SOP Abbott Afinion 2
SOP Abbott Afinion 2
COPY
Summary of Significant Changes at this Revision
Change the qpulse reference number of the external documents from MAN/7 toEXT/POCT/5
Cleaning of the analyser to be recorded on the quality control log in the column provided
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
Contents
1. Clinical Relevance/Purpose and Limitations of the Examination ......................................... 3
2. References and Definitions: ............................................................................................. 3
3. Actions and Methods:...................................................................................................... 4
4. Patient Preparation ......................................................................................................... 5
4.1. Environmental and Safety Controls ............................................................................ 5
5. Required Reagents, Quality Controls, Calibrators and Equipment Preparation..................... 5
5.1. Reagents .................................................................................................................. 5
5.2. Quality Control ......................................................................................................... 6
5.3. Installing a new lot number of iQC ............................................................................. 7
5.4. Calibrators ............................................................................................................... 8
5.5. Equipment ............................................................................................................... 8
6. Monthly Maintenance ................................................................................................... 10
6.1. Cartridge chamber cleaning: .................................................................................... 10
7. Afinion Test Cartridge .................................................................................................... 11
8. Instructions for the Performance of the Examination ....................................................... 13
9. Analysing patient sample ............................................................................................... 16
10. Procedure for Recording of Results ............................................................................. 17
11. Potential Sources of Variation and Measurement Uncertainty of Measured Values ........ 18
12. Performance characteristics and interferences ............................................................. 18
12.1. Specificity ........................................................................................................... 18
12.2. Linearity and measuring range ............................................................................. 18
12.3. Interferences ...................................................................................................... 18
13. External Quality Assessment Schemes ......................................................................... 19
14. Reference Limits, Reportable intervals and Report Interpretation ................................. 19
15. Personnel Involved in Interpretation, Authorising, Reporting and Monitoring of Reports20
16. Troubleshooting ........................................................................................................ 20
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
Definitions
QC Quality Control
EQA External Quality Assurance
IFCC International Federation of Clinical Chemistry
cITm Cobas IT Middleware
HbA1c glycated haemoglobin, haemoglobin A1c
min minute
ID identifier
Max maximum
SOP standard operating procedure
Hb haemoglobin
NEQAS National External Quality Assurance Scheme
3.1. Principle and Method of the Procedure Used for the Examination
Abbott Afinion 2 analyser is a portable analyser used for quantitative in vitro determination of
HbA1c in whole blood.
The test principle for the measurement of HbA1c using this analyser is by boronate affinity assay.
Afinion Test Cartridge contains all the necessary reagents for determination of HbA1c
concentration.
Sample material is obtained using sampling device (integrated with cartridge) and the test
cartridge is placed in the analyser.
Blood sample is automatically diluted and mixed with liquid in order to release haemoglobin from
erythrocytes. Haemoglobin then precipitates. Sample mixture is transferred to blue boronic acid
conjugate, which binds with glycated haemoglobin. This mixture is then soaked through filter
membrane and all precipitated haemoglobin (glycated and non-glycated) remains on the
membrane, excess conjugate is removed with washing reagent.
Analyser evaluates the precipitate on the membrane by measuring reflectance and evaluating
colour intensities of glycated haemoglobin (blue colour) and total haemoglobin (red). Ratio
between the colour intensities is proportional to percentage of HbA1c in the sample. Analyser
displays HbA1c concentration at the end of the assay in mmol/mol (HbA1c values are aligned to
IFCC reference method).
Whole blood specimen is required for HbA1c analysis and sample may be obtained by either finger
prick (capillary blood) or venepuncture (venous blood).
If venous blood is used the following anticoagulants are suitable: EDTA, lithium heparin, sodium
citrate and fluoride/oxalate.
Capillary on the sampling device (refer to figure 3) holds 1.5 µl of whole blood and this is minimum
volume required for the test.
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
Capillary blood samples cannot be stored and are discarded into appropriate waste once analysis
has taken place. Venous blood samples with anticoagulants can be stored in the fridge (2-80C) for
10 days or at room temperature (18-30 0C) for 8 hours. Venous blood samples should not be
stored frozen.
All samples will be identified by patient identification written on the tube to include a minimum of
3 of the following: full name, hospital number, NHS number, date of birth, address.
4. Patient Preparation
When collecting capillary blood sample make sure that the finger is clean, warm and dry. Allow for
a good drop of blood to form before sampling.
5.1. Reagents
Single Afinion test cartridge (Ref 1116795) contains all necessary reagents for one test and is
supplied ready to use (Figure 3).
Boxes of cartridges need to be acceptance tested before use with patient samples, this can be
done by running two levels of quality control on two cartridges from:
Any new lot number of cartridges ahead of going into use.
A new batch (each delivery) of cartridges regardless of lot number ahead of going into use.
When either a new lot arrives or when a delivery arrives (of same lot to previous) iQC should be
run on cartridges from these on the next Friday, this will count as your weekly iQC check. Please
write on the stock sheet that the batch or lot was accepted into use, with the date and time and
by whom.
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
Controls should be handled and disposed of as potentially biohazardous materials. Gloves should
be worn at all times when handling control material. Controls contain sodium azide as
preservative.
Storage
Control vials should be stored refrigerated at 2-80C – avoid exposure to direct sunlight and
temperatures above 250C.
Unopened control is stable until expiry date indicated on the vial when stored refrigerated.
Opened control vials are stable for 60 days when stored refrigerated – note the date of opening on
the vial. Store vials refrigerated, tightly capped and in an upright position when not in use.
Frequency
It is required to run both controls at a set time interval to ensure satisfactory performance of the
analyser. Controls will need to be run every 168 hours from last satisfactory control (approx. every
7 days). Analyser automatically locks out when this time has lapsed and it will not be possible to
run patient samples until controls are run and the result is satisfactory.
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
It will be necessary to manually enter new iQC lot numbers into the analyser’s database to ensure
controls are within the correct limits.
Control Data is encoded into a barcode and contains the following: lot number, control type,
control level, expiry date and acceptable control range. Barcode can be found on Afinion Control
Kit Package Insert.
Touch add button end enter Control Data from Kit insert – either scan
in a barcode or manually enter the long number which is located directly
underneath the barcode.
Data for both controls are required to be installed into the analyser. Repeat
the same procedure for the next control.
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
5.4. Calibrators
Abbott Afinion HbA1c does not require calibration. Analyser is originally calibrated during
manufacturing against reference system.
Test specific calibration data are established for each lot of test cartridges and stored in the
barcode label (see Figure 3). Barcode is read by integrated camera when cartridge is placed inside
the analyser and calibration data for the lot are transferred to the instrument and used for
calculating results (See Abbott Afinion cartridge kit insert EXT/POCT/5)
5.5. Equipment
Abbott Afinion 2 analyser (Figure 1) is a portable analyser used for quantitative in vitro
determination of HbA1c in whole blood.
Analyser self-tests during start up to ensure it is operating as per established specifications. This
test validates hardware and software integrity, test cartridge transport system, liquid transport
system and camera vision system. If the test fails at any point, red LED will flash and information
code will be displayed on the touch screen.
When analyser stays switched on for longer period of time, it will automatically restart once a day
to ensure self-test is done – this procedure does not interrupt analysis of the test cartridge.
Touch screen (Figure 1) allows operation of the instrument and only fingertips should be used – do
not use pens or other sharp instruments.
Screen saver will turn on after 3 minutes when touch screen is not in use – to reactivate, touch the
screen.
The lid (Figure 1) opens automatically when required, but needs closing manually – do not
attempt to open the lid manually.
Sound signals – short beep indicates completion of analysis. Two beeps mean information code is
displayed.
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
Switch analyser on by pressing ON/OFF button (Figure 1) – start up and self-test is initiated.
LED on top of analyser will flash red to indicate analyser is busy (Figure 1)
Analyser is ready to use when LED on top of analyser turns green (Figure 1)
and start up menu is displayed (Figure 2)
To switch the analyser off press ON/OFF button, wait for the touch screen
to switch off. It is now safe to unplug the analyser from the main power
supply if it needs to be moved.
When analyser is moved off site and exposed to outdoor conditions, please
allow 15 min for the analyser’s temperature to equilibrate before switching
analyser on.
Figure 2. Start-up menu
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
6. Monthly Maintenance
There are two procedures to be carried out for the monthly maintenance which are detailed out
below.
Afinion analyser cleaning kit (Ref: 1116784) should be used for cleaning the chamber – cleaning kit
is a pack of cleaning swabs. Please note the analyser must be unplugged before cleaning. Please
take care not to drip off any liquid from the swab into the analyser (this can affect the optics).
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
Test cartridge for HbA1c analysis (Ref 1116795) is supplied in a kit which contains 15 test
cartridges packaged separately in foil pouches with a desiccant bag.
Test cartridges are stable until the expiry date only when stored refrigerated (2-80C) in sealed foil
pouches.
Test cartridges can be stored in unopened foil pouches at room temperature (15-250C) and are
then stable for 90 days – note the date of removal from the refrigerator on the cartridge box.
Test cartridge must reach an operating temperature of 18-300C before use – once removed from
the fridge allow 15 minutes for the cartridge to reach desired temperature.
Open the cartridge just before use – once foil pouch is opened the test cartridge needs to be used
within 10 minutes after opening. Avoid exposure to direct sunlight and avoid touching optical area
at all times (refer to figure 3) – use designated handle to hold the cartridge.
Test cartridge is single use only – once used it is to be discarded into biohazard waste.
Open the foil pouch by using the tear strip as indicated. Grip the handle
(refer to figure 3) and remove the test cartridge from the pouch. Discard
desiccant bag and foil pouch in suitable waste container.
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
Please note that controls are potentially biohazardous materials – gloves should be worn at all
times when control material is handled.
a. Remove control material from the fridge and allow to reach room temperature (15-250C)
before use (takes about 45 minutes). Discard the vial if there is evidence of microbial or
fungal contamination and use fresh control material.
b. Control material should be mixed prior to use by vigorously shaking the vial for about 30s.
c. Refer to procedure 3.6.1 (b) on how to fill capillary with control material.
d. When cartridge is ready press the blue icon on the main screen in order to
analyse control sample – the lid will automatically open (slides out towards the operator).
i. Result of the control is checked against acceptable range for corresponding lot number:
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
j. Record result in the Quality Control log, then press to accept. Results are stored in
the analyser memory by date and time.
k. Proceed from step a. for the second level of iQC ensuring the blue icon is pressed
before running the next level.
l. Email the Internal Quality Control results to POCT email address:
[email protected]. Please include the following: IQC results, lot number of both
controls used and lot number of cartridges used.
m. The QC results will then be entered onto cITm in the Biochemistry Lab.
n. Laboratory staff: please refer to SOP/IT/IT/43 Cobas IT middleware, section 5.5 for
instructions on how to record results in cITm manually.
QC lockout
Set at 168 hours (from last satisfactory QC), after this interval the assay will be locked and not
available for patient testing. Controls must be run and be successful to reset the interval and to
unlock patient testing.
Yellow padlock appears in the upper right corner of the touch screen as a warning when
all controls are within interval, but there is ≤ 10% of the interval remaining before it
expires. Patient testing is enabled.
It is possible to check when QC interval will expire – touch the padlock and the screen will show
QC lockout status. Remaining time will be displayed next to the control (C I and C II) in the
following format DD:HH:MM (DD-days, HH- hours, MM-minutes).
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
Please note that gloves should be worn at all times when performing this procedure.
a. Refer to procedure 3.6.1 on how to fill capillary with sample - analysis of test cartridge
should start within 2-3 minutes – if stored too long before analysis the sample may dry or
coagulate and analyser will display an information code – it will be necessary to use new
cartridge and obtain fresh specimen.
b. When cartridge is ready press the orange icon on the main screen to analyse
patient sample – the lid will automatically open (slides out towards the operator).
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
An information code will be displayed (refer to Manufacturer’s manual for full list).
The result
Your signature
Time and date of measurement
The analyser the result was measured (Afinion 2)
N.B If no result is available for reporting this MUST be recorded in the patient notes.
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1
Traceability
Abbott Afinion™ HbA1c is traceable to the IFCC Reference Method for Measurement HbA1c
(EXT/POCT/5)
MoU
Refer to comparison data under properties in VAL/POCT/14 for Measurement Uncertainty
calculation for HbA1c measured on the Afinion 2 analyser.
12.1. Specificity
Abbott Afinion measures total glycated haemoglobin and reports HbA1c value. No significant
interference (<7%) was observed for samples with Hb variants and derivatives up to certain
concentration levels.
For full list please refer to Abbott Afinion Cartridge Kit Insert (QMS/EXT/POCT/5)
The measuring range for HbA1c is 20-140 mmol/mol and the assay is linear throughout this range.
12.3. Interferences
Please note: local guidelines should be considered and adhered to when interpreting results.
Reference Limits
All patients should have an individual target, balancing long term risk of complications with quality
of life and risk of hypoglycaemic events.
Expected values
Patients with HbA1c levels between 39 and 47 mmol/mol are identified as being at risk for
developing diabetes.
The diagnostic cut-off is 48 mmol/mol.
Results <20 mmol/mol are rare and may indicate that the sample contains substantial amounts of
foetal haemoglobin or that the patient may have haemolytic anaemia.
If any result appears questionable or if the clinical signs and symptoms appear inconsistent with
the result, re-test the patient and / or confirm the result by sending a labelled EDTA sample to the
Biochemistry Lab for confirmation.
The POCT coordinator is responsible for supply of and performance monitoring of the EQA for the
Afinion 2.
16. Troubleshooting
o An inadequate blood sample was obtained by finger prick - This must be recorded in the
patient’s notes - record as an incident in Datix if appropriate. A venous sample should be
collected into an EDTA tube (purple top), to be analysed on Abbott Afinion analyser (refer
to section 8) or send to the laboratory for HBA1C analysis.
Error codes
See the user manual for the full list (EXT/POCT/4)
Information code #105 = Result below the measuring range – report as <20 mmol/moL
Information code #106 = Result above the measuring range – report as >140 mmol/mol
Copy number Location held
1 Paediatric diabetes OP
2 Website
Printed copies are uncontrolled unless there is an allocated Copy Number on page 1