0% found this document useful (0 votes)
30 views

A Novel Approach To Consecutive Extraction of Drugs With Different Properties Via On Chip Electromembrane Extraction

This document describes a novel approach for the consecutive extraction of drugs with different chemical properties via on-chip electromembrane extraction coupled with high performance liquid chromatography. The chip consists of two channels carved in polymethyl methacrylate parts separated by a porous polypropylene membrane impregnated with an organic solvent. Platinum electrodes at the bottom of the channels are connected to a power supply to drive the consecutive extraction when a sample containing betaxolol, naltrexone and nalmefene is passed through. This method allows for the analysis of these drugs from biological samples.

Uploaded by

shayan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
30 views

A Novel Approach To Consecutive Extraction of Drugs With Different Properties Via On Chip Electromembrane Extraction

This document describes a novel approach for the consecutive extraction of drugs with different chemical properties via on-chip electromembrane extraction coupled with high performance liquid chromatography. The chip consists of two channels carved in polymethyl methacrylate parts separated by a porous polypropylene membrane impregnated with an organic solvent. Platinum electrodes at the bottom of the channels are connected to a power supply to drive the consecutive extraction when a sample containing betaxolol, naltrexone and nalmefene is passed through. This method allows for the analysis of these drugs from biological samples.

Uploaded by

shayan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 25

View Article Online

View Journal

Analyst
Accepted Manuscript

This article can be cited before page numbers have been issued, to do this please use: Y. Abdossalami
Asl, Y. Yamini and S. Seidi, Analyst, 2015, DOI: 10.1039/C5AN02019K.

This is an Accepted Manuscript, which has been through the


Royal Society of Chemistry peer review process and has been
accepted for publication.

Accepted Manuscripts are published online shortly after


acceptance, before technical editing, formatting and proof reading.
Using this free service, authors can make their results available
to the community, in citable form, before we publish the edited
article. We will replace this Accepted Manuscript with the edited
and formatted Advance Article as soon as it is available.

You can find more information about Accepted Manuscripts in the


Information for Authors.

Please note that technical editing may introduce minor changes


to the text and/or graphics, which may alter content. The journal’s
standard Terms & Conditions and the Ethical guidelines still
apply. In no event shall the Royal Society of Chemistry be held
responsible for any errors or omissions in this Accepted Manuscript
or any consequences arising from the use of any information it
contains.

www.rsc.org/analyst
Page 1 of 24 Analyst
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
4 A novel approach to consecutive extraction of drugs with different properties via
5
6 on chip electromembrane extraction
7
8
9
10
11
12
Yousef Abdossalami Asla, YadollahYaminia,∗, Shahram Seidib
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13
14
a
15 Department of Chemistry, Tarbiat Modares University, P.O. Box 14115-175, Tehran, Iran
16 b
Department of Analytical Chemistry, Faculty of Chemistry, K.N. Toosi University of Technology, Tehran, Iran

Analyst Accepted Manuscript


17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52 ∗
Corresponding author at: Tarbiat Modares University, Department of Chemistry, P.O. Box 14115-175, Tehran, Iran.
53
Tel.: +98 21 82883417; Fax: +98 21 82883455.
54
E-mail address: [email protected] (Y. Yamini).
55
56
57
58 1
59
60
Analyst Page 2 of 24
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
4 Abstract
5
6 In the present research, for the first time, a consecutive on chip electromembrane extraction
7
8
9 coupled with high performance liquid chromatography was introduced for analysis of betaxolol
10
11 (Bet), naltrexone (Nalt) and nalmefene (Nalm) as model analytes with different chemical properties
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 from biological samples. The chip consists of two polymethyl methacrylate (PMMA) parts which
14
15
16 two microfluidic channels are craved in each part. These channels were used as a flow path for

Analyst Accepted Manuscript


17
18 sample solution and a thin compartment for acceptor phase. A porous polypropylene sheet
19
20 membrane impregnated with an organic solvent was placed between two parts of chip device in
21
22
23 order to separate the channels. Two platinum electrodes were bent at the bottom of these channels
24
25 that are connected to a power supply providing the electrical driving force for migration of ionized
26
27
28
analytes from the sample solution through the porous sheet membrane into the acceptor phase. The
29
30 new setup provides effective and reproducible extractions by using low volume of sample solution.
31
32 Efficient parameters on consecutive electromembrane extraction of the model analytes were
33
34
35 optimized by using one variable at a time method. Under the optimized conditions, the new setup
36
37 offered a good linearity in the range of 10.0–500 µg L-1with coefficient of determination (R2) higher
38
39 than 0.9932. The relative standard deviation (RSD %) and LOD values were less than 6.8% based on
40
41
42 four replicate measurements and 10.0 µg L-1for the model analytes, respectively. The
43
44 preconcentration factors higher than 15.6-fold were obtained. Finally, the proposed method was
45
46 successfully applied for determination and quantification of the model analytes in biological
47
48
49 samples.
50
51 Keywords: Consecutive on chip extraction; Electromembrane extraction; Betaxolol;
52
53
54
Naltrexone; Nalmefene.
55
56
57
58 2
59
60
Page 3 of 24 Analyst
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
4 Introduction
5
6 Sample preparation has unique meaning and special importance in the field of analytical
7
8
9 chemistry. For several decades, liquid–liquid extraction (LLE) has been employed for sample
10
11 preparation prior to analysis by high performance liquid chromatography and gas chromatography 1-
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 5
. Recent developments of sample preparation are directed toward down scaling and automation of
14
15
16 conventional extraction methods by decreasing acceptor-to-donor ratio in order to reduce organic

Analyst Accepted Manuscript


17
18 solvent, sample solution and chemical reagents consumption 6. Liquid phase micro extraction
19
20 (LPME) is a miniaturization of the traditional LLE that performed in several modes such as single
21
22
23 drop microextraction, dispersive liquid-liquid microextraction, porous membrane based LPME and
24
25 7-13
solidification of floating organic droplet . Electromembrane extraction (EME), as one of the
26
27 14
28
LPME methods that was introduced by Pedersen-Bjergaard and coworkers in 2006 . In this
29
30 method, the mass transfer is carried out through an organic solvent immobilized within the pores of a
31
32 porous membrane by applying an electrical potential difference, as the driving force, between two
33
34
35 platinum electrodes placed on both sides of the porous membrane in donor and acceptor phases.
36
37 EME has numerous advantages in comparison with other liquid phase microextraction and provides
38
39 more efficient, fast, very selective extractions and excellent sample cleanup 15.
40
41
42 In addition to miniaturization of extraction techniques, downscaling of extraction devices has
43
44 important role in decreasing of organic solvent, sample solution and chemical reagents consumption.
45
46 Miniaturization of analytical methods and instruments for extraction researches is an area of
47
48
49 burgeoning interest. In many cases, the reduction in size of analytical procedure or technique often
50
51 translates to a reduction in analysis time and costs. Miniaturized devices integrate sample
52
53
54
preparation, separation and detection on a single small chip 16-18. A miniaturized and portable device
55
56 is more preferable for the on-site and rapid analysis application.
57
58 3
59
60
Analyst Page 4 of 24
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
The combination of microfluidic technologies and EME would be suitable for such
4
5
6 applications. For the first time, on chip electromembrane extraction (EME)as one type of
7
8 miniaturized membrane based LPME techniques was introduced by Pedersen-Bjergaard and
9
10 19
11
Rasmussen in 2010 . In this technique, a piece of porous membrane is used as the support for
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 water–immiscible organic solvent. Afterwards, several reports have been emerged on


14
15 implementation of EME as chip based devices to make more developments in this field of area 20-23.
16

Analyst Accepted Manuscript


17
18 In all of the reported papers using chip devices, one extraction chamber has been used for
19
20 extraction. Therefore, analytes with different properties including hydrophilicity, polarity and
21
22 functional groups cannot be extracted with these devices. By considering the polarity of various
23
24
25 analytes, different supported liquid membrane (SLM) compositions are required to extract the target
26
27 analytes. Adding of a carrier into SLM composition has different effects on the extraction of
28
29
analytes. It has been demonstrated that the use of carriers in SLM composition has strong negative
30
31
32 effect on extraction of nonpolar drugs such as betaxolol (Bet) 24, 25. However, the presence of di-(2-
33
34 ethylhexyl) phosphate (DEHP) as an anionic ion pairing reagent is necessary for the extraction of
35
36
37 compounds such as naltrexone (Nalt) and nalmefene (Nalm) 26. Therefore, Bet could be extracted by
38
39 use of only pure 2-nitrophenyl octyl ether (NPOE) as the SLM, whereas for the extraction of Nalm
40
41 and Nalt a mixture of NPOE and DEHP is required. Consequently, the simultaneous extraction of
42
43
44 these analytes is impossible.
45
46 In the current paper, consecutive electromembrane extraction chip as a new on chip EME setup
47
48 was introduced for extraction and determination of Nalt, Nalm and Bet from biological samples. The
49
50
51 used analytes were selected as the model compounds with different physicochemical properties to
52
53 show the ability of the proposed technique for simultaneous extraction of analytes with different
54
55
56
chemical properties. In fact, the physicochemical properties of analytes are the important keys for
57
58 4
59
60
Page 5 of 24 Analyst
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
the selection of membrane organic solvents so that with selection of a specific composition of
4
5 24
6 membrane organic solvent selective extraction can be achieved in EME . The proposed method
7
8 provides the using possibility of different compositions of membrane organic solvent and thus
9
10
11
simultaneous extraction of different analytes with different chemical properties. As well known,
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 todays many combination drugs, which are a mixture of two or more drugs, are used by the patients.
14
15 On the other hand a patient may use several drugs, simultaneously. The proposed method can be
16

Analyst Accepted Manuscript


17
18 used an efficient sample preparation method because it provides the simultaneous analysis
19
20 possibility of all drugs by a little volume of a biological sample whereas by other techniques
21
22 separated samples, and consequently larger volumes of biological samples, should be used for
23
24
25 analysis of analytes with different physicochemical properties. Here, two extraction chambers were
26
27 employed in which anode and cathode electrodes embedded at the bottom of microfluidic extraction
28
29
channels, and SLMs were separately optimized for each class of analytes. As a result, the proposed
30
31
32 system offers suitable conditions for the consecutive extraction of Bet, Nalt and Nalm as model
33
34 analytes and can be applied for the extraction of other compounds with similar properties. Effects of
35
36
37 different variables on the extraction efficiency of the analytes were studied and optimized. After
38
39 optimization, the method followed by high-performance liquid chromatography–ultraviolet detection
40
41 (HPLC–UV) was applied for the extraction and determination of Bet, Nalt and Nalm in human urine
42
43
44 and plasma samples.
45
46
47 Experimental
48
49
50 Chemicals and reagents
51
52 Betaxolol was kindly donated by Sina Darou (Tehran, Iran). Nalmefene and naltrexone were
53
54 obtained from Parand Darou (Tehran, Iran). Chemical structures, pKa and log P, which P is octanol-
55
56
57
58 5
59
60
Analyst Page 6 of 24
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
water partition coefficient, of the analytes were reported in Table 1. 2-Nitrophenyloctyl ether
4
5
6 (NPOE), 2-nitrophenyl phenylether (NPPE), tris-(2-ethylhexyl) phosphate (TEHP) and di-(2-
7
8 ethylhexyl) phosphate (DEHP) were purchased from Fluka (Buchs, Switzerland). All of the
9
10
11
chemicals used were analytical-reagent grades. The Accurel 2E HF (R/P) polypropylene membrane
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 sheet was supplied by Membrana (Wuppertal, Germany) with the thickness of 150 µm, and pore size
14
15 of 0.2 µm. Ultrapure water was prepared by a Younglin 370 series aqua MAX purification
16

Analyst Accepted Manuscript


17
18 instrument (Kyounggi-do, Korea).
19
20 A stock solution containing 2.0 mg mL−1 of Bet, Nalt and Nalm was prepared in methanol. All
21
22 standard solutions were stored at 4°C which is protected from light. Working standard solutions
23
24
25 were prepared daily by dilution of the stock solutions with ultrapure water.
26
27 Real samples
28
29
Urine sample: To plot the calibration curves and to obtain figures of merit, a human urine
30
31
32 sample was collected from a 29-year-old healthy adult male volunteer. The sampling procedure was
33
34 performed according to the guidelines for research ethics. The protocol was approved by an Internal
35
36
37 Review Board. Moreover, written consent was obtained from volunteers prior to the experiment. The
38
39 sample was filtered through 0.45-µm pore size cellulose acetate filter from Millipore (Madrid,
40
41 Spain). The filtrate was collected in a glass vial, which was carefully cleaned with hydrochloric acid
42
43
44 and washed with deionized water and stored at 4°C to prevent bacterial growth and proteolysis. The
45
46 urine sample was diluted 1:4 with ultra-pure water and the pH value was adjusted to 7.00 by adding
47
48 proper amounts of hydrochloric acid (100 mM) and/or sodium hydroxide (100 mM) solutions. Then,
49
50
51 the urine sample was spiked with a mixed standard solution to obtain the desired concentration. One
52
53 milliliter of the sample solution was subsequently submitted to on chip EME procedure. Further
54
55
56
urine sample was obtained from a healthy volunteer (29-year-old).
57
58 6
59
60
Page 7 of 24 Analyst
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
Plasma sample: Frozen drug-free human plasma sample (blood group B+) was obtained from
4
5
6 the Iranian Blood Transfusion Organization (Tehran, Iran). Prior to use, the plasma sample was
7
8 allowed to thaw at room temperature. According to literature, trifluoroacetic acid (TFA) is a suitable
9
10
11
reagent for protein precipitation and decreasing the matrix effect of real samples such as human
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 plasma 27. To this end, four drops of concentrated TFA (99 w/w%) were added into 10 mL of each
14
15 plasma sample and centrifuged for 5 minutes at 3000 rpm, afterwards. Finally, the supernatant
16

Analyst Accepted Manuscript


17
18 sample solutions were transferred to a glass container and stored in 4 °C for subsequent experiments.
19
20 The pH of 5-fold diluted human plasma samples with ultra-pure water was adjusted at 7.0 using the
21
22 proper amounts of hydrochloric acid (100 mM) and/or sodium hydroxide (100 mM) solutions. Then,
23
24
25 the plasma sample was spiked with a mixed standard solution to obtain the desired concentration.
26
27 One milliliter of the sample solution was subsequently submitted to on chip EME procedure.
28
29
30
Chromatographic apparatus
31
32 The chromatographic equipment consisted of an Agilent 1260 HPLC system, including a
33
34 quaternary pump, degasser, and UV-Vis detector (Waldbronn, Germany). Chromatographic data
35
36
37 were recorded and analyzed by using Chem Station for LC systems software (version B.04.03). The
38
39 separations were performed on an ODS-3 column (150 mm × 4.6 mm, with particle size of 5 µm)
40
41 from MZ-Analysentechnik (Mainz, Germany) via an isocratic elution at the flow rate of 1.0 mL
42
43
44 min−1. For Bet, Nalt and Nalm, the mobile phase consisted of 10 mM phosphate buffer (pH = 8.0)
45
46 and acetonitrile (55:45) respectively for 15 minutes. The detector wavelength was set at 215 nm.
47
48
49
Manufacture of on chip EME device
50
51
52 The structure of the consecutive chip device for electromembrane extraction is shown in Fig. 1.
53
54 The chip consists of two polymethyl methacrylate (PMMA) parts. In each part two long channels
55
56
57
with a length of 30 mm, depth of 200 µm and width of 1.0 mm was craved. As shown in Fig. 1, the
58 7
59
60
Analyst Page 8 of 24
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
channels, which were craved in upper part, acted as acceptor phase channels, whereas the other two
4
5
6 channels locating in lower part are a flow path for donor solution. Moreover, in each part of the
7
8 consecutive chip device, six individual holes (with I.D. of 0.5 mm) were drilled. The holes of “a”
9
10
11
and “f” were connected to inlet and outlet tubes whereas the holes of “b” and “d” were used for the
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 entrance of platinum electrodes. Holes of “c” and “e” were connected together to deliver extraction
14
15 solutions from one extraction chamber to the other one. In each part of the consecutive chip device,
16

Analyst Accepted Manuscript


17
18 two platinum electrodes were bent and mounted at the bottom of each channel along with its length
19
20 after passing through the holes of “b” and “d”. All channels and holes were milled using a CNC
21
22 micromilling machine model SMG-302 CNC milling from Sadrafan Gostar Industries (Tehran,
23
24
25 Iran). The channels of sample solution and acceptor phase were separated with two separated small
26
27 pieces of porous polypropylene sheet membrane during extraction which was impregnated with
28
29
suitable organic solvents regarding the physicochemical properties of the model analytes. The sheet
30
31
32 membranes were located between the two parts along with lengths of the channels and fixed using
33
34 four bolts and nuts during the extraction procedure. After each extraction step, the porous sheet
35
36
37 membrane was replaced with a new one. The sample solution was flowed toward inlet of the sample
38
39 channel by a syringe pump. In addition, a micro syringe was used for introducing and withdrawing
40
41 of a microliter volume of acceptor solution into the acceptor phase channel.
42
43
44 Consecutive on chip EME procedure
45
46 One milliliter of the sample solution containing the target analytes was filled into a Hamilton
47
48 syringe and then the syringe mounted on a syringe pump. To impregnate the organic solvent into the
49
50
51 pores of the sheet membrane, two 3 mm × 4 cm pieces of the porous membrane were cut and dipped
52
53 into the suitable organic solvents for 5 s and then the excess of organic solutions were gently wiped
54
55
56
away by using a piece of Kleenex. Then, the membranes were carefully placed between two parts of
57
58 8
59
60
Page 9 of 24 Analyst
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
the chip device. The upper and lower parts of the chip were fixed together using bolts and nuts.
4
5
6 Thirty microliters of 100 mM HCl (acceptor solution) was introduced into the acceptor channel by a
7
8 microsyringe. Donor phase pumped through the donor channel using a syringe pump. When the
9
10
11
extraction was completed, the acceptor solution was collected by a microsyringe and injected into
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 HPLC injection loop. After each extraction, the channels of sample solution and acceptor phase were
14
15 washed by ultrapure water and a new porous membrane impregnated with suitable organic solvent
16

Analyst Accepted Manuscript


17
18 was located and fixed between two parts of extraction chip device for subsequent experiment. Fig. 2
19
20 illustrates the extraction process, schematically.
21
22
23 Calculation of preconcentration factor, extraction recovery, and relative recovery
24
25
26 The preconcentration factor (PF) was defined as the ratio of the final analyte concentration in
27
28 the acceptor phase (Cf,a) and the initial concentration of analyte (Ci,s) in the sample solution:
29
30
,
 =
31
32 ,
(1)
33
34
35 where Cf,a was calculated from a calibration graph obtained from direct injection of standard
36
37
38 solutions of the analytes. The extraction recovery (ER%) was defined as the percentage of the
39
40 number of moles of the analyte extracted into the acceptor phase (nf,a) respected to the number of
41
42 moles originally present into the sample solution (ni,s).
43
44
45 , ×,
%= × 100 = × 100
,
46 (2)
, , ×,
47
48
49

50 % =  ,  ×  × 100 (3)
51 ,
52
53
54 where Vf,a and Vi,s are the volumes of the acceptor phase and sample solution, respectively. The
55
56 relative recovery (RR%) was calculated by the following equation:
57
58 9
59
60
Analyst Page 10 of 24
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3   
4 %= 
× 100 (4)
5
6
7
! %= % − 100 (5)
8
9
10
11 where Cfound, Creal, and Cadded are the concentrations (µg L-1) of analyte after adding a known amount
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 of standard into the real sample, the concentration of analyte in real sample, and the concentration of
14
15 a known amount of standard, which was spiked into the real sample, respectively.
16

Analyst Accepted Manuscript


17
18
19 Results and discussion
20
21
22 Selection of supported liquid membrane
23
24 Composition of the organic solvent which was used as SLM has an important role in EME. A
25
26 suitable liquid membrane enhances the extraction efficiency and selectivity. Composition of
27
28
29 membrane affects the diffusion coefficient of analyte. Furthermore, the range of applied voltage is
30
31 determined by nature of SLM composition. Considering the above points, some organic solvents
32
33 such as 2-nitrophenyl octyl ether (NPOE), NPOE containing 5, 10 and 15% (v/v) di-(2-ethylhexyl)
34
35
36 phosphate (DEHP), NPOE containing 5, 10 and 15% (v/v) tris-(2-ethylhexyl) phosphate (TEHP) and
37
38 NPOE containing 5% TEHP and 5% DEHP was investigated as SLM for the consecutive on chip
39
40
EME. As shown in Fig. 3A, NPOE containing 10% DEHP represented higher extraction efficiency
41
42
43 for Nalt and Nalm where suitable extraction efficiency for Bet was provided when pure NPOE was
44
45 used as SLM. Thus, NPOE containing 10% DEHP and NPOE were selected as the optimum SLM
46
47
48 compositions for the subsequent experiments.
49
50
51 Effect of sample solution and acceptor phase compositions
52
53 The pHs of sample solution and acceptor phase are critical parameters in extraction efficiency
54
55
56
of consecutive on chip EME. In EME, the electrokinetic migration of the analytes across the SLM
57
58 10
59
60
Page 11 of 24 Analyst
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
into the acceptor solution is affected by ion balance which is defined as the ratio of the total ionic
4
5 28
6 concentration in the sample solution to that in the acceptor solution . In EME, for basic analytes,
7
8 ion balance is mainly considered as the ratio of HCl concentration in sample solution to acceptor
9
10
11
phase 29. Based on the previous studies on the role of ion balance, it was anticipated that decreasing
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 of ion balance leads to an increase in the flux of ions through SLM. Therefore, for a basic drug, the
14
15 extraction efficiency is increased by increasing the HCl concentration into the acceptor phase and its
16

Analyst Accepted Manuscript


17
18 decreasing into the sample solution. It can be attributed to this fact that increasing of the HCl
19
20 concentration into the acceptor phase not only avoids from partial deprotonation of the analyte and
21
22 its back-diffusion into SLM but also, accelerate the analyte release from SLM into the acceptor
23
24
25 solution. On the other hand, for extraction of the basic analyses into the acceptor phase, acidification
26
27 of the sample solution is performed to keep the targeted compounds in their ionized form. However,
28
29
30
there are some limitations for increasing of HCl concentration in both donor and acceptor phase. By
31
32 applying an electrical field in EME, cations migrate toward cathode and anions migrate toward
33
34 anode through the SLM. Increasing content of ions in each of donor and acceptor phase results to
35
36
37 increasing number of ions migrate through SLM at a given moment, it causes to increasing of the
38
39 current level, increasing of Joule heating therefore instability of SLM. Increasing of the current level
40
41 between electrodes increase electrolysis reactions on the surfaces of electrodes and therefore bubble
42
43
44 formation risk. Bubble formation increases uncertainties in obtained data for EME. According to the
45
46 previous EME works that were carried out for extraction of these model analytes 24-26, 10.0 mM and
47
48
100.0 mM of HCl were selected as donor and acceptor solutions, respectively.
49
50
51
52 Effect of applied voltage
53
54 In EME, the electrical field, which is provided by applying an electrical potential between two
55
56 platinum electrodes, stimulates the transfer of analytes through the SLM into the acceptor phase.
57
58 11
59
60
Analyst Page 12 of 24
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
Therefore, mass transfer is greatly depended on the electrical field so that it is expected that
4
5
6 extraction efficiency can be increased by increasing the applied voltage. However, at the high
7
8 applied voltages, several problems such as bubble formation due to electrolysis process, SLM
9
10 29
11
punctuation, and sparking may be occurred . Therefore, selecting an optimum applied voltage is
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 necessary to succeed electromembrane extraction. Hence, a series of experiments with various


14
15 extraction voltages in the range of 0-140 V were examined to determine the most suitable voltage
16

Analyst Accepted Manuscript


17
18 while 10 mM HCl and 100 mM HCl was used as donor and acceptor solutions, respectively, for 30
19
20 min. As shown in Fig. 3B, relatively high peak areas were obtained at the voltage of 100 V.
21
22 Therefore, electrical potential of 100 V was applied for future experiments.
23
24
25
26 Effect of sample volume
27
28 Sampling is one of the main problems related to the analysis of biological real samples. Except
29
30 urine, preparing sample with large volumes for other biological samples is difficult. The main
31
32
33 purpose for design and construction of a consecutive on chip EME was low consumption of sample
34
35 solution. In the current research, by considering the above reasons, 1.0 mL of sample solution was
36
37
selected for the subsequent experiments.
38
39
40
41 Effect of sample solution flow rate and extraction time
42
43 A syringe pump was used to pump the sample solution into the on chip EME device. The flow
44
45 rate of the sample solution plays an essential role in increasing the kinetics and efficiency of
46
47
48 extraction by delivering a sample solution into extraction chambers, increasing the mass transfer and
49
30, 31
50 decreasing the thickness of the double layer around SLM . Various sample flow rates in the
51
52
range of 10–60 µL min-1 were examined. As shown in Fig. 3C, extraction efficiency increased by
53
54
55 increasing sample flow rate from 10 to 30 µL min-1, whereas a decreasing was observed by further
56
57
58 12
59
60
Page 13 of 24 Analyst
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
increasing of the flow rate. When the sample flow rate is increased considerably, target analysts do
4
5
6 not have enough time to pass through the SLM into acceptor phase. Also, perforation in the porous
7
8 membrane as well as losing of impregnated organic solvents into the pores of porous membrane
9
10
11
sheets were observed at high sample flow rates.
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 Fig. 3C also illustrates the effect of extraction time on the extraction efficiency. Extraction
14
15 time depends on the sample flow rate, inversely. Therefore, extraction time and sample flow rate was
16

Analyst Accepted Manuscript


17
18 investigated together. As can be seen, extraction time is increased by decreasing of sample flow rate.
19
20 But, stability of SLM is decreased at long extraction durations. When the sample solution is pumped
21
22 at low flow rates, accumulation of the ions around SLM is increased which leads to increasing of the
23
24
25 charge density of the double layer; increasing of electrokinetic migration of ions through SLM;
26
27 increasing of Joule heating and consequently instability of SLM. Therefore, 30 µL min-1 was
28
29
selected as the optimum sample flow rate.
30
31
32
33 Method evaluation
34
35 To verify the practical applicability of the proposed technique, figures of merit of the proposed
36
37 method were studied by using standard solutions of the analytes in drug-free plasma and urine
38
39
40 samples. Optimal conditions were applied to find out linearity, repeatability, and LODs of this
41
42 method that is summarized in Table 2. The calibration curves were linear in the range of 10.0–500
43
44
45
µg L-1 with a coefficient of determination (R2) more than 0.9932. The relative standard deviations
46
47 (RSD%) for extraction and determination of the model analytes were less than 6.8% based on four
48
49 replicate measurements. LODs less than 10.0 µg L-1 were observed for the target analytes. PF values
50
51
52 higher than 15.6-fold were obtained for the extraction of Bet, Nalt and Nalm at the concentration
53
54 level of 100 µg L-1. According to Clarke's analysis of drugs and poisons 32
, the serum therapeutic
55
56 concentration range for Bet is 5 to 50 µg L-1. After a single oral dose of 100 mg to four subjects,
57
58 13
59
60
Analyst Page 14 of 24
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
peak plasma concentrations of 10 to 60 (mean 40) µg L-1 of naltrexone were attained in 1 to 2 h. The
4
5
6 peak plasma concentration of 24 healthy males administrated with Nal, aged between 19 and 46
7
8 years, were in the range of 12-110 µg L-1. Consequently, the proposed method can be applied for
9
10
11
extraction and determination of low levels of Nalm, Nalt and Bet from biological fluids.
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13
14 Application of the proposed method for analysis of real samples
15
16 In order to illustrate the performance of the consecutive on chip EME, the procedure was used

Analyst Accepted Manuscript


17
18
19
for extraction and determination of Bet, Nalt and Nalm in urine and plasma samples. The obtained
20
21 results are presented in Table 3. As it can be seen, model analytes were not found in the investigated
22
23 urine and plasma samples. Therefore, urine and plasma samples were spiked with 100 µg L-1 of Bet,
24
25
26 Nalt and Nalm. The relative error for extraction and determination of the model analytes in urine and
27
28 plasma samples were less than 5.2%. According to literature, the mean recovery will be within 90 to
29
30 110% (or relative Error% between -10% to +10%) 33. Therefore, the results showed that this method
31
32
33 is applicable for the real samples. Fig. 4 shows the typical chromatograms of the extracted model
34
35 analytes from urine and plasma samples before and after spiking with 100 µg L-1 of Bet, Nalt and
36
37
Nalm.
38
39
40
41 Conclusions
42
43 In the present paper, for the first time, a consecutive on the chip EME extraction device was
44
45 introduced for the analysis of trace amounts of Bet, Nalt and Nalm as model analytes from low
46
47
48 volumes of urine and plasma samples. The effect of various variables affects the extraction
49
50 efficiency of the on chip EME procedure were investigated and optimized. Making the extraction
51
52
possibility from low volumes of biological fluids, providing considerable sample cleanup and
53
54
55 acceptable LODs are the advantages that can be accounted for the proposed method. The results
56
57
58 14
59
60
Page 15 of 24 Analyst
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
indicated that the consecutive on chip EME may be a good alternative for drug analysis from
4
5
6 biological samples.
7
8
9 References
10
11
12 1. H. Ebrahimzadeh, Y. Yamini, F. Kamarei and S. Shariati, Anal. Chim. Acta, 2007, 594, 93-
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13
14 100.
15
16
2. M. Rezaee, Y. Yamini, M. Moradi, A. Saleh, M. Faraji and M. H. Naeeni, J. Supercrit.

Analyst Accepted Manuscript


17
18
19 Fluids, 2010, 55, 161-168.
20
21 3. M. H. Naeeni, Y. Yamini and M. Rezaee, J. Supercrit. Fluids, 2011, 57, 219-226.
22
23
24 4. S. Seidi, Y. Yamini and M. Rezazadeh, J. Chromatogr. B, 2013, 913–914, 138-146.
25
26 5. H. Chen, Q. Fang, X.-F. Yin and Z.-L. Fang, Lab Chip, 2005, 5, 719-725.
27
28 6. A. Sarafraz-Yazdi and A. Amiri, TrAC Trends Anal. Chem., 2010, 29, 1-14.
29
30
31 7. Y. Ke, W. Li, Y. Wang, X. Hao, R. Jiang, F. Zhu and G. Ouyang, Microchem. J., 2014, 117,
32
33 187-193.
34
35
8. H. Ebrahimzadeh, Y. Yamini, A. Gholizade, A. Sedighi and S. Kasraee, Anal. Chim. Acta,
36
37
38 2008, 626, 193-199.
39
40 9. Y. Assadi, M. A. Farajzadeh and A. Bidari, in Comprehensive Sampling and Sample
41
42
43
Preparation, ed. J. Pawliszyn, Academic Press, Oxford, 2012, pp. 181-212.
44
45 10. Y.-H. Jian, Y. Hu, T. Wang, J.-L. Liu, C. Zhang and Y. Li, Chin. J. Anal. Chem., 2010, 38,
46
47 62-66.
48
49
50 11. M.-I. Leong and S.-D. Huang, J. Chromatogr. A, 2009, 1216, 7645-7650.
51
52 12. M. R. Khalili Zanjani, Y. Yamini, S. Shariati and J. Å. Jönsson, Anal. Chim. Acta, 2007, 585,
53
54 286-293.
55
56
57
58 15
59
60
Analyst Page 16 of 24
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
13. T. Honda, M. Miyazaki, Y. Yamaguchi, H. Nakamura and H. Maeda, Lab Chip, 2007, 7,
4
5
6 366-372.
7
8 14. S. Pedersen-Bjergaard and K. E. Rasmussen, J. Chromatogr. A, 2006, 1109, 183-190.
9
10
11
15. Y. A. Asl, Y. Yamini, M. Rezazadeh and S. Seidi, Anal. Methods, 2015, 7, 197-204.
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 16. Z.-X. Cai, Q. Fang, H.-W. Chen and Z.-L. Fang, Anal. Chim. Acta, 2006, 556, 151-156.
14
15 17. A. Berduque, J. O’Brien, J. Alderman and D. W. M. Arrigan, Electrochem. Commun., 2008,
16

Analyst Accepted Manuscript


17
18 10, 20-24.
19
20 18. H. Shen and Q. Fang, Talanta, 2008, 77, 269-272.
21
22 19. N. Petersen, H. Jensen, S. Hansen, S. Foss, D. Snakenborg and S. Pedersen-Bjergaard,
23
24
25 Microfluid Nanofluidics, 2010, 9, 881-888.
26
27 20. N. J. Petersen, J. S. Pedersen, N. N. Poulsen, H. Jensen, C. Skonberg, S. H. Hansen and S.
28
29
Pedersen-Bjergaard, Analyst, 2012, 137, 3321-3327.
30
31
32 21. S. Seidi, M. Rezazadeh, Y. Yamini, N. Zamani and S. Esmaili, Analyst, 2014, 139, 5531-
33
34 5537.
35
36
37 22. B. Li, N. J. Petersen, M. D. R. Payán, S. H. Hansen and S. Pedersen-Bjergaard, Talanta,
38
39 2014, 120, 224-229.
40
41 23. M. D. Ramos Payán, H. Jensen, N. J. Petersen, S. H. Hansen and S. Pedersen-Bjergaard,
42
43
44 Anal. Chim. Acta, 2012, 735, 46-53.
45
46 24. S. Seidi, Y. Yamini and M. Rezazadeh, J. Pharm. Biomed. Anal., 2011, 56, 859-866.
47
48 25. L. Arjomandi-Behzad, Y. Yamini and M. Rezazadeh, Anal. Biochem., 2013, 438, 136-143.
49
50
51 26. M. Rezazadeh, Y. Yamini and S. Seidi, J. Chromatogr. B, 2011, 879, 1143-1148.
52
53 27. M. Ghambarian, Y. Yamini and A. Esrafili, J. Chromatogr. A, 2012, 1222, 5-12.
54
55
56
57
58 16
59
60
Page 17 of 24 Analyst
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
28. T. M. Middelthon-Bruer, A. Gjelstad, K. E. Rasmussen and S. Pedersen-Bjergaard, J. Sep.
4
5
6 Sci., 2008, 31, 753-759.
7
8 29. Y. Yamini, S. Seidi and M. Rezazadeh, Anal. Chim. Acta, 2014, 814, 1-22.
9
10
11
30. M. Rezazadeh, Y. Yamini and S. Seidi, J. Sep. Sci., 2012, 35, 571-579.
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 31. A. Gjelstad, K. E. Rasmussen and S. Pedersen-Bjergaard, J. Chromatogr. A, 2007, 1174,


14
15 104-111.
16

Analyst Accepted Manuscript


17
18 32. A. C. Moffat, M. D. Osselton and B. Widdop, Clarke's analysis of drugs and poisons,
19
20 Pharmaceutical press London, 2011.
21
22 33. G. A. Shabir, JVT, 2005, 10, 314-325.
23
24
25
26 34. Guidance for Industry; Bioanalytical Method Validation, U.S. Department of Health and
27
28 Human Services, Food and Drug Administration, Rockville, 2001.
29
30
31
32
33
34
35 Figures captions
36
37
38 Fig. 1. Schematic illustration of consecutive on chip EME extraction cell with two extraction
39
40 chambers.
41
42
43
Fig. 2. Schematic illustration of consecutive on chip EME procedure.
44
45 Fig. 3. Optimization of (A) SLM composition, (B) applied voltage, (C) sample solution flow rate
46
47 and extraction time for Bet, Nalt and Nalm using consecutive on chip EME.
48
49
50 Fig. 4. Typical HPLC chromatograms obtained using consecutive on chip EME procedure for (A) a
51
52 urine sample and (B) a plasma sample after (a) and before (b) spiked with Bet, Nalt and Nalm at a
53
54 concentration level of 100 µg L-1.
55
56
57
58 17
59
60
Analyst Page 18 of 24
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
4 Table 1
5 Chemical structures, pKa and log P of the model analytes.
6
7 Name Chemical structure pKa Log P
8
9
10
11
12 Betaxolol 9.67 2.54
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13
14
15
16

Analyst Accepted Manuscript


17
18
19 Naltrexone 8.88 1.36
20
21
22
23
24
25
26
27
28
29 Nalmefene 9.57 1.95
30
31
32
33
34
35 Ref. [34]
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58 18
59
60
Page 19 of 24 Analyst
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
4 Table 2
5 Figures of merit of consecutive on chip EME for extraction of Bet, Nalt and Nalm.
6
7
LOD LOQ Linearity
8 Sample Analytes R2 PFa RSD%b
9 (µg L-1) (µg L-1) (µg L-1)
10
11 Urine Bet 3.0 10.0 10.0-500 0.9958 17.1 4.2
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 Nalt 3.0 10.0 10.0-500 0.9997 18.9 5.4


14
15 Nalm 2.0 5.0 10.0-500 0.9991 19.5 4.8
16

Analyst Accepted Manuscript


17
18
19 Plasma Bet 5.0 10.0 10.0-500 0.9932 15.6 5.4
20
21 Nalt 5.0 10.0 10.0-500 0.9974 17.5 5.6
22
23 Nalm 3.0 10.0 10.0-500 0.9942 18.9 6.8
24
a
25 Preconcentration factor at 100 µg L-1.
26 b
Based on four-replicate measurements.
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58 19
59
60
Analyst Page 20 of 24
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
4 Table 3
5 Determination of Bet, Nalt and Nalm in real samples using consecutive on chip EME.
6
7
Sample Analyte Creal Cadded Cfound RSD% Error%
8
9 (µg L-1) (µg L-1) (µg L-1) (n=4)
10
11 Urine Bet nda 100.0 96.8 4.6 -3.2
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13 Nalt nd 100.0 96.1 5.1 -3.9


14
15 Nalm nd 100.0 103.9 4.9 +3. 9
16

Analyst Accepted Manuscript


17
18
19
20
Plasma Bet nd 100.0 95.1 5.3 -4.9
21
22 Nalt nd 100.0 94.8 6.2 -5.2
23
24 Nalm nd 100.0 104.2 5.1 +4.2
25 a
26 Not detected
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58 20
59
60
Page 21 of 24 Analyst
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
4
5
6
7
8
9
10
11 Fig. 1
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13
14
15
16

Analyst Accepted Manuscript


17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58 21
59
60
Analyst Page 22 of 24
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
4 Fig. 2
5
6
7
8
9
10
11
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13
14
15
16

Analyst Accepted Manuscript


17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58 22
59
60
Page 23 of 24 Analyst
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
4 Fig. 3
5
6
7
8
9
10
11
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13
14
15
16

Analyst Accepted Manuscript


17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58 23
59
60
Analyst Page 24 of 24
View Article Online
DOI: 10.1039/C5AN02019K

1
2
3
4
5
6 Fig. 4
7
8
9
10
11
12
Published on 17 November 2015. Downloaded by Université Laval on 21/11/2015 07:30:22.

13
14
15
16

Analyst Accepted Manuscript


17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58 24
59
60

You might also like