SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis
Standard operating procedure of Out of Limit results of Environmental Monitoring and Water Analysis.
1.0 PURPOSE
To lay down the procedure for handling of out of limits (OOL) results of environmental monitoring and water system monitoring results.
2.0 SCOPE
Applicable to handling and conducting investigations when out of limit results are obtained in environmental monitoring and water system monitoring
of production and microbiology laboratory facilities.
3.0 RESPONSIBILITY
3.1 Quality Control Microbiology
Microbiology Officer is responsible to notify Microbiology Head or his designee when and alert or action limit is obtained.
Head - Microbiology or his designee is responsible to notify the QA and concerned departments and initiate the investigation in the laboratory and
concerned department.
To implement any corrective action and preventive action
3.2 Quality Assurance Department
To participate in investigations
To review and approve investigation reports
To review action taken reports
3.3 Production / Engineering Department
To participate in investigations
To implement necessary corrective and preventive actions
4.0 ACCOUNTABILITY
Head of Department
5.0 PROCEDURE
5.1 Notification and allotment of number for Out of Limit Result
5.1.1 On obtaining an out of the limit result, the microbiologist shall notify the Microbiology In - Charge or his designee and they shall notify the QA
and concerned department.
5.1.2 The details of Out of Limit results shall be entered in Log Book and a number shall be allotted as M-OOL/YY/ZZZ, where
M-OOL : Microbiological Monitoring – Out Of Limit Result
YY : Represents last two digits of current year
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 1/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
• If the sample location is in critical area, then review the entry exit logs for number of persons present in the area, their duration of stay in the area
and their personal hygiene and training/qualification status.
• Review of records of physical conditions like pressure differentials, temperature and relative humidity of the subject area on the day, days before
and after the occurrence.
• Contact Engineering department for any discrepancies in the functioning of HVAC and other systems or any maintenance activities undertaken or
due for maintenance
• If any discrepancies observed during the review, determine if it has any impact on the observed results.
• Record all the observations in the investigation report.
5.2.4.5 Based on the information gathered, determine if follow up monitoring is required or not as per Investigation report and proceed accordingly.
5.2.4.6 If the results of the follow-up monitoring performed are not satisfactory elevate the investigation to out of action limit investigations.
5.2.4.7 Review the identification results and verify if it is normal micro flora of the area. If the isolate is different or objectionable, initiate necessary
corrective actions.
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 3/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 4/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
5.3 Handling and Investigation of Out of Limits Results in Non Viable Monitoring
Note: This section is applicable for handling, out of limit results of Microbiological lab facility.
5.3.1 On obtaining any out of limit results during non-viable monitoring, immediately do the following:
5.3.2 Check the instrument is operating properly and any disturbance or changes in room conditions is observed.
5.3.3 Check activities (specifically for particle or aerosol generating or disturbance to particle counter probe) performed around the sample location
during the time of out of limit result and evaluate if it has any effect on the reported result.
5.3.4 Verify that the instrument used was within calibration and testing performed as per procedure. If appropriate perform the zero count of the
particle counter.
5.3.5 Re-sample the location after the conditions are restored and verify the results. Record the noted observation the report. If the results of re-
sample are conforming to limits, then no further action is required. If the results are still non-conforming proceed to 5.3.6
5.3.6 Review the trend for the sample location / room in question and results of other sample locations performed on the day.
5.3.7 Review department cleaning logs, room differential pressure records, number for personal in the room at the time of testing, number of
equipment and their operation status and other activities for any discrepancies.
5.3.8 Contact Engineering / Maintenance department and review logs for any discrepancies in the functioning and maintenance of HVAC and other
systems or any maintenance activities undertaken.
5.3.9 Evaluate the information gathered and determine if it has an impact on the results observed.
5.3.10 Based on the information gathered, evaluate the actions to be performed and perform re-sampling of the concerned location or room as per
Investigation report and proceed accordingly.
5.3.11 If the re-sample results conform to limits then no further action is required.
5.3.12 If the re-sample result does not conform to limits, then carry out further investigation for determining the root cause. Following activities can be
performed to determine the root cause:
• Extensive cleaning of area
• Air Velocity verification of HEPA filters
• HEPA filter integrity testing
• Air flow studies
5.4 Handling and Investigation of Out of Limits Results in Personnel Monitoring
5.4.1 Observe the plates under incubation (sampled after the date of sampling showing out of limit results) of the sample same person(s) showing
out of limits and check of for any out of limit results. Inform the observations to Head Microbiology or his designee.
5.4.2 Verify the negative control plate incubated along with test samples for any contamination. Inform the observation to Head Microbiology or his
designee.
5.4.3 Observe the other plates of environmental monitoring performed on the day of out of limit occurrence and compare the colonies with plates
showing out of limit results.
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 5/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
5.4.4 Process the out of limit results for identification along with morphologically similar colonies from environmental monitoring plates if any as
follows:
5.4.4.1 Perform gram staining of morphologically similar colonies and identification of representative isolatess based on Gram Staining Results as
per SOP
5.4.4.2 In case of out of limit results in Grade A and Grade B areas perform identification of all isolates as per SOP and also by DNA sequencing.
5.4.5 Laboratory Investigation
5.4.5.1 Interview the microbiologist who performed sampling and verify whether the sampling was performed as per SOP and if any deviations
observed during the sampling/testing.
5.4.5.2 Verify all the media used were within their shelf life and review their preparation records for results of pre-incubation and sterility checks.
5.4.5.3 If contamination in negative control plates is observed or laboratory investigation reveals fault in sampling, discrepancies in results of media
pre-incubation and sterility check are not satisfactory, then the occurrence of out of limit results could be attributed to laboratory/sampling error.
5.4.6 Facility Investigation
5.4.6.1 Review the previous data of concerned person for any out of limit results.
5.4.6.2 Verify the personal hygiene and health status of concerned person. Review the medical and training records of the concerned person.
5.4.6.3 Review the environmental monitoring data of day and days before and after the day of occurrence for out of limit results.
5.4.6.4 Review cleaning / disinfection logs, entry exits logs and other activities of subject area for any discrepancies.
5.4.6.5 Review of records of physical conditions like pressure differentials, temperature and relative humidity of the subject area on the day, days
before and after occurrence.
5.4.6.6 Review the garment preparation and sterilization records for any discrepancies.
5.4.7 Based on the information gathered, determine the actions to performed as per Investigation and Decision Flow Chart in (Annexure I) and
proceed accordingly.
5.4.8 Review the activities performed the concerned individual and if he has performed critical aseptic operations, critically review the microbiological
results of the concerned batch.
5.5 Handling and Investigation of Out of Limits Results in Microbiological Analysis of Water
5.5.1 Observe the plates of same type of water sampled and analyzed on the same day and those under incubation (sampled after the date of
sampling showing out of limit results) of the sample location / sample type showing out of limits and check of for any out of limit results. Inform the
observations to Head Microbiology or his designee.
5.5.2 Verify the negative control plate incubated along with test samples for any contamination. Inform the observation to Head Microbiology or his
designee.
5.5.3 Process the out of limit results for identification as follows:
• In case of out of alert limit results, perform gram staining of morphologically similar colonies and identification of representative isolates based on
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 6/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
• Verify the materials used for sampling and testing was properly sterilized and handled.
• Verify all the media used were within their shelf life and review their preparation records for results of pre-incubation and sterility checks.
• If contamination in negative control plates is observed or laboratory investigation reveals fault in sampling and testing results of media pre-
incubation and sterility check are not satisfactory, then the occurrence of out of limit results could be attributed to laboratory/sampling error.
5.5.5.5 Water System / Facility Investigation
Contact Engineering and Production Department for any discrepancies in the functioning of water systems or any maintenance undertaken. Verify
the operation and sanitization log books for concerned system / area.
If the out of limit result is observed in only one sample location and rest of the system is conforming to specifications, then verify the sample location
for any discrepancies in the sample/user point and the location.
If the water from the specific location was used for batch manufacturing, then verify the results in-process and finished product samples of concerned
batch(es).
• Bio-burden results of bulk sample before filtration
• Microbiological tests results of Oral Solid Dosages
5.5.5.6 Evaluate the information gathered and determine if it has any impact on the observed results.
5.5.5.7 Based on the information gathered, determine the actions to be initiated as per Investigation report and proceed accordingly.
• Review the identification results and verify if it is normal micro flora. If isolate is different, then include it micro flora stock for use in different tests.
• If isolate identified is of objectionable, then investigate the possible source of contamination and take necessary corrective and preventive actions.
• If the results of follow up sampling are satisfactory after carrying out corrective actions (if any) conclude the investigation.
• If the results of follow up sampling are not satisfactory carry out further investigation take necessary actions accordingly. Following activities can be
performed to determine the root cause:
- Sampling and analysis at different stages of generation and distribution system to identify the contamination.
- Sanitization of generation, storage, distribution and heat exchangers as applicable.
- Verification of air vent filters.
- Verification of gaskets, valves and other components.
5.5.5.8 The system can be released for use after obtaining satisfactory results for consecutive three days.
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 8/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
5.6.4 Verify the test is performed properly as per procedure for any analyst error during testing.
5.6.5 If any discrepancy is observed in glassware used for sampling and testing, then take necessary corrective actions and arrange for re-sampling
from particular sample point from glassware conforming to requirements.
5.6.6 If the any discrepancy is observed in chemicals, reagents or instruments, then take necessary corrective actions and retest using original
sample if available or with a fresh sample.
5.6.7 If the analyst error is observed, then the second analyst shall perform the test.
5.6.8 If the results of the resample conform to specifications, then record the results and water may be released.
5.6.9 If the no assignable cause is determined above or the test results show non-conformance on re-sampling and testing, then inform the
concerned department and QA.
5.6.10 Perform the investigation to determine the root cause and take necessary corrective actions. Following activities can be performed to
determine the root cause:
• Testing of input water and at different stages in the treatment and generation system.
• Cleaning and sanitation of generation, storage and distribution system.
5.6.11 The system can be released for use after obtaining satisfactory results for consecutive three days.
5.6.12 For investigation of chemical analysis follow the chart.
5.7. Investigation of Out of Limit Results in TOC
5.7.1 Investigation out of Alert Limit Results
5.7.1.1 If TOC results exceed the above Alert level then the following actions shall be initiated.
5.7.1.2 Any samples exceeding the alert limits shall be immediately informed to the Head Microbiology & QA and do not discard the original left over
the sample (if any)
5.7.1.3 Review the data for the sample location /system in question for any previous instances of out of limit results in last three months.
5.7.1.4 If the last data indicates previous occurrences of out of limits, then review the previous investigation reports to determine any similarities.
5.7.2 Laboratory Investigation
5.7.2.1 Interview the microbiologist who performed sampling and verify whether the sampling was performed as per sop and if any deviation
observed during the sampling/testing.
5.7.2.2 Verify the glassware used for sampling was properly cleaned.
5.7.2.3 Verify that sample was intact during transportation.
5.7.2.4 Verify that the instrument used was within calibration and testing performed as per procedure.
5.7.2.5 If laboratory investigation reveals a fault in sampling/glassware used/and transportation then the occurrence of out of limit results could be
attributed to laboratory /sampling error.
• An immediate repeat test of the original sample together with an additional sample from the same location shall be performed.
• If the TOC re-sample show results over the alert limit then additional user points shall be immediately sampled and tested.
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 9/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
• If results suggest that only one point is affected and it is an isolated incident then the result will be recorded and used for trending analysis &
investigate for corrective and preventive action.
• If the retest results from the additional user points are found out of Alert limit, the results shall be informed to Head QA, Production and Engineering
department to carry out the detailed investigation and take immediate corrective and preventive actions.
5.8
• If results suggest that only one point is affected and it is an isolated incident
• Investigation shall be initiated to identify the root cause.
• If the retest results from the additional user points are found out of Action limit, the results shall be informed to Head QA, Production and
Engineering department to carry out the detailed investigation and take immediate corrective and preventive actions.
• Until and unless the investigation is complete, and immediate corrective actions is completed no further batches will be manufactured.
• The system can be released for use after investigation & obtaining the satisfactory results.
• A trend of OOL shall be prepared and review for repetitive nature & Effectiveness of CAPA on half yearly basis.
6.0 ABBREVIATIONS
6.1 SOP - Standard Operating Procedure
6.2 OOL - Out of Limit
6.3 OOS - Out of Specification
6.4 TOC - Total Organic Carbon
6.5 BET - Bacterial Endotoxin Test
6.6 CAPA - Corrective and preventive action
Annexure I
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 10/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
Date of Sampling :
Sampling Room/location ID :
Sample ID :
Type of water :
A.R. No. :
Sampled By :
Test Date :
Tested by :
Observed by/Date :
Date of Mfg. :
Phase I investigation
1.0 Observation of plates under Incubation (Sampled after the date of sampling showing out of limit results of the particular sample location)
Remarks/Summary:
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 11/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
(a) Analyst trained & qualified for the particular Yes No N/A
Procedure/technique as per the SOP
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 12/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
4.4 (a) Autoclave validation Status Date of Validation Due date of Validation
Remarks/Summary:
4.5 Bio safety cabinet /LAF used for media preparation Date of Validation Due date of Validation
4.6 Review of Biosafety cabinet /LAF records in which media is poured, for velocity, Diff. Pressure & validation document etc.
Remarks/Summary:
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 13/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
(e) Review of Incubator Data used for pre incubation Date of Validation Due date
of
Validation
Remarks/Summary:
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 14/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
8.0 If assignable cause(s) identified, describe the cause(s) in details and give justification:
QC-Supervisor Head of QC
(Sign. & Date) (Sign. & Date)
Sampled by
Sampling Date
Observations during resampling
14.0 Signatures
Head of QC Head of QA
(Sign. & Date) (Sign. & Date)
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 15/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
16.0 Signatures
OC Supervisor Head of QC
(Sign. & Date) (Sign. & Date)
18.0 Conclusion
Head of QC Head of QA
(Sign. & Date) (Sign. & Date)
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 16/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
Phase II investigation
1.0 Review of Sanitization / Operation Record of concern system
Remark/Summary
2.0 Interview of Water System /Production Personnel for any discrepancies in the functioning of water system or any maintenance undertaken
Remark/Summary
Remark/Summary
4.0 If one Sample location is out of limit then verify the sample location for any discrepancies
Remark/Summary
Remark/Summary
Remark/Summary
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 17/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
10.0 If assignable cause(s) identified, describe the cause(s) in details and give justification
13.0
https://www.pharmaguideline.com/2010/01/sop-for-procedure-for-handling-of-out.html 18/19
5/16/23, 3:42 PM SOP for Handling of Out of Limit Results in Environmental Monitoring and Water Analysis : Pharmaguideline
(Sign. & Date) (Sign. & Date) (Sign. & Date) (Sign. & Date)