This document is a certificate of calibration for a digital thermometer. It provides information about the customer, calibration dates, reference standards used, calibration results, and signatures. Key details include: - The thermometer was calibrated by Phoenix Calibration in the Dominican Republic on December 27, 2022. - It was found to be out of tolerance and was adjusted. - Reference standards including temperature baths and probes were used to calibrate the thermometer. - Temperature measurements were taken at 12.5°F, 212.3°F, and 358.3°F and the thermometer passed calibration.

Uploaded by

Michael Carrion
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
53 views3 pages

Certificate of Calibration: CERT. NO.: C321923

This document is a certificate of calibration for a digital thermometer. It provides information about the customer, calibration dates, reference standards used, calibration results, and signatures. Key details include: - The thermometer was calibrated by Phoenix Calibration in the Dominican Republic on December 27, 2022. - It was found to be out of tolerance and was adjusted. - Reference standards including temperature baths and probes were used to calibrate the thermometer. - Temperature measurements were taken at 12.5°F, 212.3°F, and 358.3°F and the thermometer passed calibration.

Uploaded by

Michael Carrion
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

809 563-0457 · 786 866-5606

[email protected]
www.phoenixcalibrationdr.com
Calle D, Esquina C. El Diamante
No. 18, Zona Ind. de Herrera
Santo Domingo, Rep.Dom. 11001
@phoenixcalibrationdr

CERTIFICATE OF CALIBRATION CERT. NO.: C321923


GENERAL INFORMATION CALIBRATION INFORMATION

Customer: JOHANSEN & CO., S.A. Calibration Date: 12/27/2022


Contact: Michael Manuel Carrion Muñoz Next Due: 12/27/2023
Service No.: 46965 Cal. Freq.: 12 Months
Asset No.: 9A-9914 AsFound Condition: Out of tolerance
Description: Digital Thermometer Calibration Results: Passed
Manufacturer: THERMOPROBE INC Calibration Location: Gral. Lab
Model No.: TP9-A Calibrated by: Alberto Uribe
Serial No.: 9A-9914 Environment: 66.9°F 56%RH
Range: See Datasheet Cal. Procedure: DR-WI-0125
Location: N/D

This calibration is traceable to the International System of Units (SI), through National Metrology Institutes (NIST, PTB, NRC, NPL, etc.). This calibration
conforms to the requirements of ISO/IEC 17025:2017. Results contained in this document relate only to the items tested or calibrated. The reported
expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage factor k=2, such that the
confidence level approximates 95%. The customer is responsible for assigning the calibration frequency if needed, and it does not imply that the conformity
of the equipment with the specifications is maintained. This certificate shall not be reproduced except in full and with the written consent of Phoenix
Calibration. The statement of conformity is based on simple acceptance, whether the calibration result is within or outside the manufacturer’s
specification/acceptance limits, with the TUR expressed on the datasheet. It is the responsibility of the end user to determine if it is appropriate for your
specific application. Pass: Results within limits. Fail: Results exceed limits.

REMARKS
Performed Routine Calibration/Certification. Instrument was received out of tolerance and was adjusted accordingly. As Found Cal Correction = (Nominal -
As Found Reading) / As Left Cal Correction = (Nominal - After Adjustment Reading). Environmental Conditions Uncertainty: +/-0.22 °F & +/-1.6 %RH”

REFERENCE STANDARDS USED TO CALIBRATE EQUIPMENT

ID Inst Description Serial No. Cal. Date Due Date


10126 Temperature and Humidity Datalogger 98:8B:AD:20:EF:BE 05/05/2022 05/31/2023
5235 Water Bath -40°C to +150°C 100054027 06/26/2022 06/30/2023
9334 Temperature Indicator (-25 to 200) °C 75012050201 03/02/2022 03/31/2023
5027 Temperature Bath 125 °C 100255139 09/29/2022 09/30/2023
9016 Temperature Bath GX2010 200°C 13-882 06/25/2022 06/30/2023
6714 Platinium Probe 450°C SN906193 09/09/2022 12/31/2022

REVISION COMMENT

SIGNATURES

Calibrated by: Alberto Uribe Quality Approval: María del Rosario

Date: 12/27/2022 Date: 12/30/2022

Revision No.: 0 Page 1 of 3 DR-FM-0085 Rev. D Issue Date: 10/17/2022


809 563-0457 · 786 866-5606
[email protected]
www.phoenixcalibrationdr.com
Calle D, Esquina C. El Diamante
No. 18, Zona Ind. de Herrera
Santo Domingo, Rep.Dom. 11001
@phoenixcalibrationdr

DATASHEET CERT. NO.: C321923


Temperature - °F - Measure - Average
Units Max. Error(Tol.) Nominal Low Limit HighLimit AsFound Meas. Uncert. TUR
°F 0.2 12.52 12.32 12.72 10.72 N/A 0.49 0.4:1
After Adjustment 12.51 N/A 0.49 0.4:1
°F 0.2 212.31 212.11 212.51 210.51 N/A 0.41 0.6:1
After Adjustment 212.27 N/A 0.41 0.6:1
°F 0.5 358.31 357.81 358.81 356.52 0.43 1.6:1
After Adjustment 357.99 0.43 1.6:1

Temperature - Measure Repeatability 14


Units Max. Error(Tol.) Nominal Low Limit HighLimit AsFound Meas. Uncert. TUR
°F 0.2 12.52 12.32 12.72 10.72 N/A ** n/a
After Adjustment 12.51 N/A ** n/a
°F 0.2 12.52 12.32 12.72 10.72 N/A ** n/a
After Adjustment 12.51 N/A ** n/a
°F 0.2 12.51 12.31 12.71 10.71 N/A ** n/a
After Adjustment 12.51 N/A ** n/a
°F 0.2 12.51 12.31 12.71 10.71 N/A ** n/a
After Adjustment 12.51 N/A ** n/a
°F 0.2 12.51 12.31 12.71 10.71 N/A ** n/a
After Adjustment 12.51 N/A ** n/a
°F 0.2 12.52 12.32 12.72 10.72 N/A ** n/a
After Adjustment 12.51 N/A ** n/a
°F 0.2 12.53 12.33 12.73 10.73 N/A ** n/a
After Adjustment 12.51 N/A ** n/a
°F 0.2 12.52 12.32 12.72 10.72 N/A ** n/a
After Adjustment 12.51 N/A ** n/a
°F 0.2 12.52 12.32 12.72 10.72 N/A ** n/a
After Adjustment 12.51 N/A ** n/a

Temperature - Measure Repeatability 212


Units Max. Error(Tol.) Nominal Low Limit HighLimit AsFound Meas. Uncert. TUR
°F 0.2 212.31 212.11 212.51 210.51 N/A ** n/a
After Adjustment 212.27 N/A ** n/a
°F 0.2 212.31 212.11 212.51 210.51 N/A ** n/a
After Adjustment 212.27 N/A ** n/a
°F 0.2 212.31 212.11 212.51 210.51 N/A ** n/a
After Adjustment 212.27 N/A ** n/a
°F 0.2 212.31 212.11 212.51 210.51 N/A ** n/a
After Adjustment 212.27 N/A ** n/a
°F 0.2 212.32 212.12 212.52 210.52 N/A ** n/a
After Adjustment 212.27 N/A ** n/a
°F 0.2 212.31 212.11 212.51 210.51 N/A ** n/a
After Adjustment 212.27 N/A ** n/a
°F 0.2 212.31 212.11 212.51 210.51 N/A ** n/a
After Adjustment 212.27 N/A ** n/a
°F 0.2 212.31 212.11 212.51 210.51 N/A ** n/a
After Adjustment 212.27 N/A ** n/a
°F 0.2 212.29 212.09 212.49 210.49 N/A ** n/a
After Adjustment 212.29 N/A ** n/a

Revision No.: 0 Page 2 of 3 DR-FM-0086 Rev. D Issue Date: 12/20/2022


809 563-0457 · 786 866-5606
[email protected]
www.phoenixcalibrationdr.com
Calle D, Esquina C. El Diamante
No. 18, Zona Ind. de Herrera
Santo Domingo, Rep.Dom. 11001
@phoenixcalibrationdr

DATASHEET CERT. NO.: C321923


Temperature - Measure Repeatability 356
Units Max. Error(Tol.) Nominal Low Limit HighLimit AsFound Meas. Uncert. TUR
°F 0.5 358.39 357.89 358.89 356.59 N/A ** n/a
After Adjustment 358.05 N/A ** n/a
°F 0.5 358.38 357.88 358.88 356.58 N/A ** n/a
After Adjustment 358.03 N/A ** n/a
°F 0.5 358.34 357.84 358.84 356.59 N/A ** n/a
After Adjustment 358.02 N/A ** n/a
°F 0.5 358.30 357.80 358.80 356.50 N/A ** n/a
After Adjustment 358.00 N/A ** n/a
°F 0.5 358.32 357.82 358.82 356.52 N/A ** n/a
After Adjustment 358.00 N/A ** n/a
°F 0.5 358.32 357.82 358.82 356.52 N/A ** n/a
After Adjustment 357.98 N/A ** n/a
°F 0.5 358.32 357.82 358.82 356.52 N/A ** n/a
After Adjustment 357.96 N/A ** n/a
°F 0.5 358.21 357.71 358.71 356.41 N/A ** n/a
After Adjustment 357.96 N/A ** n/a
°F 0.5 358.21 357.71 358.71 356.41 N/A ** n/a
After Adjustment 357.94 N/A ** n/a

End of datasheet

( ) Compliance/Pass.
( ) Non-compliance/Fail.
(**) Represents Non ISO/IEC 17025 Accredited Values.

Revision No.: 0 Page 3 of 3 DR-FM-0086 Rev. D Issue Date: 12/20/2022