Install v9.10 Software
Install v9.10 Software
Purpose Install v9.10 software on an ARCHITECT Module i2000SR, ci8200, or ci16200 System
Platform F or later SCC licensed for Control Center (SCC) Platform F or
Windows 7 later licensed for Windows 7
Tools and 1. ARCHITECT System Software v9.10 Estimated Procedure: *4.00 hr.
Materials (7-20000005-02) Time Verification: 1.00 hr.
2. V9.10 Software Patch Disk Total: 5.00 hrs.
3. ARCHITECT Windows 7 Base
Configuration (7-207210-02 or above)
4. Blank CD-R compact disc (not provided)
Preparation
Action Steps Reference
Preliminary 1. This procedure provides instructions to install v9.10 software on a Refer to SCC
Information Platform F or later SCC running Windows 7. Platform
Descriptions in
2. *Estimated time includes an extra 2.5 hours for updates to Windows 7 on
Appendix A.
the SCC.
3. Read the entire procedure before beginning. It is important to follow this
procedure step by step in order to perform a successful installation.
Create Software 1. Perform the Create Software Patch Disk procedure in the section VIII:
Patch Disk PROCEDURES table of this TSB.
Note: The Create Software Patch Disk procedure only needs to be performed
once, the CD created in this procedure can be retained and be used on future
installations of TSBs 115-156, 116-128, 117-052, 126-088, 127-057, and 128-
035.
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 2 of 17
Platform F SCC
Platform F+ SCC
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 3 of 17
4. As directed on-screen:
Confirm the Edgeport is connected if required, disconnect the printer, and
then select OK.
Note: The monitor will go blank for 3 to 5 minutes during the SCC restart
process. If the “Press any key to boot from CD or DVD…” prompt is
displayed. DO NOT press a key.
5. During the Setup process, which will take approximately 30 minutes, the
system will load Windows 7 and needed drivers.
DO NOT TOUCH THE KEYBOARD, MOUSE OR TOUCHSCREEN AND
DO NOT RESPOND TO ANY PROMPTS UNTIL THE FOLLOWING
MESSAGE IS DISPLAYED (APPROXIMATELY 30 MINUTES):
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 4 of 17
Configure 1. Access the Control Panel Devices and Printers screen in Windows 7 Refer to Printer
Printer by selecting Start then Devices and Printers. Configuration in
Appendix A
2. Locate and right click on the printer to be installed and then select Printer
properties (Refer to Appendix A for more information). Rename this
printer Architect Printer.
Note: For a LEXMARK MS410dn/MS415dn printer refer to ISA ARCH-
017 LEXMARK MS410dn/MS415dn Printer to install and configure the
printer driver.
3. Right click on the Architect Printer and select Set as default printer.
Note: Once the Printer is connected, Windows 7 will hide all but the
default printer and make it more difficult to choose the appropriate printer.
If printers are already hidden, refer to Appendix A for installation
instructions.
Calibrate 1. Select Start, Control Panel, Elo Touchscreen to open the ELO Refer to P-205
Touchscreen Touchscreen application. Calibrate
Touchscreen in
2. Select the Align button and follow the onscreen instructions to perform the Service &
the touchscreen calibration. Support Manual
Note: If the Align button is unavailable, troubleshoot the touchscreen
cable.
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 5 of 17
Select Serial 1. At the Select Serial Port Settings window select one of the following. It is
Port Settings important to determine if an edgeport is present before making this
selection.
Virtual null modem port (Select this if NO Edgeport is present)
When to select this option:
• Any system that DOES NOT require an edgeport
• Do not select if the system is attached to an Accelerator track
system or is an i2000SR that has an Architect Advisor signal
tower installed
What this selection does:
• Loopback cable and physical ports are not needed to use Architect
Terminal program
• Enables virtual COM port program
• COM 3 and COM 8 will be used as a virtual loopback ports
Edgeport (Select this if an Edgeport is present)
When to select this option:
• Any system that requires an edgeport
• Select if the system is attached to an Accelerator track system
or is an i2000SR that has an Architect Advisor Light installed
What this selection does:
• Sets the CLUI_SYS.DAT file for COM 3
• Loopback cable and physical ports are needed to use the Architect
Terminal program
• The edgeport will be configured for COMs 3,4,5,6
• The 25 pin serial port connector on the SCC will be configured for
COM 7
• Disables virtual COM port program
• Allows you to use external laptop connection
2. Then select Next.
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 6 of 17
Select TSB 1. At the Select TSB Number window select the appropriate TSB number
Number from the drop down menu.
For a stand-alone i2000SR or an integrated i4000SR system, select TSB
116-128.
For an integrated ci8200 system select TSB 126-088.
For an integrated ci16200 system select TSB 127-059.
2. Then select Next.
Note: The electronic TSB tracking will only record that one TSB has been
installed. The module one product code type has been selected as the
TSB to track on integrated systems. A service ticket must be closed for
both systems.
Media 1. At the “Media Verification” window select the check box titled “Perform
Verification optional md5 verification of this disc”, and then select Install. This
(Optional) can take approximately 5 -10 minutes.
Note: This is optional and checks the integrity of the disk. It is
recommended to perform each time a new disk is used for the first time to
install software.
2. The software installation procedure will begin. A progress window will
provide information regarding the installation progress.
Note: DO NOT TOUCH THE KEYBOARD, MOUSE OR
TOUCHSCREEN DURING THE REMAINDER OF THE INSTALLATION
PROCESS! THE SYSTEM MAY REBOOT SEVERAL TIMES. WAIT
UNTIL THE INSTALLATION MESSAGES WINDOW APPEARS
(APPROXIMATELY 3.2 HOURS)!
Installation Note: An Installation Messages window will be displayed after the software
Messages installation is finished. The notes in this window will confirm the completion
(Install status of the various steps of the software installation procedure. Any steps
Completed) that failed will be displayed in red. This message window will also list any
actions (also displayed in red) that will need to be performed after the
software upgrade is completed.
1. Review this message window and make note of any post upgrade
activities messages or any installation failure errors displayed in red. Any
post upgrade activities will be performed after the v9.10 install is
complete.
If a software component did not install successfully and an installation
failure error was displayed, proceed to the next action Restart Software
Installation to recover from the failure. If there were no software
installation errors, proceed to step 2.
2. Select Next at the Installation Messages window. The final window stating
“Completing the Architect System Software Setup Wizard” indicates
installation is complete. Remove the DVD from the CD/DVD-RW drive
and then select Finish.
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 7 of 17
Restart Note: Only perform this step if software installation error messages were
Software displayed in the Installation Messages window. This step is not required if
Installation only post software upgrade activities were displayed.
(Perform only if
1. Select Next at the Installation Messages window and then at the
installation fails)
installation Wizard select Finish.
2. The SCC will reboot to the Snapshot Screen. To reload the software,
repeat this procedure by starting at the Begin Architect Software Install
action in this procedure.
3. If a software component will not install successfully after reinstalling
Architect software, perform this procedure again starting at the Install
Base Configuration action in this procedure.
Configure First Note: It is very important the Clinical Chemistry module be configured as the
Module first module when installing an integrated (ci8200/ci16200) system and also
that the correct module serial number is entered.
1. Remove the ARCHITECT System Software v9.10 DVD from the SCC
CD/DVD-RW drive
2. At the ‘Configure first module’ window press the Num Lock key to
activate the number pad and enter the FSE logon code.
3. From the Module Type pull down menu, select one of the following:
Select i2000SR if installing a stand-alone i2000SR or an integrated
i4000SR system.
Select c8000 if installing an integrated ci8200 system.
Select c16000 if installing an integrated ci16200 system.
4. Enter the Serial No. for module one of the system being installed in the
format shown in the onscreen example, and then select Done.
5. Select OK to the shutdown request. The SCC application software will
restart and show “0536 No database found….” and begin to create a new
database. This may take up to 5 minutes. DO NOT INTERRUPT THE
SCC DURING THIS PROCESS. WAIT until the database creation is
completed and the 0537 “Improper shutdown...” message appears, select
OK, and then wait for the “Log on…” window to appear.
Note: If the installation fails or corruption is suspected, refer to the
Appendix B [Option 2] Reinstall Base Configuration / ARCHITECT
System Software and restore from backup (F: or CD).
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 8 of 17
Verify UI Patch 1. In Task Manager select New Task, select or enter CMD and then select
Installation OK.
2. In the cmd.exe window type dir C:\IPLS\Controls\Release\
itxcustomlistboxexru.ocx and then press Enter.
3. Verify the displayed file attributes match the following:
Size = 243200 bytes
Date = 09.11.2015
Time = 03:41 PM
Note: The time may vary by one hour from actual time based on Daylight
Savings Time.
4. In the cmd.exe window type dir C:\IPLS\dlls\itxcustomlistboxexru.pdb
and then press Enter.
5. Verify the displayed file attributes match the following:
Size = 2,477,056 bytes
Date = 09.11.2015
Time = 03:41 PM
6. If the any of the attributes do not match then repeat the steps in the
Install v9.10 Software UI Patch and Verify Patch Installation actions.
Reboot SCC 1. In the cmd.exe windows type shutdown /r /t 0 and then press Enter to
shut down and restart the SCC computer.
2. At the Snapshot Screen, log on as FSE.
Install All M&D 1. Select System, Diagnostics, Module 5, Utilities, 6115 Install/Delete
Procedures Procedures, and then Perform F5.
2. Install All of the M&D procedures from the Hard Drive.
Note: If an integrated ci8200/ci16200 system and the ICT Module is not
to be used on the clinical chemistry processing module, perform
procedure 6115 to delete M&D procedure numbers 6305, 6306, 6019,
6300, 6062, 6063, 6309, and 2131 so the ICT maintenance items do not
show in Maintenance menus.
Additionally, if the clinical chemistry module high concentration waste
container is not to be used, then perform procedure 6115 to delete M&D
procedure 6307.B
3. When complete select Exit and then select Done.
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 9 of 17
System Configuration
Action Steps Reference
Configure 1. Select System, Configuration, System control center, and then Refer to the
System Settings Configure F6. ARCHITECT
System
2. Enter the System date and System time, and then select the desired
Operations
Date format.
Manual
3. Open the Time zone pull down list and select the local Time Zone.
4. If desired, uncheck the Automatically adjust clock for daylight
savings: check box.
5. Enter the ARCHITECT system no. and SCC serial no. These numbers
can be found on a label on the front of the SCC.
6. Set the Screen timeout to a value greater than 0 to assure that the
screen saver is activated.
7. Open the System language pull down list box; select the desired System
language, and then select Done.
8. If a language other than English was selected, then “A new system
language has been selected. The SCC will shutdown.” dialog box will
appear, select OK to confirm the SCC shutdown.
9. If the system rebooted due to a System language change then Log back
on as FSE.
Configure Note: These steps only need to be performed for a standalone i2000SR
Optics Values or an integrated i4000SR system.
(Standalone
1. Select System, Configuration, Modules, and then Configure F6.
i2000SR or
Integrated 2. Enter the Optics Normalization and Linearity values from the label on
i4000SR only) the i2000SR Optics Assembly, and then select Done.
3. Select OK at the Warning: The optics values have been edited dialog.
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 10 of 17
Configure RSHx 1. Select System, Configuration, Modules, and then select F6 Configure.
module
(Integrated 2. Open the Module number pull down list, select New Module, and then
ci8200/ci16200 select RSH Extension.
attached to a 3. Enter the Number of RSH Extension priority sections desired and
p540 Aliquoter select Add.
Only)
4. Select Done and then select OK to confirm the system reboot.
5. Log on as FSE.
Confirm Serial 1. Select System, Configuration, System control center, and then, and
Numbers then select Configure F6.
2. Verify the System no and SCC serial no match those printed on the
label located on the front of the SCC computer.
3. If the System no or SCC serial no do not match, correct the values and
then select Done.
4. Select System, Configuration, Sample handler, and then select
Configure F6.
5. Confirm the configured serial number matches the serial number on the
sample handler, correct as needed, and then select Done.
6. Select System, Configuration, Modules, and then select Configure F6.
7. Select each Module in the Module drop down box and confirm that the
serial number displayed matches the serial number on the Module. If any
of the serial numbers do not match, correct the serial number.
8. Select Done. If a processing module serial number was changed then
select OK to confirm the system reboot.
Note: The base configuration and Architect software must be reloaded to
reconfigure the Clinical Chemistry instrument as the first Module if the
i2000SR is mistakenly configured as module 1 on integrated ci8200/ci16200
system.
Configure and 1. If a hand held bar code reader is installed, use the Bar Code Scanner Refer to P-228
Verify Hand Guide to configure the reader and then perform procedure P-228 to SCC Bar Code
Held Barcode verify operation. Scanner
Scanner Verification in
the
ARCHITECT
i System
Service and
Support Manual
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 11 of 17
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 12 of 17
Perform Note: Only perform M&D 6008 if error “9838 Call Abbott. CPU firmware
Controller configuration does not match application firmware.” is displayed.
Configuration
Note: Do not perform the M&D for multiple modules and then cycle the power
on All Modules
once for all modules. This could cause the CPU board to lockup and
(If Needed)
the processing module will not come Online. If this occurs cycle power
to the SCC.
WARNING: DO NOT interrupt module power or perform any system functions
until the Controller update has completed. Interrupting power will cause
a corruption of the Controller firmware. Corrupted firmware cannot be
repaired and the controller board will need to be replaced. Do not cycle
power to the c8000/c16000 (if present) while performing this procedure.
1. Log on as FSE and with the all processing module(s) and RSH at the
Ready or Stopped state, select System, Diagnostics, Module 5,
Utilities.
2. Perform M&D 6008 Controller, enter 0 for the RSH module, and then
select Continue. Cycle power to the i2000SR as directed in the M&D to
complete the firmware update.
If a stand-alone i2000SR:
3. Perform M&D 6008 Controller, enter 1 for the i2000SR module, and then
select Continue. Cycle power to the i2000SR as directed in the M&D to
complete the firmware update.
If the system is an integrated ci8200/ci16200:
4. Perform M&D 6008 Controller Configuration, enter 2 for the i2000SR
module, and then select Continue. Cycle power to the i2000SR
processing module as directed in the M&D to complete the firmware
update.
If the system is an integrated i4000SR:
5. Perform M&D 6008 Controller Configuration, enter 1 for the first
i2000SR module, and then select Continue. Cycle power to the first
i2000SR processing module as directed in the M&D to complete the
firmware update.
6. Perform M&D 6008 Controller Configuration, enter 2 for the second
i2000SR module, and then select Continue. Cycle power to the second
i2000SR processing module as directed in the M&D to complete the
firmware update.
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 13 of 17
Install Assays 1. Before installing any assays, select the preferred language in System,
for Verification Abbott Mail, Options, Download language. If assays are loaded,
Architect software will attempt to update the Assay insert in the currently
selected language. If the language is changed after the first update, Error
Code 9518 will appear in the message history every time a new reagent
lot is used. There is no way to correct this at present. System operation is
not affected.
2. While logged on as FSE, select System, Diagnostics, Module 5,
Utilities.
3. Perform M&D 6114 Install/Delete Assays.
4. Install the following assays to be used for performance verification testing;
1 Step, 2 Step, B-hCG and TSH for the i2000SR. For a ci8200/ci16200
integrated system, reference P-159 (P-547 for ci16200) Performance
Verification Test (PVT).
5. For an integrated ci8200/ci16200, from the snapshot screen select
System, Diagnostics, Module 1, Utilities, and then perform M&D 6117
Remove Trig Cuvette SmartWash.
Perform Module 1. Attach the ARM and LIS cables at this time (if present).
Preparation and
2. If connecting to a Laboratory Information System (LIS), configure the Host
Calibrations
Query and Timeout parameters. (Obtain from customer)
3. If an ARM is present, configure the ARM parameters.
4. As instructed in the appropriate Architect Service and Support Manual
Installation instructions, complete the module preparation and robotics
calibrations.
THE FOLLOWING ACTION ITEM SHOULD BE PERFORMED AFTER CALIBRATING ALL MODULE
HARDWARE (ROBOTICS). IF AN INTEGRATED ci8200/ci16200 SYSTEM, ENSURE THAT DIAGNOSTIC 6001
WAS PERFORMED AS DETAILED ABOVE, BEFORE CONTINUING.
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 14 of 17
Set Bar Code 1. Perform option 4 of Diagnostic 3253 (System, Diagnostics, Module 5,
Transitions Utilities) to set the bar code transition value depending on the type of bar
code the customer will be using.
2. As necessary, perform the other options for Code 39 or Codabar large
intercharacter and Codabar start and stop character gap.
APS only – 1. If this module is connected to an APS track, update the following For more
Update parameters in the LASM_SYS.CAL file (located in the information see
LASM_SYS.cal C:\IPLS\icw\system directory). the latest
version of TSB
Note: This step 2. Modify or verify that the following 6 values are set as indicated.
743-004
only applies to
APS Systems a. Set Reagent Allocation Completion Code value to “0”.
with ARCHITECT b. Set Consumable Allocation Completion Code to “0”.
Analyzers at c. Set the Sample Integrity (LLS/Step Loss) Completion Code
software versions value to “3”.
8.1 or higher d. Set IA_Module to “True”.
e. Set CC_Module to “True”.
f. Set STAT BYPASS to “False”.
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 15 of 17
3. Reboot each Analyzer that had the LASM_SYS.CAL file modified prior to
attempting any system testing.
4. After reboot, open the LASM_SYS.CAL file again and verify that the
changes made are still in place.
Install Note: If this procedure has been performed to reinstall v9.10 software,
Overwritten previously installed software updates may be overwritten and must be
Software reinstalled.
Upgrades
1. Log-on as FSE at the Snapshot screen.
(if applicable)
2. Select the Updates icon in lower right-hand side of the snapshot screen.
3. Install all software upgrades (TSBs) displayed in the Available updates
window. Refer to the associated TSB number for detailed installation
instructions.
Note: If the hard drive has been replaced or D: partition deleted/corrupted,
updates may not be available. Refer to ISA/TSB database identify any
TSB software upgrades that should be reinstalled.
Review System 1. Review TSB database and perform all other applicable TSBs.
TSB Level
Create a New 1. From the Snapshot screen, select System, Utilities, and then Create ARCHITECT
System Data backup F4. System
Base Backup Operations
2. In the comment field, type After v9.10 (current date) and then select
Manual
Done. The system will create a new backup of the module calibration
files, system configuration, and database.
3. Select Exit when done.
Copy Backup Note: The following steps will create a CD safe backup to be used in the
Folder to CD – event of an unrecoverable system crash during the upgrade. If disaster
For Disaster recovery is necessary, refer to Appendix B of this TSB.
Recovery
1. Obtain a Blank CD-R compact disc (not provided).
2. Select System, Diagnostics and then select Module: 5 for the SCC.
3. Page down and select Utilities.
4. Select 6004 Copy Backup Software, Perform F5, and then select OK.
5. Select Proceed to continue.
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 16 of 17
Update TSB 1. Mark an X through the number 128 on each i2000SR processing module
Modification modification label.
Label
2. If an integrated ci8200 system, mark an X through the number 88 on the
c8000 processing module modification label.
3. If an integrated ci16200 system, mark an X through the number 59 on the
c16000 processing module modification label.
Note: Follow local procedures to ensure the TSB information is entered into
the Call Management System for each completed TSB (refer to TSB
Implementation and Coding Instructions section).
Verify 1. Using the touchscreen, move through at least four system screens to
Touchscreen verify touchscreen operation.
Calibration
Verify Printer 1. Perform P-207 ARCHITECT Print Test, in the appropriate Architect
Operation Service & Support Manual.
Verify 1. Verify SCC is communicating with the ARCHITECT ARM (i.e., actual
ARM/iARM ARM status should match what is displayed in the ARM status screen,
Operation etc.).
(if present)
For New 1. For new installations, after all Verification steps are performed, proceed
Installation - with system installation as directed in the Installation section of the
Proceed With appropriate Architect Service & Support Manual.
System
Installation
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB 116-128A ARCHITECT v9.10 Software Install v9.10 on a new Architect system Page 17 of 17
Completed
Items to Perform Comments
Preparation
Preliminary Information
Prepare SCC
Identify SCC Type
Install Windows Base Configuration
Install Windows 7 Base Configuration
Configure Printer
Calibrate Touchscreen
Install ARCHITECT Software
Install Architect Software
Install v9.10 Software UI Patch
Install M&D Procedures
Enter Key for Premium Features (Optional)
System Configuration
Configure System Settings
END OF DOCUMENT