Description Notes:
Proximity Probe and Extension Cable
1. 1-metre systems do not use an extension cable.
The 3300 XL probe and extension cable also reflect
improvements over previous designs. A patented 2. Proximitor sensors are supplied by default from the
TipLoc* molding method provides a more robust factory calibrated to AISI 4140 steel. Calibration to
other target materials is available upon request.
bond between the probe tip and the probe body.
The probe’s cable incorporates a patented CableLoc* 3. Consult Bently Nevada* Applications Note,
design that provides 330 N (75 lbf) pull strength to Considerations when using Eddy Current Proximity
more securely attach the probe cable and probe tip. Probes for Overspeed Protection Applications, when
considering this transducer system for tachometer or
You can also order 3300 XL 8 mm probes and overspeed measurements.
extension cables with an optional FluidLoc* cable
4. 3300 XL 8 mm components are both electrically and
option. This option prevents oil and other liquids
physically interchangeable with non-XL 3300 5 mm
from leaking out of the machine through the cable’s and 8 mm components. Although the packaging of
interior. the 3300 XL Proximitor Sensor differs from its
Connectors predecessor, its design fits in the same 4-hole
mounting pattern when used with the 4-hole
The 3300 XL probe, extension cable, and Proximitor mounting base, and will fit within the same mounting
sensor have corrosion-resistant, gold-plated space specifications (when minimum permissible
ClickLoc* connectors. These connectors require only cable bend radius is observed).
finger-tight torque (the connectors will "click" when 5. Mixing XL and non-XL 3300-series 5 mm and 8 mm
tight), and the specially-engineered locking system components limits system performance to the
mechanism prevents the connectors from loosening. specifications for the non-XL 3300 5 mm and 8 mm
These connectors require no special tools for Transducer System.
installation or removal. 6. The 3300-series 5 mm probe (refer to Specifications
You can order the 3300 XL 8 mm probes and and Ordering Information p/n 141605-01) uses
smaller physical packaging, but does not reduce the
extension cables with connector protectors already side view clearances or tip-to-tip spacing
installed. We can also supply connector protectors requirements as compared to an 8 mm probe. It is
separately for field installations (such as when an used when physical (not electrical) constraints
application must run the cable through restrictive preclude the use of an 8 mm probe. When your
conduit). We recommend connector protectors for application requires narrow side view probes, use the
all installations to provide increased environmental 3300 NSv* Proximity Transducer System (refer to
protection8. Specifications and Ordering Information p/n 147385-
01).
Extended Temperature Range Applications
7. 8 mm probes provide a thicker encapsulation of the
An extended temperature range (ETR) probe and ETR probe coil in the molded PPS plastic probe tip. This
extension cable are available for applications in results in a more rugged probe. The larger diameter
which either the probe lead or extension cable may of the probe body also provides a stronger, more
exceed the standard 177 °C (350 °F) temperature robust case. We recommend that you use 8 mm
probes when possible to provide optimal robustness
specification. The ETR probe has an extended
against physical abuse.
temperature rating for up to 218 °C (425 °F). The ETR
extension cable rating is up to 260 °C (500 °F). Both 8. Each 3300 XL extension cable includes silicone tape
the ETR probe and cable are compatible with that you can use instead of connector protectors. We
do not recommend silicone tape for applications that
standard temperature probes and cables, for will expose the probe-to-extension cable connection
example, you can utilize an ETR probe with the to turbine oil.
330130 extension cable. The ETR system uses the
standard 3300 XL Proximitor Sensor. Note that when
you use any ETR component as part of your system,
the ETR component limits the system accuracy to
the accuracy of the ETR system.
Page 2 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Specifications Nominal
Extension Cable
Unless otherwise noted, the following specifications DC Resistance
are for a 3300 XL 8 mm Proximitor Sensor, extension
Resistance (RCORE) from Center
cable and 8 mm probe between +18 °C and +27 °C
Conductor to Center Conductor
(+64 °F to +80 °F), with a -24 Vdc power supply, a 10
kΩ load, an AISI 4140 steel target, and a probe Length of Extension
gapped at 1.27 mm (50 mils). Performance RCORE (Ω)
Cable (m)
characteristics apply to systems that consist solely
of 3300 XL 8 mm components. The system accuracy 3.0 0.66 ± 0.10
and interchangeability specifications do not apply to 3.5 0.77 ± 0.12
transducer systems that are calibrated to any target 4.0 0.88 ± 0.13
other than our AISI 4140 steel target. 4.5 0.99 ± 0.15
7.0 1.54 ± 0.23
Electrical 7.5 1.65 ± 0.25
Proximitor 8.0 1.76 ± 0.26
Sensor Input 8.5 1.87 ± 0.28
Accepts one non-contacting
3300-series 5 mm, 3300 8 mm or Resistance (RJACKET) from Outer
3300 XL 8 mm Proximity Probe Conductor to Outer Conductor
and Extension Cable.
Length of Extension
Power RJACKET (Ω)
Cable (m)
Requires -17.5 Vdc to -26 Vdc 3.0 0.20 ± 0.04
without barriers at 12 mA 3.5 0.23 ± 0.05
maximum consumption, -23 Vdc 4.0 0.26 ± 0.05
to -26 Vdc with barriers. 4.5 0.30 ± 0.06
Operation at a more positive 7.0 0.46 ± 0.09
voltage than -23.5 Vdc can result 7.5 0.49 ± 0.10
in reduced linear range. 8.0 0.53 ± 0.11
Supply 8.5 0.56 ± 0.11
Sensitivity
Less than 2 mV change in output Extension Cable
voltage per volt change in input Capacitance
voltage.
69.9 pF/m (21.3 pF/ft) typical
Output
Field Wiring
Resistance
0.2 to 1.5 mm2 (16 to 24 AWG) .
50 Ω
Recommend using 3-conductor
Nominal Probe shielded triax cable and tinned
DC Resistance field wiring. Maximum length of
305 metres (1,000 feet) between
Resistance (RPROBE) from Center
the 3300 XL Proximitor Sensor
Conductor to Outer Conductor
and the monitor. See the
Probe Length RPROBE (Ω) frequency response graphs in
through Figure 13 (pages 28 and
0.5 7.45 ± 0.50 29) for signal rolloff at high
1.0 7.59 ± 0.50 frequencies when using longer
1.5 7.73 ± 0.50 field wiring lengths.
2.0 7.88 ± 0.50
5.0 8.73 ± 0.70
9.0 9.87 ± 0.90
Specifications and Ordering Information
Part Number 141194-01
Rev. P (11/11)
Page 3 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Linear Range Deviation from best fit straight line (DSL)
2 mm (80 mils). Linear range Standard
begins at approximately 0.25 mm 5- or 1-metre
(10 mils) from target and is from System:
0.25 to 2.3 mm (10 to 90 mils)
Less than ±0.025 mm (±1 mil) with
(approximately –1 to –17 Vdc).
components at 0 °C to +45 °C
Recommended (+32 °F to +113 °F).
Gap Setting for
Standard
Radial Vibration
9-metre
-9Vdc [approximately 1.27 mm System:
(50 mils)]
Less than ±0.038 mm (±1.5 mil)
Incremental with components at 0 °C to +45
Scale Factor °C (+32 °F to +113 °F).
(ISF)
Extended
Standard Temperature
5- or 1- metre Range 5 and
System: 9-metre
Systems:
7.87 V/mm (200 mV/mil) ± 5%
including interchangeability error Less than ±0.038 mm (±1.5 mil)
when measured in increments of with components at 0 °C to +45
0.25 mm (10 mils) over the 80 mil °C (+32 °F to +113 °F).
linear range from 0 °C to +45 °C
Performance
(+32 °F to +113 °F).
Over Extended
Standard Temperatures
9-metre
Standard
System:
5- or 1-metre
7.87 V/mm (200 mV/mil) ± 6.5% System:
including interchangeability error
Over a probe temperature range
when measured in increments of
of –35 °C to +120 °C (-31 °F to
0.25 mm (10 mils) over the 80 mil
+248 °F) with the Proximitor
linear range from 0 °C to +45 °C
sensor and extension cable
(+32 °F to +113 °F).
between 0 °C to +45°C (+32 °F to
Extended +113 °F), the ISF remains within
Temperature ±10% of 7.87 V/mm (200 mV/mil)
Range (ETR) and the DSL remains within
for 5- and ±0.076 mm (±3 mils).
9-Metre
Over a Proximitor sensor and
Systems:
extension cable temperature
7.87 V/mm (200 mV/mil) ± 6.5% range of –35 °C to +65 °C (-31 °F
including interchangeability error to +149 °F) with the probe
when measured in increments of between 0 °C to +45 °C (+32 °F to
0.25 mm (10 mils) over the 80 mil +113 °F), the ISF remains within
linear range from 0 °C to +45 °C ±10% of 7.87 V/mm (200 mV/mil)
(+32 °F to +113 °F). and the DSL remains within
±0.076 mm (±3 mils).
Page 4 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Standard Recommended
9-metre Minimum:
System:
76.2 mm (3 in)
Over a probe temperature range
When gapped at the center of the
of –35 °C to +120 °C (-31 °F to
linear range, the interaction
+248 °F) with the Proximitor
between two separate transducer
sensor and extension cable
systems (cross-talk) will be less
between 0 °C to +45°C (+32 °F to
than 50 mV on shaft diameters of
+113 °F), the ISF remains within
at least 50 mm (2 in) or greater.
±18% of 7.87 V/mm (200 mV/mil)
You should take care to maintain
and the DSL remains within
minimum separation of
±0.152 mm (±6 mils).
transducer tips, generally at least
Over a Proximitor sensor and 40 mm (1.6 in) for axial position
extension cable temperature measurements or 38 mm (1.5 in)
range of –35 °C to +65 °C (-31 °F for radial vibration measurements
to +149 °F) with the probe to limit cross-talk to 50 mV or less.
between 0 °C to +45 °C (+32 °F to Radial vibration or position
+113 °F), the ISF remains within measurements on shaft
±18% of 7.87 V/mm (200 mV/mil) diameters smaller than 76.2 mm
and the DSL remains within (3 in) will generally change the
±0.152 mm (±6 mils). scale factor.
Extended Effects of 60 Hz
Temperature Magnetic Fields
Range 5 and up to 300 Gauss
9-metre
Output Voltage in Mil pp/Gauss
Systems:
5- or
Over a probe and extension cable 9-metre
Gap 1-metre Ext.
temperature range of –35 °C to Proximitor Probe
(mil) Proximitor Cable
+260 °C (-31 °F to +500 °F) with Sensor
Sensor
the Proximitor sensor between 0
°C to +45 °C (+32 °F to +113 °F), 10 0.0119 0.0247 0.0004 0.0004
the ISF remains within ±18% of 50 0.0131 0.0323 0.0014 0.0014
7.87 V/mm (200 mV/mil) and the 90 0.0133 0.0348 0.0045 0.0045
DSL remains within ±0.152 mm
(±6 mils). Compliance and Certifications
Frequency EMC
Response
European Community Directives:
(0 to 10 kHz), +0, -3 dB, with up to EMC Directive 2004/108/EC
305 metres (1000 feet) of field Standards:
wiring. EN61000-6-2
Minimum EN61000-6-4
Target Size Maritime
15.2 mm (0.6 in) diameter (flat ABS 2009 Steel Vessels Rules
target) 1-1-4/7.7, 4-8-3/1.11.1, 4-9-7/13
Shaft Diameter
Minimum:
50.8 mm (2 in)
Page 5 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Hazardous Area Approvals 3300 XL
Note: Multiple approvals for hazardous areas certified by
Proximitor
Canadian Standards Association (C/US) in North America
Sensor, nA:
and by Baseefa for Europe and IEC Ex. II 3 G Ex nA II T4/T5 when
installed per drawing 140979.
Field Wiring Limitations:
T5 @ Ta= -35 °C to +85 °C
Type Gas Capacitance Inductance L/R T4 @ Ta= -51 °C to +100 °C
Approval: Group (µF) (mH) Ratio
3300 XL
(μH/Ω)
8mm probe,
ATEX and ia:
IEC Zone
II 1 G EEx ia IIC, Temperature
0/1 IIC 0.078 0.99 29.2 Classification per Table 1 when
IIB 0.645 7.41 117.0
installed per drawing 142491.
IIA 2.144 15.6 234.0
CSA Div 1 3300 XL
A&B 0.070 1.0 29.2 8mm probe,
C 0.600 5.0 117.0 nA:
D 2.09 11.0 234.0
CSA Div 2 II 3 G EEx nA II, Temperature
All 0.460 100.0 N/A Classification per Table 1 when
installed per drawing 142491.
North America
Brazil
3300 XL
Proximitor 3300 XL
Sensor and Proximitor
probe, ia: Sensor, ia:
Page 6 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Ii = 140 mA Li = 0 H Hazardous Area
Pi = 0.91 W Conditions of
Safe Use:
ATEX:
IEC Ex Follow the conditions of safe use
included on the Declaration of
3300 XL
Conformance sent with each
Proximitor
product.
Sensor, ia:
Canadian
Ex ia IIC T4 (-51°C ≤ Ta ≤ +100°C) /
Standards
T5 (-35ºC ≤ Ta ≤ +85ºC)
Association
Ui= -28V Ci = 0 (CSA):
Ii= 140mA Li =10µH
Division 1 (Intrinsically safe):
Pi= 0.84W
Install per Bently Nevada drawing
3300 XL
141092.
Proximitor
Sensor, nA: Division 2 (non-Incendive): Install
per Bently Nevada drawing
Ex nA II T4 (-51°C ≤ Ta ≤ +100°C) /
140979.
T5 (-35°C ≤ Ta ≤ +85°C)
IECEx:
Ui = -28V
Zone 0 (Intrinsically safe): The
3300 XL
Proximitor Sensor must be
8mm and
installed to minimize the risk of
3300 5mm
impact or friction with other
Eddy Current
metallic surfaces.
Probes, ia:
Zone 2 (non-Incendive): The
Ex ia IIC Temperature
probe must be supplied from a
Classification per Table 1.
voltage-limited source.
Ui = -28V Ci = 1.5 nF
Ii = 140 mA Li = 200 µH
Pi = 0.84 W Mechanical
Probe Tip
Material
3300 XL
8mm and Polyphenylene sulfide (PPS).
3300 5mm Probe Case
Eddy Current Material
Probes, nA:
AISI 303 or 304 stainless steel
Ex nA II for Zone 2 Temperature (SST).
Classification per Table 1.
Probe Cable
Table 1: Probe Temperature Classification
Specifications
Temperature Ambient Temperature Standard
Classification (Probe Only) cable:
T1 -51ºC to +232ºC
T2 -51ºC to +177ºC 75Ω triaxial, fluoroethylene
T3 -51ºC to +120ºC propylene (FEP) insulated probe
T4 -51ºC to +80ºC cable in the following total probe
lengths: 0.5, 1, 1.5, 2, 5, or 9
T5 -51ºC to +40ºC
metres.
Page 7 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Extended Extended
Temperature Temperature
Range cable: Range cable:
75Ω triaxial, perfluoroalkoxy (PFA) 75Ω triaxial, perfluoroalkoxy (PFA)
insulated probe cable in the insulated.
following total probe lengths: 0.5,
Minimum
1, 1.5, 2, 5, or 9 metres.
Cable Bend
Armor Radius:
(optional on
25.4 mm (1.0 in)
both):
Note: 3300 XL 8 mm components are both electrically and
Flexible AISI 302 or 304 SST with physically interchangeable with non-XL 3300 5 mm and 8
FEP outer jacket. mm components when minimum permissible cable bend
radius is observed..
Tensile Strength
(Maximum Connector
Rated): Material:
Page 8 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
5 or 9 metres (including extension Note: Exposing the probe to temperatures below –34 °C (-30 °F)
cable) or 1 metre (probe only). may cause premature failure of the pressure seal.
Environmental Limits
Probe Temperature Range Extension Cable Temperature Range
Operating and Operating and
Storage Storage
Temperature Temperature
Standard Standard
Probe: Cable:
-51 °C to +177 °C (-60 °F to +350 -51 °C to +177 °C (-60 °F to +350
°F) °F)
Extended Extended
Temperature Temperature
Range Probe: Range Cable:
-51 °C to +218 °C (-60 °F to -51 °C to +260 °C (-60 °F to +500
+425°F) for the probe tip; -51 °C to °F)
+260 °C (-60 °F to +500 °F) for the
probe cable and connector. Proximitor Sensor Temperature Range
Operating
Temperature
Specifications and Ordering Information
Part Number 141194-01
Rev. P (11/11)
Page 9 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
-51 °C to +100 °C (-60 °F to +212
°F)
Storage
Temperature
-51 °C to +105 °C (-60 °F to +221
°F)
Relative
Humidity
Less than a 3% change in
Average Scale Factor (ASF) when
tested in 93% humidity in
accordance with IEC standard
68-2-3 for up to 56 days.
Patents:
Components or procedures
described in one or more of the
following patents apply to this
product: 5,016,343; 5,126,664;
5,351,588; and 5,685,884.
Page 10 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Ordering Information Probes 3300 XL 8 mm Proximity Probes, Metric:
330103 3300 XL 8 mm Probe, M10 x 1 thread, without
3300 XL 8 mm Proximity Probes: armor2
330101 3300 XL 8 mm Probe, 3/8-24 UNF thread,
without armor2 330104 3300 XL 8 mm Probe, M10 x 1 thread, with
armor2
330102 3300 XL 8 mm Probe, 3/8-24 UNF thread, with
armor2 Part Number-AXX-BXX-CXX-DXX-EXX
A: Unthreaded Length Option Note: Unthreaded length must be at least 20 mm less than the
case length.
Note: Unthreaded length must be at least 0.8 inches less than
the case length. Order in increments of 10 mm.
Length configuration:
Order in increments of 0.1 in Maximum unthreaded length: 230
Length configurations: mm
Maximum unthreaded length: 8.8 in Minimum unthreaded length: 0 mm
Minimum unthreaded length: 0.0 in Example: 0 6 = 60 mm
Example: 0 4 = 0.4 in B: Overall Case Length Option
B: Overall Case Length Option Order in increments of 10 mm.
Order in increments of 0.1 in Metric thread configurations:
Threaded length configurations: Maximum length: 250 mm
Maximum case length: 9.6 in Minimum length: 20 mm
Minimum case length: 0.8 in Example: 0 6 = 60 mm
Example: 2 4 = 2.4 in C: Total Length Option
C: Total Length Option 05 0.5 metre (1.6 feet)
05 0.5 metre (1.6 feet) 10 1.0 metre (3.3 feet)
10 1.0 metre (3.3 feet) 15 1.5 metres (4.9 feet)
15 1.5 metre (4.9 feet) 20 2.0 metres (6.6 feet)
20 2.0 metres (6.6 feet) 50 5.0 metres (16.4 feet)
50 5.0 metres (16.4 feet)
Note: 5-metre probes are designed for use with the 5-metre
Note: 5-metre probes are designed for use with the 5-metre Proximitor sensor only.
Proximitor sensor only.
90 9.0 metres (29.5 feet)
90 9.0 metres (29.5 feet) D: Connector and Cable-Type Option
D: Connector and Cable-Type Option 01 Miniature coaxial ClickLoc
01 Miniature coaxial ClickLoc connector with connector
connector with connector protector, standard cable
protector, standard cable 02 Miniature coaxial ClickLoc
02 Miniature coaxial ClickLoc connector, standard cable
connector, standard cable 11 Miniature coaxial ClickLoc
11 Miniature coaxial ClickLoc connector with connector
connector with connector protector, FluidLoc cable
protector, FluidLoc cable 12 Miniature coaxial ClickLoc
12 Miniature coaxial ClickLoc connector, FluidLoc cable
connector, FluidLoc cable E: Agency Approval Option
E: Agency Approval Option 00 Not required
00 Not required 05 Multiple Approvals
05 Multiple Approvals
Page 11 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
3300 XL 8 mm Reverse Mount Probes C: Connector and Cable-Type Option
01 Miniature coaxial ClickLoc
330105-02-12-CXX-DXX-EXX, 3/8-24 UNF threads2
connector with connector
330106-05-30-CXX-DXX-EXX, M10 x 1 threads2 protector, standard cable
02 Miniature coaxial ClickLoc
Option Descriptions connector, standard cable
C: Total Length Option 11 Miniature coaxial ClickLoc
05 0.5 metre (1.6 feet) connector with connector
10 1.0 metre (3.3 feet) protector, FluidLoc cable
15 1.5 metre (4.9 feet) 12 Miniature coaxial ClickLoc
20 2.0 metres (6.6 feet) connector, FluidLoc cable
50 5.0 metres (16.4 feet) D: Agency Approval Option
00 Not required
Note: 5-metre probes are designed for use with the 5-metre 05 Multiple Approvals
Proximitor sensor only.
3300 XL 8 mm Extended Temperature Range
90 9.0 metres (29.5 feet)
D: Connector Option
(ETR) Proximity Probes:
02 Miniature ClickLoc coaxial 330191 3300 XL 8 mm ETR Probe, 3/8-24 UNF thread,
connector without armor
12 Miniature ClickLoc coaxial 330192 3300 XL 8 mm ETR Probe, 3/8-24 UNF thread,
connector, FluidLoc cable with armor
Note: The FluidLoc cable option –12 is not necessary on the vast Part Number-AXX-BXX-CXX-DXX
majority of 330105 and 330106 installations due to the
presence of the probe sleeve. Consider carefully the A: Unthreaded Length Option
application before ordering the FluidLoc cable option for
these probes.
Note: Unthreaded length must be at least 0.8 inches less than
E: Agency Approval Option the case length.
00 Not required Order in increments of 0.1 in
05 Multiple Approvals Length configurations:
Maximum unthreaded length: 8.8 in
3300 XL 8 mm Proximity Probes, Smooth Case: Minimum unthreaded length: 0.0 in
Example: 1 5 = 1.5 in
330140 3300 XL 8 mm Probe without armor1 B: Overall Case Length Option
330141 3300 XL 8 mm Probe with armor1 Order in increments of 0.5 in
Threaded length configurations:
Part Number-AXX-BXX-CXX-DXX Maximum case length: 9.6 in
Minimum case length: 0.8 in
Option Descriptions
Example: 2 5 = 2.5 in
A: Overall Case Length Option C: Total Length Option
Order in increments of 0.1 in 05 0.5 metre (1.6 feet)
Length configurations: 10 1.0 metre (3.3 feet)
Maximum length: 9.6 in 15 1.5 metre (4.9 feet)
Minimum length: 0.8 in 20 2.0 metres (6.6 feet)
Example: 2 4 = 2.4 in 50 5.0 metres (16.4 feet)
B: Total Length Option
Note: 5-metre probes are designed for use with the 5-metre
05 0.5 metre (1.6 feet) Proximitor sensor only.
10 1.0 metre (3.3 feet)
15 1.5 metres (4.9 feet) 90 9.0 metres (29.5 feet)
20 2.0 metres (6.6 feet) D: Agency Approval Option
50 5.0 metres (16.4 feet) 00 Not required
05 Multiple Approvals
Note: 5-metre probes are designed for use with the 5-metre
Proximitor sensor only.
Page 12 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
3300 XL 8 mm Extended Temperature Range 3300 XL 8 mm Extended Temperature Range
(ETR) Proximity Probes, Metric: (ETR) Proximity Probes, Smooth Case:
330193 3300 XL 8 mm Probe, M10 x 1 thread, without 330197 3300 XL 8 mm Probe without armor1
armor
330194 3300 XL 8 mm Probe, M10 x 1 thread, with armor 330198 3300 XL 8 mm Probe with armor1
Part Number-AXX-BXX-CXX-DXX Part Number-AXX-BXX-CXX
A: Unthreaded Length Option A: Overall Case Length Option
Order in increments of 0.5 in
Note: Unthreaded length must be at least 20 mm less than the Length configurations:
case length. Maximum length: 9.5 in
Order in increments of 10 mm. Minimum length: 1.0 in
Length configuration: Example: 3 5 = 3.5 in
Maximum unthreaded length: 230 B: Total Length Option
mm 05 0.5 metre (1.6 feet)
Minimum unthreaded length: 0 mm 10 1.0 metre (3.3 feet)
Example: 0 6 = 60 mm 15 1.5 metres (4.9 feet)
B: Overall Case Length Option 20 2.0 metres (6.6 feet)
Order in increments of 10 mm. 50 5.0 metres (16.4 feet)
Metric thread configurations: Note: 5-metre probes are designed for use with the 5-metre
Maximum length: 250 mm Proximitor sensor only.
Minimum length: 20 mm 90 9.0 metres (29.5 feet)
Example: 0 6 = 60 mm
C: Agency Approval Option
C: Total Length Option 00 Not required
05 0.5 metre (1.6 feet)
05 Multiple Approvals
10 1.0 metre (3.3 feet)
15 1.5 metres (4.9 feet) Aluminum probe clamp bracket1
20 2.0 metres (6.6 feet) 137491-AXX
50 5.0 metres (16.4 feet)
A: Mounting screw option
Note: 5-metre probes are designed for use with the 5-metre
Proximitor sensor only.
01 10-24 UNC-2A mounting
screws
90 9.0 metres (29.5 feet) 02 M5 x 0.8-6g mounting screws
D: Agency Approval Option The aluminum clamp bracket is an unthreaded
00 Not required mounting bracket designed for use with the smooth
05 Multiple Approvals case probes (330140, 330141, 330197 and 330198).
After gapping the probe, tighten the clamp bracket
by tightening the screws. The mounting screws have
3300 XL 8 mm Extended Temperature Range pre-drilled holes for safety wire.
(ETR) Reverse Mount Probes
330195-02-12-CXX-DXX, 3/8-24 UNF threads Aluminum probe threaded mounting bracket
137492-AXX
330196-05-30-CXX-DXX, M10 x 1 threads
A: Thread size
C: Total Length Option
01 3/8-24
05 0.5 metre (1.6 feet)
04 M10 x 1
10 1.0 metre (3.3 feet)
The aluminum probe threaded mounting bracket is
15 1.5 metre (4.9 feet)
the standard mounting bracket for most 3300 and
20 2.0 metres (6.6 feet)
3300 XL probe installations. The -01 option includes
50 5.0 metres (16.4 feet)
two 10-24 UNC-2A mounting screws. The -04 option
Note: 5-metre probes are designed for use with the 5-metre
Proximitor sensor only. includes two M5 x 0.8-6g mounting screws. The
90 9.0 metres (29.5 feet) mounting screws have pre-drilled holes for safety
D: Agency Approval Option wire.
00 Not required
05 Multiple Approvals
Specifications and Ordering Information
Part Number 141194-01
Rev. P (11/11)
Page 13 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Phenolic threaded probe mounting bracket
27474-AXX
A: Thread size
01 3/8-24
04 M10 x 1
We recommend the phenolic threaded mounting
bracket if your application requires additional
electric isolation from the mounting location (as in
some generator and electrical motor bearing
locations). The -01 option includes two 10-24 UNC-
2A mounting screws. The -04 option includes two M5
x 0.8-6g mounting screws. The mounting screws
have pre-drilled holes for safety wire.
Page 14 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Ordering Information Extension 3300 XL Extended Temperature Range (ETR)
Extension Cable
Cables
330190-AXXX-BXX-CXX
Note: Make sure that the extension cable length and the probe
length, when added together, equal the Proximitor Sensor
3300 XL Standard Extension Cable total length.
330130-AXXX-BXX-CXX
Note: Make sure that the extension cable length and the probe
A: Cable Length Option
length, when added together, equal the Proximitor Sensor 0 3 0 3.0 metres (9.8 feet)
total length. 0 3 5 3.5 metres (11.5 feet)
0 4 0 4.0 metres (13.1 feet)
A: Cable Length Option 0 4 5 4.5 metres (14.8 feet)
0 3 0 3.0 metres (9.8 feet) 0 7 0 7.0 metres (22.9 feet)
0 3 5 3.5 metres (11.5 feet) 0 7 5 7.5 metres (24.6 feet)
0 4 0 4.0 metres (13.1 feet) 0 8 0 8.0 metres (26.2 feet)
0 4 5 4.5 metres (14.8 feet) 0 8 5 8.5 metres (27.9 feet)
0 7 0 7.0 metres (22.9 feet) B: Cable Option
0 7 5 7.5 metres (24.6 feet) 00 Standard cable
0 8 0 8.0 metres (26.2 feet) 01 Armored cable
0 8 5 8.5 metres (27.9 feet) C: Agency Approval Option
B: Connector Protector and Cable Option 00 Not required
00 Standard cable 05 Multiple Approvals
01 Armored cable
02 Standard cable with connector
protectors
03 Armored cable with connector
protectors
10 FluidLoc cable
11 Armored FluidLoc cable
12 FluidLoc cable with connector
protectors
13 Armored FluidLoc cable with
connector protectors
C: Agency Approval Option
00 Not required
05 Multiple Approvals
Page 15 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Ordering Information Proximitor
Sensor
Page 16 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Accessories 148722-01
141078-01 3300 XL test plug. The 3300 XL
Manual. Test Plug contains 3 small test
pins attached to 3 color-coded 1-
175751 metre wires, each terminated in a
3300 XL Multi-Purpose Stainless banana plug. The 3-pin adapter
Steel Housing. 12”x12”x6”. Can plugs into the test pin holes on
hold up to 8 Proximitor Sensors in 3300 XL-style Proximitor sensors.
a DIN-mount configuration or 6 You can use this test plug to
Proximitor Sensors in a panel- check the performance of the
mount configuration. (Available Proximitor sensor from the test
with ATEX Zone 0 and Zone 1 pin holes in the terminal strip
certifications.) without removing the field wiring.
176467 04310310
Page 17 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
protectors for 3300 XL 8 mm 330153-09
probes and extension cables.
3300 XL ETR Connector Kit. Used
03839410 on ETR 3300 XL 8mm probes and
3300 ETR XL extension cables.
75Ω triaxial male connector
Contains one pair of male and
protector. Male connector
female ClickLoc connectors, two
protectors install onto the
color-coded sleeves, two pieces
extension cable and attach to the
of high temperature slit FEP
female connector protector on
tubing, and one strip of silicon
the probe, providing
tape.
environmental protection of
connectors. 163356
03839420 Connector Crimp Tool Kit.
Includes 1 set of multiconnector
75Ω triaxial female connector
inserts and connector installation
protector. Female connector
instructions. Compatible only
protectors install onto the probe
with 330153 connector kits or
lead and attach to the male
with probes shipped in 2003 or
connector protector on the
later with ClickLoc connectors
extension cable, providing
uninstalled. Supplied with
environmental protection of
carrying case.
connectors. You can also place
the connector protector onto the
extension cable to slide over the
connection to the Proximitor
sensor to protect that connection
from the environment.
04301007
3/8-24 probe lock nut with
safety wire holes. Single probe
lock nut with 2 holes drilled
through the nut in order to secure
the lock nut in place with safety
wire.
04301008
M10 x 1 probe lock nut with
safety wire holes. Single probe
lock nut with 2 holes drilled
through the nut in order to secure
the lock nut in place with safety
wire.
330153-01
3300 XL connector kit. Used on
3300 XL 8 mm probes and
extension cables. Contains 1 pair
each of male and female ClickLoc
connectors, 2 color-coded
sleeves, 2 pieces of slit FEP tubing,
and 1 strip of silicone tape.
Page 18 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Graphs
Gap (m m ) 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
4 0.10
DSL Error (mils)
2 0.05
mV/mil)
0 0.00
-2 -0.05
-4 -0.10
10
Referenced to 7.87
V/mm (200 mV/mil)
ISF Error (%)
-5
-10
-24
-22
-20
-18
Output (Volts)
-16
-14
-12
-10
-8
-6
-4
-2
0
0 10 20 30 40 50 60 70 80 90 100
Gap (m ils)
Page 19 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Gap (m m ) 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
6 0.15
mV/mil)
0 0.00
-2 -0.05
-4 -0.10
-6 -0.15
10
Referenced to 7.87
V/mm (200 mV/mil)
ISF Error (%)
-5
-10
-24
-22
-20
-18
Output (Volts)
-16
-14
-12
-10
-8
-6
-4
-2
0
0 10 20 30 40 50 60 70 80 90 100
Gap (m ils)
Page 20 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Gap (m m ) 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
6 0.15
4 0.10
mV/mil)
0 0.00
-2 -0.05
-4 -0.10
-6 -0.15
10
Referenced to 7.87
V/mm (200 mV/mil)
ISF Error (%)
-5
-10
-24
-22
-20
-18
Output (Volts)
-16
-14
-12
-10
-8
-6
-4
-2
0
0 10 20 30 40 50 60 70 80 90 100
Gap (m ils)
1m Probe @ 25 °C (77 °F) 1m Probe @ -35 °C (-31 °F)
1m Probe @ 120 °C (248 °F)
Figure 3: Typical 3300 XL 8mm Probe over API 670 Operating Range
Page 21 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Gap (m m ) 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
6 0.15
4 0.10
DSL Error (mils)
mV/mil)
0 0.00
-2 -0.05
-4 -0.10
-6 -0.15
10
Referenced to 7.87
V/mm (200 mV/mil)
ISF Error (%)
-5
-10
-24
-22
-20
-18
Output (Volts)
-16
-14
-12
-10
-8
-6
-4
-2
0
0 10 20 30 40 50 60 70 80 90 100
Gap (m ils)
Figure 4: Typical 3300 XL 8 mm 5m Proximitor Sensor with 4m Extension Cable at Tc (Probe is at 25 °C)
Page 22 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Gap (m m ) 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
8 0.20
6 0.15
DSL Error (mils)
200(mV/mil)
2 0.05
0 0.00
-2 -0.05
-4 -0.10
-6 -0.15
-8 -0.20
15
10
Referenced to
ISF Error (%)
200(mV/mil)
-5
-10
-15
-20
-18
-16
-14
Output (Volts)
-12
-10
-8
-6
-4
-2
0
Gap (m ils) 0 10 20 30 40 50 60 70 80 90 100
Figure 5: Typical 3300 XL 8 mm 5m Proximitor Sensor with 4m Extension Cable at Th (Probe is at 25°C)
Page 23 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Gap (m m )
0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
8 0.20
6 0.15
15
10
ISF Error (%)
-5
-10
-15
-22
-20
-18
-16
Output (Volts)
-14
-12
-10
-8
-6
-4
-2
0
0 10 20 30 40 50 60 70 80 90 100
Gap (m ils)
Figure 6: Typical 3300 XL 8mm 9 m Proximitor Sensor with 8m of Extension Cable at Tc (Probe is at 25 °C)
Page 24 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Gap (m m )
0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
8 0.20
6 0.15
15
10
ISF Error (%)
-5
-10
-15
-22
-20
-18
-16
Output (Volts)
-14
-12
-10
-8
-6
-4
-2
0
0 10 20 30 40 50 60 70 80 90 100
Gap (m ils)
Th=+25 °C (+77 °F) Th=+65 °C (+149 °F)
Th=+85 °C (+185 °F) Th=+100 °C (+212 °F)
Figure 7: Typical 3300 XL 8mm 9m Proximitor Sensor with 8m Extension Cable at Th (Probe is at 25 °C)
Page 25 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Gap (m m ) 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
10 0.25
8 0.20
6 0.15
15
10
ISF Error (%)
-5
-10
-15
-24
-22
-20
-18
Output (Volts)
-16
-14
-12
-10
-8
-6
-4
-2
0
Gap (m ils) 0 10 20 30 40 50 60 70 80 90 100
Figure 8: Typical 3300 XL Extended Temperature Range Probe and 4m Extended Temperature Range Extension
Cable at Th (Proximitor Sensor and Probe Tip with 1-foot Cable are at +25 °C)
Page 26 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Gap (m m ) 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
10 0.25
8 0.20
6 0.15
15
10
ISF Error (%)
-5
-10
-15
-24
-22
-20
-18
Output (Volts)
-16
-14
-12
-10
-8
-6
-4
-2
0
Gap (m ils) 0 10 20 30 40 50 60 70 80 90 100
Figure 9: Typical 3300 XL Extended Temperature Range Probe and 8m Extended Temperature Range Extension
Cable at Th (Proximitor Sensor and Probe Tip with 1-foot Cable are at +25 °C)
Page 27 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Frequency Response to Different Field Wiring Lengths
without Barriers (5 m System)
1
0
Magnitude (dB)
-1
-2
-3
-4
-5
100 1,000 10,000 100,000
Frequency (Hz)
Figure 10: Frequency Response, Typical 3300 XL 8mm 5m or 1m System with Varying Lengths of Field Wiring
Attached, No Barriers
-20
-30
-40
-50
-60
-70
-80
-90
-100
100 1,000 10,000 100,000
Frequency (Hz)
No field wiring 1000' wiring 2000' wiring
5000' wiring 12,000' wiring
Figure 11: Phase Response, Typical 3300 XL 8mm 5m or 1m System with Varying Lengths of Field Wiring Attached,
No Barriers
Page 28 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Frequency Response to Different Field Wiring Lengths without
Barriers (9 m System)
1
Magnitude (dB) -1
-2
-3
-4
-5
100 1,000 10,000 100,000
Frequency (Hz)
Figure 12: Frequency Response, Typical 3300 XL 8mm 9m System with Varying Lengths of Field Wiring Attached,
No Barriers
-20.00
-30.00
-40.00
-50.00
-60.00
-70.00
-80.00
-90.00
-100.00
100 1,000 10,000 100,000
Frequency (Hz)
Figure 13: Phase Response, Typical 3300 XL 8mm 9m System with Varying Lengths of Field Wiring Attached, No
Barriers
Page 29 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Figures
Note: All dimensions shown in millimetres (inches) except as noted.
4
2
1 3
5
2.5 (0.10) 6
7
8
9
10
Page 30 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2
3
1
4
2 (0.08)
5 (0.2) 5
6
7
8
9
Page 31 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2 4
1 3
5
2.54 (0.100)
6 7
8
9
Page 32 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2
4
83.8 83.8
1 (3.30) (3.30)
5
3
83.8 83.8
(3.30) (3.30) 7
6 8
9
2 3
1 4
Page 33 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
1. Mounting option “A”, Options –50 or -90
Figure 19: Panel Mount 3300 XL Proximitor Sensor
Page 34 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
1. Mounting option “A”, Options –50 or -90
Figure 21: Physical Mounting Characteristics Showing Interchangeability of 3300 and 3300 XL Proximitor Sensors
when 4-hole Mounting Option Is Used8
Figure Notes:
1. All dimensions on figures are in millimetres (inches) unless otherwise noted.
2. Standard mount 8 mm probes supplied with M17 or 9/16 inch lock nut.
3. Probes ordered with 5 or 9 metre integral cables have a length tolerance of +20%, -0%.
4. Reverse mount probes not available with armor or connector protector options.
6. Stainless steel armor is supplied with FEP outer jacket for standard probes, PFA outer jacket for ETR probes.
7. FEP jacket is standard non-armored portion of the cable for standard probes, PFA jacket on non-armored portion
for ETR probes.
8. Use M3.5 or #6 screws for panel-mount Proximitor Sensors (screws provided when purchasing Bently Nevada
housings).
Page 35 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
* Denotes a trademark of Bently Nevada, Inc., a wholly owned subsidiary of General Electric Company.
Page 36 of 36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Artisan Technology Group is an independent supplier of quality pre-owned equipment
Artisan Scientific Corporation dba Artisan Technology Group is not an affiliate, representative, or authorized distributor for any manufacturer listed herein.
We're here to make your life easier. How can we help you today?
(217) 352-9330 I [email protected] I artisantg .com