HT - Part Probe Calibration
HT - Part Probe Calibration
1. Set G54: Begin by setting the G54 work coordinate system. And Hit Cycle Start
FIGURE 2 FIGURE 1
1
3. Create and Load Part Probe:
a. Create a part probe (3D probe) and load it into the spindle.
4. Stylus Indication:
a. Runout Adjustment: Use a magnetic base and indictor. Utilize
four set screws to finely adjust the probe stylus for minimal
runout. (See Figure3)
b. Runout Verification: Verify the runout, aiming for a range of
2 to 4 micrometers.
FIGURE 3
FIGURE 4
FIGURE 5
2
6. Measurement Work Preparation:
a. Access the "MEAS. WORKP." (Measurement Workpiece) section.
b. Probe Calibration: Initiate the probe calibration process.
3
7. Ring Gauge Setup:
a. Place a ring gauge onto the table.
b. Position the probe in the middle and inside of the ring gauge.
c. Meas Workp. > Calibrate Probe > Radius.
d. Enter in the diameter in the field. Hit Cycle Start.
8. The Part Probe is Calibrated
4
Setting up a Part Probe for a Positive tool length Strategy with Sphere
1. Repeat Steps 1 through 6 for the probe length.
2. Mount a Tooling ball on the table and position it over the ball.
3. Meas. Workp. > Calibrate Probe > Comp to Sphere
4. Enter the Diameter into the field and hit cycle start.
1. Determine the fixed datum for reference (typically the table surface) this will be the zero location.
2. Create and load a part probe into the spindle
3. Make sure the active WC offset is set to zero
4. Manually set the tool length offset of the part probe to the table surface. (this will be a large negative number)
5. Select the length calibrate function under the calibrate probe key in Measure Workpiece
6. Have the probe positioned just above the table surface within the measuring range of MD51751
7. Set the Z0 value in the length calibrate probe screen to zero
8. Press cycle start to activate the calibration routine (this will cause the Z axis to move down)
9. An updated Tool Length offset should now be seen for the Probe