MI 3002 EurotestDL MI 3102 EurotestXE UK HW3 Ver 1.1 20751468
MI 3002 EurotestDL MI 3102 EurotestXE UK HW3 Ver 1.1 20751468
MI 3002/ MI 3102
Instruction manual
Version 1.1, HW 3; Code No. 20 750 468
Distributor:
Manufacturer:
METREL d.d.
Ljubljanska cesta 77
1354 Horjul
Slovenia
Mark on your equipment certifies that this equipment meets the requirements of the EU
(European Union) concerning safety and interference causing equipment regulations.
No part of this publication may be reproduced or utilized in any form or by any means
without permission in writing from METREL.
2
MI 3002 & MI3102 Table of contents
1 Preface ......................................................................................................... 5
2 Safety and operational considerations ..................................................... 6
2.1 Warnings and notes .................................................................................... 6
2.2 Batteries ..................................................................................................... 9
2.3 Charging ..................................................................................................... 9
2.4 Precautions on charging of new battery cells or cells unused for a longer
period ....................................................................................................... 10
2.5 Standards applied ..................................................................................... 11
3 Instrument description ........................................................................... –12
3.1 Front panel ............................................................................................... 12
3.2 Connector panel ....................................................................................... 13
3.3 Back panel ................................................................................................ 14
3.4 Bottom view .............................................................................................. 15
3.5 Carrying the instrument ............................................................................ 16
4 Instrument operation ................................................................................ 17
4.1 Meaning of symbols and messages on the instrument display .......................... 17
4.1.1 The online voltage and output terminal monitor ................................... 17
4.1.2 Message field – battery status .............................................................. 18
4.1.3 Message field – measurement warnings/messages ............................. 18
4.1.4 Result field ........................................................................................... 19
4.1.5 Other messages ................................................................................... 19
4.1.6 Sound warnings.................................................................................... 20
4.1.7 Function and parameter line ................................................................. 20
4.2 Selecting measurement function/ subfunction .......................................... 20
4.3 Setting measurement parameters and limits ............................................ 21
4.4 Help menu ................................................................................................ 21
4.5 Setup menu .............................................................................................. 21
4.5.1 Supply system setup ............................................................................ 22
4.5.2 Impedance scaling factor adjustment ................................................... 22
4.5.3 Language selection .............................................................................. 23
4.5.4 Communication port selection .............................................................. 23
4.5.5 Plug / Tip commander support ............................................................. 23
4.5.6 Recalling original settings ..................................................................... 24
4.6 Display contrast adjustment...................................................................... 25
5 Measurements ........................................................................................... 26
5.1 Insulation resistance ................................................................................. 26
5.2 Continuity.................................................................................................. 28
5.3 Testing RCDs ........................................................................................... 31
5.3.1 Limit contact voltage............................................................................. 31
5.3.2 Nominal differential trip-out current ...................................................... 31
5.3.3 Multiplier of nominal residual current .................................................... 31
5.3.4 RCD type and test current starting polarity ........................................... 32
5.3.5 Testing selective (time-delayed) RCDs ................................................ 32
5.3.6 Contact voltage .................................................................................... 32
5.3.7 Trip-out time ......................................................................................... 35
5.3.8 Trip-out current ..................................................................................... 36
5.3.9 Autotest ................................................................................................ 37
5.4 Fault loop impedance and prospective fault current ................................. 41
3
MI 3002 & MI3102 Table of contents
4
MI 3002 & MI3102 Preface
1 Preface
Congratulations on your purchase of the Eurotest instrument and its accessories from
METREL. The instrument was designed on a basis of rich experience, acquired through
many years of dealing with electric installation test equipment.
The large graphic matrix display with backlight offers easy to read results, indications,
measurement parameters and messages. The operation of the instrument is designed
to be as simple and clear as possible and no special training (except for the reading this
instruction manual) is required in order to begin using the instrument.
To become more familiar with how to perform measurements in general and in typical
applications, we recommend reading Metrel handbook Measurements on electric
installations in theory and practice.
The instrument is equipped with all accessories required in order to perform testing
comfortably. The soft carrying bag, included with the meter, protect the instrument and
keep all accessories together making it simple and easy to move between locations.
5
MI 3002 & MI3102 Safety and operational considerations
6
MI 3002 & MI3102 Safety and operational considerations
Insulation resistance
Insulation resistance measurement should only be performed on de-energized
objects!
When measuring the insulation resistance between installation conductors, all
loads must be disconnected and all switches closed!
Do not touch the test object during the measurement or before it is fully
discharged! Risk of electric shock!
When an insulation resistance measurement has been performed on a capacitive
object, automatic discharge may not be done immediately! The warning message
and the actual voltage is displayed during discharge until voltage drops below
10 V.
Do not connect test terminals to external voltage higher than 550 V (AC or DC) in
order not to damage the test instrument!
Continuity functions
Continuity measurements should only be performed on de-energized objects!
Parallel impedances or transient currents may influence test results.
Testing PE terminal
If phase voltage is detected on the tested PE terminal, stop all measurements
immediately and ensure the cause of the fault is eliminated before proceeding
with any activity!
General
The indicator means that the selected measurement cannot be performed
because of irregular conditions on input terminals.
Insulation resistance, continuity functions and earth resistance measurements
can only be performed on de-energized objects.
PASS / FAIL indication is enabled when limit is set. Apply appropriate limit value
for evaluation of measurement results.
In the case that only two of the three wires are connected to the electrical
installation under test, only voltage indication between these two wires is valid.
Insulation resistance
If voltages of higher than 10 V (AC or DC) are detected between test terminals,
the insulation resistance measurement will not be performed.
Continuity functions
If voltages of higher than 10 V (AC or DC) are detected between test terminals,
the continuity resistance test will not be performed.
Before performing a continuity measurement, where necessary, compensate test
lead resistance.
7
MI 3002 & MI3102 Safety and operational considerations
RCD functions
Parameters set in one function are also kept for other RCD functions!
The measurement of contact voltage does not normally trip an RCD. However,
the trip limit of the RCD may be exceeded as a result of leakage current flowing
to the PE protective conductor or a capacitive connection between L and PE
conductors.
The RCD trip-lock sub-function (function selector switch in LOOP position) takes
longer to complete but offers much better accuracy of fault loop resistance (in
comparison to the R L sub-result in Contact voltage function).
RCD trip-out time and RCD trip-out current measurements will only be performed
if the contact voltage in the pre-test at nominal differential current is lower than
the set contact voltage limit!
The autotest sequence (RCD AUTO function) stops when trip-out time is out of
allowable time period.
Line impedance
The low limit prospective short-circuit current value depends on fuse type, fuse
current rating, fuse trip-out time and impedance scaling factor.
The specified accuracy of tested parameters is valid only if the mains voltage is
stable during the measurement.
8
MI 3002 & MI3102 Safety and operational considerations
2.2 Batteries
2.3 Charging
The batteries will begin charging whenever the power supply adapter is connected to
the instrument. The built-in protection circuits control the charging procedure and assure
maximum battery lifetime. The power supply socket polarity is shown in figure 2.1.
- +
Note:
Use only power supply adapter delivered from the manufacturer or distributor of
the test equipment to avoid possible fire or electric shock!
9
MI 3002 & MI3102 Safety and operational considerations
In this situation, Metrel recommend the following procedure to improve the battery
lifetime:
Completely charge the batteries for at least 14h via the instrument built-in
charger.
Completely discharge the batteries (this can be performed by using the
instrument normally until the instrument is fully discharged).
Repeat the charge/discharge cycle at least 2-4 times in order to restore the
batteries to their normal capacity..
Notes:
The charger in the instrument is a pack cell charger. This means that the cells
are connected in series during the charging so all of them must be in similar state
(similarly charged, same type and age).
If even one deteriorated battery cell (or just one of a different type e.g. capacity,
chemical design) can cause disrupted charging of the entire battery pack which
could lead to overheating of the battery pack and a significant decrease in the
operating time.
If no improvement is achieved after performing several charging/discharging
cycles, the state of each individual battery cells should be determined (by
comparing battery voltages, checking them in a cell charger, etc). It is very likely
that one or more of the battery cells could have deteriorated.
The effects described above should not be mixed with the normal battery
capacity decrease over time. All charging batteries lose some of their capacity
when repeatedly charged/discharged. The actual decrease in capacity compared
to the number of charging cycles depends on the battery type. This information
is normally provided in the technical specification from battery manufacturer.
10
MI 3002 & MI3102 Safety and operational considerations
11
MI 3002 & MI3102 Instrument operation
3 Instrument description
3.1 Front panel
3
6 7 1
Figure 3.1: Front panel
Legend:
1........ ON/OFF key, to switch the instrument on and off.
The instrument will automatically switch off 10 minutes after the last key press /
function switch rotation.
2........ Function selector switch.
3........ MEM key, to access memory operations.
4 ........ HELP/CAL key, to access help menus. CAL function can be used to
compensate for the test lead resistance in low-value resistance
measurements.
5........ Jogger keypad with cursors for navigating screen menus and TEST keys for
starting tests.
The TEST key also acts as the PE touching electrode which detects potentially
dangerous voltages appearing on the PE conductor.
6……. BACKLIGHT and CONTRAST key. This is used to turn the backlight on and off
and for adjusting the level of the contrast.
The high level backlight is automatically shut off 20 seconds after the last key
press / function switch rotation in order to extend service life of the battery.
7........ 128 64 dots matrix display with backlight.
12
MI 3002 & MI3102 Instrument description
> 550V
5
6
Figure 3.2: Connector panel
Legend:
1........ Test connector.
Warning! Maximum allowed voltage between test terminals and ground is
600V! Maximal allowed voltage between test terminals is 550 V!
MI 3102 only: In the earth resistance function, the test connector terminals are
used as follows:
N/L2 blue test lead is used for the earth electrode (E).
L/L1 black test lead is used for the auxiliary earth electrode (H).
PE/L3 green test lead is used for the probe (S).
13
MI 3002 & MI3102 Instrument description
1 2
Legend:
1........ Battery/fuse compartment cover.
2........ Information label.
3........ Fixing screws for battery/fuse compartment cover.
2
1
3
Fuse Fuse
F2 F1
F3
Fuse S/N XXXXXXXX
4 6
5
14
MI 3002 & MI3102 Instrument description
Legend:
1........ Fuse F1.
2........ Fuse F2.
3........ Fuse F3.
4........ Serial number label.
5........ Battery cells (size AA).
6........ Battery holder.
3
UC: 0.00V 100.0V
Test current: min. ±200mA RS : 0.00 10.00k(RS =UC / IN)
Open-circuit voltage: 5.4VDC 9.0VDC
Continuity 7mA Fault loop resistance (without tripping RCD)
RS : 0.00 10k
R: 0.0 1999
Test current: max. 7mA DC UC: 0.00V 100.0V, (UC=RS I N)
Open-circuit voltage: 5.4VDC 7.2VDC
Tripping time
Insulation resistance (EN 61557-2) non-delayed (time-delayed) RCDs
R: 0.000M 199.9M , U N=100V DC, 250V DC t: 0ms 300ms (500ms)
R: 0.000M 999M , U N= 500V DC, 1kV DC t: 0ms 150ms (200ms)
U: 0V 1200V, t: 0ms 40ms (150ms)
Nominal voltages: 100V DC, 250V DC, 500V DC, 1kV DC UC: 0.00V 100.0V
Measuring current: min. 1mA at R N=U N 1k /V Tripping current
Short-circuit current: <3mA DC I: 0.2 IN 1.1 IN
Line resistance t: 0ms 300ms
R L-N (L): 0.00 1999 UC: 0.00V 100.0V
IPSC: 0.06A 24.4kA
IN : 10mA, 30mA, 100mA, 300mA, 500mA, 1A
Nominal voltage: 100V AC 440V AC / 45Hz 65Hz
Multiplier: 1/2 1, 2, 5
Fault loop resistance (EN 61557-3) Nominal voltage: 100VAC 264VAC / 45Hz 65Hz
R L-PE: 0.00 1999
IPFC: 0.06A 24.4kA Phase rotation (EN 61557-7)
Nominal voltage: 100V AC 440V AC / 45Hz 65Hz Nominal voltage: 100V AC 440V AC / 45Hz 65Hz
Voltage, frequency Results: 1.2.3. or 2.1.3
U: 0V 440V
f: 45Hz 65Hz
Legend:
1........ Information label.
2........ Neck strap openings.
3........ Handling side covers.
15
MI 3002 & MI3102 Instrument description
16
MI 3002 & MI3102 Instrument operation
4 Instrument operation
4.1 Meaning of symbols and messages on the instrument display
The instrument display is divided into four sections:
1
4 3
Legend:
1........ Function and parameter line.
In the top line of the display, the measuring function/sub-function and
parameters are displayed.
2........ Result field.
In this field the main result and sub-results together with the
PASS/FAIL/ABORT status are displayed.
3........ Online voltage and output monitor.
4........ Message field.
L – N polarity changed.
17
MI 3002 & MI3102 Instrument operation
18
MI 3002 & MI3102 Instrument operation
Measurement passed.
Measurement failed.
19
MI 3002 & MI3102 Instrument operation
1
2 3
Legend:
1……Main function name.
2……Function or sub-function name.
3……Measuring parameters and limit values.
20
MI 3002 & MI3102 Instrument operation
Once the measurement parameters are set, the settings are retained until new changes
are made or the original settings are recalled.
To enter the Setup menu press the BACKLIGHT key and rotate function selector switch
in any direction at the same time.
Rotate function selector switch again to leave Setup menu or setup sub-menus.
21
MI 3002 & MI3102 Instrument operation
Select SYSTEMS in the Setup menu by using and keys and press the TEST key
to enter the Supply system setup menu.
By using and keys select supply system and press the TEST key to accept the
setting.
Use the and keys to adjust the scaling factor and press the TEST key to accept the
new setting.
The impedance limit values for different overcurrent protective devices are scaled down
by a factor 0,8 or 0,75 (Z factor). This means that the fault current will still be high
enough also at increased conductor temperatures and low supply voltage. This assures
a safe operation of the overcurrent protection device in all conditions.
22
MI 3002 & MI3102 Instrument operation
By using and keys select the language you want to use and press The TEST key
to accept the new setting.
Use the and keys in the Setup menu to highlight the SET COMMUNICATION
PORT option and press the TEST key to enter the Communication menu.
By using the and keys, select communication port you want to use. If the RS232
communication port is selected, use and keys to select baud rate (The USB port
has a baud rate set to 115200 bps). Press the TEST key to accept the new setting.
Note:
Only one port can be active at one time.
If disable is selected, then keys on the Plug / Tip commander are disabled (except
backlight key). This implies that the selected test can only be started (or results can be
saved) using the keys on the instrument.
If enable is selected, the keys on Plug / Tip commander are activated and tests can be
started and results saved either from the buttons on the instrument or the buttons on the
remote commanders.
Note:
This option is intended to disable the commander’s remote keys. In the case of
high EM interfering noise the operation of the commander’s key can be irregular.
23
MI 3002 & MI3102 Instrument operation
In order to restore the original setting of the instrument, press and hold key and
switch on the instrument (the external charger must not be connected). »Hard reset«
message will be displayed at start-up.
Instrument settings, measurement parameters and limits are set to their initial values as
follows:
Instrument settings Default value
Contrast 50 %
Impedance scaling factor 0.8
Supply system TN/TT
Communication RS232, 9600 bps
Commander Enabled
Function
Parameter / limit value
Sub-function
CONTINUITY
r1, r2,rN, R1+R2, R2 High limit resistance value: 2.0
INSULATION
ISO L/N, ISO L/E, ISO N/E, ISO L/L Nominal test voltage: 500 V
Low limit resistance value: 1 M
LINE Fuse type: none selected (F)
Fuse current rating: none selected (A)
Fuse tripping current: none selected (ms)
LOOP Fuse type: none selected (F)
Zs Fuse current rating: none selected (A)
Zs (rcd) Fuse tripping current: none selected (ms)
RCD Selected function: RCD Uc
Contact voltage – RCD Uc Nominal differential current: I N =30 mA
Trip-out time – RCD t RCD type and test current starting polarity:
Trip-out current – RCD III G
Autotest – RCD AUTO Limit contact voltage: 50 V
Nominal differential current multiplier: 1
RESISTANCE TO EARTH High limit resistance value: 50
(MI 3102 only)
ILLUMINATION (MI 3102 only) Low limit illumination value: 300 lux
TRMS CURRENT (MI 3102 only) Current limit: 4.5 mA
24
MI 3002 & MI3102 Instrument operation
Use the and keys to adjust the contrast level and, when satisfied, press the TEST
key to accept the new setting.
25
MI 3002 & MI3102 Measurements
5 Measurements
5.1 Insulation resistance
The Insulation resistance measurement is performed in order to ensure safety against
electric shock. Using this measurement the following items can be determined:
Insulation resistance between installation conductors,
Insulation resistance of non-conductive rooms (walls and floors),
Insulation resistance of ground cables,
Resistance of semi-conductive (antistatic) floors.
The insulation resistance tests are carried out in the same way regardless which sub-
functions is selected. However it is important to select the appropriate sub-function in
order to classify the measurement to be correctly considered in verification documents
(Electrical Installation Certificate, Periodic Inspection Report etc.).
Step 1 Select Insulation function with the function selector switch and sub-function
with the / keys .The following menu is displayed:
Step 3 Ensure that no voltages are present on the item for testing. Connect the test
leads to the Eurotest instrument, Connect the test cables to the item under
test. (see figure 5.2) to perform insulation resistance measurement. Use the
Help function if necessary for further reference.
26
MI 3002 & MI3102 Measurements
Step 4 Check the displayed warnings and online voltage/terminal monitor before
starting the measurement. If OK, press and hold the TEST key until the result
has stabilised. Actual measured results are shown on the display during
measurement.
After the TEST key is released the last measured results are displayed,
together with the PASS/FAIL indication (if applicable).
Displayed results:
R .............Insulation resistance,
Um..........Actual voltage applied to item under test
Refer to chapter 6.1. Saving results for information on how to save the
displayed results for documentation purposes.
Warnings:
Insulation resistance measurement should only be performed on de-energized
objects!
When measuring the insulation resistance between installation conductors, all
loads must be disconnected and all switches closed!
Do not touch the test object during the measurement or before it is fully
discharged! Risk of electric shock!
When an insulation resistance measurement has been performed on a capacitive
object, an automatic discharge may not be done immediately! The warning
message and actual voltage is displayed during discharge until the voltage of
the item under test drops below 10 V.
In order to prevent damaging the test instrument, do not connect test terminals to
an external voltage higher than 600 V (AC or DC)
Note:
If a voltage higher than 10 V (AC or DC) appears between the test terminals, the
insulation resistance measurement will not be performed.
27
MI 3002 & MI3102 Measurements
5.2 Continuity
This function is used to test the resistance between two different points of the
installation to ensure that a conductive path exists between them. The test ensures that
all protective conductors, earth conductors or bonding conductors are correctly
connected, terminated and have the correct resistive value.
The Continuity measurement is performed with a test current of more than 200mA. An
automatic pole reversal of the test voltage and the test current is performed during the
test. This test checks for any components (e.g. diodes, transistors, SCRs) that may
have a rectifying effect on the circuit which could cause problems when a voltage is
applied.
Step 1 Select the Continuity function with the function selector switch and the sub-
function with the / keys. The following menus will be displayed:
Step 3 Connect test cable to the Eurotest instrument. Before performing a Continuity
measurement, compensate for the test leads resistance as follows:
28
MI 3002 & MI3102 Measurements
N/L2
PE/L3 N/L2
PE/L3
L/L1 L/L1
prolongation lead
The compensation of the test leads can be carried out in any of the Continuity
sub-functions (r1, r2, rN, R1+R2, R2). The compensation value will then be
transferred to all other subfunctions.
Step 4 Ensure that the item for testing is disconnected from any voltage source.
Connect the test cables to the item under test. Follow the connection
diagrams shown in figures 5.6 and 5.8 to perform a Continuity measurement.
Use the Help function if necessary.
r1 rN r2
Figure 5.6: Connection for testing the r1, rN, and r2 sections of the wiring
29
MI 3002 & MI3102 Measurements
Figure 5.7: Connection for testing the R2 section of the wiring and bonding connections
Figure 5.8: Connection for testing the R1+R2 section of the wiring
Step 5 Check for any warnings and the online voltage/terminal monitor on the
display before starting the measurement. If everything is OK, press the TEST
key. After performing the measurement, the results appear on the display
together with the PASS/FAIL indication (if applicable).
Displayed results:
R .............Main Continuity result (average of R+ and R- results),
30
MI 3002 & MI3102 Measurements
Refer to chapter 6.1. Saving results for information on how to save the
displayed results for documentation purposes.
Warnings:
Low-value resistance measurements should only be performed on de-energized
objects!
Parallel impedances or transient currents may influence test results.
Note:
If voltage between test terminals is higher than 10 V the measurement will not be
performed.
31
MI 3002 & MI3102 Measurements
32
MI 3002 & MI3102 Measurements
Step 1 Select RCD function with the function selector switch first. Use the / keys to
select contact voltage function (Uc). The following menu will be displayed:
Step 3 Connect the test leads to the instrument and follow the connection diagram
shown in figure 5.12 to perform contact voltage measurement. Use the Help
function if necessary.
L1
L2
L3
N
PE
L/L
PE/L3 1
N/L2
N PE L
Ro RE
Step 4 Check for any warnings and check the online voltage/terminal monitor on the
display before starting the measurement. If everything is ok, press the TEST
key. After performing the measurement, the results will be displayed along
with a PASS/FAIL indication.
Displayed results:
U .............Contact voltage.
Rl ............Fault loop resistance.
33
MI 3002 & MI3102 Measurements
Refer to chapter 6.1. Saving results for information on how to save the
displayed results for documentation purposes.
Notes:
Parameters set in this function are also kept for all other RCD functions!
The measurement of contact voltage does not normally trip an RCD. However,
the trip limit may be exceeded as a result of leakage currents flowing through the
PE protective conductor or a capacitive connection between the L and PE
conductor.
RCD trip-lock sub-function (function selector switch in LOOP position) takes
longer to complete but offers much better accuracy of a fault loop resistance
result (in comparison with the R L sub-result in Contact voltage function).
34
MI 3002 & MI3102 Measurements
Step 1 Select the RCD function with the function selector switch and use the /
keys to select the Trip-out time function (RCDt). The following menu will be
displayed:
Step 3 Connect the leads to the instrument and follow the connection diagram shown
in figure 5.12 (see the chapter 5.3.6 Contact voltage) to perform trip-out time
measurement.
Step 4 Check for any warnings and check the online voltage/terminal monitor on the
display before starting measurement. If everything is ok, press the TEST key.
35
MI 3002 & MI3102 Measurements
After performing the measurement, results will appear on the display along
with a PASS/FAIL indication.
Displayed results:
t ..............Trip-out time,
U C ...........Contact voltage.
Notes:
Parameters set in this function are also transferred onto all other RCD functions!
RCD trip-out time measurement will be performed only if the contact voltage at
nominal differential current is lower than the limit set in the contact voltage
setting!
The measurement of the contact voltage in pre-test does not normally trip an
RCD. However, the trip limit may be exceeded as a result of leakage current
flowing through the PE protective conductor or a capacitive connection between
L and PE conductors.
Step 1 Select RCD function with the function selector and use the / keys to select
the Trip-out current (RCD ) function. The following menu is displayed:
36
MI 3002 & MI3102 Measurements
Step 2 By using cursor keys the following parameters can be set in this
measurement:
Nominal residual current,
RCD type,
Test current starting polarity.
Step 3 Connect the test leads to the instrument and follow the connection diagram
shown in figure 5.12 (see the chapter 5.3.6 Contact voltage) to perform trip-
out current measurements. Use the Help function if necessary.
Step 4 Check for any warnings and check the online voltage/terminal monitor shown
on the display before starting the measurement. If everything is ok, press the
TEST key. After performing the measurement, the results will be displayed
along with a PASS/FAIL indication.
Displayed results:
I .............Trip-out current,
U Ci ..........Contact voltage,
tI .............Trip-out time.
Refer to chapter 6.1. Saving results for information on how to save the
displayed results for documentation purposes.
Notes:
Parameters set in this function are also kept for other RCD functions!
RCD trip-out current measurement will be performed only if the contact voltage at
nominal differential current is lower than set limit contact voltage!
The measurement of contact voltage in the pre-test does not normally trip an
RCD. However, the trip limit may be exceeded as a result of leakage current
flowing through the PE protective conductor or a capacitive connection between
L and PE conductors.
5.3.9 Autotest
The purpose of the autotest function is to perform a complete RCD testing and
measurement of most important associated parameters (contact voltage, fault loop
resistance and trip-out time at different fault currents) with one press of a button. If a
faulty parameter is noticed during the autotest, the test will stop to highlight the need for
further investigation.
Notes:
The measurement of contact voltage in the pre-test does not normally trip an RCD.
However, the trip limit may be exceeded as a result of leakage current flowing
37
MI 3002 & MI3102 Measurements
Step 1 Select RCD function with the function selector switch first. Use the / keys
to select RCD autotest function (AUTO). The following menu is displayed:
Step 3 Connect the test leads to the instrument and follow the connection diagram
shown in figure 5.12 (also see the chapter 5.3.6 Contact voltage) to perform
the RCD autotest. Use the Help function if necessary.
Step 4 Check for any warnings and check the online voltage/terminal monitor
displayed on the screen before starting the measurement. If everything is ok,
press the TEST key. The autotest sequence will then start to run as follows:
38
MI 3002 & MI3102 Measurements
39
MI 3002 & MI3102 Measurements
Displayed results:
x1 (left) ...... Step 1 trip-out time result, t3 (I N , 00),
x1 (right) .... Step 2 trip-out time result, t4 (I N , 1800),
x5 (left) ...... Step 3 trip-out time result, t5 (5I N , 00),
x5 (right) .... Step 4 trip-out time result, t6 (5I N , 1800),
x½ (left) ..... Step 5 trip-out time result, t1 (½I N , 00),
x½ (right) ... Step 6 trip-out time result, t2 (½I N , 1800),
Uc ...........Contact voltage.
Refer to chapter 6.1. Saving results for information on how to save the
displayed results for documentation purposes.
Note:
In case of testing the RCD type A with rated residual currents of In = 300 mA,
500 mA, and 1000 mA, the x5 Auto tests will be automatically skipped. In this
case, the auto test result passes if the t1 to t4 results pass, and on the display
are omitted t5 and t6.
40
MI 3002 & MI3102 Measurements
Where:
Nominal input voltage U N Voltage range
115 V (100 V U L-PE 160 V)
230 V (160 V U L-PE 264 V)
Step 1 Select the LOOP function with the function selector switch and use the /
keys to select the Zs sub-function. The following menu is displayed:
41
MI 3002 & MI3102 Measurements
Step 3 Connect the test leads to the instrument and follow the connection diagram
shown in the figure 5.26 to perform fault loop impedance measurement. Use
the Help function if necessary.
L1
L2
L3
N
PE
L/L
PE/L3 1
N/L2
N PE L
Ro RE
Step 4 Check for any warnings displayed on the screen and check the online
voltage/terminal monitor before starting the measurement. If everything is ok,
press the TEST key. After performing the measurement, the test results will
appear on the display together with the PASS/FAIL indication (if applicable).
Displayed results:
Z .............Fault loop impedance,
I SC ...........Prospective fault current (displayed in amps),
Lim .........High limit fault loop impedance value (if applicable).
Refer to chapter 6.1. Saving results for information on how to save the
displayed results for documentation purposes.
Notes:
The fault loop impedance limit depends on the fuse type, fuse current rating, fuse
trip-out time and Impedance scaling factor.
The specified accuracy of test parameters is valid only if mains voltage is stable
during the measurement.
The Zs Fault loop impedance measurement trips RCD protected circuits.
42
MI 3002 & MI3102 Measurements
5.4.2 The fault loop impedance test for RCD protected circuits
The fault loop impedance is measured with a low test current to avoid tripping the RCD.
This function can also be used for fault loop impedance measurement in system
equipped with RCDs which have a rated trip-out current of 10 mA and above.
Where:
Nominal input voltage U N Voltage range
115 V (100 V U L-PE 160 V),
230 V (160 V U L-PE 264 V),
Step 1 Select the LOOP function with the function selector switch and use the /
keys to select Zsrcd sub-function. The following menu is displayed:
Step 3 Connect the appropriate test leads to the instrument and follow the
connection diagram shown in figure 5.12 to perform RCD trip-lock
measurement (see chapter 5.3.6 Contact voltage). Use Help function if
necessary.
43
MI 3002 & MI3102 Measurements
Step 4 Check for warnings on the display and check the online voltage/terminal
monitor before starting the measurement. If everything is ok, press the TEST
key. After performing the measurement the results will appear on the display.
Figure 5.29: Example of fault loop impedance measurement results using trip-lock
function
Displayed result:
Z .............Fault loop impedance,
I SC ...........Prospective fault current,
Lim .........High limit fault loop impedance value (if applicable).
Refer to chapter 6.1. Saving results for information on how to save the
displayed results for documentation purposes.
Notes:
The measurement of fault loop impedance using trip-lock function does not
normally trip an RCD. However, if the trip limit may be exceeded as a result of
leakage current flowing through the PE protective conductor or a capacitive
connection between L and PE conductors.
The specified accuracy of test parameter is valid only if mains voltage is stable
during the measurement.
44
MI 3002 & MI3102 Measurements
Where:
Nominal input voltage U N Voltage range
115 V (100 V U L-PE 160 V)
230 V (160 V U L-PE 264 V)
400 V (264 V U L-PE 440 V)
Step 1 Select the LINE function with function selector switch. The following menu is
displayed:
Step 3 Connect the appropriate test leads to the instrument and follow the
connection diagram shown in figure 5.31 to perform phase-neutral or phase-
phase line impedance measurement. Use the Help function if necessary.
45
MI 3002 & MI3102 Measurements
L1
L2
L3
N
PE
N/L2
PE/L3
L/L1
L/L
PE/L3 1
N/L2
N PE L
Ro RE
Step 4 Check for warnings displayed on the screen and check the online
voltage/terminal monitor before starting the measurement. If everything is ok,
press the TEST key. After performing the measurement, the results will
appear on the display together with the PASS/FAIL indication (if applicable).
Displayed results:
Z .............Line impedance,
I SC ...........Prospective short-circuit current,
Lim .........High limit fault loop impedance value (if applicable).
Refer to chapter 6.1. Saving results for information on how to save the
displayed results for documentation purposes.
Notes:
The fault loop impedance limit depends on fuse type, fuse current rating, fuse
trip-out time and Impedance scaling factor.
The specified accuracy of the test parameter is valid only if mains voltage is
stable during the measurement.
46
MI 3002 & MI3102 Measurements
Step 1 Select the PHASE ROTATION function ( ) with the function selector switch.
The following menu is displayed.
Step 2 Connect test cable to the Eurotest instrument and follow the connection
diagram shown in figure 5.34 to test phase sequence.
L3
L2
L1
N
PE
PE/L3
PE/L3
N/L2
N/L2
L/L1
L/L1
option A 1110
result 1.2.3 result 2.1.3
Figure 5.34: Connection of universal test cable and optional three phase cable
Step 3 Check for warnings on the display and check the online voltage/terminal
monitor. The phase sequence test is a continuously running test hence the
results will be displayed as soon as the full test lead connection to the item
under test has been made.. All three-phase voltages are displayed in order of
their sequence represented by the numbers 1, 2 and 3.
47
MI 3002 & MI3102 Measurements
Displayed results:
Ph ...........Phase sequence,
1.2.3........Correct connection,
2.3.1........Invalid connection,
-.-.- ..........Irregular voltages.
Refer to chapter 6.1. Saving results for information on how to save the
displayed results for documentation purposes.
Step 1 Select the VOLTAGE function (Volt.) with the function selector switch. The
following menu is displayed:
Step 2 Connect test cable to the Eurotest instrument and follow the
connection diagram shown in figure 5.37 to perform a voltage and frequency
measurement.
L1
L2
L3
N
PE
N/L2
PE/L3
L/L1
L/L
PE/L3 1
N/L2
N PE L
Ro RE
48
MI 3002 & MI3102 Measurements
Step 3 Check the displayed warnings. The Voltage and Frequency test continually
runs, showing fluctuantions as they occur, these results are shown on the
display during measurement.
Displayed results:
Ul-n.........Voltage between phase and neutral conductors,
Ul-pe.......Voltage between phase and protective conductors,
Un-pe .....Voltage between neutral and protective conductors.
Refer to chapter 6.1. Saving results for information on how to save the
displayed results for documentation purposes.
49
MI 3002 & MI3102 Measurements
Step 1 Select the EARTH function with function selector switch. The following menu
is displayed:
Step 3 Connect the leads to the instrument and follow the connection diagram shown
in figure 5.40 to perform a resistance to earth measurement. Use the Help
function if necessary.
50
MI 3002 & MI3102 Measurements
E
H S
MPEC
Rc Rp RE
>5d d
Step 4 Check for warnings on the display and check the online voltage/terminal
monitor before starting the measurement. If everything is ok, press the TEST
key. After performing the measurement, the results will appear on the display
together with the PASS/FAIL indication (if applicable).
Displayed results:
R .............resistance to earth,
R C ...........auxiliary earth electrode resistance,
R P ...........probe resistance.
Refer to chapter 6.1. Saving results for information on how to save the
displayed results for documentation purposes.
Notes:
If a voltage higher than 30 V is present between the test terminals, the test will be
prevented from being performed.
If a noise voltage higher than approx. 5 V is present between the H and E or S test
terminals, the (noise) warning symbol “ ” will be displayed, indicating that the
test result may not be correct!
51
MI 3002 & MI3102 Measurements
Step 1 Select the TRMS CURRENT function (TRMS) with the function selector
switch. The following menu is displayed:
Step 3 Connect the test leads to the Eurotest instrument and follow the connection
diagram shown in Figure 5.43 to perform a TRMS current measurement. Use
the Help function if necessary.
L1
L2
L3
N
PE
IL
Option A 1018
I
c c
52
MI 3002 & MI3102 Measurements
Step 4 Check for any warnings on the display before starting the measurement. If
everything is ok, press the TEST key.
After starting the test, the test will run continuously until the user stops the
measurement. To stop measurement, press the TEST key again. The last
measured result will be displayed together with the PASS/FAIL indication (if
applicable).
Displayed results:
I ..............TRMS current (or TRMS leakage current).
Refer to chapter 6.1. Saving results for information on how to save the
displayed results for documentation purposes.
Notes:
Use test clamp supplied by METREL or another clamp with similar characteristics
(current output, 1000:1 clamp ratio, appropriate measuring range, consider the
error of the test clamp when evaluating the measured results)!
Current clamps Metrel A 1074 and A 1019 are suitable for use with MI 3102
EurotestXE instrument in range 0.2 A 200 A. Below 0.2 A they can be used as
indicator only. They are not suitable for leakage current measurements.
53
MI 3002 & MI3102 Measurements
Step 1 Select the SENSOR function with the function selector switch. The following
menu is displayed:
Step 3 Connect LUXmeter probe to the instrument RS232 connector and follow the
positioning diagram shown in figure 5.46 to perform illumination
measurement. Turn on the LUXmeter probe by pressing its ON/OFF key. (the
Green LED should light on the probe). Use the Help function if necessary.
54
MI 3002 & MI3102 Measurements
Step 4 Check for any warnings on the display before starting the measurement. If
everything is ok, press the TEST key to start the measurement. The results
will appear on the display along with a PASS/FAIL indication (if applicable).
After starting the test, the test will run continuously until the user stops the
measurement. To stop measurement, press the TEST key again The last
measured result will be displayed, together with the PASS/FAIL indication (if
applicable).
Displayed result:
E .............Illumination
Refer to chapter 6.1. Saving results for information on how to save the
displayed results for documentation purposes.
Notes:
For accurate measurements, make sure that the milk glass bulb is lit without any
shadows cast by hand, body or any other unwanted objects.
It is very important to know when the artificial light sources will reach full power (this
may take a period of time see technical data for light sources for more information)
and should therefore be switched on for this period of time before any
measurements are taken.
55
MI 3002 & MI3102 Measurements
Step 2 Follow the connection diagrams shown in figures 5.48 and 5.49 to test PE
terminal.
L1
N
PE
Figure 5.48: Connection of plug cable to mains outlet with reversed L and PE
conductors
56
MI 3002 & MI3102 Measurements
L1
N
PE
Reversed phase and
protection conductors!
L/
PE/L3 L1 MOST DANGEROUS
N/L2 SITUATION!
N PE L
Figure 5.49: Connection of universal test cable to load connection terminals with
reversed L and PE conductors
Step 3 Touch the PE test probe (TEST key) for a few seconds. If PE terminal is
connected to phase voltage a warning message will be displayed and
instrument buzzer will sound.
Warning:
If phase voltage is detected on tested PE terminal, stop all measurements
immediately and remove the cause of the fault before proceeding with any
activity!
Notes:
The PE terminal can only be tested with the function selector set to RCD, LOOP
or LINE!
For correct testing of PE terminals, the TEST key has to be touched for a few
seconds.
When testing, make sure the instrument operated is stood on non-isolated floor
(otherwise the test result may be wrong!)
57
MI 3002 & MI3102 Working with results
EARTH ELECTRODE
BONDING Water
BONDING Oil
BONDING Lightning
BONDING Gas
BONDING Structural Steel
BONDING Other
BONDING Earthing conductor
EARTH ELECTRODE
BONDING Water
BONDING Oil
BONDING Lightning
BONDING Gas
BONDING Structural Steel
BONDING Other
BONDING Earthing conductor
EARTH ELECTRODE
BONDING Water
BONDING Oil
BONDING Lightning
BONDING Gas
BONDING Structural Steel
BONDING Other
BONDING Earthing conductor
OBJECT 002
...
OBJECT 999
The memory structure is organized in a way that is suited to the BS 7671 and other
verification documents (Electrical Installation Certificate, Periodic Inspection Report
58
MI 3002 & MI3102 Working with results
etc.). If the results are stored in appropriate memory locations the automatic creation of
a certificate/ report is possible via the PC software.
Step 1 When the measurement has completed, press the MEM key. The SAVE
results menu is displayed with the last used memory location selected.
Step 2 The results can be saved into the selected memory location as follows:
By using the / keys, position the cursor on the Object line. Use the /
keys to select appropriate ‘Object’ memory location (three-digit number ).
By using the / keys, position the cursor on the D. Board line. Use the /
keys to select appropriate ‘Distribution block’ memory location (three-digit
number).
By using the / keys, position the cursor on the Circuit / Earth Electr. /
Bonding line.
Choose between the various options (‘Circuit’, ‘Earth Electrode’ or ‘Bonding’)
by pressing the TEST key.
If Circuit is selected, use the / keys to select the appropriate ‘Circuit’
memory location (three-digit number).
If Bonding is selected, use the / keys to select the appropriate ‘Bonding’
memory location (Water/ Oil/ Lightning/ Gas/ Str.steel/ Earth.cond).
Step 3 After the memory location is set, press the MEM key to save the results.
»Saved to memory« message will displayed to confirm that the results have
been saved. After saving the results the instrument returns to normal test
screen.
Note:
Each measurement result can be saved only once.
59
MI 3002 & MI3102 Working with results
Step 1 Select Recall results from the Memory menu by using the and keys and
press the TEST key to confirm. The Recall results menu will be displayed and
the last used memory location will be selected.
Step 2 To recall a saved result the memory location must be selected first:
The Object line can be selected with the / keys. Use the / keys to
select the appropriate ‘Object ‘memory location (three-digit number).
In the lower part of the display, two numbers will be displayed (e.g. No.:2/16
see figure 6.4):
- the first number is the number of measurements stored in the selected
‘Object’ memory location.
- the second number is the number of measurements stored in the selected
‘Object’ memory location and its sub-locations (i.e. D.board, Circuit, Earth
Electrode, Bonding memory locations).
The D. Board line can be selected with the / keys. Use the / keys to
select appropriate ‘D. Board memory location (three-digit number).
In the lower part of the display, two numbers will be displayed (e.g. 6/16 see
figure 6.4):
- the first number is the number of measurements stored in the selected ‘D.
Board’ memory location.
- the second number is the number of measurements stored in the selected
‘D. Board’ memory location and its sub-locations (Circuit, Electrode,
Bonding memory locations).
60
MI 3002 & MI3102 Working with results
The Circuit / Earth Electrode / Bonding line can be selected with the /
keys.
The appropriate memory location (‘Circuit’ or ‘Earth Electrode’or ‘Bonding’)
can be selected with the TEST key.
If the Circuit is selected, use the / keys to select the appropriate ‘Circuit’
memory location.(three-digit number). In the lower part of the display (No.
line) the number of measurements stored in the selected ‘Circuit’ memory
location is shown.
If the Bonding is selected, use the / keys to select appropriate ‘Bonding’
memory location (e.g. Water/ Oil/ Lightning/ Gas/ Str.Steel/ Earth.cond.). In
the lower part of the display the number of measurements stored in the
selected ‘Bonding’ memory location is shown.
If the Earth Electrode is selected, the number of measurements stored in the
selected ‘Electrode’ memory location will be displayed in the lower part of the
display.
Step 3 Once the memory location is set, press the MEM key and the cursor will jump
down to the No. line
Use the / keys to select the function for which you want to view results and
press the TEST key to confirm.
Under the same memory location the other saved results can be viewed by
using the / keys,
Pressing the MEM key returns the instrument back to the Memory Menu in
order selection of an alternative memory location..
61
MI 3002 & MI3102 Working with results
Step 1 Select Delete results from the Memory menu by using the and keys to
place the cursor next to the option and press the TEST key to confirm. The
Delete results menu will be displayed and the last used memory location will
be selected
Step 2 To delete saved result(s) the memory location must be selected first:
The Object line can be selected with the / keys. Use the / keys to
select the appropriate ‘Object ‘memory location (three-digit number).
In the lower part of the display, two numbers will be displayed (e.g. No.:2/16
see figure 6.8):
- the first number is the number of measurements stored in the selected
‘Object’ memory location.
- the second number is the number of measurements stored in the selected
‘Object’ memory location and its sub-locations (i.e. D.board, Circuit, Earth
Electrode, Bonding memory locations).
The D. Board line can be selected with the / keys. Use the / keys to
select appropriate ‘D. Board memory location (three-digit number).
In the lower part of the display, two numbers will be displayed (e.g. 6/16 see
figure 6.8):
- the first number is the number of measurements stored in the selected ‘D.
Board’ memory location.
62
MI 3002 & MI3102 Working with results
The Circuit / Earth Electrode / Bonding line can be selected with the /
keys.
The appropriate memory location (‘Circuit’ or ‘Earth Electrode’or ‘Bonding’)
can be selected with the TEST key.
If the Circuit is selected, use the / keys to select the appropriate ‘Circuit’
memory location.(three-digit number). In the lower part of the display (No.
line) the number of measurements stored in the selected ‘Circuit’ memory
location is shown.
If the Bonding is selected, use the / keys to select appropriate ‘Bonding’
memory location (e.g. Water/ Oil/ Lightning/ Gas/ Str.Steel/ Earth.cond.). In
the lower part of the display the number of measurements stored in the
selected ‘Bonding’ memory location is shown.
If the Earth Electrode is selected, the number of measurements stored in the
selected ‘Electrode’ memory location will be displayed in the lower part of the
display.
To exit the Delete results menu without deleting any results, rotate the function switch.
Step 1 After the memory location (Object, D. Board, Electrode, Bonding, Circuit) has
been selected, press the MEM key. The cursor will jump down to the No. line
Step 2 Use the / keys to select the results you want to delete and press the TEST
key. Press the TEST key again to confirm that the results require deletion or
press MEM key to return to the menu for the selection of different memory
location (without deleting any results).
63
MI 3002 & MI3102 Working with results
How to delete all saved results in a memory location and its sub-
locations
Step 1 After the memory location (Object, D. Board, Earth Electrode, Bonding,
Circuit) has been selected, press the HELP key. A warning will appear at the
bottom of the screen asking for confirmation.
Figure 6.9: Menus for deleting all results in memory location and its subitems
Step 2 Press the TEST key to delete the results in the selected memory location and
all of the associated sub-locations. Press the TEST key again for confirmation
or press MEM key to return to the menu for selection of memory location
without changes.
Step 1 Select Clear memory in Memory menu by using the and keys and press
the TEST key to confirm. The following menu is displayed:
Step 2 Press the TEST key again to confirm that all results require deleting or press
any cursor key (or MEM key) to return to the Memory menu without deleting
any saved results.
64
MI 3002 & MI3102 RS232 / USB communication
3 1
5
1
2 6
4
Figure 7.1: Interface connection for data transfer over PC COM port
Step 1 Connect the Eurotest instrument to PC using either the RS232 or USB cable.
Make sure that the correct communication port is selected and that the Eurotest
instrument is switched on. Refer to chapter 4.5.4 Communication port selection.
Step 3 Select the Receive results icon or Instrument / Receive results option from
the menu. The system will begin to download results saved in the instruments
memory onto the PC. After the result have been downloaded, the following
memory structure will be displayed.
65
MI 3002 & MI3102 RS232 / USB communication
Step 4 The software allows a variety of tasks to be performed including editting the
downloaded structure for documentation purposes, saving results / parameters
to an *.eul file or proceed to creation of appropriate Report / Certificate.
Note:
The USB drivers must be installed on PC before using the USB interface is used.
See accompanying CD for further instructions about USB installation.
For more information about the operation of the software application, please see
the help files available from the HELP option in the EurolinkPRO software.
66
MI 3002 & MI3102 Maintenance
8 Maintenance
8.1 Replacing fuses
There are three fuses under back battery cover of the Eurotest instrument.
F1
M 0.315 A / 250 V, 205 mm
This fuse protects internal circuitry of low-value resistance function if test probes
are connected to the mains supply voltage by mistake.
F2, F3
F 4 A / 500 V, 326.3 mm
General input protection fuses for the L/L1 and N/L2 test terminals.
Warnings:
Disconnect any measuring accessory from the instrument and ensure that
the instrument is turned off before opening the battery/fuse compartment cover,
hazardous voltage can exist inside this compartment!
Replace any blown fuses with exactly the same type of fuse. The instrument can
be damaged and/or operator’s safety impaired if this is not performed!
The Position of fuses can be seen in figure 3.4 in chapter 3.3 Back panel.
8.2 Cleaning
No special maintenance is required for the housing. To clean the surface of the
instrument use a soft cloth slightly moistened with soapy water or alcohol. Then leave
the instrument to dry totally before use.
Warnings:
Do not use liquids based on petrol or hydrocarbons!
Do not spill cleaning liquid over the instrument!
8.4 Service
For repairs under warranty, or at any other time, please contact your distributor.
Unauthorised person(s) are not allowed to open the Eurotest instrument. There are no
user replaceable components inside the instrument, except for the three fuses inside
the battery compartment, refer to chapter 8.1 Replacing fuses.
67
MI 3002 & MI3102 Technical specifications
9 Technical specifications
9.1 Insulation resistance
Insulation resistance (nominal voltages 100 V DC and 250 V DC )
Measuring range according to EN61557-2 is 0.017 M 199.9 M.
Measuring range (M) Resolution (M) Accuracy
0.000 1.999 0.001
2.00 99.99 0.01 (5 % of reading + 3 digits)
100.0 199.9 0.1
Voltage
Measuring range (V) Resolution (V) Accuracy
0 1200 1 (3 % of reading + 3 digits)
68
MI 3002 & MI3102 Technical specifications
RCD test current selection (r.m.s. value calculated to 20 ms) according to IEC 61009:
½I N 1I N 2I N 5I N RCD I
I N (mA) AC A AC A AC A AC A AC A
10 5 3,5 10 20 20 40 50 100
30 15 10,5 30 42 60 84 150 212
100 50 35 100 141 200 282 500 707
*)
300 150 105 300 424 600 848 1500
*)
500 250 175 500 707 1000 1410 2500
*) *) *)
1000 500 350 1000 1410 2000
*)
not available
Measuring range according to EN61557-6 is 3.0 V 49.0 V for limit contact voltage 25
V.
Measuring range according to EN61557-6 is 3.0 V 99.0 V for limit contact voltage 50
V.
Measuring range (V) Resolution (V) Accuracy
0.0 9.9 0.1
(-0 % / +10 %) of reading + 2
digits
10.0 99.9 0.1 (-0 % / +10 %) of reading
Accuracy applies for 1 year in reference conditions. Temperature coeficient ouside
these limits is +1 digit.
Test current ...................................... max. 0.5I N
Limit contact voltage ......................... 25 V, 50 V
UC
Fault loop resistance at contact voltage is calculated as RL .
I N
69
MI 3002 & MI3102 Technical specifications
Trip-out time
Measuring range (ms) Resolution (ms) Accuracy
0 300 1 3 ms
Contact voltage
Measuring range (V) Resolution (V) Accuracy
(-0 % / +10 %) of reading +
0.0 9.9 0.1
2 digits
10.0 99.9 0.1 (-0 % / +10 %) of reading
70
MI 3002 & MI3102 Technical specifications
71
MI 3002 & MI3102 Technical specifications
72
MI 3002 & MI3102 Technical specifications
73
MI 3002 & MI3102 Technical specifications
If voltage greater than 500 V is applied to the test terminals, online voltage monitor is
used as voltage indicator only.
Reference conditions
Reference temperature range ........ 10 OC 30 OC
Reference humidity range .............. 40 %RH 70 %RH
Operating conditions
Working temperature range ........... 0 OC 40 OC
Maximum relative humidity............. 95 %RH (0 OC 40 OC), non-condensing
Storage conditions
Temperature range ........................ -10 OC +70 OC
74
MI 3002 & MI3102 Technical specifications
The error in operating conditions could be at most the error for reference conditions
(specified in the manual for each function) + 1 % of measured value + 1 digit unless
otherwise specified.
75
MI 3002 & MI3102 Fuse base tables
76
MI 3002 & MI3102 IT supply systems
B IT supply systems
In order for the operator to be familiar with the measurements and their typical
applications on IT supply systems, it is advisable to read Metrel’s handbook
Measurements on IT power supply systems.
B.2 Fundamentals
An IT supply system is a mains supply system that is insulated from ground (PE) (i.e. it
is ungrounded supplying system). The system either has no direct connection to ground
or the connection is provided through a relatively high impedance. IT systems mainly
occur in areas where additional protection against electric shock is required. Typical
places are medical surgery rooms.
IT supply systems also has no grounding currents except leakages, and in this way they
are not suceptable to problems with step voltages, (e.g. voltage drops in a step) or high
energy sparking in extrinsic areas.
The IT system can provide an additional level of protection against electric shock. In the
situation where a failure occurs in any of the insulation between the line conductors and
PE (e.g. through equipment failure, wrong application or procedure), this system will
remain safe but will be converted into a TN / TT type system. However, additional failure
can be hazardous, which means that the insulation must be continuously checked and
repaired immediately after a failure has been detected.
Supplementary to other protection devices, an IT system normally contains insulation
monitoring devices (IMD) or systems that sound alarms when insulation resistance or
impedance drops below set threshold. The threshold value usually depends on
environment and the application. (e.g. Typical value for a medical installations is 55 k.
In some countries, it is not enough to trace the insulation resistance of the IT supply
system to ground, they require tracing of system capacitance, too.
IEC 60364-4-41 (©IEC): In IT systems, live parts shall be insulated from earth or
connected to earth through sufficiently high impedance. This connection may be made
77
MI 3002 & MI3102 IT supply systems
either at the neutral point or at midpoint of the system or at an artificial neutral point.
The latter may be connected directly to earth if the resulting impedance to earth is
sufficiently high at the system frequency. Where no neutral point or mid-point exists a
line conductor may be connected to earth through high impedance.
L1
133 V
230 V
133 V 230 V
133 V
L2
230 V L3
Optional
high N (Optional) IMD
impedance
Note:
Where a residual current operating device (RCD) is used, tripping of the RCD in
the event of a first fault cannot be excluded due to capacitive leakage currents.
78
MI 3002 & MI3102 IT supply systems
The table below contains functions of the instrument including compatibility notes
related to IT system.
B.3.1 Voltage
79
MI 3002 & MI3102 IT supply systems
RCD testing is performed in the same way as in TN/TT system (See chapter 5.3 Testing
RCDs), with the following exception:
Contact voltage measurement is not relevant.
Test circuit with bypassing principle should correspond to that in figure B.3.
Zi
RCD RCD
L1
L2
L3
PE
L/L
1
N/L2
L2 PE L1
PE/L3
RE RE2
80
MI 3002 & MI3102 IT supply systems
Step 1 Select the Insulation function with the function selector switch and use the
/ keys to select ISFL (first fault current) function. The following menu is
displayed:
Step 3 Connect test cable to the instrument and to the item under test. Follow the
connection diagram shown in figure B.5 to perform first fault current
measurement. Use the Help function if necessary.
L1
L2
L3
PE
L/L
IMD N/L2
PE/L3 1
L2 PE L1
Zi RE
Figure B.5: Connection with plug commander and universal test cable
Step 4 Check the displayed warnings and online voltage/terminal monitor before
starting measurement. If OK, press the TEST key. After performing the
measurement results appear on the display together with the PASS/FAIL
indication (if applicable).
81
MI 3002 & MI3102 IT supply systems
Displayed results:
I SC1 .........First fault current between L1-PE,
I SC2 .........First fault current between L2-PE.
Step 1 Select the Insulation function with the function selector switch and use the
/ keys to select IMD check function. The following menu is displayed:
Step 3 Connect test cable to the instrument and to the item under test. Follow the
connection diagram shown in figure B.5 to check insulation monitor device.
Use the Help function if necessary.
Step 4 Check the displayed warnings and online voltage/terminal monitor before
starting the measurement. If OK, press the TEST key. Use the / keys to
decrease indicative insulation resistance until insulation monitor device
alarms bad insulation. Indicative insulation resistance and calculated first fault
current between first live conductor (e.g. L1) and PE conductor are displayed.
After performing the measurement results with PASS/FAIL indication appear
on the display.
Step 5 Use the key to select second live line (e.g. L2). Use the / keys to
decrease indicative insulation resistance until insulation monitor device
82
MI 3002 & MI3102 IT supply systems
alarms bad insulation. Indicative insulation resistance and calculated first fault
current between second live conductor (e.g. L2) and PE conductor are
displayed.
To complete measurement press the TEST key again.
After performing the measurement results with PASS/FAIL indication appear
on the display.
Displayed results:
R1 ...........Threshold indicative insulation resistance for L1,
I1 ............Calculated first fault leakage current at treshold insulation
resistance for L1,
R2 ...........Threshold indicative insulation resistance for L2,
I2 ............Calculated first fault leakage current at treshold indicative insulation
resistance for L2.
The calculated first fault current is the maximum current that would flow when
insulation resistance decreases to the same value as the applied test
resistance, and a first fault is assumed between opposite line and PE.
Note:
It is recommended to disconnect all appliances from the tested supply net to
receive regular test results. Any connected appliance will influence the insulation
resistance threshold test.
The displayed resistances and currents are indicative only. Displayed resistance
can significantly differ from the actual resistance the Eurotest simulates. If IMD’s
with very low test currents (below 1mA) are checked the displayed resistance
value is typically lower (and current higher) than the actual simulated resistance.
The difference is lower for lower set resistances.
83
MI 3002 & MI3102 IT supply systems
84
MI 3002 & MI3102 Reduced low voltage supply systems
C.2 Fundamentals
Special supply systems are applied where inherent protection against electric shock is
required but no SELV used. Reduced low voltage supply with ground reference can be
used for this purpose.
There are two options with 110 V nominal voltage.
L1 L1
55 V 63 V
110 V
110 V 110 V
PE 63 V 63 V
55 V L2
110 V
L2 L3
Single phase with center tap Three phase star connection, center
connected to PE (i.e. 2 x 55 V). tap connected to PE (i.e. 3 x 63 V).
No neutral line. No neutral line.
85
MI 3002 & MI3102 Reduced low voltage supply systems
The table below containsEurotest functions intended for test and measurement of
supply systems with compatibility notes related to the reduced low voltage system.
C.3.1 Voltage
86
MI 3002 & MI3102 Reduced low voltage supply systems
Tests can be carried out for both combination L1-PE and L2-PE. Each individual test
result is accompanied with appropriate indication.
87
MI 3002 & MI3102 Reduced low voltage supply systems
Nominal residual current................... 10 mA, 30 mA, 100 mA, 300 mA, 500 mA,
1000 mA
Nominal residual current accuracy.... -0 / +0.1I ; I = I N , 2I N , 5I N
-0.1I / +0; I = ½I N
Maximum nominal differential
current for declared accuracy ........... 1000 mA for I N
500 mA for 2I N
100 mA for 5I N
Maximum test current ....................... 1 A (for Z LOOP < 1 )
Test current shape............................ Sine-wave (AC), pulsed (A)
DC offset for pulsed test current ....... 6 mA typical
RCD type .......................................... general (G, non-delayed), selective (S, time-
delayed)
Test current starting polarity ............. 00 or 1800
Nominal input voltage ....................... 55 V / 63 V (45 Hz 65 Hz)
Measuring range according to EN61557-6 is 3.0 V 32.6 V for limit contact voltage 25
V.
Measuring range according to EN61557-6 is 3.0 V 66.0 V for limit contact voltage 50
V.
Measuring range (V) Resolution (V) Accuracy
0.0 9.9 0.1
(-0 % / +15 %) of reading +
2 digits
10.0 99.9 0.1 (-0 % / +15 %) of reading
Accuracy applies for 1 year in reference conditions. Temperature coeficient ouside
these limits is +1 digit.
Test current ...................................... max. 0.5I N
Limit contact voltage ......................... 25 V or 50 V
The contact voltage is calculated to.. I N (general type) or to 2I N (selective type)
UC
Fault loop resistance at contact voltage is calculated as RL .
I N
88
MI 3002 & MI3102 Reduced low voltage supply systems
Trip-out time
Complete measurement range corresponds to EN61557-6 requirements. Specified
accuracies are valid for complete operating range.
Trip-out current
Trip-out time
Measuring range (ms) Resolution (ms) Accuracy
0 300 1 3 ms
Contact voltage
Measuring range (V) Resolution (V) Accuracy
(-0 % / +15 %) of reading +
0.0 9.9 0.1
2 digits
10.0 99.9 0.1 (-0 % / +15 %) of reading
89
MI 3002 & MI3102 Reduced low voltage supply systems
90
MI 3002 & MI3102 Reduced low voltage supply systems
91
MI 3002 & MI3102 Reduced low voltage supply systems
92