0% found this document useful (0 votes)
60 views

Cable Matching Info

This presentation discusses cables and cable measurements. It will focus on cable TV trunk cables and measuring structural return loss on coaxial cables. Twisted pair cables used in computer networks will also be discussed. There are different types of cables for various applications. Common cable measurements include attenuation, impedance, delay, and crosstalk. Swept frequency measurements characterize impedance, loss, return loss, phase, and delay. Structural return loss is a specialized return loss measurement.

Uploaded by

rasool wani
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
60 views

Cable Matching Info

This presentation discusses cables and cable measurements. It will focus on cable TV trunk cables and measuring structural return loss on coaxial cables. Twisted pair cables used in computer networks will also be discussed. There are different types of cables for various applications. Common cable measurements include attenuation, impedance, delay, and crosstalk. Swept frequency measurements characterize impedance, loss, return loss, phase, and delay. Structural return loss is a specialized return loss measurement.

Uploaded by

rasool wani
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 43

Solving Difficult

Cable Measurements

Slide #1 Introduce yourself and the topic for the next 40


minutes. Inquire if there are any customers that are
currently performing cable testing and determine the
level of the audience. This section is a survey rather
than a detailed discussion on cable testing.
Solving Difficult
Cable Measurements

RF and MW Device Test Seminar 1993

Let's talk about cable testing in this session. Cables


are the most common devices, except for perhaps
connectors, which may be considered as part of the
cable. Virtually every system must have cabling to
interconnect with other devices. There are many
different considerations in selecting cables and in the
related measurements to characterize them.

7-1
Solving Difficult
Cable Measurements

Slide #2 Four main topics with the most emphasis on two


network analyzer measurements: SRL on coax and
swept measurements on twisted pair.
Agenda

Cable types and trends

Measurements

Coaxial Cables

Twisted Pair cables

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

This presentation is a brief overview of cables and the


trends within the industry.

We'll review the typical measurements that are used to


specify and characterize cables.

More specifically, we will focus on Cable TV trunk


cables and the structural return loss measurement.

Twisted pair cables are widely used in computer LAN


installations and provide some additional
measurement issues.

7-2
Solving Difficult
Cable Measurements

Slide #3 There are lots of different types of cable and some


common measurements to define specifications.
Overview slide.
Cable Types

Type of Cable Application Parameters


Zc, Attenuation,
Capacitance
Coaxial Broadcast, CATV Video Velocity of Propagation
DC Reisistance
Delay
Zc, Attenuation
Audio, LAN, Capacitance
Paired Computer EIA RS232, Velocity of Propagation
RS423 DC Resistance
Crosstalk
Attenuation
Flat Computer Inductance
Instrumentation Crosstalk
FDDI, LAN Aperture, Attenuation
Fiber Optic Telecommunication -3 dB Bandwidth
Microwave Instrument Division
RF & MW Device Test Seminar
0893 cabltest.pre

There are four main types of cables, coaxial, paired,


flat and fiber optic. Each may be used within the same
system.

Specifications, and the measurements that are


performed to verify them, fall into a few basic groups.
For example, attenuation is a common concern in each
of the four types of cable. Isolation from adjacent
cables, or crosstalk is another factor in cable
selection. Basic electrical properties, such as
inductance and capacitance help define the
operational properties. Characteristic impedance, or
aperture in fiber optic systems defines their
capabilities and operational environment.

7-3
Solving Difficult
Cable Measurements

Slide #4 Current changes taking place in the marketplace.


Again, an overview.

Cable Trends
Development of insulating material
Thin cables and low-loss cables

Twisted cables
Reduction of crosstalk for paired cables
Wider bandwidth and higher speed

High speed data transfer


Paired cables for 100 MBPS for LANs
1 GHz bandwidth for Cable TV

Transition to fiber optics

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Cable manufacturing processes and materials are


always changing and evolving. New materials and
cable configurations are continually being developed.
Development of cables that are lighter, stronger and
more efficient in the system operation is always taking
place.

Twisted pair cables are being used widely in LANs


with wider bandwidth requirements and higher
operational speeds, up to 100 mega-bits. Crosstalk
between pairs is a major concern as well as return loss
at the higher frequencies. Cable TV operators are
implementing higher bandwidths as well, up to 1 GHz
and additional channels with digital transmissions.
A trend towards replacement of existing systems and
creating new installations using fiber optics is present
throughout all aspects of the marketplace.

7-4
Solving Difficult
Cable Measurements

Slide #5 Topic slide for the next three slides. This is my


attempt to group the types of measurements and
there is some overlap. In this section the focus is on
Types of Cable Measurements swept frequency measurements.

Swept frequency

Twisted pair

Physical characteristics

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Let's now discuss the types of measurements that are


performed on cables by both the manufacturers and
the end-users. There seem to be three general
categories; swept frequency measurements, twisted
pair characterization and the measurements of the
basic electrical properties. Many specifications are
given at specific frequencies. The measurements may
also be swept in frequency but specified at only a few
representative frequencies. As always with
specifications, there is an attempt to standardize on
measurements. Different types of test equipment
may be used to make the same measurements.

Twisted pair measurements are a series of


measurements that are used to define the relationships
of the pairs to each other and to other pairs within the
cable.

Physical measurements are those measurements that


define the basic electrical properties of the cable; e.g.
resistance or capacitance.

7-5
Solving Difficult
Cable Measurements

Slide #6 Refer to earlier discussions of the measurements in


the filter section. Point out that SRL is a specialized
measurements of return loss referenced to the cable
Swept Frequency Measurements impedance. There will be a discussion of time
domain and gating at the end of the SRL section.
Characteristic impedance, Zc
Insertion loss
Return loss
Structural return loss

Phase Network Analyzer


Group delay

Electrical length
Fault location TDR (FDR)

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Swept frequency measurements are used to evaluate


the usable frequency range of cable and help define the
type of environment for which the cable is suitable.
Impedance, insertion loss and return loss are the most
common measurements of a cable's characteristics.
Structural return loss is a special case of return loss
which will be discussed in greater detail later in this
section.

Phase changes through a cable and the deviation from


linear phase or, group delay, help define the cables
applications. Matched phase cables are required for
many applications.

Measurement of electrical length and determining


propagation delay provide further details on a cable's
operation.

Examining the cable for defects and continuity is


easily implemented with FDR, or Frequency Domain
Reflectometry. When measuring in the frequency
domain, the Fourier transform is used to convert the
measurement data to the time domain display of data.
More on this measurement later.

7-6
Solving Difficult
Cable Measurements

Slide #7 mention that they are made in the course of


characterizing twisted pair.

Twisted Pair Measurements


Multiple Conductor and Balanced

DC resistance unbalance

Mutual pair capacitance

Pair-ground capacitance unbalance


Impedance Analyzer
Pair-pair capacitance unbalance

Crosstalk

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

There are several measurements that are unique to


twisted pairs. These measurements define the
relationship between the two wires in a pair, between
other pairs and the ground, or common reference.
Resistance and capacitance measurements are usually
made with an impedance analyzer or an LCR
measuring device.

Crosstalk measurements are made by injecting a signal


on one pair and measuring the signal induced on
another terminated pair.

These are unique measurements based on pairs or


groups of pairs and their electrical relationship with
each other. Again, no attempt to define the tests, just

7-7
Solving Difficult
Cable Measurements

Slide #8 These are the low frequency measurements that are


usually made on short lengths of cable, usually
wavelength/8 . These are basic physical
Physical Characteristics measurements and not really mentioned again.
Fractional Wavelength Sample Lengths

Capacitance

Propagation delay

Attenuation constant
Impedance Analyzer
Phase constant

DC resistance

Core-shield resistance
Microwave Instrument Division
RF & MW Device Test Seminar
0893 cabltest.pre

Fractional wavelength samples of the cable are


measured. The lengths are determined by the test
frequency of the analyzer, usually one-eighth of the
wavelength. Using fractional wavelengths of cable
removes the presence of lumped parameters which
one would observe in a longer length of cable.

These basic electrical properties are usually measured


with an LCR tester or an impedance analyzer. These
measurements may be made at several frequencies, or
continuously, but are usually specified at single
frequencies.

The characteristic impedance may also be measured in


this fashion.

7-8
Solving Difficult
Cable Measurements

Slide #9 These are the tough measurements that will be


summarized at the end of the section. Each topic is
discussed again. The diagram illustrates the path
Measurement Issues loss through the cable and the reflections returning to
the input port.
Low-loss

Non-Insertable
Requires adapters

Balanced

Long electrical length

Broadband

Long reflection path loss

Non-standard impedances
Microwave Instrument Division
RF & MW Device Test Seminar
0893 cabltest.pre

There are several fundamental characteristics of


cables that make measurements difficult. Low loss is a
desirable feature of cables. But, mismatch with the
source and termination can cause reflections.
Calibration with the full 12-term error correction and
the use of a test set will solve this problem.

Adapters are required to perform connections with the


cable and must be accounted for in the calibration and
measurement.

Balanced pairs require a signal format conversion to


equal and opposite voltages in the pair with no relation
to ground.

Long electrical length causes propagation delays and


may require stepping the sweep to keep the source and
receiver in synchronization.

The long cables also require high dynamic range due to


path losses in the signal traveling through the cable in
a return loss measurement. The test system must have
a dynamic range greater that twice the insertion loss to
observe effects from the far end of the cable.

Cables are not always 50 ohms and may require


special test set or matching networks to measure.

7-9
Solving Difficult
Cable Measurements

Slide #10 This slide describes the typical measurement


environment for reels of cable TV cables. Long cables
mean big losses. There's a table of losses per cable
Cable TV Cables reel on the next slide. Slow sweeps or step mode are
used to compensate for the electrical delay through
Higher Bandwidths the cable.
400 MHz to 750 MHz to 1 GHz

Long lengths of cables

Insertion loss

Electrical delay

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Let's examine the measurements on cable TV cables in


detail. The same concerns and measurements would
also apply to and long cable run; e.g. antenna feeds.

Higher bandwidths are evolving and are affecting the


test requirements and test equipment needs. Higher
frequency testing will lead to more stringent
requirements in test equipment.

Long runs of cable require higher dynamic range for


both insertion loss and reflection measurements.

Insertion losses increase with higher frequencies.

Electrical delay must be taken into consideration due


to long lengths.

7-10
Solving Difficult
Cable Measurements

Slide #11 losses through the cable make insertion loss


measurements impossible for broadband systems.
Recount that dynamic range must exceed the
Cable Attenuation at 1 GHz measurement values by 20 dB to achieve +- 1 dB
accuracy. Point out that the .5 inch feeder cable will
be measured in the video demo and read the values.
Trunk Cables Feeder Cables Drop Cables
Cable Attenuation Cable Attenuation Cable Attenuation
Size db/100' db/Reel Size db/100' db/Reel Size db/100' db/Reel
0.750" 1.74 43.5 0.412" 3.10 96.1 59 7.92 79.2
0.875" 1.53 38.3 0.500 2.53 69.3 6 5.42 64.2
1.000" 1.44 36.0 0.625" 2.11 52.8 611 5.41 54.1
TX840 1.53 38.3 TX565 2.17 54.3 11 4.31 43.1
TX1160 1.20 30.0

Requires Dynamic Range!

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Here are some typical attenuation or insertion loss


values for reels of cable. High dynamic range is
required to measure these values. As discussed in the
filter section, dynamic range is a function of the
highest signal input to the sensitivity of the system.
Dynamic range must exceed the measurement
requirements by 10 dB. Simply matching the required
range yields a 100% error rate.

Consider a TDR or reflection loss measurement where


the transit loss through the cable is doubled.

Trunk and drop cables are relatively low-loss


compared to the feeder cables.

Point out that these tables are for 1 GHz, the new
upper frequency limits for the cable systems. The

7-11
Solving Difficult
Cable Measurements

Slide #12 The cable impedance does change with frequency and
is not a perfect match with the test set.

Cable Impedance

Function of Physical Properties


Dielectric
Inner and outer diameters

Trunk and feeder cable 75 ± 2 Ohms

Drop cable 75 ± 3 Ohms

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Cable impedance in a coaxial cable at higher


frequencies is primarily a function of the physical
dimensions and dielectric quality of the cable.

Trunk and feeder cables impedance tolerance is tighter


than drop cables which reflect the generally more
expensive and higher quality cable.

Impedance variation within the cable may be viewed


as a reflection or return loss.

Cable measurements are made at 75 ohms. The


cables have a large tolerance in impedance values.

7-12
Solving Difficult
Cable Measurements

Slide #13 Reflection loss measurement is just as discussed in


the earlier sections.

Return Loss Measurement

Observe discontinuities as reflections

Measure Impedance variations

Evaluate connectors and cable

Cable
Incident Reel
LOAD

Source
Reflected

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

This measurement is easily made on one end of a


cable. The amount of signal that is reflected is
compared to the signal that is injected.

Any discontinuities or physical variations are observed


as a reflected signal. Impedance variations also appear
as discontinuities as each change reflects power.

Connectors and the cable interface may also generate


reflections. Connector quality and match may be also
examined.

7-13
Solving Difficult
Cable Measurements

Slide #14 The periodic discontinuities may be distributed over


the entire cable. The best example is an extrusion
machine which has a jaw that grasps the cable and
Structural Return Loss pulls to draw out the cable. If the jaw is
Periodic discontinuity mis-adjusted, a periodic dent is placed in the cable
Manufacturing process each time the jaw seizes the cable and pulls. These
"Dropped reel" dents may be too small to observe their reflections
Dents
directly, but can cause an accumulative effect on the
Wavelength / 2 separation cable return loss quality.
Very frequency coherent

λ/2 λ/2 λ/2

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Structural return loss is a special case of return loss


measurements. Physical deformation of the cable, by
handling or manufacturing process, cause reflections.

Structural return loss occurs when these periodic


deformations sum at a half-wavelength spacing and
reflect the input signal.

A narrow spike of high return loss results from a


specific frequency component being reflected from the
cable and not transmitted through the cable.

When the effects from these small discontinuities are


summed, they represent a major effect on the
transmission capability of the cable. Periodic spacing
of irregularities in time or distance leads to a
cumulative effect in frequency.

7-14
Solving Difficult
Cable Measurements

Slide #15 realistic example. Note the scales for the plots and the
markers. The dip in the insertion loss is just a few
dB with the SRL being 15 dB or so out of spec. The
Structural Return Loss old plots, one good cable and one cable after denting,
CH1 MEM log MAG 5 dB/ REF -15 dB 1 _:-36.163 dB
56.831 001 MHz
are the same results as shown in the video tape.
*

Cor
Show the video tape at this point. Be sure to point
1
out that the dents are an attempt to simulate SRL,
but from just a few dents rather than the distributed
Hld
dents throughout the cable.
CH1 START 5.000 000 MHz STOP 200.000 000 MHz
CH2 MEM log MAG 5 dB/ REF -56 dB

Cor

Hld

CH2 START-20 ns STOP 200 ns

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Here is the plot of a network analyzer return loss


measurement and the corresponding time domain
display. A 75 ohm test set was used to match the
impedance of the trunk cable.

The return loss from a good cable and a damaged


cable are shown in the upper plot. The lower plot
shows the time domain display of the return loss
measurement. The peaks represent the reflections
from the cable. The connector is the first reflection in
both cables. The additional reflections in the bad cable
cause the increase in return loss as shown in the upper
display.

There is a new plot which shows SRL and the


insertion loss effect at the same frequency. This is a

7-15
Solving Difficult
Cable Measurements

Slide #16 this system. This is the grease pencil world, where
you adjust an attenuator to match a level on the
display.
Variable Bridge Test Setup
Variable
Sweep Attenuators
Generator
Calibrate
Display
Measure

Cable Reel
Variable 75 Ohm
Bridge

Test Port

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Structural return loss measurements were originally


made with a variable bridge, a sweeper and a detector.
The impedance of the cable is matched and measured,
by adjusting the bridge for a minimum return loss and
reading the value from the bridge dial. Cable
impedance variation may also be eliminated by
adjusting the bridge.

Once a signal level is observed, the switch is thrown


and the attenuator adjusted to match the signal level.
Subtracting the attenuator setting from the source
level determines the signal level being measured.

This manual measurement requires a skilled operator.

This is the original and most common measurement


setup for SRL on cables. Many customers still use

7-16
Solving Difficult
Cable Measurements

Slide #17 system except for attenuators and sweeper


calibration cycles.

Variable Impedance Bridges ERROR: the second sentence should read: As


frequency increases, the bridge lacks good raw
Advantages directivity. The sentence will be removed in
Adjusts to cable impedance revisions!
Rapid measurement

Limitations
Accuracy
Manual adjustment
Dynamic range
Limited frequency range

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

The variable bridge allows a rapid measurement and does a


good job of matching the cable impedance, but.

As frequency increases the bridge lacks good directivity.

Repeatability is as good as the operator and equipment.


Accuracy is a function of the calibration of the attenuators
and the bridge.

The measurements must be made manually.

Dynamic range is limited to the amount of power in the


source and the noise floor of the detector.

The frequency range of the measurements is limited by the


sweeper, bridge and detector.

Quick summary. The old system has lots of


shortcomings. No calibration for anything as a test

7-17
Solving Difficult
Cable Measurements

Slide #18 This is the proposed system for the SCTE


measurement. Better measurement but still the same
problems/limitations with the variable bridge.
Current Test Setup

Power
Splitter

Cable Reel
Variable 75 Ohm
Bridge

Test Port

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Most customers are now using the variable bridge and


the network analyzer to perform their cable testing
measurements. The network analyzer provides a
calibrated measurement and the additional frequency
range that is required.

The analyzer is set to the desired test range and the


bridge is adjusted to match and balance the cable
being tested. The measurement is performed even
more rapidly as the substitution method is not required
to establish the signal levels.

The Society of Cable Test Engineers (SCTE)


recommends this setup as part of their standard
practices for measuring structural return loss.

7-18
Solving Difficult
Cable Measurements

Slide #19 The system can be calibrated, but as soon as you


tweak the bridge, the CAL is gone! This is the current
solution and may be contested by the audience.
Network Analyzer Focus on calibration of the test set. New title on slide
to identify the variable bridge solution.
Advantages
Minimize cable mismatch
Rapid measurement
Wider frequency range

Limitations
Accuracy
Manual adjustment

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

The network analyzer adds the needed dynamic range


and a calibrated measurement to the system, but the
limitations of the variable bridge still affect the
measurement. The directivity, repeatability and
accuracy of the variable bridge technique is limited.

The bridge is usually varied to match the cable


impedance. The bridge may be calibrated to a fixed
standard, but the measurement is influenced by
adjusting the bridge. Resisting the urge to adjust the
bridge is necessary to obtain a good measurement.
Even minor adjustments, "tweaks", can significantly
affect the measurement results and cause the cable to
pass the test requirements.

7-19
Solving Difficult
Cable Measurements

Slide #20

Network Analyzer System


Cable
Incident Reel Transmitted

Source
Reflected

Signal
Separation
Incident Reflection Transmitted
RF

Receiver
IF

Display

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Using a network analyzer for the same measurement


offers several advantages. But, first let's review the
major components of the analyzer.

A 75 Ohm test set is used to provide the connections to


the cable. A bridge configuration provides both the
reflected and incident signal separation, as well as a
load for transmission testing or the insertion loss
measurement.

The receiver signal is converted to an IF signal and


then measured and displayed.

This is the same diagram used in other sessions.


Nothing new, just a review of the components.

7-20
Solving Difficult
Cable Measurements

Slide #21 This is the solution for accuracy and repeatability,


but you must correct for the cable impedance and
adjust your data to measure structural return loss.
Network Analyzer Test Set

Advantages
Repeatability
Accuracy
Automated

Limitations Network Analyzer


Must convert data to cable impedance
Fixed impedance

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Using a test set provides a standard reference for the


measurement system.

Better directivity is provided by the test set, especially


at higher frequencies.

Repeatability of data from test-to-test is excellent.

Accuracy is enhanced by the calibration with known


standards and the use of full 2-port 12 term error
correction.

Automation allows the operator to focus on the


measurement results rather than the measurement
process.

The limitations of this test setup are that the


measurement is performed to the 75 Ohm test set
reference, not the cable impedance. The data must be
compared to the impedance of the cable to conform
with the structural return loss measurement definition.

7-21
Solving Difficult
Cable Measurements

Slide #22 Re-emphasizing the features of the full 2-port


calibration and the errors that are removed. This
slide is used in other sessions and should deserve
Systematic Errors only a brief mention of the errors.

Directivity Frequency Response

R A B
Crosstalk

Source Load
Mismatch Mismatch

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Let's review the test set configuration and the


associated errors. The systematic errors present in the
test set are shown in this diagram. All of these errors
may be removed by calibration with a known set of
standards. Calibrating out these errors adds a high
degree of accuracy and repeatability to the
measurement.

Directivity error is caused by a signal path that does


not pass out through the test set port to the cable
under test. This signal passes through the test set to
the input port or is reflected back from the input port.

Crosstalk is the signal that is coupled around the test


set ports, from the output to the input port.

Source and load mismatch are the errors associated in


the quality of the test port terminations.

Frequency response is the error associated with


increasing frequency and the response of the test set.

7-22
Solving Difficult
Cable Measurements

Slide #23 transformation need not be performed. The


transformation will lower the SRL value. The
benefits of an automated search and accurate,
Impedance Transformation consistent measurements far outweigh the
conversion time. IBASIC can also perform the
Adjust data to match cable impedance
transformation in the 8751.

Measure average impedance


Vector correction to data

Requires an external controller

External complex math


Corrects data display

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Because of the fixed test set impedance, the return


loss data must be transformed to the average cable
impedance. The trace data is averaged out to obtain an
average value and this value is used to perform a
vector correction on the trace data. These values
obtained are then structural return loss and
comparable to the variable bridge technique.

The transformation must be performed with an


external computer using complex math, or in the case
of the newer analyzers, with built-in IBASIC .

If the measured trace data values are sufficiently


below the specifications, no transformations need be
performed.

The transformation is not difficult and can be done.


If the data is below a certain level, the

7-23
Solving Difficult
Cable Measurements

Slide #24

Time Domain Gating

Examine or remove response from connector

Reduce reflections

Increase resolution

View cable response only

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

As mentioned earlier, time domain gating may be used


to examine individual responses from the cable under
test. The sums of the individual components may
obscure the device being examined. For example, at
higher frequencies a poorly fitting slip-on connector
may cause a larger reflection than the cable. The cable
may be viewed independently from the connector by
gating and viewing only the cable response.

Underlying signals which are masked may now be


viewed.

Refer to the earlier discussions about SAW filter


measurements in the filters section.

7-24
Solving Difficult
Cable Measurements

Slide #25 Connector and impedance mismatches, and other


major anomalies may be ignored or bypassed with
this gating.
Time Domain Gating
CH1 RFL&M log MAG 5 dB/ REF -35 dB

Cor

Hld

CH1 START 5.000 000 MHz STOP 200.000 000 MHz

CH2 RFL&M log MAG 5 dB/ REF -56 dB

Cor

Gat 1
Hld

CH2 START 0 s STOP 500 ns


Microwave Instrument Division
RF & MW Device Test Seminar
0893 cabltest.pre

The traces shown here illustrate the effects of time


domain gating. The desired response is enclosed
within the gates in the time domain display. The return
loss data is modified to display only the signals
selected in the time domain trace. This capability
allows us to effectively remove undesirable reflections
from the connector. We can then measure the cable
and ignore the effects from the connector.

The upper trace shows the return loss measurement of


two cables, one with reflections from a bad connector.
By removing the reflections from the connector, the
second trace reveals more detail about the cable
reflections.

The first reflection in the time domain display , the


connector, is gated out, i.e. the gate is set around all of
the other reflections except the one from the
connector.

7-25
Solving Difficult
Cable Measurements

Slide #26

Agenda

Cable types and trends

Measurements

Coaxial Cables

Twisted Pair cables

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Let's now leave the coaxial cable discussion and


examine the measurements which are performed on
twisted pair cables and the techniques required to
make those measurements.

This concludes the coaxial section. Questions or


comments? Now on to twisted pairs. Breakpoint.

7-26
Solving Difficult
Cable Measurements

Slide #27 communication techniques and connections are the


deciding factor. Think of a string of Christmas tree
lights. When the connections are broken or a bulb
Why Twisted Pair? burns out, the lights all go out in a ring
configuration and only one light goes out in the star
Star vs Ring LAN connection configuration . Another comparison is to try to drive
Easier to troubleshoot coax in parallel paths with multiple splitters, versus
Easier to maintain / repair twisted pairs driven in parallel by connecting up to
Reliability terminals. Multiple twisted pairs are much easier to
connect and drive with data.
Suitable for short runs

Major trend for coax and fiber for backbones

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Twisted pair has become popular with the conversion


from ring LAN configurations to the star configuration.
Star configurations use twisted pairs for the shorter
connections from the peripheral systems to the hubs.

If a cable is damaged, or the connector torn off, only


one system will fail rather than the entire ring of
systems. It is easier to trace the location of the cable
fault and correct it in this configuration.

Twisted pair is an economical solution for short runs


of cable.

The major trend is to use coaxial LAN connections for


the major interconnection between hubs or the
backbone of the site installation. Fiber optic cable is
also being used for this type of installation.

The choices are all data or digital format related, not


the electrical properties of the cable. The

7-27
Solving Difficult
Cable Measurements

Slide #28

Twisted Pair Cables

Balanced transmission

Non-standard impedance (100 or 150 Ohm)

LAN usage
25 pair vertical
4 pair horizontal

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

There are some difficulties associated with measuring


twisted pair cables. The cables are used in a balanced
configuration and require a balun device to interface
with most standard test equipment.

Twisted pairs have an impedance other than 50 Ohms,


usually 100 or 150 Ohms.

There are a lot of wires involved. A minimum of 4


pairs are used in a horizontal cable and as many as 25
pairs are contained in a backbone cable. Each of the
pairs must be treated individually and as a group in the
measurement. Each pair must be terminated to avoid
influencing the measurements on the others.

Description of twisted pair cables used in LANs.

7-28
Solving Difficult
Cable Measurements

Slide #29
ERROR: Voltage is equal but opposite to ground at
any point not zero or canceled. The current does
Balanced vs. Unbalanced cancels at any point across the pair.

Unbalanced V
Coaxial cables 0V

+V
Balanced
Twisted pairs -V

I
Effective to 500 MHz
I

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

In a normal coaxial measurement , the signal is


referenced to the shield or ground. There is a center
conductor which is centered in a dielectric cylinder
and surrounded by a shield layer on the outside of the
cylinder.

Balanced transmission is accomplished with an equal


and opposite current flowing through the pair. The
voltage potential is between pairs, not referenced to
ground. Each line of the pair is equal and opposite in
voltage with respect to each other. At any instant in
time, the current and voltage will cancel out at a point.

Description of balanced operation. There is an error


in the handout, in the last sentence.

7-29
Solving Difficult
Cable Measurements

Slide #30 connection. Introduce the balun concept. Balun


selection criteria is discussed in a couple of slides.

Balanced Measurements

Balun
1:2 Balanced cable
Test Equipment +V +V
50 Ohm
Coaxial 0 -V
Unbal Bal 100 Ohms

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

This illustration shows a balun transformer being used


to convert a 50 ohm single ended signal to a 100 ohm
balanced twisted pair. Connectors are also required to
make the connections with the balun and the source.
The balanced end is usually a set of terminal posts.

Care should be taken in choosing the connectors as


they will affect the measurement of the pair. As the
frequency increases, the choice of connectors and
balun configuration becomes more critical.

The balun is used to perform the impedance


transformation as well as balance the twisted pair.
No ground needs to be used in the coaxial

7-30
Solving Difficult
Cable Measurements

Slide #31 The balun is added and the reflection measurement


is performed as before. Calibration removes the
measurement errors associated with the balun and
Balanced Configuration system. This is true only when the twisted pair
Reflection measurement impedance does not vary. Below 1 MHz, the
impedance varies widely and can cancel out any
calibration.

Open
Test Set
Short
Load
Calibration Plane

Balun Cable

Load

Unbal Bal

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

This is the setup to measure the return loss of a


twisted pair.

The test port of the test set is connected to the 50 ohm


side of the balun. A calibration or reference plane is
established on the output side of the balun. This port is
calibrated with a short, open and a load to match the
balun. The associated errors of the output port and the
balun are characterized with the calibration and
removed from the measurement data.

The twisted pair is then connected to the balun and


terminated with the load. The measurement is then
performed and the data interpreted as in an
unbalanced coaxial system.

7-31
Solving Difficult
Cable Measurements

Slide #32

Calibration Standards

Calibration Plane
Considerations
Accuracy Balun
Open
Repeatablity Short
Load
Unbal Bal
Geometry
Minimum lead length
Good electrical connection

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Calibration standards must be chosen carefully as the accuracy of the


calibration and measurement depend upon them. A good electrical
connection with the test connectors is most important for all three
standards. The short and open may be just that, open terminals and a
piece of wire or copper sheet. The load should be measured carefully
to provide the exact resistance , i.e. a 1% resistor reflects a 1%
accuracy in the measurement.

Lead length does not become a factor until the lengths approach a
significant portion of the wavelengths of the frequencies being
measured. With 500 MHz as the upper limit of the testing, carefully
shortened leads are the only consideration. This would not be the
case at 5 GHz! Use a connector that is reasonable for the cable being
tested and build a set of standards using the same connector. Avoid
using alligator clips and long test leads!

Balun calibration standards do not exist and must


be created for the measurement setup and fixturing.
Build connectors and standards with the care
required to reflect your measurement needs. It's not
difficult to generate an acceptable set of standards.

7-32
Solving Difficult
Cable Measurements

Slide #33 Add another balun and a 2-port cal to perform a


transmission measurement. The calibration plane is
set at the end of the cable.
Balanced Configuration
Transmission Measurement

Open Open
Short Thru Short
Load Load

Calibration Plane Calibration Plane

Balun Cable Balun

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

To perform a transmission measurement requires high


dynamic range in the measurement system and a
receiver port that is balanced as well. A second balun
is used as a termination for the twisted pair and is
connected to the input port for the analyzer. A second
calibration plane is established for the analyzer input
which removes the systematic errors from the
measurement. This configuration allows both a
transmission and reflection measurement.

A thru connection must be added to the calibration


standards. This would usually be a representative
short length of the cable pair being tested.

7-33
Solving Difficult
Cable Measurements

Slide #34 when the induced signal levels from another pair
mix with the desired signals and eventually cause
disruption of transmission. This is a complicated
Crosstalk Measurements measurement with lots of data taking and
comparisons between pairs as well as switching and
terminating.
Cable Near end

Terminations

Power
Splitter
Terminations Far end
Balun
Cable
Balun

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Crosstalk measurements are made between pairs of


pairs. One pair is energized and the other pair is
examined for signals present. Baluns are required as in
the transmission measurements. Note that a power
splitter may be used in place of the test set for this
measurement. However, a test set would also work
correctly.

Two different tests are illustrated here, a near end and


a far end crosstalk measurement. The near end refers
to the location of the measurement on the receiving
pair.

While the crosstalk tests are being performed, all of


the other pairs within the cable must be terminated in
a characteristic load

Crosstalk measurements are extremely important to


LAN applications . Crosstalk interference is defined

7-34
Solving Difficult
Cable Measurements

Slide #35 North Hills has 13 different impedance choices with


differing bandwidths and HP has 2 different models.
The references section lists the model numbers and
Balun Considerations information. Baluns can be purchased as
self-contained components or as pulse transformers,
Frequency response wire would around coils.
100 Hz to 10 MHz
100 kHz to 300 MHz Balun
Flat response

Impedance matching
50 to 75, 100 , 150

Parametric measurements
10x rule

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Baluns perform the impedance transformation and the


balanced line output very well, but there are some
considerations in selecting them for your
measurement.

There are two different types of transformers


associated with baluns, low frequency and high
frequency. Both are fairly wideband, but over different
frequency ranges. The balun should have a flat
frequency response, or 3 dB bandwidth, greater than
the frequency range of testing.

Impedance matching is usually available in a ratio or


for typical impedances, such as 50, 75 or 100 ohms.

Be careful when making physical measurements that


the balun is not affecting your measurements. A
general rule is to use baluns whose impedance is at
least ten times larger than the largest value of the
cable or ten times smaller that smallest of the cable
values that you are trying to measure. Let's examine
this rule in more detail in the next slide.

7-35
Solving Difficult
Cable Measurements

Slide #36 How do you measure the open and shorted


impedance? Why, with our analyzers!

Balun Selection Criteria ERROR: last sentence. An error ratio of ten, 20 dB,
Balun still yields a 1 d
Zbs Short

Zbs Balun's shorted impedance


Zbs < 1/10 Zcs
Zcs Cable's shorted impedance
Balun

Zbo Open

Zbo Balun's open impedance


Zbo > 10x Zco
Zco Cable's open impedance

Zco = 100 ohms Zbo > 1k ohms


Zcs = 100 ohms Zbs < 10 ohms
Microwave Instrument Division
RF & MW Device Test Seminar
0893 cabltest.pre

Measure the open and short-circuit impedance of the


balun and observe the ten times rule. The short-circuit
impedance of the balun should be as low as possible,
at least 1/10 or less than the impedance of the cable
being measured. The open circuit impedance of the
balun must be at least ten times the maximum value of
impedance of the cable.

The balun impedance must be ten times greater or ten


time less than the values you are trying to measure. By
observing this rule, you are eliminating the effects of
the balun impedance on the measurement. However, a
ratio of ten still yields a 20% error in the measurement.

For example, if you are measuring a cable which has


an impedance of 100 Ohms, you should choose a balun
that has an open-circuit impedance of at least 1 k Ohm
and a short-circuit impedance of less than 10 Ohms.

7-36
Solving Difficult
Cable Measurements

Slide #37 transition. There isn't time to explain them. The


latest proposal, June 93 , for TIA/EIA has some
Standards interesting new measurements, such as power sum.
Standards Organizations
Each of the lines in the crosstalk measurement must
TIA: Telecommunications Industry Association be measured against all the others and then the
EIA: Electronic Industries Association results combined mathematically for a worst case
NEMA: National Electrical Mfg's. Association
ASTM: American Society for Testing and Materials
condition. The new trends only lead more toward
automation and fixturing for the measurements.
Standards Specifications
UTP/STP UTP/STP UTP/STP
Product Spec's Test Spec's Test Methods STP Shielded Twisted Pair
TIA/EIA - 568 UTP Un-shielded Twisted Pair
NEMA WC - 63
ASTM 4566 - 90

Independent Product Testing


ETL, UL
Microwave Instrument Division
RF & MW Device Test Seminar
0893 cabltest.pre

Standards have been defined for the manufacture of


twisted pair cable and measurement of the associated
parameters. Here are a number of the current
standards that must be observed and their relationship
to each other. A number of measurements are required
to meet these standards.

Note that the NEMA specifications specify the entire


range of product specifications, test specifications and
the test methods. Both STP, (Shielded Twisted Pair),
and UTP, (Unshielded Twisted Pair) are included in
the specifications.

This table illustrates that there are standards in


existence which are continually evolving and in

7-37
Solving Difficult
Cable Measurements

Slide #38

Test Requirements

100 Ω UTP 150 Ω UTP 100 Ω UTP


Levels 3 - 5 Type A
DC Resistance
DC Resistance Unbalance
Mutual Capacitance
Capacitance Unbalance
Characteristic Impedance
Balanced Attenuation
Near End Crosstalk
Structural Return Loss
Common Mode Attenuation
Core Shield Resistance

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

This table shows the series of tests that must be


performed to certify that a cable meets the standards.
Note the mix of physical parameters, twisted pair and
swept frequency measurements.

Again, just to show examples of the types of tests that


are specified. Point out the mix of the three types of
measurements.

7-38
Solving Difficult
Cable Measurements

Slide #39 layout allows equal lengths of pairs from the cable to
the fixture.

Test Fixturing

100 Ohm 100 Ohm

6 7 6 7
5 8 5 8
4

3
4 9

10
3
9
25 Pair Level 5 UTP
10
2
2 11
11
Interface Test
1 12 1 12

13 25
Adaptor (ITA)
14 13
25 24

24 15 23 14

23 16 22 15

22 17 21 16
211 18 20 17
20 19 19 18

Near End Far End

Top View

Side View
Microwave Instrument Division
RF & MW Device Test Seminar
0893 cabltest.pre

A large number of connections must be made with


baluns and the loads for the cables not being tested as
well as the pair(s) being measured. Here is the test
fixture layout for a 25 pair backbone LAN cable.

Each terminal pair has an associated balun and load.


Switching and termination of the pairs is controlled by
a large switch matrix which connects the desired pair
to an LCR meter or a network analyzer for performing
the desired tests.

A circular layout geometry is desired to allow


uniformity of the cable pair lengths and the stripped
pairs for insertion into the fixture. Note there is a near
end and a far end set of terminals.

This is an example of a proposed test fixture to


connect the pairs to a switch matrix for testing, This

7-39
Solving Difficult
Cable Measurements

Slide #40 manufacturers have anted up $150K and only one


more is needed. Have the salesman contact Pete
Johnson if there is any interest from a customer in
Cable Tester Physical Layout purchasing or specifying a system.

HP 8751A Peter Johnson


HP 87512A Project Manager
System Engineering Business Unit
UTP 25 Pair Hewlett-Packard Co.
ITA
29 Burlington Mall Road
Burlington, MA 01803
STP 6 Pair
ITA
Direct (617) 221-4674
FAX (617) 221-5240
120 Ohm ITA
HP 4263A
ITA

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

The test rack is shown here to illustrate the test


equipment required and the cable connection and
fixturing. A PC is used to control the operation and
provide a data base of test results. Test that might
require weeks manually can be accomplished in a
matter of hours.

An HP 8751 network analyzer and an HP 4263 RLC


meter are the test instruments in the system.

This is a proposed system that the Application


Center in Burlington, MA is planning to build. Four

7-40
Solving Difficult
Cable Measurements

Slide #41 that can take a week to perform can be accomplished


in a few hours. Imagine all the connections and
terminations that have to be made and changed for
Cable Test Systems each test.

Throughput: measurement speed

Test matrix switching and fixturing

Automated measurements - focus on results

Database of results and trends

Microwave Instrument Division


RF & MW Device Test Seminar
0893 cabltest.pre

Manufacturers of twisted pair cable will require an


automated solution for testing their products to the
industry standards.

This test system must provide a large matrix for


switching
test instruments, loads and baluns for the cable pairs.

Automation will provide a consistent measurement


with accurate results. Much of the tedium of the
measurement is reduced and the operator can focus on
the test results, instead of which pair is being tested.

The volume of data acquired and analyzed for each


test cable is quite large. Maintaining a data base of
results to verify the quality of the manufacturing
process requires a data processing solution that can be
provided by HP.

The proposed test system features big increases in


testing throughput and data documentation. Tests

7-41
Solving Difficult
Cable Measurements

Slide #42 Summary of the section. Briefly remind the


customers of the solutions for each major problem.
Questions if time?
Cable Test Summary

Non-standard impedance
Transform and calibrate

Balanced measurements
Baluns

Low loss devices


Full 2-port error correction

Long path loss


Dynamic range
Electrical delay
Microwave Instrument Division
RF & MW Device Test Seminar
0893 cabltest.pre

Cable testing and measurement provides a number of


challenges that we discussed in this section.

Non-standard impedances may be measured in a


calibrated environment.

Baluns can provide the test interface to twisted pair


measurements

Cables as low loss devices require a full 2-port 12-term


error correction to remove the effects of impedance
mismatch with the test ports. These can be a large
source of error in low loss measurements.

The long path loss associated with testing cable


requires a test system with a high dynamic range, at
least ten dB greater than the expected result.
Electrical delay, the time for the signal to travel
through the cable, may cause measurement problems
in long cables. Slow the sweep down to compensate
for the delay.

7-42
Solving Difficult
Cable Measurements

References

Hewlett-Packard Company, Vector Seminar, Vector Baluns


Measurements of High Frequency Networks
(HP publication number 5958-0387). Hewlett-Packard Company
Model No. Impedance Freq. Range Connectors
Hewlett-Packard Company, Scalar Seminar, Scalar 16316A 50 : 100 100 Hz - 10 MHz BNC - BPs
Network Measurements (HP publication number
16317A 50 : 600 100 Hz - 10 MHz BNC - BPs
5954-1586).

Society of Cable Television Engineers Engineering


committee, Interface procedures subcommittee North Hills Electronics
recommended practices and test methods for the Model No. Impedance Freq. Range Connectors
CATV Industry, IPS-TP-007 Preliminary test method 0101BB 50 : 75 0.1 - 125 MHz BNC - lugs
for Structural Return Loss.
0300BB 50 : 100 0.1 - 100 MHz BNC - lugs
Application Notes 0400BB 50 :150 0.1 - 100 MHz BNC - lugs
And a wide assortment of other configurations
Measuring the Impedance of Balanced Cables,
Application Note 339-4 (HP publication number North Hills Electronics A Porta Systems Company
5950-2918). 575 Underhill Boulevard, Syossett, NY, USA
Tel. 516-682-7740 FAX 516-682-7704
Measuring Cable Parameters, Application Note 380-2
(HP Literature number 5950-2399).

Balanced Cable Measurement with an Impedance


Analyzer / LCR Meter / Network Analyzer, Application
Note 346-2 (HP Literature number 5091-4480).

Video Tapes

Balanced Cable Measurement,


(HP Part number 90468T).

7-43

You might also like