits from left to right) stably show "5P" and the right two digits (2 digits from right to
(220V/50Hz) of recorder is in trouble, so that you should check that whether any outage, short circuit, left) show the Date by flashing (01 by default); press "Temp. Adjst" once, 1 minute will be added, and
or loose wiring happens If the main power is repaired, then the recorder could work under normal it could show in a cycle from 00 to 59; and then press "Set" to save after setting is completed. If there
condition, and please install the spare battery now. If the LED light stop flashing, it means that the is no pressing within 10s after the setting process is completed, it will exit and automatically save the
spare power is normal. If the LED light keeps flashing, it means that the spare battery is low in power,
26 31
2 After 3 seconds, the stylus will move from -40℃ to -20℃ and then stay there for 3s, and you could and the spare battery should be replaced until the LED green light is on without flashing.
tune the pointer by means above mentioned at this time; 2. After the spare battery is removed, if the LED light continue to flash and the recorder works
3. After that, the stylus will move from -20℃ to 0℃ and also stay there for 3s, and you could also tune normally, it means that the main power is normal but the spare battery is low in power, which should
the pointer as the same way; be replaced until the LED green light is on without flashing.
4. After 3s, the stylus will move the scale corresponding to the temperature sensed currently and then • Replacing the record paper
the calibration is completed. For replacing the recorder paper, refer to the following steps:
• Replacing the record paper 1. Find out the "Press Out" button (3#) on the upper left in the front of recorder panel;
After one recording cycle, you must replace a new piece of paper by the following steps: 2. Press (3#) once and then the recorder pen will be moved from the left of paper;
1. Press 3# once and then the stylus will move to the outside; 3. When the recorder pen is moved out of the paper completely, please loosen the central bolt and
2. After the stylus moved to the outside of the paper completely, you could loosen the tightened bolt remove the old paper and then replace with the new one, the time line of which should be aligned
and then replace a new piece. Please carefully align the scale of initial time on paper with the base with the time slot on the panel (which refers to a small groove on the left side of panel);
line on panel; 4. Tighten the central bolt again and then press 3# once to reset the recorder pen and then start
3. After tightening the bolt again, you could press 3# once again, and then the recorder will enter into recording temperature;
the mode of paper calibration. You may calibrate the paper by reference to the instructions stated in 5. Check that whether the pen could be is usable on such paper, otherwise, you could adjust the pen
the last page; beam in order to keep the nib contacting with the paper.
4. After the calibration is completed, the temperature recorder will be under normal operation. (Note: Do not damage the nib and beam. If it is difficult for adjustment, you could remove the beam
• Time setting of recorder and ID (number of recorder) setting with screwdriver and then slightly bend it and install it back again, and then repeat the operation
1. Insert the memory card into the recorder, and the indicating light of memory card will flash; above mentioned)
2. Press B. If time and ID have already been set in the memory card, the recorder will update its time 6. In order to keep the records accurate, the precision of pen should be calibrated after paper
and ID, and will delete all records in it such as temperature and time; replacement by means as follow: press the button (3#) until the pen is away from the paper, and then
3 After completion, the indicating light of memory card will stay on, otherwise it will keep flashing; press (3#) again until the pen is back to the paper again. At that time the pen will stay on the outer
temperature scale mark of paper (which might not be marked with temperature value, but it would
• Data reading of recording apparatus
be the scale mark at the outer circle). If the pen will not stay on the site above mentioned, you may
1. Insert the memory card into the recorder, and the indicating light of memory card will flash; use 1# or 2# arrow key to adjust the pen's location within 5 seconds, in order to align the nib with the
2. Press A. If there is enough space in memory card, data will be sent to it, and then after completion outer scale mark. If you could not adjust the pen's location within 5 seconds, please repeat Step 6.
of transmission, indicating light will stay on; and if transmission is not successful, the indicating light • Calibration of recorder
will keep flashing.
The recorder has already been precisely adjusted before delivery, and will not be affected even it is
3. Unplug the memory card and connect the supporting USB kit to the computer for checking data. outage. If it is sure for adjustment, please adjust it as follows:
• Recorder will save data once per hour, and could save up to 2 weeks of data. 1. Make the freezer under operation until the operation keeps stable (the temperature curve becomes
When the space is fully occupied, the records of the earliest 1h will be covered. regular), and the recorder has already recorded temperature curve for at least 2 hours;
• When the data within the memory card is read, it will still be kept in the recorder; 2. Put a standard thermo-couple in a bottle with 100ml of 10% glycerite, and then put such bottle and
but while ID or real time has been set again, the data will be eliminated. the sensor of recorder into the bin together (the recorder sensor could not contact with any solution);
• Calibration of recorder 3. About 4 hours after putting such bottle in the bin and when the temperature of the glycerite is the
same as that within the bin, you could compare with the temperature value of the standard thermo-
After the power of temperature recorder is on, put the sensor probe into an environment with stable
couple and that of the recorder. If they are different, you could adjust the temperature value shown
temperature (such as 0℃ ice-water mixture). After the recorder enters into the temperature status,
on the recorder by pressing the adjustment button on the left (1#) or right (2#) of the panel in order to
you could press 1# or 2# (for about 5 seconds), then the temperature recorder will enter into the
make it consistent with the indicating value of thermo-couple.
mode of temperature calibration after a "Beep" sound. You could press 1# and 2# to adjust it to the
left or right in order to align the pointer with the corresponding temperature scale on paper. After that, Note: The pen could be moved in 5 seconds after the button is loosened.
you could press 3# to confirm or press it for 5 seconds until a "Beep" sound, then the recorder will • Recording pen begins to move after loosening button 5s.
exit the mode of temperature calibration and return to the mode of recording. • The recorder has been calibrated before leaving factory. Do not adjust the
• Recorder has already been calibrated before delivery, please do not arbitrarily recorder optionally unless special circumstances.
adjust it except any special condition.
• If the freezer is not in use, please switch the recorder to "0" in order to prevent
the built-in battery from over discharge.
30 27
7. Recharge indicating light--When it keeps on, it means that it is under fast charge; when it is
Temperature recorder with USB interface flashing, it means that the fast charge is completed, and it is under trickling charge; and when it is off,
it means that charge is completed.
Temperature recorder with USB interface could be optionally equipped with Type DW-40L278/278J/2 8. Button--this switch is the one for rechargeable battery. For normal use, please switch it on, so
78(a)/348/348J/508/508J/508(a)/420F and DW-30L420F to record the change of inner temperature. that the recorder could continue to record temperature for about 24 hours when the battery is fully
In order to use the recorder in an accurate and reasonable way, please read the following instructions recharged but external power is cut.
seriously. • First use
• Understand your temperature recorder 1. After normal power on, please open the door of recorder and switch the battery to "I" to connect
with the back battery;
2. A piece of paper is attached to the recorder. Please check up the paper. If it is needed to replace it,
please refer to instructions on P30;
3. In order to protect the recording pen, please cover its nib with the plastic cover, and take it before
using it;
4. According to the machine and your demand, please adjust the temperature range of recorder and
the recording time cycle. The temperature range shows the range of temperature recorder, while the
time cycle refers to the time when the temperature recorder take a full piece of paper of records. After
completion of adjustment, the recorder will automatically record current setting position and cycle
without repeated adjustment.
• Adjustment of temperature range and cycle
1. After normal power on, press 3# for about 5 seconds, and then the recorder will cease for moving
after a sound of "Di" is sent from the recorder, which means that the recorder enters into the setting
Standard 7-inch paper is adopted for the disc type temperature recorder, and the data could be mode;
exported by paper and U disk. Please understand your temperature recorder by referring to the graph
2. Press 1# could switch the range. Press once, the color of indicating light below will change once,
above firstly:
and different color refers to different position, and the colors and positions could be related as follows:
1. Top cover and lock latch--relatively protect the temperature recorder. You could open the top cover
Green: -40℃ ~+30℃ (applicable to this product, which should be selected)
by slightly lift the latch and pull the cover, and could close the top cover tightly by press the latch after
closing the door; Yellow: -115℃ ~+50℃
2. Thin film button--it could be used for setting and other basic operation; Red: -200℃ ~+0℃
1# button: select the temperature range or adjust to left Note: The recorder will start working until the inner temperature falls in the recording range based on
the temperature range above.
2# button: select the recording cycle or adjust to right;
3. Press 2# to select the recording cycle. There is two cycle for choosing, which would be
3# button: setting button
differentiated by the sound made by the machine:
A button: read the data in U disk;
When the cycle is 24 hours: the sound will quick but short;
B button: time setting;
When the cycle is 7 days: the sound will short but continuous (this is recommended, for paper could
3. Working indicating light--it will show different colors of light to refer to different ranges of be saved)
temperature test. For normal use, it will show green light; and when it flashes, it means that the
4. After adjust the range and cycle in place, press 3# once to exit the setting mode.
recorder is recording the temperature;
• Paper calibration
4. Stylus--the nib of stylus could record the data by tightly laying on the paper. The metal pole
of stylus is relatively flexible, which could keep the curve clear. The stylus will be covered with 1. After the power is on and the position and cycle have already been set, the equipment should
a protective cover when delivery, which should be taken when use. It is suggested to keep the calibrate the paper before validate the setting of position and cycle in order to keep the test accurate.
protective cover in order to protect the stylus when it would be a long time for no use; If the position and cycle has not been adjusted 5 seconds after the power is on, the machine will
enter into the mode of paper calibration automatically by the means as follows:
5.USB interface--Matched memory card could be inserted in for setting time and exporting data;
After pressing 3# once, the stylus will be moved to the site marked with -40℃ and stayed there for 3
6. Press out he nut and paper--screw and tighten the nut clockwise, and replace the paper by
seconds, and at this time you could check that whether the pointer aligns with the scale on paper or
removing the nut;
not. If not, you could tune the pointer by pressing 1# or 2#;
28 29