E1782-Standard Test Method For Determining Vapor Pressure by Thermal Analysis
E1782-Standard Test Method For Determining Vapor Pressure by Thermal Analysis
Copyright © ASTM International, 100 Barr Harbour Dr., PO box C-700 West Conshohocken, Pennsylvania 19428-2959, United States
Copyright by ASTM Int'l (all rights reserved); Tue Oct 25 05:52:16 EDT 2011
1
Downloaded/printed by
University of British Columbia Library pursuant to License Agreement. No further reproductions authorized.
E1782 – 08
inert gas for elevated pressures or for back-filling after evacu-
ation of the sample chamber is recommended to minimize the
risk of oxidation.
6.3 Partial blockage of the pinhole in the DSC containers
could occasionally be encountered. This may be observed as
noise spikes on the boiling endotherm and shall not be
considered a valid pressure-temperature datum point.
7. Apparatus
7.1 The essential equipment required to provide the mini-
mum instrument capability of this test method includes (see
Fig. 2):
7.1.1 Differential Scanning Calorimeter (DSC) or Differen-
tial Thermal Analyzer (DTA), consisting of:
7.1.1.1 DSC/DTA Test Chamber, composed of a furnace(s)
to provide uniform controlled heating of a specimen and
reference at a constant rate within the 273 to 773 K temperature
range of this test method; a temperature sensor to provide an
indication of the specimen/furnace temperature to 61 K; a
differential sensor to detect a difference (temperature or heat
flow) between the specimen and reference equivalent to 5 mW;
and a means of sustaining an inert gas or vacuum test chamber
environment at pressures above and below ambient.
7.1.1.2 Temperature Controller, capable of executing a
specific temperature program by operating the furnace(s)
6. Interferences
6.1 This test method is limited to materials that exhibit a
singular sharp boiling endotherm under the conditions outlined
in this test method.
6.2 Oxidation, pyrolysis, or polymerization of condensed
organic materials retained at temperatures above their ambient
boiling point may be encountered at the elevated pressures of N OTE —“A”, DSC/DTA instrument; “B,” pressure transducer;
this method. This will be observed as an exotherm or a “C,”pressure/vacuum source; “D,” pressure stabilizer; “E,” pressure
significantly broadened endotherm, or both, and shall not be regulator; and “F,” relief valve.
considered a valid pressure-temperature datum point. Use of an FIG. 2 Schematic of Apparatus
Copyright by ASTM Int'l (all rights reserved); Tue Oct 25 05:52:16 EDT 2011
2
Downloaded/printed by
University of British Columbia Library pursuant to License Agreement. No further reproductions authorized.
E1782 – 08
between selected temperature limits to 61 K at a rate of 7.2.2 A balance to weigh specimens or specimen containers,
temperature change of 5 K/min constant within 61 %. or both, to 60.1 mg.
7.1.1.3 Recording Device, to provide a means of acquiring, 7.2.3 A syringe or micropipet to deliver liquid specimens of
storing and displaying measured or calculated signals or both. 1 to 5 µL 610 %.
The minimum output signals are heat flow, temperature and 7.2.4 Pressure relief valve to prevent accidental overpres-
time. surization in the pressure system. A rating of 10 % in excess of
7.1.2 Pressure/Vacuum System, consisting of: the upper use pressure is suggested provided it does not exceed
7.1.2.1 A pressure vessel, or similar means of sealing the the maximum working pressure rating of any individual
test chamber at any applied absolute pressure within the 5 kPa component in the system.
to 2 MPa range of this test method.
7.1.2.2 Source of Pressurized Gas, or vacuum capable of 8. Precautions
sustaining a regulated inert gas pressure to the test chamber of 8.1 Safety Precautions:
between 5 kPa and 2 MPa. 8.1.1 Pressures in addition to ambient are employed in this
7.1.2.3 Pressure Transducer(s), to measure the pressure in test method. Ensure that the pressure/vacuum system is certi-
the test chamber to within 1 % including any temperature fied for operation at the extremes of pressure encountered with
dependence of the transducer(s) over the range of 5 kPa to 2 this test method. Incorporation of a pressure relief device is
MPa. recommended.
NOTE 1—Distance (or dead volume) between the pressure transducer 8.1.2 Adequate provisions shall be available for retention
and the specimen in the test chamber should be minimized to ensure and disposal of any spilled mercury if mercury-containing
accurate recording of the pressure at the time of boiling. pressure devices are employed.
7.1.2.4 Pressure Regulator, or similar device to adjust the
9. Sampling
applied pressure in the test chamber to 62 % of the desired
value. 9.1 Typical specimen sizes used for individual pressure
7.1.2.5 Ballast, or similar means to maintain the applied measurements are 1 to 5 mg of solid or 1 to 5 µL of liquid.
pressure in the test chamber constant to 61 %. Similar size specimens should be used for each individual
7.1.2.6 Valves, to control delivery of the inert gas/vacuum to measurement of a given sample.
the test chamber or to isolate components of the pressure/ 9.2 Samples are assumed to be tested as received. Report
vacuum system, or both. Valves shall be rated in excess of the any special sampling or pretreatment used with this test
2 MPa upper pressure limit of this test method. method.
7.1.3 Containers, (pans, capillary tubes, etc.) that are inert
to the specimen and reference materials and which are of 10. Calibration
suitable structural shape and integrity to contain the specimen 10.1 Perform calibration according to Practice E967, using
and reference in accordance with the following specific re- the heating rate and specimen containers intended for this test
quirements: method. Accomplish temperature calibration at ambient pres-
7.1.3.1 It is imperative that the containers used in this test sure.
method be capable of retaining the specimen in a manner that NOTE 3—The effect of pressure on the melting temperature of pure
minimizes sample loss through vaporization prior to boiling materials used to calibrate the temperature axis has been shown to be
and that promotes the development of vapor-liquid equilibrium <0.01 K at the maximum pressure of this method (4). The effects of
at boiling. When both conditions are met a sharp endotherm vacuum on the heat transfer characteristics and subsequent thermal lag of
with little or no baseline curvature at the onset will be various differential thermal instruments (DSC and DTA) have not been
observed. established. From general experiences these effects should not alter the
temperature axis calibration by more than 1 K at the minimum pressure of
NOTE 2—Studies by ASTM task group E37.01.05 and others (2, 3) have this test method.
determined glass cylindrical containers of 2 to 4 mm inside diameter by 25
10.2 Calibrate the pressure transducer according to the
mm long are suitable for thermocouple inserted style DTA instruments;
and a hermetic sealable pan (approximately 40 µL vol) with a single recommendations of the manufacturer or similar appropriate
pinhole in the center of the lid of #125 µm diameter is suitable for DSC procedure.
instruments. For purposes of this test method, pinhole diameters of 50 to
75 µm are recommended. These pinhole dimensions for DSC containers 11. Procedure
were established specifically for use with heating rates of nominally 5 11.1 Place the specimen and inert reference in suitable
K/min. Use of heating rates other than 5 K/min are not recommended for
containers (see 7.1.3) into the test chamber.
this test method. Higher rates may result in some self-pressurization of the
specimen and lesser rates will extend measurement times and will tend to NOTE 4—If hermetic sealable DSC pans with pinholed lids are used,
promote preboiling vaporization. make sure there is no sample material on the outer surfaces of the
7.2 Auxiliary equipment considered useful in conducting container and that a good hermetic seal is accomplished. Either will result
in preboiling vaporization that at least partially negates the function of the
this test method include: pinhole. Be certain, also, that the pinhole is free of dirt or debris.
7.2.1 A coolant system that can be coupled directly with the
controller to the furnace to hasten its recovery from elevated 11.2 Seal the test chamber and apply the desired pressure.
temperatures or to sustain a subambient temperature to within NOTE 5—It is recommended to flush residual oxygen from the test
61 K of a lower limit temperature. chamber by either purging for several minutes with inert gas or by
Copyright by ASTM Int'l (all rights reserved); Tue Oct 25 05:52:16 EDT 2011
3
Downloaded/printed by
University of British Columbia Library pursuant to License Agreement. No further reproductions authorized.
E1782 – 08
evacuation and back-filling with inert gas. NOTE 9—Deviations from linearity (curvature) due to expected de-
creases in enthalpy of vaporization with temperature (Antoine equation
11.3 Allow the pressure to stabilize and equilibrate the test
“C” constant negative) should not be confused with the abrupt deviation
chamber at a start temperature which shall be at least 30 K due to decomposition or polymerization. Curvature of normal data is
below the expected boiling temperature to ensure stable barely perceptible.
temperature control and baseline.
12.3 Calculate the Antoine vapor pressure equation con-
11.4 Heat the specimen and reference at a constant rate of 5
stants: A, B, and C retaining all available decimals using a
K/min, recording the DSC/DTA curve until the vaporization is
nonlinear least-squares regression program to fit the Antoine
complete.
equation, Log10 P = A − B/(T + C) to the corrected pressure-
11.5 Record the absolute pressure in the test chamber at the
temperature data points. Data for which any of the fitted
time the boiling endotherm is observed.
Antoine equation constants fall outside of ranges given in Note
NOTE 6—Most pressure gauges report pressure relative to ambient 10 shall be rejected.
pressure. In such cases, the measured pressure shall be added to or
subtracted from atmospheric pressure measured by a barometer to obtain NOTE 10—This test method uses the SI units of Kelvin for temperature
absolute pressure. instead of the traditional use of Celsius. The only effect of this change is
to make the “C” constant of the Antoine equation a negative value.
11.6 Restore the test chamber to ambient conditions upon Antoine equation constants (for log10, kPa, and K units) typically fall in
completion of the heating curve. these ranges:4 A, 4.9 to 7.8; B, 750 to 3000; and C, −235 to −3. If a
11.7 Repeat 11.1-11.6 with a new specimen at each of four nonlinear least-squares regression program is not available, a multilinear
or more additional pressures throughout the pressure range least-squares regression program may be used for Antoine equation fitting
capability of the equipment. by making these variable transformations:5 a = A; b = A * C − B; and
c = −C. The fitting equation, now linear in the parameters, is:
NOTE 7—A minimum of five measurements at different pressures are t*log10P5a*t 1 b 1 c*log10P
required for this test method. Additional measurements should improve
the fit of the Antoine vapor pressure equation and reduce the uncertainty 12.4 Using the Antoine equation with constants determined
of the Antoine constants used to calculate the vapor pressure curve. in 12.3, compute a table of pressure-temperature data pairs
encompassing the range of conditions tested including the
12. Calculation boiling point temperatures for pressures of 1000, 101.32 (760
12.1 Determine and tabulate each boiling temperature (as To torr), and 10 kPa and the vapor pressure at 293.15 K (20 °C).
or Te in Fig. 3) along with its corresponding observed pressure,
P. 13. Report
NOTE 8—Traditionally, the extrapolated onset temperature (To) of an 13.1 Report the following information:
endotherm recorded by DSC is used as the transition temperature; the peak 13.1.1 Sample identity, purity (if known or determined, as
maximum temperature (Te) is used for thermocouple inserted DTA well as how it was determined), and source,
configurations. The convention employed during temperature calibration 13.1.2 A table of the corrected pressure-temperature data
shall be used for these calculations. points,
12.2 If necessary, correct the observed pressures and tem- 13.1.3 The Antoine vapor pressure equation constants A, B,
peratures by the amount determined from the calibrations. Plot and C rounded to the nearest 0.000001,
the logarithms of pressure (log10P) versus the reciprocal of the 13.1.4 A table of the computed pressure-temperature data
absolute temperature (1/T, K−1). Examine this plot for any including the boiling point temperatures at 1000, 101.32 (760
abrupt deviation from linearity as evidence of instability. torr), and 10 kPa rounded to the nearest 0.1 K, and the vapor
pressure at 293.15 K (20 °C) rounded to the nearest 0.1 kPa.
13.1.5 A graphic presentation of the vapor pressure curve
including the regression fit to the Antoine vapor pressure
equation, and
13.1.6 Any abrupt deviation from linearity in the vapor
pressure curve and the temperature of its occurrence.
13.1.7 The specific dated version of this test method used.
14. Precision and Bias
14.1 Interlaboratory Study (ILS)—An interlaboratory study
for measurement of vapor pressure by this test method was
conducted in 1995. Two materials were studied: n-heptane and
distilled water. Each of six laboratories measured vapor pres-
sure data for n-heptane; each of six laboratories measured
vapor pressure data for distilled water. Practice E691 was
followed for the ILS design and for the analysis of the data.
4
These ranges were determined by examination of Antoine equation literature
FIG. 3 Boiling Endotherm from DSC/DTA Instrumentation with: and databanks, for example, see Refs (1) and (5).
5
extrapolated Onset, To and Peak Maximum, Te This procedure was described in Ref (6).
Copyright by ASTM Int'l (all rights reserved); Tue Oct 25 05:52:16 EDT 2011
4
Downloaded/printed by
University of British Columbia Library pursuant to License Agreement. No further reproductions authorized.
E1782 – 08
14.2 Test Results—The precision information given below Boiling point (K) 320.4 4.9 5.1 319.0
is for duplicate determinations of vapor pressure data calcu- at 10.0 kPa
Vapor pressure (kPa) 2.1 1.2 1.2 2.3
lated from the derived Antoine equations as described in 12.3. at 93.15 K
14.3 Precision—The terms repeatability and reproducibility
given below are used as specified in Practice E177. Standard
deviations among study results may be calculated by dividing 14.4 Bias:
the third and fourth table columns by 2.8. 14.4.1 Bias for n-Heptane—The values listed in the NIST
14.3.1 Precision for n-Heptane: column in 14.3.1 can be used as accepted reference values as
ILS Repeat- Repro- NIST defined in Practice E177. The deviation of the study results
Average ability duc- Data
ibility from the NIST Data is less than the reproducibility bounds.
Boiling point (K) at 1000 kPa 467.7 11.3 19.1 474.3 14.4.2 Bias for Distilled Water—The values listed in the
Normal Boiling point (K) at 371.2 2.0 3.5 371.6
101.325 kPa NIST column in 14.3.2 can be used as acceptable reference
Boiling point (K) at 10.0 kPa 308.3 3.2 3.6 308.5 values as defined in Practice E177. The deviation of the study
Vapor pressure (kPa) at 4.9 0.9 1.0 4.7
293.15 K
results from the NIST Data is less than the reproducibility
bounds.
14.3.2 Precision for Distilled Water:
ILS Repeat- Repro- NIST
Average ability duc- Data
15. Keywords
ibility
Boiling point (K) 450.9 3.5 10.0 453.1
15.1 Antoine equation; boiling pressure; boiling tempera-
at 1000 kPa ture; differential scanning calorimetry (DSC); differential ther-
Normal Boiling point (K) 373.6 1.0 2.9 373.2 mal analysis (DTA); vapor pressure
at 101.325 kPa
APPENDIX
(Nonmandatory Information)
Copyright by ASTM Int'l (all rights reserved); Tue Oct 25 05:52:16 EDT 2011
5
Downloaded/printed by
University of British Columbia Library pursuant to License Agreement. No further reproductions authorized.
E1782 – 08
REFERENCES
(1) Thompson, G. W., Chemical Reviews, Vol 38, 1946, pp. 1 – 39. International Temperature Scale of 1990 (ITS-90),” NIST-Technical
(2) Perrenot, B., de Valliere, P., and Widman, G., Journal of Thermal Note 1265.
Analysis, Vol 38, 1992, pp. 381 – 390. (5) Boublik, T., Fried, V., and Hala, E., The Vapor Pressures of Pure
(3) Contreras, M. D., Girela, F., and Parera, A., Thermochimica Acta, Vol Substances, Elsevier, New York, 1973.
219, 1993, pp. 167 – 172.
(4) Mangum, B. W., and Furukawa, G. T., “Guidelines for Realizing the (6) Willingham et al., J. Res. NBS, Vol 35, 1945, pp. 219 – 244.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/
COPYRIGHT/).
Copyright by ASTM Int'l (all rights reserved); Tue Oct 25 05:52:16 EDT 2011
6
Downloaded/printed by
University of British Columbia Library pursuant to License Agreement. No further reproductions authorized.