Pharmaceutical corporations are tasked with testing their products temperature uniformity throughout the cavity. Because the heating
extensively prior to distribution. Testing is performed on raw element is in the side of the furnace walls, it will not damage
materials, excipients, and during the quality control process for metallic crucibles, which can be used in the Phoenix BLACK. The
the finished product. Traditional techniques require significant user Phoenix BLACK is compatible with all crucible types, including
interaction and are subject to error. The Phoenix BLACK™ sulfated quartz, platinum, and ceramic.
ashing furnace and SMART Q™ rapid moisture analyzer provide
safe, automated, and reliable alternatives and meets all muffle All temperature measurements are performed using NIST-traceable
furnace and Loss on Drying (LOD) requirements for most global thermocouples. The exterior of the furnace is cool enough to touch,
pharmacopoeias, including those for the United States, Britain, even when the inside is at temperature. In order to access the
Japan, and Europe. Traditional techniques require hours in a drying interior of the furnace, the furnace door is placed in a holder that
oven for moisture, or Bunsen burners, heat lamps, hot plates, and prevents the user from being exposed to the hot door surface. An
muffle furnaces continuously maintained at high temperature. This optical sensor monitors the presence of the interior furnace door
equipment poses significant time loss for batch releases, as well and will activate safety locks to ensure the cavity door is held open
as safety risks, as users are directly exposed to hot surfaces and until the furnace door is put back in place. Upon method completion,
open flames. The risk of experimental error is also high, as weight active ventilation rapidly cools the furnace.
must be taken, transferred manually, and calculated. Other rapid
All sulfated ashing steps can be performed in the Phoenix BLACK
LOD and furnace systems exist, but analysis time, temperature
with the sulfated ashing option, eliminating the need for additional
uniformity, temperature measurement, and safety standards are
equipment and minimizing operator input. Acid can be added to the
all inferior to the instruments discussed herein. The SMART Q
sample in the crucible and placed directly into the Phoenix BLACK.
and Phoenix BLACK minimize risk and maximize lab efficiency by
improving the LOD and sulfated ash procedures. A lower temperature sulfating/ashing step with active vapor
scrubbing is automatically completed, prior to the high-temperature
Safer and More Efficient Sulfated dry ashing of the organic sample.
Ashing The system can be programmed with a pause step to allow the
The Phoenix BLACK is a muffle furnace that meets all microwave sulfated crucible to be removed and weighed, prior to the dry ashing
and standard muffle furnace requirements listed in ASTM, USP step. The Phoenix BLACK sulfated ashing option neutralizes toxic
(<281> ROI & <733> LOI), AOAC, FDA, ISO, DIN, ASTM D5630- acid vapor through a series of scrubbers for a more environmentally
94, and ASTM D1506-94b. The system is easily programmable, sound, laboratory-friendly solution with no emission of hazardous
contains a library of preprogrammed methods, and can ash up fumes. The system also helps the laboratory comply with ISO
to 15 samples at a time. The continuous silicon-carbide heating 14000 Environmental Management requirements to eliminate
element is encased in a quartz-fiber furnace, which maintains greenhouse gases.
© 2020 CEM Corporation
White Paper Page 2 of 3
Better Proximate Testing of Pharmaceuticals ap0232v1
The Phoenix BLACK workstation allows quantification for document The Phoenix BLACK results closely match the results of the
control and automatically determines the % ROI, % LOI or % Ash. conventional muffle furnace. All of the tests were complete in
Once generated, sample information cannot be modified, aiding in significantly less time than the traditional muffle furnace. It should
21 CFR Part 11 compliance. All weighing and method information be noted that the time for the conventional muffle furnace does
is included in the data file, as well as the final result, user-entered not include additional operator actions, such as using a hot plate
sample ID, and a date/time stamp indicating when the sample or Bunsen burner, which may require an additional 30 minutes
was ashed. All methods are stored automatically and the system of operator time, prior to insertion of the sample into the muffle
provides the security of a password lock to reduce the possibility of furnace. In most laboratories, this step requires constant attention,
erroneous method modification. Data is easily exported to a LIMS as the technician cannot leave an open flame. The Phoenix BLACK
system for even easier data management. sulfated ashing option automates the sulfating/ashing step
with sulfuric acid, reducing the amount of operator attention and
Faster Loss on Drying for all creating a safer laboratory environment.
Percent Moisture
Sample SMART Q Air Oven Difference
Acetaminophen 1.364 1.359 0.005
Gelatin Supplement 10.276 10.275 0.001
Vitamin C 0.557 0.550 0.007
Vapor Rub 11.621 11.623 -0.002
Nutritional Shake 31.265 31.265 0.000