CHE Notes Unit - 5-1
CHE Notes Unit - 5-1
A) UV-visible spectroscopy
Electromagnetic radiation: As light travels in the form of waves, when it is associated with electrical
and magnetic fields it is called as electromagnetic radiation.
(i) All the electromagnetic radiations travel at 3 × 10 8 m/sec in vacuum but the velocity decreases
when the wave interacts with electrons of matter.
(ii) The speed (V) of the wave is given by, V = wavelength × frequency
(iii) They posses properties of both wave and particle.
(iv) Energy in the photons of the EMR is given by, E = h − c / λ.
Fundamentals of Spectroscopy
Spectroscopy deals with interaction of electromagnetic radiation with matter. Due to these
interactions,
energy is absorbed or emitted by the matter. Measurement of radiation frequency indicates the
change in energy involved.
2) Emission spectroscopy: In this technique, sample is subjected to intense source of energy like
electric arc. As a result, sample is vaporized and electrons in the sample are excited to higher
energy state. When these electrons come back to ground state energy level, they emit radiations.
These radiations are analyzed by spectrophotometer to generate spectrum that gives information
about atoms present in the sample.
Like atoms, even molecules possess a set of discrete energy levels, which are of three types:
a) Electronic energy levels: At room temperature the molecules are in the lowest energy levels E 0.
When the molecules absorb UV-visible light from electromagnetic radiation, one of the outermost
electrons from pi - bond, sigma bond or a lone pair is promoted to higher electronic energy state.
b) Vibrational energy levels: Vibrational energy levels are of less energy than electronic energy
levels. The spacing between vibrational energy levels are relatively small. When infrared radiation is
absorbed molecules are excited from one vibrational level to another. Thus, during vibrational
excitation molecule begins to vibrate.
c) Rotational energy levels: The spacing between these levels is even smaller than vibrational energy
levels. The order of spacing of energy levels is as follows.
This wavelength at which there is a maximum absorption is called l max. This is a characteristic
property of molecule from sample.
Lambert’s Law
The rate of decrease in intensity of radiation is directly proportional to path length of solution, when
the monochromatic light passes through a solution of constant concentration.
Where, I0 = Intensity (radiant power) of the incident radiation
dx = Small thickness of solution or path length
Beer’s Law dI = Small decrease in intensity of light = I0 – It
The rate of decrease in intensity of light is proportional to concentration of solution for a fixed
thickness of solution.
3
By taking unit concentration of sample in solution (C = 1 mole/lit) and unit length of path (x = 1
cm), the absorbance observed will be the value of the constant in Lambert-beer’s equation.
A=∈·x·C
is known as molar absorptivity, or molar extinction coefficient constant.
Principle:
Following principles are involved in UV-visible spectrophotometry:
1. Beer’s law : The absorbance of the monochromatic light from UV-visible region is proportional
to concentration of the solution, for the constant path length.
2. The absorption of λmax from UV-visible region causes excitation of bonding electrons in a
covalent bond to higher energy antibonding orbitals e.g. σ σ*, π π*, non-bonding electrons
(lone pair) to π* etc.
(The electrons in a stronger covalent bond require higher energy for excitation which is available
from a shorter wavelength UV light and electrons in a weaker covalent bond require lower energy
for excitation which is available in visible light wavelengths.)
4
1) Light source (Source of radiation): The source of radiation should provide sufficient intensity over
the wavelength region. It has high intensity. The most common radiation sources used are hydrogen
discharge lamp, quartz halogen lamps, deuterium gas discharge tube, mercury arc.
3) Sample cell: It holds the sample either in the form of solution or as such. Quartz cuvettes are used
in UV spectroscopy but for visible spectroscopy glass cuvettes can be used.
4) Detector: Transmitted light from sample cell falls on the detector where it is converted into
electric current. It converts light energy directly into electrical energy. The output current of photo
detector is directly proportional to the intensity of light falling on it.
For UV spectroscopy: Photoelectric cell, photomultiplier tube, photovoltaic cell can be used as
photodetector.
5) Recorder: It is a display device, which automatically draws the spectrum. The signal from
monochromator is recorded onx-axis while the absorbance of the sample is displayed on y-axis. The
absorption of light (EMR) by a substance is governed by certain laws, namely, Beer's law and
Lambert's law.
Theory of Electronic Transitions: Energy absorbed from the ultraviolet –visible region causes
excitation of electrons to higher energy state.
1) σ → σ* transition : These transitions can occur in compounds, in which all the electrons are
involved in single (sigma) bonds and there are no lone pair of electrons on any atom in the molecule.
e.g. saturated hydrocarbons like ethane, methane. As the energy required for the transition is very
large, the absorption band occurs in the far ultraviolet region (100 to 135 nm).
Eg. C = 𝑁̈ , C = 𝑂̈ , 𝑁̈ = 𝑂̈ show this transition. In aldehydes, ketones, cyanides, -N=O, etc the
transition occurs due to excitation of electron from p-orbital (lone pair of electron) to antibonding π*
orbital. [at longer wavelength around 300 nm].
3) n → σ* transition : Saturated compounds with atoms having lone pair of electrons can undergo
this transition, and the bands are observed in the near UV region (150-225 nm). If this transition is
observed in the compound having π π* transition, then the λmax for this transition shifts to the
longer wavelength.
Alkenes generally absorb in the region 170 - 205 nm and are difficult to record by
spectrophotometer. However, molecules containing two or more conjugated double bonds absorb
above 200 nm and can be easily recorded.
2. Auxochrome: The functional groups attached to a chromophore which modifies the ability of
the chromophore to absorb light, altering the wavelength or intensity of absorption.
OR The functional group with non-bonding electrons that does not absorb radiation in near UV
region but when attached to a chromophore alters the wavelength & intensity of absorption.
6
3. Bathochromic Shift (Red Shift): When absorption maxima (λmax) of a compound shifts to
longer wavelength, it is known as bathochromic shift or red shift
• The effect is due to presence of an auxochrome or by the change of solvent. e.g. An auxochrome
group like –OH, -OCH3 causes absorption of compound at longer wavelength.
E.g.; In alkaline medium, p-nitrophenol shows red shift. Because negatively charged oxygen
delocalizes more effectively than the unshared pair of electron.
4. Hypsochromic Shift (Blue Shift): When absorption maxima (λmax) of a compound shifts to
shorter wavelength, it is known as hypsochromic shift or blue shift.
• The effect is due to presence of an group causes removal of conjugation or by the change of
solvent.
E.g; Aniline shows blue shift in acidic medium, due to loss of conjugation.
5. Structural information : From the λmax which is a characteristic value for different chromophores
and chromophore – auxochrome the structural parts in organic molecules, can be known.
6. Dissociation constants of weak acid or weak base : From the absorption of max by known conc.
of weak acid, ratio [A –] / [HA] can be calculated to find the Ka for weak acid.
Infrared Spectroscopy is the study of interaction of infrared light with a molecule. This can be analyzed
in three ways by measuring absorption, emission and reflection. It is used by chemists to determine
functional groups in molecules. IR Spectroscopy measures the vibrations of atoms, and based on this it
is possible to determine the functional groups.
Infrared radiation is largely thermal energy. It induces stronger molecular vibrations in covalent bonds.
Principle:
If a molecule absorbs IR radiation, it gets excited to higher vibrational energy level (vo → v1). The type
of IR radiation absorbed by molecules is unique and depends on nature of chemical bonds and functional
groups present. On absorption, stretching or bending type of vibrational changes occur in molecules
which provides information about the presence or absence of certain functional groups.
The atoms in molecules undergo a number of different vibrations. Vibrations fall into the two main
categories of A) stretching and B) bending.
A) Stretching:
In this type of vibrations, the bond length is increased or decreased at regular intervals. There are two
types of stretching vibrations - Symmetrical stretching and asymmetrical stretching.
i) Symmetric stretching: These involve relative movement of atoms in same plane and in same
direction
i.e. in symmetrical pattern
ii) Asymmetric stretching: These involve relative movement of atoms in same plane but in
opposite direction i.e. in symmetrical pattern
1
B) Bending:
In this type of vibrations, a change in bond angle occurs between bonds with a common atom, or there is
a movement of a group of atoms with respect to the remaining molecule without movement of the atoms
in the group with respect to one another.
The bending vibrations are also called as deformation vibrations. They are of two types.
a) In-plane Bending vibrations b) Out of plane Bending vibrations
a) In plane bending- In these types of vibrations, there is a change in bond angle. This type of
bendingtakes place within the same plane. In plane bending are of two types.
b) Out of plane bending- This type of bending takes plane outside of the plane of molecule.
i. Wagging: in this both atoms move to one side of the plane
ii. Twisting: in this one atom is above the plane and the other is below the plane
Vibration modes are simple to calculate for diatomic or triatomic molecules but are somewhat more
complex in polyatomic molecules.
Number of fundamental vibrations in the molecule can be calculated from its degree of freedom.
For a non-linear molecule, out of the 3N degrees of freedom, three describe rotation and three describe
translation and remaining (3N-6) degrees vibration degrees of freedom.
In a linear molecule, only two degrees of freedom are required to describe rotation and three describe
translation. Thus the remaining (3N-5) degrees are vibration degrees of freedom.
Energy of IR photon is insufficient to cause electronic excitation but can cause vibrational or
rotational excitations.
1) Dipole moment (Electric dipole): A bond or molecule must have a permanent dipole moment. If not,
then, some of its vibrations must produce an induced dipole moment in order to have an absorbance in
the IR spectrum.
A molecule can only absorb IR radiation when its absorption causes a change in dipole moment (electric
dipole). Such vibrations are said to be IR active.
When the molecule having dipole moment is kept in the beam of IR radiation, the field will exert forces
on the electric charges in the molecule. Opposite charges will experience force in opposite directions.
This tends to decrease in separation of atoms. Thus, dipole moment changes.
Symmetrical diatomic molecules like H2, Cl2, O2, N2 do not possess electrical dipole. They cannot be
excited by infrared radiations and thus do not give rise to IR spectra. Such molecules are IR inactive.
2) Correct frequency of radiation: A molecule absorbs radiations only when the natural frequency of
vibration of some part of a molecule (i.e. atom or group of atoms present in molecule) is same as the
frequency of incident radiation.
When frequency of vibration of a bond and frequency of IR radiation used for excitation, match
perfectly then only IR energy is absorbed.
Example: natural frequency of vibration of HCl molecule is about 8.7x1013 sec-1. When IR radiation is
passed through HCl sample and transmitted radiation is analysed by IR spectrophotometer, it is observed
that part of radiation which has frequency 8.7x1013 sec-1 is missing from transmitted radiation. Thus the
part of radiation with that frequency has been absorbed by HCl molecule and remaining frequencies of
radiation transmitted. Thus, the frequency 8.7x1013 sec-1 is characteristic of HCl molecule.
After absorbing the correct frequency of radiation the molecule vibrates at increased amplitude. This
occurs at the expense of the energy of IR radiation which has been absorbed.
3
Symmetrical molecules do not absorb IR radiation since they do not have dipole moment (O 2,
F2, H2, Cl2)
Diatomic molecules with dipole moment are IR-active (HCl, HF, CO, HI)
Molecules with more than two atoms may or may not be IR active depending on whether they
have permanent net dipole moment
4
c) Incandescent wire: It is a spiral wire of nichrome.
d) Mercury arc: High pressure mercury arc which consists of quatz tube containing mercury vapours can
be used.
2) Monochromator: Prisms or gratings are used as monochromators. They are used to select
radiation of desired frequencies. Prisms are normally made of alkali metal halides like NaCl, KBr.
3) Sample cells and sampling of substances : Infrared spectra can be obtained for solid, liquid or
gases (neat or in solution form). Material containing sample must be transparent to the IR
radiation, so the material made out of certain salts like NaCl, KBr, LiF etc. are used.
Sampling of liquids: Liquid sample can be sandwiched using sample cells of KBr or NaCl.
Sampling of gases: Sample cells made of KBr or NaCl can be used for the purpose.
a) Thermocouple: It uses two wires of different materials welded together. One of the two
welded joints is hot junction, becomes hotter because of IR radiations than the other joint.
A small electrical potential develops between the joints due to temperature difference. The
resulting current is directly proportional to the IR radiation falling on detector.
5) Amplifier and Recorder: Amplifier amplies the signals from detector and sends to recorder.
Sophisticated recorders give IR spectra of a compound as a graph of percentage transmission of IR
radiation versus wavelength or frequency or wave number (IR spectrum).
5
6
7
Parts of IR
Spectrum
The electromagnetic radiations having wavelength range 0.8 to 200 µ or wave number range 12800 to
50 cm-1 are infrared radiations.
Far infrared region infrared radiations.
The infrared region is further divided into three regions- Near infrared region, middle infrared region
and far infrared region.
Middle infrared region is further subdivided into functional group region, fingerprint region,
regioncharacteristic for aromatic compounds.
1. Functional group region: The region from 4000- 1500 cm-1 (2.5-6.5 µ) is called functional group
region. This part of IR spectrum contains absorption bands due to stretching vibrations of
functional groups like O - H, N - H, C -H. etc
The common functional groups show absorption bands in this region. The exact position of
absorption depends on force constant, masses of atoms, environment of bond.
8
9
1
0