TFS Assets - CMD - Technical Notes - TN 708 Cds Automate System Suitability Testing tn70092 en
TFS Assets - CMD - Technical Notes - TN 708 Cds Automate System Suitability Testing tn70092 en
Key Words
Chromeleon Chromatography Data System, (CDS), System Suitability
Testing, Chromatographic Suitability, SST
Goal
To illustrate and define the tools used by the Chromeleon Chromatography
Data System (CDS) to perform Automated System Suitability Testing
Introduction
System Suitability Testing (SST) is commonly used by
laboratories to ensure that the complete analytical system
(including instrument, reagents, columns and analysts) is
suitable for the intended application.
The general goal of system suitability testing is to monitor
The United States Pharmacopeia (USP) Chromatography chromatographic results to ensure chromatographic
General Chapter states: suitability (e.g. by testing tailing factor, column efficiency
and resolution of critical peak pairs) and consistent
“System Suitability Tests are an integral part of gas and
system performance (e.g. by using replicate injections
liquid chromatographic methods. They are used to verify
of test standards).
that the resolution and reproducibility of the chromato-
graphic system are adequate for the analysis to be done. SST calculations are usually performed on standards
The tests are based on the concept that the equipment, before any samples are analyzed to ensure that potentially
electronics, analytical operations and samples to be valuable samples are not injected into an unsuitable
analyzed constitute an integral system that can be system. In a manual environment, this can require user
evaluated as such.” interaction with every batch of samples and result in
considerable throughput delays. Furthermore, if calcula-
tions are done manually, there is the potential for error
and incorrect pass or fail results.
Figure 1. A Chromeleon processing method is shown above, configured with FDA recommended suitability tests
method, the “Fail-Action” column 3
directs that if the test fails to meet
this criterion, then sequence
acquisition will be halted.
Reporting Results
Chromeleon comes with a compre-
hensive set of default reports,
which can be used as they are, or
can be easily customized to meet
the specialized needs of the lab.
Predefined report tables that include
system suitability related information Figure 2. An example of an SST report is shown above.
can be inserted in any report. Figure 2
shows a customized system suitability
injection report that includes the SST
results, along with an overall pass/fail
determination.
Te ch ni cal N ote 1 0 0 6
allows the user to define what test
result should be reported if the test
cannot be evaluated and, during
sequence acquisition, what action
Chromeleon should take if the
test fails.
Completed Wizard
Upon completion of the Wizard, the
new test is added to the SST table of
the method. Double clicking on any
item in the test row will reopen the
appropriate wizard page (Figure 12).
Figure 11. Conditional actions for the test results may be specified.
Conclusion • Reduced sample loss. Chromeleon • The functionality can also be used
Chromeleon 7 offers many significant can automatically stop sequences to generate control charts to
advantages for System Suitability as soon as it detects a System monitor system performance in the
Testing: Suitability failure. This prevents laboratory and quickly identify
samples from being injected on to a systems that continually produce
• Improved reliability of System
system that is not suitable for the high levels of failure.
Suitability Tests results. Chrome-
analytical method and eliminates • By automating some of the most
leon performs all calculations,
sample loss. repetitive and time consuming
eliminating any errors that can be
caused by manual calculation steps. • The automated SST functionality steps found in laboratories, the
can be extended to perform other Automated System Suitability Test
• Faster generation of System
types of tests, such as comparing functionality saves time and money
Suitability Test results. As soon as
sample results against and increases laboratory
the peak is detected, Chromeleon
specifications. productivity.
will automatically recalculate the
system suitability results. This References
increases throughput by reducing 1. Center for Drug Evaluation and
delays created by having to Research (CDER) Reviewer
manually recalculate the results. Guidance Validation of
Chromatographic Methods,
November 1994.
thermoscientific.com
©2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its
subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult
your local sales representative for details.
U.S./Canada (847) 295 7500 Brazil (55) 11 3731 5140 India (91) 22 2764 2735 Singapore (65) 6289 1190
Australia (61) 2 9420 5233 China (852) 2428 3282 Ireland (353) 1 644 0064 Sweden (46) 8 473 3380
Austria (43) 1 616 51 25 Denmark (45) 36 36 90 90 Italy (39) 02 51 62 1267 Switzerland (41) 62 205 9966
Benelux (31) 20 683 9768 France (33) 1 39 30 01 10 Japan (81) 6 6885 1213 Taiwan (886) 2 8751 6655
(32) 3 353 42 94 Germany (49) 6126 991 0 Korea (82) 2 2653 2580 United Kingdom (44) 1276 691722
TN70092_E 07/16S