Immunity Test System
Immunity Test System
PARTNER
Accredited Calibration
Quality at EMC PARTNER is based on an ISO 9001 management sys-
tem. This is the foundation for an ISO/IEC 17025 accreditation
verified by the Swiss Calibration Service (SCS). SCS No. 146 is the
accredita-tion number of EMC PARTNER AG. Locally accredited but
recognized worldwide through affiliation with the ILAC organisation
THE CONDUCTED IMMUNITY TEST SYSTEM
The ideal is a flexible test equipment that combines any or all of these
requirements into one single solution.
3
CUSTOM IS STANDARD
IMU and DOW generators are designed with the user in
mind. A modular architecture allows configuration of test
circuits to meet any requirement. A solution that can
easily be extended with more modules when additional
tests are needed. IMU-MGE
DOW-CG1
IMU-MGS
SLOW 100kHz & 1MHz / 4.4kV Surge / CWG 4kV / 5 kV / 6kV / 8kV
FAST 3MHz, 10MHz & 30MHz / 4 kV Ring Wave 6kV / 8kV
CDN 690Vac & 500Vdc / 32A Telecom 6kV / 8kV
Insulation 1.2/50us 500Ω / 0.5J 8kV EFT 4kV / 5kV / 6kV / 7kV
Common Mode 35V / 330V
AC & DC Interrupts 16A / 32A / 75A
AC DIPS & Variation 16A
Differential Mode voltage / current
ESD 16kV
WHY
MODULARITY ?
You get a cost-effective
solution that fits your test
requirements. Also, you
can extend your system
for other tests after initial
purchase and on-site
whenever you have new
test requirements.
4
AVAILABLE TESTS
IMU Series
DOW Series
5
Enhanced safety is standard
UNIQUE FEATURES
Leading technology - New designs take advantage of latest innovations.
Link together many different test types Generate test reports via USB interface
into one sequence. or built in webserver as csv, html and
pdf formats.
6
EPOS - BEYOND BETTER
EMC PARTNER Operating System (EPOS) is an independant software with free-of-charge up-
dates for lifetime. EPOS is based on a full colour graphic interface and easy to follow on-
screen graphics. Pop-up help gives information when needed, directly during the setting
process. EPOS is full of features found only in top of the range instrumentation.
Pre-programmed tests for all basic and Save time with the latest in intuitive
generic standards included. menu structures.
Active monitor and control of EUT power. Select between English, German,
EUT current limit to prevent damage. French, Italian, Spanish, Russian,
Chinese (traditional and simplified).
7
TEMA3000 SOFTWARE SUITE
The best solution for professional EMC Test Labs enables comfortable test setups, easy
parameter changes and customizable test reports and DSO integration.
Want to know more? Further information and demo version available on the website
CDNS
IEC 61000-4 -2 -4 -5 -8 -9 -11 -12 -16 -18 -19 -29 -34
Supply Lines
CDNs for burst, Surge and Ring wave
4 kV / 6 kV line
CDN-M-6-32
CDN-A-6-32
CDN-A-6-63
CDN-A-6-125
CDN-A-6-200
7.6 kV line
CDN-M-8-32
CDN-A-8-32
CDN-A-8-63
CDN-A-8-125
CDN-A-8-200
7 kV burst
CDN-F-125
CN16DC
DN16-1P6
DN16-1P16
I/O Lines
CDN-KIT1000 ED3
CDN-DATA-4L
CDN-DATA-8L
CDN-UTP ED3
CDN-UTP8 ED3
CN16T, CN16T4
CN-R40C05 other
CN-R40C05 8 other
CDN-DOW-DATA-LF
CDN-DOW-DATA-HF18
ESD-TARGET2
ESD-VERI-V
ESD-STAND Ed2
EARTH CABLE
ESD-VCP50
CN-EFT1000
VERI50 EFT
VERI1K EFT
VERI-CP-EFT
ADAPTER EFT-CDN
CN-BALUN AC ANSI
EFT-INSULATION
V-PROBE-SI
I-PROBE-P101
MF1000-1
MF1STAND
MF1000-2
MF1000-3
MF3STAND
VAR-EXT1000
SRC16-1P
VERI-DIPS
DIPS100E
PFS32
SRC32-18UH
SRC32-AMD1-18 UH
PFS75
PFS75-690V
SRC75-18UH
SRC75-690V-18UH
PS3
RS485-RS232 ADAPT.
EXT-IMU C-SHORT
VERI01 OSI
VERI10/50
EXT-IMU D-29D
EXT-IMU D-29I
EMERGENCY STOP
WARNING LAMP
TC-ST
TEMA3000 SOFTWARE
IMU-MGS
IMU-MGS Mainframe
EUT power input 1 (CDN) max. AC/DC 300V, 16A (fused 16A) with P module
EUT power input 2 (CDN) max. AC/DC 300V, 16A (not fused) with P module
Internal CDN freq. range DC, 50 Hz, 60 Hz
Power freq. synchr. 50/60 Hz
Coupling Burst L, N, PE, L+N, L+PE, N+PE, L+N+PE, direct out
Coupling Surge 2 Ω: L-N, direct out, 12 Ω: L-PE, N-PE
IMU-MGS mainframe
Operating voltage 100 – 240 V (50/60 Hz) ± 10 %
Power consumption ON < 150 VA, standby < 15 VA
Weight 16 kg (weight of modules not included)
Wxdxh 520 x 433 x 180 mm
Version 19“ unit, 4 UH
Temperature range 10 – 35 °C
Humidity < 80 % non-condensing
Air pressure 86 – 106 kPa
Included articles
Power cord with country plug
Supply connection 3 cables x 2 m, banana plugs
User manual digital format (PDF)
Calibration certificate factory calibration
Ethernet cable 1 piece
USB stick 1 piece
CDN burst cal. adapter 1 piece
IMU-MGE Mainframe
EUT power input 1 (CDN) max. AC/DC 300V, 16A (fused 16A) with P module
EUT power input 2 (CDN) max. AC/DC 300V, 16A (not fused) with P module
Internal CDN freq. range DC, 50 Hz, 60 Hz
Power freq. synchr. 50/60 Hz
Coupling Burst L, N, PE, L+N, L+PE, N+PE, L+N+PE, direct out
Coupling Surge 2 Ω: L-N, direct out, 12 Ω: L-PE, N-PE, direct out
Coupling Ring wave 12 Ω and 30 Ω: L-N, N-PE, L-PE, direct out
IMU-MGE module: EXT-IMU3000 R6, 100 kHz Ring wave extension 6.6 kV
Standard IEC 61000-4-12 latest edition
Output impedance 12 Ω, 30 Ω
Voltage OC 0.25 kV - 6.6 kV ± 10 %, decay as in IEC, ANSI
Voltage rise time/ osc. freq. 0.5 µs ± 30 % / 100 kHz ± 10 %
Current SC into 12 Ω 20.833 A – 550 A ± 10 %
Current SC into 30 Ω 8.3 A – 220 A ± 10 %
Current rise time 0.2 - 1 µs
Pulse repetition up to 1 / s @ 1 kV, 1 / s @ 6.6 kV
Polarity positive, negative, alternating
Synchronization 0 – 359°, step 1°
Ramps voltage, synchronisation angle
Optional accessories CDNs, CDNs for I/O Lines, Accessories
IMU-MGE mainframe
Operating voltage 100 – 240 V (50/60 Hz) ± 10 %
Power consumption ON < 150 VA, standby < 15 VA
Included articles
Power cord with country plug
Supply connection 3 cables x 2 m, banana plugs
User manual digital format (PDF)
Calibration certificate factory calibration
Ethernet cable 1 piece
USB stick 1 piece
CDN burst cal. adapter 1 piece
DOW-CG1 Mainframe
Built-in CDN three phase CDN integrated
EUT voltage AC max. 3 x 690 V L-L (400 V L-N), 50 / 60 Hz
EUT current AC max. 3 x 32 A
EUT protection AC over-current automatic prot., Type C charact.
EUT power DC max. 500 V, 32 A L-L or L-PE
EUT power switch ON / OFF switch on the touch screen
Internal CDN freq. range DC, 50 Hz, 60 Hz
Power freq. synchr. 50 / 60 Hz for both slow and fast
L1-L2, L1-L3, L2-L3, L1-N, L2-N, L3-N,
Coupling slow DOW L1-PE, L2-PE, L3-PE, N-PE, L1+N to PE, L1+L2+L3+N
to PE
L1-PE, L2-PE, L3-PE, N-PE, L1+N to PE, L1+L2+L3+N
Coupling fast DOW
to PE
DOW-CG1 mainframe
Operating voltage 100 – 240 V (50/60 Hz) ± 10 %
Power consumption ON < 400 VA, standby < 15 VA
Temperature range 10 – 35 °C
Humidity < 80 % non-condensing
Air pressure 86 – 106 kPa
Included articles
Power cord with country plug
Supply connection 5 cables x 2 m, banana plugs
User manual digital format (pdf)
Calibration certificate factory calibration
CDN Calibration adapter 1 piece
Ethernet cable 1 piece
USB stick 1 piece
CDN-M-6-32
Standards IEC61000-4-4, IEC61000-4-5, other
Can be upgraded for IEC61000-4-12 (for ANSI see automatic version)
Type 3-ph., manual
EUT voltage AC max. 3 x 690V L-L, 50 / 60 Hz
EUT current AC max. 3 x 32A (covers 0 – 32 A as per standard)
EUT voltage DC max. 500V (can be upgraded, see options)
EUT current DC max. 32A (covers 0 – 32 A as per standard)
Current flow bi-directional (source to EUT and EUT to source)
Test level burst (as per IEC) max. 6 kV
Test level surge (as per IEC) max. 6 kV
Surge waveform as required for current EUT range 0 – 32A
Residual voltage typically < 5 % for burst and < 12 % for surge
Dimensions 19” unit, basic 4 UH
Weight depending on options
Included in delivery 10 connectors (power, EUT), 5 EUT cables
adapter for burst calibration
Generators IMU-MGS, IMU-MGE
CDN-A-6-32
Standards IEC61000-4-4, IEC61000-4-5, other, latest editions
Can be upgraded for IEC61000-4-12, ANSI C62.45 (see options)
Type 3-ph., automatic
EUT voltage AC max. 3 x 690V L-L, 50 / 60 Hz
EUT current AC max. 3 x 32A (covers 0 – 32 A as per standard)
EUT voltage DC max. 500V (can be upgraded, see options)
EUT current DC max. 32A (covers 0 – 32 A as per standard)
Current flow bi-directional (source to EUT and EUT to source)
Test level burst (as per IEC) max. 6 kV
Test level surge (as per IEC) max. 6 kV
Surge waveform as required for EUT current range 0 – 32A
Residual voltage typically < 5 % for burst and < 12 % for surge
Dimensions 19” unit, basic 4 UH (with options max. 8 UH)
Weight depending on options
Included in delivery EUT connection cables (5 pieces), additionally
10 connectors (power, EUT), 5 EUT cables
adapter for burst calibration
Generators IMU-MGS, IMU-MGE
CDN-M-8-32
Standards IEC61000-4-4, IEC61000-4-5, other, latest editions
Can be upgraded for IEC61000-4-12 (for ANSI see automatic version)
Type 3-ph., manual
EUT voltage AC max. 3 x 690V L-L, 50 / 60 Hz
EUT current AC max. 3 x 32A (covers 0 – 32 A as per standard)
EUT voltage DC max. 500V (can be upgraded, see options)
EUT current DC max. 32A (covers 0 – 32 A as per standard)
Current flow bi-directional (source to EUT and EUT to source)
Test level burst (as per IEC) max. 7.1 kV with IMU-MGE
Test level surge (as per IEC) max. 8 kV supported (with IMU-MGEmax. 7.6 kV)
Surge waveform as required for current EUT range 0 – 32A
Residual voltage typically < 5 % for burst and < 12 % for surge
Dimensions 19” unit, basic 4 UH
Weight depending on options
Included in delivery 10 connectors (power, EUT), 5 EUT cables
adapter for burst calibration
Generators IMU-MGE
CDN-A-8-32
Standards IEC61000-4-4, IEC61000-4-5, latest editions
Can be upgraded for IEC61000-4-12, ANSI C62.45 (see options)
Type 3-ph., automatic
EUT voltage AC max. 3 x 690V L-L, 50 / 60 Hz
EUT current AC max. 3 x 32A (covers 0 – 32 A as per standard)
EUT voltage DC max. 500V (can be upgraded, see options)
EUT current DC max. 32A (covers 0 – 32 A as per standard)
Current flow bi-directional (source to EUT and EUT to source)
Test level burst max. 7.1 kV with IMU-MGE
Test level surge max. 8 kV supported (with IMU-MGE max. 7.6 kV)
Surge waveform as required for EUT current range 0 – 32A
Residual voltage typically < 5 % for burst and < 12 % for surge
Dimensions 19” unit, basic 4 UH (with options max. 8 UH)
Weight depending on options
Included in delivery 10 connectors (power, EUT), 5 EUT cables
adapter for burst calibration
Generators IMU-MGE
CN16
CN16-300
Standard IEC 61000-4-16 latest edition
Type manual, for 3 power lines
EUT voltage AC max. 3 x 480 V L-L (280 V L-N), 50 / 60 Hz
EUT current AC limited only by decoupling
EUT power DC max. 115 V (AC/RF disturbance) no current limit
Coupling paths 3 x 300 Ω, 1 µF per line
Test level power tests 330 V @ 16.67 Hz, 50 Hz, 60 Hz, 115 V @ DC
Test level sweep test 35 V
Dimensions 28 x 18 x 11 cm
Weight 3 kg
Optional PS3, EXT-IMU C-SHORT
DN16-1P6 or DN16-1P16 decoupling transformer
Generators IMU-MGS, IMU-MGE
CN16T
See section “Other CDNs for I/O Lines”
CN16-22-7C
Standard IEC60255-26 latest edition
Type manual, for 2 power lines, common mode test
EUT power AC 415 V, 50 / 60 Hz, 16 A
EUT power DC 300 V, 16 A
Coupling path 2 x 220 Ω ± 5 %, 0.47 µF ± 5 % per line
Test level max. 300 V
Dimensions 28 x 18 x 11 cm
Weight 3.3 kg
Requires PS3, EXT-IMU C-SHORT
Generators IMU-MGS, IMU-MGE
CN16-22-7D
Standard IEC60255-26 latest edition
Type manual, for 2 power lines, diff. mode test
EUT power AC 415 V, 50 / 60 Hz, 16 A
EUT power DC 300 V, 16 A
Coupling path 1 (class A) 2 x 100 Ω ± 5 %, 0.1 µF ± 5 % per line
Coupling path 2 (class B) 2 x 100 Ω ± 5 %, 0.47 µF ± 5 % per line
Test level max. 300 V
Dimensions 28 x 18 x 11 cm
Weight 3.5 kg
Requires PS3, EXT-IMU C-SHORT
Generators IMU-MGS, IMU-MGE
DN16-1P16
Standard IEC 61000-4-1 latest edition
Type 1P decoupling transformer for power lines
EUT power AC (see table) 230 V, 50 / 60 Hz, 16 A
EUT power DC not suitable for DC applications
Test level max. 330 V
Insulation ≥1 kV
CM decoupling > 60 dB in the range 15 Hz – 150 kHz
Dimensions 19 “ unit, 4 UH
Weight 50 kg
Requires CN16 or CN16DC
Generators IMU-MGS, IMU-MGE
Type of lines Line CDN-KIT1000 ED3 CDN-DATA-4L CDN-DATA-8L CDN-UTP ED3 CDN-UTP8 ED3
Unsymmetrical, Yes, Yes, max 4 Yes, max 8 Yes, max 2 Yes, max 4
unshielded max 2 lines lines lines lines lines
Unsymmetrical,
shielded
No CDN required, surge and ring wave are
applied on the shield directly from generator
Symmetrical,
shielded
CDN-KIT1000 ED3
CDN-DATA-4L
Standards IEC 61000-4-5, -4-12 latest editions
Application surge (fig. 9), ring wave on 4 asym. lines
EUT voltage per line max. 200 V DC or 240 V peak
EUT current per line max. 4 A cont. or 5 A for 5 min.
Line speed max. 100 kHz as per IEC 61000-4-5
Coupling path surge 1 40 Ω + 0.5 µF capacitor
Coupling path surge 2 40 Ω + 33 V bipolar diode
Coupling path surge 3 40 Ω + GDT 90 V, or any external element
Coupling path ring wave 1 33 V bipolar diode
Coupling path ring wave 2 any external element
Decoupling 20 mH per line, as per IEC 61000-4-5, -4-12
Test level surge max. 6 kV, coupling L-L, L-PE
Test level ring wave max. 6 kV, coupling L-L, L-PE (only IMU-MGE)
CDN-DATA-8L
Standards IEC 61000-4-5, -4-12 latest editions
Application surge (fig. 9), ring wave on 8 asym. lines
EUT voltage per line max. 200 V DC or 240 V peak
EUT current per line max. 4 A cont. or 5 A for 5 min.
Line speed max. 100 kHz as per IEC 61000-4-5
Coupling path surge 1 40 Ω + 0.5 µF capacitor
Coupling path surge 2 40 Ω + 33 V bipolar diode
Coupling path surge 3 40 Ω + GDT 90 V, or any external element
Coupling path ring wave 1 33 V bipolar diode
Coupling path ring wave 2 any external element
Decoupling 20 mH per line, as per IEC 61000-4-5, -4-12
Test level surge max. 6 kV, coupling L-L, L-PE
Test level ring wave max. 6 kV, coupling L-L, L-PE (only IMU-MGE)
Dimensions 19“ unit, 4 UH
Weight 25 kg
Generators IMU-MGS, IMU-MGE, other EMCP models
CDN-UTP ED3
Standards IEC 61000-4-5, IEC 61000-4-12 latest editions
Application 1 (1.2 / 50 µs) surge (fig. 10) on up to 4 sym. lines
Application 2 (1.2 / 50 µs) surge (fig. 9) on up to 2 asym. lines
Application 3 ( 10 / 700 µs) telecom surge (fig. A.4) on up to 4 asym. lines
Aplication 4 (0.5 µs /100kHz) ring wave as per fig. 8, 9 and 10
EUT voltage per line max. 300 V DC or peak
EUT current per line max. 1 A cont., total for all lines max. 2A
EUT line(s) characteristics high speed, over 100 Mbps on 4 wires (2pairs)
Example of EUT I/O lines RS485, USB, Ethernet 4 wires, CAN bus, etc.
Coupling path surge 1 2 x 80 Ω for 2 lines or 4 x 160 Ω for 4 lines
Coupling path surge 2 1 x 40 Ω + 0.5 µF
Coupling path tel. surge 3 2 x 25 Ω for 2 lines or 4 x 25 Ω for 4 lines
Coupling path ring wave 2 x GDT 90 V or 4 x GDT 90 V or 1 x 0.5 µF
Coupling elements 2 x GDT 90 V or 4 x GDT 90 V or 1 x 0.5 µF
Decoupling up to 4 x 20 mH per line, current compensated
AE protection 4 x GDT 90 V or customer defined
Test level surge, ring wave max. 6 kV, coupling CM and DM
Test level telecom surge max. 6 kV, coupling CM (only IMU-MGE)
Dimensions 19“ unit, 4 UH
Weight 40 kg
Included Mentioned coupling elements, AE protection
Optional ADAPTER BOX RJ45 (ask for details)
Other relevant standards ITU-T K20, K21, K44, FCC part 68 / D
Generators IMU-MGS, IMU-MGE, other EMCP models
CN-R40C05
Standard EN50121-4 latest edition
Application surge on 2 unsymmetrical lines
Test level surge max. 6 kV
Low speed I/O + Railway unsymmetrical, coupling with capacitor
Coupling path 1 40 Ω + 0.5 µF capacitor or simply 40 Ω
EUT current determined by CDN used for decoupling
EUT voltage 440V AC/DC
Dimensions 1 module
Weight 1 kg
Requires IMU internal CDN or external CDN
Generators IMU-MGS, IMU-MGE, other EMCP models
CN16T4
Standard IEC 61000-4-16 latest edition
Type manual, for 4 symmetrical I/O lines
EUT voltage max. 63 V (AC), max. 100 V (DC)
EUT current max. 0.5 A
Coupling path 4 x 400 Ω, + 4.7 µF per line
Test level continuous 35 V
Test level short duration 330 V (bridge open), 115 V (bridge closed)
Dimensions 30.5 x 23 x 11 cm
Weight approx 5 kg
Generators IMU-MGS, IMU-MGE
CDN-DOW-DATA-HF18
Standard IEC61000-4-18 SLOW DOW latest edition
Application manual CDN for applying SLOW DOW
on max. 8 sym. high speed lines (1 Gbps)
Frequency DOW 100 kHz, 1 MHz ± 10 %
Rise time 75 ns ± 20 %
Source impedance 200 Ω
Decaying Pk5 to Pk1 > 50 %, Pk10 to Pk1 < 50 %
Number of data lines 4 lines / 2 pairs, 8 lines / 4 pairs
EUT voltage per line max. 60 V DC (PoE) or signal voltage
EUT current per line max. 1 A (PoE all versions)
Coupling as per IEC61000-4-18
Decoupling > 1.5 mH per line
Voltage test level 200 V - 4000 V ± 10 %
Short circuit current 1 A – 20 A ± 20 %
Dimensions 30.5 x 23 x 11 cm
Weight 3.7 kg
Generators DOW-CG1, all models with SLOW module
Optional accessories ADAPTER BOX RJ45-8L (2 pieces needed)
ESD-TARGET2
Application current target for calibration of ESD generator
Input impedance 2Ω
Input voltage max. 10 kV CD
Frequency range ± 0.5 dB up to 1 GHz, ± 1.2 dB up to 4 GHz
Current range 0 – 50 A standard, could be extended
Transfer function 0.2 V / 1 A with 20 dB
Diameter 70 mm
Thickness 40 mm
Weight 1 kg
Fixing 8 x M3 screws, not included in delivery
Included 20 dB att., 50 Ω coax. cable (1 m) with BNC out
ESD-STAND Ed2
Application stand for supporting ESD gun, fixed point test
Height 50 – 180 cm, adjustable
Position 360˚ adjustable
Dimensions 64 x 17 x 12 cm (packed)
Weight 4 kg
Included cable holder for calibration
EARTH CABLE
Application connection of HCP or VCP to ground plane
Impedance 2 x 470 kΩ
Length 2m
Connectors 2 x banana plugs
ESD-VCP50
Application indirect ESD application as per standard
Spacer in between 10 cm plastic spacer
Coupling plane 50 x 50 cm
Application points one on each side
Dimensions 50 x 50 x 10 cm
Weight 8 kg
Included 2 m earth cable (with 2 x 470 kΩ)
CN-EFT1000
Application capacitive coupling clamp for burst on I/O lines
Coupling plate dimensions as per IEC 61000-4-4 latest edition
Waveform into 50 Ω 5 ± 1.5 ns / 50 ± 15 ns
Insulation 5/50 ns up to 8 kV
Insulation DOW up to 5kV 3 MHz - 30 MHz
Insulation 1.2/50 µs up to 5 kV
Usable cable diameter up to 70 mm
Coupling capacitance typically 100 pF – 1000 pF
Dimensions 114 x 15 x 10 cm
Weight 4 kg
Included high voltage cable for connection to IMU
Other applications IEC 61000-4-18, ANSI C37.90
CN-BALUN-AC
Application differential mode burst test adapter
Standards applicable ANSI C37.90, fig. 6, ISO7637-4
Input 7.1 kV Burst signal, common mode
Output 7.1 kV signal, differential mode, line to line
EUT voltage max. 480 V L-L @ 50, 60 Hz
Decoupling 2 x 66 nF built-in
Dimensions 18 x 10 x 8 cm
Weight 2 kg including accessories
Included 2 x test tips and cables, cable to generator
VERI50 EFT
Application 50 Ω calibration load for burst
Input impedance 50 Ω ± 2 %
Input voltage max. 7.1 kV, max. 350 pulses/s @ 7.1 kV
Tolerance up to 100 MHz ± 1 dB, as per standard
Tolerance up to 400 MHz ± 3 dB in the range 100 MHz – 400 MHz
Power dissipation max. 3 W, no EUT power during calibration
Ratio 50 Ω DSO input 1 :1000, 60 dB
Ratio 1MΩ DSO input 1:500, 54 dB
Dimensions 15.5 x 2.5 x 2.5 cm
Weight 0.2 kg
Other applications IEC 61000-4-18
VERI-CP-EFT
Application calibration transducer plate for CN-EFT1000
Plate width 120 ± 1 mm, as per norm
Plate length 1050 ± 5 mm, as per norm
Insulation 5/50 ns up to 7.1 kV
Insulation 1.2/50 µs minimum 2.5 kV
Dimensions 113 x 7 x 5 cm
Weight 1 kg
Requires VERI50 EFT
Other applications IEC 61000-4-18
EFT-INSULATION
Application EUT support for burst test
Number of plates 2 pieces
Height of plates 10 cm
Surface of both plates 100 x 40 cm
Maximum EUT weight 50 kg
Dimensions 100 x 40 x 10 cm (both plates)
Weight 16 kg (both plates)
Other applications IEC 61000-4-9, IEC 61000-4-10
WARNING-LAMP
Cable length 5m
Dimensions diameter 7x cm x 25 cm
Weight 0.5 kg
EMERGENCY-STOP
Cable length 5m
Dimensions 8 cm x 8 cm x 10cm
Weight 0.3 kg
TC-ST
Insulation withstand 36 kV 1.2/50 µs impulse
max. EUT dimension 20 cm x 20 cm x 30 cm
Material 7 mm acrylic glass
Safety Features locked cover during test
External Warning Lamp M12 / 5-pol connector
Dimension 47 cm x 43.5 cm x 25.4 cm
Weight 8.5 kg
MF1000-1
Application coil for AC magnetic field, magnetic pulses
Coil type 1 x 1 m, one turn, as per standard
Coil factor 0.87
Magnetic field 50, 60 Hz 1 – 160 A / m, continuous (IEC 61000-4-8)
Duration continuous MF 29999 s
Magnetic pulse 8/20 µs 93 – 1521 A / m or 1855 A / m (IEC 61000-4-9)
with 2 coils up to 1000 A / m (≥ 5kV surge)
Magnetic pulse DOW up to 220 A / m (IEC 61000-4-10)
Dimensions 120 x 100 x 10.5 cm
Weight 6 kg
MF1STAND
Application stand for MF1000-1
Height 0.2 - 1.8 m
Adjustable on all 3 directions, 360˚
Dimensions 60 x 50 cm stand
Weight 16 kg
MF1000-2
Application coil for AC magnetic field, magnetic pulses
Coil type 1 x 2.6 m, one turn, as per standard
Coil factor 0.66
Magnetic field 50, 60 Hz 1 – 110 A / m, continuous (IEC 61000-4-8)
Duration continuous MF 29999 s
Magnetic pulse 8/20 µs 67 – 1103 A / m or 1345 A / m (IEC 61000-4-9)
Magnetic pulse DOW up to 110 A / m (IEC 61000-4-10)
Dimensions 260 x 100 x 10.5 cm
Weight 24 kg
MF1000-3
Application coil for short duration magnetic field
Coil type 1 x 1 m, one turn, as per standard
Coil factor 0.87
Magnetic field 50, 60 Hz 150 – 500 A / m, continuous (IEC 61000-4-8)
Duration continuous MF 29999 s
Magnetic field 50, 60 Hz 150 – 1100 A / m, short term (IEC 61000-4-8)
Duration short term MF 3s
Dimensions 100 x 100 x 13 cm
Weight 18 kg
VAR-EXT1000
VERI-DIPS
Application calibration of inrush current before dips test
Current range 1 A – 1 kA
Capacitance 1700 µF ± 20 %
Discharge resistor 4.7 kΩ ± 10 %
DSO output 1 V = 400 A, BNC connector
Dimensions 20 x 10 x 11 cm
Weight 1 kg
DIPS100E
Application calibration of dips switch time
Resistor 100 Ω ± 5 %, non-inductive
Power max. 1 kW
Impulse insulation 3 kV
Dimensions 65 x 12 x 8 cm
Weight 3.8 kg
PFS32
Application 3P AC dips generator, IEC 61000-4-34
EUT AC voltage L-L 3 x 200 V – 3 x 480 V, 50 or 60 Hz
EUT AC current max. 3 x 32 A
Output connection star or delta connection possible
Overcurrent protection 50 A per phase continuous, 220 A short term
Rise & fall time into 100 Ω 1 – 5 µs for AC
Inrush current capability > 500 A
Dip / interruption duration 50 µs – 60 s
Synchronization 0 – 359˚, resolution 1˚
Output monitor BNC 3 x I (1 V : 100 A), 3 x U (1 V : 100 V)
Dimensions 19“ unit, 4 UH
Weight 26 kg
Controlled by IMU-MGS D, IMU-MGE D
Requires SRC32-18UH, SRC32-AMD1/AMD1-36UH
SRC32-AMD1-18UH / 36UH
Application 3P tapped transformer for PFS32
Construction type automatic switch between dip levels
EUT AC voltage L-L 3 x 200 V – 3 x 480 V, 50 or 60 Hz, Y or Δ
EUT AC current max. 3 x 32 A at nominal voltage
Overcurrent protection 50 A per phase continuous, 220 A short term
Dip levels 0 %, 40 %, 50 %, 70 %, 80 %, selectable
Dimensions 19” rack (wheels), 18 UH
Weight 317 kg
Alternative rack 36 UH rack available (ask for details)
Controlled by PFS32
PFS75
Application 3P AC dips generator, IEC 61000-4-34
EUT AC voltage L-L 3 x 200 V – 3 x 480 V, 50 or 60 Hz
EUT AC current max. 3 x 75 A
Output connection star or delta connection possible
Overcurrent protection 220 A short term (< 2 s)
Rise & fall time into 100 Ω 1 – 5 µs for AC
Inrush current capability > 1000 A
Dip / interruption duration 50 µs – 60 s
Synchronization 0 – 359˚, resolution 1˚
Output monitor BNC 3 x I (1 V : 100 A), 3 x U (1 V : 100 V)
Dimensions 19“ unit, 8 UH
Weight 40 kg
Controlled by IMU-MGS D, IMU-MGE D
Requires SRC75-18 UH
PFS75-690V
Application 3P AC dips generator, IEC 61000-4-34
EUT AC voltage L-L 3 x 200 V – 3 x 690 V, 50 or 60 Hz
EUT AC current max. 3 x 75 A
Output connection star or delta connection possible
Overcurrent protection 220 A short term (< 2 s)
Rise & fall time into 100 Ω 1 – 5 µs for AC
Inrush current capability > 1000 A
Dip / interruption duration 50 µs – 60 s
Synchronization 0 – 359˚, resolution 1˚
Output monitor BNC 3 x I (1 V : 100 A), 3 x U (1 V : 100 V)
Dimensions 19“ unit, 8 UH
Weight 54 kg
Controlled by IMU-MGS D, IMU-MGE D
Requires SRC75-690V
SRC75-690V-18UH
Application 3P tapped transformer for PFS75-690V
Construction type automatic switch between dip levels
EUT AC voltage L-L 3 x 200 V – 3 x 690 V, 50 or 60 Hz, Y or Δ
EUT AC current max. 3 x 75 A at nominal voltage
Overcurrent protection 220 A short term (< 2 s)
Dip levels 0 %, 40 %, 70 %, 80 %, selectable
Inrush current > 1000 A
Dimensions 19” rack (wheels), 36 UH
Weight 332 kg
Controlled by PFS75-690V
EXT-IMU C-SHORT
Application extension for IMU -C module, power tests
Output impedance 50 Ω ± 10 %
Output voltage 30 V – 330 V for 1000 seconds
Power frequencies DC, 16.67 Hz, 50 Hz, 60 Hz
Disturbance duration 1 s – 1000 s to full test level
Dimensions 19“ unit, 4 UH
Weight 54 kg
Controlled by IMU-MGS C, IMU-MGE C
Requires PS3, CN16 or CN16DC or CN16T
PS3
Application 1P power source for IEC 61000-4-16, 300V test
Type programmable, pre-programmed buttons
Input AC 100 V – 240 V, 47 – 63 Hz
Output voltage AC 50 – 250 V, DC 24 – 350 V
Output frequency DC – 400 Hz
Output current max. 16A @ 115V / 60Hz, 10 A @ 230V / 50Hz
Output power max. 3 kW or 3 kVA
Dimensions 19“ unit, 2 UH
Weight 18 kg
Controlled by IMU-MGS C or IMU-MGE C for power test
Requires EXT-IMU C-SHORT, RS485-RS232 ADAPT.
Other applications IEC 61000-4-19, IEC 61000-4-29
VERI50 EFT
See VERI50 EFT technical specification.
VERI1K EFT
See VERI1K EFT technical specification.
VERI01 OSI
Application 0.1 Ω shunt for 3, 10, 30 MHz DOW SC current
Input impedance 0.1 Ω ± 2 %
Input voltage max. 6.3 kV
3 dB bandwidth > 400 MHz
Power dissipation max. 3 W, max. 1000 spikes/s @ 4.4 kV
Measurement ratio 1 V @ 10 A in 1 MΩ ± 2 %
Dimensions 8.5 x 2.5 x 2.5 cm
Weight 0.1 kg
At CDN output Calibration adapter delivered with DOW-CG1
VERI10-50
Application 10 / 50 Ω calibration balun for SLAVE SMART
Input impedance 2 x 10 Ω ± 10 %
Output impedance 2 x 50 Ω ± 10 %
Frequency response 2 kHz – 200 kHz ± 0.5 dB
Damping limit 10 dB @ 2 kHz to 50 dB@ 150 kHz, linear var.
Dimensions 24 x 10 x 8 cm
Weight 1.2 kg
See PS3 technical specification. 1 x PS3 power source needed for DC interruptions,
2 x PS3 power sources needed for DC dips.
EXT-IMU D-29D
EXT-IMU D-29I
Package including adjustment and calibration of IMU-MGS D, IMU-MGE D together
with 1 x PS3 and 1 x RS485-RS232 ADAPTER in order to ensure compliance of the test
system with IEC 61000-4-29 requirements for DC interruptions.
SOFTWARE
TEMA3000
Application modular control software for IMU / DOW system
License 1 license for 1 generator
TEMA3000 basic remote control of generator, single tests
TEMA3000-SEQUENCE more single tests linked in a sequence
TEMA3000-REPORT generation of customized test reports
TEMA3000-DSO control of DSOs via Ethernet, data in report
TEMA3000-LIBRARY collection of norm pre-programmed tests
Compatibility Windows XP, 7, 8, 10
For generators DOW, IMU Series, other
IMMUNITY TESTS
Transient Test Systems for all EMC tests on electronic equipment. ESD,
EFT, surge, AC dips, AC magnetic field, surge magnetic field, common
mode, damped oscillatory and DC dips. According to IEC and EN 61000-
4-2, -4, -5, -8, -9, -10, -11, -12, -13, -14, -16, -18, -19, -29.
LIGHTNING TESTS
Impulse test equipment and accessories for aircraft, military and tel-
ecom applications. Complete solutions for RTCA / DO-160 and EURO-
CAE / ED-14 for indirect lighting on aircraft systems, MIL-STD-461 tests
CS106, CS115, CS116, CS117, CS118 and Telecom, ITU-T .K44 basic and
enhanced tests for impulse, power contact and power induction.
COMPONENT TESTS
Impulse generators for testing varistors, gas discharge tubes (GDT),
surge protective devices (SPDs), X / Y capacitors, circuit breakers,
electricity meters, protection relays, insulation material, suppressor
diodes, connectors, chokes, fuses, resistors, emc-gaskets, cables, etc.
EMISSION MEASUREMENTS
Measurement of Harmonics and Flicker in 1-phase and 3-phase
electrical and electronic products according to IEC /EN 61000-3-2 and
61000-3-3 . HARCS Immunity software adds interharmonic tests, voltage
variation according to IEC/EN 61000-4-13, -4-14.
SYSTEM AUTOMATION
A full range of accessories enhance the test systems. Test cabinets, test
pistols, adapters and remote control software, simplify interfacing with
the EUT. Programmable PSU, EMC hardened for frequencies from 16.7Hz
to 400Hz. PS3-SOFT-EXT complies with IEC / EN 61000-4-14 and -4-28.
SERVICE
Our committment starts with a quality management system backing up
our ISO 17025 accreditation. With the SCS number 146, EMC PARTNER
provide accredited calibration and repairs. Our customer support team
is at your service!
For further information please do not hesitate to contact your local EMC PARTNER AG representative.
Visit our website for more information and contact details.
For further information please do not hesitate to contact your local EMC PARTNER AG repre-
sentative.
www.emc-partner.com
Your will find a complete list of representatives on our website
www.emc-partner.com
EMC PARTNER AG
Baselstrasse 160
CH - 4242 Laufen
Information and specifications in this document are an indication of capability only. Version 4.0 .Subject to change without notice. EMC PARTNER AG publishes only the
English version of this document. Translation into other languages is not guaranteed to be a true representation of content or specification.
© by EMC PARTNER AG. No changes or reproduction without permission of EMC PARTNER AG allowed.