L2. Quantitative Analysis
L2. Quantitative Analysis
Basic concepts
Dr. Le Hong Luyen
Department LS, USTH
Email: [email protected]
1
Content
1. Introduction
2. Criterion of QA
3. Properties of QA
4. Error resources
5. Validation method
6. Sample preparation
2
1. Introduction
• Qualitative analysis simply determines if an
analyte is present of not in the sample
• E.g.: Test for illicit substances in an athlete
3
1. Introduction
• Quantitative analysis answers the question:
4
1. Introduction
• A quantitative analysis is generally a targeted analysis
• One or a very limited number of compounds are quantified in a
sample
• For each quantitative analysis, we normally know:
• Range of concentration of the analyte
• Specific matrix (plasma, blood, urine…)
• Developed method
5
2. Criterion for a good quantitative analysis
•An analytical method is good only for
• A specific analyte: each type of compound must follow a specific method of analysis
• A specficic matrix: A perfectly good method for one matrix may not be acceptable for another
• A specific species: same molecular formula but different stereostructures will need different
analytical methods
• A specific range of concentrations: ‘’validation range’’
•Several criterion:
• Accuracy
• Precision
• Sensitivity
• Specificity
6
2.1. Accuracy
7
2.2. Precision
•
8
2.2. Precision
How to compare the precision of 2 sets of measurements?
• Sample A: 32.56, 32.55, 32.48, 32.49, 32.48.
• Sample B: 15.38, 15.37, 15.36, 15.33, 15.32.
🡪 Measure the range (the difference between the highest and lowest
scores).
9
2.2. Precision
How to compare the precision of 2 sets of measurements?
• Sample A: 32.56, 32.55, 32.48, 32.49, 32.48.
• Sample B: 15.38, 15.37, 15.36, 15.33, 15.32.
🡪 Measure the range (the difference between the highest and lowest
scores).
11
Difference: accuracy and precision?
• The level of accuracy and precision shows whether or not quantitative
analysis is sufficient for the measurement of a particular analyte.
12
Difference: accuracy and precision?
The shooter is aiming for the right corner of the back of the net.
Which one of the following shows:
- high accuracy and high precision?
- low precision but high accuracy
A B
C D
13
2.3. Sensitivity
• Sensitivity: the ability to distinguish the smallest difference that we can
measure between two signals.
What is needed is the right sensitivity for the range of concentrations measured.
14
2.4. Specificity
• A method is specific if it measures only the analyte targeted.
• No interference from other compounds.
• For example:
• A method that cannot differentiate between a drug and its metabolites is not
specific.
• Weighing may be accurate, reproducible and sensitive but is not specific
because the mass of a sample does not tell you if the sample is pure or not.
15
False positive/ False negative result
• A method can give a false positive result if it is not specific. It means
that the signal is not due (entirely or partially) to the analyte.
16
False positive/ False negative result
• False negative: When the test says you don’t have it but you actually
have it.
• False positive: When the test says you have it but you actually don’t
have it.
17
Signal vs Concentration
• A signal (S) in a quantitative analysis is an experimental measurement that is
a function of concentration (c) of the analyte
• S = f(c)
• An analytical method gives a total signal S.
• S = f(c) + N
• N: noise
• The noise has to be subtracted to have accuracy on the response due to the
analyte
18
Signal vs Concentration
The signal may be: Absorbance (Spectroscopy), peak height or area (Chromatography),
intensity (Mass spectrometry)…
19
Signal vs Concentration
Linear
Curve S = f(c)
Exponential
Logarithmic
20
Limit of detection
• The limit of detection (LOD) is defined as the lowest quantity
of a substance that can be detected with confidence.
• It is usual assessed by the measurement of the signal/noise
ratio (S/N).
• There is no formal value but we may reasonably say a peak is
a peak if the S/N ratio ≥ 3.
21
Limit of quantification
• Lower limit of quantification (LLOQ): the lowest
concentration that can be quantitated with acceptable
accuracy and precision (≤ 20% CV).
• S/N ≥ 5-10 (but normally S/N ≥ 10)
23
Calibration curve
When we have a calibration curve we determine the concentration ci from the
signal Si obtained with the sample i. However, this determination can be valid only
if LLOQ < ci < ULOQ.
ULOQ
LLOQ
Increasing concentration
24
Quality control sample
• In order to assure that a test run is valid and results are reliable,
Quality control (QC) samples should be used in the performance of
each assay.
• QC samples should be treated in the exact same manner as the test
samples
• QC samples are samples in the same matrix with known
concentrations: LLOQ < QCs < ULOQ
• QC samples are processed side-by-side with test samples
25
The standard addition method
• Matrix effect or matrix interference: components in the matrix affect the
analyte’s signal (e.g.: endogenous matrix…)
🡪 Couldn’t use the traditional calibration curve to determine the analyte’s
concentration.
signal
Analyte in matrix in equal volumes
27
External vs internal standard
• External standard (ES) is not present in the sample.
• It is a standard solution containing a known amount of analyte, prepared and
analyzed separately from samples containing the analytes.
• Usually used for quality controls
• Rather it is run alone, as a sample, and usually at different concentrations, so you
can generate a standard curve.
• External standards do not correct for losses that may occur during preparation of
the sample, such as extraction, centrifugation, evaporation, etc.
28
External vs internal standard
• An internal standard (IS) is added in a constant amount to samples.
• A compound that must be show similar behaviour to the analyte.
• To correct analyte losses during sample preparation (extraction, centrifugation,
evaporation…)
• This substance can be used for calibration by plotting the ratio of the analyte
signal to the internal standard signal as a function of the analyte standard
concentration: S=f(ca/cIS)
• The internal standard used needs to provide a signal that is similar to the analyte
signal in most ways but sufficiently different so that the two signals are readily
distinguishable by the instrument.
29
What is a good internal standard?
• Structure similar to the analyte
• Never found in the sample (to keep theirs concentration is constant)
• Well resolved
• Stable (If it is isotopically labelled)
• Pure enough not to interfere
• Same behavior with the analyte: extraction, chromatography,
response
30
What is a good internal standard?
• The Internal Standard corrects for many things but it also adds to the variability
(in reality, you cannot add a same amount of IS to every samples).
• Error (As/ IS) = error Aa + error IS
• The concentration of the IS should be appropriate so that it introduces as little as
possible an additional error.
31
Peak height vs peak area
• Peak height is proportional to the concentration, but it varies with the chromatographic
conditions.
• In fact, peak area is preferred if a minimum of rules are respected.
• Peak area is defined as the AUC (area under the curve).
32
Peak height vs peak area
• The baseline selection may affect the value.
• When the peaks are not well resolved, the results may
be drastically different.
• The peak area represents the amount of an analyte, so if
the retention time changes, the AUC doesn’t change.
33
Peak area: unresolved peaks
34
Linear relationship: Regression analysis
• We take the simple case of a regression analysis: linear relationship.
• We try to fit a straight line with the data points (this line is close to most of data
points),
• We get a function y = ax + b with the parameters a and b which minimize the sum
of the square (yi – y)2.
35
Linear relationship: Regression analysis
• Practically 5-6 data points are sufficient
• The goodness of fit of the regression line can be assessed by
calculating the square of coefficient of correlation R2
• Perfectly, R2 = 1 but we never get 1.
• R2 depends on the variability, is not a good judge of linearity.
• R2 is sometimes used to compare 2 quantification methods.
36
Linear relationship: Regression analysis
• Anscombe’s quartet is 4 datasets with 11
data points:
• The same coefficient of correlation
• The same regression line
but showing clearly different graphs.
37
Outlier ?
• What is an outlier?
• An outlier is a data point that significantly differs from the
other data points in a sample
• Causes of outlier:
• Human mistake
• Instrument malfunction
38
Outlier: To remove or to keep?
39
Outlier
•How to Calculate Outliers?
• Biostatistics
• Or see more details on
http://www.wikihow.com/Calculate-Outliers
40
Method validation
• A method is the application of a technique to a specific analyte in a specific
matrix
• Before we run the actual samples we have to validate the method.
• We need to check the performance of the method with the quality control samples (QC
samples) and a simple statistical analysis.
• Expression of the results: The results should be given with only significant numbers.
With the uncertainty, we have to write: +/- SD, CV. Results expressed usually as a
concentration.
41
Method validation
• Criteria for validation:
• Interval of concentration: LLOQ, ULOQ
• Accuracy within the accepted value (e.g. 15%)
• Precision: within the accepted value (e.g. 15%)
• Specificity
• Sensitivity
• Reproducibility
• Stability of the sample before and during the analytical phase.
42
Sample preparation
Protein precipitation
Liquid-liquid extraction
Liquid-solid extraction
43
Introduction
• Sample preparation is an essential part of most bioanalytical methods
• Biological samples: not introduced directly into analytical instruments
• Commonly applied before LC-MS methods
• Main issues when analyzing biological samples:
o The analytes are in low concentration
o Unwanted compounds can contribute to the noise and suppress the signal of
the analytes
o Interferences can contaminate/damage the instruments
44
Goals of sample preparation
• To remove interferences for:
o Better instrumental method (chromatography)
o More confident analytical results
o Longer column lifetime
o Less instrument downtime
• To enrich sample for:
o Higher detection sensitivity
• To make your analytical lab more productive:
o Lower limits of detection
o Run more samples with less time
o Minimize costs in equipment maintance
45
1/ Protein precipitation
• Two opposite goals:
• To remain proteins for future analysis (e.g. proteomics)
• To remove proteins for regular bioanalysis
• In 1st case: the precipitation should be gentle and not to alter the
protein structure
• In 2nd case: don’t care what happen to the proteins. The protein may
be denaturated
46
1/ Protein precipitation
• The most used biofluid is plasma from human or a variety of animal
species containing high protein content
• Protein precipitation technique used before injecting samples
• Rapidly performed
• Simple equipment and reagents
• Good extraction recovery
• Remove 90-98% of the proteins from the samples
47
1/ Protein precipitation
• 4 main approaches to precipitate plasma protein:
• Organic solvents: methanol, ethanol, acetonitrile (the most used, 1:5 v:v)
• Acids: trichloroacetic acids (TCA), perchloric acid (HClO4)
• Metal ions: Zn(OH)2, copper sulphate-sodium tungstate (Na2WO4)
• Salts (less efficient): (NH4)2SO4
• After precipitation, removing the protein by centrifugation
(10,000-15,000xg) or filtration (using special membrane) to leave a
clear supernatant used for analysis
48
2/ Liquid-liquid extraction (LLE)
• The analyte (drug) usually in aqueous phase (plasma, urine) is
transferred into a water-immiscible organic solvent
• The partitioning depends on:
• The lipophilicity of the analyte
• The pH of the aqueous phase
• The nature of the organic solvent
• The relative volume of the two phases
49
2/ Liquid-liquid extraction (LLE)
• Between two immiscible phases (aqueous-organic phase)
51
Properties of organic solvents
52
Polarity of solvents
53
Acid-base extraction
• Useful for isolating and purifying amines, carboxylic acids and
phenols.
• All these functional groups can be interconverted from non-ionic
organic-soluble forms to water-soluble ionic forms by changing the pH
54
Acid-base extraction
• Ladder diagram is useful
• Ex: Phenol (pKa = 10). When extracted at pH = 7.4 (plasma)
55
Back extraction
• We may use the acid-base form to selectively back-extract a mixture
from an organic phase into an aqueous phase
• The charged specie is more soluble in water phase
• Example:
56
Back extraction
57
Distribution ratio
• At a certain temperature, the ratio of concentration of the solute between
two phases is constant
• For a given molecule, this ratio is dependent on the nature of solvent 1
(biological fluid or water phase) and solvent 2 (organic phase)
• Higher D value, higher solute’s extraction from organic solvent
58
Extraction yield
(= D)
(= D)
59
Effect of volume of the solvent 2
Case 1
D • The distribution ratio is constant
• But MORE solute is extracted with a
larger volume of solvent
Case 2
D
60
Effect of repetitive extraction
Question: Which give the better extraction (D=2)?
- Case 1: extraction 1 time with 1 volume of solvent 2
Case 2
- Case 2: extraction 1 time with 2 volume of solvent 2
- Case 3: extraction 2 time with 1 volume of solvent 2
Case 3
Extraction yield in each case?
61
Effect of repetitive extraction
Question: Which give the better extraction (D=2)?
- Case 1: extraction 1 time with 1 volume of solvent 2
Case 2
- Case 2: extraction 1 time with 2 volume of solvent 2
- Case 3: extraction 2 time with 1 volume of solvent 2
Case 3
- Case 1: % extracted = 66.6 %
- Case 2: % extracted = 80 %
- Case 3: % extracted = 89 %
AH B
AH A - + H+ BH+ B + H+
Acid Base
Water phase (1) Water phase (1)
63
How to calculate extraction yield
Weak acid Weak base
Organic phase
AH
AH A - + H+
Acid
Water phase
BH+ B + H+
Base pH = pKa - 1
Water phase
64
Partition coefficient P and Distribution ratio D
• In medicinal chemistry, we usually use logP
• P is the partition coefficient between n-octanol and water of an
uncharged molecule
• P is obtained at equilibrium and room temperature
• Relationship between logD (for ionized species) and logP (for
unionized species)
65
3/ Liquid-Solid extraction
• The solutes in aqueous phase are retained on a solid support
o Usually a column, or cartridge, flat disk
o Then eluted with an appropriate solvent
• Usually we want to remove from
o Salts
o Proteins
o Polar endogenous compounds
66
Principle
4 main steps:
• Conditioning the column using an
appropriate solvent
• Adding sample: percolation of the
samples through the solid sorbent
• Washing impurities
• Eluting the analyte
67
On-line SPE
• SPE is easily coupled with an HPLC or HPGC
system
• Beneficial when the amount of solutes is limited
• Advantages:
o Sensitivity,
o Reliability
o Speed
69
Advantages of SPE
• Its main advantages:
o Amenable to automatization
o Small volume of solvents used
o Many choices of sorbents for specific needs
o Efficient and rapid
o Preventing emulsion and disposal of large quantity of organic solvents
• Inconveniences:
o Applied for relatively small volumes of solvent
70