CT Notes
CT Notes
2
COMPUTED TOMOGRAPHY
o Increased radiation intensity toward rotation of the x-ray source, proving a self-
the edges of the beam calibrating system. Third generation system are
Compensation: used of bow- calibrated only once every few hours.
tie filter
Bow-tie Filter: equalize the radiation FIFTH GENERATION (early 1980’s)
intensity that reaches the detector array Electron beam CT
Ultrafast CT scanner
THIRD GENERATION X-ray tube rotation is mechanical
Characteristic: rotate/rotate No moving parts
X-ray Beam Shape: wide fan beam Electron Gun: produces a focused electron
Detector: curvilinear detector array (30) beam that generates a rotating x-ray fan
Imaging Time: <1 second beam after being steered along tungsten
Advantages: target rings
o Better x-ray beam collimation
o Decreased scattered radiation SIXTH GENERATION (early 1990’s)
o Good image reconstruction Helical/Spiral CT
Disadvantage: ring artifacts Volumetric scanners
o Compensation: software connected Introduced by Will Kalender and Kazuhiro
image reconstruction algorithm Katada
Uses slip ring technology
NOTE: Third generation designs have the Excels in 3D multiplanar reformation
advantage that thin tungsten SEPTA can be place Slip Ring: electromechanical device that
between each detector in the array and focused on conduct electricity and electric signals
the x-ray source to REJECT SCATTERED through rings and brushes across a rotating
RADIATION surface onto a fixed surface
Subsystems: x-ray tube, detector array, high and change them to electronic
voltage generator, patient couch and signals
mechanical support o Low or no after glow
A brief, persistent flash of
X-RAY TUBE scintillation that must be
Special requirements: taken into account &
o Power capacity: must be high subtracted before image
>120 kVp reconstruction
400 mA o High scatter suppression
o High speed rotors: for heat o High stability
dissipation Allows a system to be used
o Anode heat capacity: 7 MHU without interruption of
(Spiral CT) frequent calibration
o Heat storage capacity: 8 MHU o Types:
o Anode cooling rates: 1MHU/min o Gas-filled detector – previously used
o Focal spot size: small o Scintillation & solid-state detectors –
Takenote: CT scanners recently used
designed for high spatial o Gas detector:
resolution imaging not for o Basis: ionization of gas
direct projection imaging o Three types:
o Limiting characteristics: Ionization chamber
Focal spot design: must be Proportional counter
robust or strong Geiger-Muller counter
Heat dissipation o Characteristics:
o X-ray tube life: approx. 5000 Excellent stability
exposures (Conventional CT) Large dynamic range
Focal-Spot Cooling Algorithms: Low quantum efficiency
o Design to predict the focal spot o Scintillation detector:
thermal state o Characteristic: high x-ray detection
o To adjust the mA setting accordingly efficiency (90%)
Reduces patient dose
DETECTORY ARRAY Allows faster imaging time
o The entire collection of detectors Improves image quality
o Group of detectors o Crystals used: Sodium iodide
o The image receptor in CT o Replaced by:
o Detector: absorbs radiation and converts it Bismuth germinate (BGO)
to electrical signal Cesium iodide (CsI)
o Optimal Characteristics: o Current crystal of choice:
o High detector efficiency Cadmium tungstate (CdWO4)
The ability of the detector to Special ceramics
capture transmitted photons
4
COMPUTED TOMOGRAPHY
o Photodiode: converts light into electrical Value <1: non ideal detector system
signal o Result: increased in patient dose to
o Characteristics: maintain image quality
Small,
Cheap CAPTURE EFFICIENCY
Does not require power Refers to the ability with which the detector
supply obtains photons that have passed through the
patient
THREE IMPORTANT FACTORS
CONTRIBUTING TO DETECTOR ABSORPTION EFFICIENCY
EFFICIENCY Refers to the number of photons absorbed
by the detector
1.) GEOMETRIC EFFICIENCY Depends on: physical properties of the
The area of the detectors sensitive to detector face
radiation as a fraction of the total exposed o Thickness
area o Material
The amount of space occupied by the
detector collimator plates relative to the RESPONSE TIME
surface area of the detector The time required for the signal from the
2.) QUANTUM EFFICIENCY detector to return to zero after stimulation of
The fraction of incident x-rays on the the detector by x-ray radiation so that it is
detector that are absorbed and contribute to ready to detect another x-ray event
the measured signal A function of the detector design
Patient dose
HIGH VOLTAGE GENERATOR Predetector/Post Patient Collimator
o High frequency power o Restricts the x-ray beam viewed by
o High voltage step-up transformer the detector array
o Power: 50 kW o Purpose:
o Accommodates higher x-ray tube rotor To decrease scattered
speeds radiation
o Accommodates instantaneous power surges To improved contrast
characteristic of pulsed system o Determines:
Slice thickness
PATIENT COUCH Sensitivity profile
o Supports the patient comfortably
o Construction: low-Z material (Carbon OPERATING CONSOLE
fiber) Contains meters and controls
o Rationale: it does not interfere with o For selection of proper imaging
x-ray beam transmission & patient technique factors
imaging o For proper mechanical movement of
o Features: should be the gantry and patient couch
o Smoothly and accurately motor o For the use of computer commands
driven Allow image reconstruction
Rationale: precise and transfer
positioning is possible 2-3 operating consoles
o Capable of automatic indexing o 2 for CT radiologic technologists
Rationale: operator does not 1st: To operate imaging
have to enter the room system
between each scan 2nd: to post-process images
for filming and filing
COLLIMATION o 1 for physician
Restricts the volume of tissue irradiated To view the images
Purpose: To manipulate contrast, size
o Reduces patient dose & general visual appearance
o Improved image contrast Accepts the reconstructed
Types: post patient & prepatient collimator image from operator’s
Prepatient Collimator console
o Limits the area of the patient that Displays reconstructed image
intercepts the useful beam for viewing and diagnosis
o Mounted on the x-ray tube housing Two monitors:
or adjacent to it o 1st: provided for operator
o Purpose: to decrease patient dose To annotate patient data on
o Determines: the image (e.g. hospital
Dose profile
6
COMPUTED TOMOGRAPHY
SPATIAL FREQUENCY
FIVE PRINCIPAL CHARACTERISTICS OF Used to describe CT spatial resolution
CT IMAGE Low SF: represents large objects
1.) Spatial resolution High SF: represents small objects
2.) Contrast resolution
3.) Noise EDGE RESPONSE FUNCTION (ERF)
4.) Linearity Mathematical expression of the ability of the
5.) Uniformity CT scanner to reproduced a high-contrast
edge with accuracy
SPATIAL RESOLUTION
Ability to image small object that have high MODULATION TRANSFER FUNCTION
subject contrast (MTF)
Expressed in: linepairs/millimeter (lp/mm) Mathematical expression for measuring
A function of pixel size resolution
Takenote: SR for a CT image is limited to The ratio of the image to the object as a
the size of the pixel function of spatial frequency
Image reconstruction and postprocessing Used to describe CT spatial resolution
tasks: powerful way to affect SR MTF = 1: faithfully represents the object
Formula: SR (cm) = ½ {1/SF (lp/cm)} MTF = 0: image is blank and contain no
information
FACTORS AFFECTING/INFLUENCING MTF = intermediate values: intermediate
SPATIAL RESOLUTION levels of fidelity
1.) Pixel size
2.) Slice thickness CHARACTERISTICS OF CT IMAGING
3.) Voxel size SYSTEM CONTRIBUTING TO IMAGE
4.) Design of prepatient and predetector collimators DEGRADATION
5.) Detector size 1.) Collimation
2.) Detector size and concentration
EFFECT IN SPATIAL 3.) Mechanical/electrical gantry control
FACTORS
RESOLUTION 4.) Reconstruction algorithm
Thick slice thickness Poor SR
Thin slice thickness Better SR
IMAGE FIDELITY
Large pixel size Poor SR
Measured by determining the optical density
Small pixel size Better SR
Large voxel size Poor SR along the axis of the image
Small voxel size Better SR
Large detector size Poor SR LIMITING RESOLUTION
Small detector size Better SR Spatial frequency at MTF equal to 0.1
13
COMPUTED TOMOGRAPHY