RS80A - v3.03.01 - v3.01.17 - Technical Note - Eng - 20211025
RS80A - v3.03.01 - v3.01.17 - Technical Note - Eng - 20211025
17)
Date 2021-02-09 Written by Jason Jung
Model RS80A Approved by Donghyun Kim
1. Summary
RS80A v3.03.01, v3.01.17 release
2. Details
1) Software Information
As-Is To-Be S/W Compatibility
v3.03.00 v3.03.01 Compatible
v3.01.16 v3.01.17 Compatible (Only China)
* Before you use this SW version, you should check if it is approved in your country.
2) Change List
① MR (Modification Request)
Booting memory optimization
Improved : intermittent touch malfunction creates printer output screen from Patient screen
Improved : intermittent comment not entered on report screen
Function added: "Ductus Venosus" waveform detect function
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
Both OS & UAI installation
v1.00.00 ~ v1.00.07
(Prestige Platform is necessary)
v3.03.01
v2.00.00 ~ v3.02.13 Both OS & UAI installation
v3.03.00 Only UAI installation
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Both OS & UAI installation
v1.00.00 ~ v3.01.15
v3.01.17 (Prestige Platform is necessary)
v3.02.10 RS80A_V3.02.10.0805_UAI.iso
v3.02.11 RS80A_V3.02.11.1217_UAI.iso
v3.02.12 RS80A_V3.02.09.0624_OS.iso
RS80A_V3.02.12_V3.01.14.3002_UAI_0x941E19D6.iso
v3.01.14
v3.02.13
RS80A_V3.02.13_V3.01.15.3001_UAI_0x854742A9.iso
v3.01.15
v3.03.00
RS80A_V3.03.00_V3.01.16.3354_UAI_0xB6D027C0.iso
v3.01.16
RS80A_V3.03.00_OS_0XBE9B7616.iso
v3.03.01
RS80A_V3.03.01_V3.01.17_UAI_0xBF53C7F1.iso
v3.01.17
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v3.03.00, v3.01.16)
Date 2020-08-21 Written by Heechul Kim
Model RS80A Approved by Donghyun Kim
RS80A_V3.03.00_OS_0XBE9B7616.is UA RS80A_V3.03.00_V3.01.16.3354_UAI_0xB6D027C0.is
OS
o I o
1. Summary
RS80A v3.03.00, v3.01.16 release
2. Details
1) Software Information
As-Is To-Be S/W Compatibility
v3.02.13 v3.03.00 Compatible
v3.01.15 v3.01.16 Compatible (Only China)
* Before you use this SW version, you should check if it is approved in your country.
2) Change List
① Add new Probes
EV2-10A Probe
④ MR (Modification Request)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Modified: Access violation Error message appeared during EzExam+
Modified: Entering to ‘Report’ was occasionally slow after measuring.
Modified: Previous ‘Review Image’ appeared for a short time before entering to ‘Sonoview’.
Modified: the issue with Apply & Start Exam when the physician name is over 193 characters has
been resolved
Modified: System had shut down after 20 secs after installing S/W with DVD.
(Normal time is 5 secs)
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
v1.00.00 ~ v3.02.13 v3.03.00
(Prestige Platform is necessary)
v3.01.16 OS + UAI
v1.00.00 ~ v3.01.15
(Only China) (Prestige Platform is necessary)
v3.02.10 RS80A_V3.02.10.0805_UAI.iso
v3.02.11 RS80A_V3.02.11.1217_UAI.iso
v3.02.12 RS80A_V3.02.09.0624_OS.iso
RS80A_V3.02.12_V3.01.14.3002_UAI_0x941E19D6.iso
v3.01.14
v3.02.13
RS80A_V3.02.13_V3.01.15.3001_UAI_0x854742A9.iso
v3.01.15
v3.03.00
RS80A_V3.03.00_OS_0XBE9B7616.iso RS80A_V3.03.00_V3.01.16.3354_UAI_0xB6D027C0.iso
v3.01.16
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v3.02.13, v3.01.15)
Date 2020-06-10 Written by Jason Jung
Model RS80A Approved by Donghyun Kim
1. Summary
RS80A v3.02.13, v3.01.15 release
2. Details
1) Software Information
As-Is To-Be S/W Compatibility
v3.02.12 v3.02.13 Compatible
v3.01.14 v3.01.15 Compatible (Only China)
* Before you use this SW version, you should check if it is approved in your country.
2) Change List
① Changed H/W
Monitor Size Change (23” → 23.8”)
- It will be applied after running out of stock. It depends on the regulation of each country
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
v1.00.00 ~ v3.02.08
v3.02.13 (Prestige Platform is necessary)
OS + UAI
v1.00.00 ~ v3.01.13
v3.01.15 (Prestige Platform is necessary)
(Only China)
v3.01.14 UAI Only
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
v3.02.10 RS80A_V3.02.10.0805_UAI.iso
v3.02.11 RS80A_V3.02.11.1217_UAI.iso
v3.02.12
RS80A_V3.02.12_V3.01.14.3002_UAI_0x941E19D6.iso
v3.01.14
v3.02.13
RS80A_V3.02.13_V3.01.15.3001_UAI_0x854742A9.iso
v3.01.15
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v3.02.12, v3.01.14)
Date 2020-05-25 Written by Heechul Kim
Model RS80A Approved by Donghyun Kim
1. Summary
RS80A v3.02.12, v3.01.14 release
2. Details
1) Software Information
As-Is To-Be S/W Compatibility
v3.02.11 v3.02.12 Compatible
v3.01.13 v3.01.14 Compatible (Only China)
* Before you use this SW version, you should check if it is approved in your country.
2) Change List
① Add S/W function and modification.
Information of S/W version is set as ‘System Region setting’.
- Global setting : v3.02.12,
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
v1.00.00 ~ v3.02.08 v3.02.12
(Prestige Platform is necessary)
v3.01.14 OS + UAI
v1.00.00 ~ v3.01.13
(Only China) (Prestige Platform is necessary)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
v3.02.11 RS80A_V3.02.11.1217_UAI.iso
v3.02.12
RS80A_V3.02.12_V3.01.14.3002_UAI_0x941E19D6.iso
v3.01.14
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v3.02.11)
Date 2020-01-09 Written by Heechul Kim
Model RS80A Approved by Donghyun Kim
1. Summary
RS80A v3.02.11 release
2. Details
1) Software Information
As-Is To-Be S/W Compatibility
v3.02.10 v3.02.11 Compatible
* Before you use this SW version, you should check if it is approved in your country.
2) Change List
② Add S/W function and modification
‘ElastoScan function’ has been added with CA1-7A in Abdomen-General
- It’s only supported in Global Setting(It’s not supported in Region1)
[ Defective Symptom ]
Improved: The system was locked up after the message appeared(“Eternal exception E06D7363”)
When User changed the probe
3) Installation Information
SW Version in USE Target Version For SW Upgrade
v3.02.11
v1.00.00 ~ v3.02.08 OS + UAI
(Prestige Platform is necessary)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v3.02.10)
Date 14-Aug-2019 Written by Jason Jung
Model RS80A Approved by Donghyun Kim
1. Summary
RS80A v3.02.10 release
2. Details
1) SW Information
As-Is To-Be S/W Compatibility
v3.02.09 v3.02.10 Compatible
* Before you use this SW version, you should check if it is approved in your country.
2) Details
④ Improved Image performance and function
Improved the accuracy for AutoCalc HR automatic measurement
Improved L3-12A Vasc (SHAR mode)
⑥ MR (Modification Request)
Improved the intermittent lock up issue after running ElastoScan
3) Installation Information
SW Version in USE Target Version For SW Upgrade
v3.02.10
v1.00.00 ~ v3.02.08 OS + UAI
(Prestige Platform is necessary)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v3.02.09)
Date 17-Jul-2019 Written by Jason Jung
Model RS80A Approved by Donghyun Kim
1. Summary
RS80A v3.02.09 release
2. Details
1) SW Information
As-Is To-Be S/W Compatibility
v3.02.08 v3.02.09 Compatible
* Before you use this SW version, you should check if it is approved in your country.
2) Details
① H/W change
VGA and OS change due to VGA discontinuance
- Before : GTX 750Ti After : GTX 1050 (Same with RS85)
- Components : Total 5
VGA Card Card, Bracket HDMI to HDMI Cable Cable Tie Screw
3) Installation Information
SW Version in USE Target Version For SW Upgrade
v3.02.09
v1.00.00 ~ v3.02.04 OS + UAI
(Prestige Platform is necessary)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v3.02.08)
Date 10- May-2019 Written by Heechul Kim
Model RS80A Approved by Duke Choi
1. Summary
RS80A v3.02.08 release
2. Details
1) SW Information
As-Is To-Be S/W compatibility
v3.02.07 v3.02.08 Compatible
* Before you use this SW version, you should check if it is approved in your country.
2) Details
① Add and Modify ‘S/W function’
Improved, Taking a long time when User unlocks ‘S/W Option Unlock’
- There is no delay time when Engineer fail to unlock ‘S/W Option Unlock’ until do totally four times
② MR (Modification Request)
Fixed, Can not see the created protocol when User enters ‘EZ Exam protocol’ after upgrade
3.02.05
Improved, Collect S/W Log regarding TR Control
Fixed, The saved ‘mbd file’ in USB gets bigger(↑1GB) abnormally when User backs-up Exam
Fixed, Translation error for German about S-Detect
3) Installation Information
SW Version in USE Target Version For SW Upgrade
v3.02.08
v1.00.00 ~ v3.02.04 OS + UAI
(Prestige Platform is necessary)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v3.02.07)
Date 2019-04-03 Written by Heechul Kim
Model RS80A Approved by Duke Choi
1. Summary
RS80A v3.02.07 release
2. Details
1) S/W information
As-Is To-Be S/W compatibility
v3.02.05 v3.02.07 Compatible
* Before you use this SW version, you should check if it is approved in your country.
2) Details
① Add S/W function and modification
ECG performance is improved
It is possible to use E-Strain. Even though User set region as America
② MR (Modification Request)
The issue that Color noise happens intermittently is improved
The issue that Error code(105) happens during operating is improved
3) How to upgrade
SW Version in USE Target Version For SW Upgrade
v3.02.07
v1.00.00 ~ v3.02.04 OS + UAI
(Prestige Platform is necessary)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v3.02.05)
Date 2019-01-08 Written by Sean Park
Model RS80A Approved by Duke Choi
1. SW Information
RS80A v3.02.05 release
2. Details
1) SW Information
AS-IS TO-BE S/W compatibility
v3.02.04 v3.02.05 Compatible
* Before you use this SW version, you should check if it is approved in your country.
2) Details
① New Features
BitLocker support
- New PC Main board (MI59-01118G) is required due to PC BIOS Update (R1.00.E14 -> R1.00.E15).
- TPM(Trusted Platform Module: MI59-01259A) board is also required for this function.
② Improvements
Eable to configure Measure Result for each Application
Mitshubishi P95D driver update
③ MR (Modification Request)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
When: Setup > DICOM > DICOM Compresstion: Uncompressed > S-Detect > Image Save > DICOM transfer
Fixed: Can’t apply or start EXAM when Patient ID is longer than 65 letters
Fixed: Previous image saved in Thumbnail
When: Press P1 (Save) button twice very quickly during scanning > Annotation > Press P1 (Save) button
3) Installation Information
SW Version in USE Target Version For SW Upgrade
v3.02.05
v1.00.00 ~ v3.01.xx OS + UAI
(v1.xx users should buy Prestige option.)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v3.02.04, v3.01.11, v3.00.07)
Date 2018-10-22 Written by Sean Park
Model RS80A Approved by Duke Choi
1. SW Information
RS80A v3.02.04, v3.01.11, v3.00.07 release
2. Details
1) SW Information
AS-IS TO-BE S/W compatibility
v3.02.03 v3.02.04 Compatible
v3.01.10 v3.01.11 Compatible
v3.00.06 v3.00.07 Compatible
* Before you use this SW version, you should check if it is approved in your country.
2) Details
① MR (Modification Request)
Fixed: No biopsy needle displayed at 0 degree with V5-9 and EV3-10B
3) Installation Information
기존 SW version 대상 SW version SW Upgrade 방법
v3.00.07
v1.00.00 ~ v2.xx.xx OS + UAI
(v1.xx users should buy Prestige option.)
v3.01.11
v1.00.00 ~ v3.01.06 OS + UAI
(v1.xx users should buy Prestige option.)
v3.02.04
v1.00.00 ~ v3.01.xx OS + UAI
(v1.xx users should buy Prestige option.)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v3.02.03)
Date 2018-08-10 Written by Sean Park
Model RS80A Approved by Duke Choi
1. SW Information
RS80A v3.02.03 release
2. Details
1) SW Information
AS-IS TO-BE S/W compatibility
v3.02.02 v3.02.03 Compatible
* Before you use this SW version, you should check if it is approved in your country.
2) Details
① MR (Modification Request)
Improved: Method of QR code generation for OTP (One Time Password)
Old) QR code = System Serial Number + “:” + Random 4-digit number + 1 alphabet letter
New) QR code = Software Serial Number + “:” + Random 4-digit number + 1 alphabet letter
3) Installation Information
SW Version in USE Target Version For SW Upgrade
v3.02.03
v1.00.00 ~ v3.02.00 OS + UAI
(Prestige option is required)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v3.02.02)
Date 2018-06-18 Written by Jason Jung
Model RS80A Approved by Duke Choi
1. SW Information
RS80A v3.02.02 release
2. Details
1) SW Information
AS-IS TO-BE S/W compatibility
v3.02.00 v3.02.02 Compatible
* Before you use this SW, you should check if it is approved in your country.
2) Details
① MR (Modification Request) modification
UAI installation method Temporary change.
- Caution:
OS file is not added or changed, but if the version is raised to 3.02.02, the OS file must be reinstalled.
If the version is raised using UAI instead of reinstalling the OS file in 3.02.00 version, the DB and
UAI V3.02.00 S/W Improvement of defect which does not booting after installing.
Modify the Pulser control sequence to improve image line loss.
Modify parts of the self-prognostic that may cause errors during CW test
Fixed a problem that system is slowed when CW probe change.
- Occurrence condition: 2D → Dual → Dual → Dual → CW Probe Preset Load
3) Installation Information
SW Version in USE Target Version For SW Upgrade
v3.02.02
v1.00.00 ~ v3.02.00 OS + UAI Install files
(Prestige option is needed)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v3.02.00)
Date 2018-05-29 Written by Kelvin Cho
Model RS80A Approved by Duke Choi
1. Summary
RS80A v3.02.00 release
2. Details
1) SW Information
AS-IS TO-BE S/W compatibility
v3.01.09 v3.02.10 Compatible
* Before you use this SW, you should check if it is approved in your country.
2) Details
① Addition or change of S/W function
Cyber Security (Account Management)
- Separate and limited permissions according to user accounts (separate manual will be provided)
Added the function to automatically delete the exam at a given time for storage capacity
- Setup → System → General Tab > Auto Delete “ON”, “run date” and “delete interval setting” available
- On the date set, may delete the data or postpone the date
- Server: DICOM Setup -> check "Use TLS" -> check "Verify Client Certificate" (option)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
UAI manager update
- Compare (verify the compatibility) the installed OS and UAI version and determine whether to install
For bradycardia patients, it is difficult to measure AutoCalc. because the PW waveform period is long.
3) Installation Information
SW Version in USE Target Version For SW Upgrade
v3.02.00
v1.00.00 ~ v3.01.09 OS + UAI Install files
(Prestige option is needed)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v3.01.09)
Date 2018-03-21 Written by Kelvin Cho
Model RS80A Approved by Duke Choi
1. Summary
RS80A v3.01.09 release
2. Details
1) SW Information
AS-IS TO-BE S/W compatibility
v3.01.04 v3.01.09 Compatible
* Please check the SW version for your country before you use.
2) Details
① Addition or change of S / W function
Anti-Virus program (Avast Pro, separate purchase is needed) support
- Enable button added in “System Setup” (search for “Avast Pro installation Manual” in BTSP system)
Continue Exam without end exam → Patient OK → Start with patient B, not patient A
Modify DICOM Q / R so that it can be seen in Study List without sending Modality information
at Server when searching
Improved the problem that Indicator Cursor is saved together when Image Save is done in
annotation state
3) Installation Information
SW Version in USE Target Version For SW Upgrade
v3.01.09
v1.00.00 ~ v3.01.04 OS + UAI Install files
(Prestige option is needed)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v2.00.08)
Date 2017-12-22 Written by Dean Kim
Model RS80A Approved by Duke Choi
1. Summary
RS80A v2.00.08 release
2. Details
1) SW Information
AS-IS TO-BE S/W compatibility
v2.00.07 v2.00.08 Compatible
* Please check the SW version for your country before you use.
2) Change List
① The function of 'S-Detect' was improved.
Able to select 'BI-RADS 2003' or 'BI-RADS 2013' from Classification
Axillary and Probe marker added to Marker
Able to adjust a probe angle
When Classification setting is restored after a revision, 'Revised' mark will be removed.
If malignant, Sensitivity and Specificity can be adjusted.
② MR (Modification Request)
Intermittent ‘touch screen not working’ problem is solved.
Intermittent ‘Cursor pop-up’ problem is solved.
When ‘Print’ and ‘Save’ are assigned to User defined key, the system will internally do ‘Save’
and ‘Print’, so that the current image will be printed.
Mis-printing issue when pushing ‘Exit’ from Report during printing is solved.
3) Installation Information
SW Version in USE Target Version For SW Upgrade
v1.00.00 ~ v1.00.06 v1.00.08 OS + UAI
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.