The document reports the results of testing a disinfectant's efficacy against various microorganisms on different surfaces. Tests were conducted on the disinfectant's ability to achieve 3-6 log reductions of bacteria, fungi and environmental isolates within 5 minutes on filtration, walls, ceilings and doors.

Uploaded by

Solomon
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (1 vote)
350 views27 pages

SOP For Disinfectants Efficacy Test

The document reports the results of testing a disinfectant's efficacy against various microorganisms on different surfaces. Tests were conducted on the disinfectant's ability to achieve 3-6 log reductions of bacteria, fungi and environmental isolates within 5 minutes on filtration, walls, ceilings and doors.

Uploaded by

Solomon
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 27

Humanwell Pharmaceutical Ethiopia PLC

Form No.: SOP/SL/003/F/01


Disinfectants Validation Report Form

Attachment (A)

DISINFECTANT VALIDATION RESULTS

Disinfectant: ______________________ _________________Lot No.:

Supplier:

Analyzed on:(provide date) Analyzed by: (provide name)

Tested for: Disinfectant Efficacy

The test has been carried out at the predetermined concentration of


manufacturer (_____%)

Filtration Technique

Acceptance
Organism Criteria Result

Pseudomonas aeruginosa ATTCC Six-log reduction


9027 within 5 minutes

Escherichia coli ATCC 8739 Six-log reduction


within 5 minutes

Bacillus subtilis ATCC 6633 Six-log reduction


within 5 minutes

Staphylococcus aureus ATCC 6538 Six-log reduction


within 5 minutes

Clostridium sporogenes ATCC 11437 Six-log reduction


within 5 minutes

Candida albicans ATCC 10231 Six-log reduction


within 5 minutes
Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Aspergillus niger ATCC 16404 Six-log reduction


within 5 minutes

Environmental isolates from wall Six-log reduction


(micrococcus lutes) within 5 minutes

Environmental isolates from operator’s Six-log reduction


fingerprint (Bacillus spp.) within 5 minutes

Environmental isolates from floor Six-log reduction


(Aspergillus) within 5 minutes

Environmental isolates from glass Six-log reduction


window (Staphylococcus epidermis) within 5 minutes

Analyzed by: Microbiologist Checked by: QC Manager

Signature/date: _____________________ Signature/date: ___________________


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Attachment (B)

DISINFECTANT VALIDATION RESULTS

Disinfectant: ______________________ _________________Lot No.:

Supplier: _________________________________________________

Analyzed on:(provide date) Analyzed by: (provide name)

Tested for: Disinfectant Efficacy

The test has been carried out at the predetermined concentration of


manufacturer (_____%)

Surface Technique
Efficacy Test on Wall (Sandwich Panel) (Coupons)

Acceptance
Organism Criteria Result

Pseudomonas aeruginosa Three-log


ATTCC 9027 reduction within
5 minutes

Escherichia coli ATCC 8739 Three-log


reduction within
5 minutes

Bacillus subtilis ATCC 6633 Three-log


reduction within
5 minutes
Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Staphylococcus aureus ATCC Three-log


6538 reduction
within 5 minutes

Clostridium sporogenes ATCC Three-log


11437 reduction within
5 minutes

Candida albicans ATCC 10231 Three-log


reduction within
5 minutes

Aspergillus niger ATCC 16404 Three-log


reduction within
5 minutes

Environmental isolates from wall Three-log


(micrococcus lutes) reduction within
5 minutes

Environmental isolates from Three-log


oper- reduction within
ator’s fingerprint (Bacillus spp.) 5 minutes

Environmental isolates from floor Three-log


(Aspergillus) reduction within
5 minutes

Environmental isolates from glass Three-log


window (Staphylococcus reduction within
epidermis) 5 minutes

Analyzed by: Microbiologist Checked by: QC Manager


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Signature/date: _____________________ Signature/date: ___________________


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Attachment (C)

DISINFECTANT VALIDATION RESULTS

Disinfectant: ______________________ _________________Lot No.:

Supplier: _________________________________________________

Analyzed on:(provide date) Analyzed by: (provide name)

Tested for: Disinfectant Efficacy

The test has been carried out at the predetermined concentration of


manufacturer (_____%)

Efficacy Test on Ceiling (Cleanroom Design) (Coupons)


Organism Acceptance Criteria Result

Pseudomonas aeruginosa Three-log reduction


ATTCC 9027 within 5 minutes

Escherichia coli ATCC 8739 Three-log reduction


within 5 minutes

Bacillus subtilis ATCC 6633 Three-log reduction


within 5 minutes

Staphylococcus aureus ATCC Three-log reduction


6538 within 5 minutes

Clostridium sporogenes Three-log reduction


ATCC 11437 within 5 minutes

Candida albicans ATCC Three-log reduction


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

10231 within 5 minutes

Aspergillus niger ATCC Three-log reduction


16404 within 5 minutes

Environmental isolates from Three-log reduction


wall (micrococcus lutes) within 5 minutes

Environmental isolates from Three-log reduction


oper-ator’s fingerprint within 5 minutes
(Bacillus spp.)

Environmental isolates from Three-log reduction


floor (Aspergillus) within 5 minutes

Environmental isolates from Three-log reduction


glass window within 5 minutes
(Staphylococcus epidermis)

Analyzed by: Microbiologist Checked by: QC Manager

Signature/date: _____________________ Signature/date: ___________________


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Attachment (D)

DISINFECTANT VALIDATION RESULTS

Disinfectant: ______________________ _________________Lot No.:

Supplier: _________________________________________________

Analyzed on:(provide date) Analyzed by: (provide name)

Tested for: Disinfectant Efficacy

The test has been carried out at the predetermined concentration of


manufacturer (_____%)

Efficacy Test Doors (Galvanized Steel Finished with White Powder) (Coupons)

Organism Acceptance Criteria Result

Pseudomonas aeruginosa Three-log reduction


ATTCC 9027 within 5 minutes

Escherichia coli ATCC 8739 Three-log reduction


within 5 minutes

Bacillus subtilis ATCC 6633 Three-log reduction


within 5 minutes

Staphylococcus aureus ATCC Three-log reduction


6538 within 5 minutes

Clostridium sporogenes ATCC Three-log reduction


11437 within 5 minutes

Candida albicans ATCC 10231 Three-log reduction


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

within 5 minutes

Aspergillus niger ATCC 16404 Three-log reduction


within 5 minutes

Environmental isolates from wall Three-log reduction


(micrococcus lutes) within 5 minutes

Environmental isolates from Three-log reduction


operator’s fingerprint (Bacillus within 5 minutes
spp.)

Environmental isolates from floor Three-log reduction


(Aspergillus) within 5 minutes

Environmental isolates from glass Three-log reduction


window (Staphylococcus within 5 minutes
epidermis)

Analyzed by: Microbiologist Checked by: QC Manager

Signature/date: _____________________ Signature/date: ___________________


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Attachment (E)

DISINFECTANT VALIDATION RESULTS

Disinfectant: ______________________ _________________Lot No.:

Supplier: _________________________________________________

Analyzed on:(provide date) Analyzed by: (provide name)

Tested for: Disinfectant Efficacy

The test has been carried out at the predetermined concentration of


manufacturer (_____%)

Efficacy Test on Glass Windows (Coupons)

Organism Acceptance Criteria Result

Pseudomonas aeruginosa Three-log reduction


ATTCC 9027 within 5 minutes

Escherichia coli ATCC 8739 Three-log reduction


within 5 minutes

Bacillus subtilis ATCC 6633 Three-log reduction


within 5 minutes

Staphylococcus aureus ATCC Three-log reduction


6538 within 5 minutes

Clostridium sporogenes ATCC Three-log reduction


11437 within 5 minutes

Candida albicans ATCC 10231 Three-log reduction


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

within 5 minutes

Aspergillus niger ATCC 16404 Three-log reduction


within 5 minutes

Environmental isolates from wall Three-log reduction


(micrococcus lutes) within 5 minutes

Environmental isolates from Three-log reduction


operator’s fingerprint (Bacillus within 5 minutes
spp.)

Environmental isolates from floor Three-log reduction


(Aspergillus) within 5 minutes

Environmental isolates from glass Three-log reduction


window (Staphylococcus within 5 minutes
epidermis)

Analyzed by: Microbiologist Checked by: QC Manager

Signature/date: _____________________ Signature/date: ___________________


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Attachment (F)

DISINFECTANT VALIDATION RESULTS

Disinfectant: ______________________ _________________Lot No.:

Supplier: _________________________________________________

Analyzed on:(provide date) Analyzed by: (provide name)

Tested for: Disinfectant Efficacy

The test has been carried out at the predetermined concentration of


manufacturer (_____%)

Efficacy Test on Concrete Walls Finished with Antifungal Oil Paint (Coupons)

Organism Acceptance Criteria Result

Pseudomonas aeruginosa Three-log reduction


ATTCC 9027 within 5 minutes

Escherichia coli ATCC 8739 Three-log reduction


within 5 minutes

Bacillus subtilis ATCC 6633 Three-log reduction


within 5 minutes

Staphylococcus aureus ATCC Three-log reduction


6538 within 5 minutes

Clostridium sporogenes ATCC Three-log reduction


11437 within 5 minutes

Candida albicans ATCC 10231 Three-log reduction


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

within 5 minutes

Aspergillus niger ATCC 16404 Three-log reduction


within 5 minutes

Environmental isolates from wall Three-log reduction


(micrococcus lutes) within 5 minutes

Environmental isolates from Three-log reduction


operator’s fingerprint (Bacillus within 5 minutes
spp.)

Environmental isolates from floor Three-log reduction


(Aspergillus) within 5 minutes

Environmental isolates from glass Three-log reduction


within 5 minutes
window (Staphylococcus
epidermis)

Analyzed by: Microbiologist Checked by: QC Manager

Signature/date: _____________________ Signature/date: ___________________


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Attachment (G)

DISINFECTANT VALIDATION RESULTS

Disinfectant: ______________________ _________________Lot No.:

Supplier: _________________________________________________

Analyzed on:(provide date) Analyzed by: (provide name)

Tested for: Disinfectant Efficacy

The test has been carried out at the predetermined concentration of


manufacturer (_____%)

Efficacy Test on Floor (Epoxy) (Coupons)

Organism Acceptance Criteria Result

Pseudomonas aeruginosa Three-log reduction within 5


ATTCC 9027 minutes

Escherichia coli ATCC 8739 Three-log reduction within 5


minutes

Bacillus subtilis ATCC 6633 Three-log reduction within 5


minutes

Staphylococcus aureus ATCC Three-log reduction within 5


6538 minutes

Clostridium sporogenes ATCC Three-log reduction within 5


11437 minutes

Candida albicans ATCC 10231 Three-log reduction within 5


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

minutes

Aspergillus niger ATCC 16404 Three-log reduction within 5


minutes

Environmental isolates from wall Three-log reduction within 5


(micrococcus lutes) minutes

Environmental isolates from Three-log reduction within 5


operator’s fingerprint (Bacillus minutes
spp.)

Environmental isolates from floor Three-log reduction within 5


(Aspergillus) minutes

Environmental isolates from glass Three-log reduction


within 5 minutes
window (Staphylococcus
epidermis)

Analyzed by: Microbiologist Checked by: QC Manager

Signature/date: _____________________ Signature/date: ___________________


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Attachment (H)

DISINFECTANT VALIDATION RESULTS

Disinfectant: ______________________ _________________Lot No.:

Supplier: _________________________________________________

Analyzed on:(provide date) Analyzed by: (provide name)

Tested for: Disinfectant Efficacy

The test has been carried out at the predetermined concentration of


manufacturer (_____%)

Efficacy Test on Machines (Product Contact Stainless Steel 316) (Coupons)

Organism Acceptance Criteria Result

Pseudomonas aeruginosa Three-log reduction within


ATTCC 9027 5 minutes

Escherichia coli ATCC 8739 Three-log reduction within


5 minutes

Bacillus subtilis ATCC 6633 Three-log reduction within


5 minutes

Staphylococcus aureus ATCC Three-log reduction within


6538 5 minutes

Clostridium sporogenes ATCC Three-log reduction within


11437 5 minutes

Candida albicans ATCC 10231 Three-log reduction within


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

5 minutes

Aspergillus niger ATCC 16404 Three-log reduction within


5 minutes

Environmental isolates from wall Three-log reduction within


(micrococcus lutes) 5 minutes

Environmental isolates from Three-log reduction within


operator’s fingerprint (Bacillus 5 minutes
spp.)

Environmental isolates from floor Three-log reduction within


(Aspergillus) 5 minutes

Environmental isolates from glass Three-log reduction


within 5 minutes
window (Staphylococcus
epidermis)

Analyzed by: Microbiologist Checked by: QC Manager

Signature/date: _____________________ Signature/date: ___________________


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Attachment (I)

DISINFECTANT VALIDATION RESULTS

Disinfectant: ______________________ _________________Lot No.:

Supplier: _________________________________________________

Analyzed on:(provide date) Analyzed by: (provide name)

Tested for: Disinfectant Efficacy

The test has been carried out at the predetermined concentration of


manufacturer (_____%)

Efficacy Test on Manufacturing Vessels (Non-Product-Contact Stainless Steel 304)


(Coupons)
Organism Acceptance Criteria Result

Pseudomonas aeruginosa Three-log reduction within 5


ATTCC 9027 minutes

Escherichia coli ATCC 8739 Three-log reduction within 5


minutes

Bacillus subtilis ATCC 6633 Three-log reduction within 5


minutes

Staphylococcus aureus ATCC Three-log reduction within 5


6538 minutes

Clostridium sporogenes ATCC Three-log reduction within 5


11437 minutes
Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Candida albicans ATCC 10231 Three-log reduction within 5


minutes

Aspergillus niger ATCC 16404 Three-log reduction within 5


minutes

Environmental isolates from wall Three-log reduction within 5


(micrococcus lutes) minutes

Environmental isolates from Three-log reduction within 5


operator’s fingerprint (Bacillus minutes
spp.)

Environmental isolates from floor Three-log reduction within 5


(Aspergillus) minutes

Environmental isolates from glass Three-log reduction


within 5 minutes
window (Staphylococcus
epidermis)

Analyzed by: Microbiologist Checked by: QC Manager

Signature/date: _____________________ Signature/date: ___________________


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Attachment No. 1700.140(J)

DISINFECTANT VALIDATION RESULTS

Disinfectant: ______________________ _________________Lot No.:

Supplier: _________________________________________________

Analyzed on:(provide date) Analyzed by: (provide name)

Tested for: Disinfectant Efficacy

The test has been carried out at the predetermined concentration of


manufacturer (_____%)

Efficacy Test on Washing Basin (Stainless Steel 304) (Coupons)

Organism Acceptance Criteria Result

Pseudomonas aeruginosa Three-log reduction within 5


ATTCC 9027 minutes

Escherichia coli ATCC 8739 Three-log reduction within 5


minutes

Bacillus subtilis ATCC 6633 Three-log reduction within 5


minutes

Staphylococcus aureus ATCC Three-log reduction within 5


6538 minutes

Clostridium sporogenes ATCC Three-log reduction within 5


11437 minutes

Candida albicans ATCC 10231 Three-log reduction within 5


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

minutes

Aspergillus niger ATCC 16404 Three-log reduction within 5


minutes

Environmental isolates from wall Three-log reduction within 5


(micrococcus lutes) minutes

Environmental isolates from Three-log reduction within 5


operator’s fingerprint (Bacillus minutes
spp.)

Environmental isolates from floor Three-log reduction within 5


(Aspergillus) minutes

Environmental isolates from glass Three-log reduction


within 5 minutes
window (Staphylococcus
epidermis)

Analyzed by: Microbiologist Checked by: QC Manager

Signature/date: _____________________ Signature/date: ___________________


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Attachment No. 1700.140(K)

DISINFECTANT VALIDATION RESULTS

Disinfectant: ______________________ _________________Lot No.:

Supplier: _________________________________________________

Analyzed on:(provide date) Analyzed by: (provide name)

Tested for: Disinfectant Efficacy

The test has been carried out at the predetermined concentration of


manufacturer (_____%)

Efficacy Test on Bench (Stainless Steel 304) (Coupons)

Organism Acceptance Criteria Result

Pseudomonas aeruginosa Three-log reduction within 5


ATTCC 9027 minutes

Escherichia coli ATCC 8739 Three-log reduction within 5


minutes

Bacillus subtilis ATCC 6633 Three-log reduction within 5


minutes

Staphylococcus aureus ATCC Three-log reduction within 5


6538 minutes

Clostridium sporogenes ATCC Three-log reduction within 5


11437 minutes

Candida albicans ATCC 10231 Three-log reduction within 5


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

minutes

Aspergillus niger ATCC 16404 Three-log reduction within 5


minutes

Environmental isolates from wall Three-log reduction within 5


(micrococcus lutes) minutes

Environmental isolates from Three-log reduction within 5


operator’s fingerprint (Bacillus minutes
spp.)

Environmental isolates from floor Three-log reduction within 5


(Aspergillus) minutes

Environmental isolates from glass Three-log reduction


within 5 minutes
window (Staphylococcus
epidermis)

Analyzed by: Microbiologist Checked by: QC Manager

Signature/date: _____________________ Signature/date: ___________________


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Attachment (L)

DISINFECTANT VALIDATION RESULTS

Disinfectant: ______________________ _________________Lot No.:

Supplier: _________________________________________________

Analyzed on:(provide date) Analyzed by: (provide name)

Tested for: Disinfectant Efficacy

The test has been carried out at the predetermined concentration of


manufacturer (_____%)

Efficacy Test on Tables (Stainless Steel 304) (Coupons)

Organism Acceptance Criteria Result

Pseudomonas aeruginosa Three-log reduction


ATTCC 9027 within 5 minutes

Escherichia coli ATCC 8739 Three-log reduction


within 5 minutes

Bacillus subtilis ATCC 6633 Three-log reduction


within 5 minutes

Staphylococcus aureus ATCC Three-log reduction


6538 within 5 minutes

Clostridium sporogenes ATCC Three-log reduction


11437 within 5 minutes

Candida albicans ATCC 10231 Three-log reduction


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

within 5 minutes

Aspergillus niger ATCC 16404 Three-log reduction


within 5 minutes

Environmental isolates from wall Three-log reduction


(micrococcus lutes) within 5 minutes

Environmental isolates from Three-log reduction


operator’s fingerprint (Bacillus within 5 minutes
spp.)

Environmental isolates from floor Three-log reduction


(Aspergillus) within 5 minutes

Environmental isolates from glass Three-log reduction


within 5 minutes
window (Staphylococcus
epidermis)

Analyzed by: Microbiologist Checked by: QC Manager

Signature/date: _____________________ Signature/date: ___________________


Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Attachment (M)

DISINFECTANT VALIDATION RESULTS

Disinfectant: ______________________ _________________Lot No.:

Supplier: _________________________________________________

Analyzed on:(provide date) Analyzed by: (provide


name)

Tested for: Disinfectant Efficacy

The test has been carried out at the predetermined concentration of


manufacturer (_____%)

Efficacy Test on Doors/Lockers (Melamin) (Coupons)

Organism Acceptance Criteria Result

Pseudomonas aeruginosa Three-log reduction within


ATTCC 9027 5 minutes

Escherichia coli ATCC 8739 Three-log reduction within


5 minutes

Bacillus subtilis ATCC 6633 Three-log reduction within


5 minutes

Staphylococcus aureus ATCC Three-log reduction within


6538 5 minutes

Clostridium sporogenes ATCC Three-log reduction within


11437 5 minutes
Humanwell Pharmaceutical Ethiopia PLC
Form No.: SOP/SL/003/F/01
Disinfectants Validation Report Form

Candida albicans ATCC 10231 Three-log reduction within


5 minutes

Aspergillus niger ATCC 16404 Three-log reduction within


5 minutes

Environmental isolates from wall Three-log reduction within


(micrococcus lutes) 5 minutes

Environmental isolates from Three-log reduction within


operator’s fingerprint (Bacillus 5 minutes
spp.)

Environmental isolates from floor Three-log reduction within


(Aspergillus) 5 minutes

Environmental isolates from glass Three-log reduction


within 5 minutes
window (Staphylococcus
epidermis)

Analyzed by: Microbiologist Checked by: QC Manager

Signature/date: _____________________ Signature/date: ___________________