Manual
Manual
Nicolet iS5
User Guide
Microsoft, Windows and Excel are either trademarks or registered trademarks of Microsoft Corporation in the
United States and/or other countries. All other trademarks are the property of Thermo Fisher Scientific Inc.
and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error-
free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
For Research Use Only. This instrument or accessory is not a medical device and is not intended to be used
for the prevention, diagnosis, treatment or cure of disease.
Contents
Chapter 1 Welcome. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Conventions Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
About the Help System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Finding Other PDF Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Ordering Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Contacting Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Chapter 2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Electrical and Communication Cabling. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Turning the Power On and Off. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Installing an Accessory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Installing Integrated Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Installing Baseplate-mounted Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Installing Slide-mounted Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Collecting Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Setting the Purge Gas Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Aligning Your Spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Verifying the Laser Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Using the iD1 Transmission Accessory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Installing Samples in the iD1 Transmission Accessory . . . . . . . . . . . . . . . . . . 19
Collecting Data With the iD1 Transmission Accessory . . . . . . . . . . . . . . . . . 19
Entering Serial Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
Welcome
The Thermo Scientific Nicolet™ iS™5 Fourier transform infrared (FT-IR) spectrometer lets
you perform chemical analyses of sample materials by collecting data in the mid-IR spectral
range using a variety of accessories, including transmission, attenuated total reflection (ATR),
diffuse reflection (DRIFTS), and specular reflection (SR). The system has integrated
verification features, a powerful software suite, and many other features that make data
collection easy. You can install optional hardware and perform several service and
maintenance procedures yourself. This document, or other provided documentation, contains
the information you need.
Be sure to read the safety guide that came with the system before operating the instrument.
Note We recommend that you maintain seal and desiccation and/or purge your
instrument at all times. Equipment damage due to failure to maintain seal and desiccation
and/or purge is not covered by the warranty. If you have questions about this requirement,
please contact us.
Conventions Used
Safety precautions and other important information use the following format:
WARNING Indicates a hazardous situation which, if not avoided, could result in death or
serious injury.
CAUTION Indicates a hazardous situation which, if not avoided, could result in minor or
moderate injury.
NOTICE Follow instructions with this label to avoid damaging the system hardware or
losing data.
Ordering Parts
To order parts, contact us.
If you need to send the instrument or an accessory to us for repair, call or e-mail us first for
any shipping requirements or other instructions.
Contacting Us
For Technical Support, please contact: www.thermofisher.com
If you have problems with your system, refer to the Troubleshooting information for help, or
call Technical Support.
Note To expedite and assist in problem diagnosis, please be prepared to send us a copy of
your diagnostics log file. The diagnostic log includes a history of important system
parameters and events that occur during system operation. To copy the diagnostics log to
your Windows desktop, start OMNIC software and choose File > Copy Diagnostics File
to Desktop. Attach this file to an e-mail message to our Technical Support group.
Operation
This section provides instructions for setting up your instrument and installing accessories.
Contents
• Electrical and Communication Cabling
• Turning the Power On and Off
• Installing an Accessory
• Collecting Data
• Setting the Purge Gas Controls
• Aligning Your Spectrometer
• Verifying the Laser Frequency
• Using the iD1 Transmission Accessory
• Entering Serial Numbers
Your instrument is designed for indoor use at altitudes up to 2,000 m (6,500 ft). It operates
reliably at a constant temperature between 15 °C and 35 °C (60 °F and 95 °F). Temperature
changes may result in drift in the system response.
NOTICE Your instrument must be protected from excessive humidity, since the
beamsplitters, detectors, and other components may corrode with exposure to moisture.
Maintain humidity in the range of 20% to 80% noncondensing. To accomplish this, this
instrument is sealed and desiccated. For additional humidity protection, the optional
purge configuration and/or zinc selenide (ZnSe) windows can be used.
CAUTION If a protective cover on a system component appears damaged, turn off the
system and secure it against any unintended operation.
Power
Computer
Computer (USB)
cable
Power cable
Check the Preparing Your Site guide on your documentation disk for electrical and other
system requirements before you connect the power cord. The wall outlet must be equipped
with a 3-wire line: live, neutral, and ground.
WARNING
• To ensure a good ground connection and avoid shock hazard, do not use an outlet
that has ground connected to a conduit ground. The ground must be a
non-current-carrying wire connected to earth ground at the main distribution box.
• Do not block access to the power supply, AC power cord or power switch.
Power switch
When the system is powered on, the blue power indicator on the main cover flashes while the
instrument warms up and performs diagnostics (about 2 minutes) and then the indicator stays
lit. During initialization, OMNIC software displays the number of seconds remaining and the
Collect menu is disabled; the system cannot collect data until initialization is completed.
Power indicator
NOTICE Do not place anything on top of the vents at the back of the instrument.
Electronic components can overheat if these vents are blocked.
Installing an Accessory
The Nicolet iS5 spectrometer includes the iD1 transmission accessory which accommodates a
wide variety of samples and slide-mount sample holders for transmission experiments.
Integrated iS5 accessories for common ATR analyses are also available. Contact us for details.
Mounting baseplates are available to accommodate a wide variety of standard accessories from
us and other manufacturers.
All of our integrated accessories and baseplates are easy to install and remove and are
recognized automatically by the software.
Contents
• Installing Integrated Accessories
• Installing Baseplate-mounted Accessories
• Installing Slide-mounted Accessories
WARNING Prevent fire and explosion. The infrared source inside the instrument is an
ignition source. If you are using volatile solvents, provide a fume hood or other active
venting system that is free of spark and other ignition sources and prevents flammable
vapors from collecting in the atmosphere surrounding the instrument.
These measures will help prolong the life of your instrument and will eliminate the possibility
of spectral interference caused by volatile solvent vapors.
Using solvents that may produce HCl or HF vapors in the sample compartment may severely
damage the system. If you use halogenated solvents, purging the instrument with dry, clean air
or nitrogen is strongly recommended. See the Preparing Your Site guide for a details.
CAUTION Materials such as hydrochloric acid, hydrofluoric acid and phosgene are highly
toxic. If you plan to regularly use solvents containing halogenated hydrocarbons, be sure
your work area is properly ventilated.
The accessory fits over two alignment pins on the spectrometer baseplate and is held in place
by magnets.
Magnets
Pins
If your instrument is purged, attach the purge line labeled “Accessory” to the purge inlet on
the back of the accessory. See Installing a Purge Kit for more information.
Purge inlet
Note To remove an accessory, disconnect the purge line if necessary and then grasp the
accessory by the handles and lift it up. When not using an accessory, store it in a dust-free
environment like a cabinet or box.
iD Base Adapter
The iD Base Adapter provides a place to mount large transmission accessories and other
specialized accessories.
Note Avoid installing the iD Base Adapter without an attached accessory. Very strong
magnets hold the adapter in place; you may need to use a tool to pry it free.
For slide-mounted accessories that are too big for the iD1 Transmission Accessory, a slide
mount that fits the iD base is available.
Slide mount
The iD Base Adapter has two sets of screw holes so the slide mount can be installed facing
either direction. To install the slide mount, position it over the screw holes, and then insert
and tighten the two washers and screws.
Use the large center screw to raise or lower a sample or accessory to align it in the infrared
light beam.
To install an accessory attached to the iD Base Adapter, remove any accessory from the iS5
spectrometer and place the adapter over the pins on the spectrometer baseplate. Magnets hold
the adapter in the proper position.
Here is the iD Base Adapter and attached slide mount with a cuvette holder installed in the
Nicolet iS5 spectrometer.
iD Foundation Adapter
The iD Foundation Adapter fits our Foundation Series Swap-Top Modules. For information
about these modules, see the Foundation Series User Guide on your documentation disk.
Tighten screws to secure
the Foundation accessory
To install the adapter, remove any accessory from the iS5 spectrometer and place the adapter
over the pins on the spectrometer baseplate. Magnets hold the adapter in the proper position.
Slide any Foundation Series Swap-Top Module into the holder, rotate the accessory to line up
the notches and pins and then hand tighten the screws (see the image above). For diffuse
reflection accessories, remove the two screws and lower panel to accommodate the sample
slide as shown in the image below.
Here is a Foundation Multi-bounce ATR accessory installed in the Nicolet iS5 spectrometer.
1. Slide the accessory into one pair of slots in the iD1 Transmission Accessory.
Choose a pair of slots that will place the center of the accessory at the beam focus
(designated by arrow heads on the accessory housing).
Beam focus
Collecting Data
For information about collecting data with your instrument, see the provided Nicolet iS5
Getting Started and Fast Facts manuals.
For best results the purge gas should be dried to a dew point of -70 °C (-94 °F) or below.
WARNING Never use a flammable, combustible, or toxic gas to purge this instrument.
The purge gas must be free of oil and other reactive materials. Heat from the source or
from laser absorption may ignite flammable gases or reactive materials in purge gas. Use
only dried air or nitrogen to purge your instrument.
1. Open the main valve to start the flow of purge gas through the regulator.
2. Adjust the pressure regulator for the spectrometer until the gauge indicates that the
pressure is 5 psig (34 ).
Set accessory pressure Set spectrometer pressure
NOTICE Flow rates greater than recommended values can cause vibration which can
affect data quality.
Note The instrument is aligned automatically each time you run a Performance
Verification (PV) test. We recommend running a PV text at least once a month.
Use the Align button on the Diagnostic tab of the Experiment Setup dialog box (available
through the Collect menu in OMNIC software). For more information, see OMNIC Help
Topics in the Help menu.
For optimum system performance, verify the laser frequency according to the following
schedule:
Note The laser is verified automatically each time you run a Performance Verification
(PV) test. We recommend running a PV text at least once a month.
1. From the OMNIC software main screen, choose Collect > Experiment Setup.
2. Click Laser Verification on the Diagnostic tab.
3. As prompted, remove all accessories and samples from the instrument and then click OK.
4. Click Start Verification.
When the verification is completed, one of the following occurs:
• A message indicates the laser frequency has been verified. Click OK to finish.
• The newly measured laser frequency appears below the previously measured
frequency and a message indicates the laser frequency has been optimized. Click OK
to finish.
Beam focus
The wide slot provides clearance for liquid and gas cell flanges.
1. Choose Advanced Diagnostics in the Collect menu. The Diagnostics screen appears.
Verify that all components are working properly. (If a component malfunctions, a red
slash mark appears over the corresponding icon on this screen. See Troubleshooting or
contact us for assistance.)
2. Click the Installation icon. An information screen appears.
Update any pertinent information, enter today’s date and then click Save.
Maintenance
This section provides instructions for installing and replacing hardware and maintaining your
instrument.
WARNING When performing any of the maintenance procedures, always follow the
safety precautions included in the user guide or Help system and in any of the other
documentation that came with your instrument.
Contents
• Installing or Replacing Hardware
• Maintaining Your Instrument
WARNING The following parts must be replaced only with replacement parts
supplied by us:
• Infrared source
• Desiccant
• Sample compartment windows
• Power supply
Set up needed:
• Source of dry air or nitrogen that meets Thermo Scientific specifications for the
Nicolet iS5 spectrometer. See the Preparing Your Site manual on your documentation
disk for more information.
Note For best results dry the purge gas to a dew point of -70 °C (-94 °F) or below.
Tools needed:
• A 3/4-inch open-ended wrench
• An 11/16-inch open-ended wrench
• A #2 Phillips head screw driver
• Purge regulator valve (with a 1/4-inch male or 3/8-inch female fitting)
• Pipe tapes
• Purge kit with sealed desiccant packet (open just prior to installation)
WARNING Never use a flammable, combustible, or toxic gas to purge this instrument.
The purge gas must be free of oil and other reactive materials. Heat from the source or
from laser absorption may ignite flammable gases or reactive materials in purge gas. Use
only dried air or nitrogen to purge your instrument.
1. Install the regulator valve and either a 1/4-inch male fitting or a 3/8-inch female fitting
on the purge gas source. (Choose a valve and fittings that are appropriate for the purge gas
source.)
3. Wrap the reducing nipple or the 1/4-inch male fitting with Teflon pipe tape, and then
install the pressure coupling. Use a 3/4-inch open-ended wrench to tighten the
connection.
4. Snap the male inlet of the wall plumbing assembly firmly into the pressure coupling.
Refer to this drawing when performing this and subsequent steps:
Pressure coupling
To accessory To spectrometer
5. Turn off the instrument power, unplug the power cord and remove all accessories and
samples from the system. See Turning the Power On and Off and Installing an Accessory
for instructions.
WARNING Before opening the desiccant compartment, turn off the instrument
power, unplug the power cord and remove all accessories and samples from the
system. Entry of flammable liquids or gases into the desiccant compartment is an
explosion hazard. If such entry occurs, contact us immediately and do not apply
power to the instrument until the condition has been corrected.
6. Loosen the two captive Phillips head screws from the standard desiccant cartridge, and lift
the cartridge straight up and out of the instrument.
Note If the standard desiccant cartridge is stored in a dry box or sealed bag (you can
use the bag the purge cartridge came in), the cartridge can be reused.
Note Be careful not to drop the sealing O-ring into the desiccant compartment.
8. Open the purge kit and then open the sealed desiccant packet (must be sealed to keep the
desiccant dry).
9. Remove the O-ring from the packet and install it on the instrument base plate (see the
previous image). Press (rather than roll) the new O-ring into place to be sure it is properly
seated in the groove.
10. Remove the desiccant cartridge with purge fittings from the packet, insert the cartridge in
the desiccant compartment, make sure it is seated properly over the O-ring and then
secure it using the two screws.
Note the correct orientation of the cartridge (the label on the cartridge should be readable
from the front of the instrument when the cartridge is installed).
Purge inlet Relief valves
Purge Desiccant
Cartridge
11. Connect the right-angle flow coupler from the wall plumbing assembly (labeled “to
spectrometer”) to the purge inlet on the instrument.
15. Plug in the power cord to the instrument and turn on the power.
16. Wait 30 to 60 minutes for the instrument to fully purge.
Note If the instrument is powered off without purge, you can use either desiccant
cartridge (with or without the purge connectors) to maintain the required humidity.
Make sure you check the humidity indicator once a month (see Checking the
Humidity Indicators) and replace the desiccant when necessary (see Replacing the
Desiccant).
WARNING Prevent fire and explosion. The infrared source inside the instrument is an
ignition source. If you are using volatile solvents, provide a fume hood or other active
venting system that is free of spark and other ignition sources and prevents flammable
vapors from collecting in the atmosphere surrounding the instrument.
The sample compartment windows are either potassium bromide (KBr) with a hygroscopic
protective coating, or non-hygroscopic zinc selenide (ZnSe) with an anti-reflection coating
(for extremely humid environments). Both types of windows should be clear and clean (no
fingerprints).
NOTICE
• Do not allow liquids to come into contact with the sample compartment windows.
• Windows can be scratched and ruined very easily. Do not touch or attempt to clean
them. Dust will not affect the signal, but fingerprints can degrade the performance of
the instrument and permanently damage mirrors or windows. If you wish to remove
dust from a mirror or window, blow it off with a gentle stream of clean, dry air or
nitrogen (Do not use compressed air from a can or contaminants may cause damage.)
WARNING Replace the sample compartment windows only with replacement parts
supplied by us.
Tools needed:
• iS5 window replacement tool (provided)
• iS5 replacement windows (2)
• latex gloves
Replacement windows
1. Turn off the instrument power. See Turning the Power On and Off.
Note If your instrument is purged, you can leave the purge on while you replace the
windows to prevent ambient air from entering the spectrometer.
2. Remove any sampling accessory from the instrument. See Installing an Accessory.
3. Line up the three ribs in the window replacement tool with the notches in the rim of the
installed window.
Sample
compartment window
Note The window fittings are tight. Use one hand to stabilize the instrument and the
other to remove the window.
4. Press firmly on the tool and turn it counterclockwise to loosen the window.
5. Remove the tool and continue loosening the window by hand until it releases.
Note Be sure to use gloves to handle replacement windows and handle them by the
rims. (Avoid touching the windows even when wearing gloves.)
6. Gently insert the new window into the opening and turn it clockwise until hand tight.
The window should screw in easily at the start; if it doesn’t, remove the window and
restart to avoid stripping the threads.
7. Using the window replacement tool, slowly and firmly turn the window as far as it will go
(you will feel the O-ring compress fully when the window rim contacts the housing).
8. Turn on the instrument power and start OMNIC software.
9. Make sure the window type is properly configured in OMNIC software.
Choose Advanced Diagnostics in the Collect menu. The Diagnostics screen is displayed.
Select the Purge Window type (KBr or ZnSe), enter today’s date and then click Save.
10. Make sure the system has been on for at least 15 minutes (one to six hours for best
results), and then run the Performance Verification (PV) test.
From the OMNIC main screen, choose View > System Status. The System Status
Overview dialog box is displayed.
Note To order parts, call our Technical Support. See Contacting Us.
WARNING Before replacing the power supply, turn off the instrument. Replace the power
supply only with replacement parts supplied by us.
1. Turn off the instrument power. See Turning the Power On and Off.
2. Disconnect the old power supply from the wall outlet or power strip and from the
instrument power connector.
Power switch
Power
connector
Note The power supply replacement kit does not include the AC power cord, which
is specific for the location of your installed instrument. Save the cord and use it with
the new power supply. If the power cord becomes damaged, call our Technical
Support to order a new one. See Contacting Us.
WARNING To ensure a good ground connection and avoid shock hazard, do not use
an outlet that is connected to a conduit ground. The ground must be a
noncurrent-carrying wire connected to earth ground at the main distribution box.
Type the serial number for the new power supply, enter today’s date and then click Save.
8. Make sure the system has been on for at least 15 minutes (one to six hours for best
results), and then run the Performance Verification (PV) test.
From the OMNIC main screen, choose View > System Status. The System Status
Overview dialog box is displayed.
Tools needed:
• A #1 Phillips head screwdriver (provided with the instrument)
WARNING Before replacing the source, turn off the instrument. Replace the source only
with replacement parts supplied by us.
WARNING The source becomes extremely hot while you are using the instrument.
Always allow the source to cool for at least 10 minutes after it is turned off before you
work with it.
NOTICE Never touch a source element with your bare fingers. Skin oils or other deposits
on the element will shorten its life. Always use clean finger cots, gloves, or clean laboratory
tissue when handling a source element.
1. Turn off the instrument (see Turning the Power On and Off) and unplug its power cord.
2. Disconnect the cables from the back of the instrument.
3. If the spectrometer or an installed accessory is purged, disconnect the purge lines from the
spectrometer and the accessory (quick connect fittings stop the flow automatically). See
Installing a Purge Kit for more information.
4. Remove any installed accessories from the instrument (see Installing an Accessory).
Note Make sure at least 10 minutes have passed since you turned off the instrument
before you perform the next step.
5. Carefully tip the instrument back until it rests on the four feet on the rear panel.
6. Use a #1 Phillips head screwdriver to loosen the three captive screws that hold the source
in place.
7. Grasp the bale firmly and pull the source straight out of the instrument.
8. Record the serial number for the new source on a piece of paper. You will need to enter it
after the source is installed.
9. Insert the new source.
Make sure the three captive screws are straight and then slowly insert the source into the
cavity until the source mount is flush with the instrument. Tighten the screws to secure
the source and snap the bale into its storage position.
10. Carefully return the instrument to its upright position and reconnect the cables you
removed.
11. If your instrument is not purged, replace the desiccant. See Replacing the Desiccant for
instructions.
WARNING To ensure a good ground connection and avoid shock hazard, do not use
an outlet that is connected to a conduit ground. The ground must be a
noncurrent-carrying wire connected to earth ground at the main distribution box.
12. Plug in the power cord to the instrument, turn on the power switch and start OMNIC
software.
13. Wait until the power indicator stops flashing.
14. Enter the serial number for the new source in OMNIC software.
Choose Advanced Diagnostics in the Collect menu. The Diagnostics screen appears.
Verify that the new source is working properly. (If it does not work properly, a red slash
mark appears over the Source icon on this screen. Contact Technical Support for
assistance; see Contacting Us.)
Click the Installation icon. A diagnostic information screen appears.
Type the serial number for the new source (recorded in step 8), enter today’s date and
then click Save.
15. Make sure the system has been on for at least 15 minutes (one to six hours for best
results), and then align the instrument (see Aligning Your Spectrometer).
CAUTION To avoid a shock hazard, do not allow liquid to run into the power supply or
the back of the instrument.
NOTICE
• Do not use harsh detergents, solvents, chemicals or abrasives; these can damage the
finish. Do not allow liquid to contact any windows in the sample compartment.
• Windows can be scratched and ruined very easily. Do not touch or attempt to clean
them. Dust will not affect the signal, but fingerprints can degrade the performance of
the instrument and permanently damage mirrors or windows. If you wish to remove
dust from a mirror or window, blow it off with a gentle stream of clean, dry air or
nitrogen. Never allow any liquid to come into contact with a window or optical
component in the instrument.
NOTICE
• Before you disconnect the power supply, always discharge any static electricity you
may have accumulated by touching the perforated metal shield over the power supply.
• Do not touch any printed circuit board in the instrument (such as the circuit board
on the detector).
• Do not remove replacement components from their protective packaging until you
are ready to install that component in the instrument.
Follow these steps to monitor the humidity level inside the instrument.
Contents
• Checking Humidity When the Instrument is Powered Off
• Checking Humidity When the Instrument is Powered On
As long as the humidity indicator on the instrument is blue, the desiccant is not saturated and
does not need replacement. When the desiccant becomes saturated, the humidity indicator
turns pink and then light pink or almost white. This means you must replace the desiccant
and the humidity indicator. See Replacing the Desiccant for instructions.
NOTICE Check the paper humidity indicator at least once a month when the instrument
is powered off.
1. From the OMNIC software main screen, choose Advanced Diagnostics in the Collect
menu. The Diagnostics screen appears.
3. Verify that the current Internal Humidity value is between 2 and 50%.
Note When the internal humidity is above 50%, OMNIC software displays a
message to replace the desiccant (see Replacing the Desiccant) or increase the purge
flow (see Setting the Purge Gas Controls). If the humidity is above a critical level, the
System Status will report an instrument failure.
Tools needed:
• Desiccant replacement kit (or desiccant canisters that have been regenerated)
• A 0.05-inch Hex key
• A #2 Phillips head screw driver
• Gloves, finger cots, or laboratory tissue (to handle the humidity indicator)
WARNING Before opening the desiccant compartment, turn off the instrument power,
unplug the power cord and remove all accessories and samples from the system. Entry of
flammable liquids or gases into the desiccant compartment is an explosion hazard. If such
entry occurs, contact us immediately and do not apply power to the instrument until the
condition has been corrected.
NOTICE Make sure nothing falls into the instrument while the desiccant cover is
removed.
WARNING Replace the desiccant only with replacement parts supplied by us.
1. Turn off the instrument (see Turning the Power On and Off) and unplug the power cord.
2. Turn off the purge, if applicable, and disconnect the purge line to the instrument and any
accessory.
3. Remove any installed accessory and all samples from the instrument (see Installing an
Accessory).
Note You can use either desiccant cartridge (with or without the purge connector) to
maintain the required humidity when the instrument is powered off. The illustrations
below show the nonpurged version of the desiccant cartridge. To see the purged
version, see Installing a Purge Kit. The instructions for replacing the desiccant
canisters are the same for both cartridges.
4. Use the #2 Phillips head screw driver to loosen the two captive screws from the desiccant
cartridge, and lift the cartridge straight up and out of the instrument.
Note Be careful not to drop the sealing O-ring into the desiccant compartment.
6. Remove the new O-ring from the packet and install it on the instrument base plate (see
the previous image).
Press (rather than roll) the new O-ring into place to be sure it is properly seated in the
groove.
7. Use the 0.05-inch Hex key to loosen (approximately two turns) the eight screws that hold
the two saturated desiccant canisters to the desiccant cartridge. Remove the canisters.
Note The desiccant canisters can be regenerated. See Regenerating the Desiccant for
details.
Humidity indicator
Desiccant cartridge
Channel
Desiccant canisters
8. Turn the desiccant cartridge upside down and peel the old humidity indicator off the
window. Discard the used indicator.
NOTICE Always wear lab gloves or finger cots, or use laboratory tissue, when
handling a new humidity indicator. Oil or moisture from skin can discolor the
indicator.
9. Open the sealed packet that contains the new desiccant (must be sealed to maintain the
desiccant) and remove the new humidity indicator.
10. Turn the desiccant cartridge upside down and press the blue indicator onto the window
so that the flat edges of the indicator are aligned with the edges of the cartridge. The text
should be visible through the window when you turn the cartridge right side up.
NOTICE Make sure that the indicator is pressed tightly against the window so that it
does not come loose in the desiccant compartment when you reinstall the cartridge.
11. Install the new desiccant canisters. See the previous illustration for this step.
Insert the canisters into the channel of the desiccant cartridge and tighten the eight screws
(approximately 2 turns) until they are flush with the cartridge channel.
12. Insert the desiccant cartridge into the spectrometer so that the text on the humidity
indicator can be read by someone standing in front of the instrument.
Make sure the cartridge is seated properly over the O-ring and then secure it using the
two screws.
WARNING To ensure a good ground connection and avoid shock hazard, do not use an
outlet that is connected to a conduit ground. The ground must be a noncurrent-carrying
wire connected to earth ground at the main distribution box.
13. Plug in the power cord to the instrument and turn on the power.
14. Record the desiccant replacement date in OMNIC software.
Start OMNIC software and then choose Advanced Diagnostics in the Collect menu.
The Diagnostics screen appears.
Enter today’s date for the desiccant replacement and then click Save.
15. Reconnect the purge line to the instrument, if applicable, and turn on the purge.
16. Make sure the system has been on for at least 15 minutes (one to six hours for best
results), and then run the Performance Verification (PV) test.
From the OMNIC main screen, choose View > System Status. The System Status
Overview dialog box is displayed.
Tools needed:
• Vented oven
• Insulated cloth or hot pad
Note If you need to replace the humidity indicator, you must order new desiccant
canisters. A new humidity indicator is included with the new desiccant canisters.
NOTICE
• If you are going to dry and reuse saturated desiccant canisters, make sure you have
fresh desiccant canisters you can place in the instrument while the saturated canisters
dry.
• Do not leave canisters in the oven for more than three hours and do not exceed a
temperature of 150 °C (300 °F). Allow the canisters to cool completely in a dry
environment before you install them in the cartridge.
• Make sure the insulated cloth or hot pad you use to handle the regenerated desiccant
canisters is not contaminated with substances that could be absorbed by the desiccant
and subsequently released into the instrument.
1. Place the saturated desiccant canisters in a vented oven at 150 °C (about 300 °F) for three
hours.
WARNING To avoid a burn hazard, use an insulated cloth or hot pad to handle the heated
desiccant canisters, and always wait until the canisters have cooled to room temperature
before reinstalling them in the instrument.
2. Use an insulated cloth or hot pad to remove the desiccant canisters from the oven.
Place the hot canisters in a dry environment such as a desiccator and allow them to cool
to room temperature. As soon as the canisters reach room temperature, you must either
install them in the instrument or seal them in an airtight container or bag. Otherwise,
they will absorb moisture from the air in the room and quickly become saturated again.
If the filter is clean and white, it does not need replacement. If the filter is yellow, or otherwise
discolored, or if it is contaminated with debris or foreign particles, replace the filter.
1. Turn off the purge gas at the main valve. Do not turn down the flow meter or the pressure
regulator.
2. Remove the plastic bowl that houses the filter and then remove the filter. (You can
unscrew them both by hand.)
Bowl
Filter
Note You may notice increased levels of water and carbon dioxide in spectra collected
immediately after you have had the purge gas turned off. If this interferes with your
data, wait 15 to 60 minutes until purge is completely re-established.
Replacing Fuses
The fuses in your instrument are in the power supply and are not user-replaceable. If the fuses
blow, you must replace the power supply (see Replacing the Power Supply). If you need
assistance, contact Technical Support; see Contacting Us.
Troubleshooting
Contents
• Using the Troubleshooting Topics
• Hardware Problems
• Error Messages
Hardware Problems contains topics that address specific hardware problems you may
encounter. Each topic lists steps you can take to try to resolve the problem.
Error Messages contains topics that deal with error messages you may see. Each topic lists
steps you can take if you see such error messages.
To expedite and assist in problem diagnosis, please be prepared to send us a copy of your
diagnostics log file. The diagnostic log includes a history of important system parameters and
events that occur during system operation. To copy the diagnostics log to your Windows
desktop, start OMNIC software and choose File > Copy Diagnostics File to Desktop.
Attach this file to an e-mail message to our Technical Support group. See Contacting Us for
our e-mail addresses.
Note If you are having a problem with applications or software, refer to the electronic or
printed software documentation that came with your system.
WARNING When performing any of the procedures in the troubleshooting topics, always
follow the safety precautions included in this document or Help system and in any of the
other documentation that came with your instrument.
Hardware Problems
Contents
• Spectral Peaks Appear Shifted in Wavenumber Axis
• Fails Laser Frequency Verification
• The Power Indicator Does Not Light When You Turn on Your Instrument
• The Power Indicator Blinks Continuously
• The System Scans Normally but the Signal Intensity Is Low
• The Spectra Have Noise Spikes
• The Baseline Is Not Stable
• The Alignment Fails
• Data Cannot Be Collected (or Other Software Problems)
The Power Indicator Does Not Light When You Turn on Your Instrument
1. Turn the instrument off and check the connections between the instrument and the
power supply.
2. Make sure the power cord is securely connected to the power supply and the power
supply is connected to a properly grounded and working outlet.
3. Check the power cord descriptions and part numbers to make sure the power cord is
appropriate for your local AC power source. Order a new power cord if necessary.
4. Check the ground prong on the power cord. If the ground prong has been removed or is
defective, replace the power cord.
5. Contact Technical Support for assistance; see Contacting Us.
Note If you are using purge and the accessory cover was recently opened (or a service
representative opened the main cover), allow the instrument to equilibrate for 1 to
6 hours after the cover is closed.
8. Open Experiment Setup and click the Bench tab. Click the Laser icon and check the laser
indicators. If a laser measurement is out of tolerance, contact Technical Support for
assistance.
9. Contact Technical Support for assistance; see Contacting Us.
6. Make sure the system has been on for at least 15 minutes (one hour for best results), and
then click Align on the Diagnostic tab.
7. Make sure any accessories are installed according to the instructions that came with them.
8. Contact Technical Support for assistance; see Contacting Us.
Error Messages
Contents
• The Computer Cannot Communicate With the Instrument
• The Instrument Temperature Is Out of Specification
• OMNIC Cannot Communicate With the Detector
• The Instrument Is Not Scanning
• The Laser Voltage Is Out of Specification
• The Source Voltage Is Out of Specification
• The Power Supply Voltages Are Out of Tolerance
2. If the problem continues, close OMNIC software and turn off the instrument power.
Wait for a few seconds, and then turn on the instrument power. Restart OMNIC
software.
3. If the Power indicator on the instrument does not light, make sure the power supply is
connected to the instrument and a properly grounded wall outlet or power strip.
4. Turn off the computer and then turn off the instrument power. Check the USB data
cable. If the cable is damaged, replace it. If the cable is loose or disconnected, reconnect it.
5. Turn on the instrument power. Turn on the computer and restart OMNIC software.
6. Remove any USB computer accessories that may be interfering with communication with
the spectrometer.
7. Close any applications that might interfere with OMNIC software.
8. Contact Technical Support for assistance; see Contacting Us.
4. Turn off the power to the computer and the instrument. Leave the power off for
15 seconds. Turn on the instrument power. Turn on the computer and restart OMNIC
software. If the instrument does not resume scanning, proceed to the next step.
5. Select View Reference Detector on the Diagnostic tab.
If the instrument scans normally (an interferogram should appear in the window in the
Experiment Setup dialog box), the problem is probably related to the sample or the
detector. Make sure that the sample compartment beam is not blocked, and check the
Diagnostic tab of the Experiment Setup dialog box to make sure the detector is
recognized. If it is not, contact Technical Support for assistance; see Contacting Us.
6. Check all of the status indicators on the Diagnostic tab of the Experiment Setup dialog
box. Click any indicator that has a red slash through it to view diagnostic information
about the component.
If the laser frequency is outside the specified range, contact Technical Support to have the
electronics module replaced. If the laser voltage or current is outside the specified range,
contact Technical Support for assistance.
If the temperature is not in the acceptable range, make sure the cooling vents on the back
of the instrument are not blocked and nothing is stacked on top of the instrument.
Maintain at least 30 cm (12 in) of clearance behind the instrument and 64 cm (25 in) of
clearance above the table. Make sure the ambient temperature is between 18 °C and
35 °C (64 °F and 95°F).
7. Contact Technical Support for assistance; see Contacting Us.
2. If other values on the Power diagnostic screen are out of their specified ranges, contact
Technical Support for assistance; see Contacting Us.
Index
A collection failure 56
communication cabling 4
accessory
attaching purge to 8 computer
installing base-plate mounted 9 communication error 56
installing integrated 7 contact information 2
installing slide-mounted 14 conventions 1
removing 8
storing 8 D
adapter
data cable 4
installing iD base 9
data collection 15, 19
aligning
spectrometer 17 data collection failure 56
alignment failure 55 desiccant
checking 39
altitude
regenerating 46
recommended for instrument 3
replacing 42
desiccant indicator
B proper handling 44
baseline desiccation
unstable 55 requirement 1
baseplate mounted accessory detector
installing 9 replacing 49
beam focus 19 detector communication error 57
documentation 2
C
cabling 4 E
calibrating electrical cabling 4
laser 17 electricity
caution static 39
damaged cover 3 electronics module
power supply 30 replacing 49
toxic materials 7 error
using liquids 38 computer communication 56
cleaning detector communication 57
iD1 Transmission Module 38 humidity 59
instrument 38 interferometer temperature 59
collecting data 15 laser frequency 59
collecting data with the iD1 19 laser temperature 59
manuals 2 fuses 50
messages power supply 30
error 56 source 33
the detector 49
the electronics module 49
N the laser 49
narrow slot 19
noise spikes 54
S
sample
O focus 19
OMNIC software installing film 19
and instrument initialization 6 installing in iD1 Transmission Module 19
ordering parts 2 scanning error 57
outlet serial numbers
power 4 entering 19
shifting
P of peaks in the spectrum 52
signal
parts
too low 53
ordering 2
slide-mounted accessory
PDF documents 2
installing 10, 14
PDFs 2
slots
peaks in iD1 Transmission Module 19
shifted 52
solvents
Performance Verification test 17
warning 7
power
source
turning on or off 5
replacing 33
power indicator 6 source voltage error 58
blinks continuously 53
specification
does not light 52
for purge gas 15
power outlet
specifications
specifications for 4
for power outlet 4
power supply
spectrometer
replacing 30
aligning 17
power supply voltage error 58 cleaning 38
purge installing integrated accessory in 7
changing the filter 48 laser 17
for accessories 8 power 5
requirement 1 purging 15, 22
setting 15 troubleshooting 52
purge gas spectrum
specifications 15 noise spikes in 54
purge kit shifted peaks 52
installing 22 unstable baseline 55
static electricity precautions 39
R storing
regenerating desiccant 46 an accessory 8
removing Swap-Top Module
an accessory 8 installing 12
replacing
desiccant 42
T
technical support 2
Technical Support contact information 2
telephone numbers 2
temperature
recommended for instrument 3
temperature error 57
toxic materials
warning 7
transmission cell
installing 14
troubleshooting 51
U
unstable baseline 55
using
iD1 Transmission Module 18
V
vents
access to 6
verifying laser 17
vibration
and flow rate 16
W
warming up the instrument 6
warning
access 4
desiccant harmful if swallowed 44
for performing maintenance 21
hot source 33
open desiccant compartment 24, 42
power supply 30
purge gas 16, 22
replacement parts 21
use proper grounding 4
volatile solvents 7
wide slot 19