DMA-1 Application Book
DMA-1 Application Book
Application Book
Application field: Beverage
Milestone Srl – Helping Chemists www.milestonesrl.com [email protected] Rev. 01/13 Page 1/10
Application field: Cosmetic
Milestone Srl – Helping Chemists www.milestonesrl.com [email protected] Rev. 01/13 Page 2/10
Application field: Food/Feed
Application DMA1/FO‐01: Fresh Milk
Milestone Srl – Helping Chemists www.milestonesrl.com [email protected] Rev. 01/13 Page 3/10
Application DMA1/FO‐22: Fresh Meat
Milestone Srl – Helping Chemists www.milestonesrl.com [email protected] Rev. 01/13 Page 4/10
Application DMA1/GE‐12 : GSR‐3 Basalt
Milestone Srl – Helping Chemists www.milestonesrl.com [email protected] Rev. 01/13 Page 5/10
Application field: Pharmaceutical
Milestone Srl – Helping Chemists www.milestonesrl.com [email protected] Rev. 01/13 Page 7/10
Application DMA1/SC‐13: BCR‐184
Milestone Srl – Helping Chemists www.milestonesrl.com [email protected] Rev. 01/13 Page 8/10
Application Note: DMA‐1/BE01 Field: Beverage Spirit (alcohol 40%)
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Spirit can be performed using DMA‐1 Direct and weigh 500 mg of Add. A.
Mercury Analyzer. Such an instrument requires no 2. Add the sample.
sample wet chemistry or pre‐treatment. 3. Introduce the into sample lift.
4. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temp: 200°C
Instrumentation Purge: 60 sec
Additive:
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Wine can be performed using DMA‐1 Direct and weigh 500 mg of Add. A.
Mercury Analyzer. Such an instrument requires no 2. Add the sample.
sample wet chemistry or pre‐treatment. 3. Introduce the boat into sample lift.
4. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temp: 200°C
Instrumentation
Purge: 60 sec
Additive:
Summary
Procedure
Precise and rapid determination of total mercury in
Wine can be performed using DMA‐1 Direct 1. Place a boat on the balance plate, tare it
Mercury Analyzer. Such an instrument requires no 2. Add the sample.
sample wet chemistry or pre‐treatment. 3. Introduce the boat into sample lift.
4. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 250°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temp: 250°C
Purge: 60 sec
Instrumentation
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Results μg/kg Statistical Data
with DMA‐1 software or DMA‐1 PC software, metal Sample 1 avg: sd: rsd:
boats. 1.22‐1.32‐1.14 1.23 μg/kg 0.07 μg/kg 5.76 %
Sample 2 avg: sd: rsd:
Analytical balance, spatula, pipette, or
1.50‐1.48‐1.28 1.42 μg/kg 0.04 μg/kg 6.87 %
appropriate mechanical pipette and volumetric flask
(Class A), 50 or 100 ml.
Conclusion
Sample weight:
The DMA‐1 Mercury Analyzer successfully processed
Up to 100 mg (max) Sugar sample. Total analysis time per sample was less
than 6 minutes, including the time employed to weigh
Additive: each sample into the boat.
No
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Wine can be performed using DMA‐1 Direct 2. Add the sample.
Mercury Analyzer. Such an instrument requires no 3. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 4. Run the DMA‐1 program to completion.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
AgNO3 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 0.85
with DMA‐1 software or DMA‐1 PC software, metal
2 1.03
boats.
3 1.00
Analytical balance, spatula, pipette, or 4 0.98
appropriate mechanical pipette and volumetric flask 5 0.82
(Class A), 50 or 100 ml.
Avg: 0.94 µg/kg SD: 0.08 µg/kg RSD: 8.73 %
Sample weight :
80‐ 100 mg
Conclusion
The sample has been mixed and then introduced The DMA‐1 Mercury Analyzer successfully processed
into the metal boat. AgNO3 sample. Total analysis time per sample was less
than 6 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
NaOH and NaOCL can be performed using DMA‐1 and weigh 300 mg of additive.
Direct Mercury Analyzer. Such an instrument 2. Add the sample.
requires no sample wet chemistry or pre‐treatment. 3. Introduce the boat into sample lift.
4. Run the DMA‐1program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temp: 200°C
Instrumentation Purge: 60 sec
Analytical balance, spatula, pipette, or NaOCL 1.70 – 1.80 μg/kg rsd: 2.92 %
appropriate mechanical pipette and volumetric flask
(Class A), 50 or 100 ml.
Conclusion
Sample weight :
The DMA‐1 Mercury Analyzer successfully processed
Up to 50‐100 mg (max)
NaOH and NaOCl sample. Total analysis time per
Additive: sample was less than 7 minutes, including the time
employed to weigh each sample into the boat.
Silica gel
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
FeCl3 can be performed using DMA‐1 Direct and weigh 500 mg of additive.
Mercury Analyzer. Such an instrument requires no 2. Add the sample.
sample wet chemistry or pre‐treatment. 3. Introduce the boat into sample lift.
4. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temp: 200°C
Instrumentation Purge: 60 sec
Sample weight : Avg: 93.57 µg/kg SD: 2.59 µg/kg RSD: 2.77 %
Up to 50‐100 mg (max)
Additive: Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Aluminum oxide, activity grade: Super I FeCl3 sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
KBr can be performed using DMA‐1 Direct Mercury and weigh 500 mg of additive.
Analyzer. Such an instrument requires no sample 2. Add the sample.
wet chemistry or pre‐treatment. 3. Introduce the boat into sample lift.
4. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temp: 200°C
Instrumentation Purge: 60 sec
Sample weight : Avg: 0.42 µg/kg SD: 0.02 µg/kg RSD: 4.97 %
Up to 50‐100 mg (max)
Additive: Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Aluminum oxide, activity grade: Super I KBr sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
KIO3 can be performed using DMA‐1 Direct and weigh 500 mg of additive.
Mercury Analyzer. Such an instrument requires no 2. Add the sample.
sample wet chemistry or pre‐treatment. 3. Introduce the boat into the sample lift.
4. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temp: 200°C
Instrumentation Purge: 60 sec
Sample weight : Avg: 10.38 µg/kg SD: 0.41 µg/kg RSD: 3.99 %
Up to 50‐100 mg (max)
Additive: Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Aluminum oxide, activity grade: Super I KIO3 sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
CaSO4 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 11.8
with DMA‐1 software or DMA‐1 PC software, metal
2 10.4
boats.
3 11
Analytical balance, spatula, pipette, or 4 10.2
appropriate mechanical pipette and volumetric flask 5 10.7
(Class A), 50 or 100 ml.
Avg: 11 µg/kg SD: 0.1 µg/kg RSD: 5.78 %
Sample weight :
50‐ 100 mg
Conclusion
The sample has been mixed and then introduced The DMA‐1 Mercury Analyzer successfully processed
into the metal boat. CaSO4 sample. Total analysis time per sample was less
than 6 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
MgO can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 42.2
with DMA‐1 software or DMA‐1 PC software, metal
2 45.6
boats.
3 43.1
Analytical balance, spatula, pipette, or 4 45.7
appropriate mechanical pipette and volumetric flask 5 44.3
(Class A), 50 or 100 ml.
Avg: 44.2 µg/kg SD: 1.5 µg/kg RSD: 3.47 %
Sample weight :
50‐ 100 mg
Conclusion
The sample has been mixed and then introduced
The DMA‐1 Mercury Analyzer successfully processed
into the metal boat. MgO sample. Total analysis time per sample was less
than 6 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
MgO can be performed using DMA‐1 Direct and weigh 500 mg of additive.
Mercury Analyzer. Such an instrument requires no 2. Add the sample.
sample wet chemistry or pre‐treatment. 3. Introduce the boat into sample lift.
Once a weighed sample portion is introduced into 4. Run the DMA‐1 program to completion.
the instrument, analysis is completed in six
DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition,
N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic
1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA
2 00:02:00 650°C
7473 and is validated for laboratory as well as field
3 00:01:00 650°C
analysis.
Max start temp: 200°C
Instrumentation Purge: 60 sec
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Whole Blood can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Sample ID Hg Statistical data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
with DMA‐1 software or DMA‐1 PC software, metal Sample 1 8.19 ± 0.06 µg/kg rsd: 0.73 %
Sample 2 30.3 ± 0.03 µg/kg rsd: 0.99 %
boats.
Instrumentation RESULTS
Direct Mercury Analyzer apparatus and supplies Sample ID Hg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Sample 1 65.49 ± 0.9 µg/kg
with DMA‐1 software or DMA‐1 PC software, metal
Sample 2 1.16 ± 0.17 µg/kg
boats.
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Whole Blood sample. Total analysis time per sample
was less than 7 minutes, including the time employed
to weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Human Hair can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° mg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 0.42
with DMA‐1 software or DMA‐1 PC software, metal
2 0.41
boats.
3 0.42
Analytical balance, spatula, pipette, or 4 0.40
appropriate mechanical pipette and volumetric flask 5 0.41
(Class A), 50 or 100 ml.
Avg: 0.41 mg/kg SD: 0.006 mg/kg RSD: 1.59 %
Sample weight :
Conclusion
10‐50 mg
The DMA‐1 Mercury Analyzer successfully processed
Human Hair sample. Total analysis time per sample
was less than 7 minutes, including the time employed
to weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Beard can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° mg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 0.52
with DMA‐1 software or DMA‐1 PC software, metal
2 0.55
boats.
3 0.50
Analytical balance, spatula, pipette, or 4 0.49
appropriate mechanical pipette and volumetric flask 5 0.49
(Class A), 50 or 100 ml.
Avg: 0.51 mg/kg SD: 0.022 µg/kg RSD: 4.38 %
Sample weight :
Conclusion
5‐10 mg
The DMA‐1 Mercury Analyzer successfully processed
Beard sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Nails can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results mg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
0.33 – 0.32 – avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, metal
0.33 – 0.34 0.30 mg/kg 0.018 mg/kg 2.65 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Urine can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Lipstick can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
Max start temp: 200°C
7473 and is validated for laboratory as well as field
Purge: 60 sec
analysis.
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies N° µg/kg
1 1.45
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
2 1.74
with DMA‐80 software or DMA‐80 PC software,
3 1.59
quartz boats. 4 1.32
5 1.07
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric flask
(Class A), 50 or 100 ml. Avg: 1.44 µg/kg SD: 0.23 µg/kg RSD: 16.02 %
Sample weight :
Conclusion
Up to 20 mg (max)
The DMA‐1 Mercury Analyzer successfully processed
The sample has been mixed and then directly Lipstick sample. Total analysis time per sample was less
introduced into the metal boat. than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Face Powder can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
Max start temp: 200°C
7473 and is validated for laboratory as well as field
Purge: 60 sec
analysis.
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies N° µg/kg
1 64.74
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
2 60.47
with DMA‐80 software or DMA‐80 PC software,
3 63.79
quartz boats. 4 57.48
5 60.88
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric flask
(Class A), 50 or 100 ml. Avg: 61.48 µg/kg SD: 2.58 µg/kg RSD: 4.20 %
Sample weight :
Conclusion
Up to 20 mg (max)
The DMA‐1 Mercury Analyzer successfully processed
The sample has been mixed and then directly Face Powder sample. Total analysis time per sample
introduced into the metal boat. was less than 6 minutes, including the time employed
to weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Shampoo can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
Max start temp: 200°C
7473 and is validated for laboratory as well as field
Purge: 60 sec
analysis.
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies N° µg/kg
1 0.57
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
2 0.69
with DMA‐80 software or DMA‐80 PC software, 3 0.60
quartz boats. 4 0.50
5 0.54
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric flask
(Class A), 50 or 100 ml. Avg: 0.58 µg/kg SD: 0.06 µg/kg RSD: 10.89 %
Sample weight :
Conclusion
Up to 50 mg (max)
The DMA‐1 Mercury Analyzer successfully processed
The sample in liquid form has been directly Shampoo sample. Total analysis time per sample was
introduced into the metal boat and then less than 6 minutes, including the time employed to
weigh each sample into the boat.
analyzed.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Eye‐Shadow can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
Max start temp: 200°C
7473 and is validated for laboratory as well as field
Purge: 60 sec
analysis.
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies N° µg/kg
1 0.79
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
2 0.69
with DMA‐80 software or DMA‐80 PC software, 3 0.80
quartz boats. 4 0.75
5 0.73
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric flask
(Class A), 50 or 100 ml. Avg: 0.75 µg/kg SD: 0.04 µg/kg RSD: 5.50 %
Sample weight :
Conclusion
Up to 50 mg (max)
The DMA‐1 Mercury Analyzer successfully processed
The sample in liquid form has been directly Eye‐Shadow sample. Total analysis time per sample
introduced into the metal boat and then was less than 6 minutes, including the time employed
to weigh each sample into the boat.
analyzed.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Coal can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° mg/kg N° mg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 0.15 ± 0.02 5 0.10 ± 0.02
with DMA‐80 software or DMA‐80 PC software,
2 0.148 ± 0.02 6 0.10 ± 0.02
metal boats.
3 0.039 ± 0.02 7 0.09 ± 0.02
Analytical balance, spatula, pipette, or 4 0.049 ± 0.02 8 0.037 ± 0.02
appropriate mechanical pipette and volumetric flask
(Class A), 50 or 100 ml. Conclusion
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Feathers can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:00:10 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Analysis result Expected Statistical data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
mg/kg mg/Kg (RSD)
with DMA‐80 software or DMA‐80 PC software,
22 ±1 27 5.17 %
metal boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Paint Chips can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
DMA‐80 EPA EPA Method
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Sample ID
Method 7473 7471
with DMA‐80 software or DMA‐80 PC software, 98‐3898 4.5 ± 1ppm 0.19
metal boats. 98‐3902 0.41 ± 0.06 ppm 0.17
98‐3912 0.19 ± 0.03 ppm <0.08
Analytical balance, spatula, pipette, or
98‐3924 10 ± 3 ppm 7.00
appropriate mechanical pipette and volumetric flask
98‐3954 0.21 ± 0.05 ppm <0.085
(Class A), 50 or 100 ml.
98‐3956 0.23 ± 0.04 ppm <0.086
Sample weight :
Conclusion
Up to 100 mg (max)
The DMA‐1 Mercury Analyzer successfully processed
The sample has been chipped to approx. 0.5 cm and the Paint Chips samples using oxygen as the carrier
due to the organic content of the sample, the largest gas. Since this method does not require sample
sample size should be analyzed by this instrument is preparation, the total analysis time per sample was
100 mg less than 10 minutes, including the time taken to weigh
each sample into the boat. The scatter in the reported
data is due to inhomogeneity of the original site
samples. Even gross visual examination revealed that
each site sample represented a distribution of material
thickness and number of layers of paint and, therefore,
the potential mercury content.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Scrubber Sludge can be performed using DMA‐1 and weigh the sample.
Direct Mercury Analyzer. Such an instrument 2. Introduce the boat into sample lift.
requires no sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:00:10 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Sample ID Statistical Data
mg/kg
with DMA‐80 software or DMA‐80 PC software,
0.194
metal boats. Scrubber Sludge avg: 0.193±0.001 rsd:
0.194
dried at 200 °Cs mg/kg 0.49 %
0.192
Analytical balance, spatula, pipette, or
Flue Gas 0.025
appropriate mechanical pipette and volumetric flask avg: 0.025±0.002 rsd:
Desulfurization 0.027
(Class A), 50 or 100 ml. mg/kg 6.73 %
Scrubber Sludge 0.023
Sample weight :
Conclusion
Up to 250 mg (max)
The DMA‐1 Mercury Analyzer successfully processed
Scrubber Sludge dried at 200 °C and Flue Gas Scrubber Sludge sample. Total analysis time per
Desulphurisation Scrubber Sludge samples were sample was less than 6 minutes, including the time
used for this application. The sample has been employed to weigh each sample into the boat.
directly introduced into the metal boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Wastewater can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Statistical Data
μg/kg
with DMA‐80 software or DMA‐80 PC software,
5.82
quartz boats. avg: sd: rsd:
5.23
5,65 μg/kg 0.03μg/kg 5.33 %
5.91
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric flask
(Class A), 50 or 100 ml. Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Sample weight :
Wastewater sample. Total analysis time per sample
Up to 250 mg (max) was less than 7 minutes, including the time employed
to weigh each sample into the boat.
Instrumentation Results
Statistical Data
μg/kg
Direct Mercury Analyzer apparatus and supplies
9.96
avg: sd: rsd:
9.97
Milestone DMA‐1, 640‐12 and 1640‐12 terminal 9.78 μg/kg 0.20 μg/kg 1.57 %
9.58
with DMA‐80 software or DMA‐80 PC software,
quartz boats.
Conclusion
Analytical balance, spatula, pipette, or
The DMA‐1 Mercury Analyzer successfully processed
appropriate mechanical pipette and volumetric flask
Wastewater with CaCl2 sample. Total analysis time per
(Class A), 50 or 100 ml.
sample was less than 7 minutes, including the time
Sample weight : employed to weigh each sample into the boat.
Up to 200 mg (max)
Procedure
Milestone Srl – Helping Chemists www.milestonesrl.com [email protected] Rev. 01/13 1
Application Note: DMA‐1/EN06 Field: Environmental Wood
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Wood can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Statistical Data
μg/kg
with DMA‐1 software or DMA‐1 PC software, metal
7.72
boats.
7.73
avg: sd: rsd:
8.03
Analytical balance, spatula, pipette, or 7.79 μg/kg 0.13 μg/kg 1.71 %
appropriate mechanical pipette and volumetric flask 7.65
(Class A), 50 or 100 ml.
7.84
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Soil can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:00:10 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results mg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
56.86 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, metal
54.08 55.27 mg/kg 1.39 mg/kg 2.50 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Moss can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
24.29 ‐ 23.49
with DMA‐1 software or DMA‐1 PC software, metal 23.77 ‐ 23.60
boats. avg: sd: rsd:
24.11 ‐ 23.13
23.84 μg/kg 0.36 μg/kg 1.51 %
23.60 ‐ 24.29
Analytical balance, spatula, pipette, or 23.92 ‐ 24.11
appropriate mechanical pipette and volumetric flask
(Class A), 50 or 100 ml.
Conclusion
Sample weight : The DMA‐1 Mercury Analyzer successfully processed
Moss sample. Total analysis time per sample was less
Up to 50 mg (max)
than 7 minutes, including the time employed to weigh
The sample in powder from has been mixed and each sample into the boat.
then directly introduced into the metal boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Soil can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:00:10 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results mg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
38.21
with DMA‐1 software or DMA‐1 PC software, metal avg: sd: rsd:
40.90
boats. 23.84 mg/kg 0.36 mg/kg 1.51 %
39.38
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Seaweed powder can be performed using DMA‐1 and weigh the sample.
Direct Mercury Analyzer. Such an instrument 2. Introduce the boat into sample lift.
requires no sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
10.30 ‐ 10.29
with DMA‐1 software or DMA‐1 PC software, avg: sd: rsd:
10.68 ‐ 9.47
metal boats. 10.39 μg/kg 0.57 μg/kg 5.48 %
11.21
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Slag can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 700°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 700°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
2.27 – 2.46 – 2.50 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, metal
– 2.33 – 2.11 2.33 μg/kg 0.14 μg/kg 5.98 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Saline Water can be performed using DMA‐1 Direct and weigh 200 mg of Additive.
Mercury Analyzer. Such an instrument requires no 2. Add the sample.
sample wet chemistry or pre‐treatment. 3. Introduce the boat into sample lift.
4. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the DMA-1 program
integrated sequence of Thermal Decomposition,
Catalyst Conversion, Amalgamation, and Atomic N° step Time Temperature
Absorption Spectrophotometer, is described in EPA 1 00:01:00 200°C
7473 and is validated for laboratory as well as field 2 00:03:30 650°C
analysis. 3 00:03:00 650°C
Max start temperature: 200°C
Instrumentation Purge: 60 sec
Direct Mercury Analyzer apparatus and supplies
Results
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
with DMA‐1 software or DMA‐1 PC software, Results μg/kg Statistical Data
quartz boats. avg: sd: rsd:
1.10 – 0.96 – 1.03
1.03 μg/kg 0.06 μg/kg 5.56 %
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric
flask(Class A), 50 or 100 ml. Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Sample weight : Saline Water sample. Total analysis time per sample
Up to 200 mg (max) was less than 10 minutes, including the time
employed to weigh each sample into the boat.
Additive:
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Sludge can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 700°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 700°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
25.34 – 26.46 –
with DMA‐1 software or DMA‐1 PC software, avg: sd: rsd:
25.18 – 27.11–
metal boats. 25.94 μg/kg 0.82 μg/kg 3.16 %
25.59
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Sediment can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 250°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 250°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
11.3 – 11.7 – 11.1 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, metal
– 12.2– 12.0 11.65 μg/kg 0.46 μg/kg 3.97 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Fresh milk can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 0.0
with DMA‐1 software or DMA‐1 PC software,
2 0.0
quartz boats.
3 0.0
Analytical balance, spatula, pipette, or 4 0.0
appropriate mechanical pipette and volumetric flask 5 0.0
(Class A), 50 or 100 ml.
Expected Hg content: 0.0 μg/kg
Sample weight :
Conclusion
Up to 100 mg (max)
The DMA‐1 Mercury Analyzer successfully processed
The sample has been cut in small size and then Fresh milk sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
directly introduced into the metal boat.
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Fresh Yogurt can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
0.18 – 0.18 – 0.11 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, metal
0.12 – 0.14 0.15 μg/kg 0.03 μg/kg 20.36 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Cocoa can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
0.22 – 0.39 – 0.25 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, metal
– 0.37 – 0.36 0.32 μg/kg 0.07 μg/kg 21.70 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Coffee can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1.18 – 1.49 – 1.46 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, metal
– 1.21 – 1.31 1.33 μg/kg 0.12 μg/kg 9.29 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Cornflakes can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
0.81 – 0.74 – 0.52 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, metal
– 0.63 – 0.67 0.68 μg/kg 0.10 μg/kg 14.54 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Dried tea leaves can be performed using DMA‐1 and weigh the sample.
Direct Mercury Analyzer. Such an instrument 2. Introduce the boat into sample lift.
requires no sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1.28 – 1.42 – 1.49 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, metal
– 1.34 – 1.24 1.36 μg/kg 0.09 μg/kg 6.68 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Honey can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
0.40 – 0.26 – 0.30 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, metal
0.34 – 0.31 0.29 μg/kg 0.09 μg/kg 32.36 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Fresh Banana can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
0.13 – 0.15 – 0.09 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, metal
– 0.16 – 0.17 0.11 μg/kg 0.03 μg/kg 21.53 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Lemon Peels can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
0.80 – 0.99 – 0.68 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, metal
0.71– 0.90 0.82 μg/kg 0.12 μg/kg 14.08 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Dried Apple can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 e 1640‐12 terminal with
2.83 – 2.70 – 2.34 avg: sd: rsd:
DMA‐1 software or DMA‐1 PC software, metal
– 2.54 – 3.49 2.78 μg/kg 0.39 μg/kg 14.09 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Fresh Grapefruit Peels can be performed using and weigh the sample.
DMA‐1 Direct Mercury Analyzer. Such an 2. Introduce the boat into sample lift.
instrument requires no sample wet chemistry or 3. Run the DMA‐1 program to completion.
pre‐treatment.
DMA-1 program
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six N° step Time Temperature
minutes. Direct analysis of mercury, using the 1 00:01:00 200°C
integrated sequence of Thermal Decomposition, 2 00:02:00 650°C
Catalyst Conversion, Amalgamation, and Atomic 3 00:01:00 650°C
Absorption Spectrophotometer, is described in EPA Max start temp: 200°C
7473 and is validated for laboratory as well as field Purge: 60 sec
analysis.
Instrumentation Results
Direct Mercury Analyzer apparatus and supplies Results μg/kg Statistical Data
5.14 – 4.87 – 5.23 avg: sd: rsd:
Milestone DMA‐80, 640‐12 and 1640‐12 terminal
5.05 5.07 μg/kg 0.13 μg/kg 2.60 %
with DMA‐1 software or DMA‐1 PC software, metal
boats.
Conclusion
Analytical balance, spatula, pipette, or
The DMA‐1 Mercury Analyzer successfully processed
appropriate mechanical pipette and volumetric flask
Fresh Grapefruit Peels sample. Total analysis time per
(Class A), 50 or 100 ml.
sample was less than 6 minutes, including the time
Sample weight : employed to weigh each sample into the boat.
Up to 85 mg (max)
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Fresh Pear can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1.85 – 1.52 – 1.91 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, metal
1.57 – 1.56 1.68 μg/kg 0.16 μg/kg 9.66 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Rice Flour can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
3.74 ‐ 3.67 ‐ 3.82
with DMA‐1 software or DMA‐1 PC software, metal 4.60 ‐ 4.17 ‐ 4.31 avg: sd: rsd:
boats. 3.81 ‐ 4.21 ‐ 3.96 4.06 μg/kg 0.20 μg/kg 7.18 %
4.34
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric flask
Conclusion
(Class A), 50 or 100 ml.
The DMA‐1 Mercury Analyzer successfully processed
Sample weight : Rice Flour sample. Total analysis time per sample was
Up to 110 mg (max) less than 6 minutes, including the time employed to
weigh each sample into the boat.
The sample has been cut in small size and then
directly introduced into the metal boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Pasta can be performed using Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1.67 – 1.69 – 1.63 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, metal
1.65 – 1.69 1.76 μg/kg 0.92 μg/kg 1.33 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Fresh Tomato Pulp can be performed using DMA‐1 and weigh the sample.
Direct Mercury Analyzer. Such an instrument 2. Introduce the boat into sample lift.
requires no sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
0.10 – 0.13 – 0.14 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, metal
0.20 – 0.12 0.14 μg/kg 0.03 μg/kg 23.52 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Alligator Fat can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 160°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results mg/kg Expected Hg Statistical data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
rsd:
with DMA‐1 software or DMA‐1 PC software, 0.324 ± 0.1 0.275 mg /kg
4.5 %
quartz boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Canned Tuna Fish in Oil can be performed using and weigh the sample.
DMA‐1 Direct Mercury Analyzer. Such an 2. Introduce the boat into sample lift.
instrument requires no sample wet chemistry or 3. Run the DMA‐1 program to completion.
pre‐treatment.
DMA-1 program
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six N° step Time Temperature
minutes. Direct analysis of mercury, using the 1 00:01:00 160°C
integrated sequence of Thermal Decomposition, 2 00:02:00 650°C
Catalyst Conversion, Amalgamation, and Atomic 3 00:01:00 650°C
Absorption Spectrophotometer, is described in EPA Max start temp: 200°C
7473 and is validated for laboratory as well as field Purge: 60 sec
analysis.
Instrumentation Results
Direct Mercury Analyzer apparatus and supplies Results mg/kg Statistical data
0.57 ± 0.2 rsd: 4.5 %
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
with DMA‐80 software or DMA‐1 PC software,
quartz boats. Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Analytical balance, spatula, pipette, or
Canned Tuna Fish in Oil sample. Total analysis time per
appropriate mechanical pipette and volumetric flask
sample was less than 6 minutes, including the time
(Class A), 50 or 100 ml.
employed to weigh each sample into the boat.
Sample weight :
Up to 100 mg (max)
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Fish Tissue can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-80 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 160°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results mg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
0.149 ± 0.006
with DMA‐1 software or DMA‐1 PC software, quartz
boats. 0.130 ± 0.006
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Fresh Salmon can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
0.129–0.136
with DMA‐1 software or DMA‐1 PC software, quartz avg: sd: rsd:
0.115–0.127
boats. 0.129 μg/kg 6.66 μg/kg 5.14 %
0.139
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Oyster and Mussel Tissues can be performed using and weigh the sample.
DMA‐1 Direct Mercury Analyzer. Such an 2. Introduce the boat into sample lift.
instrument requires no sample wet chemistry or 3. Run the DMA‐1 program to completion.
pre‐treatment.
DMA-1 program
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six N° step Time Temperature
minutes. Direct analysis of mercury, using the 1 00:01:00 200°C
integrated sequence of Thermal Decomposition, 2 00:02:00 650°C
Catalyst Conversion, Amalgamation, and Atomic 3 00:01:00 650°C
Absorption Spectrophotometer, is described in EPA Max start temp: 200°C
7473 and is validated for laboratory as well as field Purge: 60 sec
analysis.
Instrumentation Results
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Bovine Muscle and Liver can be performed using and weigh the sample.
DMA‐1 Direct Mercury Analyzer. Such an 2. Introduce the boat into sample lift.
instrument requires no sample wet chemistry or 3. Run the DMA‐1 program to completion.
pre‐treatment.
DMA-1 program
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six N° step Time Temperature
minutes. Direct analysis of mercury, using the 1 00:01:00 200°C
integrated sequence of Thermal Decomposition, 2 00:02:00 650°C
Catalyst Conversion, Amalgamation, and Atomic 3 00:01:00 650°C
Absorption Spectrophotometer, is described in EPA Max start temp: 200°C
7473 and is validated for laboratory as well as field Purge: 60 sec
analysis.
Instrumentation Results
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Fresh Meat and liver can be performed using DMA‐ and weigh the sample.
1 Direct Mercury Analyzer. Such an instrument 2. Introduce the boat into sample lift.
requires no sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:02:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
2.40 – 2.82 – 2.48 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, quartz
1.43 – 1.59 2.14 μg/kg 0.54 μg/kg 25.05 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Fresh Eggs can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Expected μg/kg
Milestone DMA‐1, 640‐12 e 1640‐12 terminal with
0.0 – 0.0 – 0.0
DMA‐1 software or DMA‐1 PC software, quartz 0.0
0.0 – 0.0
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Fresh Cheese can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
4.23 – 4.58 – 4.44 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, quartz
3.54 – 4.28 4.22 μg/kg 0.36 μg/kg 8.57 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Black Pepper can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
10.44 – 9.98
with DMA‐1 software or DMA‐1 PC software, quartz avg: sd: rsd:
9.80 – 10.07
boats. 9.71 μg/kg 0.32 μg/kg 3.30 %
9.46
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Fresh garlic can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
0.93 – 1.00 – 1.13 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, quartz
0.75 – 1.07 0.98 μg/kg 0.13 μg/kg 13.46 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Fresh onion can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
0.22 – 0.13 – 0.26 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, quartz
0.19 – 0.14 0.19 μg/kg 0.05 μg/kg 26.57 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Fresh zucchini can be performed using DMA‐1 and weigh the sample.
Direct Mercury Analyzer. Such an instrument 2. Introduce the boat into sample lift.
requires no sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
0.73 – 0.58 – 0.55 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, quartz
0.71 – 0.66 0.65 μg/kg 0.07 μg/kg 10.82 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Fresh fennel can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
0.68 – 0.58 – 0.63 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software,
0.69 – 0.78 0.67 μg/kg 0.07 μg/kg 9.75 %
quartz boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Animal Feed can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Results μg/kg Statistical Data
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
7.65 – 7.19 – 6.50 avg: sd: rsd:
with DMA‐1 software or DMA‐1 PC software, quartz
7.07 – 7.17 7.12 μg/kg 0.37 μg/kg 5.14 %
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Copper ore can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 2 00:03:00 650°C
7473 and is validated for laboratory as well as field Max start temperature: 300°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
Sample ID Hg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
with DMA‐1 software or DMA‐1 PC software, metal Copper ore 1110 ± 0.05 mg/kg
boats.
Copper ore diluted* 5.94 ± 0.06 mg/kg
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric
*Data of Copper ore diluted:
flask(Class A), 50 or 100 ml.
Dilution ( 0.05 gram of sample + 10,2680 grams of silica
Sample weight : gel 70‐230mesh).
Direct analysis after dilution by silica gel
Copper ore: Up to 0.6 mg (max) Copper ore= Avg of Copper ore x DF
Avg of Copper ore after dilution=5.94
Copper ore diluted*: Up to 60 mg (max)
DF= 10,2680/0.0555=185,00
Due to big amount of Hg content in Copper ore has Copper ore= 5.94 X 185=1098 mg/Kg
been preferred a dilution* of sample by silica gel Recovery : 98.9%
before analysis. After dilution the sample has been
directly introduced into the metal boat. Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Copper ore diluted sample. Total analysis time per
sample was less than 7 minutes, including the time
taken to weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Manganese can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:00:10 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 750°C
7473 and is validated for laboratory as well as field 3 00:01:00 750°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
The sample has been mixed and then introduced Manganese is not found as the free metal in nature,
into the metal boat. however manganese minerals consisting of oxides,
silicates, and carbonates are the common. Most
Manganese is obtained from ores in Australia, Brazil,
Gabon, India, Russia, and South Africa. Manganese
nodules on ocean floors holds contain about 24%
manganese.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Clay can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:00:10 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 700°C
7473 and is validated for laboratory as well as field 3 00:01:00 700°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Results μg/kg Statistical Data
with DMA‐1 software or DMA‐1 PC software, metal 6.05 – 6.67 – 6.46 avg: sd: rsd:
boats. 6.34 – 6.58 6.42 μg/kg 0.22 μg/kg 3.36 %
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Sulphur can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Add 100 mg of Additive.
sample wet chemistry or pre‐treatment. 3. Introduce the boat into sample lift.
4. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:00:10 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 700°C
7473 and is validated for laboratory as well as field 3 00:01:00 700°C
analysis. Max start temperature: 200°C
Instrumentation Purge: 60 sec
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Cement Raw can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:00:10 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 700°C
7473 and is validated for laboratory as well as field 3 00:01:00 700°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
BE‐N can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:00:10 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Results μg/kg Statistical Data
with DMA‐1 software or DMA‐1 PC software, metal 14.15‐13.94
avg: sd: rsd:
boats. 13.76 ‐14.16
13.94 μg/kg 0.16 μg/kg 1.17 %
13.76
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric Conclusion
flask(Class A), 50 or 100 ml.
The DMA‐1 Mercury Analyzer successfully processed
Sample weight : BE‐N sample. Total analysis time per sample was less
than 6 minutes, including the time taken to weigh each
Up to 100 mg (max) sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
DRN can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Results μg/kg Statistical Data
with DMA‐1 software or DMA‐1 PC software, metal 99.07 ‐ 93.28
avg: sd: rsd:
boats. 92.95 ‐ 96.37
94.26 μg/kg 3.22 μg/kg 3.42 %
89.61
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric Conclusion
flask(Class A), 50 or 100 ml.
The DMA‐1 Mercury Analyzer successfully processed
Sample weight : DRN sample. Total analysis time per sample was less
than 6 minutes, including the time taken to weigh each
Up to 100 mg (max) sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
FK‐N can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Results μg/kg Statistical Data
with DMA‐1 software or DMA‐1 PC software, metal 21.69 ‐ 22.07
avg: sd: rsd:
boats. 22.16 ‐ 21.74
21. 98 μg/kg 0.20 μg/kg 0.89 %
22.20 ‐ 21.99
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric Conclusion
flask(Class A), 50 or 100 ml.
The DMA‐1 Mercury Analyzer successfully processed
Sample weight : FK‐N sample. Total analysis time per sample was less
than 6 minutes, including the time taken to weigh each
Up to 100 mg (max) sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
GH can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
GSD‐1 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Results μg/kg Statistical Data
with DMA‐1 software or DMA‐1 PC software, metal 12.60‐13.31
avg: sd: rsd:
boats. 13.11‐12.09
12.55 μg/kg 0.62 μg/kg 4.93 %
11.65
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric Conclusion
flask(Class A), 50 or 100 ml.
The DMA‐1 Mercury Analyzer successfully processed
Sample weight : GSD‐1 sample. Total analysis time per sample was less
than 6 minutes, including the time taken to weigh each
Up to 100 mg (max) sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
GSN can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample tray.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Results μg/kg Statistical Data
with DMA‐80 software or DMA‐80 PC software, 47.97‐ 51.12
avg: sd: rsd:
metal boats. 46.46 ‐48.83
49.00 μg/kg 1.72 μg/kg 3.50 %
50.62
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric Conclusion
flask(Class A), 50 or 100 ml.
The DMA‐1 Mercury Analyzer successfully processed
Sample weight : GSN sample. Total analysis time per sample was less
than 6 minutes, including the time taken to weigh each
Up to 100 mg (max) sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
GSR‐3 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Results μg/kg Statistical Data
with DMA‐1 software or DMA‐1 PC software, metal 30.50‐32.00
avg: sd: rsd:
boats. 32.05‐29.44
31.27 μg/kg 1.12 μg/kg 3.59 %
32.41
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric Conclusion
flask(Class A), 50 or 100 ml.
The DMA‐1 Mercury Analyzer successfully processed
Sample weight : GSR‐3 sample. Total analysis time per sample was less
than 6 minutes, including the time taken to weigh each
Up to 100 mg (max) sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
GXR‐1 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Results μg/kg Statistical Data
with DMA‐1 software or DMA‐1 PC software, metal 3951.61 ‐ 3874.14 avg: sd:
rsd:
boats. 3765.27 ‐ 3883.56 3837.76 85.91
2.24 %
3714.22 μg/kg μg/kg
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric Conclusion
flask(Class A), 50 or 100 ml.
The DMA‐1 Mercury Analyzer successfully processed
Sample weight : GXR‐1 sample. Total analysis time per sample was less
than 6 minutes, including the time taken to weigh each
Up to 100 mg (max) sample into the boat.
The sample has been mixed and then introduced
into the metal boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
GXR‐2 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Results μg/kg Statistical Data
with DMA‐1 software or DMA‐1 PC software, metal 3061.80‐3150.73 avg: sd:
rsd:
boats. 3033.03‐2997.39 3063.15 50.98
1.66 %
3072.79 μg/kg μg/kg
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric Conclusion
flask(Class A), 50 or 100 ml.
The DMA‐1 Mercury Analyzer successfully processed
Sample weight : GXR‐2 sample. Total analysis time per sample was less
than 6 minutes, including the time taken to weigh each
Up to 100 mg (max) sample into the boat.
The sample has been mixed and then introduced
into the metal boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
GXR‐3 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Results μg/kg Statistical Data
with DMA‐1 software or DMA‐1 PC software, metal 370.30‐320.83 avg: sd:
rsd:
boats. 345.00‐ 334‐73 342.68 16.18
4.72 %
342.59 μg/kg μg/kg
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric Conclusion
flask(Class A), 50 or 100 ml.
The DMA‐1 Mercury Analyzer successfully processed
Sample weight : GXR‐3 sample. Total analysis time per sample was less
than 6 minutes, including the time taken to weigh each
Up to 100 mg (max) sample into the boat.
The sample has been mixed and then introduced
into the metal boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
GXR‐4 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Results μg/kg Statistical Data
with DMA‐1 software or DMA‐1 PC software, metal 108.44‐111.93 avg:
sd: rsd:
boats. 107.18‐104.42 107.76
2.46 μg/kg 2.28 %
106.81 μg/kg
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric Conclusion
flask(Class A), 50 or 100 ml.
The DMA‐1 Mercury Analyzer successfully processed
Sample weight : GXR‐4 sample. Total analysis time per sample was less
than 6 minutes, including the time taken to weigh each
Up to 100 mg (max) sample into the boat.
The sample has been mixed and then introduced
into the metal boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
GXR‐5 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-80 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Results μg/kg Statistical Data
with DMA‐1 software or DMA‐1 PC software, metal 184.83‐175.11 avg:
sd: rsd:
boats. 176.55‐181.15 178.89
3.60 μg/kg 2.01%
176.78 μg/kg
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric Conclusion
flask(Class A), 50 or 100 ml.
The DMA‐1 Mercury Analyzer successfully processed
Sample weight : GXR‐5 sample. Total analysis time per sample was less
than 6 minutes, including the time taken to weigh each
Up to 100 mg (max) sample into the boat.
The sample has been mixed and then introduced
into the metal boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Leather can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-80 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal 1 0.93
with DMA‐1 software or DMA‐1 PC software, 2 0.74
quartz boats. 3 0.81
4 0.74
Analytical balance, spatula, pipette, or 5 0.88
appropriate mechanical pipette and volumetric flask
(Class A), 50 or 100 ml. Avg: 0.82 µg/kg SD: 0.08 µg/kg RSD: 9.23 %
Sample weight :
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Crude oil can be performed using DMA‐1 Direct and weigh 300 mg of Add. A.
Mercury Analyzer. Such an instrument requires no 2. Add the sample.
sample wet chemistry or pre‐treatment. 3. Add 200 mg of Add. A covering all the
sample and then press it by spatula to obtain a
Once a weighed sample portion is introduced into
sandwich.
the instrument, analysis is completed in six
4. Introduce the boat into sample lift.
minutes. Direct analysis of mercury, using the 5. Run the DMA‐1 program to completion.
integrated sequence of Thermal Decomposition,
Catalyst Conversion, Amalgamation, and Atomic DMA-1 program
Absorption Spectrophotometer, is described in EPA
7473 and is validated for laboratory as well as field N° step Time Temperature
analysis. 1 00:01:00 200°C
Instrumentation
2 00:02:00 650°C
3 00:01:00 650°C
Direct Mercury Analyzer apparatus and supplies Max start temperature: 200°C
Purge: 60 sec
Milestone DMA‐1, 640‐12 and 1640‐16 terminal
with DMA‐1 software or DMA‐1 PC software, quartz
Results
boats.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Naphtha can be performed using DMA‐1 Direct and weigh 500 mg of Add. A.
Mercury Analyzer. Such an instrument requires no 2. Add the sample.
sample wet chemistry or pre‐treatment. 3. Introduce the boat into sample lift.
4. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:02:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Instrumentation Purge: 60 sec
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Virgin Naphtha can be performed using DMA‐1 and weigh 500 mg of Add. A.
Direct Mercury Analyzer. Such an instrument 2. Add the sample.
requires no sample wet chemistry or pre‐treatment. 3. Introduce the boat into sample lift.
4. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:02:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Kerosene can be performed using DMA‐1 Direct and weigh 500 mg of Add. A.
Mercury Analyzer. Such an instrument requires no 2. Add the sample.
sample wet chemistry or pre‐treatment. 3. Introduce the boat into sample lift.
4. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:02:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Additive:
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Hydrocarbon mixture can be performed using DMA‐ and weigh 500 mg of Add. A.
1 Direct Mercury Analyzer. Such an instrument 2. Add the sample.
requires no sample wet chemistry or pre‐treatment. 3. Introduce the boat into sample lift.
4. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-80 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:02:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Additive:
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Gasoline can be performed using DMA‐1 Direct and weigh 500 mg of Add. A.
Mercury Analyzer. Such an instrument requires no 2. Add the sample.
sample wet chemistry or pre‐treatment. 3. Introduce the boat into sample lift.
4. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-80 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:02:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Purge: 60 sec
Instrumentation
Additive:
Summary Procedure
Precise and rapid determination of total mercury in 1. Introduce the sorbet trap A and B into sample
Gases via Sorbent trap can be performed using lift.
DMA‐1 Direct Mercury Analyzer. Such an 2. Introduce the volume amount.
instrument requires no sample wet chemistry or 3. Run the DMA‐1 program to completion.
pre‐treatment.
Once a sorbent trap is introduced into the DMA-1 program
instrument, analysis is completed in five minutes.
Direct analysis of mercury, using the integrated N° step Time Temperature
sequence of Catalyst Conversion, Amalgamation, 1 00:01:00 200°C
and Atomic Absorption Spectrophotometer. 2 00:02:00 650°C
3 00:01:00 650°C
Instrumentation
Max start temp: 200°C
Direct Mercury Analyzer apparatus and supplies Purge: 60 sec
Sorbent trap :
Sample 2 ng µg/m³
Trap A 52.32 0.0727
Trap B 0.1 ‐
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Gases via Sorbent trap. Total analysis time per sample
Trap A: Hg collection
was less than 5 minutes, including the time employed
Trap B: blank
to introduce the sample through the DMA‐1.
Flow rate during sampling: 200 ml/min
Total time (hours) : 6
Total volume (liters) : 72
Note: use a vacuum pump capable operating
within the system’s flow range.
Summary Procedure
Precise and rapid determination of total mercury in 1. Inject 100 ml of gas by syringe into DMA‐80 gas
GPL can be performed using DMA‐1 Direct Mercury inlet.
Analyzer. Such an instrument requires no sample 2. Introduce the volume amount.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to competion.
Once a weighed sample portion is introduced into DMA-80 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:30 50°C
Catalyst Conversion, Amalgamation, and Atomic Max start temperature: 80°C
Absorption Spectrophotometer, is described in EPA Purge: 60 sec
7473 and is validated for laboratory as well as field
analysis.
Results
Instrumentation
Sample 1 μg/m3
Direct Mercury Analyzer apparatus and supplies 100 ml 0.045
100 ml 0.041
Milestone DMA‐1, 640‐12 and 1640‐12 terminal 100 ml 0.043
with DMA‐1 software or DMA‐1 PC software, quartz
boats.
Sample 2 μg/m3
Carrier gas (SUGGESTED): air (DMA‐AC, Air 100 ml 0.070
Compressor supplied by Milestone) 100 ml 0.065
100 ml 0.062
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric Conclusion
flask(Class A), 50 or 100 ml.
The DMA‐1 Mercury Analyzer successfully processed
Gas sampling: GPL. Total analysis time per sample was less than 3
minutes, including time employed to introduce the
Flow rate during sampling: 200ml/min sample in the DMA‐1.
Total time (minutes): 25
Total volumes (litres): 5
Note: Use a vacuum pump capable operating
within the system flow range
Note the above mentioned parameters: flow rate
time and volume are guidelines and may be
modified according to your needs.
Summary Procedure
Direct Mercury Analyzer apparatus and supplies 3. Run the DMA‐1 program to completion.
Milestone DMA‐1, 640‐12 and 1640‐12 terminal DMA-1 program
with DMA‐1 software or DMA‐1 PC software
N° step Time Temperature
Carrier gas (SUGGESTED): air (DMA80‐AC, Air
1 00:00:00 20°C
Compressor supplied by Milestone).
2 00:01:30 20°C
Analytical balance, spatula, pipette, or Max start temp: 50°C
appropriate mechanical pipette and volumetric flask Purge: 60 sec
(Class A), 50 or 100 ml, syringe: 100 ml. Additional valves : on
Sample volume :
Results
Up to 100 cm³ (max)
Results μg/m³ Statistical Data
0.093‐0.086‐0.083 avg: rsd:
0.089‐ 0.080 0.086 μg/m³ 5.22 %
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Butane. Total analysis time per sample was less than 6
minutes, including the time employed to introduce the
sample through the injection module.
Summary Procedure
Direct Mercury Analyzer apparatus and supplies 3. Run the DMA‐1 program to completion.
Milestone DMA‐1, 640‐12 and 1640‐12 terminal DMA-1 program
with DMA‐1 software or DMA‐1 PC software.
N° step Time Temperature
Carrier gas (SUGGESTED): air (DMA80‐AC, Air
1 00:00:00 20°C
Compressor supplied by Milestone).
2 00:01:30 20°C
Analytical balance, spatula, pipette, or Max start temp: 50°C
appropriate mechanical pipette and volumetric flask Purge: 60 sec
(Class A), 50 or 100 ml, syringe: 100 ml. Additional valves: on
Sample volume :
Results
Up to 100 cm³ (max)
Results μg/m³ Statistical Data
0.027‐0.029‐0.032 avg: rsd:
0.031‐0.029 0.030 μg/m³ 5.07%
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Methane. Total analysis time per sample was less than
6 minutes, including the time employed to introduce
the sample through the injection module.
Summary Procedure
Precise and rapid determination of total mercury in 1. Introduce the sorbet trap A into the System.
Methane via Sorbent trap can be performed using 2. Introduce the volume amount.
DMA‐1 Direct Mercury Analyzer. Such an 3. Run the DMA‐1 program to completion.
instrument requires no sample wet chemistry or 4. Introduce the sorbet trap B into the System.
pre‐treatment. 5. Run the DMA‐1 program to completion.
Once a sorbent trap is introduced into the
instrument, analysis is completed in five minutes. DMA-1 program
Direct analysis of mercury, using the integrated
sequence of Catalyst Conversion, Amalgamation, N° step Time Temperature
and Atomic Absorption Spectrophotometer. 1 00:01:00 200°C
2 00:02:00 650°C
Instrumentation
3 00:01:00 650°C
Direct Mercury Analyzer apparatus and supplies Max start temp: 200°C
Purge: 60 sec
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
with DMA‐1 software or DMA‐1 PC software,
Results
Milestone sorbent trap
Sorbent trap :
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Methane via Sorbent trap. Total analysis time per
sample was less than 5 minutes, including the time
employed to introduce the sample through the DMA‐1.
Trap A: Hg collection
Trap B: blank
Flow rate during sampling: 200 ml/min
Total time (hours) : 20
Total volume (liters) :4
Note: use a MFC (Mass Flow Control) to determine
the volume of gas.
Note the above mentioned parameters: flow rate
time and volume are guidelines and may be
modified according to your needs.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Naphtha can be performed using DMA‐1 Direct and weigh 300 mg of Adsoquick.
Mercury Analyzer. Such an instrument requires no 2. Add the sample.
sample wet chemistry or pre‐treatment. 3. Introduce the boat into the sample lift.
4. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:02:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Instrumentation Purge: 60 sec
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Crude oil can be performed using DMA‐1 Direct and weigh 300 mg of Adsoquick.
Mercury Analyzer. Such an instrument requires no 2. Add the sample.
sample wet chemistry or pre‐treatment. 3. Introduce the boat into the System.
4. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six DMA-1 program
minutes. Direct analysis of mercury, using the
integrated sequence of Thermal Decomposition, N° step Time Temperature
Catalyst Conversion, Amalgamation, and Atomic 1 00:01:00 200°C
Absorption Spectrophotometer, is described in EPA 2 00:02:00 650°C
7473 and is validated for laboratory as well as field 3 00:01:00 650°C
analysis. Max start temperature: 200°C
Instrumentation
Purge: 60 sec
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Nimesulide can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
Nimesulide A 0.94 – 0.69
with DMA‐1 software or DMA‐1 PC software, metal
Nimesulide B 1.57 – 1.37
boats.
Nimesulide C 0.75 – 0.83
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric flask Conclusion
(Class A), 50 or 100 ml.
The DMA‐1 Mercury Analyzer successfully processed
Sample weight : Nimesulide sample. Total analysis time per sample was
less than 7 minutes, including the time employed to
Up to 100 mg (max) weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Paracetamol can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 0.00
with DMA‐1 software or DMA‐1 PC software, metal
2 0.00
boats.
3 0.00
Analytical balance, spatula, pipette, or 4 0.00
appropriate mechanical pipette and volumetric flask 5 0.00
(Class A), 50 or 100 ml.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Acetylsalicylic acid can be performed using DMA‐1 and weigh the sample.
Direct Mercury Analyzer. Such an instrument 2. Introduce the boat into sample lift.
requires no sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 1.43
with DMA‐1 software or DMA‐1 PC software, metal
2 1.29
boats.
3 1.20
Analytical balance, spatula, pipette, or 4 1.48
appropriate mechanical pipette and volumetric flask 5 1.32
(Class A), 50 or 100 ml.
Avg: 1.35 μg/kg SD: 0.1 μg/kg RSD: 7.52 %
Sample weight :
30‐50 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Acetylsalicylic acid sample. Total analysis time per
sample was less than 6 minutes, including the time
employed to weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Dietary Supplement can be performed using DMA‐1 and weigh the sample.
Direct Mercury Analyzer. Such an instrument 2. Introduce the boat into sample lift.
requires no sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 1.49
with DMA‐1 software or DMA‐1 PC software, metal
2 1.16
boats.
3 1.11
Analytical balance, spatula, pipette, or 4 1.18
appropriate mechanical pipette and volumetric flask 5 1.14
(Class A), 50 or 100 ml.
Avg: 1.22 μg/kg SD: 0.14 μg/kg RSD: 11.26 %
Sample weight :
30‐50 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Dietary Supplement sample. Total analysis time per
sample was less than 6 minutes, including the time
employed to weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Lactic Ferments can be performed using DMA‐1 and weigh the sample.
Direct Mercury Analyzer. Such an instrument 2. Introduce the boat into sample lift.
requires no sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 0.58
with DMA‐1 software or DMA‐1 PC software, metal
2 0.48
boats.
3 0.58
Analytical balance, spatula, pipette, or 4 0.66
appropriate mechanical pipette and volumetric flask 5 0.59
(Class A), 50 or 100 ml.
Avg: 0.58 μg/kg SD: 0.058 μg/kg RSD: 10.00 %
Sample weight :
30‐50 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Lactic Ferments sample. Total analysis time per sample
was less than 6 minutes, including the time employed
to weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Homeopathic Antihistamine can be performed and weigh the sample.
using DMA‐1 Direct Mercury Analyzer. Such an 2. Introduce the boat into sample lift.
instrument requires no sample wet chemistry or 3. Run the DMA‐1 program to completion.
pre‐treatment.
DMA-1 program
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six N° step Time Temperature
minutes. Direct analysis of mercury, using the 1 00:01:00 200°C
integrated sequence of Thermal Decomposition, 2 00:02:00 650°C
Catalyst Conversion, Amalgamation, and Atomic 3 00:01:00 650°C
Absorption Spectrophotometer, is described in EPA Max start temp: 200°C
7473 and is validated for laboratory as well as field Purge: 60 sec
analysis.
Instrumentation Results
Sample weight :
Conclusion
50‐100 mg (max)
The DMA‐1 Mercury Analyzer successfully processed
Homeopathic Antihistamine sample. Total analysis
time per sample was less than 6 minutes, including the
time employed to weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Homeopathic Laxative can be performed using and weigh the sample.
DMA‐1 Direct Mercury Analyzer. Such an 2. Introduce the boat into sample lift.
instrument requires no sample wet chemistry or 3. Run the DMA‐1 program to completion.
pre‐treatment.
DMA-1 program
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six N° step Time Temperature
minutes. Direct analysis of mercury, using the 1 00:01:00 200°C
integrated sequence of Thermal Decomposition, 2 00:02:00 650°C
Catalyst Conversion, Amalgamation, and Atomic 3 00:01:00 650°C
Absorption Spectrophotometer, is described in EPA Max start temp: 200°C
7473 and is validated for laboratory as well as field Purge: 60 sec
analysis.
Instrumentation Results
Sample weight :
Conclusion
50‐100 mg (max)
The DMA‐1 Mercury Analyzer successfully processed
Homeopathic Laxative sample. Total analysis time per
sample was less than 6 minutes, including the time
employed to weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Homeopathic Antihistamine from Stone can be and weigh the sample.
performed using DMA‐1 Direct Mercury Analyzer. 2. Introduce the boat into sample lift.
Such an instrument requires no sample wet 3. Run the DMA‐1 program to completion.
chemistry or pre‐treatment.
DMA-80 program
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six N° step Time Temperature
minutes. Direct analysis of mercury, using the 1 00:01:00 200°C
integrated sequence of Thermal Decomposition, 2 00:02:00 650°C
Catalyst Conversion, Amalgamation, and Atomic 3 00:01:00 650°C
Absorption Spectrophotometer, is described in EPA Max start temp: 200°C
7473 and is validated for laboratory as well as field Purge: 60 sec
analysis.
Instrumentation Results
Sample weight :
Conclusion
50‐100 mg (max)
The DMA‐1 Mercury Analyzer successfully processed
Homeopathic Antihistamine from Stone sample. Total
analysis time per sample was less than 6 minutes,
including the time employed to weigh each sample
into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Homeopathic Detoxificant can be performed using and weigh the sample.
DMA‐1 Direct Mercury Analyzer. Such an 2. Introduce the boat into sample lift.
instrument requires no sample wet chemistry or 3. Run the DMA‐1 program to completion.
pre‐treatment.
DMA-1 program
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six N° step Time Temperature
minutes. Direct analysis of mercury, using the 1 00:01:00 200°C
integrated sequence of Thermal Decomposition, 2 00:02:00 650°C
Catalyst Conversion, Amalgamation, and Atomic 3 00:01:00 650°C
Absorption Spectrophotometer, is described in EPA Max start temp: 200°C
7473 and is validated for laboratory as well as field Purge: 60 sec
analysis.
Instrumentation Results
Sample weight :
Conclusion
50‐100 mg (max)
The DMA‐1 Mercury Analyzer successfully processed
Homeopathic Detoxificant sample. Total analysis time
per sample was less than 6 minutes, including the time
employed to weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Homeopathic Painkiller can be performed using and weigh the sample.
DMA‐1 Direct Mercury Analyzer. Such an 2. Introduce the boat into sample lift.
instrument requires no sample wet chemistry or 3. Run the DMA‐1 program to completion.
pre‐treatment.
DMA-1 program
Once a weighed sample portion is introduced into
the instrument, analysis is completed in six N° step Time Temperature
minutes. Direct analysis of mercury, using the 1 00:01:00 200°C
integrated sequence of Thermal Decomposition, 2 00:02:00 650°C
Catalyst Conversion, Amalgamation, and Atomic 3 00:01:00 650°C
Absorption Spectrophotometer, is described in EPA Max start temp: 200°C
7473 and is validated for laboratory as well as field Purge: 60 sec
analysis.
Instrumentation Results
Sample weight :
Conclusion
30‐50 mg (max)
The DMA‐1 Mercury Analyzer successfully processed
Homeopathic Painkiller sample. Total analysis time per
sample was less than 6 minutes, including the time
employed to weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Teflon can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 5.2
with DMA‐1 software or DMA‐1 PC software, metal
2 5.9
boats.
3 6.2
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric flask Avg: 5.77 µg/kg SD: 0.42 µg/kg RSD: 7.27 %
(Class A), 50 or 100 ml.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
PC can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 1.50
with DMA‐1 software or DMA‐1 PC software, metal
2 1.29
boats.
3 1.51
Analytical balance, spatula, pipette, or 4 1.82
appropriate mechanical pipette and volumetric flask 5 1.77
(Class A), 50 or 100 ml.
Avg: 1.58 µg/kg SD: 0.19 µg/kg RSD: 12.36 %
Sample weight :
30‐50 mg (max)
Conclusion
PC has been cut in small pieces. The DMA‐1 Mercury Analyzer successfully processed
PC sample. Total analysis time per sample was less
than 6 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
PE can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 3.15
with DMA‐1 software or DMA‐1 PC software, metal
2 3.71
boats.
3 3.08
Analytical balance, spatula, pipette, or 4 3.28
appropriate mechanical pipette and volumetric flask 5 2.90
(Class A), 50 or 100 ml.
Avg: 3.29 µg/kg SD: 0.27 µg/kg RSD: 8.84 %
Sample weight :
30‐50 mg (max)
Conclusion
PE has been cut in small pieces The DMA‐1 Mercury Analyzer successfully processed
PE sample. Total analysis time per sample was less
than 6 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
PET can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 11.98
with DMA‐1 software or DMA‐1 PC software, metal
2 10.78
boats.
3 10.41
Analytical balance, spatula, pipette, or 4 13.65
appropriate mechanical pipette and volumetric flask 5 11.76
(Class A), 50 or 100 ml.
Avg: 11.72 µg/kg SD: 1.13 µg/kg RSD: 9.64 %
Sample weight :
15‐30 mg (max)
Conclusion
PET has been cut in small pieces. The DMA‐1 Mercury Analyzer successfully processed
PET sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
PFA can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 3.35
with DMA‐1 software or DMA‐1 PC software, metal
2 3.02
boats.
3 2.35
Analytical balance, spatula, pipette, or 4 2.59
appropriate mechanical pipette and volumetric flask 5 3.69
(Class A), 50 or 100 ml.
Avg: 3.00 µg/kg SD: 0.49 µg/kg RSD: 12.36 %
Sample weight :
20‐30 mg (max)
Conclusion
PFA has been cut in small pieces. The DMA‐1 Mercury Analyzer successfully processed
PFA sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
PP can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 4.33
with DMA‐1 software or DMA‐1 PC software, metal
2 4.91
boats.
3 5.67
Analytical balance, spatula, pipette, or 4 4.06
appropriate mechanical pipette and volumetric flask 5 4.62
(Class A), 50 or 100 ml.
Avg: 4.72 µg/kg SD: 0.55 µg/kg RSD: 11.72 %
Sample weight :
10‐30 mg (max)
Conclusion
PP has been cut in small pieces. The DMA‐1 Mercury Analyzer successfully processed
PP sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
PTFE can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐80, 640‐12 and 1640‐12 terminal
1 2.56
with DMA‐1 software or DMA‐1 PC software, metal
2 2.26
boats.
3 2.20
Analytical balance, spatula, pipette, or 4 1.79
appropriate mechanical pipette and volumetric flask 5 2.21
(Class A), 50 or 100 ml.
Avg: 2.20 µg/kg SD: 0.25 µg/kg RSD: 11.18 %
Sample weight :
Conclusion
20‐40 mg (max)
The DMA‐1 Mercury Analyzer successfully processed
PTFE has been cut in small pieces. PTFE sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
PVC can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 6.28
with DMA‐1 software or DMA‐1 PC software, metal
2 5.66
boats.
3 6.01
Analytical balance, spatula, pipette, or 4 6.04
appropriate mechanical pipette and volumetric flask 5 5.62
(Class A), 50 or 100 ml.
Avg: 5.84 µg/kg SD: 0.19 µg/kg RSD: 3.28 %
Sample weight :
Conclusion
20‐30 mg (max)
The DMA‐1 Mercury Analyzer successfully processed
PVC has been cut in small pieces. PVC sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Tygon can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐1 and 1640‐1 terminal with
1 13.96
DMA‐1 software or DMA‐1 PC software, metal
2 13.78
boats.
3 13.80
Analytical balance, spatula, pipette, or 4 14.41
appropriate mechanical pipette and volumetric flask 5 13.88
(Class A), 50 or 100 ml.
Avg: 13.97 µg/kg SD: 0.23 µg/kg RSD: 1.67 %
Sample weight :
30‐50 mg (max)
Conclusion
Tygon has been cut in small pieces. The DMA‐1 Mercury Analyzer successfully processed
Tygon sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
CSM can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐1 and 1640‐1 terminal with
1 15.70
DMA‐1 software or DMA‐1 PC software, metal
2 17.29
boats.
3 16.02
Analytical balance, spatula, pipette, or 4 15.68
appropriate mechanical pipette and volumetric flask 5 16.16
(Class A), 50 or 100 ml.
Avg: 16.17 µg/kg SD: 0.59 µg/kg RSD: 3.66 %
Sample weight :
20‐30 mg (max)
Conclusion
The sample has been cut in small pieces. The DMA‐1 Mercury Analyzer successfully processed
CSM sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
EPDM can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 2.49
with DMA‐1 software or DMA‐1 PC software, metal
2 2.27
boats.
3 2.18
Analytical balance, spatula, pipette, or 4 2.84
appropriate mechanical pipette and volumetric flask 5 2.50
(Class A), 50 or 100 ml.
Avg: : 2.46 µg/kg SD: 0.23 µg/kg RSD: 9.29 %
Sample weight :
Conclusion
30‐50 mg (max)
The DMA‐1 Mercury Analyzer successfully processed
The sample has been cut in small pieces. EPDM sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
FPM can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 5.39
with DMA‐1 software or DMA‐1 PC software, metal
2 5.32
boats.
3 5.20
Analytical balance, spatula, pipette, or 4 5.33
appropriate mechanical pipette and volumetric flask 5 5.20
(Class A), 50 or 100 ml.
Avg: 5.28 µg/kg SD: 0.06 µg/kg RSD: 1.23 %
Sample weight :
30‐60 mg (max)
Conclusion
The sample has been cut in small pieces. The DMA‐1 Mercury Analyzer successfully processed
FPM sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Koflon can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 2.10
with DMA‐1 software or DMA‐1 PC software, metal
2 2.31
boats.
3 2.44
Analytical balance, spatula, pipette, or 4 2.18
appropriate mechanical pipette and volumetric flask 5 2.66
(Class A), 50 or 100 ml.
Avg: 2.34 µg/kg SD: 0.20 µg/kg RSD: 8.53 %
Sample weight :
30‐40 mg (max)
Conclusion
The sample has been cut in small pieces. The DMA‐1 Mercury Analyzer successfully processed
Koflon sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
MFQ can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 640‐12 terminal with
1 2.95
DMA‐1 software or DMA‐1 PC software, metal
2 3.68
boats.
3 2.80
Analytical balance, spatula, pipette, or 4 3.42
appropriate mechanical pipette and volumetric flask 5 2.56
(Class A), 50 or 100 ml.
Avg: 3.09 µg/kg SD: 0.41 µg/kg RSD: 13.25 %
Sample weight :
30‐50 mg (max)
Conclusion
The sample has been cut in small pieces. The DMA‐1 Mercury Analyzer successfully processed
MFQ sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
MQ can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 2.50
with DMA‐1 software or DMA‐1 PC software, metal
2 2.43
boats.
3 2.19
Analytical balance, spatula, pipette, or 4 1.95
appropriate mechanical pipette and volumetric flask 5 2.62
(Class A), 50 or 100 ml.
Avg: 2.34 µg/kg SD: 0.24 µg/kg RSD: 10.24 %
Sample weight :
30‐50 mg (max)
Conclusion
The sample has been cut in small pieces. The DMA‐1 Mercury Analyzer successfully processed
MQ sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
NBR can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 4.15
with DMA‐1 software or DMA‐1 PC software, metal
2 3.95
boats.
3 3.39
Analytical balance, spatula, pipette, or 4 4.38
appropriate mechanical pipette and volumetric flask 5 4.15
(Class A), 50 or 100 ml.
Avg: 4.00 µg/kg SD: 0.34 µg/kg RSD: 8.43 %
Sample weight :
20‐30 mg (max)
Conclusion
The sample has been cut in small pieces. The DMA‐1 Mercury Analyzer successfully processed
NBR sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Nitril can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 5.25
with DMA‐1 software or DMA‐1 PC software, metal
2 3.84
boats.
3 4.83
Analytical balance, spatula, pipette, or 4 7.01
appropriate mechanical pipette and volumetric flask 5 7.54
(Class A), 50 or 100 ml.
Avg: 5.70 µg/kg SD: 1.38 µg/kg RSD: 24.26 %
Sample weight :
20‐30 mg (max)
Conclusion
The sample has been cut in small pieces. The DMA‐1 Mercury Analyzer successfully processed
Nitril sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
POM can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 7.03
with DMA‐1 software or DMA‐1 PC software, metal
2 5.95
boats.
3 6.56
Analytical balance, spatula, pipette, or 4 6.73
appropriate mechanical pipette and volumetric flask 5 6.23
(Class A), 50 or 100 ml.
Avg: 6.50 µg/kg SD: 0.38 µg/kg RSD: 5.80 %
Sample weight :
20‐30 mg (max)
Conclusion
The sample has been cut in small pieces. The DMA‐1 Mercury Analyzer successfully processed
POM sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
PSU can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 7.99
with DMA‐1 software or DMA‐1 PC software, metal
2 6.15
boats.
3 5.57
Analytical balance, spatula, pipette, or 4 5.39
appropriate mechanical pipette and volumetric flask 5 6.30
(Class A), 50 or 100 ml.
Avg: 6.28 µg/kg SD: 0.92 µg/kg RSD: 14.68 %
Sample weight :
20‐30 mg (max)
Conclusion
The sample has been cut in small pieces. The DMA‐1 Mercury Analyzer successfully processed
PSU sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
PU can be performed using DMA‐1 Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 19.84
with DMA‐1 software or DMA‐1 PC software, metal
2 23.23
boats.
3 20.73
Analytical balance, spatula, pipette, or 4 24.78
appropriate mechanical pipette and volumetric flask 5 21.39
(Class A), 50 or 100 ml.
Avg: 21.99 µg/kg SD: 1.78 µg/kg RSD: 8.11 %
Sample weight :
20‐30 mg (max)
Conclusion
The sample has been cut in small pieces. The DMA‐1 Mercury Analyzer successfully processed
PU sample. Total analysis time per sample was less
than 7 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Polyol can be performed using DMA‐1 Direct and weigh 300 mg of Add. A.
Mercury Analyzer. Such an instrument requires no 2. Add the sample.
sample wet chemistry or pre‐treatment. 3. Add 200 mg of Add. A covering all the
sample and then press it by spatula to obtain a
Once a weighed sample portion is introduced into
sandwich.
the instrument, analysis is completed in six
4. Introduce the boat into sample lift.
minutes. Direct analysis of mercury, using the 5. Run the DMA‐1 program to completion.
integrated sequence of Thermal Decomposition,
Catalyst Conversion, Amalgamation, and Atomic
Absorption Spectrophotometer, is described in EPA
7473 and is validated for laboratory as well as field DMA-1 program
analysis.
N° step Time Temperature
Instrumentation 1 00:01:00 200°C
2 00:03:30 650°C
Direct Mercury Analyzer apparatus and supplies
3 00:03:00 650°C
Milestone DMA‐1, 640‐12 and 1640‐12 terminal Max start temperature: 200°C
with DMA‐1 software or DMA‐1 PC software, quartz Purge: 60 sec
boats.
Results
Analytical balance, spatula, pipette, or
appropriate mechanical pipette and volumetric N° µg/kg
flask(Class A), 50 or 100 ml. 1 6.87
2 7.29
Sample weight :
3 7.60
Up to 30 mg (max) 4 7.56
5 7.68
Additive:
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Polyol sample. Total analysis time per sample was less
than 10 minutes, including the time employed to weigh
each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Nist 1568a can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 5.57
with DMA‐1 software or DMA‐1 PC software, metal
2 5.86
boats.
3 6.05
Analytical balance, spatula, pipette, or 4 6.10
appropriate mechanical pipette and volumetric flask 5 6.05
(Class A), 50 or 100 ml.
Certified value of Hg : 5.8 ± 0.5 μg/kg
Sample weight :
Avg: 5.92 µg/kg SD: 0.19 µg/kg RSD: 3.24 %
80‐100 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Nist 1568a sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Nist 1573a can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 30.16
with DMA‐1 software or DMA‐1 PC software, metal
2 28.38
boats.
3 30.62
Analytical balance, spatula, pipette, or 4 33.97
appropriate mechanical pipette and volumetric flask 5 29.94
(Class A), 50 or 100 ml.
Certified value of Hg : 34 ± 4 μg/kg
Sample weight :
Avg: 30.61 µg/kg SD: 1.76 µg/kg RSD: 5.69 %
80‐100 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Nist 1573a sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Nist 1630a can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 96.31
with DMA‐1 software or DMA‐1 PC software, metal
2 98.50
boats.
3 94.73
Analytical balance, spatula, pipette, or 4 95.17
appropriate mechanical pipette and volumetric flask 5 96.99
(Class A), 50 or 100 ml.
Certified value of Hg : 93.8 ± 3.7 μg/kg
Sample weight :
Avg: 96.35 µg/kg SD: 1.35 µg/kg RSD: 1.40 %
80‐100 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Nist 1630a sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Nist 1632c can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 94.41
with DMA‐1 software or DMA‐1 PC software, metal
2 92.43
boats.
3 95.20
Analytical balance, spatula, pipette, or 4 93.87
appropriate mechanical pipette and volumetric flask 5 94.13
(Class A), 50 or 100 ml.
Certified value of Hg : 93.8 ± 3.7 μg/kg
Sample weight :
Avg: 94.01 µg/kg SD: 0.91 µg/kg RSD: 0.96 %
80‐100 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Nist 1632c sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Nist 1633b can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:00:10 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 144.24
with DMA‐1 software or DMA‐1 PC software, metal
2 142.91
boats.
3 143.45
Analytical balance, spatula, pipette, or 4 141.84
appropriate mechanical pipette and volumetric flask 5 142.28
(Class A), 50 or 100 ml.
Certified value of Hg : 141 ± 19 μg/kg
Sample weight :
Avg: 142.95 µg/kg SD: 0.85 µg/kg RSD: 0.59 %
80‐100 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Nist 1633b sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Nist 2709 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:00:10 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° mg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 1.378
with DMA‐1 software or DMA‐1 PC software, metal
2 1.404
boats.
3 1.406
Analytical balance, spatula, pipette, or 4 1.424
appropriate mechanical pipette and volumetric flask
(Class A), 50 or 100 ml. Certified value of Hg: 1.40 ± 0.8 mg/kg
Sample weight : Avg: 1.40 mg/kg SD: 0.2 mg/kg RSD: 1.06 %
80‐100 mg
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Nist 2709 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Nist 2711 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:00:10 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° mg/kg
Milestone DMA‐1, 640‐12 e 1640‐12 terminal with
1 6.13
DMA‐1 software or DMA‐1 PC software, metal
2 6.24
boats.
3 6.27
Analytical balance, spatula, pipette, or 4 6.22
appropriate mechanical pipette and volumetric flask 5 6.13
(Class A), 50 or 100 ml.
Certified value of Hg: 6.25 ± 0.19 mg/kg
Sample weight :
Avg: 6.20 mg/kg SD: 0.058 mg/kg RSD: 0.94 %
80‐100 mg
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Nist 2711 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
aqueous standard can be performed using DMA‐1 and weigh the sample.
Direct Mercury Analyzer. Such an instrument 2. Introduce the boat into sample lift.
requires no sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 e 1640‐12 terminal with
1 103.02
DMA‐1 software or DMA‐1 PC software, metal
2 101.23
boats.
3 101.42
Analytical balance, spatula, pipette, or 4 103.35
appropriate mechanical pipette and volumetric flask 5 103.87
(Class A), 50 or 100 ml.
Certified value of Hg : 100 ± 5 μg/kg
Sample weight :
Avg: 102.57 µg/kg SD: 1.43 µg/kg RSD: 1.39 %
100 mg
Conclusion
The DMA‐80 Mercury Analyzer successfully processed
aqueous standard sample. Total analysis time per
sample was less than 6 minutes, including the time
employed to weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
BCR 61 can be performed using Direct Mercury and weigh the sample.
Analyzer. Such an instrument requires no sample 2. Introduce the boat into sample lift.
wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐1640 terminal with
1 217.44
DMA‐1 software or DMA‐1 PC software, metal
2 219.62
boats.
3 223.43
Analytical balance, spatula, pipette, or 4 224.31
appropriate mechanical pipette and volumetric flask 5 220.58
(Class A), 50 or 100 ml.
Certified value of Hg : 210 – 250 µg/kg
Sample weight :
Avg: 221.08 µg/kg SD: 2.51 µg/kg RSD: 1.14 %
30‐100 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
BCR 61 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
BCR 141 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:00:10 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 52.84
with DMA‐1 software or DMA‐1 PC software, metal
2 53.91
boats.
3 54.02
Analytical balance, spatula, pipette, or 4 52.30
appropriate mechanical pipette and volumetric flask 5 54.14
(Class A), 50 or 100 ml.
Certified value of Hg : 52.5 – 61.1 µg/kg
Sample weight :
Avg: 53.44 µg/kg SD: 0.74 µg/kg RSD: 1.38 %
40‐100 mg
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
BCR 141 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
BCR 145R can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:00:10 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° mg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 2.02
with DMA‐1 software or DMA‐1 PC software, metal
2 2.10
boats.
3 2.07
Analytical balance, spatula, pipette, or 4 2.08
appropriate mechanical pipette and volumetric flask 5 2.09
(Class A), 50 or 100 ml.
Certified value of Hg : 17.9 – 2.23 mg/kg
Sample weight :
Avg: 2.07 µg/kg SD: 0.028 µg/kg RSD: 1.38 %
80‐130 mg
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
BCR 145R sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
BCR 150 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 9.39
with DMA‐1 software or DMA‐1 PC software, metal
2 8.89
boats.
3 9.37
Analytical balance, spatula, pipette, or 4 9.41
appropriate mechanical pipette and volumetric flask 5 9.14
(Class A), 50 or 100 ml.
Certified value of Hg : 7.7 – 11.1 µg/kg
Sample weight :
Avg: 9.24 µg/kg SD: 0.21 µg/kg RSD: 2.16 %
80‐120 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
BCR 150 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
BCR 184 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 3.25
with DMA‐1 software or DMA‐1 PC software, metal
2 3.02
boats.
3 3.10
Analytical balance, spatula, pipette, or 4 3.05
appropriate mechanical pipette and volumetric flask 5 3.03
(Class A), 50 or 100 ml.
Certified value of Hg : 2.0 – 3.2 µg/kg
Sample weight :
Avg: 3.09 µg/kg SD: 0.08 µg/kg RSD: 2.74 %
30‐100 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
BCR 184 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
BCR 191 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 25.31
with DMA‐1 software or DMA‐1 PC software, metal
2 24.26
boats.
3 24.81
Analytical balance, spatula, pipette, or 4 24.87
appropriate mechanical pipette and volumetric flask 5 22.89
(Class A), 50 or 100 ml.
Certified value of Hg : NC
Sample weight :
Avg: 24.43 µg/kg SD: 0.84 µg/kg RSD: 3.42 %
40‐100 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
BCR 191 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
BCR 279 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 58.77
with DMA‐1 software or DMA‐1 PC software, metal
2 58.46
boats.
3 57.94
Analytical balance, spatula, pipette, or 4 57.58
appropriate mechanical pipette and volumetric flask 5 57.16
(Class A), 50 or 100 ml.
Certified value of Hg : (50 µg/kg)
Sample weight :
Avg: 57.99 µg/kg SD: 0.58 µg/kg RSD: 1.00 %
40‐100 mg
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
BCR 279 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
BCR 397 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° mg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 12.49
with DMA‐1 software or DMA‐1 PC software, metal
2 12.89
boats.
3 12.69
Analytical balance, spatula, pipette, or 4 12.78
appropriate mechanical pipette and volumetric flask 5 12.26
(Class A), 50 or 100 ml.
Certified value of Hg : 11.8 – 12.8 mg/kg
Sample weight :
Avg: 12.61 mg/kg SD: 0.20 mg/kg RSD: 1.64 %
10‐30 mg
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
BCR 397 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
BCR 414 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 275.59
with DMA‐1 software or DMA‐1 PC software, metal
2 274.78
boats.
3 280.25
Analytical balance, spatula, pipette, or 4 275.25
appropriate mechanical pipette and volumetric flask 5 281.88
(Class A), 50 or 100 ml.
Certified value of Hg : 258 – 294 µg/kg
Sample weight :
Avg: 277.15 µg/kg SD: 2.93 µg/kg RSD: 1.05 %
50‐100 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
BCR 414 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
BCR 422 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐1640 terminal with DMA‐1
1 561.11
software or DMA‐1 PC software, metal boats.
2 553.23
Analytical balance, spatula, pipette, or 3 569.15
appropriate mechanical pipette and volumetric flask 4 546.81
(Class A), 50 or 100 ml. 5 563.33
Sample weight :
Certified value of Hg : 543 – 575 µg/kg
50‐100 mg (max) Avg: 558.73 µg/kg SD: 7.85 µg/kg RSD: 1.40 %
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
BCR 422 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
BCR 680 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° mg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 25.57
with DMA‐1 software or DMA‐1 PC software, metal
2 26.07
boats.
3 25.74
Analytical balance, spatula, pipette, or 4 26.61
appropriate mechanical pipette and volumetric flask 5 25.05
(Class A), 50 or 100 ml.
Certified value of Hg : 24.3‐26.3 mg/kg
Sample weight :
Avg: 25.81 mg/kg SD: 0.52 mg/kg RSD: 2.01 %
10‐15 mg
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
BCR 680 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
BCR 681 can be performed using DMA‐80 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° mg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 4.32
with DMA‐1 software or DMA‐1 PC software, metal
2 4.34
boats.
3 4.57
Analytical balance, spatula, pipette, or 4 4.35
appropriate mechanical pipette and volumetric flask 5 4.18
(Class A), 50 or 100 ml.
Certified value of Hg : 4.350 – 4.650 mg/kg
Sample weight :
Avg: 4.39 mg/kg SD: 0.12 mg/kg RSD: 2.73 %
30‐40 mg
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
BCR 681 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
IAEA‐086 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 585.69
with DMA‐1 software or DMA‐1 PC software, metal
2 573.02
boats.
3 580.69
Analytical balance, spatula, pipette, or 4 580.57
appropriate mechanical pipette and volumetric flask 5 550.68
(Class A), 50 or 100 ml.
Certified value of Hg : 534 – 612 µg/kg
Sample weight :
Avg: 574.13 µg/kg SD: 12.40 µg/kg RSD: 2.16 %
30‐50 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
IAEA‐086 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
JSAC 0601‐01 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° mg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1‐2 1.19‐1.17
with DMA‐1 software or DMA‐1 PC software, metal
3‐4 1.15‐1.18
boats.
5‐6 1.15‐1.17
Analytical balance, spatula, pipette, or 7‐8 1.18‐1.17
appropriate mechanical pipette and volumetric flask 9‐10 1.17‐1.15
(Class A), 50 or 100 ml.
Certified value of Hg : 1.0 – 1.2 mg/kg
Sample weight :
Avg: 1.17 mg/kg SD: 0.014 mg/kg RSD: 1.17 %
30‐40 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
JSAC 0601‐01 sample. Total analysis time per sample
was less than 6 minutes, including the time employed
to weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
LGC 7162 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 30.77
with DMA‐1 software or DMA‐1 PC software, metal
2 31.77
boats.
3 30.04
Analytical balance, spatula, pipette, or 4 30.73
appropriate mechanical pipette and volumetric flask 5 31.62
(Class A), 50 or 100 ml.
Certified value of Hg : 27 ± 6 μg/kg
Sample weight :
Avg: 30.87 µg/kg SD: 0.52 µg/kg RSD: 1.70 %
80‐100 mg
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
LCG 7162 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
CRM‐31 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:0:10 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 5.75
with DMA‐1 software or DMA‐1 PC software, metal
2 5.97
boats.
3 5.46
Analytical balance, spatula, pipette, or 4 5.64
appropriate mechanical pipette and volumetric flask 5 5.39
(Class A), 50 or 100 ml.
Working Value of Hg : Value not certificated
Sample weight :
Avg: 5.64 µg/kg SD: 0.21 µg/kg RSD: 3.70 %
30‐100 mg
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
CRM‐31 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
CRM‐39 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:0:10 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 39.70
with DMA‐1 software or DMA‐1 PC software, metal
2 37.65
boats.
3 36.15
Analytical balance, spatula, pipette, or 4 37.26
appropriate mechanical pipette and volumetric flask 5 37.53
(Class A), 50 or 100 ml.
Working Value of Hg : Value not certificated
Sample weight :
Avg: 37.66 µg/kg SD: 1.15 µg/kg RSD: 3.06 %
30‐100 mg
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
CRM‐39 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Nist 1515 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 45.88
with DMA‐1 software or DMA‐1 PC software, metal
2 45.97
boats.
3 43.81
Analytical balance, spatula, pipette, or 4 43.78
appropriate mechanical pipette and volumetric flask 5 43.96
(Class A), 50 or 100 ml.
Certified value of Hg : 44 ± 4 μg/kg
Sample weight :
Avg: 44.68 µg/kg SD: 1.01 µg/kg RSD: 2.28 %
60‐100 mg
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Nist 1515a sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
Nist 1575a can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 40.22
with DMA‐1 software or DMA‐1 PC software, metal
2 40.15
boats.
3 38.33
Analytical balance, spatula, pipette, or 4 38.74
appropriate mechanical pipette and volumetric flask 5 38.69
(Class A), 50 or 100 ml.
Certified value of Hg : 39.9 ± 0.7 μg/kg
Sample weight :
Avg: 39.23 µg/kg SD: 0.79 µg/kg RSD: 2.03 %
60‐100 mg
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
Nist 1575a sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.
Summary Procedure
Precise and rapid determination of total mercury in 1. Place a boat on the balance plate, tare it
DORM‐3 can be performed using DMA‐1 Direct and weigh the sample.
Mercury Analyzer. Such an instrument requires no 2. Introduce the boat into sample lift.
sample wet chemistry or pre‐treatment. 3. Run the DMA‐1 program to completion.
Once a weighed sample portion is introduced into DMA-1 program
the instrument, analysis is completed in six
minutes. Direct analysis of mercury, using the N° step Time Temperature
integrated sequence of Thermal Decomposition, 1 00:01:00 200°C
Catalyst Conversion, Amalgamation, and Atomic 2 00:02:00 650°C
Absorption Spectrophotometer, is described in EPA 3 00:01:00 650°C
7473 and is validated for laboratory as well as field Max start temp: 200°C
analysis. Purge: 60 sec
Instrumentation
Results
Direct Mercury Analyzer apparatus and supplies
N° µg/kg
Milestone DMA‐1, 640‐12 and 1640‐12 terminal
1 359
with DMA‐1 software or DMA‐1 PC software, metal
2 356
boats.
3 357
Analytical balance, spatula, pipette, or 4 350
appropriate mechanical pipette and volumetric flask 5 351
(Class A), 50 or 100 ml.
Certified value of Hg : 382 ± 60 μg/kg
Sample weight :
Avg: 354.60 µg/kg SD: 3.91 µg/kg RSD: 1.04 %
80‐100 mg (max)
Conclusion
The DMA‐1 Mercury Analyzer successfully processed
DORM‐3 sample. Total analysis time per sample was
less than 6 minutes, including the time employed to
weigh each sample into the boat.