LCMS Notice
LCMS Notice
Jan.2022
LabSolutions™ LCMS
Read the instruction manual thoroughly before using the product. Keep this
instruction manual handy so that you can refer to it when necessary.
<< Notices >>
1) If the user or usage location changes, be sure this Instruction Manual is always kept
together with the product.
2) To ensure safe operation, contact your Shimadzu representative for installation,
adjustment, or re-installation after moving the instrument to a different site.
3) The content of this manual is subject, without notice, to modifications for the sake of
improvement.
4) Every effort has been made to ensure that the content of this manual was correct at
the time of creation. However, in the event that any mistakes or omissions are
discovered, it may not be possible to correct them immediately.
5) The contents of the hard disk in a PC can be lost due to an accident. To protect your
important data from accidents, be sure to back up your data.
6) Microsoft, Windows, Windows 10, and Windows 7 are registered trademarks of
Microsoft Corporation in the USA and other countries.
7) Third-party trademarks and trade names may be used in this publication to refer to
either the entities or their products/services, whether or not they are used with
trademark symbol "TM" or "®".
8) Screenshots of Microsoft products are used in accordance with guidelines from
Microsoft Corporation.
Be advised that the same results are not printed out when a data file is reprocessed as the
quantitative calculation formula and the display format are different if the data processing
settings are different on the two PCs. Please unify the data processing settings in the
[Data Processing Settings] window and use the PCs in the laboratory.
Note. For MS data, quantitative calculation is always performed by a method with the
[Apply] setting even if a method with the [Not Used] setting is selected. Therefore,
run the calculation with the default setting (method where the dilution factor is used)
unless otherwise necessary.
To change the setting, perform the procedure described below.
-1-
2. LC Units
2.1 ROM Versions
Please check the ROM version of each LC unit and consider an upgrade to a later
version to obtain the functionality as described in the table below.
Modules listed in the table below are also controlled via system controller
SCL-40 / CBM-40 / CBM-40lite
Model Name ROM Version Remarks
LC-30AD 2.10 or later
LC-30ADSF 1.00 or later
LC-20ADXR 1.30 or later
LC-20AD 1.10 or later
LC-20AT 1.07 or later
LC-20AB 1.10 or later
LC-20AP 1.10 or later
LC-20AR 1.00 or later
LC-20Ai 3.11 or later
LC-10Ai 3.1 or later
LC-6AD/7A 1.4 or later
LC-8A 1.5 or later
SIL-30AC 3.11 or later
SCL-40/CBM-40/CBM-40lite Ver. 1.40 or later
SIL-30ACFV 3.20 or later
is necessary.
Rack Changer II 2.00 or later
SIL-30ACMP 1.00 or later
SFE-30A 1.00 or later
SIL-20AXR/20ACXR,
1.22 or later
SIL-20A/20AC
-2-
SIL-10Ai 3.50 or later
SIL-10AF 4.10 or later
SIL-10AP 4.10 or later
CTO-30A 2.00 or later
CTO-30AS 1.00 or later
CTO-20A/20AC 1.07 or later
CTO-10ASvp 5.28 or later
SPD-20A/20AV 1.10 or later
RF-20A/20AXS 1.00 or later
RID-20A 1.00 or later
RF-10A 3.0 or later
RF-10AXL 3.32 or later
CDD-10Avp/sp 5.00 or later
FRC-10A 3.30 or later
SFC-30A 1.00 or later
Option Box VP 5.22 or later
b. i-Series LC-2060
Model Name ROM Version Remarks
LC-2060C 3.10 or later
LC-2060C 3D 3.10 or later
LC-2060C 3D MT 3.10 or later
RF-20A/20AXS 1.00 or later
RID-10A 3.0 or later
RID-20A 1.00 or later
-3-
LC-2040C 3D MT /
2.10 or later
LC-2040C 3D MT Plus
RF-20A/20AXS 1.00 or later
RID-10A 3.0 or later
RID-20A 1.00 or later
e. Nexera X2 series
Model Name ROM Version Remarks
LC-30AD 2.01 or later
CBM-20A/lite Ver. 2.10 or later is
LC-30AD LPGE mode 2.10 or later necessary.
Support for 10mL/min flow rate.
CBM-20A/lite Ver. 3.20 or later is
LC-30ADSF 1.00 or later
necessary.
Ver. 3.11 or later is necessary for Nexera
SIL-30AC 2.00 or later
UC.
CBM-20A/lite Ver. 3.60 or later is
SIL-30ACFV 3.20 or later
necessary.
Rack Changer II 2.00 or later
CBM-20A/lite Ver. 2.20 or later is
SIL-30ACMP 1.00 or later
necessary.
CBM-20A/lite Ver. 3.20 or later is
SFE-30A 1.00 or later
necessary.
CTO-30A 2.00 or later
CBM-20A/lite Ver. 2.20 or later is
CTO-30AS 1.00 or later
necessary.
CBM-20A/lite Ver. 2.40 or later is
SPD-M30A 3.00 or later
necessary.
CBM-20A/lite Ver. 3.30 or later is
SPD-30AM 4.00 or later
necessary.
CBM-20A/lite Ver. 3.20 or later is
SFC-30A 1.00 or later
necessary.
* Use CBM-20A/lite for the system controller.
f. Nexera XR series
Model Name ROM Version Remarks
LC-20ADXR 1.21 or later
CBM-20A/lite Ver. 2.50 or later is
LC-20ADXR LPGE mode 1.30 or later
necessary.
LC-20ADXR CBM-20A/lite Ver. 2.60 or later is
1.30 or later
Compressibility Setting necessary.
SIL-20AXR/20ACXR 1.22 or later
* Use CBM-20A/lite for the system controller.
g. Prominence series
Model Name ROM Version Remarks
CBM-20A/lite 2.02 or later
LC-20AD 1.10 or later
LC-20AT 1.07 or later
LC-20AB 1.10 or later
LC-20AP 1.00 or later
CBM-20A/lite Ver. 2.60 or later is
LC-20AP LPGE mode 1.10 or later
necessary.
-4-
CBM-20A/lite Ver. 2.90 or later is
LC-20AR 1.00 or later
necessary.
CBM-20A/lite Ver. 3.10 or later is
LC-20Ai 1.00 or later necessary.
LPGE mode is not supported.
SIL-20A/20AC 1.22 or later
Rack Changer/C 1.11 or later
CTO-20A/20AC 1.07 or later
SPD-20A/20AV 1.04 or later
SPD-20A/20AV 100Hz sampling 1.10 or later
SPD-M20A 1.08 or later
SPD-M20A 100Hz sampling 1.10 or later
RF-20A/20AXS 1.00 or later
CBM-20A/lite Ver. 2.80 or later is
RID-20A 1.00 or later
necessary.
i. Nexera Prep
Model Name ROM Version Remarks
LH-40 1.00 or later
FRC-40 1.00 or later
j. LC-2010
Model Name ROM Version Remarks
LC-2010A/C 1.13 or later
2.11,
LC-2010AHT/CHT
3.10 or later
k. LC-VP
Model Name ROM Version Remarks
5.43, 6.22 or
SCL-10Avp
later
SCL-10Asp 6.03 or later When use the CDD-10Avp/sp
LC-10ADvp 5.31 or later
LC-10ATvp 5.31 or later
Ver. 5.35 or later is necessary when it is
SIL-10ADvp 5.34 or later
used with the CBM-20A/lite.
SCL: 6.22 or
later
SIL-HTA/C
SIL: 6.06 or
later
CTO-10Avp/10ACvp 5.27 or later
CTO-10ASvp 5.28 or later
SPD-10Avp/10AVvp 5.24 or later
SPD-M10Avp 3.30 or later Available only for 32bit Windows.
CDD-10Avp/sp 5.00 or later
Option Box VP/Subcon VP 5.22 or later
-5-
l. LC-10A series
Model Name ROM Version Remarks
LC-10AD 3.1 or later
LC-10AT 3.1 or later
LC-10Ai 3.1 or later
LC-10AS 3.1 or later
LC-6AD/7A 1.4 or later
LC-8A 1.5 or later
When the CBM-20A is used, 4.20 cannot
SIL-10A/10Ai 3.50 or later
be used.
When the CBM-20A is used, 4.20 cannot
SIL-10AF 4.10 or later
be used.
When the CBM-20A is used, 4.20 cannot
SIL-10AP 4.10 or later
be used.
SIL-10AXL 3.61 or later
CTO-10A/10AC 3.0 or later
SPD-10A/10AV 3.0 or later
SPD-10Ai/10AVi 3.0 or later
RID-10A 3.0 or later
RF-10A 3.0 or later
RF-10AXL 3.32 or later
Option Box L, Valve I/F 3.2 or later
When the CBM-20A is used, 3.60 cannot
FRC-10A 3.30 or later be used. 3.61 or later version is
necessary.
m. Others
Model Name ROM Version Remarks
Ver. 1.5 or The analog input board (PC-55N) is
ELSD-LT II
later separately necessary.
Ver. A.0.7 or
ELSD-LT III
later
Analog input board (PC-55N) - SCL-10Avp/sp or CBM-20A is required.
Analog input board Available only for
-
(for LC-2030/2040/2050/2060) LC-2030/2040/2050/2060
The system configuration setting up
CBM-201m 1.0 or later CBM-201m and other devices together is
not supported.
Centrifugal field-flow fractionation (CF3)
FFF-C8030 1.01 or later
unit
-6-
Solvent blending LC-30AD Ver. 2.20 or later
The concentration of the mobile
phase can be controlled Ver. 2.40 or later
using the valve attached to the
pump.
Switching injection method SIL-30AC Ver. 3.15 or later
for SIL-30AC with SFC
system
Ver. 3.40 or later
The injection method (Direct or
Loop(SFC) injection) can be
specified in the method.
* When the system controller is SCL-40/CBM-40/CBM-40lite, there is no limit on the
ROM version of the system controller.
Note. We recommend that you ask Shimadzu service associate to perform ROM
upgrades.Please note that this service is not covered under warranty and is billed
separately.
-7-
2.2 RS-232C Communication Settings
(1) SCL-10Avp/SIL-HT
LabSolutions communicates with SCL-10Avp/SIL-HT via an Enhanced Level protocol.
Use the following procedure to set the protocol mode to “Enhanced”.
1. Turn ON the power supply on the SCL-10Avp/SIL-HT.
2. From the main menu, press "F4 (SYSTEM)" and "F3 (NEXT)" to open the
COMMUNICATION SETTING screen.
3. Set the communication settings as follows.
CLASS-VP: 5.x
Interface: RS-232C
Baud rate: 19200
Level: Enhanced
4. Turn OFF the SCL-10Avp/SIL-HT power supply, then re-start.
The Enhanced mode cannot be used with the HIC-SP Suppressor Ion Chromatograph
(CDD-6A + SCL-10Asp Ver 1.xx). Please set the Level to “Standard” as described in
the procedure below.
1. Turn ON the SCL-10Asp power supply.
2. From the main menu, press "F4 (SYSTEM)" and "F3 (NEXT)" to open the
COMMUNICATION SETTING screen.
3. Set the communication settings as follows.
CLASS-VP: 5.x
Interface: RS-232C
Baud rate: 19200
Level: Standard (or Level2)
4. Turn OFF the SCL-10Asp power supply, and then re-starts.
-8-
2.3 SPD-M10Avp Photo Diode Array Detector
Note. SPD-M10Avp cannot be used in Windows 10.
Note. SPD-M10Avp cannot be used with a 64-bit OS, so perform the following procedures
with a 32-bit OS.
(1) Driver
It is necessary to install SPD-M10Avp driver when using SPD-M10Avp detectors on
LabSolutions. Follow the next procedure to install the SPD-M10Avp driver.
1. Shut down the PC after LabSolutions is installed. Connect the SPD-M10Avp
detectors to the PC and turn on the power of the detectors.
2. Turn on the PC after the initialization of the SPD-M10Avp detectors is finished.
In the case of Windows 7, the installation is automatically complete by the
hardware wizard.
Note. Even if LabSolutions is not installed, the hardware wizard starts the installation of
the SPD-M10Avp driver when the Windows starts and the installation fails. In this
case please cancel the hardware wizard and close it, then install LabSolutions and
reboot PC.
Note. If the ROM is older than version 3.3, it is necessary to upgrade it. We recommend
that you ask Shimadzu service associate to perform ROM upgrades for
SPD-M10Avp model. Please note that this service is not covered under warranty
and is billed separately.
-9-
5. Select the [SHIMADZU SPD-M10Avp] item, and click the [OK] button to close the
[SCSI Instruments] sub-window.
6. Click the [OK] button to close the [Instrument] sub-window.
7. Double click the [Detector] icon in Available Modules to open the instrument tree
and move SPD-M10Avp to Modules Used for Analysis.
8. Click the [Auto Configuration] button.
9. Double click the [SPD-M10Avp] icon to open the [PDA Detector] sub-window and
check the ROM version.
10. Click the [OK] button to close the [System Configuration] window.
-10-
3. Using the Online Manual
3.1 Manual
All information necessary for using LabSolutions LCMS is provided by the documents
contained in the C:\LabSolutions\Manual folder ( on the LabSolutions LCMS Installation
disc: \LabSolutions\ManualFL\en).
*1 The LCMS-2050 software instruction manual is provided in common with the hardware.
Refer to LabSolutions LCMS Operators Guide for LCMS-2020 from LCMS-2050
instruction manual.
*2 Select [Print As Image] in [Print] dialog box before printing.
*3 In terms of applicable instruments, "LCMS-SQ" is indicated in the shared content for the
LCMS-2020/2050", LC/MS/MS" is indicated in the shared content for the
LCMS-8030/LCMS-8040/LCMS-8045/LCMS-8050/LCMS-8060(NX), and "LCMS
Q-TOF" is indicated in the content for the LCMS-9030.
*4 It is on the LabSolutions LCMS Installation disc: \BridgeSoftwareProMass.
*5 It is on the LabSolutions LCMS Installation disc:
\LabSolutions\Supplement\LogCollectionTool\Manual.
No. Documents Instruments*3 Files
LabSolutions LCMS Installation LCMS-SQ LCMS_Installation&
1 and Maintenance Manual LC/MS/MS MaintenanceGuide.pdf
LCMS Q-TOF
LabSolutions LCMS Operation LCMS-2020 LCMS_OperatorsGuide.pdf
2 Manual*1 LC/MS/MS LCMSMS_OperatorsGuide.pdf
LCMS Q-TOF
LabSolutions LCMS LCMS-SQ SystemUsersGuideFile.pdf
3 Administration Manual LC/MS/MS
LCMS Q-TOF
LabSolutions LCMS Data LCMS-SQ DataAcquisition&Processing
4 Acquisition and Processing LC/MS/MS TheoryGuide.pdf
Theory Manual LCMS Q-TOF
LabSolutions LCMS Quick LCMS-2020 LCMS_GettingStartedGuide.pd
Manual*1 f
5
LC/MS/MS LCMSMS_GettingStartedGuide
.pdf
LabSolutions LCMS Software LCMS-SQ LCMS_ReleaseNotes.pdf
6 Release Notes LC/MS/MS
LCMS Q-TOF
Software License Authentication LCMS-SQ LicenseManageToolManual.p
7 Instructions LC/MS/MS df
LCMS Q-TOF
Quick Guide for Connection of LCMS-SQ QuickGuideforConnectionof
8 CBM-20A (lite) and LC/MS/MS CBM-PDA.pdf
SPD-M20A/SPD-M30A LCMS Q-TOF
Prominence-i / Nexera-i LCMS-SQ QuickGuideforICM.pdf
9 Interactive Communication LC/MS/MS
Mode(ICM) LCMS Q-TOF
LabSolutions LCMS Readme LCMS-SQ ReadMe.pdf
10 LC/MS/MS
LCMS Q-TOF
Notice before using LabSolutions LCMS-SQ LCMS_Notice.pdf
11 LCMS (This sheet) LC/MS/MS
LCMS Q-TOF
-11-
LabSolutions Certificate of LCMS-SQ CertificateOfCompliance.pdf
12 Compliance*2 LC/MS/MS
LCMS Q-TOF
LabSolutions LCMS i-PeakFinder LCMS-SQ i-PeakFinder_Tips&Tricks.pdf
13 (New Peak Integration) LC/MS/MS
Guide to Adjusting Parameters LCMS Q-TOF
LabSolutions LCMS LCMS-SQ UpdateManual.pdf
14 Update Program Instruction LC/MS/MS
Manual
LabSolutions LCMS LCMS Q-TOF LCMS_Q-TOF_QuickReferenc
15 Quick Reference e.
(for LCMS Q-TOF) pdf
LabSolutions Direct Setup Guide LCMS-SQ LabSolutionsDirectSetupGuide.p
16 LC/MS/MS df
LCMS Q-TOF
LabSolutions Direct Quick Guide LCMS-SQ LabSolutionsDirectQuickGuide.p
17 LC/MS/MS df
LCMS Q-TOF
How to use ProMass Bridge LCMS-SQ ProMassBridge.pdf
18 software*4 LC/MS/MS
LCMS Q-TOF
Shimadzu Bridge program for LCMS-SQ ProMassBridge_QuickGuide.p
19 ProMass Quick Start Guide*4 LC/MS/MS df
LCMS Q-TOF
Notice before using LabSolutions LCMS-SQ GPC_Notice.pdf
20 GPC LC/MS/MS
LCMS Q-TOF
LabSolutions GPC Instruction LCMS-SQ GPC_OperatorsGuide.pdf
21 Manual LC/MS/MS
LCMS Q-TOF
LabSolutions GPC Quick Manual LCMS-SQ GPC_QuickGuide.pdf
22 LC/MS/MS
LCMS Q-TOF
LabSolutions Log Collection Tool LCMS-SQ LabSolutionsLogCollectionTool
23 Instruction Manual*5 LC/MS/MS Manual.pdf
LCMS Q-TOF
-12-
4. Sample Files and Report Template Files
4.1 Sample Files
The sample data files and sample report files are copied to the following folder after
installation of LabSolutions LCMS.
C:\LabSolutions\Sample\LC (for LC)
C:\LabSolutions\Sample\MS (for LCMS-2020)
C:\LabSolutions\Sample\MSMS (for LCMS-8030 / LCMS-8040 / LCMS-8045 /
LCMS-8050 / LCMS-8060(NX))
C:\LabSolutions\Sample\Q-TOF (for LCMS-9030)
(”C:\LabSolutions” means target folder for installation.)
-13-
5. CLASS-Agent Manager
5.1 Setting the instrument information files of the CLASS-Agent Manager
Several items about LabSolutions's data are added in the CLASS-Agent Manager.
Notice the installation order of the CLASS-Agent Manager and LabSolutions.
Please install the LabSolutions after the CLASS-Agent Manager. When the CLASS-Agent
Manager is installed after the LabSolutions, please add the instrument information files of
the CLASS-Agent Manager, which set the display of LCsolution’s data in the
CLASS-Agent Manager.
Procedure
1. Please copy the new instrument information files on the LabSolutions Installation disc
folder:
“\LabSolutionsLCMS\Supplement\Agent\inf2\en”.
2. Please paste the files to the \Agent2\inf2 folder of the drive that installs the
CLASS-Agent manager.
If the instrument information files are shared in the network, please confirm where the
currently used instrument information files by selecting [Tools]-[Configuration]-[Common
Info File] in the CLASS-Agent Manager.
Batch conversion via the Data Explorer window is recommended for conversion
from the CLASS-LC10.
-14-
(2) Select a method file or files (xxxx.M??) for the CLASS-LC10.
(3) From the menu displayed by right-clicking in the Data Explorer window,
choose [File Conversion]-[Convert CLASS-LC10 (…) File to LabSolutions
(LC) File].
(4) Specify the destination folder and click the [OK] button.
(5) The files will be converted into the LabSolutions format as the ones named
"xxxx_M??.LCM."
Note. When method files of UV detector with dual mode are converted, data
analysis parameters of Ch1/Ch2 are converted to the parameters of
detector A (Ch1)/B (Ch1) in LabSolutions.
-15-
6.2 Procedure for converting data files
6.2.1 Converting chromatograms
a. Batch conversion via the Data Explorer window
(1) Select the All Files tab of the Data Explorer window.
Note. In the conversion including methods and calculation results, files can be
converted only when there are a D-file, C-file, and M-file as a set.
-16-
a. Batch conversion via the Data Explorer window
(1) Select the All Files tab of the Data Explorer window.
(2) Select a calculation data file or files (xxxx.D??) for the CLASS-LC10.
(3) From the menu displayed by right-clicking in the Data Explorer window,
choose [File Conversion]-[Convert CLASS-LC10 (…) File to LabSolutions
(LC) File].
(4) Specify the destination folder and then click [OK] button.
(5) The file(s) will be converted into the LabSolutions format as the ones
named "xxxx_D??.LCD."
-17-
7.1 Handling CLASS-VP files
The CLASS-VP files that may be used in/converted into the LabSolutions are as
follows.
CLASS-VP file Files name on LabSolutions
Method file (*.MET) Method file (*_MET.lcm)
Data file (*.DAT, etc. ) Data file (*_DAT.lcd)
When method/data files for the CLASS-VP are used on the LabSolutions, the
following restriction will apply.
(1) The conversion or seamless loading of data files for the CLASS-VP
requires that CLASS-VP Version 5.04, Version 6.12, or later version be
installed in the computer (this is not applicable when the already converted
data is reused). In addition, CLASS-VP Version 6.13 or later version (when
using Version 5 series, Version 5.05 or later version in Version 5.X) is
necessary when converting CLASS-VP PDA data.
(2) The print format is different between the CLASS-VP and the LabSolutions.
Therefore, no report format information for the CLASS-VP will be converted.
A report format file for the LabSolutions must be created separately.
(3) The following data in a method file may be converted:
Comments
Instrument parameters
Integration parameters (CHROMATOPAC algorithm only)
Column performance parameters
Compound table (described later)
No other data (properties, calibration curve information, identification/
-18-
quantification parameters, advanced settings, custom reports, and system
suitability) will be converted.
(4) In the conversion of a compound table from the CLASS-VP into the
LabSolutions, the following data marked with Yes will be converted from a
method file and data file. For the data that cannot be converted (marked
with No), the default values will be assumed.
Method file conversion Data file conversion
Conversion Default value Conversion Default value
Compound name Yes Yes Identified
peaks only
Type Yes No Target
Analysis channel Yes Yes
Retention time Yes Yes
Concentration (1-) Yes No 1
Bandwidth Yes No 0.01
Identified peak selection -- Closest Peak -- Closest Peak
Concentration unit Yes Yes
Calculated by Yes No Area
Calibration curve Yes No Linear
Zero Yes No Not Forced
Weighting Yes No None
Window/band -- Band -- Band
Spiked No 0 No 0
(5) In the conversion of data files for the CLASS-VP, chromatograms, the last
analysis result, and the method and configuration settings in the original
data will be converted.
(6) The original method used for data acquisition will be converted. Therefore,
if the data processing parameters or PDA multi-chromatogram settings are
modified, the settings in the method within the converted data may not
conform to the chromatograms and calculation results.
In this case, separately convert the method used for Postrun analysis into
the LabSolutions format and then create a batch table to perform Postrun
batch processing. In the LC/PDA data analysis, no method file can be
specified to make a Postrun analysis.
-19-
LabSolutions installed. If no configuration settings file is available, the
method file will not be converted properly.
7.3 Procedure for converting method files
a. Batch conversion via the Data Explorer window
(1) Select the All Files tab of the Data Explorer window.
Note. If the correct configuration settings file is not selected, the instrument parameters
-20-
will not be converted properly. Be sure to select the same configuration settings
file as the one used for data acquisition (or the one that has been similarly set
up)
(5) Specify the destination folder and then click [OK] button.
(6) The files will be converted into the LabSolutions format as the ones named
"xxxx_met.lcm."
Note. If the correct configuration settings file is not selected, the instrument parameters
will not be converted properly. Be sure to select the same configuration settings
file as the one used for data acquisition (or the one that has been similarly set
up)
(3) Choose [Files]-[Open Method File] in the Data Acquisition window from the
[Analysis] window or in the [Calibration Curve] window from the Postrun
window.
(4) Select [CLASS-VP Method File] as Type on the Open Method File window.
-21-
(5) Select a method file from the list of files and then click the [Open] button.
(6) The method file for the CLASS-VP will be loaded.
(2) Select a data file or files (xxxx.dat, etc.) for the CLASS-VP.
(3) From the menu displayed by right-clicking the Data Explorer window,
choose [File Conversion]-[Convert CLASS-LC10 (…) File to LabSolutions
(LC) File].
-22-
(4) Specify the destination folder and then click [OK] button.
(5) The file(s) will be converted into the LabSolutions format as the ones
named "xxxx_dat.lcd."
(3) Select data files (xxxx.dat, etc.) from the list of files and then click the
[Open] button.
(4) The data files for the CLASS-VP will be loaded.
-23-
8. Notice for PDF File Output
8.1 Notice for PDF Output Settings
To output PDF file, perform PDF out directly from the menu (for example, the [File]-[Data
Report]-[PDF Output] menu in the [Data Analysis] window) or select the Redirect Report to
PDF file check box in the PDF Output tab of the [Tools]-[Options] menu.
(Although SkyPDF Pro Driver is displayed in the Printer Name list box in the Printer Setup
window, do not select SkyPDF Driver in the list box.)
-24-
15 PDA Data Analysis Graph Image Current project {Data filename}.pdf
(Chromatogram) folder
16 PDA Data Analysis Graph Image Current project {Data filename}.pdf
(Spectrum) folder
17 PDA Data Analysis Spectrum Index Current project {Data filename}.pdf
folder
18 PDA Data Analysis UV Library Search Current project LSS UV Library Search.pdf
Results folder
19 PDA Data Analysis Data Report Current project {Data filename}.pdf
folder
20 MS Data Analysis Graph Image Current project {Data filename}.pdf
(Chromatogram) folder
21 MS Data Analysis Graph Image Current project {Data filename}.pdf
(Spectrum) folder
22 MS Data Analysis MS Library Search Current project {Data filename}.pdf
Results folder
23 MS Data Analysis Data Report Current project {Data filename}.pdf
folder
24 UV Library Editor Library Report Folder of UV {Library filename}.pdf
library file
25 MS Library Editor Library Report Folder of MS {Library filename}.pdf
library file
26 Quant Browser Print Table Current project (LC)
folder LSS Quantitative Browser
Table.pdf
(MS)
LSS Quantitative Browser
MS Table.pdf
27 Quant Browser Quant. Report for Current project (LC)
Current Data folder LSS Quantitative Browser
Data.pdf
(MS)
LSS Quantitative Browser
MS Data.pdf
28 Quant Browser Data Report Current project {Data filename}.pdf
folder
29 Data Browser Print Image of all cells Current project {Layout filename}.pdf
folder or Data Browser Image.pdf
30 Data Browser Graph Image Current project {Data filename}.pdf
folder
31 Data Browser Print Table Current project {Data filename}.pdf
folder
32 Data Browser Print Sample Current project {Data filename}.pdf
Information folder
33 Data Browser Data Report Current project {Data filename}.pdf
folder
-25-
9. About the setting of Windows
9.1 Precautions for Windows Update
When PC is connected to the network, Windows can recognize, download and install
important updates(programs) for the security of Windows, by using the "Windows
Updates" function. In this function, the Windows reboot after the installation is complete
despite the status of applications, which might discard the data in analysis or files being
edited.
To prevent this sort of problem, change the “Windows Updates” setting as below.
< Windows 7 >
Specify “Check for updates (choose whether to download and install them)” as the
Windows Update setting. Do not specify “Install updates automatically”.
< Windows 10 >
On the Windows 10, it is difficult to prevent from updating automatically. Specify the time
when the PC can reboot at the settings of “Update & security” – “Notify to schedule restart”
or “Change active hours”. In addition, check on the “Defer upgrades” in the advanced
options.
There is a case that when PCs participate in a network (domain), the setting of "Windows
Updates" is forced by the domain or the network configuration. In such a case, please
check with your network administrator.
In addition, change the settings as below when you restrict the Sleep function.
-26-
1) Select "Power Options" in the Control Panel.
2) Click "Change plan settings" for the current plan.
3) Click "Change advanced power settings".
4) In the tree, select "Never" for the "Sleep" setting.
-27-
10.3 Notice for installed files and folders
The installer copies the LabSolutions LCMS system files to the following folders.
LabSolutions LCMS cannot work if the files on these folders are removed, moved, or
renamed.
C:\LabSolutions\System (“C:\LabSolutions” means target folder for installation.)
(In the case of 64bit OS)
C:\Program Files (x86)\LabSolutions
C:\Program Files (x86)\Shimadzu
(In the case of 32bit OS)
C:\Program Files\LabSolutions
C:\Program Files\Shimadzu
(“C:” means target drive for installation in Windows System.)
-28-
11.2 How to convert MS library to (MSn) MS library file
If you would like to use .lib library file for new (MSn) MS library feature, you can convert .lib
file to .mlb file by the following procedure.
In the .MLB file converted from .LIB file, information which is not included .LIB file (e.g.
collision energy (CE) ) is set to default value. For example, polarity information does not
include .LIB file, this value is set to [Positive] in converted .MLB file. Also, you need to set
MS stage if original .LIB spectrum was acquired MS2.
-29-
13. Others
13.1 Cautions When Adding LCMS for the First Time
When adding LCMS to an existing LabSolutions system for the first time, after adding the LCMS license,
click [Administration Tools]-[Customization Information] and initialize the customization information.
If you do not initialize the customization information, the LCMS item will not be displayed on the toolbar in
the [Report Format] window.
-30-