Manual Tuxon C Uo3r
Manual Tuxon C Uo3r
Tuxon- C
Ver. 2 / 2022 / 01
Table
1.General Information ...................................................... .- 1 -
1.1 Function ...............................................................- 1 -
1.2 Description...........................................................- 2 -
1.3 Rear illustration....................................................- 3 -
1.4 Specification ........................................................- 4 -
1.5 Dimension............................................................- 5 -
2. Wiring ............................................................................- 6 -
2.1 Installation............................................................- 6 -
2.2 Power connection ................................................- 7 -
2.3 Load cell connection............................................- 7 -
2.4 I/O terminals ........................................................- 9 -
2.5 Analog output connection ..................................- 10 -
2.6 Serial port connection........................................ - 11 -
3. Calibration...................................................................- 12 -
3.1 Illumination.........................................................- 12 -
3.2 Calibration flow chart .........................................- 12 -
3.3 Millivolt display...................................................- 15 -
3.4 Calibration without weights................................- 16 -
3.5 Quick zero/gain calibration ................................- 18 -
3.6 Parameter table .................................................- 18 -
3.7 Parameter record...............................................- 18 -
4. Parameter setting........................................................- 19 -
4.1 Defination...........................................................- 19 -
4.2 Operation parameter table.................................- 21 -
5. Operation ....................................................................- 24 -
5.1 Operating status ................................................- 24 -
5.2 Operating procedure..........................................- 25 -
5.3 Manual zeroing ................................................. .- 25 -
5.4 I/O testing ..........................................................- 26 -
5.5 I/O defination .....................................................- 27 -
5.6 Display testing ...................................................- 29 -
5.7 Reset .................................................................- 30 -
5.8 Backup...............................................................- 31 -
5.9 Restore backup .................................................- 31 -
5.10 Analog calibration ............................................- 31 -
5.11 Password .........................................................- 33 -
5.12 Password setting .............................................- 33 -
6. Serial port communication ..........................................- 34 -
6.1 RS protocol ........................................................- 34 -
6.2 RE protocol ........................................................- 45 -
6.3 MODBUS protocol .............................................- 47 -
7. Error and alarm message ...........................................- 59 -
1 General information
Tuxon-C weighing indicator is a small-volume display
specially for weighing and controlling in industrial field, which
features are abundant communication, high precision, strong
function and easy operation etc., thus to be widely used in
weighing and controlling equipment, such as concrete mixer,
asphalt batching, feeding stuff, metallurgy converter and so
on.
1.1 Function:
Small volume, unique design, easy operation;
Suit to all of load cells at bridge-type resistance strain
gauge;
Multi digital filter;
5set points are used widely to meet more requirement;
RS232/RS485 (optional) to ensure stable communication;
Abundant communication protocols(RS/RE/MODBUS
optional) to communicate easier with host computer;
High precision 16bit DA analog output;
Digital calibration with or without weights;
Password protection for calibration, operation parameters
etc.;
Automatic zeroing when power up;
Automatic zero tracking.
-1-
1.2 Description
P 1-1
Keypad:
-2-
Set point display:1 digit, display set point numbers.
-3-
1.3 Rear illustration
P 1-2
① I/O input terminals
② I/O output terminals
③ Load cell terminals
④ Serial port / analog terminals
⑤ Power connection
⑥ Fuse socket
-4-
1.4 Specification
Common:
Power supply:AC90V~260V 50Hz (or 60Hz) ± 2%
Filter:Inside
Working temperature:-10~40℃
Max humidity:90%R.H without dew
Power consumption:About 10W
Dimension:105×151×57mm
Analog:
Load cell power:DC5V 300mA (Max)
Input impedance:10MΩ
Zero steady range:0.02~9mV
Input sensitivity:0.01uV/d
Gain input range:0.2~10mV
Transfer mode:Sigma - Delta
A/D conversion speed:120times/sec
Non-linearity:0.01%F.S
Gain drift:10PPM/℃
Display Precision:1/30000 Digital:
Weight display:6digits red high-brightness LED
Minus display:“-”
Overload display:“OFL”
Decimal point:5kinds (optional)
-5-
1.5 Dimension
-6-
2 Wiring
2.1 Installation
3. Install indicator in
the front of control box.
4. Insert mounting
plates on both sides
and then screw tightly.
P 2-1
-7-
2.2 Power connection
Tuxon-C weighing indicator connects power as follows:
P 2-2
1) 6 wires connection:
Load cell
2-3
-8-
2) 4 wires connection
P 2-4
-9-
2.4 I/O terminals
Tuxon-C Weighing indicator transfer data by
optoelectronic isolation technology, thus need 24 V DC
outside as power supply, which anode connect with indicator
24 V+, and cathode with 24V- through the COM1 (DC+)
and the COM2 ( DC-). The input signal is low level
effective, and the output is transistor open-collector output,
which driving current can reach 500m A.
I/O definition as follows:
P 2-5
Indicator input terminal connection:
P 2-6
Indicator output terminal connection:
P 2-7
- 10
I/O tolerant definition as follows:
Output Input
OUT1 SP1 IN1 Zero
P 2-8
- 11 -
2.6 Serial port connection
Serial port(RS485 or RS232) connection as follows:
P 2-9
RS232 connection:
P 2-10
RS485 connection:
Controller 1
Computer
.
.
. GND is ground of RS485, it can very
mu ch i mp rov e co mmu n i cat i o n
quality via connecting with GND by
Controller n
- 12 -
P 2-11
- 13 -
3 Calibration
3.1 Instruction
Tuxon-C Weighing indicator need calibrate atbeginning
of operation or changing any parts on weighing system or
non-suitable application for user. The calibration can define
radix point, minimum scale division, maximum
scale capacity, zero and gain, etc.
User can press to enter next or press to
save setting to only change one parameter, and then
press to exit.
- 14 -
1) Press set radix
point (0~0.0000 5 kinds
optional) , press to
save and enter mini division.
2) If not change radix point,
press directly to enter
mini division.
- 15 -
1) When load cell output
millivolt value, press to
enter zero calibration directly.
2) The display value is similar
with millivolt from SIG+/SIG-
terminals. See Character 3.3
for details.
- 16 -
3.3 Millivolt display
This function is mainly used to test weighing system,
four-corner position-error of force-transmitted equipment and
load cell’s linearity.
1. Test weighing system
(1) If the indicator display variational millivolt value
according to add weight, thus indicate load cells are
connected correctly and force-transmitted equipment is
operated normally.
(2) If the indicator display OFL (or –OFL), thus indicate
the weight is too heavy (or too light), then need unload (or
add) the weight. But if always display OFL (or –OFL) after
changing weight, maybe the following reason cause:
a. Please check force-transmitted equipment.
b. Please check the load cell’s connection.
c. Please check Load cells.
2. Test four-corner position-error of force-transmitted
equipment
Please load same weight on scale’s every corner and
record individual millivolt value. If the values are obviously
different, please adjust force-transmitted equipment.
3. Test load cell’s linearity
Please load several same weight values within the
weighing capacity and record relevant millivolt value after
zeroing every time; if the discrepancy is bigger among
millivolt values, thus indicate that the load cell’s linearity is not
good, need change new load cell or adjust force-transmitted
equipment.
- 17 -
3.4 Calibration without weights
Calibration without weights is only for urgency. If change
new load cells or indicator, or adjust weighing system,
calibration without weights is not correct.
- 18 -
- 19 -
3.5 Fast zero / gain calibration
In weighing status, press for long time, when
display password input, then input password correctly and
enter zero calibration. See Character 5.10 for details.
3.6 Calibration parameter table
Initial
Sign Parameter Type Division
value
Decimal 0,0.0,0.00,0.000,
Point 5 0.0000 0
point
1d= Mini division 6 1,2,5,10,20,50 1
Max
CP ≤mini division×30000 10000
capacity
System
St
millivolt
SE Sensitivity 2 2,3 2(mV/V)
E SCAL Zero
AddLd(d) Gain
Set
PASS 000000
password
- 20 -
4 Parameter setting
4.1 Defination
Press to select parameters, then press and
to set parameters, after that, press to save
setting.
- 21 -
1) Optional parameter setting
Set“parameterF1.1”(ON)for example:
- 22 -
4.2 Operation parameter table
No. Para. Init ial Description
F1 None None No.1 parameter
Automatic zeroing when power
F1.1 ON/OFF OFF
on
Zero tacking range(00 ~ 99d
F1.2 00~99 01 optional).
If setting 0, not zero tracking.
Zeroing range(01% ~ 99% of
F1.3 00~99 05
full capacity).
Stability criterion range(01 ~
F1.4 00~99 01 99d optional). If 0, always
stability and the light are on.
Digital filter parameter
F1.5 0~9 4
0:no filter;9:strongest filter
Stability filter
F1.6 0~9 7
0:no filter;9:strongest filter
F2 None None No.2 parameter
Set point 1 Set point 1 to 5
F2.1 ×××××× 0 (maximum. according to
weight data) weight value
from max. to
F2.2 ×××××× 0 Set point 2 mini. Set weight
F2.3 ×××××× 0 Set point 3 value as need,
if no use, keep
F2.4 ×××××× 0 Set point 4
0. Note:Do not
Set point 5 set 0 as weight
F2.5 ×××××× 0 (minimum value between
weight data) set points.
- 23 -
F3 None None No.3 parameter
Analog output:
4-20 /
4-20:4-20mA
0-20 /
0-20:0-20mA
0-24 /
0-24:0-24mA
F3.1 0-5 / 4-20
0-5 :0-5V
0-10 /
0-10:0-10V
-5-5 /
-5-5 :-5-5V
-10-10
-10-10 :-10-10V
- 24 -
MODBUS communication
F4.4 Rtu/Asc Rtu mode (Available when F4.3 is
Bus)
7-E-1/
7-O-1/
7-N-2/ Data format(data bit + parity bit
F4.5 8-E-1/ 8-E-1 + stop bit) , only the last 4
8-O-1/ format is for MODBUS.
8-N-1/
8-N-2
F5 None None No.5 parameter
OFF : Not need stability to
adjudge whether output or not
ON/OF from set point;
F5.1 OFF
F ON:Need stability to adjudge
whether output or not from set
point.
- 25 -
DA:Sub-display show analog
value;
SP:Sub-display show present
weight among different set
points.
If present weight<mini set
F5.2 DA/SP DA point, sub-display show weight
value at mini set point; if
present weight>set point 1,
sub-display show weight value
at set point 1; others sub-
display show nearest set point-
present weight.
F6 None None No.6 parameter
ON /
F6.1 OFF Password switch
OFF
Password setting. See
F6.2
Character 5.11 for details.
5 Operation
5.1 Operation status
1. When power on, buzzer hoot, main and sub-display
flash “8”,and instruction lights also sparkle.
2. After finished, main display
show “8806-A” , and sub-display
show “software no. ”.
3. After 3 seconds, main display
show present weight value, andsub-
display show related setting in
Character F5.2. See Character F5.2
for details.
- 26 -
5.2 Working instrument
For examples: F2.1=SP1 、 F2.2=SP2 、 F2.3=SP3 、
F2.4=SP4、F2.5=SP5,present weight W compare with SPx
to control I/O output.
Set point
Weight I/O output
display
W>= SP1 1 SP1 output effectively
SP2<=W<SP1 2 SP2 output effectively
SP3<=W<SP2 3 SP3 output effectively
SP4<=W<SP3 4 SP4 output effectively
SP5<=W<SP4 5 SP5 output effectively
W<SP5 6 SP6 output effectively
- 27 -
5.4 I/O testing
Press to switch among OUT1~OUT10 and
press to return. Please refer following flow chart:
Stop
Sub-display indicate F
means IN1~IN4 from
Press , left to right,when input
main-display effectively, F change to
show “inPUt” O.
- 28 -
5.5 I/O definition
When stop, user can define I/O ports by and
to return.
The flow chart is to define OUT3 as O5(SP5) for
example:
- 29 -
Effective output when set
O2 Sp 2 point 1 >present weight>=set
point 2.
Effective output when set
O3 Sp 3 point 2 >present weight>=set
point 3.
Effective output when set
O4 Sp 4 point 3 >present weight>=set
point 4.
Effective output when set
O5 Sp 5 point 4 >present weight>=set
point 5.
- 30 -
5.6 Display testing
The following flow chart is to test lights on main-display
and sub-display, status and instruction:
Press 9 times in stop status till
main display show “tStdip”.
- 31 -
5.7 Reset
Reset present parameters to initial setting.
Reset “I/O definition (io)” for example:
- 32 -
5.8 Backup
User can make all of parameters backup.
Make“all of parameters(ALL)”backup for example:
1. When stop , press till main
display show “bAC”. And then press
- 33 -
2) 12mA:Main display shows CAL DA, sub-display 32768,
ID indicates 2;
3) 20mA:Main display shows CAL DA, sub-display 54613,
ID indicates 3;
maximum current:Main display shows CAL DA, sub-display
24.000.
For example: calibrate “12mA output” and “Max. current” :
1. When stop, press till main display
show CALdA.
2. Press to input calibration
password.
3. Press ,ID indicate from 1 to 2,
which means from 4mA to 12mA.
4. Calibrate when analog output is not
12mA by multi-meter.
Press , sub-display sparkle , thenpress
or to change DA digits to adjust analog
output 12mA,press tofinish.
User also can change DA digits according to
the highest digit: Press when
calibrate,then present highest digit sparkle,
so press to select and press and
in turn to change digits.
5. If ID doesn’t display, which indicate the
present value is maximum analog output,
user can input present analog value now.
6. Press to exit.
1. Press long time in calibration, sub-display
value for all of analog output restore to initial value.
- 32 -
5.11 Password
All initial passwords are :000000。Note:Calibration
password is same password as reset, backup, restore, I/O
definition and analog calibration. Whencalibration password
changed, others also changed.
Password operation as follows:
1. Press to enter password.
Correct response:
Scale
STX R S 000 Status +/- Value CRC CR LF
no.
Wrong response:
Correct response:
Wrong response:
Here:
Spx —— 1bit,1- 5 means 31H、32H、33H、34H、35H
DDDDDD —— 6bits,set point value.
Read command on set point 1 for example:
02 30 31 52 31 33 30 0D 0A
Correct response:
02 30 31 52 31 30 30 30 37 30 30 32 35 0D 0A
Means set point 1 value is 700.
3) Host computer read radix point
- 36 -
Read command:
STX Scale no. R P CRC CR LF
Correct response:
Here:
P —— 1bit,50H
DDDDDD —— 6bits,range at 0- 4,means radix point,for
example, 30H 30H 30H 30H 30H 34H for 4bits radix point
Read radix point for example:02 30 31 52 50 36 31 0D 0A
Correct response:
02 30 31 52 50 30 30 30 30 30 31 35 30 0D 0A
Means: 1# scale radix point 1bit.
4) Host computer read operation parameters
Read command:
STX Scale no. R F Parameters 0 CRC CR LF
Correct response:
STX Scale no. R F Parameters 0 DDDDDD CRC CR LF
Wrong response:
- 37 -
Here:
F —— 1bit,46H
Parameters —— 3bits, for example: digit filter F1.5 for 31H
35H 30H
0 —— 1bit,30H
DDDDDD —— 6bits,operation parameter data
Read command operation parameter F1.5 for example:
02 30 31 52 46 31 35 30 30 34 39 0D 0A
Correct response:
02 30 31 52 46 31 35 30 30 30 30 30 30 30 34 34 31 0D 0A
Means: 1# scale F1.5 is 4.
5) Host computer write set point value
Write command:
STX Scale no. W Spx DDDDDD CRC CR LF
Correct response:
Wrong response:
Here:
W —— 1bit,57H
Spx —— 1bit,1- 5 means 31H、32H、33H、34H、35H
- 38 -
DDDDDD —— 6bits,set point value
K —— 1bit,4BH
Write command set point 1 at 1500 on #1 scale for example:
02 30 31 57 31 30 30 31 35 30 30 32 39 0D 0A
Correct response :
02 30 31 57 31 4F 4B 38 39 0D 0A
Means: #1 scale have saved written data.
6) ) Host computer write operation
parameters Write command:
STX Scale no. W F Parameters 0 DDDDDD CRC CR LF
Correct response:
Wrong response:
Here:
W —— 1bit,46H
Parameters —— 3bits,for example, digit filter F1.5 31H
35H 30H
O —— 1bit,30H
DDDDDD —— 6bits,operation parameter value
Write command F1.5 at 7 on #1 scale for example:
- 39 -
02 30 31 57 46 31 35 30 30 30 30 30 30 30 37 34 39 0D 0A
Correct response:02 30 31 57 46 4F 4B 31 30 0D 0A
Means: #1 scale has saved F1.5 at 7.
7)Host computer zero calibration with weights
Write command:
STX Scale no. C Z CRC CR LF
Correct response:
Wrong response:
Here:
C —— 1bit,43H
Z —— 1bit,5AH
Write command zero calibration with weights on #1 scale for
example: 02 30 31 43 5A 35 36 0D 0A
Correct response:02 30 31 43 5A 4F 4B 31 30 0D 0A
Means command has been performed correctly. 8
)Host computer zero calibration without weights
Write command:
STX Scale no. C Y DDDDDD CRC CR LF
Correct response:
- 40 -
STX Scale no. C Y O K CRC CR LF
Wrong response:
Here:
Y —— 1bit,59H
DDDDDD —— 6bits,related zero millivolt,fixed 3bit radix
point.
For example:Send 1.500mA as zero calibration to #1 scale
02 30 31 43 59 30 30 31 35 30 30 34 39 0D 0A
Correct response:02 30 31 43 59 4F 4B 30 39 0D 0A
Means data has been saved in #1 scale.
9)Host computer calibrate radix point
Write command:
STX Scale no. C P Radix point CRC CR LF
Correct response:
Wrong response:
Here:
P —— 1bit,59H
- 41 -
Radix point —— 1bit,range at 0~4
For example: calibrate radix point 3 on #1 scale
02 30 31 43 50 33 39 37 0D 0A
Correct response: 02 30 31 43 50 4F 4B 30 30 0D 0A
Means: Data has been saved in #1 scale.
10) Host computer calibrate minimum division and
maximum capacity
Write command:
STX Scale no. C M DD DDDDDD CRC CR LF
Correct response:
Wrong response:
Here:
M —— 1bit,4DH
DD —— 2bits,division 1、2、5、10、20、50
DDDDDD —— 6bits,maximum capacity
For example: Send division 1 and maximum capacity 10000
to #1 scale.
02 30 31 43 4D 30 31 30 31 30 30 30 30 32 39 0D 0A
Correct response:02 30 31 43 4D 4F 4B 39 37 0D 0A
- 42 -
Means: Data has been saved in #1 scale.
Correct response:
Wrong response:
Here:
G —— 1bit,47H
DDDDDD —— 6bits,weights value
Send command gain calibration to #1 scale for example:
02 30 31 43 47 30 31 30 30 30 30 32 36 0D 0A
Correct response: 02 30 31 43 47 4F 4B 39 31 0D 0A
Means: Written 10000 have been saved in #1 scale.
12)Host computer gain calibration without weights
Write command:
Scale
STX C L D1D1D1D1D1D1 D2D2D2D2D2D2 CRC CR LF
no.
Correct response:
- 43 -
STX Scale no. C L O K CRC CR LF
Wrong response:
Here:
L —— 1bit,4CH
D1D1D1D1D1D1 —— 6bits,gain millivolt
D2D2D2D2D2D2 —— 6bits,weight value
For example:Calibrate 4.110mA as 10000
02 30 31 43 4C 30 30 34 31 31 30 30 31 30 30 30 30 32 35
0D 0A
Correct response:02 30 31 43 4C 4F 4B 39 36 0D 0A
Means:Written data has been saved in #1 scale.
13)Host computer zeroing operation
Write command:
STX Scale no. C C CRC CR LF
Correct response:
Wrong response:
Here:
R —— 52H
E —— 45H
A —— 41H
D —— 44H
CR —— 0DH
LF —— 0AH
Response data format is same as data format in continue
mode.
For example
Command: 52 45 41 44 0D 0A
Response: 53 54 2C 47 53 2C 2B 30 31 31 2E 31 32 30 4B
67 0D 0A
Means:stable、positive data、present weight value is 11.
120kg.
- 46 -
6.3 MODBUS protocol
6.3.1 Communication mode
MODBUS communication by two mode:RTU and Asc。
When parameter F4.4 is RTU,every 8-bit byte of the
message are divided into 2pcs of 4-bit hexadecimal
characters to transmit at binary code. Data format:8- E-1,
8- O-1,8- n-1,8- n-2。
When parameter F4.4 is ASCII,every 8-bit byte of the
message is transmitted as 2pcs ASCII characters at ASCII
code, which can support all of data format in parameter F4.5.
6.3.2 MODBUS communication address
PLC Display
Description
address address
The following items are only-read register(code 0x03)
40001 0000 Present weight value(4bits including
sign characters, the high bit is in the
40002 0001 front)
D0(0 stable、1 unstable);D1(0
normal 、 1 overflow ); D2 ( 0
none-zero、1 zero)
D4(0 positive sign、1 negative sign);
D5 ( 0 sp1 ineffective 、 1 sp1
40003 0002 effective);D6(0 sp2 ineffective、1
sp2 effective);D7(0 sp3 ineffective、
1 sp3 effective ); D8 ( 0 sp4
ineffective、1 sp4 effective)
;
D9 ( 0 sp5 ineffective 、 1 sp5
effective) ;D10(0 sp6 ineffective、1
sp6 effective);
- 47 -
40004 0003
Reserve(permit to read, reading
…... ……
value is 0 )
4007 0006
The following items are two bytes and are available to read
and write (write code 0x06,read code 0x03)
Automatically zeroing when power on.
40008 0007 (F1.1)
- 50 -
SP6 output effective(0 ineffective;1
00066 0065 effective)
- 51 -
For example:Inquiry weighing indicator 01 to read 40- 43 coil
1) RTU communication mode:
Inquiry command:
Function Start Coil CRC
Address
code address quantity parity
1byte 1byte 2byte 2byte 2byte
Correct response:
Function Account CRC
Address code byte Data area parity
1byte 1byte 1byte 1byte 2byte
Inquiry command:01 01 00 28 00 04 BD C1
Correct response:01 01 01 02 D0 49(Coil 43- 40 related
state:0- 0- 1- 0)
2) ASCII communication mode:
Inquiry command:
Function Start Coil LRC
Start Address Stop
code address quantities parity
1byte 2byte 2byte 4byte 4byte 2byte 2byte
Correct response:
Function Account Data LRC
Start Address Stop
code byte area parity
1byte 2byte 2byte 2byte 2byte 2byte 2byte
Inquirycommand:3A 30 31 30 31 30 30 32 38 30 30 30 34 44
32 0D 0A
Correct response:3A 30 31 30 31 30 31 30 32 46 42 0D 0A
(Coil 43- 40 related status:0- 0- 1- 0)
- 52 -
◆ 03 read register
Inquiry:Regulate to read register start address and register
quantities.
Response:Regulate read register byte quantities, every
register relate 2bytes respectively;Message also has data
from every read register.
For example:read register 0007、0008。
1) RTU communication mode:
Inquiry command:
Inquiry
Function Start CRC
Address register
code address parity
quantity
1byte 1byte 2byte 2byte 2byte
Correct response:
Function Account Register Register CRC
Address
code byte (0007) data (0008) data parity
1byte 1byte 1byte 2byte 2byte 2byte
Inquiry command:01 03 00 07 00 02 75 CA
Correct response:01 03 04 00 00 00 05 3A 30(Register
0007、0008 respective data:0(Hex:0000H)、5(Hex:
0005H)
2) ASCII communication mode:
Inquiry command:
Inquiry
Function Start LRC
Start Address register Stop
code address parity
quantity
1byte 2byte 2byte 4byte 4byte 2byte 2byte
- 53 -
Correct response:
Register
Function Account Register LRC
Start Address (0007) Stop
code byte (0008) data parity
data
1byte 2byte 2byte 2byte 2byte 2byte 2byte 2byte
Inquirycommand:3A 30 31 30 33 30 30 30 37 30 30 30 32 46
33 0D 0A
Correct response:3A 30 31 30 33 30 34 30 30 30 30 30 30
30 35 46 33 0D 0A(register 0007、0008 respective data :
0(Hex:0000H)、5(Hex:0005H)
◆ 05 force single coil
Force:Regulate force coil address;and one normal item in
force data area,regulate requested coil ON/ OFF state: FF00
for ON,0000H for OFF,other data is ineffective to coil.
Response:Return normal response for force coil.
For example:Force weighing indicator 01, 0056 coil is ON
1)RTU communication mode:
Force command:
Function Coil CRC
Address Force data
code address parity
1byte 1byte 2byte 2byte 2byte
Correct response:
Function Coil CRC
Address Force data
code address parity
1byte 1byte 2byte 2byte 2byte
- 54 -
Force command:01 05 00 38 FF 00 0D F7
Correct response:01 05 00 38 FF 00 0D F7(Coil 0056 is
ON)
2)ASCII communication mode:
Force command:
Function Coil Force LRC
Start Address Stop
code address data parity
1byte 2byte 2byte 4byte 4byte 2byte 2byte
Correct response:
Function Coil Force LRC
Start Address Stop
code address data parity
1byte 2byte 2byte 4byte 4byte 2byte 2byte
Force command:3A 30 31 30 35 30 30 33 38 46 46 30 30 43
33 0D 0A
Correct response:3A 30 31 30 35 30 30 33 38 46 46 30 30
43 33 0D 0A(Coil0056 is ON)
◆ 06 preset single register
Preset:Regulate to preset single register address and data.
Response:Return normal response after register is preset.
For example:Request 0009 register in weighing indicator 01
is preset 0005H
1) RTU communication mode:
Preset command:
Function Preset register Preset CRC
Address
code address data parity
1byte 1byte 2byte 2byte 2byte
- 55 -
Correct response:
Function Preset register Preset CRC
Address
code address data parity
1byte 1byte 2byte 2byte 2byte
Preset command:01 06 00 09 00 05 99 CB
Correct response:01 06 00 09 00 05 99 CB(register 0009data:5
(Hex:0005H)
)
2) ASCII communication mode:
Preset command:
Preset
Function Preset LRC
Start Address register Stop
code data parity
address
1byte 2byte 2byte 4byte 4byte 2byte 2byte
Correct response:
Preset
Function Preset LRC
Start Address register Stop
code data parity
address
1byte 2byte 2byte 4byte 4byte 2byte 2byte
Preset command: 3A 30 31 30 36 30 30 30 39 30 30 30 35
45 42 0D 0A
Correct response:3A 30 31 30 36 30 30 30 39 30 30 30 35
45 42 0D 0A(register 0009 data: 5(Hex:0005H))
◆ 16(10 hex) Preset several registers
Preset:Regulate to preset register start address and preset
value.
- 56 -
Response:Return address, function code, start address and
preset register quantity.
For example:Preset 2 registers in weighing indicator 01:
start register is 0030. Preset value is 0001H and 7318H 1
)RTU communication mode:
Preset command:
Function Start Register Account Preset CRC
Address
code address quantity byte data parity
1byte 1byte 2byte 2byte 1byte 4byte 2byte
Correct response:
Function Start Register CRC
Address
code address quantity parity
1byte 1byte 2byte 2byte 2byte
Preset command:01 10 00 1E 00 02 04 00 01 73 18 07 D5
Correct response:01 10 00 1E 00 02 21 CE
2)ASCII communication mode:
Preset command:
Function Start Register Account Preset LRC
Start Address Stop
code address quantity byte data parity
1byte 2byte 2byte 4byte 4byte 2byte 8byte 2byte 2byte
Correct response:
Function Start Register LRC
Start Address Stop
code address quantity parity
1byte 2byte 2byte 4byte 4byte 2byte 2byte
Preset command:3A 30 31 31 30 30 30 31 45 30 30 30 32
- 57 -
30 34 30 30 30 31 31 43 39 36 31 38 0D 0A
Correctresponse:
3A 30 31 31 30 30 30 31 45 30 30 30 32 43 46 0D 0A
6.3.4 MODBUS communication error message
When weighing indicator check other error message except
parity code( CRC or LRC),indicator will send message to
host computer, the highest in function code is 1,which
means function code from weighing indicator to host
computer is added 128 based on function code from host
computer(For example: read register command 03H will be
changed to 83H).
Error code:
02: error data address;Received data address is not
permitted by weighing indicator.
03: error data;Inquiry data is not permitted by weighing
indicator.
For example:Host computer read coil (0040) by function
code 03.
Error message format:
1) RTU communication mode:
Address Function code Error code CRC parity
1byte 1byte 1byte 2byte
2) ASCII communication mode:
- 58 -
Function Error LRC
Start Address Stop
code code parity
1byte 2byte 2byte 2byte 2byte 2byte
- 59 -