User's Manual: Oxygen Concentrator
User's Manual: Oxygen Concentrator
Model:7F-3 / 7F-5B
User’s Manual
SAFETY NOTES………………………………………………………………….2-6
FEATURES………………………………………………………………………..7-9
HANDLING………………………………….……………………………………10
OPERATING & INSTALLATION……...........................………………………11-16
MAINTENANCE…………………………………………………………………17-18
TROUBLESHOOTING……………………..……………………………………19-20
-1-
SAFETY NOTES
WARNING:
INDICATION FOR USE: This device designed to extract oxygen from atmospheric air. It
will typically be an electrically-powered molecular sieve (artificial zeolite) used to separate
nitrogen from room air. The oxygen concentration is variable, depending on the flow rate
being utilized.
INTENDED USE: This oxygen concentrator is intended for individual use as an oxygen
supplement device in the professional healthcare facility and home healthcare environment. It
provides high concentration of oxygen to patients with respiratory disorders e.g. COPD
(Chronic Obstructive Pulmonary Disease), LTOT (Long Term Oxygen Therapy) etc.
Users who require continuous oxygenation must plan for alternate reserve sources
of power and oxygen in the event of a failure or loss of power and oxygen. This
device is to be used as an oxygen supplement and is NOT considered
life-supporting or life-sustaining.
WARNING:
CONTRACTIONS: Oxygen poisoning and oxygen allergy user/patient DO NOT
using this oxygen concentrator.
SIDE EFFECT: No side effect.
Symbol Description
Describes a hazard or unsafe practice that if not avoided can
result in severe bodily injury, death or property damage.
Describes a hazard or unsafe practice that if not avoided can
result in minor bodily injury or property damage.
Ⅰ. IMPORTANT MESSAGE
To reduce the risk of electrical shock, DO NOT remove the cabinet. Refer service to
qualified service personnel.
Before operating the device, read and understand this manual.
The patient is an intended operator.
Ⅱ. BEFORE INSTALLATION
The concentrator should always be kept upright to prevent cabinet damage while being
transported.
If the electrical source power voltage is unstable that beyond normal power rang, please
add the voltage stabilizer.
Please use eligible, safe power set and junction box.
Non-professional people DO NOT open oxygen concentrator’s cabinet.
Ⅲ.LOCATING
You may select a room in your house where using your oxygen concentrator would be
most convenient. Your concentrator can easily move from one room to other room by
the wheels.
-2-
SAFETY NOTES
Be certain to place the device so all sides are at least 10 cm (4 inches) away from walls,
draperies, furniture, or other obstruction. MUST place the oxygen concentrator in a
well-ventilated area.
Always place the device in the expected position that the operator can hear the audio
alarming.
Oxygen concentrator MUST locate so as to avoid pollutants or fumes.
Oxygen concentrator must avoid heat source, fire source, wetness, exorbitant or
over-low extra ventricular condition.
Sundries and vessel do not be placed on top of the concentrator.
NEVER block the air openings of the unit or place it on a soft surface, such as a bed or
couch, where the air opening may be blocked. Keep the openings free from lint, hair and
the like.
Ⅳ.USING
The use of oxygen therapy requires that special care be taken to reduce the risk of fire.
Users MUST NOT SMOKE while using this device. Keep all matches, lighted
cigarettes or other sources of ignition out of the room in which this product is located.
NO SMOKING signs should be prominently displayed. Textiles and other materials
that normally would not burn are easily ignited and burn with great intensity in oxygen
enriched air. Failure to observe this warning can result in severe fire, property damage
and cause physical injury or death.
It can take up to 10 minutes to stabilize to deliver the set flowrate and concentration of
oxygen to user after switching on the oxygen concentrator.
For optimum performance, do not open or turn off the concentrator frequently, reset
after 3~5 minutes. Shorter periods of operation may reduce maximum product life.
A spontaneous and violent ignition may occur if oil, grease or greasy substances come
in contact with oxygen under pressure. These substances MUST be kept away from the
oxygen concentrator, tubing and connections, and all other oxygen device.
DO NOT use any lubricants unless recommended by manufacturer.
Ⅴ. MAINTENANCE
The oxygen concentrator was specifically designed to minimize routine preventive
maintenance at intervals of once per year. Only professionals of the healthcare field or
persons fully conversant with this process such as authorized or factory trained personnel
should perform preventive maintenance or performance adjustments on the oxygen
concentrator.
-3-
SAFETY NOTES
Ⅵ. RADIO FREQUENCY INTERFERENCE
This equipment generates, uses and can radiate radio frequency energy and, if not installed
according with the instructions, may cause harmful interference to other devices in the
vicinity. However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to other devices, which can
be determined by turning the equipment off and on, the user is encouraged to try to correct
the interference by one or more of the following measures:
• Reorient or relocate the receiving device.
• Increase the separation distance between the equipment.
• Connect the equipment into an outlet on a circuit different from that which the other
device(s) are connected.
• Consult the manufacturer or service technician for help.
-4-
SAFETY NOTES
DO NOT connect the oxygen concentrator in parallel or series with other oxygen
concentrators or oxygen therapy devices.
Use of some administration accessories or certain humidifiers not specified for use with
oxygen concentrator may impair the performance.
In certain circumstances oxygen therapy can be hazardous. Manufacturer recommends
that you seek medical advice before using this product.
Avoid creation of any spark near medical oxygen equipment. This includes sparks from
static electricity created by any type of friction.
The oxygen delivery settings of the oxygen concentrator should periodically reassessed
for the effectiveness of the therapy.
To reduce the risk of infecting another user or operator of reuse the oxygen concentrator,
the cabinet should be cleaned by a mild household cleaner before reuse. And all external
tubing, cannula, humidifier or other accessory should be replaced.
The shelf life of the device is 3 years after manufacturing (includes all components,
excepts single use consumable materials). And the Service Life is equal to the shelf
life.
Be aware that the power cord and/or tubing could present a tripping or strangulation
hazard. Always place the power cord and/or tubing in a manner that prevents
crushing by casters or others.
DO NOT modify the oxygen concentrator in any way. Modifications could result in
hazards to the user.
-6-
FEATURES
Ⅰ. SUMMARY
The oxygen concentrator is intended for individual use in the professional healthcare
facility and home healthcare environment. It is an electronically operated device that
separates oxygen from room air. It provides high concentration of oxygen directly to
you through a nasal cannula. Clinical studies have documented that oxygen
concentrators are therapeutically equivalent to other types of oxygen delivery systems.
This user’s manual will tell you about your concentrator and will serve as a reference
as you use your concentrator.
Ⅱ. CHARACTERS
Complete plastic out shell, safe and reliable.
Timing accumulates function, show total operating hours through the display screen.
Timing turns off function using convenience.
Compressor pressure relief valve help the device safer.
Power interruption alarm function.
Device failure alarm function (including pressure failure, compressor failure, low
oxygen concentration).
Compressor with over heat protect function to gain in safety of the compressor and
the concentrator.
Ⅲ. SPECIFICATIONS
1. Power Supply: ~230V, 50Hz
2. Input Power: 【7F-3】320VA;【7F-5B】400VA
3. Maximum Recommend Flow Rate: 【7F-3】3L/min;
【7F-5B】5L/min
-8-
FEATURES
15. Safety System:
Over Current or Connection Loosen (Loss Power): Alarming and Shut Down
Compressor Over Hot: Unit Shut Down
Pressure Failure: Alarming and Shut Down
Compressor Failure: Alarming and Shut Down
Low Oxygen Concentration
16. Normal Operating Condition (with Oxygen Concentration Status Indicator):
Temperature range: 5℃~40℃
Relative humidity: 15%~90%
Atmosphere pressure: 860hPa~1060hPa
Note: When operating condition over the rated ranges of ambient temperature, humidity
and atmosphere pressure, the performance of the oxygen may be dropped.
17. Oxygen Output Temperature: ≤ 46℃
Applied part Temperature(Nasal Cannula): ≤ 41℃ in normal condition
Applied part Temperature(Nasal Cannula): ≤ 43℃ in single fault condition
18. Cannula length DO NOT more than 15.2m and no twist.
19. Storage and Transportation Condition:
Temperature range: -20℃~60℃
Relative humidity: 10%~93% Non-condensing.
Atmosphere pressure: 700hPa~1060hPa
The device should be stored with no strong sunlight, no corrosive
gas and well ventilated indoor area. The device must be
transported and used in the vertical position only.
-9-
HANDLING
Ⅰ. UNPACKING
Unless use the oxygen concentrator, retain containers and
packing materials for storage until use of the concentrator is
required.
1. Check for any obvious damage to the carton or its contents. If damage is evident, please
notify the carrier or local dealer.
2. Remove all loose packing from the carton.
3. Carefully remove all the components from the carton.
Ⅱ. INSPECTION
1. Examine exterior of the oxygen concentrator for nicks, dents, scratches or other
damages.
2. Inspect all components.
Ⅲ. STORAGE
1. Store the repackaged oxygen concentrator in a dry area.
2. DO NOT place other objects on top of the repackaged concentrator.
- 10 -
OPERATING & INSTALLATION
Ⅰ. FEATURE VIEW
Figure 1: view of the machine
- 11 -
OPERATING & INSTALLATION
Ⅱ. PREPARE WORK
1. Open the transparent door, unscrew the bottle from the humidifier counter-clockwise. Fill
the bottle with pure water (or distilled water) to the level indicated by the manufacturer.
2. Screw securely the bottle back counterclockwise.
(Recommend to use Yuwell Humidifier p/n: 20005727 and its preferred location is shown as
Figure 1.)
3. Close the transparent door.
4. Connect the power supply:Ensure that the power switch is off; plug the concentrator’s
AC connector into a properly power outlet.
5. If you want to remove the concentrator , pull out the four caster lockers. ( figure 3 )
figure 3
pull out the locker to remove push down the locker to fix
NOTE: Concentrator may be used during the initial start warm-up time (approximately 30
min) while waiting for the O2 purity to reach maximum.
Ⅲ. OXYGEN ABSORBING OPERATION
1. Turn the power switch to the“|”position. The LCD shows “HELLO” , the green , yellow
indicator all light means concentrator is normal. About 1 second
later , only green indicator lights. After 4 seconds , the LCD
shows elapsed time .
2. Turn the flowrate knob to the setting prescribed by your
physician or therapist. ( Figure 4 )
4
- 12 -
OPERATING & INSTALLATION
NOTE: To properly read the flowrate, locate the prescribed flowrate line on the flowmeter.
Next, turn the flow knob until the ball rises to the line.
Now, center the ball on the L/min line prescribed.
Nasal cannula
oxygen outlet
Figure 5
Note: The firesafe valve is a thermal fuse designed to extinguish an oxygen delivery tube
fire and stop the flow of oxygen if the tube is accidentally ignited. And the firesafe valve is
flow direction-sensitive component, the install direction must correct.
Note: The firesafe valve is maintenance free and has an intended life of 4 years. Dispose of
any firesafe valve 4 years after the manufacturing date state on the device label. Once
actuated, the firesafe valve cannot be reset and must be discarded.
Note: There is another firesafe valve installed close to the oxygen outlet and inside the
cabinet of the oxygen concentrator.
Note: The proper placement and positioning of the prongs of the nasal cannula in the nose
is critical to the amount of oxygen delivered to the respiratory system of the patient.
Note: The nasal cannula is single use. Re-using the cannula may increase the re-infection
risk.
- 13 -
OPERATING & INSTALLATION
Ⅳ. ALARM SIGNAL
Concentrator has following failure Alarm function:
1) Pressure failure 2) Compressor failure
3) Low oxygen concentration 4) Over temperature
5) Low flowrate 6) Power loss
Note: All the alarms are LOW PRIORITY.
Alarm can only be ceased by turning off the device.
Initial startup of the concentrator
When the unit is turned on, the green/yellow lights and alarm will come on about 1
second to ensure the function of the indicators, then the yellow lights and alarm will
come off. After 5 minutes later, the oxygen sensor will be operating normally and will
control the indicator lights depending on oxygen concentration values. The explanations
of the indicator light functions are as follows.
Explanation of the indicator light and alarm
Indicator
Symbol Status alarm
lights
I/O SYSTEM OKAY
(O2 concentration is greater than 82%.) Green —
SYSTEM FAILURE
(Pressure failure alarm; Or, Compressor
Yellow Alarm
circuit failure alarm ; Or , over
temperature alarm,low flowrate)
- 14 -
OPERATING & INSTALLATION
3. Pressure failure alarm - yellow light illuminates, alarm sounds, and panel shows word
“E1” or “E2”. Total unit shutdown. Switch immediately to backup oxygen supply.
Call supplier immediately.
4. Compressor failure alarm - yellow light illuminates, alarm sounds, and panel shows
word“E3” or “E4”. Total unit shutdown. Switch immediately to backup oxygen
supply. Call supplier immediately.
5. Over temperature alarm - yellow light illuminates, alarm sounds, and panel shows
word“E3”. Total unit shutdown. Switch immediately to backup oxygen supply. Call
supplier immediately.
6. Low flowrate alarm - yellow light illuminates, alarm sounds, and panel shows word
“LL”. Total unit shutdown. Switch immediately to backup oxygen supply. Call
supplier immediately.
7. Power loss alarm - alarm sound , no display , no machine working. Please check the
power input.
NOTE: Concentrator will reach to most stable state after warm-up (approximately 30
minutes).
Ⅴ. TIME SETTING
User can press the timing button to setting the operating time in 1~99 minutes.
When it start working, the display shows “elapsed time” and " setting time ——",
it means that timing function is closed. It will work continually until power supply
cuts off.
Press the “+” button once, operation time increases 1 minute, holding the button
more than 1.5 seconds it will increase continually.
Press the “-” button once, operation time decrease 1 minute, holding the button more
than 1.5 seconds it will decrease continually.
The device will auto turn off and the display will show “00” when finishing timing. Reset
timing function.
Ⅵ. TURNNING OFF
Take off nasal oxygen cannula from oxygen outlet first, turn off power switch, and then
cut off the power source.
- 15 -
OPERATING & INSTALLATION
Ⅶ. SYMBOLS
No open flame:
Fire, open ignition
No smoking
source and
smoking prohibited
Keep Up Fragile
European
Manufacturer
Representative
- 16 -
MAINTENANCE
NOTE: Power should be disconnected before beginning preventive maintenance on the
concentrator. The concentrator is specifically designed to minimize routine preventive
maintenance at intervals of once per year. In places with high dust or soot levels,
maintenance may need to be performed more often. The following must be performed at a
minimum of one year in service to assure years of additional reliability.
Ⅰ. CLEAN CABINET
CUT OFF the power supply first to avoid electrical shock.
DO NOT remove device cabinet.
Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge at
least one time a month. Do Not drop any liquid into the seam of the device.
Ⅱ. CLEAN OR REPLACE FILTER
Please clean or replace the filters in time, it’s very important to protect compressor and
to extend the device life.
1. Clean the I-stage air inlet once a month. ( Figure 6 )
Replace or clean the filter immediately if the mesh turns black. (Figure7)
1) Clean the filter with a soft cleaner or wash in warm soapy water and rinse thoroughly.
2) DRY the filter thoroughly before reinstallation.
3) The filter must clean or replace after operating 100 hours operating at least or as
necessary.
- 17 -
MAINTENANCE
DO NOT operate the concentrator without installing filter, or while
filter is wet. These actions could permanently damage the
concentrator.
3. Clean Humidifier
1) Remove the humidifier bottle from the humidifier cap then clean the bottle.
2) Remove the humidifier tube and diffuser then clean them.
3) The humidifier must clean or replace once a week at least or as necessary.
4) Replace clean water into humidifier every day before using.
Note: During both normal condition and single fault condition, the cabinet, humidifier and
nasal cannula can become contaminated with body fluids or expired gases. To reduce the
infection risk, please perform the maintenance routinely
- 18 -
TROUBLESHOOTING
Use the table below to take actions when the oxygen concentrator indicates an
abnormal condition.
TROUBLESHOOTING GUIDE
- 19 -
TROUBLESHOOTING
Symptom Probable cause Solution
1) Stop using the device, please contact
1) Compressor circuit is opened.
the supplier immediately.
Concentrator is not
working, the yellow light 2) fan failure ; or concentrator air inlet
2) inside temperature is too high ;
illuminate, alarm, the and air outlet blocked ;
display show “E3”.
3) If the phenomenon is still appeared, stop using the device, please contact the
supplier immediately.
Concentrator is not 1) Compressor circuit is closed. 1) Stop using the device, please contact
working, the yellow light the supplier immediately.
illuminate, alarm, the 2) If the phenomenon is still appeared, stop using the device, please contact the
display show “E4”. supplier immediately.
1) oxygen outlet or tubing accessories
Concentrator is not 1) oxygen flowrate is too low. is blocked or kinked; or flowrate
working, the yellow light knob is turn down.
illuminate, alarm, the
2) If the phenomenon is still appeared, stop using the device, please contact the
display show “LL”.
supplier immediately.
1) Poor ventilation around the 1) Make sure the device is at least 10
around the device, so operating cm away from the walls, other jam
temperature is too high. or heater.
2) The water which added to
The nasal cannula is 2) Add cold water to humidifier.
humidifier is too hot.
fogged with vapor or full
of liquid drops. 3) Water added should between the
3) Add too many water into
(Remove the vapor or maximum and minimum levels in
humidifier.
liquid drops method: humidifier.
Turn on the device and 4) Turn off the device immediately.
4) The device stops suddenly
connect the tubing to the Restart the device to remove the
during operating.
oxygen outlet. Use moisture and liquid.
your finger to block and 5) Accessory is connected to the
5) Smooth the nasal cannula and
release the end of the device. The tubing is kinked
tubing.
tubing several times to and the device stops suddenly.
remove the moisture and 6) Cooling fan which mounted
6) Replace the cooling fan.
liquid.) inside the device stop running or
Note: Cut off the power first. And
rotate slowly, make the
perform by qualified personnel only.
operating temperature is too hot.
7) If the phenomenon is still appeared, the device can be used but please
contact the supplier.
If you have all other problems, TURN OFF the concentrator first,
use your reserve oxygen supply, and contact the supplier
immediately.
- 20 -
OTHER ATTENTION ITEMS
Ⅰ. GAS PASS OPERATION SKETCH MAP
Ⅱ. ELECTRICAL RATIONALE
Oxygen
Main Board
~ Sensor
Assemb
ValveⅡ
- 21 -
OTHER ATTENTION ITEMS
Ⅲ. PACKING LIST
1. Oxygen Concentrator: 1 unit
2. User & Technical Manual: 1 piece
3. Fuse tube:T5AH250V~/φ5×20,2pcs
4. Firesafe Cannula Valve [ Bpr firesafe ] :1 piece
5. connect tube :φ5×φ9×40mm silicone tube
6. Nasal Oxygen Cannula 2m:1 piece
Ⅳ.DEVICE DISPOSAL
Contact the local cit or town offices for proper disposal instructions for oxygen
concentrator.
- 22 -
EMC Information
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Emission of harmonics
Class A
IEC 61000-3-2
Voltage fluctuations /
Complies
flicker emissions IEC 61000-3-3
- 23 -
EMC Information
Surge
±1 kV line to line
IEC 61000-4-5
0% UT 0.5 cycle at
0°,45°,90°,135°,180°,225°,270°and
315°
Voltage dips, short
interruptions and Voltage variations on 0% UT 1 cycle
power supply input lines 70% UT 25/30 cycles at 0°
IEC 61000-4-11
Voltage short interruptions :
0% UT 250/300 cycles at 0°
Power frequency
(50 Hz) magnetic 30 A/m
IEC 61000-4-8
10V/m
Radiated RF EM fields
80MHz – 2.7GHz
IEC 61000-4-3
80% AM at 1kHz
3V/m
Conducted disturbances induced by 0.15MHz – 80MHz
RF fields 6V in ISM and amateur radio bands
IEC 61000-4-6 between 0.15MHz and 80MHz
80% AM at 1kHz
Note: UT is the a.c. mains voltage prior to application of the test level.
- 24 -
EMC Information
Pulse modulation
385 380 - 390 TETRA 400 1,8 0,3 27
18 Hz
FM
GMRS 460,
450 430 - 470 ±5 kHz deviation 2 0,3 28
FRS 460
1 kHz sine
710
LTE Band Pulse modulation
745 704 - 787 0,2 0,3 9
13,17 217 Hz
780
GSM 1800;
1720
CDMA 1900;
GSM 1900;
Pulse modulation
1845 1700 - 1990 DECT; 2 0,3 28
217 Hz
LTE Band
1970 1,3,4,25;
UMTS
Bluetooth,
WLAN,
Pulse modulation
2450 2400 - 2570 802.11 b/g/n, 2 0,3 28
217 Hz
RFID 2450,
LTE Band 7
5240
WLAN 802.11 Pulse modulation
5500 5100 - 5800 0,2 0,3 9
a/n 217 Hz
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is
permitted by IEC 61000-4-3.
- 25 -
Jiangsu Yuyue Medical Equipment & Supply Co., Ltd.
Address: Yunyang Industrial Park, Danyang, Jiangsu, P.R.China
TEL: 86-511-86900833
Notified Body:
- 26 -