Volume 3, Chapter 5. Clinical Visual Electrophysi
Volume 3, Chapter 5. Clinical Visual Electrophysi
GENERAL CONSIDERATIONS
ELECTRORETINOGRAM
ELECTRO-OCULOGRAM
VISUALLY EVOKED POTENTIALS
MULTIFOCAL TECHNIQUES IN ELECTROPHYSIOLOGY
CLINICAL APPLICATIONS
REFERENCES
Back to Top
VISUALLY EVOKED POTENTIALS ELICITED BY PATTERNED
STIMULI
Patterned VEPs can be elicited by either pattern reversal (or
counterphase reversal) or pattern appearance/disappearance
(onset/offset, or on/off) using checkerboards, square wave
gratings, or sine wave gratings as pattern stimuli. Probably the
most frequently employed stimulus in clinical practice is a slow
(one to five reversals per second), square wave reversal of high-
contrast checkerboards with checks of 30 arc min or greater.
Under these conditions, the transient response waveform consists
of an early positivity at about 50 msec, a negative component at
about 70 msec, a major positive deflection at about 100 msec
followed by a second negativity at about 140 msec, and
sometimes a second positivity at about 155 msec.
The response parameters that are most frequently used clinically
are the time from stimulus onset to peak (latency) of the largest
positive component, the P100 or P2, and its amplitude. Defects of
the optic nerve, such as those encountered in a variety of optic
neuropathies, may produce prolonged latencies and/or decreased
amplitude of this component (Fig. 20).
Fig. 20. Prolonged pattern visually
evoked potential (VEP) following
optic nerve involvement. The VEPs
were elicited by 30-minute checks
reversing at 2 Hz and were recorded
with the active electrode 5 cm above
the inion referred to the right ear.
The patient, a 27-year-old white
male, had a history of optic neuritis in the right eye. The VEP
elicited from the right eye (OD) is prolonged in latency relative
to laboratory norms and the left eye (OS).
The pattern reversal evoked potential is a complicated response
that reflects the summation of several underlying processes,
depending on the exact stimulus parameters, such as check size,
reversal rate, and mean luminance. These include the summation
of local luminance and contrast responses, the summation of
pattern appearance and disappearance responses, and the
summation of movement and contrast responses. Given the
multiple processes summated in pattern reversal VEPs, there are
multiple cortical origins of pattern reversal VEPs. Responses
elicited by reversal stimulation tend to be more affected by eye
movement disorders, such as nystagmus. However, one possible
advantage of the pattern reversal VEP is that its basic waveform
remains constant as a function of age, unlike the pattern
appearance VEP.104 The major change involves a reduction in the
latency of the major positive component from about 200 msec in
early infancy to about normal adult values by 2 to 3 years of
age105.
Back to Top
VISUALLY EVOKED POTENTIALS ELICITED BY LUMINANCE
CHANGE
If a luminance VEP is elicited by a luminance pulse or flash, the
typical waveform shows positive peaks at about 45, 100, and 190
msec and negative peaks at about 65 and 150 msec, followed by
what has been termed after discharge, which represents
restitution of cortical electrical activities to prestimulus conditions
(see Fig. 16).106
The initial period before about 100 msec, termed the primary
evoked response, most likely represents activity in area 17 of the
striate cortex and seems more closely related to central visual
function because the fovea and macula are disproportionately
displayed near the tip of the occipital cortex. The electrical
activity after about 100 msec, termed the secondary evoked
response, most likely represents spread of information to areas
18 and 19 as well as to other associational areas governing eye
movements, visual memory, and other poorly understood
activities.106,107
Periodic luminance stimulation (flicker) can be provided in one of
several ways: xenon flashes or lights modulated with sine or
square waves. Stimulation that is more rapid tends to collapse
the major components of the VEP into a simpler waveform. There
are three peaks in the function relating VEP amplitude to
frequency of stimulation: one at about 10 Hz, one at about 20 Hz,
and one at about 40 Hz. These three frequency regions have
different functional properties, suggesting that they represent
functionally separate neural systems.108 For example, scalp
topography of the response suggests that VEPs elicited by
stimulation of 30 Hz and higher are limited to the striate cortex
and probably reflect the same mechanisms as those that
generate the flash VEP negative-positive complex at about 40 to
70 msec. The 20-Hz peak appears to reflect the mechanisms
present in the complex at about 100 msec and the 10-Hz peak
appears to reflect the later, more diffuse activity of the transient
luminance VEP.
If patients are tested with closed eyelids at the rate of 10 Hz, the
waveform becomes a double-peaked response (Fig. 20). Most
likely, the smaller peak represents the primary evoked response
and the larger peak represents the secondary evoked response
(Fig. 21).38,39 Accordingly, a train of responses to 10-Hz
stimulation usually places the smaller peak approximately 80 to
100 msec following the preceding flash, with the larger peak
occurring between 110 and 130 msec following the preceding
flash (Fig. 22). As with all VEPs, binocular stimulation produces
larger responses if the visual system is otherwise normal. This
occurs because most cortical neurons are binocularly innervated.
Fig. 21. Visually evoked potential
recorded in response to 10-Hz
stimulation in normal right eye and in
left eye with reduced visual acuity as
a result of macular degeneration. Note
marked reduction in amplitude of
primary evoked response (smaller wave) with mild reduction in
amplitude of secondary evoked response (larger wave).
Fig. 22. Responses to 10-Hz
stimulation. A. Normal visually evoked
potential (VEP) with double-peaked
responses. B. VEP recorded in subject
with decreased visual acuity,
indicating a lack of smaller primary
evoked responses. C. Theoretic
primary evoked response. D.
Theoretic secondary evoked response.
Arrows show time at which
stimulation is delivered.
Back to Top
INTERNATIONAL STANDARDS FOR VISUALLY EVOKED
POTENTIALS
The current ISCEV VEP standards and their revision, which is in
press,19recognize three types of stimuli: flash, pattern reversal,
and pattern onset/offset. All standard stimuli are transient with
two or less flashes, reversals, or appearances per second. A
standard flash, as defined in the ERG standards, should elicit the
flash VEP, although a wide-field flash is acceptable. The pattern
reversal stimulus consists of black and white checks or black and
white gratings that abruptly alternate without an overall change
in luminance. At least two pattern element sizes should be 10-
and 15-minute checks, or 1.0 and 4.0 cycles per degree gratings.
The visual field stimulated should exceed 15 degrees. The pattern
onset/offset stimulus should abruptly appear and disappear from
a diffuse background, which has the same space-averaged
luminance as the pattern. The recommended pattern/blank
screen sequence is a 200-ms pattern separated by at least 400-
ms diffuse background. The analysis time should include both
onset and offset responses. All VEPs should be recorded with
normal pupils monocularly.
Two general purposes are recognized for performing VEPs:
assessment of prechiasmal lesions and assessment of
postchiasmal lesions. For both purposes, pattern reversal is the
preferred stimulus. Flash stimuli are preferred only in the
presence of media opacities. Pattern onset/offset is mainly of
benefit if one wishes to assess acuity. To detect prechiasmal
dysfunction, it is essential that monocular stimulation be
performed. Prechiasmal defects can be detected using a single
channel with the active electrode placed over Oz; therefore, one
channel is required. However, if a postchiasmal problem is
present, it cannot be detected. Three channels are suggested as
preferable, with electrodes placed at Oz, O4, and O3 and referred
to Fz. To detect chiasmal or postchiasmal defects, recordings
must be performed over both cerebral hemispheres. The active
electrodes should be placed at locations Oz, O4, and O3 and
should be referred to a common reference at Fz. Additional active
electrodes at O1 and O2 are suggested as useful. Pattern reversal
stimulation is preferred.
Back to Top
BEYOND THE STANDARD VISUALLY EVOKED POTENTIAL
As is the case of the ERG standard, the VEP standard represents
a minimum set of stimulus and response conditions that can be
clinically useful. Numerous variations on the VEP, however, can
prove to be useful in specialized clinical situations. These include,
but are not necessarily limited to, sweep VEPs, dichoptic VEPs,
and channel-specific VEPs.
Sweep Visually Evoked Potentials
Sweep VEPS are a special class of steady state VEPs in which
some stimulus parameter is changed during the course of a
recording period or sweep.109–111 Most frequently, the parameter
varied is spatial frequency or contrast. Using a standard set of
criteria, one can use this rapidly acquired information to
determine a threshold. In patients who are nonverbal or
preverbal, or whose verbal responses cannot otherwise be
trusted, these responses can be a useful means of determining
acuity or other thresholds.
Dichoptic Visually Evoked Potentials
Dichoptic VEPs are elicited when the stimuli presented to the two
eyes differ in some parameter. In the most recent past, the
stimulus parameters, which have been of greatest interest, have
been temporal frequency or phase. When one stimulates each
eye with a stimulus of different frequency (f1 and f2),
intermodulation frequencies (f1 ± f2) are generated.108,112,113
These intermodulation frequencies depend on cortical interactions
and are abnormal in many types of abnormal binocular
interactions.114–116 Similarly, phase differences in input yield
very different predictions of the type of response that will be
generated, depending on the type of binocular interaction
present.113,117,118 Dichoptic VEPs recorded in animal primate
models also indicate their utility in understanding developmental
processes of binocularity.119,120
Channel-Specific Visually Evoked Potentials
Channel-specific VEPs are VEPs elicited by stimuli that are
designed to stimulate primarily one channel or pathway. For
example, isoluminant stimuli are presumed to stimulate primarily
the parvocellular system121–123 and motion stimuli are presumed
to stimulate primarily the magnocellular system.124–129 Similarly,
patterns that appear by increasing in mean luminance and those
that appear by decreasing mean luminance appear to stimulate
the on and off pathways specifically.130,131 Although the
separation of pathways is seldom, if ever, complete, the selective
stimulation of one of several pathways permits a clearer
identification of the mechanisms of some diseases and their more
precise diagnosis.122,123,127
Back to Top
MULTIFOCAL TECHNIQUES IN ELECTROPHYSIOLOGY
For many years, clinical electrophysiologists attempted to develop
an objective map of visual function. There are many reasons for
desiring an objective map of visual function. A map of visual
function should correlate better with localized lesions than does
the standard ERG and should provide information about central
retinal function, which the standard ERG does not. A perfect
example of this usefulness of the multifocal ERG (mfERG) is
provided in Figure 23, in which a macular dystrophy is clearly
visualized on the mfERG. Lastly, ophthalmologists are familiar
with looking at visual fields and thinking of variations in function
as correlating with specific diseases.
Fig. 23. Multifocal electroretinogram
(mfERG) from a patient with
Stargardt's disease. A. Fundus. B.
Standard electroretinograms (ERGs),
which are within normal limits. C.
mfERG. First-order kernels are
presented above and second-order
kernels presented below. The left
column presents pseudocolor
representations, the middle column represents individual
tracings, and the right column presents average responses from
rings that begin in the center for the top waveform and go to the
periphery in the bottom tracing. The central ERGs are smaller in
amplitude than those in the periphery in the raw signals. This is
presented dramatically in the pseudocolor map and the raw
waveforms. Averages of concentric rings indicate that the more
central rings (two upper tracings) are smaller than the more-
peripheral rings (four lower tracings). (OPs, oscillatory
potentials)
Initial efforts to create visual function maps using
electrophysiology involved sequential stimulation of locations in
the visual field and recording the response of each location in its
turn. As one might imagine, such efforts were long and tedious
for both the patient and the electrophysiologist. Tagging different
locations in the visual field, usually by using a different frequency
for each locus, improved the ability to gain information about
several different areas of the visual field simultaneously.132
Technical limitations usually limited such efforts to two or four
visual field regions. However, Erich Sutter133–135 clearly made
true visual fields using electrophysiologic measures possible with
his introduction of the multifocal electroretinogram in the 1990s,
following almost a decade of developmental work. Once Dr. Sutter
showed the way, several groups have followed his example and
introduced their own versions of multifocal techniques. Presently,
at least four systems are available internationally: two
manufactured by Roland Consultant of Germany, one by
Metrovision of France, and one by ObjectiVision of Australia. Each
system uses a different approach. Although we present some
information about each, the clear emphasis is on the visually
evoked response imaging system (VERIS) manufactured by Dr.
Sutter's company, Electro-Diagnostic Imaging (EDI). This is not
intended as a commercial endorsement but an acknowledgment
that more is known about the operation and performance of the
VERIS than about the other systems. Readers interested in a
more detailed but still basic account of kernel analysis, are
referred to Odom, 1995,136 and those interested in a solid
introduction to the interpretation of multifocal kernels and
summary of the recent state of knowledge are referred to a
review by Donald Hood.137
GENERAL PRINCIPLES OF MULTIFOCAL SYSTEMS
At their most basic level, all multifocal systems have three basic
components: a method of tagging multiple locations in the visual
field by stimulating each location in a different manner;
extraction of the response elicited at each location, using the tag
as an identifier; and a system of displaying a field map of the
extracted responses. Current systems differ largely in their
strategy for tagging and extracting responses from different
locations. There are two basic strategies: one based on frequency
analysis and one based on time series analysis. The majority of
current research and clinical activity using multifocal systems has
been with the multifocal electroretinogram; however, the basic
strategy may be used to map any visual function—namely,
retinal, cortical, pupillary, or neuroimaging—with appropriate
adjustments for the function. Similarly, most stimuli have
involved luminance modulation; however, pattern or chromatic
modulations are possible.
PRINCIPLES OF FREQUENCY-BASED APPROACHES
In frequency-based strategies, each location to be tested is
modulated at a different temporal frequency. The recorded
response is then Fourier analyzed and each frequency of
stimulation is extracted. Although, theoretically, it should be
possible to use a wide range of frequencies, practically different
frequencies preferentially stimulate different visual subsystems;
therefore, from a pragmatic point of view, it is preferable to use a
set of stimuli that are close in temporal frequency. Roland Consult
has constructed a system that consists of a set of spaced light-
emitting diodes (LEDs) that are each modulated at a different
frequency near 30 Hz. Responses are Fourier analyzed with a
resolution of about 0.01 Hz, and the amplitudes and phases of
the first and second harmonic are calculated. Waveforms and
values can be displayed in a field map. An example of the display
and traces are shown in Figure 24. At this point, we are unaware
of published studies that have used the Roland Consult
frequency-based system. In the absence of knowledge to the
contrary, it seems reasonable to assume that the origins of these
first and second harmonic responses are the same as those for
the full-field ERG, namely, bipolar cell level for the first harmonic
and ganglion cell and amacrine cell layer for the second harmonic
(see Fig. 1).
Fig. 24. Hertz flicker
multifocal results from a
patient with Stargardt's
disease. A. Raw data. B.
Numerical results from
rings representing
distance from the fovea.
Note that as in the Stargardt's patient in Figure 23, the central
responses are nonrecordable or greatly reduced. (Used with
permission of Roland Consult, Wiesbaden, Germany:
http://www.roland-consult.de.)
PRINCIPLES OF THE TIME SERIES–BASED APPROACHES
The available time series–based systems employ pseudorandom
binary sequences. The EDI system employs an m sequence
(although modified m sequences are possible in the scientific
version of the instrument). The Roland Consult and Metrovision
systems also use m sequences, whereas the ObjectiVision uses a
subset of an m sequence termed a Kasumi sequence. The precise
algorithm varies between systems, but the general principle is
that the sequences are constructed to be independent for each
location stimulated. Responses, or kernels, are extracted for each
location by cross-correlating the response with the pseudorandom
stimuli. Most commonly, only a first- and a second-order kernel
are extracted, although in theory it is often possible to extract
even higher-order kernels and to extract “cross-kernels” that
would describe the effects of one stimulus region on another.
Kernels are calculated by cross-correlating the response with the
stimulus; therefore, they themselves are not responses in the
same sense that a standard ERG or VEP is a response. However,
they do represent and predict responses. Figure 25 provides an
illustration of what kernels represent. The first-order kernel
represents the best approximation under the stimulation
conditions of the linear response of the system. The second-order
kernel represents an estimate of the deviation from the prediction
of this linear approximation. As such, it represents the effect of
one stimulus on another.
Fig. 25. First- and second-order kernels.
A. Stimulus consisting of two impulses
with a fixed separation. Within the
context of the multifocal
electroretinogram, this might be thought
of as two frames of light at a particular
location. B. Response to a single pulse, linear prediction of the
response to two pulses with the delay illustrated in A (the
response of two single flashes added together with the
appropriate delay), and the obtained response. C. Nonlinear
behavior of the retina, that is, the difference between the linear
prediction and the obtained response. D. Second-order kernel.
The second-order kernel has three dimensions. Time from pulse
1 is on the abscissa, and time from pulse 2 is on the ordinate.
The difference between linear predictions and obtained results
(as in C) could be plotted either on the z axis (not displayed) or
as contour lines on the xy coordinates. Most multifocal systems
present the z axis representation. Off diagonals represent the
differences between predicted and obtained responses for a
specific difference in time between the two pulses. In C, The first
diagonal slice reflects the interaction of two consecutive stimuli
separated by the minimum time between stimuli. C presents this
slice for the delay illustrated in A and B viewed as a z-axis plot.
Rectifiers are physical systems with second-order amplitude–
dependent nonlinearities. (Adapted from Odom.136)
Interpretation of mfERGs and multifocal VEPs (mfVEPs) also
requires an understanding of the visual system. Although the
first-order kernel represents a best approximation of the linear
behavior of a system, the visual system is highly nonlinear.
Therefore, even a best approximation of linear behavior will
contain nonlinear contributions (Fig. 26). When we think of the
visual system, we recognize that it is highly nonlinear and that
those nonlinearities are different for the retina and the cortex. For
example, at the retinal level, one of the earliest nonlinearities is
the fact that the visual response reveals considerable attenuation
at higher luminance levels. Although there is considerable
variation with eccentricity in retinal response, this inhomogeneity
of the visual field is magnified at the cortical level because of
cortical magnification. As a result, failure to maintain fixation has
a more dramatic effect for cortical than retinal responses.
Fig. 26. Linearity and the first-order kernel of a
nonlinear system. A strong reliable first-order
kernel does not mean that the system or the
response is linear. Linear estimates of a nonlinear
process are highly dependent on the stimulus
values used, such as mean luminance or contrast. Consequently,
different experiments can yield very different estimates of the
first-order kernel. The solid line represents a logarithmic
nonlinearity such as might be imagined for the retina. The dotted
lines indicate the limits of the stimulus conditions (abscissa) and
the observed responses (ordinate). The straight lines or “linear
estimates” fitted through the two regions are very different,
reflecting the underlying nonlinearity of the function they are
estimating. Similarly, the underlying nonlinearity of the visual
system implies that first-order kernels obtained under different
conditions will differ because of the nonlinear behavior they
estimate. This limitation does not deny clinical utility of the
technique. (Adapted from Odom.136)
Knowing the stimulus is important in interpreting kernels and
understanding multifocal results (Table 2). The most common
implementation of the multifocal technique employs a video
monitor running at 75 Hz, with a space averaged mean luminance
of about 100 cd·m-2. The video frames are about 13.3 msec.
Each lighted period is only a few milliseconds in duration and
separated from the next frame by a dark remainder of the frame.
There are several consequences of this stimulus arrangement.
First, sequences that involve several lighted frames represent
several sequential flashes rather than a continuous luminance
onset period,138 which has implications for efforts to record
scotopic ERGs139 or long-duration flash ERGs.140
http://www.octopus.ch/products/fr3_veris_description.htm.) B.
Subject looking into one of the newer VERIS systems that
permits fixation monitoring and is therefore appropriate for use
with children as well as adults.
The overall stimulus pattern should subtend a visual angle of 20
to 30 degrees on either side of fixation. Contrast between the
lighted and darkened stimulus elements should be 90% or
greater, with a mean luminance of 50 to 100 cd·m-2. The region
of the CRT beyond the area of stimulus hexagons should have a
luminance equal to the mean luminance of the stimulus array.
Central fixation stimuli (dots or crosses) should cover as little as
possible of the central stimulus element to avoid diminishing the
response. Electrodes that contact the cornea are recommended
for mfERG recording as for the full-field ERG. The optical opening
must be clear to allow good visual acuity and refraction.
The default response of the default mfERG is the first-order
kernel, termed K1 or FOK. It is a biphasic wave with an initial
negative deflection followed by a positive peak. A second
negative deflection may occur after the peak. These three peaks
are, respectively, N1, P1, and N2. The recommendations extend
only to the most frequent default conditions and not to higher-
order kernels.
Back to Top
CLINICAL APPLICATIONS
Electrophysiologic testing is most frequently used in five clinical
situations: diagnosis of retinal disease; diagnosis of optic nerve
disease; determination of organic versus functional visual loss
and identifying the locus of organic loss; determination of visual
function in nonverbal patients (mainly children); and assessing
visual system integrity behind medial opacities.
DIAGNOSIS OF RETINAL DISEASE
One of the most obvious applications of electroretinography is in
the study of hereditary and constitutional disorders of the retina.
These include partial and total color blindness (achromatopsia),
night blindness, and retinal degenerations.
In the past, electrophysiology diagnosed certain diseases, such as
retinitis pigmentosa. Because these diseases were not treatable,
electrophysiology was frequently used for diagnosis and genetic
counseling. However, as our knowledge progresses, it is becoming
increasingly clear that electrophysiology does not relate in a one-
to-one manner with genetically defined diagnoses. In the
relatively near future, many previously untreatable retinal
diseases will become treatable. Correct diagnosis will then
become crucial to the treatment. Diagnosis is likely to be
dependent on genetic testing. This will not eliminate the need for
electrophysiologic testing, however. Electrophysiology is likely to
become an important initial tool in characterizing the phenotype
of the disease so that appropriate genetic testing can be selected
to provide a definitive, treatment-related diagnosis. Not all
diseases will become treatable. Diagnosis of retinal diseases is
important even though they often are untreatable. Correct
diagnosis is essential for genetic counseling and for counseling
patients on the likely progression of their disease. When a
disorder involves primarily the rod system or primarily the cone
system, the ERG shows corresponding abnormalities that may be
important in counseling the patient (Fig. 28).
Fig. 28. ISCEV standard electroretinogram
responses in achromatopsia. The dark-
adapted responses are entirely normal. The
light-adapted response is nonrecordable.
This particular patient with achromatopsia
had no photophobia yet no residual cone
electroretinogram. Based on the response to psychophysical
tests, the patient was classified as a blue-cone monochromat.
Nearly the first dictum to emerge with the development of clinical
electroretinography was that the ERG was “extinguished” or
nonrecordable in patients with retinitis pigmentosa. With
improvements in recording techniques, however, small responses
have often been revealed. These responses are due to the cones,
which may still function even when all rod function has ceased
(Fig. 29). Retinitis pigmentosa is not the only disorder in which
the ERG is usually nonrecordable or greatly reduced in amplitude.
Other chorioretinal degenerations or inflammations, such as
choroideremia, Spielmeyer-Vogt disease, and luetic
chorioretinitis, may also result in virtually complete destruction of
the photoreceptors. This is also true in Leber's congenital
amaurosis, which is due to dysgenesis of the rods and cones.
Therefore, the nonrecordable ERG cannot be considered
pathognomonic of any of these conditions, and it is not helpful in
distinguishing between them. However, a consideration of the
presenting symptoms and fundus appearance can usually
distinguish between them.
Fig. 29. Electroretinogram
(ERG) responses in early
retinitis pigmentosa and
congenital stationary night
blindness (CSNB). Responses
in retinitis pigmentosa are
typically small or nonrecordable under all stimulus conditions. A.
Responses using ISCEV standard conditions. B. Standard ERG
responses of a patient with CSNB. The distinction between CSNB
and retinitis pigmentosa is very important in the clinical retina.
In other degenerative states of the retina, the standard ERG may
be normal; this is true in Tay-Sachs disease, in which the lesion is
located in the ganglion cells. Use of the photopic negative
response is too new to have proven useful in these cases. In
patients with paraneoplastic disease, the on response may be
significantly affected (Fig. 30).
Fig. 30. On-off response deficit in a patient with
paraneoplastic night blindness with cutaneous
malignant melanoma. The two upper tracings
present the on and off responses of a patient with metatstatic
cutaneous malignant melanoma. The on responses are greatly
reduced as compared with responses of the right eye (OD) of a
normal patient (lower tracing). (OS, left eye) (Courtesy of
Philippe Kestelyn, University of Ghent, Belgium.)
Retinal vascular disorders may have profound effects on various
ERG components. Retinal ischemia may result from many
different disease processes, such as arteriosclerosis, giant cell
arteritis, occlusion of a retinal artery or vein, and carotid artery
insufficiency. All result in a diminished b wave and a
proportionately larger a wave due to “unmasking” of the process
(PIII) responsible for the a wave (Fig. 31). Changes related to the
sensitivity of the dark-adapted ERG to light, such as latency of
the b wave or 30-Hz flicker and the intensity that generates a
half-amplitude b wave, are predictive of complications in central
retinal vein occlusion. OPs are altered in diabetic retinopathy and
may be a valuable predictor of proliferative changes.
Fig. 31. Electroretinographic responses in ischemic
vascular disease. The patient had visual loss due
to central retinal vein occulusion. Note deep a
wave and diminished b wave in the tracing for the
left eye.
Toxic states of the retina may be accompanied by ERG changes.
Siderosis produces ERG responses larger than normal in its early
stages; low-voltage responses are produced later in its course
(Fig. 32). The ERG in patients with chalcosis does not appear to
evolve through the early supernormal phase. Administration of
many drugs that may produce retinal damage, such as
chloroquine and quinine, results in a corresponding lowering of
the ERG responses (Fig. 33). ERGs have been used to assess the
extent of retinal damage in birdshot chorioretinopathy.149 In
addition, measuring the initial level of functional retinal damage,
ERGs may be used to monitor the need for and the effect of
treatment.
Fig. 32. Electroretinographic changes in
siderosis. These responses were obtained
on three different occasions from a
patient with an intraocular foreign body
containing iron. Note progressive
reduction of b-wave amplitude over a 15-
month period in both dark- and light-
adapted conditions.