3216 22917 1 PB
3216 22917 1 PB
4, 480-486
DOI: 10.11152/mu-3216
1Department Allgemeine Innere Medizin (DAIM) Kliniken Hirslanden Beau Site, Salem und Permanence, Bern,
Switzerland, 2Department for Internal Medicine, Krankenhaus Märkisch Oderland GmbH Strausberg/ Wriezen and
Brandenburg Institute of Clinical Ultrasound (BICUS) at Medical University Brandenburg „Theodor Fontane“,
Neuruppin, Germany, 3Ruprecht Karls University Heidelberg Medical School, Heidelberg, Germany, 4Department
of Internal Medicine 2, Caritas Krankenhaus, Bad Mergentheim, Germany, 5Imaging Department, Imperial College
Healthcare NHS Trust, Charing Cross Hospital, Fulham Palace Road, London UK, 6Department of Medical Ultra-
sound, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China,
7Department of Ultrasound, Zhongshan Hospital, Fudan University, Shanghai, China, 8Harvard Medical School,
Department of Emergency Medicine, Beth Israel Deaconess Medical Center, Boston, USA
Abstract
Ultrasonography is a ubiquitous and indispensable diagnostic and therapeutic tool in medicine. Due to modern equipment
and automatic image optimization, nowadays the introduction of ultrasound imaging requires only little technical and physi-
cal knowledge. However, profound knowledge of the device function repertoire and underlying mechanisms are essential for
optimal image adjustment and documentation. From a medical as well as an aesthetic point of view, the goal should always be
to achieve the best possible image quality. This article provides an overview of handling of ultrasound systems, fundamental
adjustments and their optimization in Doppler ultrasound.
Keywords: guideline; Doppler; ultrasound; perfusion; vascularity
Fig 3. Measurement of flow velocities in the superior mesenteric artery (SMA): Following adjustment of scale to the velocity range
typical for superior mesenteric artery (95.9 cm/s) and a correction of the Doppler angle to 45° correct measurement of the normal
peak systolic velocity is possible: 145 cm/s (a); in case of no correction of Doppler angle an erroneous low peak systolic velocity is
measured (b; 90 cm/s); improper positioning of the US probe results in a Doppler angle of 72° (c), and correct measurement of flow
velocity is impossible (erroneous measurement of a high peak systolic velocity in the range of mild stenosis)
flow downstream, which is also independent of the angle different colours (usually red and blue) and colour bright-
of insonation, is the Pulsatility Index [PI = (peak systolic ness and shading convey different flow velocities. A third
flow – peak diastolic flow) / (mean flow)]. If it is not fea- colour (green) displays the velocity dispersion. Thus, tur-
sible to adjust the Doppler angle below 60°, one should bulences are visualized. In contrast to the modality men-
refer to Doppler angle independent measurements like RI tioned above, the intensity mode (Amplitude-Doppler /
or PI [8-10]. Power-Doppler / Angiomode) displays only the flow am-
Inversion plitude without flow direction. This allows for a higher
The “Inversion” function interchanges the mapping sensitivity in detecting low-velocity blood flow [11].
of the flow direction as part of the Doppler spectrum. Val- There are two main objectives of CDUS. The first
ues are displayed above or below the baseline depending is to detect the presence of flow, its direction and ac-
on the flow towards or away from the probe. celerations within larger vessels (macro-CDUS). The
Post-Processing second is to visualize perfusion of organs or tumors by
As part of post-processing, the brightness of the im- detecting low-velocity flow in small parenchymal ves-
age can be varied by adjusting the Doppler gain, which sels (micro-CDUS). Micro-CDUS techniques close the
allows for better noise suppression. For arteries, the gain gap to contrast-enhanced ultrasound (CEUS). Different
should be set in a way that a frequency-free window is settings and modalities are used for macro- and micro-
recognizable. In post-processing, the colour of the image CDUS [12,13]. CDUS must be distinguished from other
can also be changed (e.g., monochromatic or polychro- flow imaging modalities. Microvascular Flow Imaging
matic illustration), and the angle correction can be carried (MFI) for example displays blood flow in real-time on
out as described above. In newer machines, the user can the B-scan image. Echoes are encoded, decoded and fil-
edit sweep speed and baseline as well. The knowledge tered in such a way that the significantly weaker echoes
of post-processing alterations is essential particularly in of blood cells are visualized. By comparing successive
measurements acquired in breath hold while examining frames, these weaker echoes are displayed as a flow. Cur-
abdominal vessels. rently, manufacturers have established this technology to
Automatic Image Optimization varying degrees. The advantages are significantly better
Most high-end ultrasound devices can automatically spatial resolution similar to B-scan sonography and bet-
adjust the Doppler settings by changing the parameters ter representation of slow flow with fewer artifacts than
outlined above. Nevertheless, optimum image settings CDUS. Owing to limitations such as a lack of velocity
are best achieved manually. measurement and penetration depth, it is likely that mi-
croflow imaging will increasingly complement CDUS in
2. Colour and Power Doppler US the future without replacing it [14,15].
Other types of MFI make use of traditional Doppler
CDUS integrates B-scan and colour-coding of flow ultrasound techniques with enhanced filters and higher
information. It is often complemented by PWD. CDUS frame rates improving the discrimination between slow
is therefore primarily used for targeted vascular diagnos- flowing signals and background noise. These advanced
tics. Two modes are distinguished. In velocity mode, the Doppler techniques have acronyms such as Superb Mi-
flow direction in relation to the transducer is coded by crovascular Imaging (SMI) (Canon Medical systems)
484 Axel Löwe et al „Knobology“ in Doppler Ultrasound
Fig 7. Adjustment of velocity scale in colour Doppler of a focal thrombus of the femoral vein: low scale (2.3 cm/s), aliasing with
thrombus (*) poorly visibly (a); high scale (62.3 cm/s), nearly no flow is visible and complete thrombosis is feigned (b); adequate
scale (11.7 cm/s) showing venous blood surrounding the focal thrombus (c)
Med Ultrason 2021; 23(4): 480-486 485
Duplex- and Triplex-Mode 5. Atkinson NSS, Bryant RV, Dong Y, et al. How to perform
gastrointestinal ultrasound: Anatomy and normal findings.
CDUS and PWD may be used together with B-mode World journal of gastroenterology 2017;23:6931-6941.
6. Jenssen C, Tuma J, Moller K, et al. [Ultrasound artifacts
imaging. While “Duplex imaging” incorporates grey-
and their diagnostic significance in internal medicine and
scale imaging and either CDUS or PWD, the examiner
gastroenterology - part 2: color and spectral Doppler arti-
makes use of all three modalities in what is termed “Tri- facts]. Z Gastroenterol 2016;54:569-578.
plex imaging”. This modality can be applied to detect the 7. El Haddad M, Backer Td, Buyzere Md, et al. Grading of
vessel of interest by first using CDUS and then perform mitral regurgitation based on intensity analysis of the con-
an additional flow measurement by pressing the “Update” tinuous wave Doppler signal. Heart (British Cardiac Soci-
button. Here, only PWD is active in real time while the ety) 2017;103:190-197.
CDUS image is frozen. Newer devices can also display 8. Ignee A, Boerner N, Bruening A, et al. Duplexsonography
B-scan, CDUS and PWD simultaneously in real time in of the mesenteric vessels - a critical evaluation of inter ob-
Triplex mode. However, as this is accompanied by a loss server variability. Z Gastroenterol 2016;54:304-311.
of image quality and a significantly reduced frame rate, 9. Dietrich CF, Jedrzejczyk M, Ignee A. Sonographic assess-
ment of splanchnic arteries and the bowel wall. Eur.J.Radiol
Duplex mode is generally preferred.
2007;64:202-212.
10. Dietrich CF, Ignee A, Seitz KH, Caspary WF. [Du-
3. Contrast enhanced Doppler ultrasound plex sonography of visceral arteries]. Ultraschall Med
2001;22:247-257.
The introduction of US contrast agents (USCA) has
11. Evans DH, Jensen JA, Nielsen MB. Ultrasonic colour Dop-
improved the potentials of colour Doppler US in difficult pler imaging. Interface Focus 2011;1:490-502.
situations, e.g., to visualize the renal arteries [17,18], and 12. Herbay Av, Häussinger D. Der Einsatz der farbkodierten
to reduce the number of equivocal examinations [19]. The Duplexsonographie (FKDS) erhöht die diagnostische Si-
same is true for other organs including pancreatic ves- cherheit in der Abdomensonographie. Ultraschall Med
sel evaluation for the important differential diagnosis of 2002;23:168-175.
chronic pancreatitis and ductal adenocarcinoma [20,21]. 13. Martinoli C, Derchi LE, Rizzatto G, Solbiati L. Power Dop-
pler sonography: general principles, clinical applications,
Conclusion and future prospects. Eur Radiol 1998;8:1224-1235.
14. Wachsberg RH. B-Flow Imaging of the Hepatic Vascula-
The article summarizes knowledge, handling and re- ture: Correlation with Color Doppler Sonography. Ameri-
commendations for colour and spectral Doppler ultra- can Journal of Roentgenology 2007;188:W522-W533.
sound parameter adjustment to achieve the best possible 15. Calliada F, Campani R, Bottinelli O, Bozzini A, Sommaru-
image quality. The most important parameters include ga MG. Ultrasound contrast agents: Basic principles. Eur J
transmission frequency, pulse repetition frequency (scale), Radiol 1998;27:S157-S160.
baseline, wall filter, gain, sweep speed, size of gate, Dop- 16. Lim AKP, Satchithananda K, Dick EA, Abraham S, Cos-
grove DO. Microflow imaging: New Doppler technology
pler angle, inversion mode, post-processing parameters,
to detect low-grade inflammation in patients with arthritis.
automatic image optimization, colour Doppler box size
Eur Radiol 2018;28:1046-1053.
and angle steering as well as duplex- and triplex-mode. 17. Dietrich CF, Ignee A, Hocke M, Schreiber-Dietrich D,
Greis C. Pitfalls and artefacts using contrast enhanced ul-
Conflict of interest: none
trasound. Z.Gastroenterol 2011;49:350-356.
References 18. Brkljacic B, Castellani S, Deane C, Dietrich CF. Doppler
ultrasound of renal vessels. In: Dietrich CF, ed. EFSUMB
1. Dietrich CF, Shi L, Wei Q, et al. What does liver elastog- Course Book on Ultrasound. London, 2012; 716-736.
raphy measure? Technical aspects and methodology. Min- 19. Dietrich CF, Trenker C, Fontanilla T, et al. New Ultrasound
erva Gastroenterol Dietol 2020, DOI: 10.23736/S1121- Techniques Challenge the Diagnosis of Sinusoidal Obstruc-
421X.20.02787-7 tion Syndrome. Ultrasound Med Biol 2018;44:2171-2182.
2. Dietrich CF, Averkiou M, Nielsen MB, et al. How to per- 20. Hocke M, Ignee A, Topalidis T, Stallmach A, Dietrich CF.
form Contrast-Enhanced Ultrasound (CEUS). Ultrasound Contrast-enhanced endosonographic Doppler spectrum
Int Open 2018;4:E2-E15. analysis is helpful in discrimination between focal chronic
3. Lim AKP, Moran CM, Dietrich CF. Ultrasound “Knobol- pancreatitis and pancreatic cancer. Pancreas 2007;35:286-
ogy”. In: Dietrich CF, ed. EFSUMB Course Book, 2nd Edi- 288.
tion. London: EFSUMB, 2020;1-24. 21. Hocke M, Schulze E, Gottschalk P, Topalidis T, Dietrich
4. Zander D, Hüske S, Hoffmann B, et al. Ultrasound Im- CF. Contrast-enhanced endoscopic ultrasound in discrimi-
age Optimization (“Knobology”): B-Mode. Ultrasound Int nation between focal pancreatitis and pancreatic cancer.
Open 2020;6:E14-E24. World J.Gastroenterol 2006;12:246-250.