Neutron Imaging and Its Applications - Book - 2009
Neutron Imaging and Its Applications - Book - 2009
1 Neutron Imaging
and Applications
Neutron Imaging and Applications
1 23
Neutron Imaging and Applications
Neutron Scattering Applications and Techniques
Series Editors:
Ian S. Anderson Alan J. Hurd
Neutron Sciences Directorate Lujan Neutron Scattering Center at LANSCE
Oak Ridge National Laboratory Los Alamos National Laboratory
Building 8600, MS 6477 PO Box 1663, MS H805
Oak Ridge, TN 37831 Los Alamos, NM 87545
USA USA
[email protected] [email protected]
Robert L. McGreevy
ISIS
Science and Technology Facilities Council
Rutherford Appleton Laboratory
Harwell Science and Innovation Campus
Chilton, Didcot OX11 0 QX
UK
[email protected]
Hassina Z. Bilheux
Editors
Neutron Imaging
and Applications
A Reference for the Imaging Community
13
Editors
Ian S. Anderson Robert L. McGreevy
Oak Ridge National Laboratory Rutherford Appleton Laboratory,
Oak Ridge, TN, USA Oxfordshire, UK
[email protected] [email protected]
Hassina Z. Bilheux
Oak Ridge National Laboratory
Oak Ridge, TN, USA
[email protected]
Cover illustration: Cover graphics courtesy of Dr John Katsaras, National Research Council, Chalk
River Laboratories, and Dr Taner Yildirim, NIST Center for Neutron Research.
springer.com
Preface
The first successful experiments in Neutron Radiography were carried out (to
our knowledge) in 1935, just a few years after the discovery of the neutron, by
H. Kallmann and E. Kuhn using a small neutron generator [1]. Not
surprisingly, the field has developed and diversified over the last 70 years so
that neutron imaging, in the broad sense, is now routinely used in a wide range
of applications. The aim of this book (one of a series on the applications of
neutron scattering [2]) is to introduce to the reader, whether novice or
experienced researcher, the basic techniques used to image objects using
neutron beams and to give a flavor of the vast range of applications where
these imaging capabilities provide unique insight (Fig. 1).
Fig. 1 Left to right: Pressure gauge with metal backplate; fire hydrant and test tubes filled with H2O
and D2O imaged with gamma-rays (top) and neutrons (bottom) [3]
v
vi Preface
itself. Even the most powerful neutron sources in existence today have a source
brightness that is comparable to a simple X-ray tube and many orders of
magnitude lower than a third-generation synchrotron X-ray source. Hence,
while synchrotron X-ray sources provide the capability of imaging single nano-
particles with nanometer resolution, or dynamic images of larger objects with
micrometer spatial resolution and microsecond time resolution, neutrons are
presently limited to static images with spatial resolutions of the order of tens of
microns, or dynamic images of 100 microns and microsecond exposure times
for stroboscopic processes. Neutron imaging of smaller objects can only be
achieved indirectly, using scattering techniques from an ensemble of particles.
Despite these limitations, the following chapters give a flavor of the wide
range of applications that presently (or will potentially) benefit from neutron
imaging techniques. The book is organized into three major sections.
Section A provides a comprehensive overview of basic neutron techniques
aimed more specifically at a non-specialist audience. Frequent reference is made
to the two introductory chapters in the first book of this series [2] by Roger
Pynn (Neutron Scattering – a Non-Destructive Microscope for Seeing Inside
Matter), and Helmut Schober (Neutron Scattering Instrumentation). Both these
chapters are freely available at www.springerlink.com. In Chapter 1 Kenneth
Herwig summarizes the essential neutron properties and techniques which are
relevant to the majority of neutron imaging applications. Masatoshi Arai and
Kent Crawford provide, in Chapter 2, an excellent survey of the different types
and characteristics of neutron sources, including nuclear reactors, high-power
spallation sources, and portable generators, which are typically used nowadays
for neutron imaging. Although there is considerable overlap in the use of these
sources, each type of source has specific advantages for certain types of applica-
tions. Due to the low-intrinsic brightness of neutron sources, efficient optical
systems are imperative, so Ken Andersen’s chapter (Chapter 3) presents the basic
concepts of the neutron optics that are typically used on imaging beam lines.
Finally in this section, Lowell Crow (Chapter 4) examines modern neutron detec-
tion methods for imaging. The first section reviews neutron capture converters
which form the basis for thermal neutron detection, and the following sections
examine detector systems with an emphasis on imaging applications.
Section B focuses on the neutron beam implementation of some well-known
imaging techniques. Arthur Heller and Jack Brenizer (Chapter 5) present a
summary of the history, methods, and related variations of neutron radio-
graphy techniques including a section on the application of standards. Even
today, conventional film radiography remains the mainstay of high-resolution,
large field-of-view, neutron imaging. In Chapter 6, Wolfgang Treimer extends
the basic theories and applications to include three-dimensional tomography
and introduces some of the newer methods for enhancing contrast such as
wavelength dependent (‘‘Bragg edge’’) imaging and small-angle scattering.
Kenneth Tobin et al. (Chapter 7) provide a review of neutron image formation,
resolution analysis concepts, and methods for both the design and
characterization of radiography systems and conclude with a discussion of
x Preface
References
1. H. Kallman, Research 1, 254 (1947).
2. http: //www.springer.com/series/8141, this series of books.
3. O. Peter, Naturforsch. I(10), 557 (1946).
4. G. B. Winkelman, C. Dwyer, T.S. Hudson, D. Nguyen-Manh, M. Döblinger, R.L. Satet,
M.J. Hoffmann, D.J.H. Cockayne, Phil. Mag. Lett. 84, 755–62 (2004).
Contents
Section A
3 Neutron Optics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
K. Andersen
Section B
5 Neutron Radiography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
A.K. Heller and J.S. Brenizer
6 Neutron Tomography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
W. Treimer
xiii
xiv Contents
Section C
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 339
Contributors
xv
xvi Contributors
K.W. Herwig
1.1 Introduction
As one of the constituents of the atom, the neutron was discovered well
after its electrically charged companions the electron, by J. J. Thomson in
1897 [1], and the proton, for which there is some debate over any individual
discoverer. A close combination of an electron and a proton was proposed
as a neutral doublet by E. Rutherford in 1920, which he postulated would
have ‘‘novel properties’’ including the ability to ‘‘move freely through mat-
ter,’’ be difficult to detect, and perhaps impossible to contain in a vessel [2].
The term ‘‘neutron’’ was apparently first recorded in the literature in 1921
by W. D. Harkins in a discussion on classifying isotopes [3]. However,
conclusive evidence for the existence of the neutron waited until 1932
when J. Chadwick reported first the ‘‘possible existence’’ [4] and shortly
thereafter ‘‘the existence’’ of the neutron [5]. Chadwick was recognized for
his discovery by receiving the 1935 Nobel Prize in Physics quite simply ‘‘for
h 395:6
l¼ ¼ (1:1)
mn v v
1
E ¼ mn v2 ¼ 5:2270 106 v2 : (1:2)
2
I(x) I(x+dx)
dx
Fig. 1.1 Illustration of the attenuation of an incident beam of idealized neutrons by a thin
sample
6 K.W. Herwig
attenuation of the incident neutron beam is then given by the fractional area
occupied by the classical atoms, dx n p r2 and
Iðx þ dxÞ ¼ IðxÞ 1 dx n p r2 ¼ IðxÞð1 dx nÞ : (1:3)
In this classical picture, the microscopic cross section, , is just the cross
sectional area of the individual sample particles, p r2, and has units of cm2.
More generally, the microscopic cross section is a measure of the effective
interaction area for the neutron with a nucleus.
From Eq. (1.3), the rate of change of I(x) is then
dIðxÞ
¼ n I ðxÞ ; (1:4)
dx
where the sum is over all isotopes in the sample and the transmitted neutron flux
may be written as
t ¼ s þ a (1:8)
where s and a are the scattering and absorption cross sections, respectively.
An in-depth introduction to neutron scattering is provided by R. Pynn [9], so
only a few relevant terms are presented here. Because the neutron interaction
is with the nucleus, the strength of the scattering cross section varies non-
uniformly across the periodic table and depends not only on the individual
element but also on the particular isotope of that element. Figure 1.2 shows
the total neutron scattering and absorption cross sections for bound atoms
across the periodic table as derived from the tabulation in [10]. In practice, the
scattering cross sections are typically taken as constant in the epithermal,
thermal, and into the cold neutron energy ranges. In general, however, these
cross sections have an energy dependence with gadolinium (due to the close
proximity of a nuclear resonance) and bound hydrogen (e.g., in the case of
water) that exhibits significant variation. (See [11] for a short discussion of this
effect.)
The neutron scattering cross section can be further broken down into
coherent and incoherent components. In a coherent scattering process,
neutron waves that are scattered by different nuclei combine with one
another to produce an interference pattern that depends on the relative
locations of the atoms in the material. Incoherent scattering may arise
when more than one isotope is present in the sample (whether the same
element or not) or when a given isotope has a non-zero nuclear spin [12]. In
Fig. 1.2 Neutron scattering and adsorption cross sections for thermal neutrons (v = 2200 m/s)
as tabulated in [10]. Open circles are adsorption cross sections and are plotted on the logarithmic
scale of the right y-axis. Closed circles are the bound atom scattering cross sections plotted on
the linear scale of the left y-axis
8 K.W. Herwig
2 H atoms
n¼ NA ¼ 6:691 1022 ; (1:9)
M cm3
where M is the molecular weight (18 g mol–1 for water), 2 is for the two
hydrogen atoms/water molecule, and NA is Avogadro’s number, 6.022 1023
mol–1. The macroscopic cross section due to scattering from hydrogen in water
is SH = 5.353 cm–1. In comparison, the macroscopic scattering cross sections
are SFe = 0.99 cm–1and SAl = 0.091 cm–1 for iron and aluminum, respectively.
Figure 1.3 shows the attenuation of a neutron beam due to these scattering cross
sections as a function of depth x into a sample. What is very clear from Fig. 1.3
is that the presence of hydrogen in a specimen gives rise to a very rapid
attenuation of the neutron beam. This has been used to great advantage recently
in real-time identification of the formation and motion of water in operating
proton exchange membrane fuel cells (see Chapter 11 and, e.g., [13]). The high
penetration depth of the neutron in aluminum and iron is also illustrated
(Fig. 1.1 of [9]).
Real materials are much more complicated than the cases determined above
for simple attenuation; this is particularly evident in the energy dependence of
the total scattering cross section for polycrystalline materials. Figure 1.4 shows
the measured elastic cross section for polycrystalline iron [14]. The coherent
nature of the scattering and the regular positions of the iron atoms in the
1 Introduction to the Neutron 9
Fig. 1.3 Attenuation as a function of depth into samples of water, iron, or aluminum due to
neutron scattering
Fig. 1.4 The total neutron scattering cross section for polycrystalline iron
crystalline lattice give rise to abrupt changes in the cross section and associated
neutron transmission. Associated with a given spacing between planes of atoms
in the material, d is a Bragg reflection whose scattered intensity is determined by
the local arrangement of atoms within a unit cell. The condition for neutron
scattering is determined by Bragg’s law
nl ¼ 2d sinðÞ ; (1:10)
10 K.W. Herwig
where is half the angular separation between the incident and scattered
neutron beams. There is clearly a maximum neutron wavelength lmax for any
given d at which the condition of Eq. (1.10) is satisfied, namely lmax = 2d at
= 908. For shorter wavelengths and lower scattering angles, there will also be
neutron scattering that satisfies this condition. Thus there is a sharp edge at lmax
for each family of reflections allowed by the crystal structure, followed by a
decreasing tail at shorter l as seen in Fig. 1.4. This wavelength dependence of
the neutron cross section and associated transmission can be used to great
advantage in optimizing the conditions for high contrast from various materials
in a sample (Neutron Tomography (Chapter 6) and Material Science and
Engineering with Neutron Imaging (Chapter 12) and, e.g., [15]). Radiographs
at different neutron wavelengths can be obtained through either time-of-flight
means (spallation neutron sources) or monochromatization (reactor sources).
Neutron absorption, another neutron–nuclei interaction that attenuates an
incident neutron beam, is a mechanism whereby contrast can be obtained. In
this interaction, a neutron is absorbed by a nucleus, often resulting in an
unstable, radioactive nucleus that will decay with a particular half-life. The
half-life is the time required for the radioactivity to decay to half of its original
value. As the nucleus decays, it may emit a variety of secondary radiation,
including an -particle, which is identical to a helium nucleus and carries two
units of positive charge; a -particle, which is an energetic electron or positron;
-rays, which are high-energy photons originating in the nucleus; and others.
An -particle interacts strongly with matter and is rapidly attenuated even in
air, whereas the latter two can penetrate relatively deeply into matter. Emitted
–rays have a well-defined and catalogued energy spectrum that can act as a
signature for the decay of particular isotopes. One mature technique is neutron
activation analysis, whereby decay radiation, either delayed or prompt, is
detected from nuclei that have absorbed a neutron. Prompt decay gives rise to
an imaging paradigm whereby a sample rastered through a neutron beam can
enable a quantitative measure of the presence of particular isotopes in the
volume sampled by the neutron beam (Novel Neutron Imaging Techniques
for cultural Heritage Objects (Chapter 13)). An energy sensitive -detector can
register the energy of the promptly emitted , yielding a fingerprint of the
isotope. In the low epithermal through cold neutron energy ranges, these
absorption processes follow a 1/v probability law for most nuclei. The value
varies across the periodic table as demonstrated in Fig. 1.2 for a neutron
velocity of 2200 m/s. The probability of absorption then increases linearly
with increasing neutron wavelength. Exceptions include gadolinium, which
has many resonances; the lowest occurs at approximately 2 eV, with a corre-
sponding nonlinear response in the scattering cross section as well.
The half-life of the activated isotopes may be long enough that it is practical
to remove an object exposed to a neutron beam and use its residual activity to
expose either a photographic film or an image plate. The radiation emitted from
particular isotopes will be present to a lesser or greater degree after neutron
irradiation, depending on their respective half-lives and the intervening time
1 Introduction to the Neutron 11
before the film or image plate is exposed by the sample. Multiple exposures at
subsequent times will be more sensitive to the longer-lived isotopes as the
shorter-lived ones decay to low levels of activity, giving rise to an additional
means of obtaining isotopic contrast.
An additional neutron–nuclei interaction involves a high-energy neutron
(MeV) that scatters inelastically, imparting a relatively large energy to the
nucleus. The excited nucleus may give up the energy in a quantized fashion
through -emission. Similarly to prompt detection, an energy-sensitive detec-
tor can be used to identify the signatures of particular isotopes. This is the
underlying principle of neutron stimulated emission computed tomography,
described in Chapter 15 of this book.
Finally, to this point, only neutron interactions with nuclei have been con-
sidered; but, as mentioned, the neutron also carries a magnetic moment that can
interact with magnetic fields. This provides an additional means by which
neutrons can be scattered out of an incident beam as they interact with the
magnetic fields produced by unpaired electrons in materials. This interaction
may be elastic, giving rise to magnetic Bragg peaks that yield information on the
magnetic order in a sample, or inelastic, giving insight into fluctuations. A
discussion of the use of magnetic scattering and the production and use of
polarized neutron beams is given by R. Pynn in his presentation on neutron
scattering techniques [9]. Some examples of applications to neutron imaging are
given in Chapter 10.
In conclusion, the neutron has several intrinsic properties that make it very
useful for imaging purposes and result in a variety of contrast mechanisms. It is
uncharged and is able to penetrate deeply into matter in a non-destructive
fashion. The neutron cross sections for interactions with nuclei are both isotope
and neutron energy dependent and include processes that attenuate a neutron
beam by either scatter or absorption. It has a wavelength/energy relationship
such that the neutron simultaneously responds to both length and time scales
relevant to condensed matter at room temperatures. Finally, the activation of a
nucleus upon neutron absorption can yield emission of -rays having energies
that are isotope dependent.
References
1. J.J. Thomson, Cathode Rays, The Electrician 39, 104–109 (1897).
2. E. Rutherford, Proc. Roy. Soc. A 97, 374–400 (1920).
3. W.D. Harkins, Natural systems for the classification of isotopes, and the atomic weights of
pure atomic species as related to nuclear stability, J. Am. Chem. Soc. 43, 1038–1060 (1921).
4. J. Chadwick, Possible existence of a neutron, Nature 129, 312 (1932).
5. J. Chadwick, The Existence of a neutron, Proc. Roy. Soc. A 136, 692–708 (1932).
6. All values for fundamental constants taken from the NIST Reference on Constants, Units,
and Uncertainty, available at http://physics.nist.gov/cuu/Constants/.
7. D.L. Price and K. Sköld in Neutron Scattering, edited by K. Sköld and D.L. Price
(Academic Press, Orlando, Fl., 1986), Vol. A.
12 K.W. Herwig
Abstract This chapter provides a brief survey of the types of neutron sources
that are available for imaging applications, primarily focusing on high-flux
sources such as reactors or spallation sources but also including smaller or
portable sources based on radioactive decay or small accelerators. Although
nearly all neutron imaging studies performed so far have used portable or
reactor-based sources, spallation sources will be of increasing importance in
the future. Their designs are less familiar and somewhat more complex, so the
chapter concentrates on them.
2.1 Introduction
One might think it is not necessary to have any detailed knowledge of the
neutron source in order to use the techniques of neutron scattering or imaging.
To an extent this is true. However, the distribution of neutrons from the
source in terms of energy and time, and the distribution of ‘‘background’’
(fast neutrons, delayed neutrons, gammas, . . .), has a direct bearing on the
design of neutron scattering and imaging instruments and their performance,
on how measured ‘‘raw’’ data must be corrected to make them scientifically
meaningful, and on the types of measurements that can be undertaken. Even if
the average user does not design the instrument, the user needs to choose the
source/instrument to be used for a particular experiment or measurement, so
M. Arai (*)
Japan Proton Accelerator Research Complex Center, Japan Atomic Energy Agency,
Japan
e-mail: [email protected]
some knowledge of the different types of sources is advisable. There are two
kinds of neutron sources for neutron scattering and high-resolution/high-flux
imaging facilities: (1) reactor sources and (2) large accelerator-based neutron
sources. However, low-resolution/low-flux imaging can also be carried out
with small radioactive or accelerator-based sources, which can be fixed or
portable.
Nuclear reactors use the fission process to produce neutrons. Most of the
current reactor sources for scattering applications were built in the 1960s and
1970s and were primarily designed for materials testing for the nuclear
industry, providing medium flux. The best reactor source, optimized for
neutron scattering applications, is still the High-Flux Reactor (HFR) at the
Institut Laue-Langevin (ILL), built in 1972 in Grenoble, France [1, 2].
Reactors coming into operation more recently include JRR3 at the Japan
Atomic Energy Agency (1990) [3], HANARO at the Korea Atomic Energy
Research Institute (1997) [4], FRM-II in Munich, Germany (2004) [5], OPAL
at the Australian Nuclear Science and Technology Oganisation (2006) [6],
and the China Advanced Research Reactor in Beijing, China (potential
operation in 2008) [7]. With the exception of the HFR, these are all med-
ium-sized research reactor sources (typical power 10–20 MW) built with
advanced technology.
Electron accelerators can produce neutrons in a target material using the
Bremsstrahlung photo-neutron reaction. Although electron accelerators are
relatively inexpensive to construct, the large amount of heat dissipated in
the target per neutron produced severe limits on the potential performance.
Neutrons can also be produced by the spallation process, in which high-energy
protons strike a solid target. The development of proton accelerator technology,
driven by other applications such as particle physics, helped to increase the
potential power and hence the neutron flux of spallation sources. The pulsed
nature of most accelerator-based neutron sources can offer a significant advan-
tage in experiments using the time-of-flight (TOF) method, in which the speed of
the neutron is measured by timing its flight from the source to the detector. Pulsed
proton-driven neutron sources recently completed or under construction include
the Spallation Neutron Source (SNS) in the United States [8, 9], the Japanese
Spallation Neutron Source (JSNS) of the Japan Proton Accelerator Research
Complex (J-PARC) project in Japan (2008) [8, 9], and the China Spallation
Neutron Source (CSNS) in China (under construction) [10, 11].
Both fission and spallation produce neutrons in the megaelectron volt energy
range as discussed in Section 2.2. However, neutron scattering and most
neutron imaging applications require neutrons at electron volt or lower
energies. Moderators are used to slow the neutrons to these energies, as
described in the section ‘‘Moderation Mechanisms.’’ Reactor technology is
highly developed and has not changed significantly for many years. The tech-
nologies for accelerator-driven sources offer many more parameters that can be
varied to optimize the source for different purposes, so more space will be
devoted to describe these options in this chapter.
2 Neutron Sources and Facilities 15
2.2.1 Reactors
235
U þ n ! X þ Y þ 2:5 n ;
(2:1)
ð 200 MeV total energy release; 2 MeV per neutronÞ
(a)
(b)
Fig. 2.1 (a) Layout of the reactor pool of FRM-II. The reactor core is very compact—24 cm in
diameter—so that the maximum neutron flux is achieved at the moderator position, as shown
in (b), giving a very high flux for instruments [5]
magnitude less heat must be dissipated than in a fission reaction producing the
same time-averaged neutron flux. Spallation reactions occur for proton ener-
gies above 100 MeV. High-energy neutrons, pions, and spalled nuclei cause
inter-nuclear cascades followed by low-energy neutron evaporation from the
excited nuclei, as illustrated in Fig. 2.2.
Table 2.1 Existing medium- and high-flux reactor sources and their respective parameters
United United
Country States States Canada France France Germany Germany Australia Korea Japan
Neutron HFIR NBSR NRU HFR ORPHEE BENSC FRM-II OPAL HANARO JRR-3 M
source
Organization Oak Ridge National Atomic Institut Laue- Laboratoire Helmholtz- Technische Australian Korea Atomic Japan
National Institute of Energy of Langevin Léon Zentrum Universitat Nuclear Energy Atomic
Laboratory Standards and Canada Brillouin Berlin Munchen Science and Research Energy
Technology Limited Technology Institute Agency
Organization
Power (MW) 85 20 120 58 14 10 20 20 24 (present) 30 20
(designed)
Flux (ncm–2 1.51015 31014 31014 1.51015 31014 21014 81014 31014 21014 31014
2 Neutron Sources and Facilities
s–1)
Number of 1/0 1/0 0/0 2/1 1/1 1/0 1/1 1/0 1(planned)/0 1/0
cold/hot
sources
Number of 9(present) 24 5 26 22 22 20 (present) + 6 6 24
instruments +6 10 (under
(planned construction)
by 2012)
Existing BT-2 [14] [15] CONRAD ANTARES NR-port [18] TNRF [19]
neutron [16] [17] And
imaging TNRF-2
instrument [20]
Facility 1967 1970 1957 1972 1980 1973 2004 2006 1997 1990
operating (refurbished
since 1993)
HFIR: High-Flux Isotope Reactor [21]; NBSR: National Bureau of Standards Reactor [22]; NRU: National Research Universal Reactor, Chalk River, Canada [23]; HFR:
High-Flux Reactor at ILL [1, 2]; ORPHEE: reactor at LLB [24]; BENSC: Berlin Neutron Scattering Centre [25]; FRM-II: Forschungsneutronenquelle Heinz Maier-Leibnitz
[5]; OPAL: Open Pool Australian Light-water Reactor [6]; HANARO: High-flux Advanced Application Reactor [4]; JRR-3 M: Japan Research Reactor No. 3 Modified [3].
Consult the web sites for these facilities to obtain additional information and current details. A number of smaller research reactors, primarily at universities, are not listed here.
17
18 M. Arai and K. Crawford
Fig. 2.3 Neutron energy spectrum from mercury and PbBi alloy targets [27]
stopping length of a proton is greater than three times its mean free path,
neutron production efficiency becomes close to 100%, so the proton energy
should be as high as 1 GeV or greater.
In the early days of design work on spallation sources, proton energies were
less than 1 GeV because of the lack of experimental experience at higher energies.
However, more sophisticated codes and experiments in the 1990s demonstrated
that the neutron production rate is almost proportional to the accelerator power
even at 12 GeV [28, 29], as shown in Fig. 2.4. This resulted in flexibility in
optimizing accelerator and neutron target design. Proton current and energy
are equally optimized for beam experiments. Current and energy for spallation
sources are listed in Table 2.2 along with some other parameters to be discussed.
For a short-pulse neutron source, the repetition rate of the proton acceleration
is an important parameter to be considered. When a long neutron flight path is
used for improved TOF resolution, a slow repetition rate is important to mini-
mize frame overlap (where the fast neutrons from one pulse overlap with the slow
neutrons from the previous pulse). However, if the same time-averaged power is
maintained, a lower repetition rate requires more power per pulse, creating a
more difficult engineering problem for the accelerator and target.
Most accelerator-based neutron sources are pulsed, and heat is produced in
the target only during the pulses. This allows the heat to dissipate slowly in the
period between pulses, so the instantaneous power and neutron flux can be very
high. However, thermal shock in the target remains a problem to be overcome
at the highest levels of proton power. Building spallation neutron sources
Fig. 2.4 Neutron yield from a lead target as a function of proton energy. Neutron yield is
almost proportional to proton energy at up to 12 GeV. Here the symbols show experimental
results, whereas the lines show the results of various calculations [28, 29]
20
Table 2.2 Past, existing, and future spallation source and their respective parameters
United United United
Country States States States U.K. Switzerland China Europe Japan Japan
Neutron source IPNS LANSCE SNS ISIS SINQ CSNS ESS KENS JSNS
Organization Argonne Los Alamos Oak Ridge Rutherford Paul Scherrer Institute Undecided High Energy Japan
National National National Appleton Institute of High Accelerator Atomic
Laboratory Laboratory Laboratory Laboratory Energy Research Energy
Physics Organization Agency
Proton energy (MeV)/ 450/15 800/70 1000/1400 800/200 590/1500 1600 1333/7500 500/9 3000/333
Current (mA)
Proton beam power 7 kW 56 kW 1.4 MW 160 kW 1 MW 100 kW 5 MW 4.5 kW 1 MW
Repetition rate (Hz) 30 20 60 50/10 (2 Continuous 25 16 (long 20 25
targets) pulse)
Target material Depleted Tungsten Mercury Tantalum Zircaloy Tungsten Mercury Tungsten Mercury
Uranium
Moderator S-CH4/L- L-H2/H2O L-H2/H2O L-H2/L-CH4/ L-D2/D2O H2OL- L-H2 S-CH4/H2O L-H2
CH4 H2O CH4L-
H2
Number of 12 7 24 (beam 22 (TS1) 15 20 (beam 15 23 (beam
instruments ports) 7 (TS2) ports) ports)
Existing neutron NEUTRA
imaging instrument [30] and
ICON [31]
Facility operating 1981 (closed 1983 2006 1985 (TS1) 1996 2014 Under 1980 (closed 2005) 2008
since or planned to 2008) 2008 (TS2) planning
operate in
IPNS: Intense Pulsed Neutron Source [32]; LANSCE: Los Alamos Neutron Science Center [33]; SNS: Spallation Neutron Source [8, 9]; ISIS: [34, 35]; SINQ: Swiss Spallation
Neutron Source [36, 37]; CSNS: Chinese Spallation Neutron Source [10, 11]; ESS: European Spallation Source [38, 39]; KENS: Koh-Energy-ken Neutron Source [40, 41];
JSNS: Japanese Spallation Neutron Source [8, 9]. Consult the websites for these facilities to obtain additional information and current details.
M. Arai and K. Crawford
2 Neutron Sources and Facilities 21
Storage
ring
Linac
Target
Station
Fig. 2.5 A typical short-pulse spallation neutron source, the SNS facility, (Oak Ridge
National Laboratory, United States) [8, 9]
Long-pulse sources, typically having 1-ms proton pulse widths, are another
option rapidly gaining favor for neutron sources not requiring such high timing
resolution. No accumulator ring is required for such sources, so a very high-
intensity direct proton beam can be delivered from the linac to the neutron target
[43]. Continuous cyclotron-based spallation sources, such as that at the SINQ at
the Paul Scherrer Institute in Switzerland [36, 37], produce continuous neutron
beams; their performance is similar to that of a medium-flux reactor. Table 2.2
shows spallation neutron sources either in operation or being planned.
The cross section for neutron-induced fission is much higher for thermal
neutrons (100 meV) than for the fast neutrons (1–2 MeV) that are produced.
For a reactor to achieve a self-sustaining chain reaction from a small mass of
fissile material, and to obtain suitable neutrons for neutron scattering, the fast
neutrons within the core must be ‘‘slowed down.’’ This is done using a
22 M. Arai and K. Crawford
moderator, which is usually also the coolant. A moderator reduces the neutron
energy by inelastic scattering and so is preferably a material containing light
elements, e.g., H2O or D2O. Hydrogen has a large scattering cross section and a
large absorption cross section, so when H2O coolant is used, the core needs to be
relatively compact to achieve a high-flux density at the neutron beam tubes.
These beam lines are directed tangentially to the cylinder of maximum thermal
flux in order to reduce the background of high-energy neutrons and gammas, as
illustrated in Fig. 2.1. With D2O, a larger core can be used (meaning easier
cooling because heat can be dissipated more easily). Surrounding the moderator
there may be ‘‘reflector’’ materials that do not provide much moderation but
scatter or reflect some of the fast neutrons back into the moderator to enhance
the thermalized neutron flux output.
The neutron energy distribution can be altered from that produced by the
reactor coolant/moderator or reflector/moderator materials and design by stra-
tegic placement of special moderators, shifting the neutron energy spectrum to
either slightly lower energies (a ‘‘cold source’’—e.g., liquid hydrogen) or higher
energies (a ‘‘hot source’’—e.g., carbon heated by gamma radiation from the
reactor). Such special moderators expand the range of usable neutron energies
and hence expand the research capabilities of the reactor facility. Table 2.1 shows
which of the research reactor facilities provide such capabilities.
Fig. 2.6 The neutron energy distribution (flux) of the J-PARC neutron source for coupled,
decoupled, and poisoned decoupled moderators. The flux consists of a Maxwell distribution
at low energies and a 1/E region at higher energies [45]
The pulse width is a key parameter for short-pulse spallation sources and is
directly influential on neutron beam instrument performance. It is almost
proportional to the neutron wavelength in the 1/E region of flux, is broadened
in the thermal equilibrium region, and then saturates in the very low-energy
region. The broadening starts to occur at about 300 meV (neutron wavelength
of 0.5 Å) for an ambient-temperature moderator and at about 15 meV (2.5 Å)
for a methane moderator at 20 K. This is clearly seen in Fig. 2.7. In the 1/E
region for each moderator, the pulse width, t, is proportional to wavelength,
l, as [46]
2
t½s pffiffiffiffiffiffiffiffiffiffiffiffi 7l½— : (2:2)
E½eV
24 M. Arai and K. Crawford
Fig. 2.8 Typical pulse structure from a short-pulse spallation source. The sharp rising edge
comes from the slowing down component followed by the storage component with a long time
tail [47]
Cold coupled moderators have been extensively optimized for pulsed sources
by Watanabe and Kiyanagi, greatly enhancing neutron flux from the cold
neutron moderator while minimizing heat deposition in the moderator
(Fig. 2.9) [48].
Fig. 2.9 Design of a coupled moderator for J-PARC. The hydrogen moderator is surrounded
by a water premoderator, which reduces neutron energy in the first stage and removes heat
before neutrons go into the hydrogen moderator [48]
26 M. Arai and K. Crawford
Diffraction peaks
Flight length
Scattering
angle
Detector
L2
L2
Sample
Sample
Detectors
L1
L1
Source
Time of Flight
half-life of 85.5 years. The average neutron energy is 2.14 MeV, and the rate is
2.34 1012 ns–1 g–1. Such sources are produced by irradiation in a high-flux
reactor.
A more common method is neutron production as a secondary process, as
used by Chadwick when he first discovered the neutron. For example, in an -n
source such as 241Am/Be, 241Am undergoes -decay; the -particle can be
absorbed by a light element such as beryllium, which then decays by neutron
emission. This can be written as
241
Am ! 237 Np þ 4 He ð5:6 MeVÞ followed by
(2:3)
9
Be þ 4 He ! 12 C þ n ðfew MeVÞ :
The half-life is 433 years, and these sources can produce 106–108 n s–1 g–1 Am.
An alternative is a -n source, for example
124
Sb !124 Te þ hð1:7 MeVÞ followed by 9 Beþ
(2:4)
h ! 2 4 He þ n ðfew MeVÞ :
The half-life is 60 days, and such sources can produce 109–1010 n s–1.
Radioactive decay sources have the advantage of being small and highly
portable, but they have low intensity and are always ‘‘on.’’ They can be used for
testing (e.g., of neutron detectors), in medicine (e.g., activation analysis, cancer
treatment with 252Cf needles), and for low-resolution/low-flux radiography.
Higher intensities can be produced by small accelerator-based neutron
sources. Over the years, these have evolved sufficiently that compact portable
sources are now commercially available from a range of vendors. They are
normally based on the ‘‘D-T’’ reaction:
2
Hð 150 keVÞ þ3 H ! 4 He þ n ð14:2 MeVÞ : (2:5)
References
1. W.J. Drexel, IEEE Trans. Nucl. Sci. NS-29, 123–126 (1982).
2. C.J. Carlile, Physica B 385–386, 961–965 (2006).
3. Fumio Sakurai, Yoji Horiguchi, Shinsho Kobayashi, and Masaji Takayanagi, Physica B
311, 7–13 (2002).
4. Chang-Hee Lee, Young-Hwan Kang, and Il-Hiun Kuk, ICANS-XV Proceedings of the
15th Meeting of the International Collaboration on Advanced Neutron Sources, Tsukuba,
Japan, 146–154 (2000).
5. J. Neuhaus and W. Petry, Neutron News 18:2, 13–15 (2007).
6. Shane J. Kennedy, Physica B 385–386, 949–954 (2006).
7. D.F. Chen, Y.T. Liu, C. Gou, and C.T. Ye, Physica B 385–386, 966–967 (2006).
8. T.E. Mason, et al., Physica B 385–386, 955–960 (2006).
9. Tony A. Gabriel, John R. Haines, and Thomas J. McManamy, J. Nucl Mater. 318, 1–13
(2003).
10. J. Zhang, Q.W. Yan, C. Zhang, P.L. Zhang, S.N. Fu, F.W. Wang, Z. Zhang, and S.X.
Fang, J. Neutron Res. 13:1, 11–14 (2005).
11. Q.W. Yan, W. Yin, and B.L. Yu, J. Nucl. Mater. 343, 45–52 (2005).
12. Colin D. West, Physica B 174, 430–437 (1991).
13. A. Lawler, Science 279, 470–471 (1998).
14. D. Hussey et al., Nucl. Instrum. Methods Phys. Res. A 542 (2005) 9–15.
15. http://www-llb.cea.fr/neutrono/nr1.html.
16. A. Hilger et al., Physica B 385–386 (2006) 1213–1215.
17. E. Calzaba et al., Nucl. Instrum. Methods Phys. Res. A 542 (2005) 38–44.
18. I.-C. Lim, M.-S. Kim, K.-Y. Nam, C.-M. Sim, B.-C. Lee, H.-Y. Choi, and S.-Y.Hwang,
Characteristics of neutron beam from IR beam port of HANARO for its application to
dynamic neutron radiography, in Proceedings of the 2002 ANS RPD Topical Meeting, Int.
Conf. New Frontiers of Nuclear Technology, Seoul, Korea, Oct. 2002.
19. M. Matsubayashi and A. Tsuruno, JRR-3 Neutron Radiography Facility, Proceedings of
the 4th World onf. Neutron Radiography, San Francisco, California, May 10–16, 1992, p.
415, Gordon & Breach Science Publishers.
20. M. Matsubayashi et al., Nucl. Technol. 132 (2000) 309.
21. S.E. Nagler and H.A. Mook, Neutron News 19:2, 12–13 (2008); also other articles this
issue.
22. R.L. Cappelletti, C.J. Glinka, S. Krueger, R.A. Lindstrom, J.W. Lynn, H.J. Prask,
E. Prince, J.J. Rush, J.M. Rowe, S.K. Satija, B.H. Toby, A. Tsai, and T.J. Udovic,
J. Res. Natl. Instrum. Stand. Technol. 106, 187–230 (2001).
23. J. Root, Neutron News 14:2, 12–13 (2003).
24. P. Monceau and M. Alba, Neutron News 14:3, 10–12 (2003).
25. R. Michaelsen, Neutron News 12:3, 7–9 (2001).
26. N. Watanabe, Rep. Prog. Phys. 66:339–381 (2003).
27. M. Teshigawara et al., JAERI-Res. 99–0101 (1999).
28. M. Arai et al., J. Neutron Res. 8, 71–83 (1999).
29. J.M. Carpenter, et al., Physica B 270, 272–279 (1999).
30. E. Lehmann, et al., Nucl. Instrum. Meth. A 377 (1996) 11.
31. G. Kuehne et al., Nucl. Instrum. Methods Phys. Res. A 542 (2005) 264–270.
30 M. Arai and K. Crawford
32. Raymond G. Teller, James W. Richardson, and John M. Carpenter, Neutron News 15:3,
9–13 (2004).
33. J.A. Roberts, Neutron News 10:4, 11–14 (1999).
34. R. Eccleston and C. Wilson, Neutron News 15:1, 15–18 (2004).
35. Penfold, Neutron News 15:2, 9–12 (2004).
36. W. Wagner, J. Mesot, P. Allenspach, G. Kuehne, and H.M. Rønnow, Physica B 385–386,
968–971 (2006).
37. W. Wagner and H. Hayck, Proc. ICANS-XVI, ISSN 1433-559X, 53–66 (2003).
38. E. Cartlidge, Phys. World 10, June, 10 (2007).
39. D. Richter, Appl. Phys. A 74 [Supl.], S18–S22 (2002).
40. Ishikawa, ICANS-IV Proceedings of the 4th Meeting of the International Collaboration on
Advanced Neutron Sources, Tsukuba, Japan, 89–101 (1980).
41. Ikeda, ICANS-XV Proceedings of the 15th Meeting of the International Collaboration on
Advanced Neutron Sources, Tsukuba, Japan, 93–96 (2000).
42. Shoji Nagamiya, J. Neutron Res. 13:1, 7–10 (2005).
43. F. Mezei, Proceedings 12th meeting of International Collaboration on Advanced Neutron
Sources ICANS-XII, Abingdon, Oxfordshire, UK, I-377–384 (1993).
44. Takashi Ino, Motoki Ooi, Yoshiaki Kiyanagi, Yoshimi Kasugai, Fujio Maekawa,
Hiroshi Takada, Guenter Muhrer, Eric J. Pitcher, and Gary J. Russell, Nucl. Instrum.
Meth. A 525, 496–510 (2004).
45. Tamura et al., JAERI-Report 2003–008 (2003).
46. F.R. Mildner and R.N. Sinclair, J. Nucl. Energy 6, 225 (1979).
47. Ikeda and J. Carpenter, Nucl. Instrum. Meth. Phys. Res. A239 536–544 (1985).
48. Watanabe et al., Proceedings of the 10th meeting of International Collaboration on
Advanced Neutron Sources ICANS-X, Los Alamos, USA, pp.787–797 (1988).
49. Leuschner, D.V. Baxter, J.M. Cameron, V. Derenchuk, C. Lavelle, A. Lone, H. Nann,
T. Rinckel, and W.M. Snow, J. Res. Natl. Instrum. Stand. Technol. 110, 153–155 (2005).
Chapter 3
Neutron Optics
K. Andersen
3.1 Introduction
K. Andersen (*)
Institut Laue-Langevin, 6 rue Jules Horowitz, B.P. 156, 38042 Grenoble Cedex 9,
France
e-mail: [email protected]
Neutron sources are typically 10–20 cm in size in all directions. They consist of
an efficient moderator material (usually containing hydrogen or deuterium) at
a well-defined temperature to produce a quasi-Maxwellian wavelength spectrum.
Their performance is characterised by their brightness (or brilliance) d2 =d dl
(i.e., flux per unit solid angle and per unit wavelength l). It is usually given in
units of neutrons/sec/cm2/sterad/Å. Brightness curves for the ILL vertical cold
source are shown in Fig. 3.1. For a pulsed source, the brightness can be given as
the peak brightness (i.e., the instantaneous brightness at its maximum in time) or
integrated over the full time spectrum of the moderator. These curves are also
shown in Fig. 3.1 for the liquid H2 moderator on the first target station at ISIS.
The brightness is a property of the source and does not change with distance.
The neutron flux at the sample is obtained by integrating over the source solid
angle seen by the sample and over the desired wavelength range. It then needs to
be scaled down by any imperfections in the transmission optics such as guide or
monochromator reflectivities or chopper transmissions.
The efficiency of any neutron optical device is limited by Liouville’s theorem,
also known as the ‘no free lunch’ theorem, which states that phase space density
in a conservative system is conserved. What this means is that any compression
in one dimension, such as focusing the source onto a spot smaller than the
source, is accompanied by an equivalent defocusing in another dimension, in
this case the divergence. This can be verified by looking at the standard lens
equation illustrated in Fig. 3.2: w1 =L1 ¼ w2 =L2 (i.e., the size of the focal spot
scales with the distance to the lens).
1012
Brightness (n/cm2/s/A/ster)
1010
108
source L1 L2 sample
It can easily be verified that the reduction in the size of the focal spot is
accompanied by an increase in the divergence hitting the focal spot, in accor-
dance with Liouville’s theorem. Liouville’s theorem should really be treated as a
best-case scenario. No optics system is fully conservative: some neutrons are
always lost from absorption, scattering, or other imperfections.
where L is the guide length, w is the guide width, and Rc is the radius of
curvature. Curving the guide also preferentially reduces the transmission of
short-wavelength neutrons. The characteristic wavelength l* (in Å) of a curved
guide is given by
575 pffiffiffiffiffiffiffiffiffiffiffiffiffiffi
l ¼ 2w=Rc ; (3:4)
m
Fig. 3.3 A supermirror guide on the ISIS second target station. The guide is viewed end-on,
showing the multiple reflections from the side and top and bottom walls. Photo courtesy of the
Science & Technology Facilities Council (UK)
nl ¼ 2d sin : (3:5)
2 2
@l @l
l2 ¼ d þ ¼ ð2 sin dÞ2 þð2d cos Þ2
@d @ ; (3:6)
) l=l cot
where we have dropped the first term which is usually small. From here we can
see that the diffracted wavelength spread is proportional to cot, which
becomes very large for small take-off angles. This means that when the crystal
is used to obtain wavelengths which are significantly shorter than the d-spacing,
the wavelength resolution becomes quite poor, which can be useful if you wish
to maximise flux at the sample at the expense of resolution. At the other end of
Table 3.1 Typical performance of some commonly used monochromator crystals and their
reflections. Heusler is the Cu2MnAl alloy, which provides a polarised beam. The peak
reflectivity is given for a crystal in reflection geometry
Peak reflectivity Peak reflectivity
Material Reflection d-spacing at l ¼ 2 Å at l ¼ 4 Å Mosaic
Silicon 111 3.135 Å 80% 90% bent
perfect
Graphite 002 3.355 Å 70% 80% 0.58
Germanium 111 3.266 Å 60% 80% 0.28
Copper 111 2.087 Å 60% 80% 0.38
Heusler 111 3.437 Å 25% 35% 0.38
38 K. Andersen
the range, when the Bragg angle approaches 908, the cot term becomes small
and very high resolution measurements are possible.
Monochromators usually comprise several crystals assembled together to
form a continuous surface, which can focus on to the sample. Large focusing
monochromators can give very large flux gains at the sample by accepting large
amounts of solid angle from the source. Figure 3.4 shows two large focusing
monochromators built for the IN8 triple-axis spectrometer at the ILL [4].
Vertical focusing contributes little to instrumental resolution as the vertical
tilt does not affect the Bragg angle to a first approximation. Horizontal focusing
of a parallel beam invariably entails varying the Bragg angle across the surface
of the monochromator and thus tends to degrade resolution to some extent. The
resolution degradation can be strongly reduced by arranging the instrument
optics in order to obtain focusing in Q or energy, as illustrated in Fig. 3.5.
Fig. 3.4 Graphite (left) and copper crystals on the IN8 monochromator. Both faces consist of
99 mosaic crystals and have variable horizontal and vertical curvature
3 Neutron Optics 39
θA θA
A A
θB θB
B B
Fig. 3.5 Q-focusing and time-focusing. The left-hand frame shows Q-focusing as employed on
powder diffractometers on continuous sources. The right-hand frame shows time-focusing as
employed on time-of-flight spectrometers at continuous sources
In both cases, the monochromator focuses on the sample in space and the
wavelength selected by the monochromator varies with position on the mono-
chromator. As a result, the incident angle on the sample is strongly correlated
with the wavelength. In the case of Q-focusing, the angular difference between
the short- and long-wavelength neutrons coming from the monochromator is
cancelled out by the Bragg scattering at the sample, and the result is a
scattered beam which is as well collimated as the beam incident on the
monochromator. Q-focusing works only when the direction of scattering at
the sample is the opposite of that at the monochromator, and is most efficient
when the d-spacing measured at the sample is similar to that of the
monochromator.
In the case of time-focusing, a Fermi chopper (a rotating collimator) pulses
the beam before the sample. It spins in the direction which lets through
the slower neutrons (B) before the faster ones (A), so that if they transfer
the same amount of energy at the sample, they arrive at the detector at the
same time.
When Q- or time-focusing are properly optimised they can provide the same
energy resolution as what would be obtained with a flat monochromator, but
with a large gain in flux at the sample.
When using mosaic crystals, the reflected beam is somewhat smeared by
the misorientation of the crystallites of the monochromator crystals. The
left-hand panel of Fig. 3.4 illustrates Q-focusing starting from a parallel
beam, which can be considered as a point-source at an infinite distance. In
general, a focusing monochromator can be optimised for a point-source at
any distance by adjusting the curvature. The finite size of the real source
then needs to be optimised to match the mosaic spread of the crystals and
the sample size. Neutrons originating outside this optimum source size will
not hit the sample and therefore only contribute to background. Hence the
concept of the ‘virtual source’ in which a diaphragm is used to reduce the
beam size at the source, giving significant gains in signal-to-background
ratios.
40 K. Andersen
Since the refractive index is smaller than that of vacuum, a focusing lens must
have a concave shape, unlike for the case of light.
A magnetic field gradient can also deviate the neutron trajectory:
@2r ~
¼ rB ; (3:8)
@t2 m
where is the magnetic moment of the neutron and m is its mass. The plus/
minus sign refers to the two spin states of the neutron: antiparallel or parallel to
the magnetic field, respectively.
Unfortunately for neutron optics purposes, the amount of focusing which
can be achieved by refraction and magnetic-field gradients is usually quite
small. Currently, used lens devices are restricted to small beam sizes and tend
to suffer from long focal lengths (of the order of tens of metres) and chromatic
aberrations, though work is underway to address these problems both for
refractive [5] and for magnetic [6] lenses.
Considerable technical advances have been made in polarisation optics over the
last 10 years or so. For this reason, I go into a little more detail here than in the
other sections. The reader is referred to [7] for a good overview of the technique.
Each individual neutron in a neutron beam has a spin of ½. When placed in a
magnetic field, this defines a quantisation axis with respect to which there are
only two possible orientations: parallel and antiparallel or ‘up’ and ‘down’. The
beam polarisation can be expressed as a scalar P which is usually defined to vary
between 0 and 1:
N" N#
P¼ ; (3:9)
N" þ N#
where there are N" (N#) neutrons with spin up (down). The beam polarisation is
often expressed in terms of the flipping ratio F given by
3 Neutron Optics 41
N" 1 þ P F1
F¼ ¼ )P¼ : (3:10)
N# 1 P Fþ1
1
Tn ðlÞ ¼ T" þ T# ¼ expðOðlÞÞ coshðOðlÞPHe Þ ; (3:13)
2
P, T, P2 T
an unpolarised beam are
shown, as is the standard
figure of merit P2T, which 0.4
display a maximum at the
optimum opacity
0.2
0
0 10 20 30 40 50 60
Opacity (bar.cm.A)
Fig. 3.7 Examples of glass 3He polariser cells with silicon entrance and exit windows. Photo-
graph courtesy of David Jullien (ILL)
3 Neutron Optics 45
been developed to maximise the relaxation time due to field gradients. Their
geometries are optimised for the particular instrument and are usually based on
solenoids, Helmholtz coils, or magnetised mu-metal. Screening from external
fields can be provided by steel, mu-metal, or superconducting Meissner screens.
References
1. ILL Neutron Data Booklet, Albert-José Dianoux and Gerry Lander (ed.), OCP Science
(July 2003).
2. V.F. Sears, Neutron Optics: An Introduction to the Theory of Neutron Optical Phenomena
and Their Applications, Oxford University Press, (1989).
3. V.F. Sears, Acta Cryst. A 53, 35 (1997).
4. A. Hiess et al., Physica B 385–386, 1077 (2006).
5. S.-M. Choi et al., J. Appl. Cryst. 33, 793 (2000).
6. H.M. Shimizu, Physica B 385–386, 989 (2006).
7. W.G. Williams, Polarized Neutrons, Oxford University Press, 1988.
8. A. Freund et al., Physica B 120, 86 (1983).
9. T.R. Gentile et al., J. Res. N.I.S.T. 110, 299 (2005).
10. A.K. Petoukhov et al., Physica B 385–386, 1146 (2006).
Chapter 4
Neutron Detectors for Imaging
L. Crow
Abstract This chapter examines neutron detection methods for imaging. The
first section reviews neutron capture converters, which form the basis for
thermal neutron detection, and the following sections examine various detector
systems with an emphasis on imaging applications and the choices available for
imaging at time-of-flight facilities.
L. Crow (*)
Neutron Facilities Development Division, Oak Ridge National Laboratory,
Oak Ridge, TN 37831, USA
e-mail: [email protected]
Table 4.1 Thermal neutron cross sections and charged particle production for some materials
used as neutron converters [4–6]
Charged
Neutron absorption cross particles and Gamma-ray
Isotope Reaction section (barns) (at 2,200 m/s ) energies (keV) production?
3 3
He He(n,p)3H 5333 p: 573, 3H: 191 No
6 6
Li Li(n, )3H 940 3
H: 2727, : 2055 No
10 10
B B(n, )7Li 3835 : 1472, 7Li: 480 Yes
nat nat
Gd Gd(n, ) 49700 Conversion Yes
electron
29–191
157 157
Gd Gd(n, ) 158Gd 259000 Conversion Yes
electron
29–182
235 235
U U(n,f) 681 Various fission Yes
products
Both the nuclear properties and the material characteristics affect the choice
of converter for a particular application. Helium-3 and 6Li have relatively large
cross sections, but the most important feature is that these capture reactions
release all of their energy in charged particles [6]. This makes these converters
the best choices for low-noise detectors which need to operate at high rates in
accessible areas. Helium-3 gas proportional counters have modest position
resolution, but very low gamma-ray sensitivity, and are the most widely used
detectors for neutron scattering. Lithium-6 is a very efficient scintillation con-
verter and appears in the most widely used thermal neutron scintillators, 6Li
glass and 6LiF:ZnS [7]. Boron-10 has a larger cross section than 6Li, and unlike
3
He can be in solid materials with very high atom density. The 10B capture
reaction has the disadvantage of producing a gamma-ray and heavier, shorter-
ranged charged particle products ( and 7Li). The compound 10BF3 has long
been used in proportional counters, and 10B is used in surface conversion gas
detectors. Boron-10 is also used in semiconductor detectors, microchannel plate
detectors, and a few scintillators.
Gadolinium has a number of disadvantages as a converter but is of interest in
imaging because of the enormous resonance at thermal neutron energies of
157
Gd (155Gd also has a very large cross section). The 157Gd capture results in a
spectrum of high energy gamma rays, which are difficult to count or shield,
along with conversion electrons. Gadolinium-based detectors rely on the inter-
nal conversion electrons to transfer the energy to a phosphor, gas, or lattice.
The cross section does not have the inverse velocity dependence of most con-
verters. This feature makes comparison of energy-dependent neutron response
difficult, but with white-beam imaging, the preferential response to thermal
neutrons helps suppress fast and epithermal neutron background.
Neutron capture reactions where the primary products are gamma rays may be
used for applications such as prompt or decay gamma activation analysis. The
4 Neutron Detectors for Imaging 49
characteristic gamma rays identify the isotopes which have absorbed the neutrons.
This method may be used for chemical analysis or for neutron integrated flux
measurements. Isotopes with higher energy absorption resonances have applica-
tions in energy-specific neutron detection [8]. A significant effort for chemical
spectroscopy and imaging is discussed in Chapter 13 of this volume. Isotopes
which capture neutrons and produce short-lived radioactive products maybe used
in indirect imaging or autoradiography, using the decay beta particles.
(a)
(b)
Fig. 4.1 The figure illustrates the dominant processes in direct and indirect screen capture for
two common screen choices. (a) A direct Gadolinium screen and X-ray film neutron imaging
process is shown schematically. The neutrons are captured by the Gd nuclei [157Gd (n, )
158
Gd]. The high energy prompt gamma rays escape, but some of the internal conversion
electrons deposit energy in the film and contribute to the photographic image. (b) In an
indirect process, a dysprosium screen captures neutrons during irradiation [164Dy (n, )
165
Dy]. The screen is removed from the neutron beam and placed in contact with film. An
image is produced by - particles from dysprosium decay [165Dy ( -, 165Ho), t1/2 ¼ 2.33 h]
usually a few tens of microns, is determined by the grain sizes in the converter
screen and the film, and the efficiency is dependent on the gadolinium conver-
sion electron capture in the film. Newer methods are more efficient, but the
position resolution of these screens remains competitive [11]. This detector
system requires a mechanism to open and close the incident beam to define
the exposure time and allow removal and replacement of the film and converter
screen and a facility to develop the film. This works well at continuous sources
with relatively long exposure times but is difficult to adapt for fast sample
changes.
4 Neutron Detectors for Imaging 51
Fig. 4.2 The figure shows a scale drawing and a perspective view of the microtomography
scintillator and CCD detector system at the Paul Scherrer Institut, Villigen, Switzerland. The
scintillator screen, a 458 mirror, the lens system, and the cooled CCD camera form the neutron
detector system. (This image was provided by Eberhard Lehmann of PSI by private
communication)
(a) (c)
(b) (d)
Fig. 4.3 (a) The figure shows an optical detector assembly of an Anger camera module for
the SNAP high pressure diffractometer at the Spallation Neutron Source (SNS) at Oak
Ridge National Laboratory. Nine 88 channel position-sensitive photomultiplier tubes are
assembled to form a single Anger module. (b) A SNAP module is shown with the bezel,
which contains a 15 cm square 6Li glass scintillator. This detector has a two-dimensional
position resolution of 1.3 mm full width at half maximum (FWHM). (c) A crossed
wavelength-shifting fiber assembly for a detector module to be mounted on the Vulcan
engineering diffractometer at SNS is shown. The wavelength-shifting fibers collect the blue
scintillation light from a 6LiF/ZnS(Ag) scintillator screen and guide the shifted green light
to a coded photomultiplier array. (d) A completed Vulcan module operates for initial
calibration. This 80 cm 40 cm detector module has 5 mm wide by 37.5 mm tall pixels,
with resolution matched to the diffractometer
Anger camera, achieve resolution approaching 1 mm [47], but it is not likely that
these technologies can achieve the submillimeter resolution needed to compete in
radiography or tomography.
Storage phosphors and image plate readers: Storage phosphors are function-
ally similar to scintillators, except that instead of a prompt light emission, energy
is stored in the lattice. The phosphor is read by photostimulation (typically by
scanning with a laser), inducing light emissions. Neutron image plates using a
gadolinium converter with a storage phosphor (Fig. 4.4), developed in the 1990s
4 Neutron Detectors for Imaging 55
(a)
(b)
Fig. 4.4. (a) A 157Gd-based image plate accumulates energy from internal conversion
electrons produced in neutron capture in its storage phosphor. The image plate is
transferred to a reading system. A laser scans the image plate and photostimulates the
release of blue photons from the storage phosphor in the image plate. (b) A 25 cm square
neutron image plate is shown
[48, 49], have become commercially available [50] in the last decade and are now
in regular use for diffraction [51] and radiography [52] at steady-state sources.
Image plates are much more efficient than any of the photographic film
56 L. Crow
In a gas detector, ions produced within the gas accelerate due to a charge bias
applied within the gas volume. If the acceleration is large enough, the signal is
amplified as the ions collide with gas atoms and produce additional ionization.
In this way, an ionization event, such as a neutron capture, will produce an
electron ‘‘charge cloud,’’ which drifts to the biased anode. Depending on the
bias and the gas, the charge may be equivalent to the original energy deposition
(ionization mode), amplified proportional to the original energy (proportional
mode), or a nonproportional avalanche (Geiger-Müller mode). Neutron detec-
tors with gas converters such as 10BF3 [54] and 3He [55, 56] usually operate in
the proportional range to take advantage of the excellent discrimination
allowed by the energetic charged particles.
Helium-3 proportional counters are the most widely used detector systems
for neutron scattering, where resolution can be matched to the requirements
of the instruments. The development of position-sensitive proportional coun-
ters has led to imaging detectors for neutron-scattering applications such as
small-angle neutron scattering (SANS), neutron reflectometry, and diffrac-
tion. Readout systems for area detectors include delay line [57], charge divi-
sion [58], and the now more widely used multiwire centroid coincidence
method [59–61]. Linear position-sensitive detectors using charge division
[62] may be arranged to form area detectors for SANS [63, 64] and a wide
variety of neutron-scattering applications [65, 66] (Fig. 4.5). The counting
readout systems are readily adaptable to time-of-flight work. Gas converter
detectors cannot compete easily for submillimeter position resolution because
of the thick gas volume and the long capture product ranges (typically on the
order of a millimeter or longer). The best wire chamber detector resolution to
date is about 400 mm [67]. Gas detector development continues with efforts to
overcome the present count rate limitations of wire chambers, typically
between 40,000 and 100,000 counts per second per chamber due to drift
time. A new high-performance multiwire gas chamber detector known as
MILAND1 [68] is expected to achieve 1 mm resolution at a count rate of up
1
MILAND (for Millimeter Resolution Large Area Neutron Detector) is a joint research
activity of the European Union supported under the Fifth Framework Program (FP5).
4 Neutron Detectors for Imaging 57
(a) (b)
(c)
Fig. 4.5 (a) An array of 8 mm diameter 1 m long linear position-sensitive 3He proportional
tubes forms an area detector for small angle neutron scattering, with resolution of about
8 mm. The modules shown here will be installed on EQSANS at SNS. (b) A 20 cm 17 cm
multiwire proportional counter manufactured by Brookhaven National Laboratory operates
in the SNS liquid reflectometer. (c) This multiwire detector, with 6.5 bar of 3He, has a
resolution of 1.3 mm FWHM, as shown in this calibration mask image
to 1,000,000 counts per second over the entire detector area using readout with
separate amplifiers on each wire.
Strategies for higher performance gas detectors use micropattern elec-
trode structures to effectively reduce drift distances and times and in the
process divide the readout of a detector chamber into smaller segments.
58 L. Crow
These detectors may use gas converters or solid converters within the
counting gas volume. In microstrip gas counters (MSGC) [69–72], the
anode and cathode strips are tens of microns apart on a glass substrate,
reducing the drift times to a few hundred nanoseconds. Gas scintillation
proportional counters are variants of the MSGC that use the light from the
charge amplification to localize the neutron events [73–75]. Gas electron
multipliers [76–78] use biased stacks of perforated kapton foils to achieve
high gas amplification in a very small distance. Another in this same
subclass of gas detectors, the Micromegas (for MicroMEsh Gaseous
Structure), has a converter layer and a very small amplification gap of
about 100 mm [79–80] between two parallel glass or epoxy plates.
(a) (b)
Fig. 4.6 (a) A silicon wafer detector element has been perforated with grooves at 100 mm
spacing. Each groove has been filled with 6LiF. The pixel size is 100 mm by 4 cm (the detector
height). (b) The detector is mounted at the center of the circuit board; the application-
specific integrated circuits to operate the detector are shown above and below the detector
window
Fig. 4.7 A 40 mm diameter microchannel plate detector built by NOVA Scientific, similar to
the detector described in [91], is shown. The microchannel plate assembly is visible through the
sapphire neutron window. The delay line readout enables counting operation at high rates
with 40 mm resolution
4.7 Conclusion
The current state of neutron imaging detectors presents a range of options for
continuous sources and some promising paths for achieving the timing capabil-
ities needed for pulsed sources. The neutron storage phosphor image plates
yield excellent resolution at modest operating costs. The CCD camera systems
are fast and flexible and are currently the primary detection technique for most
major imaging user facilities. The potential for improvement in scintillators or
cameras puts pulsed source operation within reach, but gating and camera
readout at high rates are big challenges. The microchannel plate detectors
operate at very high count rates, are operable as counting detectors now, and
may have the shortest path to operation as timing detectors.
References
1. J.A. Harvey and N.W. Hill, Nucl. Instrum. Meth. 162, 507 (1979).
2. G.F. Knoll, Radiation Detection and Measurement, 3rd ed. Wiley, New York (2000).
4 Neutron Detectors for Imaging 61
36. Helena Pleinert, Eberhard Lehmann, and Sonja Körner, Nucl. Instrum. Meth. A399, 382
(1997).
37. D.S. Hussey, D.L. Jacobson, M. Arif, P.R. Huffman, R.E. Williams, and J. C. Cook,
Nucl. Instrum. Meth. A542, 9 (2005).
38. B. Schillinger, E. Calzada, F. Grunauer, and E. Steichele, Appl. Radiat. Isot. 61, 253 (2004).
39. M. Dierick, J. Vlassenbroeck, B. Masschaele, V. Cnudde, L. Van Hoorebeke, and A.
Hillenbach, Nucl. Instrum. Meth. A 542, 296 (2005).
40. B. Schillinger, H. Abele, J. Brunner, G. Frei, R. Gähler, A. Gildemeister, and A.
Hillenbach, E. Lehmann, P. Vontobel, Nucl. Instrum. Meth. A 542, 142 (2005).
41. Masahito Matsubayashi and Kazuhiko Soyama, Nucl. Instrum. Meth. A 529, 384 (2004).
42. E. Lehmann and P. Vontobel, Appl. Radiat. Isot. 61, 567 (2004).
43. Mirko Estermann and Joe Dubois, IEEE Trans. Nucl. Sci., 52, 356 (2005).
44. Mirko Estermann, Gabriel Frei, Eberhard Lehmann, and Peter Vontobel, Nucl. Instrum.
Meth. A 542, 253 (2005).
45. N.J. Rhodes, A.G. Wardle, A.J. Boram, and M.W. Johnson, Nucl. Instrum. Meth. A 392,
315 (1997).
46. N.J. Rhodes, E.M. Schooneveld, and R.S. Eccleston, Nucl. Instrum. Meth. A 529, 243 (2004).
47. M. Katagiri, T. Nakamura, M. Ebine, A. Birumachi, S. Sato, E.M. Schooneveld, and N.J.
Rhodes, Nucl. Instrum. Meth. A 573, 149, (2007).
48. Nobuo Niimura,Yuuko Karasawa, Inchiro Tanaka, Junji Miyahara, Kenji Takahashi,
Hiroki Saito, Satoshi Koizumi, and Masanori Hidaka, Nucl. Instrum. Meth. A 349, 521
(1994).
49. Jens Hofmann and Christian Rausch, Nucl. Instrum. Meth. A 355, 494 (1995).
50. Fuji Film, I & I-Imaging Inform. (1995).
51. D.A.A. Myles, C. Bon, P. Langan, F. Cipriani, J.C. Castagna, M.S. Lehmann, and C.
Wilkinson, Physica B 241-243, 1122 (1998).
52. Shigenori Fujine, Kenji Yoneda, Masahiro Kamata, and Masahiro Etoh, Nucl. Instrum.
Meth. A 424, 200 (1999).
53. B. Schillinger, J. Baumann, H. Gebele, R. Schaetzing, H. Schaller, and M. Schuster, App.l
Radiat. Isot. 61, 451 (2004).
54. S.A. Korff and W.E. Danforth, Phys. Rev. 55, 980 (1939).
55. R. Batchelor, R. Aves and T.H.R. Skyrme, Rev. Sci. Instrum. 26, 1037 (1955).
56. W.R. Mills, R.L. Caldwell, and I.L. Morgan, Rev. Sci. Instrum. 33, 866 (1962).
57. C.J. Borkowski and M.K. Kopp, Rev. Sci. Instrum. 46, 951 (1975).
58. V. Radeka, IEEE Trans. Nucl. Sci. 21, 51 (1974).
59. G. Charpak and F. Sauli, Nucl. Instrum. Meth. 113, 381 (1973).
60. Veljko Radeka and Robert A. Boie, Nucl. Instrum. Meth. 178, 543 (1980).
61. R. Kampmann, M. Marmotti, M. Haese-Seiller, and V. Kudryashov, Nucl. Instrum.
Meth. A 529, 342 (2004).
62. R. Berliner, J.S. King and D.F.R. Mildner, Nucl. Instrum. Meth. 152, 431 (1978).
63. K. Abend, W. Schmatz, J. Schelten and K.D. Müller, Nucl. Instrum. Meth. 83, 111 (1970).
64. R. Berliner, D.F.R. Mildner, O.A. Pringle and J.S. King, Nucl. Instrum. Meth. 185, 481
(1981).
65. R.K. Crawford, J.R. Haumann, J.E. Epperson, P. Thiyagarajan, A.J. Schultz, G.P.
Felcher, D.G. Montague and R.J. Dejus, Nucl. Instrum. Meth. A 299, 17 (1990).
66. R.G. Cooper, Nucl. Instrum. Meth. A 529, 394 (2004).
67. V. Radeka, N.A. Schaknowski, G.C. Smith, and B. Yu, Nucl. Instrum. Meth. A 419, 642
(1998).
68. B. Guerard, http://jra2.neutron-eu.net/jra2/ (2008).
69. A. Oed, Nucl. Instrum. Meth. A 263 351(1988).
70. J.F. Clergeau, P. Convert, D. Feltin, H.E. Fischer, B. Guerard T. Hansen, G. Manzin,
A. Oed, and P. Palleau, Nucl. Instrum. Meth. A 471, 60 (2001).
4 Neutron Detectors for Imaging 63
71. Hiroyuki Takahashi, Prasit Siritiprussamee, Kaoru Fujita, Hisako Niko, Masaharu
Nakazawa, Shunji Kishimoto, Takashi Ino, Michihiro Furusaka, and Hidenori Toyo-
kawa, Nucl. Instrum. Meth. A 573, 175 (2007).
72. B. Gebauer, S.S. Alimov, A.Yu. Klimov, F.V. Levchanovski, E.I. Litvinenko, A.S.
Nikiforov, V.I. Prikhodko, G. Richter, V.V. Rogov, Ch. Schulz, V.I. Shashkin, M.
Wilhelm, and Th. Wilpert, Nucl. Instrum. Meth. A 529, 358 (2004).
73. Masayo Suzuki, Tan Takahashi, Hidekazu Kumagai, Masayasu Ishihara, and Hisao
Kobayashi, Nucl. Instrum. Meth. A 333, 484 (1993).
74. G. Manzin, B. Guerard, F.A.F. Fraga, and L.M.S. Margato, Nucl. Instrum. Meth. A 535,
102 (2004).
75. J.C. Buffet, J.F. Clergeau, R.G. Cooper, J. Darpentigny, A. De Laulany, C. Fermon, S.
Fetal, F. Fraga, B. Guerard, R. Kampmann, A. Kastenmueller, G.J. Mc Intyre, G. Manzin,
F. Meilleur, F. Millier, N. Rhodes, L. Rosta, E. Schooneveld, G.C. Smith, H. Takahashi, P.
Van Esch, T.L. Van Vuure, and K. Zeitelhack, Nucl. Instrum. Meth. A 554, 392 (2005).
76. M. Klein, H. Abele, D. Fiolka, and Chr. J. Schmidt, AIP Conf. Proc. 596, 193 (2001).
77. D.D. DiJulio, A.I. Hawari, and R. Berliner, Nucl. Instrum. Meth. A 579, 71 (2007).
78. F.A.F. Fraga, L.M.S. Margato, S.T.G. Fetal, M.M.F.R. Fraga, R. Ferreira Marques, A.
J.P.L Policarpo, B. Guerard, A. Oed, G. Manzini, and T. van Vuure, Nucl. Instrum. Meth.
A 478, 357 (2002).
79. S. Andriamonje, D. Cano-Ott, A. Delbart, J. Derre, S. Diez, I. Giomataris, E.M.
Gonzalez-Romero, F. Jeanneau, D. Karamanis, A. Lepretre, I. Papadopoulos, P. Pavlo-
poulos, and D. Villamarin, Nucl. Instrum. Meth. A 481, 120 (2002).
80. Charles L. Britton, Jr, William L. Bryan, Alan L. Wintenberg, Robert J. Warmack,
Timothy E. McKnight, Shane S. Frank, Ronald G. Cooper, Nancy J. Dudney, Gabriel
M. Veith, and Andrew C. Stephan, IEEE Trans. Nucl. Sci. 51, 1016 (2004).
81. Y. Kumashiro, and Y. Okada, Appl. Phys. Lett. 47, 64 (1985).
82. D.S. McGregor, T.C. Unruh, and W.J. McNeil, Nucl. Instrum. Meth. A 591, 530 (2008).
83. D.S. McGregor, M.D. Hammig, Y.-H. Yang, H.K. Gersch, and R.T. Klann, Nucl.
Instrum. Meth. A 500, 272 (2003).
84. J. Uher, T. Holy, J. Jakubek, E. Lehmann, S. Pospisil, and J. Vacik, Nucl. Instrum. Meth.
A 542, 283 (2005).
85. W.J. McNeil, S.L. Bellinger, B.J. Blalock, C.L. Britton, Jr., J.L. Britton, S.C. Bunch, S.A.
Cowley, C.M. Henderson, T.J. Sobering, R.D. Taylor, and D.S. McGregor, Nuclear
Science Symposium Conference Record, 2007. NSS ’07. IEEE 3, 2340 (2007).
86. Takekazu Ishida, Masatoshi Nishikawa, Shigehito Miki, Hisashi Shimakage, Zhen
Wang, Kazuo Satoh, Tsutomu Yotsuya, Masahiko Machida, and Masaru Kato, Physica
C 460–462, 618 (2007).
87. G.W. Fraser and J.F. Pearson, Nucl. Instrum. Meth. A 293, 569 (1990).
88. R.M. Ambrosi, G.W. Fraser, R.A. Street, J.I.W. Watterson, R.C. Lanza, J. Dowson, A.
F. Abbey, B. Feller, G. Downing, P. White, and T. Stevenson, Nucl. Instrum. Meth. A 542,
271 (2005).
89. Oswald H.W. Siegmund, John V. Vallerga, Adrian Martin, Bruce Feller, Muhammad
Arif, Daniel S. Hussey, and David L. Jacobson, Nucl. Instrum. Meth. A 579, 188 (2007).
90. A.S. Tremsin, J.V. Vallerga, J.B. McPhate, O.H.W. Siegmund, W.B. Feller, L. Crow, and
R.G. Cooper, Nucl. Instrum. Meth. A 592, 374 (2008).
Section B
Chapter 5
Neutron Radiography
5.1 Introduction
1
British Experimental Pile ‘0,’ the UK’s second reactor.
5 Neutron Radiography 69
5.3.1 Sources
incident particle and target material that can be used, a list of which may be found in
von der Hardt [14]. Accelerators produce thermal neutrons in the range from 107 to
1010 n/cm2/s and offer the benefit of intermittent operation and portability. Sealed-
neutron generator tubes are easy to use but generally produce low fluxes compared
with high current accelerators, which are moderately complex to operate and have a
target output that may deteriorate with use. Large spallation sources can achieve fluxes
of more than 1014 n/cm2/s, but such sources are very complex and not portable.
A radioisotope-based neutron source makes use of radioactive isotope decay
to generate its neutron flux in much the same way as accelerators: by allowing
the gamma rays or alpha particles emitted by a radioactive isotope to bombard
a neutron emitting target. A number of (,n) and (,n) radioisotopic neutron
sources exist, tables of which may be found in von der Hardt [14] and Berger [8].
Radioisotope-based neutron sources have low thermal neutron fluxes, within
the range of 105–109 n/cm2/s, and also suffer from decreasing output, from both
deterioration of the target and decay of the source. The inability to turn off the
radiation is often considered a disadvantage. The major advantages these
systems possess are their simple operation and ease of portability.
Nuclear reactor-based neutron sources provide intense neutron beams and
thus high-quality neutron radiographs. Thermal neutron fluxes obtainable in
such facilities range from 1010 to 1015n/cm2/s or even higher. It should be noted
that large spallation neutron facilities may produce beams with intensities equal
to or higher than those produced by nuclear reactors. However, reactors
typically provide a lower cost per neutron than accelerators. Their major dis-
advantages are the high costs associated with construction; licensing and regula-
tory requirements; complexity to operate; and lack of portability.
5.3.2 Moderation
The neutrons born in each of the sources discussed above possess high energies
with a continuous spectrum of energies peaking from 0.85 MeV from fission (in
reactors) up to 14 MeV (in accelerators). Conventional neutron radiography,
however, requires neutrons in the thermal/epithermal energy range of
0.025 eV–10 keV. Thus, some form of moderator with low neutron absorption
cross section (to maximize flux) and high scattering cross section (to maximize
energy loss) is required to slow down the neutrons to this energy range. The
often-used moderator materials of water, heavy water, graphite, beryllium, and
polyethylene meet these criteria. In this, the nuclear reactor has an inherent
advantage: the moderation of its core already produces a low-energy spectrum
resulting in fewer neutrons lost in the moderation process.
5.3.3 Collimation
Once low-energy neutrons are produced, they must be formed into a usable
beam. Neutrons are emitted and then scattered randomly in the moderator and,
5 Neutron Radiography 71
because of their neutral charge, they cannot be focused like electrons. Those
neutrons traveling in the desired solid angle can, however, be selected by the
introduction of a tube into or adjacent to the moderator. This has the effect of
allowing neutrons to stream down the tube axis toward the object being radio-
graphed. The walls of the collimator tube are lined with a neutron opaque
material having a high absorption cross section (such as boron, gadolinium,
and cadmium), which prevents stray neutrons from entering and also reduces
low-angle scattering within the collimator. The most common collimator design
is a divergent collimator (Fig. 5.1) with a small entrance aperture and a larger
exit. This maximizes the neutron flux and permits a larger field at the imaging
plane. The angular spread of the emerging beam is dependent upon the ratio of
the collimator tube length (L) to its aperture diameter (D), referred to as the L/D
ratio. A higher L/D results in a narrower beam spread at the expense of a lower
neutron flux. This ratio is a characteristic parameter of each collimator. Extensive
discussions regarding collimator design may be found in von der Hart [14]
and Domanus [21].
5.3.4 Detectors
After a neutron’s birth in the source, its moderation to thermal energies, and its
escape along the collimator tube, it will encounter the object to be radiographed.
Any neutrons that successfully penetrate the object must then be detected to
produce the radiograph. In neutron radiography, a detector collectively refers to
both an intermediate medium, called a converter (which emits an alpha, beta,
gamma, or light when neutrons are absorbed) and the sensor used to detect this
emitted radiation, called the image recorder. The converter material is used
because it emits a much more readily detectable radiation.
When the image recorder is film, one possible converter material is a gado-
linium foil which emits an electron with every absorbed neutron. The converter
foil is placed in direct contact with the film’s emulsion and the emitted electrons
expose the emulsion, producing an image. A typical spatial resolution using a
single-coated fine-grain radiographic film, a vapor deposited gadolinium con-
verter, and a vacuum cassette is 10 mm.
Another possible converter is a scintillation screen, which will expose the
film’s emulsion with light. A scintillation screen is 30–100 times faster at
producing an image on radiographic film than a gadolinium foil. However,
72 A.K. Heller and J.S. Brenizer
due to light spread within the scintillator, the spatial resolution is reduced.
A typical spatial resolution for a scintillator and film is 100 mm. Both of these
converters continuously emit radiation for the duration of neutron exposure
and can therefore be used in low neutron flux environments with long integra-
tion exposures.
An electronic form of imaging where a scintillation converter is optically
coupled with an analog or digital camera image recorder can also be used. This
allows the rapid capture of successive neutron radiographic images that can
be viewed directly and stored on videocassette or digital media, preserving
dynamic information. This electronic form of producing radiographic images
is referred to as radioscopy.
Other types of analog and electronic detectors exist and vary depending
upon the imaging technique used. They are discussed in later sections of this
chapter.
To summarize, a neutron radiography system consists of a neutron source, a
moderator to thermalize the neutrons, an aperture and a collimator to organize
the neutrons into a beam, and a detector to visualize the image (Fig. 5.2).
Moderator
Image
Aperture Recorder
Object
Neutrons
Neutron Source
Converter
¼ o ei t : (5:1)
5 Neutron Radiography 73
De ¼ G ðlog EÞ ; (5:2)
where Io is the incident light (such as from a light box) and I is the transmitted
light through the film.
In nearly all forms of digital imaging, the resulting grey level value of any
pixel making up the image may be described by
G ¼ C þ Goffset ; (5:4)
where G is the numerical grey level value of the pixel within an image, C is the
electronic gain of the camera or imaging system (a constant), is the trans-
mitted neutron intensity and Goffset is the dark current, an additive offset due to
electronic noise.
These equations form the basis of all radiographic image analysis. With
them, one may manipulate images to isolate terms and perform quantitative
analyses or provide the basis for qualitative comparisons. Extensive descrip-
tions and applications of these equations may be found in [14, 20–22].
developed. Converter screens of this type are usually 25-mm thick and are either
gadolinium foils laminated to aluminum or vapor-deposited gadolinium on
aluminum for ease of handling. Vacuum cassettes are used to ensure good contact
between the converter and image recorder, which is vital in reducing image blur.
The resolution of this converter–imager recorder combination is 10–20 mm. This
resolution is quite impressive, considering it is obtainable over very large areas
using films up to 35.5 cm by 43 cm.
In a similar fashion, a light-emitting scintillation screen may be used as the
converter. In these screens, a neutron-absorbing material that yields a charged
particle is mixed with a phosphor material that produces light. An example is a
mixture of lithium-6 and ZnS(Ag). Lithium-6 emits an alpha particle when
absorbing a neutron, and it is the kinetic energy of the alpha particle that causes
the ZnS(Ag) to emit light. This light then exposes the film emulsion. Scintilla-
tion screens are thicker than gadolinium foils and, because of the light spread
within them, the obtainable resolution (75–100 mm with the best screens) is less
than that of gadolinium foils.
The direct imaging method has a major disadvantage when it involves
nuclear applications. A radioactive object emitting gamma rays as it decays
or a neutron beam contaminated with gamma rays will directly produce an
image on the film. Since the gamma rays are from a different source than the
neutrons, the images will be different and the film will be blurred or what is
known as ‘‘gamma fogged.’’ Another disadvantage of the method is the time
associated with the film development, which inherently prohibits the investiga-
tion of dynamic processes.
There are, however, a wide variety of applications for which this method
is well suited. In the area of turbine manufacturing, the direct radiographic
method can easily perform a quality control check on the presence of residual
materials used in the blade’s production (Fig. 5.3). In fact, the technique has
been used for quality control inspections for moisture, corrosion, adhesive
defects, proper lubrication, and quality of seals in the aerospace and automotive
industries. It is also used in the health monitoring and maintenance of in-service
components such as aircraft flight control surfaces [23]. The direct method also
plays a major role in the field of research, where it has been used to investigate
two-phase flow behavior in heat pipes, water distribution in fuel cells, and water
permeability in concrete, to list a few examples.
The indirect radiographic method applies when the converter screen is not in
direct contact with the image recorder. This is also referred to as the transfer
technique because of the manner in which the image is produced. A metal foil
converter is placed in the beam independent of the image recorder. The foil
builds up a radioactivity through neutron absorption, producing what is
called an activation image. After being removed from the beam, the activated
foil is placed in contact with the image recorder and the decay radiation
emitted (low-energy electrons) ‘‘transfers’’ the activation image to the image
recorder. In this method, the image recorder is film, just as with the direct
radiographic method, but the converter screen used is indium, dysprosium, or
gold because of the need for the foil to rapidly activate for image transfer and
rapidly decay for reuse. The image quality produced by these techniques is the
same as that of the direct techniques and provides spatial resolutions of
10–20 mm.
The technique is slower than the direct method, but one major advantage is
that the converter foil used is insensitive to gamma radiation from the object
or beam, making it an ideal candidate for use in nuclear applications such as
the investigation of spent fuel rods. Gamma rays produced by radioactive
decay in these applications will not fog up the radiographic image. The fact
that this technique is slower than the direct method does not, however,
preclude it from being applied to the same fields as the direct radiographic
method. Indeed, the indirect radiographic method is equally applicable to the
same industries, research, and quality control as the direct radiographic
method.
radiation damage within the film is made visible by etching it in a hot base
solution, such as sodium hydroxide. A vacuum cassette is used to ensure tight
contact between the film and the converter screens.
The track-etch method is insensitive to gamma rays and also to visible light,
allowing the etching to be performed in daylight. However, the exposure time
needed for this technique is longer than the direct and indirect methods, and the
contrast is also weaker. A unique advantage is the ability to stop the etching
process at intermediate stages, which can then be continued after evaluating
the resultant neutron radiograph. In the end, several radiographs of varying
densities and contrasts can be had. Because of the gamma-ray insensitivity,
radiography of radioactive objects, such as spent nuclear fuel, is the primary
application for this method.
Film is not the only form of image recorder possible. There are techniques that use
electronic means, both analog and digital, for the image recorder. The production
of radiographic images via electronic image recorders, initially referred to as real-
time neutron radiography, is now known as radioscopy [24]. Electronic image
recorders depend upon the capture of light to record an image and, because they
have the ability to rapidly acquire these images, they are universally coupled with
a converter that has an equally rapid (if not faster) light response: a scintilla-
tion screen. Thus, these methods are frequently used to view and store dynamic
radioscopic images.
In electronic imaging systems, the image recorder can be a video camera if an
analog system, or a charge-coupled device (CCD) camera if a digital one. In the
case of an analog system, the camera is connected to a television and VCR for
the real-time viewing and recording of the dynamic radioscopic images. In a
digital system, the CCD camera is connected to a computer that displays and
stores the radioscopic images in a digital format. A combination of these systems
may also exist wherein the analog television camera can be connected to a
computer via an analog-to-digital converter card which will digitize the camera’s
analog signal and store it.
The camera, be it analog or digital, is coupled to the scintillation screen via a
light-tight box to prevent polluting the radioscopic image with outside light.
The intensity of light produced by the scintillation screen is linear with the
intensity of incident neutron flux, and there are a number of reactor facilities
where the neutron flux intensity is sufficiently large for the light produced by the
scintillation screen to be seen on a TV monitor. However, there are many
facilities that do not possess a neutron flux of such intensity, and in these
circumstances an image intensifier is placed within the light-tight box between
the scintillation screen and the camera. The image intensifier boosts the light
from the scintillation screen by about 104, still providing good image definition
5 Neutron Radiography 77
5.10 Standards
5.11 Conclusions
Neutron radiography and its related 2D neutron imaging techniques have
established themselves as invaluable nondestructive inspection methods and
quantitative measurement tools. They have been used in a wide variety of
applications ranging from inspection of aircraft engine turbine blades to study-
ing two-phase fluid flow in operating proton exchange membrane fuel cells.
Advances in digital cameras, CR using imaging plates, and the development of
more intense and robust neutron generator tubes will expand the use of neutron
imaging. These advances will also increase portability, making it possible to take
imaging systems to the field rather than bringing objects to a fixed imaging
5 Neutron Radiography 79
References
1. J. Chadwick, Possible Existence of a Neutron, Nature 129, 312 (1932).
2. H. Kallmann, Neutron Radiography Research 1, 254 (1948).
3. O.Z. Peter, Neutronen-Durchleuchtung, Naturforsch. 1, 557 (1946).
4. J. Thewlis and R.T.P. Derbyshire, Report AERE M/TN 37, U. K. Atomic Energy
Research Establishment, Harwell England (1956).
5. H. Berger, Proceedings Symposium on Physics and Nondestructive Testing, Argonne,
Illinois ANL-6346, Argonne National Laboratory, Argonne Illinois (1960).
6. H. Berger, Neutron Radiography: Methods, Capabilities and Applications, Elsevier Pub-
lishing Company, New York (1965).
7. H.V. Watts, Report ARF 1164-6, Armour Research Foundation, Chicago, Illinois (1960).
8. J.P. Barton, Introductory Remarks Neutron Radiography: Proceedings at the Second World
Conference on Neutron Radiography, Paris, France (June 16–20, 1986), J. Barton, J. Farny,
J. Person, H. Röttger, D. Reidel, eds. Pub. Co., Dordrecht, Holland, p. 13 (1987).
9. E.L. Criscuolo and D. Polansky, Proceedings Missiles and Rockets Symposium, Concorde,
California, 1961, U. S. Naval Ammunition Depot, Concord, California, p. 112 (1961).
10. A.W. Schultz and W.Z. Leavitt, Report WALTR 142.67,Watertown Arsenal Labora-
tories, Watertown Massachusetts (1961).
11. M.R. Hawkesworth (technical editor) Radiography with Neutrons, Conference held
10–12 September, 1973, at the University of Birmingham, London, British Nuclear
Energy Society (1975).
12. H. Berger (editor) Practical Applications Of Neutron Radiography And Gauging, ASTM
STP 586 (1976).
13. J.P. Barton and P. von der Hardt (editors) Neutron Radiography Proceedings of the First
World Conference, San Diego, California, U.S.A. , December 7–10, 1981. D. Reidel
Publishing Company, Dordrecht: Holland/Boston: U.S.A./London, England (1983).
14. P. von der Hardt and H. Röttger (editors), Neutron Radiography Handbook, D. Reidel
Publishing Company, Dordrecht: Holland/Boston: U.S.A./London, England (1981).
15. J.S. Brenizer and M.F. Sulcoski Real-Time Neutron Radiography at the University of
Virginia in Use and Development of Low and Medium Flux Research Reactors Harling,
O.K., Clark, L., and von der Hardt, P., eds., Karl Theiemig Graphische, W. Germany,
p. 958 (1984).
16. J.D. Jones, J.T. Lindsay, C.W. Kaufmann, A.T. Vulpeti and B. Peters, Real Time
Neutron Imaging Applied to Internal Combustion Engine Behavior SAE Technical
Paper Series No. 850560, International Congress and Exposition (1985).
17. M. Tamaki, K. Ohkubo, Y. Ikeda, and G. Matsumoto Analysis of Two Phase Flow in
Heat Pipe by Neutron Radiography Proceedings of the Second World Conference on
Neutron Radiography, Paris, France, June 16–20, 1986 Barton, G. Farny, J. Person, H.
Röttger, D. Reidel, eds. Pub. Co., Dordrecht, Holland, pp. 609–616 (1987).
80 A.K. Heller and J.S. Brenizer
18. J.S. Brenizer, B. Hosticka, R.W. Jenkins Jr and D.D. McRae An Advanced Video System
for Real-Time Neutron Radiography Proceedings of the Second World Conference on
Neutron Radiography, Paris, France, June 16–20, 1986, Barton, G. Farny, J. Person, H.
Röttger, D. Reidel, eds. Pub. Co., Dordrecht, Holland, pp. 571–578 (1987).
19. R. Polichar and D. Shreve Processing of Real-Time Images for Quantitative Neutron
Radiography Proceedings of the Second World Conference on Neutron Radiography, Paris,
France, June 16–20, 1986, J. Barton, G. Farny, J. Person, H. Röttger, D. Reidel, eds. Pub.
Co., Dordrecht, Holland, pp. 587–593 (1987).
20. T. Hibiki and T. Mishima Prediction of Measurement Error Due to Low Gray Scale
and Spatial Resolution of an Imaging System on Quantification of Neutron Radiogra-
phic Image Nuclear Instruments and Methods in Physics Research Section A, vol. 338,
pp. 204–211 (1997).
21. J.C. Domanus (editor) Practical Neutron Radiography, Kluwer Academic Publishers
Dordrecht: Holland/Boston: U.S.A./London: England (1992).
22. R.A. Quinn and C.C. Sigl (editors) (1980) Radiography in Modern Industry, 4th edition,
Eastman Kodak Company, Rochester NY.
23. T.R. Chalovich, L.G.I. Bennett, W.J. Lewis, and J. Brenizer, Jr., Development of
Neutron Radioscopy for the Inspection of CF188 Flight Control Surfaces, Applied
Radiation and Isotopes, vol. 61, pp. 693–700 (2004).
24. E07 on Nondestructive Testing ASTM E1316 Standard Terminology for Nondestructive
Examinations, Annual Book of ASTM Standards vol 03.03 (2007).
25. E07 on Nondestructive Testing E2007 Standard Guide for Computed Radiology (Photo-
stimulable Luminescence (PSL) Method) Annual Book of ASTM Standards, vol. 03.03
(2007).
26. M. Thoms, D. Myles and C. Wilkinson Neutron Detection with Imaging Plates Part I.
Image Storage and Readout, Nuclear Instruments and Methods in Physics Research A, vol.
424, pp. 26–33 (1999).
27. M. Thoms Neutron Detection with Imaging Plates Part II. Detector Characteristics
Nuclear Instruments and Methods in Physics Research A, vol. 424, pp. 34–39 (1999).
28. H. Kolbe, E Lehmann, W. Gunia and S. Körner Applications and Characteristics of
Imaging Plates as Detector in Neutron Radiography at SINQ Nuclear Instruments and
Methods in Physics Research A, vol. 424, pp. 40–47 (1999).
29. J.J. Haskins ASTM Activities in Neutron Radiography Practical Applications of Neutron
Radiography and Gauging, ASTM STP 586, H. Berger (editor) American Society for
Testing and Materials, Philadelphia, pp. 106–113 (1976).
30. J.J. Haskins Neutron Radiography Standards in the United States of America Neutron
Radiography Proceedings of the First World Conference, San Diego, California, U.S.A. ,
December 7–10, 1981, Barton JP and von der Hardt P (editors) D. Reidel Publishing
Company. Dordrecht: Holland/Boston: U.S.A./London: England, pp. 985–991 (1983).
31. R.L. Newacheck and R.R. Tsukimura Current Status of the ASTM E545 Image Quality
Indicator System Neutron Radiography (3) Proceedings of the Third World Conference on
Neutron Radiography Osaka, Japan (May 14–18, 1989), S. Fujine, K. Kanda, G. Matsumoto,
J. Barton, eds. Pub. Co. Kluwer Academic, Dordrecht, The Netherlands, pp. 875–883 (1990).
32. E07 on Nondestructive Testing E748 Standard Practices for Thermal Neutron Radio-
graphy of Materials Annual Book of ASTM Standards, vol. 03.03 (2007).
33. E07 on Nondestructive Testing E545 Standard Test Method for Determining Image
Quality in Direct Thermal Neutron Radiographic Examination Annual Book of ASTM
Standards, vol. 03.03 (2007).
34. E07 on Nondestructive Testing E803 Standard Test Method for Determining the L/D
Ratio of Neutron Radiography Beams Annual Book of ASTM Standards, vol. 03.03 (2007).
35. E07 on Nondestructive Testing E1496 Standard Test Method for Neutron Radiographic
Dimensional Measurements Annual Book of ASTM Standards, vol. 03.03 (2007).
Chapter 6
Neutron Tomography
W. Treimer
W. Treimer (*)
Department of Mathematics, Physics and Chemistry, University of Applied Sciences
(TFH) Berlin, Luxemburgerstr. 10, D-13353 Berlin; Helmholtz Center Berlin for
Materials and Energy, Department SF1, Glienicker Str. 100, D-14106, Berlin
e-mail: [email protected], [email protected]
6.1 Introduction
Tomography and radiography are well known from their uses in medicine, and
many people have already come into contact with one or more of these meth-
ods. They are closely related because tomography is based on radiography,
which is a two-dimensional (2D) attenuation coefficient distribution of a ray-
path-integrated projection from a three-dimensional (3D) object (Chapter 5). In
the past, X-ray film-converter systems were used for neutron imaging; but since
the mid-1990s digital imaging systems, e.g., image plates or charge-coupled
device (CCD) cameras with scintillator screens, have become increasingly avail-
able (Chapter 4). The quality of digital images is now comparable to that
obtained using film. The advantages of CCD camera-based imaging systems
are easy image processing, high bit depth (optical density), and real-time
imaging and information handling (delivery, storage, post-processing). Thus
radiographic and tomographic techniques for neutron imaging have become
easier to handle and to apply with higher performance since the advent of PC-
based electronic media. It is only a short step from a radiograph to a tomogram,
so some fundamental parameters for both techniques are briefly summarized
here.
For neutron radiography and tomography, four parameters characterizing
the neutron beam are of principal interest: the incident flux intensity , the
wavelength spectrum spread l/l, the divergence , and the given spin state of
the neutron beam in the case of polarized neutrons (Chapter 10). The geometry
required for a radiography setup, for example the distance between sample and
detector, is influenced by the size of the sample and by the pixel size (smallest
scanning part) of the detector. The flux and the wavelength spread are directly
connected because the smaller the l/l, the lower the neutron flux. This also
holds true in principle for the achievable spatial resolution—the smaller the
pixel size (i.e., the higher the spatial resolution), the smaller the integrated flux
at the pixel. Therefore, for conventional radiography and tomography (i.e.,
attenuation-based neutron imaging techniques), the full spectrum (white beam)
is used to get acceptable results in short times. For various other cases
of tomography, the use of monochromatic neutrons is necessary, so a certain
l/l, typically 1–5%, is required to determine structures in samples that are not
revealed by conventional tomography.
The basic properties of the neutron, for example, the relationships between
wavelength, energy, and velocity, and the scattering and absorption cross
sections, have already been introduced in Chapter 1. Scattering processes are
unwanted effects in a classical radiograph or tomogram because they cause a
decrease of sharpness in the image. However, it must be noted that scattering
processes can also be used as imaging signals using different imaging techniques
(Section 6.4). X-ray scattering cross sections generally have an atomic
number–dependent relationship, while neutron cross sections typically have a
‘‘statistical’’ behavior, as shown in Fig. 6.1 (see also Fig. 2.2).
6 Neutron Tomography 83
Fig. 6.1 X-ray and neutron cross sections for different elements. The sizes of the circles
symbolize the strengths of the scattering cross sections
Fig. 6.2 Geometric point broadening: A point P in the object plane is enlarged (blurred) up to
a size d as a result of a given L/D
84 W. Treimer
can be measured), the incident intensity I0 in front of a sample can be calculated from
a given source flux Isource. To realize a point-to-point image, the size of the source and/
or the beam divergence must be kept small. This means that a large L/D ratio
corresponds to a small horizontal and vertical divergence. Furthermore, one must
consider that a point in an object is also enlarged and blurred to an area in the image
by the object-detector (e.g., film, CCD camera) distance, and by scattering effects.
Every point in a sample is enlarged to an area with diameter d on the detector
as d ¼ l/(L/D) if L/D defines the (inverse) divergence h,v (Fig. 6.2). One can
calculate that even for L/D ¼ 1000 (divergence h,v 0.0578 ¼ 0.001 rad), a
point 50 mm from the detector is enlarged to 50 mm. Additional broadening of
an object point is also caused by the process of converting neutrons into visible
light; for example, the thickness of the scintillator material and the optical
system focusing the point onto the CCD chip might inherently decrease the
sharpness of the image. However, it must be noted that in the case of refraction,
or of phase radiography and tomography, the general rule that a shorter
distance between object and detector means a sharper image does not hold.
The imaging system, i.e., the imaging of a given point in a sample, can be
described by the so-called point spread function (PSF) because it involves not
only the interaction with the sample but also the detector resolution (pixel size),
the spread of radiation due to divergence and scintillation (detection) processes,
and optical systems (CCD cameras). The PSF is also known as the impulse
response (Chapter 7). The spatial 2D resolution of a system is best measured by
determining its modulation transfer function (MTF; Fig. 6.3). The MTF can be
derived from the Fourier transform of the PSF, i.e., |MTF| ¼ FT{PSF(x,y)},
the PSF being calculated as [2]
2 =2p ;
PSFðx; yÞ ¼ (6:1)
3
1 þ d2 fx2 þy2 g 2
Fig. 6.3 The point spread function of an imaging system can be derived from the first derivative
of a ‘‘real’’ (blurred) edge (left). The MTF is derived from the Fourier transform of the PSF as
jMTFj ¼ FTfPSFg (right)
6 Neutron Tomography 85
where ¼ 1/dconv, dconv being the distance of the converter foil to the image
plane. Usually dconv can be taken as the thickness of the converter if the film is
attached to it or the CCD camera is focused on it.
So the measured image I(x,y) is the convolution of the 2D PSF with the 2D
projection of the object function OP(x,y) (the 2D Radon transform of the 3D
object), plus background (bg), normalized to the pixel-efficiency "pixel (x,y) of
the detector:
Fig. 6.4 Left: Neutrons hitting the ZnS screen are converted into visible light that is collected
on the CCD of a camera. Right: Light-proofed box containing mirror and CCD optics and
camera. For details see, for example, Lehmann et al. [3]
86 W. Treimer
detectors has nearly superseded the film-converter technique, because all the
image-processing techniques known from medical imaging can also be applied
to neutron radiography and tomography [3].
The number of applications of neutron tomography is quite large, and there
are several papers and books on this topic. However, some new developments
are significant. The experimental conditions have improved, and there is an
increased need for applications of neutron tomography related to industry and
cultural heritage. The new developments are energy-dependent radiography,
(rapid) real-time radiography, refraction or phase contrast radiography and
tomography, small-angle scattering radiography and tomography, and radio-
graphy and tomography using polarized neutrons (the latter is covered in
Chapter 10). However, these are currently not routine methods, and their
potential must be further evaluated.
Fig. 6.5 Scanning geometry of an object (I ) in the {x-y-z} coordinate system. A slice in the
(x–y) plane is scanned through angle (II) and the transmitted intensity is registered in the t-s
rotated coordinate system (II, III). For details, see text
6 Neutron Tomography 87
where I0 is the intensity of the incident neutron beam and f ¼ m(x,y) is the 2D
function to be reconstructed. A new rectangular coordinate system (t,s) is
In(
I0 /
I)
θ
t
Pa
ral
lel S
sca
no
f th
es t = x.cos(θ) + y.sin(θ)
ou
rce
Fig. 6.6 Scanning of a slice in the x–y plane (compare with Fig. 6.5). Note that z is the axis of
rotation, which coincides with the z-axis of the {s-t-z} system
88 W. Treimer
defined that expresses the rotated detector system (t,s) with respect to the fixed
sample system (x,y) (or vice versa, if the sample is rotated as is done in neutron
tomography). Transforming the (x,y) system into the (t,s) system, t can be
expressed as
The set of all projections P (t) of m(x,y) is called the Radon transform of
m(x,y). From these projections, a 2D image can be reconstructed using the
‘‘Fourier slice theorem.’’ This states that the Fourier transform of a projection
P(t) of m(x,y) is a subset of a function M(u,v), which is the Fourier transform of
m(x,y), i.e.,
Z1 Z1
Mðu; vÞ ¼ ðx; yÞ e2piðuxþyvÞ dxdy: (6:7)
1 1
To prove the slice theorem, consider the simple case of ¼ 0. In this case, the
two coordinate systems (x,y) and (u,v) coincide, so the projection P(t) is simply
the integral (Eq. [6.10]):
Z1
P¼0 ðtÞ ¼ ðx; yÞ dy : (6:8)
1
Z1 Z1
Mðu; 0Þ ¼ ðx; yÞ e2piðuxÞ dxdy
1 1
8 9 (6:9)
Z1 < Z1 =
¼ ðx; yÞ dy e2piðuxÞ dx :
: ;
1 1
Z1
Mðu; 0Þ ¼ P¼0 ðxÞ e2piux dx : (6:10)
1
Z1
S ðwÞ ¼ P ðtÞ e2piwt dt ; (6:11)
1
then S¼0(w) is equal to M(u,0). Because the orientation of the object in the (u,v)
system is arbitrary with respect to the (x,y) system, this relation allows it to be
used for all angles , keeping in mind that the Fourier transform conserves the
rotation made in real space. The Fourier transform of P(t) yields the values of
the Fourier transform of m(x,y). A dense set of values in the Fourier space (given
in polar coordinates because of the polar scanning of the object) can be
approximated to rectangular coordinates and back-transformed into real
space. This may sometimes lead to diffuse reconstructions as a result of incom-
plete fitting for high frequencies in the Fourier domain (Fig. 6.7).
V V
u u
Fig. 6.7 Plot of Fourier-transformed projection P (left) and interpolation of the data onto a
rectangular coordinate system (right). The interpolation for higher frequencies (large (u,v)
values) becomes imprecise
90 W. Treimer
How many projections yield the best result? This is answered by the Shannon
theorem, which states that a unique reconstruction from a sampled object is
obtained if the object is sampled with a frequency twice as high as the highest
frequency in the Fourier-transformed image. For parallel mode scanning, using
N steps for one projection, it can easily be derived that M projections (angles )
are required to fulfill the Shannon theorem for tomography. If D is the diameter
of the object to be scanned, and x is the difference between two scanning
points, then the number of scanning points per projection is
D
N¼ ; (6:12)
x
fulfilling the Shannon condition. The rotation of the sample around 1808 means
a path length y for a point sitting on the circle with the radius D/2. We require
that x ¼ y, so that the number of projections M must be
D
2 p Dp
M¼ ¼ : (6:13)
y 2 y
p
M N : (6:14)
2
N and M determine the quality of the reconstruction. If M does not fulfill the
inequality, i.e., if fewer than M projections are registered, then the reduced D ¼
D* that fulfills the Shannon condition can be calculated for a smaller number of
projections.
Z1 Z1
ðx; yÞ ¼ Mðu; vÞ e2piðuxþvyÞ dxdy : (6:15)
1 1
Z2p Z1
ðx; yÞ ¼ Mðw; Þ eþ2piwðx cosðÞþy sinðÞÞ w dw d : (6:16)
0 1
The transformation of the (x,y) system into the (w,) system uses the relations
u ¼ w cosðÞ
v ¼ w sinðÞ (6:17)
du dv ¼ w dw d :
Zp Z1
ðx; yÞ ¼ Mðw; Þ eþ2piwt jwj dw d
0 1
2 3 (6:18)
Zp Z1
¼ 4 S ðwÞ eþ2piwt jwj dw5d :
0 1
The integral given in the square brackets can be considered as the Fourier
transform of P(t). However, S(w) is multiplied by the |w| function, which
is a special filtering function of the frequencies in the frequency space. We
define
Z1
Q ðtÞ ¼ S ðwÞ jwj eþ2piwt dw : (6:19)
1
Fig. 6.8 Scanning of a simple 2D object (square and circle) and corresponding projections
(figure courtesy of T. Donath, GKSS)
6 Neutron Tomography 93
Fig. 6.9 The sinogram of Fig. 6.8. The y-axis shows the projection as gray values of Fig. 6.8
This section summarizes some techniques that are becoming increasingly rele-
vant in neutron tomography and that will be used for a range of applications in
the future.
(a) (b)
Fig. 6.12 (a) Attenuation of an element (polycrystalline) as a function of neutron wavelength
(b) attenuation coefficient for aluminum in the range 2–6 Å [8]
6 Neutron Tomography 95
I1
I01 I1 I02 eSðl1 Þd
I2
¼ ¼ ¼ e½Sðl2 ÞSðl1 Þ d : (6:23)
I02
I2 I01 eSðl2 Þd
In the case of S(l1) S(l2), for example, as shown in Fig. 6.12 (a), the
exponential function becomes nearly unity and there will be no contribution of
this material to the normalized image; i.e., this material becomes transparent in this
image.
Original sample
Radiograph with
2.62 Å
Fig. 6.13 Radiographs of a
bent metal plate (50 10
5 mm, top picture) Radiograph with
performed at two different 3.55 Å
wavelengths (2.62 and
3.55 Å). For details see [8]
96 W. Treimer
(a) (b)
Fig. 6.15 Ratios of properly normalized radiographs of polyethylene and metal wedges for
different wavelengths. Some materials become transparent for particular wavelength ratios,
for example, copper in (a) and steel in (b) [8, 9]
6 Neutron Tomography 97
Rr 1
AFZ ¼ pl : (6:24)
Rþr sinðÞ
98 W. Treimer
Here R is the distance from the point source to the object, r is the distance
between object and detector, and is the angle between the ray propagation
vector and the plane of observation (usually ¼ 908). (Compare with draw-
ings in classical optic books). The size AFZ determines the coherent source
for any interference effects to be observed. For standard geometries in
neutron tomography (and radiography), R 500 cm, r 1 cm, and l
0.5 nm, so AFZ is 1.6 10–7 cm2 and the diameter of the coherent source
can roughly be estimated as 4 mm. To use the coherent properties of such a
beam, a point-like source must be created by a small pinhole, which deter-
mines—together with the divergence of the beam—the coherent area of the
real neutron beam. The diameter dcoh characterizing the coherent size of a
source can be estimated by simple geometric considerations as given by [13].
1L
dcoh ¼ l (6:25)
2D
To achieve a reasonable dcoh, the source (i.e., D) must be kept small and
L must be large (meters).
To understand the phase contrast, consider a (plane) wave traveling through
an object that could be smaller or larger than the coherent wave front of the
wave (Fig. 6.16).
Phase contrast originates from a coherent superposition of diffracted
waves; the resulting intensity variation can be observed only in the coherent
part of the wave that propagates through the object and experiences position-
dependent phase shifts caused by different indices of refraction in the sample.
This modulates the wave front, and a change of propagation direction will be
observed. The strongest phase contrast is observed in the near field region, z <
dcoh2/l, where the modulation of the amplitude is best; it broadens with
increasing z [13]. (z is the distance from object to detector and dcoh dimen-
sion of the coherence width.) Fringes can be registered that are caused, for
example, by an edge that would be washed out if the beam divergence and/or
(a) (b)
Fig. 6.16 An object is illuminated by a coherent wave front, which can be (a) larger or (b) smaller
than the object
6 Neutron Tomography 99
source size were too large. It can be calculated that the optimal contrast occurs
at z ¼ dcoh2/2l [13]. One approach for realizing phase contrast was performed
by neutron radiography some years ago [14]. The first experiments based on
phase effects (refraction) were performed in 1989 using a double crystal
diffractometer [15]. The first neutron radiography based on phase contrast
was performed in 2000 at the National Institute for Standards and Technol-
ogy (USA) using a pinhole and a wavelength of 4.43 Å and with the sample
placed l.8 m from the source [16, 17]. However, phase contrast imaging is
much more often used with synchrotron radiation because of the higher
brilliance of the source.
One can calculate the phase contrast, determining the Fourier-transform of
the structure or of the phase boundary of two structures. This can be done by
ray tracing, assuming that a certain width of the edge is coherently illuminated.
Here we assume that an incident plane wave hits a phase step (boundary of
two different materials with a difference of 10 mm in thicknesses). However, in
general, one can also use ray-tracing, adding up all amplitudes emerging from
the structure that cause the phase gradient that contributes to a measuring point
showing an intensity distribution as plotted in Fig. 6.17, top.
1
500 550 600 650 700
Fig. 6.17 Calculated phase contrast (top; x-scale in mm, y-scale in arbitrary units) for a phase
step consisting of silicon and aluminum (bottom) [12]
100 W. Treimer
The phase development quite close behind the sample yields strong inten-
sity oscillations that can be measured at different distances from the object
[18]. Very often, another concept is used to describe this phase contrast
correctly via geometric considerations of refraction by edges [19]. The latest
developments use a phase grating-based shearing interferometer; expressed
simply, this senses the phase-modulated (refracted) beam behind the sample
using an analyzing grating. The measured intensity depends on the differ-
ence in phase shift introduced by two adjacent regions in the sample that are
coherently illuminated. Several images must be measured with varying phase
grating positions and the phase change due to the object then calculated.
Additional details concerning this technique can be found in Chapter 8 and
in [20, 21].
X
n
¼ i : (6:26)
i¼1
Fig. 6.18 Neutrons are refracted several times in the sample (e.g., by inhomogeneities) and
exit with a deviation angle
The change of the phase ’(x,y) (here is not the Laplacian operator) can
be expressed as the sum of all deviations that the ray experiences:
Z
’ðx; yÞ ¼ k ðx; yÞ ds : (6:28)
path
1
ð; tÞ ¼ r? ðjðx; yÞÞ : (6:29)
k
So P(t) becomes
Z t
P ðtÞ ¼ ðx; yðx; ; t0 ÞÞ dt0
0
Z 1 Z 1
(6:30)
¼ Dirac ðx cosðÞ þ y sinðÞ tÞ dx dy :
1 1
l2 Nbc
n¼1 (6:31)
2p
where is the scattering cross section, N(x,y) is the particle density distribution,
and R is a parameter with the dimension of a length specifying an average size or
correlation length in the scattering object (corresponding to the Gaussian
approximation). Now one can use B or B2 as an imaging signal to use (ultra)
(a) (b)
The most astonishing result was the detection of clusters of particles (ß-
carotene) having a size of approximately 150 nm, dissolved in D2O ( 5–12%).
In this experiment all effects—refraction, USANS, and absorption—were mea-
sured simultaneously with the double crystal diffractometer. The shift of the RC
yields the final deviation and determines the refraction, the broadening of the
small-angle scattering, and the decrease of the integral intensity of the absorp-
tion of the neutron beam. All these data are derived from a single curve.
Combining USANS and refraction data yields a complete reconstruction.
With pure attenuation, and only incomplete refraction or USANS data, imper-
fect images are reconstructed [24, 27].
Chapter 16 describes the use of modeling to reconstruct the shapes of biolo-
gical macromolecules with a typical size of <100 nm based on small-angle
scattering data. In combination with the methods described here, this offers the
possibility of covering the full range of length scales from macroscopic to nano-
scopic with neutron imaging methods.
6.5 Outlook
There are several other fields of application of neutron tomography that should
be mentioned here. In stress and strain investigations (Fig. 6.13), the investiga-
tion time can be drastically reduced by 3D scanning of a sample using Bragg-
edge and energy-resolved techniques (Chapter 12). There seems to be a great
interest in investigating quasi-in-situ water distribution in operating fuel cell
stacks by neutron tomography (Chapter 11 and e.g., [30]), and this also
holds true for applications in the fields of moisture detection and geological
sampling (Chapter 17) and noninvasive investigation of cultural heritage obj-
ects (Chapter 13 and, e.g., [3]). Another interesting topic will be the use of
polarized neutrons (Chapter 10). The imaging signal is the change of the
spin direction with respect to the orientation in front of the sample; i.e., the
2D-depolarization image of a neutron beam will be measured. The neutron
depolarization technique is well established; the problem is to calculate a unique
reconstruction, for example, of two-dimensionally (three-dimensionally) ran-
domly arranged small magnetic fields in samples, such as complicated magnetic
flux distributions in solids, magnetic domains, etc. Several attempts to solve this
problem have been published [31]; however, a reconstruction technique applied
to refraction or small-angle tomography similar to that discussed in [22] or
iterative techniques seem to be most challenging and promising.
References
1. S.G. Glasstone, M.C. Edlund, Kernreaktortheorie, Wien, Springer-Verlag, p. 30 (1961).
2. A.A. Harms, D.R. Wyman, Mathematics and Physics of Neutron Radiography, D. Reidel
Publishing Company, Dordrecht, Holland (1986).
6 Neutron Tomography 107
27. W. Treimer, M. Strobl, A. Hilger, H.J. Peschke, Neutron tomography using small angle
scattering data, IEEE 52:1, 386–288 (2005).
28. N. Kardjilov, I.Manke, M. Strobl, A, Hilger, W. Treimer, M. Meissner, T. Krist,
J. Banhart, Three-dimensional imaging of magnetic fields with polarized neutrons. Nat.
Phys. 4, 399–403 (01 May 2008).
29. W. Treimer, BMBF project 03TR7TFH 2007–2010.
30. I. Manke, Ch. Hartnig, M. Grünerbel, J. Kaczerowski, W. Lehnert, N. Kardjilov,
A. Hilger, J. Banhart, W. Treimer and M. Strobl, Quasi-in situ neutron tomography on
polymer electrolyte membrane fuel cell stacks, Appl. Phys. Lett. 90, 184101 (2007).
31. E. Jericha, R. Szeywerth, H. Leeb, G. Badurek, Reconstruction techniques for tensorial
neutron tomography, Physica B 397, 1–2, 159–161 (2007).
Chapter 7
Mathematics of Neutron Imaging
Abstract Imaging with neutrons at macro-world scales (e.g., >10 mm) requires
particular understanding of the non-diffracting or refracting interactions that these
electrically neutral particles have with their environment. While image formation
with neutrons shares some commonality with other radiation sources such as X-rays
and gamma rays, neutrons provide complementary interaction mechanisms to these
techniques that can uniquely probe materials and structure. Neutron sources are
also historically fraught with issues regarding low source intensities, challenging
beam configurations and resolution limitations that are all closely linked. In this
chapter we will present a mathematical construct and methods compatible with the
design and characterization of radiography systems for volumetric imaging. We
begin with a review of neutron image formation, provide resolution analysis con-
cepts and methods for both the design and the characterization of radiography
systems, and conclude with a review and discussion of volumetric reconstruction
techniques using analytic or iterative computed tomography algorithms.
7.1 Introduction
heat exchanger systems [2, 10], development of hydrogen fuel cells (Chapter 11),
inspection of cultural heritage objects (Chapter 13), and interrogating biologi-
cal systems [28]. Newer, intense sources of neutrons from spallation facilities are
providing the potential to interrogate time and energy-dependent phenomena
as well. Therefore, a mathematical representation of the neutron image forma-
tion process can provide insight into the potential application of time- and
energy-dependent, volumetric radiography to the characterization and quanti-
tation of complex material structures.
The mathematics of neutron imaging is similar to that of other non destruc-
tive methods that use non-refracting or diffracting radiation to interrogate
materials or composites of materials [3]. Figure 7.1 represents the basic elements
of a generic neutron radiography system beginning with a source of neutrons,
beam conditioning and shaping, neutron conversion and detection, and sam-
pling and signal generation.
For the purposes of this discussion, neutron image formation will be treated
as an encoding of the material characteristics of a three-dimensional (3D) object
projected onto a two-dimensional (2D) detection array, typically comprising a
converter followed by an imaging sensor as described in Chapter 5. Material
characteristics of the object are represented by the total macroscopic interaction
cross-section, T, which is a sum of the absorption and scattering cross-
sections, T = a+s (i.e., ignoring the potential for neutron capture and
fission cross-sections in some materials). The interaction of the neutron beam,
o, with the 3D object produces a total 2D field at the converter plane, T, that
can be decomposed into the uncollided flux, u, and the scattered flux, s, as
T = u+ s. The uncollided flux arises from neutrons that arrive at the con-
verter plane unperturbed by the imaged object while the scattered flux considers
only neutrons that have been scattered by the object toward the converter plane.
To understand the dependencies of the scattered flux on the image formation
process, we introduce a buildup factor [20, 24] defined as B = T/u= 1+s/u.
Rearranging to solve for T, we can represent the total flux at the converter
plane as T = Bu, i.e., a multiplicative build up of the uncollided flux due
to object scattering. This is represented using the well-known neutron
θ
z
n
n γ
Beam conditioning/shaping φ T (E,θ,ΣT ,x,y,t) I(E,θ,ΣT ,x,y,t)
Projected flux Signal
where x and y are spatial position in the plane perpendicular to the ray from
source to detector, z represents position along ray from source to detector and E is
the neutron energy. In this relationship, corresponds to the angular orientation
of the object under test (Fig. 7.1) and is included as a prelude to later discussions
regarding volumetric reconstruction from projections. Also note that the buildup
factor is a value B 1. It is functionally dependent on the material composition
and geometry presented by the object under test and is therefore difficult to
predict or accommodate under typical radiographic conditions. Buildup can be
significant for materials containing, e.g., H, Si, Ni, Cu, and other highly scattering
metals, and should be considered a source of error in quantitative analysis. For
example, imaging geometries that move the object away from the detection plane
can reduce the number of neutrons scattered into the detector, but this may also
reduce the system resolution due to the characteristics of the beam aperture. The
beam aperture and length is typically characterized by the L/D ratio, where L
denotes the distance from the aperture to the object and D is the diameter of the
aperture. This effect will be detailed further in Section 7.2.
Note that a primary goal of image analysis can be formulated as one of
measuring the interaction cross-section of the materials in the imaged object,
T(E,x,y,z). Direct measurement of T, for example, provides indirect informa-
tion regarding other material properties since = N = NA/M, where is
the microscopic cross-section, N is the number of atoms per cm2, is the material
density, NA is Avagadro’s number, and M is the atomic weight [25].
The electronic signal produced by the system is further impacted beyond the
geometry effects of L/D by the response of the converter, optical components,
and sensor (e.g., a charge-coupled device, CCD). Making the practical assump-
tion of a linear shift invariant (LSI) system [14], we can describe the time-,
space-, and energy-dependent 2D electronic image, I, in terms of a convolution
of the modulated (by the object) neutron field at the converter, T, with a system
impulse response function, h, to give,
Z Z
IðE; ; T ; x; y; tÞ ¼ T ðE; ; T ; ; ; tÞhðx ; y Þdd
(7:2)
¼ T ðE; ; T ; x; y; tÞ hðx; yÞ
In Section 7.2 we will model and analyze the impulse response of a radio-
graphy system to estimate system design resolution and characterize resolution
of an existing radiography system through empirical means. In Section 7.3 we
will describe considerations and methods required to perform volumetric recon-
struction and estimation of T using projection data and computed tomogra-
phy based on both analytic and iterative techniques.
112 K.W. Tobin et al.
Scintillator
z
n
Optical
Beam line system
L d
x '' x' x
Fig. 7.2 Representation of a basic radiographic image formation system used to estimate
resolution performance during the design stage. Model accounts for geometry, aperture size,
scintillator resolution, and optical system resolution
7 Mathematics of Neutron Imaging 113
0 x
x 00 x
hD ðx 00 Þ ¼ rect ; h ðx 0 Þ ¼ Gauss ; hCCD ðxÞ ¼ rect ; (7:3)
D s
where sinc(x) = sin(x)/x. Note that this expression has been put into a form
explicitly containing the L/D ratio.
In Eq. (7.5), the sinc () term drives much of the response typically observed
in a radiography system. For example, as L/D !1, this term broadens out in
frequency representing improved resolution. Also, as the object moves closer to
the detection plane, i.e., as d! 0, resolution also improves. These are both
effects that are readily observed in radiography systems.
An example of the MTF is shown on the left-hand-side of Fig. 7.3 for a system
with the model parameters: L/D=150, d=10 cm, =50 mm, MCCD=0.25. MTF
is a measure of system response versus frequency content of the object being
imaged. In this figure, the X axis represents frequency, with units of cycles per
millimeter often referred to as line pairs per millimeter. The Y axis represents the
transmission of this frequency to the image. In this example, the response to a
sinusoid with 10 cyc/mm drops to approximately 55% of the sinusoid’s amplitude.
Applying the Rayleigh criterion of 10% MTF [16] to this example, indicates
that the highest resolvable frequency is 17 cyc/mm. System resolution is defined
114 K.W. Tobin et al.
0.8 0.8
L /D = 150
0.6 0.6
MTF (μ)
Fig. 7.3 MTF modeling results for typical radiography system parameters. The left plot shows
the frequency response of constituent system components at a specified L/D ratio, while the
right plot shows the system response for a variety of L/D ratios
0.8
0.6
MTF
0.4
0.2
Projection Reconstruction 0
0 0.1 0.2 0.3 0.4 0.5
Frequency (lp/pixel)
Fig. 7.4 Spatial resolution from CT reconstruction using ASTM E1695 standard
116 K.W. Tobin et al.
75
Central
Contrast sensitivity (%)
reconstructed 60
slice
Tiling 45
30
15
1/3 radius
0
1 2 3 4 5 6 7
Region size in pixels (N×N)
Fig. 7.5 Contrast discrimination measurement geometry (left) and CDF results (right)
7 Mathematics of Neutron Imaging 117
where I0 and I refer to the beam intensity before and after interaction with
the object, K denotes the path of the beam, and m represents the material-
dependent macroscopic cross-sections. To reconstruct m, which corresponds to
T, in the preceding section for neutron radiography, we apply log-normal-
ization to isolate the cross-section (Eq. (7.1), assuming B1),
Z
I
mðÞ d ¼ log : (7:8)
K I0
118 K.W. Tobin et al.
We note that in practice, the result of a dark field scan (where the source is
turned off) is often subtracted from both I0 and I to compensate for noise
generated by physical processes that occur inside the imager (CCD).
Different imaging geometries require different reconstruction algorithms.
We consider both parallel and divergent beam collimation of the neutron
beam as shown in Fig. 7.6. We assume that the entire object is visible in each
projection image and that it is placed on a turntable to allow for a circular orbit-
type scan. Generalization to a helical scan is possible but beyond the scope of
the present exposition.
I(θ,s) I(θ,s(1+d/L))
s s
y y
x x
θ θ
u u
L d
Fig. 7.6 Parallel beam (left) and fan beam (right) sampling geometries. Cone beam sampling
extends the fan beam geometry axially
Z2p
1
mðx; y; zÞ ¼ g2 ð; x; yÞ~
pð; gð; x; yÞðx cos þ y sin ÞÞ d
2
0
(7:13)
gð; x; yÞ ¼L=ðL þ x sin y cos Þ
pffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
p~ð; sÞ ¼L= L2 þ s2 pð; sÞ hRAMP ðsÞ
where L refers to the distance from the neutron beam aperture to the isocenter
of the object (c.f. Fig. 7.6). The detector is assumed to be located at the isocenter
for mathematical simplicity (i.e., d=0). Notice that the backprojection takes
place over 2p instead of just p as used for the parallel beam geometry. Also note
that L! 1 leads to the parallel beam algorithm.
120 K.W. Tobin et al.
Notice that z=0 leads to the fan-beam algorithm. Also note that filtering is
not applied in the axial dimension.
We mentioned that the fan-beam backprojection equation indicates that the
source trajectory data must cover 2p. However, by introducing a weight func-
tion that effectively limits each projection ray to be considered just once, Parker
[34] showed that it is possible to stop when p+ has been covered where
denotes the fan angle. The idea, which is heuristic, can be extended to apply to
cone-beam data as well.
Filtered backprojection algorithms, whether exact or approximate, are based on
continuous mathematics. When implemented for practical use, all integrals
are replaced by Riemann sums and nonintegral projection data lookup indices are
handled using some form of interpolation. The filtering can take place in either the
spatial domain or the frequency domain. When based on fast fourier transforms, the
latter can be done very efficiently. In that case, backprojection along with interpolation
dominates the computational cost. Still, given recent advances in computer technol-
ogy, it is possible to reconstruct even large images in a very short amount of time.
The image quality depends on taking enough projections and having good
count statistics to work with. Not taking enough projections invariably leads to
streak artifacts. Indeed, they are almost always present to some degree. Poor
count statistics results in a low-contrast, grainy image. Under normal imaging
conditions, however, good images are produced.
7 Mathematics of Neutron Imaging 121
Henceforth, let vectors x and b refer to the image representing the macroscopic
cross-sections and the log-normalized projection data, and let matrix A denote
the discretization of the projection integral connecting the two. Reconstruction
can then be viewed as a matter of solving Ax=b. Numerous applicable iterative
algorithms exist ranging from Gauss–Seidel, Jacobi, and other stationary
methods to conjugate gradient (CG) and other Krylov subspace methods.
Stationary methods can be derived using matrix splitting [36]. For example,
A=MN leads to the iteration scheme,
where x(0) is given, d(0)=r(0) = b Ax(0), |r|2 = rTr, and |d|A2= dTAd.
Whether a stationary method or a Krylov subspace method is used, the
search for x is terminated either when the residual norm associated with the
current iterate is deemed to be small enough, or when a fixed number of
iterations have been executed.
In most imaging applications, the linear system is under-or overdetermined
and a least squares problem must be solved instead. Let this be denoted by
normal equations which form a linear system that can be solved by any of the
above-mentioned methods, namely,
AT Ax ¼ AT b (7:18)
where R(x) denotes the penalty term and is a free parameter that controls the
influence exerted thereby on the solution. From a computational point of view,
linear penalty terms are preferable as they can be handled by straightforward
row-expansion of A and b.
7 Mathematics of Neutron Imaging 123
Many platforms exist that can provide the compute power needed in order for
a volumetric image to be reconstructed in a reasonable amount of time. A
relatively small cluster of multi-core PCs may suffice to solve even a rather
large problem in a matter of minutes [18]. From a programming point of view
this is conceptually straightforward as it merely involves writing multi-
threaded code for a distributed memory environment. In recent years, com-
modity graphics processing units (GPUs) have seen a rapid increase in both
capability and programmability to the point that they are now being used for
general purpose computing and thus also tomographic image reconstruction
[31]. Papers often report order-of-magnitude performance gains over opti-
mized CPU applications. However, GPU programming is not for the novice.
Also, floating-point arithmetic may not follow IEEE standards, which could
be a problem. Field-programmable gate-arrays (FPGAs) provide a different
but equally powerful platform from a data throughput point of view. How-
ever, FPGAs are notoriously difficult to program, especially since floating-
point arithmetic may not be supported to the extent needed. Finally, we
mention the cell processor, which conceptually is somewhat similar to a
multi-core CPU only orders-of-magnitude faster. High-level programming is
possible, e.g., in C, but some thought must be devoted to memory handling as
the processing units only have a limited amount of memory available. None-
theless, the cell may represent the fastest platform currently on the market for
volumetric imaging [22].
The ability to model, design, and measure the resolving power of a radiography
system is critically important to the experimental design required to achieve
relevant measurement characteristics of complex, material structures. The focus
of this chapter has been on methods to design and characterize radiography
systems according to experimental goals for volumetric measurements (i.e., CT
recording and reconstruction methods). This focus has been primarily related to
spatial resolution and the estimation of macroscopic material interaction cross-
sections.
New or modified radiography facilities that generate cold neutrons [38]
provide wavelength tunable monochromaters [39] or provide time-of-flight
(TOF) measurements [27], facilitating the radiographic capture of images that
take I(,T,x,y,z) to I(E,,T,t,x,y,z). The ability to better characterize
the energy–time–space dependencies of materials, T(E,t,x,y,z), will produce
better material resolution.
As an example of how to adapt this discussion to higher resolution measure-
ments, consider the example MTF shown in Fig. 7.7. Using Eq. (7.5), we have
7 Mathematics of Neutron Imaging 125
MTF (μ)
comparable to the
scintillator limit. How do we
achieve sufficient neutron 0.4 Total MTF
counting statistics? Beam Aperture
Scintillator
0.2
CCD/Optics
10% Rayleigh criterion
0
0 20 40 60 80
μ (cyc/mm)
The point is that these analysis tools can be adapted to new, more stringent
resolution requirements by applying creative ideas and methods.
For another example, the steady-state flux of a conventional reactor cannot
readily interrogate time-dependent phenomena as deeply as the high-intensity
and energy-dependent beams available at a spallation neutron source. Using a
spallation source, the potential exists to develop new time-synchronized,
energy-resolved methods to better resolve subtle material differences. This is
potentially achievable by taking advantage of the Bragg edges in materials [27]
or of the fact that absorption cross-sections, a at low neutron energies
p
are typically proportional to 1/ E. Using prior knowledge of the material
characteristics under test should facilitate the collection of a series of
126 K.W. Tobin et al.
References
1. R. Accorsi, F. Gasparini, et al., A Coded aperture for high-resolution nuclear medicine
imaging with conventional anger camera. IEEE Trans. Nucl. Sci. 48:6, 2411 (2001).
2. H. Asano, N. Takenaka, et al., Visualization and void fraction measurement of gas-liquid
two-phase flow in a commercial plate heat exchanger by thermal neutron radiography.
IEEE Trans. Nucl. Sci. 52:1, 280–284 (2005).
3. M. Balasko, E. Svab, et al., Comparison of neutron radiography with other nondestruc-
tive methods. IEEE Trans. Nucl. Sci. 52:1, 330–333 (2005).
4. L.G.I. Bennett, T.R. Chalovich, et al., Comparison of neutron radiography with other
non-destructive techniques for the inspection of cf188 flight control surfaces. IEEE Trans.
Nucl. Sci. 52:1, 334–337 (2005).
5. M. Bierlaire, P.L. Toint, et al., On iterative algorithms for linear least squares problems
with bound constraints. Linear Algebra Appl. 143, 111–143 (1991).
6. P. Bingham, L. Arrowood, et al., Calibration and performance testing for reconfigurable
computed tomography systems. Mat. Eval. 65, 1102–1107 (2007).
7. G.D. Boreman, Modulation Transfer Function in Optical and Electro-Optical Systems.
Bellingham, WA, SPIE Press (2001).
8. C.Bouman and K. Sauer, A generalized Gaussian image model for edge-preserving MAP
estimation. IEEE Trans. Image Process. 2, 296–310 (1993).
9. S. Casalta, G.G. Dawuino et al., Digital image analysis of X-ray and neutron radiography
for the inspection and the monitoring of nuclear materials. NDT&E Int. 36, 349–355
(2003).
10. J.M. Cimbala, J.S. Brenizer, et al., Study of a loop heat pipe using neutron radiography.
Appl. Radiat. Isot. 61, 701–705 (2004).
11. A. Dempster, N. Laird, et al., Maximum likelihood from incomplete data via the EM
algorithm. J. R. Stat. Soc. B 39, 1–38 (1977).
12. R.O. Duda, P.E. Hart, et al., Patt. Class.. New York, John Wiley & Sons, Inc. (2001).
13. L.A. Feldkamp, L.C. Davis, et al., Practical cone beam algorithm. J. Opt. Soc. Am. A 1,
612–619 (1984).
14. J.D. Gaskill, Linear Systems, Fourier Transforms, and Optics, John Wiley & Sons, Inc.
(1978).
15. P. Gilbert, Iterative methods for the three-dimensional reconstruction of an object from
projections. J. Theor. Biol. 36, 105–117 (1972).
16. J.W. Goodman, Introduction to Fourier Optics, Roberts & Company Publishers (2004).
17. S.R. Gottesman, and E.E. Fenimore, New family of binary arrays for coded aperture
imaging. Appl. Opt. 28:20, 4344 (1989).
18. J. Gregor and T. Benson, Computational analysis and improvement of SIRT. IEEE
Trans. Med. Imaging 27, 918–924 (2008).
7 Mathematics of Neutron Imaging 127
19. J. Gregor and F.R. Rannou, Least squares framework for projection MRI reconstruction.
Medical Imaging, San Diego, CA, Proceedings of SPIE (2001).
20. A.A. Harms and D.R. Wyman, Mathematics and Physics of Neutron Radiography.
Boston, D. Reidel Publishing Company (1986).
21. T. Hebert and R. Leahy, A generalized EM algorithm for 3-D Bayesian reconstruction
from Poisson data using Gibbs priors. IEEE Trans. Med. Imaging 8, 194–202 (1989).
22. M. Kachelriess, M. Knaupp, et al., Hyperfast parallel-beam and cone-beam backprojec-
tion using the Cell general purpose hardware. Med. Phys. 34, 1471–1486 (2007).
23. A. Kak and M. Slaney. Principles of Computerized Tomographic Imaging, IEEE Press
(1988).
24. G.F. Knoll, Radiation Detection and Measurement, John Wiley & Sons, Inc. (2000).
25. J.R. Lamarsh, Introduction to Nuclear Engineering, Addison-Wesley Publishing Com-
pany (1983).
26. C.L. Lawson and R.J. Hanson, Solving Least Squares Problems, Prentice-Hall (1974).
27. E. Lehmann, P. Vontobel, et al., Neutron imaging – present status and options with tof
methods – TOF neutron imaging. J. Neutron Res. 13, 27–31 (2005).
28. U. Matsushima, Y. Kawabata, et al., Measurement of changes in water thickness in plant
materials using very low-energy neutron radiography. Nucl. Instrum. Meth. Phys. Res.
A 542, 76–80 (2005).
29. S.L. Meeks, J.F. Hamon, et al., Performance characterization of megavoltage computed
tomography imaging on a helical tomotherapy unit. Med. Phys. 32, 2673–81 (2005).
30. J.J. Milczarek, A. Czachor, et al., Dynamic neutron radiography observations of water
migration in porous media. Nucl. Instrum. Meth. Phys. Res. A 542, 232–236 (2005).
31. K. Müller and F. Xu, Accelerating popular tomographic reconstruction algorithms on
commodity PC graphics hardware. IEEE Trans. Nucl. Sci. 3, 654–663 (2005).
32. M. Ohkubo, S. Wada, et al., An effective method to verify line and point spread functions
measured in computed tomography. Med. Phys. 33, 2757–64 (2006).
33. C.C. Paige and M.A. Saunders, LSQR: An algorithm for sparse linear equations and
sparse least squares. ACM Trans. Math. Software 8, 43–71 (1982).
34. D. Parker, Optimal short-scan convolution reconstruction for fanbeam CT. Med. Phys. 9,
254–257 (1982).
35. A. Rosenfeld and A. Kak, Digital Picture Processing, Academic Press (1976).
36. Y. Saad, Iterative Methods for Sparse Linear Systems, PWS Publishing Co (1996).
37. E. Samei and M.J. Flynn, A method for measuring the presampled MTF of digital
radiographic systems using an edge test device. Med. Phys. 25, 102–113 (1998).
38. M. Tamaki, Conceptual monochromatic digital neutron radiography using continuous
cold neutron beam. Nucl. Instrum. Meth. Phys. Res. A 542, 32–37 (2005).
39. W. Treimer, M. Strobl, et al., Wavelength tunable device for neutron radiography and
tomography. Appl. Phys. Lett. 89, 203504 (2006).
40. H. Tuy, An inversion formula for cone-beam reconstruction. SIAM J. Appl. Math. 43,
546–552 (1983).
Chapter 8
Neutron Phase Imaging
F. Pfeiffer
8.1 Introduction
From a classical physics point of view, neutrons are small massive particles with a
confinement radius of about 1015 m and a distinct internal quark-gluon structure.
In quantum mechanics, neutrons are described by de Broglie [1] wave packets
whose spatial extent may be large enough to show interference effects similar to
what can be observed with visible laser light or highly brilliant X-rays from
synchrotron sources. Measurements of the neutron wave packet’s phase shift
induced by different interaction potentials have a long and distinguished history
in the exploration of the fundamental principles of quantum mechanics [2, 3]. If
such phase-sensitive measurements are further combined with neutron-imaging
approaches, two or even three dimensionally resolved spatial information on the
quantum mechanical interactions of massive particles with matter can be obtained.
In the following, the term ‘‘phase imaging’’ is used to describe not only phase
contrast methods, which simply record some form of contrast generated due to
a phase shift induced by the object, but also methods that determine the phase
F. Pfeiffer (*)
Department of Physics (E17), Technical University Munich,
85747 Garching/Munich, Germany
e-mail: [email protected]
8.2 Principles
where (x,y,z) describes the decrement from unity of the real part of the
refractive index and (x,y,z) is the imaginary part of the refractive index. A
conventional neutron attenuation-based projection image maps the imaginary
part of n and is given by
Z Z
Iðy; zÞ
¼ exp ðx; y; zÞdx ¼ exp 4pðx; y; zÞ=ldx ; (8:2)
I0 ðy; zÞ
n with = l/4p . Generally contains not only the specific neutron capture
cross-section of the material, but also the effect that neutrons are removed from
the beam traversing the specimen to coherent or incoherent scattering
processes.
In analogy to the description of attenuation-based projections, the projected
phase shift of an object can be written as [3]
Z
2p
ðy; zÞ ¼ ðx; y; zÞdx : (8:3)
l
A phase shift projection image thus can give direct information on the real part
of the refractive index distribution in the object and carries additional and
complementary information if compared to an attenuation image.
Unlike X-rays, where the values of and are linked through the X-ray
optical properties of the materials, the corresponding values for neutrons can be
very different. Table 8.1 shows values for several materials for a neutron
wavelength of 4Å. It is interesting to note that, for example, titanium and
copper have very similar values, but opposite signs in . Moreover, we observe
that aluminum and silicon are characterized by very small values and thus
yield very little contrast in a conventional attenuation image. Their relatively
large values, on the other hand, make them easily detectable in phase-sensitive
imaging (see below). A graphical overview of the distribution of values and
corresponding attenuation coefficients is shown in Fig. 8.1.
Table 8.1 Values for the complex neutron refractive index n = 1-+i of several representa-
tive materials for neutrons with a wavelength of 4Å [4, 5]
Material /
Al 5.3 106 9.8 1011 5.4 104
Si 5.3 106 6.0 1011 8.7 104
Ti 5.0 106 2.4 109 2.0 103
Mn 7.8 106 7.7 109 1.0 103
Cu 1.7 105 2.3 109 7.3 103
Gd 5.0 106 1.1 10 5
4.8 101
Pb 7.9 106 4.0 1011
2.0 105
In a more fundamental way, the decrement of the real part of the neutron
refractive index is related to the interaction potential that causes the phase shift.
For most materials, the nuclear interaction potential produces the strongest
phase shift and thus determines the value. For an idealized slab of homo-
genous material with a thickness d, the phase shift due to the nuclear interaction
potential can be expressed by [3]
¼ Nbc ld ; (8:4)
132 F. Pfeiffer
where N is the atom number density and bc the neutron nuclear scattering
length. This is consistent with Eq. (8.3), when is expressed by
l2 Nbc
¼ : (8:5)
2p
Fig. 8.1 Semilogarithmic plot of linear attenuation coefficient versus the real part of the
refractive index for different elements for thermal neutrons with a wavelength l ¼ 1.8Å [4, 5]
(courtesy of K. Lorenz)
The second most important interaction potential causing a phase shift is the
magnetic interaction. In this case can be written as
Bmld
¼ ; (8:6)
h2
2p
where is the neutron dipole moment, B the applied magnetic field, m the
neutron mass, and d the sample thickness.
In addition to the phase shift effect caused by the nuclear and magnetic
interactions, several other interaction potentials exist that cause a measurable
phase shift. Among those are the phase shift effects caused by the Coriolis [6],
the Aharanov–Casher [7], the scalar Aharanov–Bohm [8], the magnetic Joseph-
son [9], the Fizeau [10], and the geometric (Berry) [11] interactions.
8.2.2 Phase Tomography
Fig. 8.2 Projection geometry for tomographic reconstruction. The three different projection
functions correspond to the three main modalities for obtained phase contrast images. G (y 0 , ):
direct measurement on the projected phase shift induced by the object, D (y 0 , ): measurement
of the gradient of the phase shift r, and L (y 0 , ): measurement of a projection image
containing information on the Laplacian of the wave front phase r2
134 F. Pfeiffer
Z 1
Gðy 0 ; Þ ¼ gðx 0 ; y 0 Þdx 0 ; (8:7)
1
where Gðve 0 ; Þ is the Fourier transform of Gðy 0 ; Þ, v 0 the reciprocal space
coordinate corresponding to y 0 ¼ x sin þ y cos , kðve 0 Þ the Fourier represen-
1
tation of the filter function in the FBP, and F T the inverse Fourier trans-
form. For continuously sampled projection functions Gðy 0 ; Þ, the filter is
e 0 Þ jv 0 j. In practice, where the projections are sampled with a
defined as kðv
finite resolution, y 0 , the filter should be truncated at the Nyquist frequency,
1=2y 0 , and becomes [13]
e 0Þ jv 0 j; jv 0 j 1=2y 0
kðv : (8:9)
0; jv 0 j41=2y 0
Fig. 8.3 Backprojection filter functions for (a) simple projection line integrals, (b) gradient
projection integrals, and (c) projection image containing information on the Laplacian of the
object
8 Neutron Phase Imaging 135
Z 1
@gðx 0 ; y 0 Þ 0
Dðy 0 ; Þ ¼ dx : (8:10)
1 @y 0
We note that the z-axis, which was omitted for reasons of simplicity for the
treatment above, was reintroduced here. This is because projection images
containing the Laplacian of the phase shift cannot simply be factored into
single tomographic slices. The filter applied to the individual projections before
the back projection intrinsically becomes two-dimensional (2D) in this case.
Analogous to the derivation of the correct filter function, using the Fourier
derivative theorem above, we find that the tomographic reconstruction of the
object is given by
Z p
gðx; y; zÞ ¼ e 0 ; w; Þ e
F T 1 ½Lðv lðv 0 ; wÞd ; (8:14)
0
136 F. Pfeiffer
e 0
1
4p2
v 02jvþwj 2 jv 0 j 1=2y 0
lðv ; wÞ ; (8:15)
0; jv 0 j41=2y 0
where a0 and a1 are the average intensity and fringe contrast, respectively. In
practice the interferometer has some residual strain that normally generates a
built-in fringe pattern, and the influence is involved in inst.. Often this effect is
not negligible, making the interference pattern too complicated for grasping
structural information by eye. One way to separate the effect of the sample from
the characteristics of the interferometer is using a ‘‘phase-stepping’’approach.
This phase-stepping method processes multiple interference patterns observed
by varying the phase difference between the two arms, for instance by using a
tunable phase shifter. When interference patterns
are obtained with M 3, both with and without the sample in place, (y,z) and
inst. can uniquely be processed out from the data set. For phase tomography,
this measurement is repeated while rotating the sample.
Figure 8.5 shows an example [19] where an aluminum screw has been
imaged in an aluminum block of slightly different composition by using
neutron crystal interferometry. From the image without sample (Fig. 8.5a),
we can deduce that the phase of the empty interferometer is reasonably con-
stant over the total projection area of about 10 mm width and 25 mm height.
With sample (Fig. 8.5b), the modulation of the phase induced by the sample is
clearly recognizable.
By essentially taking the difference and correcting for phase-wrapping
errors, one obtains a projected phase map of the specimen, as shown in
Fig. 8.6a. Because the measured phase projection images are of a quanti-
tative nature, phase tomography reconstructions can be performed using
Eq. (8.8). One such reconstruction is shown in Fig. 8.6b, in the form of a
3D rendering of the object (see [19] for more details). Other examples of
the application of neutron interferometry for phase imaging are reported
in [20, 21, 22, 23, 24].
138 F. Pfeiffer
Fig. 8.5 Raw images behind a neutron crystal interferometer of an aluminum screw in an
aluminum block. (a) Reference image without sample and (b) interferometer image with
sample in the beam path (images from [19])
a b
Given the reflection curve R(g) of the analyzer and the complex transmis-
pffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
sion function Iðy; zÞ=I0 ðy; zÞ exp [i(y,z)] of the specimen, the image
I(y,z,g0) of the beam reflected by the analyzer is given by
140 F. Pfeiffer
pffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
Iðy; z; g0 Þ ¼ Iðy; zÞ=I0 ðy; zÞR½g0 aðy; zÞ ; (8:19)
where g0 is the incident angle without the sample. Similar to the phase-stepping
procedure in single-crystal interferometry, the values a (y,z) can be determined
more precisely by measuring several points along the crystal rocking curve and
subsequent data processing. Tomographic reconstruction of an object from a
data set containing measured projections of g (y,z) is also possible. Since these
projections basically represent gradient projections of the object phase shift, CT
reconstruction can be achieved by applying Eq. (8.11).
An example of neutron phase tomography using such a crystal analyzer-
based method is shown in Fig. 8.8 (from [26]). We see tomographic images of an
aluminum test sample imaged with conventional attenuation-based neutron
tomography (Fig. 8.8a) and in phase tomography (Fig. 8.8b). Because alumi-
num exhibits only a very small neutron capture cross-section, it produces
negligible contrast in absorption. In the phase reconstruction, however, the
internal details are clearly visible. Moreover, it has recently been shown that
these reconstructions are in quantitative agreement with the expected literature
values for the neutron refractive index of materials (see [27] for more details).
Other examples and possible future applications of analyzer-based neutron
phase imaging are reported in [28, 29, 30, 31].
Fig. 8.8 Reconstructed tomographic slices obtained for an aluminum test sample with holes.
(a) Conventional attenuation-based reconstruction; (b) phase contrast-based reconstruction
(images from [26]). The size of the sample is approximately 25 25 mm2
8 Neutron Phase Imaging 141
Fig. 8.9 (a,b) Principle setup of a grating-based neutron shearing interferometer. (c) The
phase grating (G1) forms a periodic interference pattern in the plane of the analyzer grating. A
phase object in the incident beam will cause a slight refraction, which results in changes of the
locally transmitted intensity through the analyzer
The source grating (G0), an absorbing mask with transmitting slits, typically
placed close to the neutron beam exit port, creates an array of individually
coherent but mutually incoherent sources. Each individual line source provides
enough spatial coherence for the image formation process [32, 33, 34]. As the
source mask G0 can contain a large number of individual lines, each creating a
virtual source, efficient use can be made of typical neutron source sizes of more
than a few square centimeters. To ensure that each of the line sources contributes
constructively to the image formation process, the geometry of the setup should
satisfy the condition p0 = p2 l/d, where p0 (p2) is the period of G0 (G2), l the
distance between G0 and G1, and d the distance between G1 and G2 (Fig. 8.9b).
142 F. Pfeiffer
The DPC image formation process achieved by the two gratings G1 and G2
is similar to Schlieren imaging [25] or to a Shack–Hartmann wave front sensor
[35]. The second grating (G1) acts as a beam splitter and divides the incoming
beam essentially into the two first diffraction orders. Since the wavelength l of
the illuminating neutron radiation ( 1010 m) is much smaller than the grating
pitch ( 106 m), the angle between the two diffracted beams is so small that
they overlap almost completely.
Downstream of the beam-splitter grating, the diffracted beams interfere and
form linear periodic fringe patterns in planes perpendicular to the optical axis
[33, 36]. For a phase grating with a phase shift of p, the periodicity p2 of the
fringe pattern equals half the period of G1. Neither the period nor the lateral
position of these fringes depends on the wavelength of the radiation used.
Perturbations of the incident wave front, such as those induced by refraction
on a phase object in the beam, lead to local displacement of the fringes
(Fig. 8.9c).
The fundamental idea of the method presented here is to detect the local
positions of the fringes and determine from these the phase shift induced by the
object. However, since the spacing of the interference fringes does not exceed a
few microns, an area detector placed in the detection plane will generally not
have sufficient resolution to resolve the fringes, let alone the exact position of
their maxima. Therefore, a grating G2 with absorbing lines and the same
periodicity and orientation as the fringes is placed in the detection plane,
immediately in front of the detector. This analyzer grating acts as a transmis-
sion mask for the detector and transforms local fringe positions into signal
intensity variations. The detected signal profile thus contains quantitative
information about the phase gradient of the object.
To separate this phase information from other contributions to the signal
such as attenuation in the sample, inhomogeneous illuminations, or imperfec-
tions of the gratings, the phase-stepping approach used in other interferometry
methods [3] was adapted to this setup. When one of the gratings is scanned
along the transverse direction yg (Fig. 8.9c), the intensity signal I (y,z) in each
pixel (y,z) in the detector plane oscillates as a function of yg (Fig. 8.10e). The
interferometer phases #ðy; zÞ of the intensity oscillations in each pixel are
related to the wave front phase profile (y,z), the neutron wavelength l, and
the period p2 of the absorption grating by [25]
ld @
#¼ : (8:20)
p2 @y
Fig. 8.10 Principle of phase stepping. (a–d) Raw image data of a square molybdenum (Mo)
metal rod, taken at different positions yg ¼ y1, . . ., y4 of the analyzer grating G2. (e) Intensity
oscillation in two detector pixels as a function of yg (images from [33])
were obtained at the ICON beam line of the SINQ (Paul Scherrer Institut,
Switzerland). Due to their specific neutron capture cross-sections and incoher-
ent scattering cross-sections, differences in the transmitted intensity through the
rods are observed in Figure 8.11a. Because the vector component of the phase
gradient in the direction perpendicular to the grating lines is constant, but of
opposite signs for both sides of the rods, the rods appear black and white in the
DPC image. It is interesting to note that titanium has a negative neutron
scattering length density [5], and consequently, a negative phase shift is mea-
sured in the material (see Fig. 8.11c). Furthermore, we note that particularly for
weakly absorbing materials, like magnesium and lead (Fig. 8.11a), the DPC
(Fig. 8.11b) and the integrated phase shift (Fig. 8.11c) signal yield a higher
contrast in the image.
As the phase contrast obtained with a grating interferometer can be
described quantitatively, it can further be used for neutron phase tomography.
Figure 8.12 shows one exemplary processed DPC image (a), a 3D rendering of
the reconstructed coherent scattering length density distribution (b), and 2 (out
of 512) reconstructed tomographic slices (c,d) of a sample comprising titanium
144 F. Pfeiffer
a b
10 mm Ti Pb Mo Mg
Fig. 8.11 Linear contour plots and section profiles of processed data for a test object
comprising titanium (Ti), lead (Pb), molybdenum (Mo), and magnesium (Mg) metal rods
with a square profile. (a) Conventional neutron transmission image. (b) Differential phase
contrast image d /d y. (c) Integrated phase shift (y,z) retrieved from d/d y by integration.
(d,e,f) Section profiles through the corresponding image data (images from [33])
Fig. 8.12 Projection images and tomographic reconstruction for a sample comprising lead
(black) and titanium (white) rods with diameters of 6.35 mm. (a) Differential phase contrast
projection image. (b) 3D rendering of the reconstructed coherent scattering length
distribution in the sample. (c,d) 2 (out of 512) reconstructed tomographic slices (images
from [33])
8 Neutron Phase Imaging 145
In the classical sense, the distance between an object and a neutron image
detection plane should be as short as possible to avoid blurring. However,
under (partially) coherent illumination, Fresnel or Fraunhofer diffraction is
observed at a distance away from an object even with neutrons. Then, an edge-
enhanced contrast is observed without optical elements even for a pure phase
object, as illustrated in Fig. 8.13. The choice of the propagation distance has a
direct influence on the characteristics of the observed fringe contrast produced
by the edges of the object.
The effect can be described by considering the propagation of a wave field
along a certain direction in space (Fig. 8.13). It can be written under the paraxial
approximation by [38, 39]
lx 2
Iðx; y; zÞ ¼ 1 þ r ð0; y; zÞ ; (8:21)
2p
where we have assumed a propagation along the x direction and r denotes the
2D gradient operator acting in the (y,z) plane. This expression can be derived
from the so-called transport of intensity equation [38, 39], assuming a nonab-
sorbing, weak-phase object. The character of the Laplacian in Eq. (8.21)
suggests the generation of the contrast particularly outlining surfaces and
structural boundaries, where the refractive index changes abruptly.
In propagation-based phase contrast radiography, the distance between the
sample and the image detector is large so that the weak lateral deflections due to
Fig. 8.13 Illustration showing the generation of an edge-enhanced contrast of a phase object
due to propagation
146 F. Pfeiffer
Fig. 8.14 A series of neutron radiographs showing the development of the phase contrast of a
lead sinker mounted on an aluminum screw. The images, counterclockwise from top left to
top right (a–e), correspond to propagation distances of approximately 0.02, 0.22, 0.45, 0.90,
and 1.8 m, respectively. The visibility of the edge detail within the sinker, screw, and shavings
trapped in the axial hole are greatly enhanced (images from [40])
8 Neutron Phase Imaging 147
To generate contrast with phase information, a wave front with a defined phase is
required. This means that a certain degree of coherence is required in the neutron
beam. The coherence can be evaluated perpendicular to the optical axis and along
the optical axis. The first case refers to the so-called spatial or transverse coher-
ence and the second to the temporal or longitudinal coherence [49, 50].
Let us first consider the coherence properties of a source in a plane transverse
to the propagation direction of the radiation. A good measure for the degree of
coherence is the interference contrast recorded in the far-field diffraction pattern
of an object in the beam. Let us take as an object Young’s double slit arrangement
[25]. Illuminating the double slit with a monochromatic point source results in the
far field in a set of interference fringes with fully developed maxima and zero
intensity in the minima, corresponding to a fringe visibility of 100%.
In practice, however, the source is not point-like, but has some lateral
extension. It is easy to see that this reduces the visibility of the interference
fringes. Although the fringe patterns resulting from different points of the
extended source are identical, they emerge at slightly shifted diffraction angles.
Their superposition, therefore, causes the observed pattern to be somewhat
washed out (i.e., the fringe visibility to be reduced). The effect of the extension
of the source can be readily quantified. Let us consider two narrow slits at
distance d, illuminated by a uniform line source of height w (Fig. 8.15).
The interference fringes produced by an infinitesimally small emitting source
element located on the optical axis appear at diffraction angles a = ml /d, with
m = 0, 1, 2, . . . for the maxima and m ¼ 12, 32 ; 52 for the minima.
Here, the small-angle approximation has been made. The fringes produced by
an emitting element at the border of the source are found at the shifted angles
ml /d + w/2R, with R the distance between the source and the slits. It is easy to
see that the maxima from a border element of the source coincide with the
minima from the central element at a slit separation d = l R/w. This distance is
defined as the transverse coherence length
lR
t ¼ : (8:22)
w
Fig. 8.15 Diffraction patterns from two narrow slits at distance d, originating from the central
part of the source (solid curve) and from the edge of the source at height w/2 (dashed curve).
The slit distance d is such that the two patterns are in antiphase. This occurs for d¼l R/w.
Angles and distances are not to scale (adapted from [50])
The degree of coherence of the radiation along its propagation direction (i.e.,
its longitudinal [or temporal] coherence) enters the imaging process in a different
way. Let us consider two wave fronts, one at wavelength l and the other at a
slightly different wavelength l+l, which simultaneously depart from a
single-point. Let us assume that after some distance l both wave fronts are in
antiphase (Fig. 8.16).
Fig. 8.16 Propagation of two waves with wavelengths l and l+l. The longitudinal coher-
ence length l is defined as the distance over which the phase difference of the two waves has
become p (adapted from [50])
8 Neutron Phase Imaging 149
If the first wave has made N oscillations over that distance, the second wave
must have made N 12 oscillations. One therefore has Nl ¼ ðN 12Þ (l+l).
Solving for N and substituting in l ¼ Nl we find for this distance
1 l2
l ’ : (8:23)
2 l
The precise prefactor is dependent on the spectral power density of the source;
for a Lorentzian spectrum one has l ’ ð2=pÞ l2 =l [51].
In practice the intrinsic coherence properties of a neutron spallation or
reactor source are relatively low. This is because such sources typically are a
few centimeters in size and emit a broad polychromatic spectrum. But the
coherence can be improved by adapting the experimental setup accordingly.
As can easily be seen from Eq. (8.22), the transverse coherence can be
increased by inserting a small aperture into the neutron beam at a large distance
upstream from the sample. For example, if we insert an aperture with a 1 mm
diameter at a distance of 10 m upstream from the sample, we will obtain a
transverse coherence length on the order of a few microns for thermal neutrons
with a typical wavelength of a few angstroms. The propagation-based phase
contrast imaging method, for example, typically requires this amount of trans-
verse coherence (see also Section 8.3.3).
As the longitudinal coherence depends on the wavelength spread of the neutron
energy spectrum [Eq. (8.23)], it can be improved by reducing the bandwidth of the
wavelength distribution. This can be achieved, for example, by using a double-crystal
silicon monochromator. With a typical bandwidth of l=l 104 , a longitudinal
coherence length of a few microns can be achieved for thermal neutrons with a
typical wavelength of a few angstroms. Such a beam preparation is necessary, for
example, in the case of the crystal analyzer-based method (see Section 8.3.2.1).
Table 8.2 contains a comparison of the coherence requirements for the phase
imaging methods described above.
References
1. A. F. Kracklauer, On the theory of quanta (a translation of the thesis of de Broglie,
‘‘Recherches sur la théorie des quanta’’), Ann. de Phys. 10, t. III (1925).
2. A. G. Klein, S. A. Werner, Rep. Prog. Phys. 46, 259 (1983).
3. H. Rauch, S. A. Werner, Neutron Interferometry, Oxford University Press, Oxford, (2000).
4. http://www.ncnr.nist.gov/resources/sldcalc.html.
5. V. F. Sears, Neutron News, 3, 26 (1992).
6. M. Zawisky, U. Bonse, F. Dubus, et al., Europhys. Lett. 68, 337–343 (2004).
7. A. Cimmino, G. I. Opat, A. G. Klein, H. Kaiser, Phys. Rev. Lett. 68, 380 (1989).
8. B. E. Allman, A. Cimmino, A. G. Klein, Phys. Rev. Lett. 68, 2409 (1992).
9. G. Badurek, H. Rauch, D. Tuppinger, Phys. Rev. A 34, 2600 (1986).
10. A. G. Klein, G. I. Opat, A. Cimmino, et al., Phys. Rev. Lett. 46, 1551 (1981).
11. A. G. Wagh, V. C. Rakhecha, J. Summhammer et al., Phys. Rev. Lett. 78, 755 (1997).
12. J. Radon, Ber. Verh. Sächs. Akad. Wiss. Leipzig, Math–Nat. 69, 262 (1917).
13. G.T. Herman, Image reconstruction from projections, Academic Press, New York (1980).
14. A. C. Kak M. Slaney, Principles of Computerized Tomography IEEE Press, New York,
(1987).
15. G. W. Faris, R. L. Byer, Appl. Opt., 27, 5202–5212 (1988).
16. U. Bonse, M. Hart, Appl. Phys. Lett. 6, 155 (1965).
17. H. Rauch, W. Treimer, U. Bonse, Phys. Lett. A 47, 369–371 (1974).
18. W. Treimer, Krist. Tech. Crystal Res. Technol. 13, 1105–1116 (1978).
19. F. Dubus, U. Bonse, M. Zawisky, et al., IEEE Trans. Nucl. Sci. 52, 364–370 (2005).
20. G. Badurek, M. Hochhold, H. Leeb, Physica B 234, 1171–1173 (1997).
21. G. Badurek, R. J. Buchelt, H. Leeb, et al., Physica B 335, 114–118 (2003).
22. E. Jericha, R. Szeywerth, H. Leeb, et al., Physica B 397, 159–161 (2007).
23. J. Rehacek, Z. Hradil, M. Zawisky, et al., Phys. Rev. A, 71, 023608 (2005).
24. S. A. Werner, J. L. Staudenmann, R. Colella, Phys. Rev. Lett. 35, 1053 (1979).
25. M. Born, E. Wolf, Principles of Optics, Pergamon Press, Oxford (1980).
26. W. Treimer, M. Strobl, A. Hilger, et al., Appl. Phys. Lett. 83, 398–400 (2003).
27. M. Strobl, K. Staack, W. Treimer, et al., Meas. Sci. Technol. 19, 034020 (2008).
28. M. Strobl, W. Treimer, A. Hilger, Nucl. Instrum. METHODS Phys. Res. B 222, 3–4
(2004).
29. M. Strobl, W. Treimer, A. Hilger, Phys. B, 385–386, 1209–1212 (2006).
30. M. Strobl, W. Treimer, P. Walter, et al., Appl. Phys. Lett. 91, 254104 (2007).
31. W. Treimer, M. Strobl, A. Hilger, et al., IEEE Trans. Nucl. Sci., 52, 386–388 (2005).
32. F. Pfeiffer, O. Bunk, C. Schulze–Briese, et al., Phys. Rev. Lett. 94, 164801 (2005).
33. F. Pfeiffer, C. Grunzweig, O. Bunk, et al., Phys. Rev. Lett., 96, 215505 (2006).
34. F. Pfeiffer, T. Weitkamp, O. Bunk, et al., Nat. Phys. 2, 258–261 (2006).
35. J. Hartmann, Z. Instrumentenkd. 8, 2 (1990).
36. T. Weitkamp, A. Diaz , C. David, et al., OPT. Express, 13, 6296–6304 (2005).
37. F. Pfeiffer, C. Kottler, O. Bunk, et al., Phys. Rev. Lett. 98, 108105 (2007).
38. J. M. Cowley, Diffraction Physics, Elsevier, Amsterdam (1995).
39. M. R. Teague, J. Opt. Soc. Am. 73, 1434 (1983).
40. B. E. Allman, K. A. Nugent, Physica B, 385–86, 1395–1401 (2006).
41. B. E. Allman, P. J. McMahon, K. A. Nugent, et al., Nature 408, 158–159 (2000).
42. J. W. Chen, H. Y. Gao, H. F. Zhu, et al., Acta Phys. Sin. 54, 1132–1135 (2005).
43. F. Fiori, A. Hilger, N. Kardjilov, et al., Meas. Sci. Technol. 17, 2479–2484 (2006).
44. D.L. Jacobson, B.E. Allman, J. McMahon, et al., Appl. Radiat. Isot. 61, 547–550 (2004).
45. E. Lehmann, K. Lorenz, E. Steichele, et al., Nucl. Instrum. Methods Phys. Res. A 542,
95–99 (2005).
46. P. J. McMahon, B. E. Allman, D. L. Jacobson, et al., Phys. Rev. Lett. 91, 145502 (2003).
47. B. Schillinger, E. Calzada, K. Lorenz, Mat. Trans., Proc. 112, 61–71 (2006).
8 Neutron Phase Imaging 151
48. S. C. Mayo, T. J. Davis, T. E. Gureyev, et al. Opt. Express 11, 2289–2302 (2003).
49. J. Als-Nielsen, D. McMorrow, Elements of Modern X–Ray Physics, John Wiley & Sons
Ltd, New York (2001).
50. F. van der Veen, F. Pfeiffer, J. Phys. Condens. Matter 16, 5003–5030 (2004).
51. J. W. Goodman, Statistical Optics, Wiley, New York, (1985).
Chapter 9
Thermal Neutron Holography
Abstract X-ray and neutron diffraction techniques have for almost a century
produced results that provide important insights into materials of interest to
a wide range of scientific and technological disciplines. However, traditional
diffraction techniques have their limits, and these limits are best exemplified
by the fact that certain important materials (e.g., integral proteins) are
difficult if not impossible to crystallize — diffraction techniques usually
require high-quality single crystals. Recently developed atomic resolution
X-ray and neutron holography techniques offer the promise to resolve the
structures of difficult-to-crystallize materials to atomic resolution. This chap-
ter will discuss the latest developments in neutron holography and the
challenges that must be overcome to make the technique a viable tool.
9.1 Introduction
Holography can be traced to Bragg’s [1] X-ray work and Gabor’s [2] electron
interference microscope. Although the technique lay dormant for a number of
decades, there has been increasing interest in its development as implied by the
increasing number of publications on atomic-resolution holography using
either electrons [3–7] or hard X rays [8–13]. Moreover, the feasibility of
atomic-resolution thermal neutron holography, either the ‘‘inside source’’ or
‘‘inside detector’’ type, was experimentally demonstrated [14–16], and a theory
was developed showing that the diffraction pattern of plane waves incident on
a sample with a uniformly random distribution of incoherent scatterers
J. Katsaras (*)
National Research Council, Chalk River Laboratories, Chalk River, ON K0J 1J0,
Canada
e-mail: [email protected]
(e.g., 1H in a polymeric material) per unit cell is the same as that from a sample
with a single incoherent scatterer per unit cell [17]. There is interest in the
possibility of using neutron holography to reconstruct, to atomic resolution,
the three– dimensional (3D) structure of materials rich in hydrogen that are
presently difficult to crystallize (e.g., membrane-associated proteins).
Dennis Gabor (1900–1979) developed holography as a technique to improve
the resolution of the electron microscope while working for the British Thomson-
Houston Company (Rugby, England). A patent was filed in 1947 [18], but the
general field did not advance until the invention of the laser [19, 20] (the process
of stimulated emission was first theorized by Einstein in 1917) [21]. The first 3D
holograms were made by Denisyuk [22] in 1962 (reflection hologram) and later
that year by Leith and Upatnieks at the University of Michigan (transmission
hologram) [23]. Since then, optical holography has been extensively used in
optical devices (e.g., holographic lens for heads-up displays), security (e.g., credit
cards), scanners, computer memory storage, medicine, and a variety of other
applications. For example, holographic interferometry is a precise technique for
measuring changes in the dimensions of an object, and digital cameras with
holographic capabilities are able to focus in poor contrast conditions.
For many decades, X-ray diffraction has been relied upon to determine the
3D structure of a number of crystalline materials, the prerequisite being a very
high-quality single crystal. However, even when the requisite sample was made
available, the technique suffered from the so-called ‘‘phase problem’’ — that is,
the phase information was not recorded. As a result, a number of isomorphous
crystals had to be grown to allow for the direct solution of molecular structure —
at increased cost and time. It should also be pointed out that, in some cases,
isomorphous derivatives were impossible to produce. Recently, however, syn-
chrotron sources with their tunable X-ray beams have revolutionized crystal-
lography by allowing the development of multiple-wavelength anomalous
dispersion (MAD), a technique used to obtain reliable phase information.
Now a single crystal serves the purpose of several different metal-containing
crystals and has the added advantage of faster data collection times. Despite
these advances, the need for high-quality crystals remains.
Both holography and crystallographic techniques, such as MAD, solve the
phase problem in the case of 3D materials. Holographic techniques, however,
have the added advantage that only orientational order is required; that is,
translational symmetry is not necessary. In addition to imaging crystalline
materials, holography is also suitable for imaging noncrystalline materials (e.g.,
nematic liquid crystals) or poorly crystallized membrane-associated proteins.
Although the idea of atomic-resolution holography has its roots in Bragg’s
[1] X-ray work and Gabor’s [2] electron interference microscope, atomic-
resolution holography did not evolve until Szöke [24] pointed out that photo-
excited atoms within a sample emit highly coherent outgoing electrons or
fluorescent X rays. The interference between the unperturbed wave from the
photoexcited atom (reference wave) and waves scattered from neighboring
atoms (object waves) thus permitted the realization of atomic-resolution
9 Thermal Neutron Holography 155
ik r −r
b(r1 )
i kin ⋅r1 e 2 1
e
r2 − r1
ik R−r
× b(r2 )
e 2
R − r2
r2 = ΨII (R )
ΨI (r2 ) =
ik r −r
b(r1 )
i kin ⋅r1 e 2 1
e
r2 − r1
r1 ΨI (R ) =
ik R−r
b(r1 )
i kin ⋅r1 e 1
e
R − r1
Fig. 9.1 (A) Schematic depiction of second-order plane-wave scattering. The incident neutron
plane wave inc interacts with an atom at r1, producing a primary spherical wave (S wave) I.
This primary S wave interacts with a second atom at r2, producing a secondary S wave II.
The interference between the primary and secondary S waves leads to an intensity modulation
(hologram) detected at R. The detected intensity (evaluated for incident and detected wave
vectors, kin and kout) is given as I ¼ I I þ 2Re I II þ II II . Figure adapted from Sur
et al. [17]. See Ref. [17] or Sect. 9.5 of this paper for details
can either pass through the sample unperturbed (reference wave) or be scattered
by the surrounding atoms (object wave). The interference between the reference
and object beams produces the hologram.
In the inside detector method developed by Gog et al. [9, 27], the inside source
holography beam paths can be reversed and the source and detector exchanged. In
other words, an incident plane wave can reach the photoexcitable atom in a crystal
unperturbed (reference wave) or after it has scattered off other atoms in the sample
(object wave). The interference between these two beams dictates the fluorescence
intensity given off by the photoexcitable atom, which is now the detector. When the
incident beam is scanned, the phase relation between the reference and object
waves leads to oscillations of the fluorescent intensity [28] and forms a hologram.
9.3 Neutrons
In 1932, James Chadwick [29, 30] discovered the neutron — first postulated by
Rutherford [31]—a neutral, subatomic elementary particle found in all atomic
nuclei except hydrogen nuclei. A neutron’s mass is similar to that of a proton,
but a neutron possesses a nuclear spin of ½ and a magnetic moment. Neutrons
are stable only when bound in an atomic nucleus; an unstable free neutron has a
mean lifetime of approximately 900 s and decays into a proton, an electron, and
an antineutrino [32].
Even though the neutron’s interaction with atomic nuclei is weak, the
scattering power (cross section) of an atom is not related to its atomic number.
Therefore, neighboring elements in the periodic table can have substantially
different scattering cross sections. It is important to note that the interaction of
a neutron with the nucleus of an atom allows for an element’s isotopes to be
differentiated. An example is the isotopic substitution of 1H (hydrogen) for 2H
(deuterium), a technique commonly used in the study of polymeric and bio-
logically relevant materials. Hydrogen has a negative coherent scattering length
(bcoh = –3.74 10–15 m), lending it ‘‘contrast’’ when surrounded by other,
positive-scattering-length atoms. For biological samples, intrinsically rich in
hydrogen, judicious substitution of 2H (bcoh = +6.67 10–15 m) for 1H
provides a powerful method for selectively tuning the ‘‘contrast’’ of a given
macromolecule. One can therefore accentuate or nullify the scattering from
particular parts of a macromolecular complex by selective deuteration [33]. Of
importance to holography is that 1H has a large incoherent cross section
for thermal neutrons (i = 80 barns), producing S waves similar to the de-
excitation of a photoexcitable atom in X-ray holography. Unlike forms of
electromagnetic radiation that interact primarily with electrons, neutrons inter-
act directly with nuclei and are, to first order, equally capable of imaging light
atoms and heavy atoms. Consequently, atomic-resolution holographic images
of materials comprising entirely light atoms can be obtained, a great advantage
in the study of polymers and biologically relevant materials.
9 Thermal Neutron Holography 157
lines were first observed by Kikuchi [25] in electron scattering studies using mica
plates of varying thicknesses. Thus such lines are generally referred to as
Kikuchi lines. In the present case, thermal neutron S waves are generated by
incoherent elastic scattering from hydrogen. The fundamental scattering pro-
cess that gives rise to the characteristic lines — S wave interaction with the
crystal lattice — is the same as for Kossel and Kikuchi lines; therefore, they are
collectively referred to as K lines.
The K line intensity pattern depends on the relative position of the periodic
scatterers with respect to the S wave source (i.e., the crystallographic phase of
the scattering structure function). Thus K lines provide the signal for direct 3D,
atomic-resolution imaging techniques (holography) for single crystals via the
inside source or inside detector concept.
Atomic nuclei are, to a very good approximation, point or pure S wave
scatterers of thermal neutron waves. Hydrogen atoms have a large bound, spin-
incoherent cross section, i = 80 barns, for unpolarized thermal neutrons. An
unpolarized, parallel, thermal neutron beam interacting with a single crystal
containing hydrogen atoms should therefore give rise to both Bragg reflections,
from the incident plane-wave neutron beam, and K lines, from the S waves
produced by the incoherent scattering from the 1H atoms.
9.5 Theory
To gain an understanding of the process, consider a single-point S wave source
of amplitude a at the origin surrounded by a distribution of weak point or
purely S wave scatterers described by a coherent scattering length density b(r).
Then, in the Born approximation, the unperturbed or reference wave is
aeijkjjrj
ðrefÞ ðrÞ ¼ ; (9:1)
j rj
The total wave amplitude, tot for a detector at R is given by ref (R) + obj
(R). The detected intensity, I1S (R), is then *tot tot. Using the far field
diffraction condition (i.e., R >> ^r) and defining the outgoing wave vector in
the detector direction as kout ¼ k R, the detected intensity is then given as
a a
I1S ffi ½1 þ 2Re½ðkout Þ þ jðkout Þj2 ; (9:3)
R2
9 Thermal Neutron Holography 159
In the case of inside source holography, the scattering length distribution b(r)
is reconstructed by a Fourier inversion of the observed modulation intensity [14].
To gain an understanding of K lines, consider this expression [Eq. (9.4)]. In
reciprocal space, the modulation function (aside from constants) is given by
Z
BðqÞ
ðkout Þ ¼ 2
dq ; (9:5)
jqj 2kout q
For every pair of discrete points t hkl located within a sphere of radius 2k,
the zeros in the denominator of Eq. (9.6) cause the modulation to go to infinity
in kout directions whose locus is a cone with axis along t hkl and full opening
angle 2, satisfying the relation 2k cos = hkl. Viewed on the surface of a
hemisphere of radius kout, the intersections of the cones with the hemisphere are
K line rings centered on t hkl. These K line rings can readily be calculated for any
hkl relative to the crystal coordinate system of a known structure. For example,
a selection of K lines expected for the structure of KDP in a particular sample
orientation is shown in Fig. 9.2A on the surface of a hemisphere kout/| kout| with
| kout| = 4.8 Å–1 [15]. The incident plane wave, kin, will also produce Bragg
peaks, which will appear as discrete points on the surface of the hemisphere,
when condition kout – kin = t hkl is satisfied (Fig. 9.2A).
In a high-resolution experiment, K lines are readily identified by their character-
istic fine structure [27]. However, K line fine structure is affected by factors such as
sample size and shape;
dynamics or higher-order partial wave scattering in both the coherent and
incoherent scatterers, leading to a broadening of the signal and a nonuni-
form reference wave intensity; and
wavelength spread and angular resolution.
All of these effects conspire to mute the sharp modulations characteristic of
K lines, but they still produce modulations along the lines (Fig. 9.2B).
160 B. Sur et al.
Fig. 9.2 (A) Calculated K lines for the various KDP lattice planes for l ¼ 1.30 Å neutrons. In
addition to the K lines, the locations of Bragg reflections are shown as white spots. The
orientation of the crystal with respect to the sphere-of-scattering (kin || kz) is depicted by the
crystallographic axes a, b, and c. (B) KDP holographic data corrected for background,
Debye-Waller factor, and sample shape. K lines appear as bands of scattering
corresponding to the white lines in (A). Figure adapted from Sur et al. [15]
According to Eqs. (9.3) and (9.4), the experimentally observed hologram mod-
ulation function (k) produced by the interaction of a weak S wave coherent
scattering length distribution b(r) with either a single incoherent point scatterer
or with a uniform distribution of incoherent scatterers is given, aside from
constants, by
9 Thermal Neutron Holography 161
Z
bðrÞ ½iðkrkrÞ
ðkÞ Re e dr þ O b2 : (9:7)
r
all space
The conditions for pure S wave weak scattering are particularly well satisfied
for unpolarized thermal neutron scattering from nuclei. Additionally, for most
nuclei, the thermal neutron coherent scattering lengths are almost purely real
and have the same sign. The real part of the coherent scattering length distribu-
tion can be directly reconstructed from the measured hologram modulation
function using the following relations.
Z
breal ðrÞ þ breal ðrÞ
b~even ðrÞ ¼ cosðkrÞ / ðkÞ cosðk rÞdk : (9:8)
kr
constantjkj
Z
breal ðrÞ breal ðrÞ
b~odd ðrÞ ¼ sinðkrÞ / ðkÞ sinðk rÞdk : (9:9)
kr
constantjkj
Strictly speaking, for a single wavelength or constant k, the even and odd
parts are reconstructed at nonoverlapping points in space (i.e., the zeros of the
even part reconstruct at turning points of the odd part and vice versa). There-
fore, it is possible to eliminate the conjugate image breal (r) from the recon-
struction of a single hologram data set by a suitable combination of b~odd ðrÞ and
b~even ðrÞ. For instance, b~odd ðrÞ sin(kr) and b~even ðrÞ cos(kr) can be summed with a
moving box average of dimensions l/2 to give b(r). A consequence of this
procedure is that the resolution of the reconstructed image is broadened by
l/2. Alternatively, the quality of the reconstructed image can be improved by
combining holograms obtained at several different wavelengths.
For a single-point incoherent scatterer per unit cell, the above formulation
will reconstruct the coherent scattering length with the incoherent scatterer
located at the origin. For a uniformly random distribution of point incoherent
scatterers, the reconstruction origin will be the ‘‘center of illumination,’’ which
by definition occupies exactly the same volume as the sample.
The first inside source neutron holography experiments were carried out at the
N5 triple-axis spectrometer (Fig. 9.3A) located at the NRU reactor at Chalk
River Laboratories, Ontario. Those experiments used 1.3-Å neutrons with a
l/l 1.5% using the (113) reflection from a germanium (Ge) single-crystal
monochromator. The sample-to-detector angular resolution was limited by
distance collimation, defined by a combination of an aperture in neutron-
absorbing cadmium masks and the size of the sample. Sample orientation ()
162 B. Sur et al.
Fig. 9.3 (A) The N5 triple-axis spectrometer on which the first neutron holography
experiments were carried out, located at the National Research Universal reactor, Chalk
River Laboratories. (B) Plan view photograph and schematic of the neutron holography
experimental setup at the N5 triple-axis spectrometer. The incident 1.3-Å neutron beam
(originating from the bottom of the photograph) was parallel to the axis of rotation, . The
angle between the detector and the axis is denoted by . In order to efficiently obtain the
hemisphere of scattering, the crystal was rotated, in optimal steps of 28/sin , from 0 to 2p for
a given (178 838). The lower value for (178) was dictated as a result of the physical
interference between the scattered beam collimator and the sample rotation spindle. The
scattered neutrons were recorded using a 3He detector. Figure adapted from Sur et al. [14]
9 Thermal Neutron Holography 163
and detector angle () were manipulated in such a way that nearly a hemisphere
of scattering was recorded (Fig. 9.3B). However, as a result of poor scattered
intensity and a single-point detector, collecting good statistical quality data
(total scattering and background) over 1.7p radians in scattering wavevector
(kout) took 10 days (Fig. 9.4A).
The structure of simpsonite contains two layers of oxygen atoms perpendi-
cular to the c axis of the crystal lattice. The first layer, which is closest to the 1H
atom, has an oxygen atom located directly above (positive direction) the 1H
Fig. 9.4 (A) Raw hologram data containing 4,334 28 by 28 pixels plotted on a surface of
constant scattered wave vector magnitude (|kout| = 4.8 Å–1). Because of the experimental
geometry, data collection was limited between 17 and 838 in , and about 13% of the pixels
containing Bragg reflections were excluded from the reconstruction. The remaining data
contained the hologram and -dependent background from the vanadium sample holder
and aluminum wires (e.g., aluminum powder lines at = 32, 37, 54, and 648) used to secure
the simpsonite crystal to the vanadium pedestal. (B) Reconstructed plane located
approximately + 0.9 Å from the hydrogen atom at the origin showing the seven oxygen
atoms. The plane is roughly coplanar to the basal plane of the crystal. The central spot is the
oxygen atom located directly above the origin. The triangles are used to indicate the two
triplets of oxygen atoms located slightly below this central oxygen atom. (C) Reconstructed
plane located approximately –1.4 Å from the hydrogen atom at the origin showing the
positions of the three oxygen atoms. The distortion of the triangles was attributed to
limitations in the quality of the demonstration data. Figure adapted from Sur et al. [14]
164 B. Sur et al.
atom along the direction of the c axis. This oxygen atom is surrounded by two
triangular sublattices of oxygen atoms slightly below the position of the central
oxygen atom. In the second layer, located below (negative direction) the 1H
atom, the oxygen atoms form triangular lattices about the perpendicular pro-
jection of the 1H atom onto this layer.
Figure 9.4B shows a reconstructed plane, which is approximately coplanar to
the basal plane, at a distance of +0.9 Å from the 1H atom, which is at the origin
of the reconstruction. There is a systematic distortion of the triangular lattices
(Fig. 9.4B) owing to the just adequate counting statistics. Nevertheless, all of
the expected oxygen atoms, six in the two triangles and the central atom, are
present in the reconstruction. Figure 9.4C shows the next closest plane of
oxygen atoms found at a distance of –1.4 Å from the origin. From the recon-
struction, it is easily seen that these oxygen atoms form a triangle of nearest
neighbors, albeit distorted.
The example discussed used a large incoherent neutron scatterer (i.e., hydrogen)
to produce nearly S waves, ‘‘imaging’’ its surrounding atoms. By using
the principle of optical reciprocity, the inside detector concept interchanges
the positions of source and detector. In 2002, Cser et al. [16] demonstrated the
feasibility of the inside detector technique to image lead (Pb) nuclei in a single
crystal of Pb doped with the strong neutron absorber cadmium (Cd). In this
case, Cd atoms act as point-like detectors, with a neutron wave reaching these
detector nuclei either directly (reference beam) or after scattering from the Pb
atoms in the crystal (object beam). The capture of neutrons by Cd results in
the emission of radiation. The intensity of the radiation is directly propor-
tional to the neutron intensity experienced by the detector nuclei. When
recorded as a function of sample orientation with respect to the incident
neutron beam, this fluctuating radiation intensity is effectively the
hologram.
To demonstrate inside detector neutron holography, Cser et al. [16] used a
7-mm-diameter spherically shaped single crystal of Pb0.9974Cd0.0026 (the use of
spherical samples greatly simplifies the data analysis). Cadmium atoms are
strong neutron absorbers with a thermal absorption cross section over four
orders of magnitude greater than that of Pb atoms [40]; this accounts for their
ability to act as efficient detectors. The crystal structure of the Pb0.9974Cd0.0026
alloy is the same (face-centered cubic, fcc) as for pure lead [16], with the Cd
atoms randomly distributed. As Cd was used in low concentrations, on average,
every Cd atom was surrounded by 12 nearest-neighbor Pb atoms.
The (220) reflection from a Cu single crystal was used to select 0.8397-Å
neutrons. Prompt radiation from the Cd nuclei was recorded using two
scintillation counters shielded by 20 cm of lead. The sample was rotated
9 Thermal Neutron Holography 165
about its axis () in 38 steps through 3508 and tilted (), also in 38 steps,
through 1208. Figure 9.5A shows the reconstruction from the hologram of the
12 Pb atoms surrounding the detector Cd atom plotted in and coordinates,
and Fig. 9.5B shows the Cd atom and its 12 neighboring Pb atoms plotted on
the surface of a sphere of radius 3.49 Å. The x-axis is the direction of the
incident neutron beam.
Fig. 9.5 (A) Reconstruction showing the positions of the 12 Pb atoms surrounding the detector
Cd atom plotted in and coordinates. (B) The 12 Pb atoms surrounding the Cd detector atom
plotted on the surface of a sphere of radius 3.49 Å. The x-axis denotes the direction of the
incident neutron beam. Data were collected using the D9 diffractometer located at the Institute
Laue-Langevin (Grenoble, France). Figure adapted from Cser et al. [16]
166 B. Sur et al.
The most recent inside detector neutron holography study by Cser et al. [41]
demonstrated for the first time the atomic positions with precision down to
1 picometer. To date, this is the only work based on a neutron holography
experiment that contains information not previously made available by other
experimental techniques.
This brief review has presented the current state of neutron holography.
Neutron holography is a nondestructive, in situ technique capable of atomic-
resolution measurements of biologically relevant materials and appears to be an
excellent option for the solution of poorly crystallized or quasicrystal proteins.
However, an experimental demonstration of ‘‘diffuse source’’ atomic-resolution
holography, utilizing incoherent scattering from the large numbers of hydrogen
atoms in biological materials, is key to future developments. Future neutron
holography experiments will benefit greatly from spallation source instruments
capable of acquiring neutron scattering data with simultaneous large solid angle
coverage and at multiple wavelengths, capabilities that will become available at
the Spallation Neutron Source.
References
1. W.L. Bragg, The X-ray microscope. Nature 149, 470 (1942).
2. D. Gabor, A new microscopic principle. Nature 161, 777–778 (1948).
3. G.R. Harp, D.K. Saldin, and B.P. Tonner, Atomic resolution holography in solids with
localized sources. Phys. Rev. Lett. 65, 1012–1015 (1990).
9 Thermal Neutron Holography 169
28. G. Faigel, and M. Tegze, X-ray holography. Rep. Prog. Phys. 62, 355–393 (1999).
29. J. Chadwick, Possible existence of a neutron. Nature 129, 312 (1932).
30. J. Chadwick, The existence of a neutron. Proc. Roy. Soc. (London) A136, 692–708 (1932).
31. E. Rutherford, Nuclear constitution of atoms. Proc. Roy. Soc. A97, 374–400 (1920).
32. W. Mampe, P. Ageron, C. Bates, J.M. Pendelbury, and A. Steyerl, Neutron lifetime
measured with stored ultracold neutrons. Phys. Rev. Lett. 63, 593–596 (1989).
33. B. Jacrot, The study of biological structures by neutron scattering from solution. Rep
(sp. nov.) from Tabba Tabba, western Australia. J. R. Soc. W. Aust. Prog. Phys. 39,
911–953 (1976).
34. L. Cser, G. Krexner, and Gy. Török, Atomic resolution neutron holography. Europhys.
Lett. 54, 747–752 (2001).
35. H. Bowley, Simpsonite (sp. nov) from Tabba Tabba, western Australia. J. R. Soc. W. Aust.
25, 89–92 (1939).
36. L.E.R. Taylor, X-ray studies of Simpsonite. J. R. Soc. W. Aust. 25, 93–97 (1939).
37. T.S. Ecrit, Ćerný, P., and F.C. Hawthorne, The crystal chemistry of Simpsonite. Can.
Mineral. 30, 663–671 (1992).
38. G.L. Clark, and W. Duane, A new method of using x-rays in crystal analysis. Proc. Natl.
Acad. Sci. U.S.A. 8, 90–96 (1922).
39. M. v. Laue, Die fluoreszensröntgenstrahlung von einkristallen. Ann. Physik 23, 705–746
(1935).
40. V.F. Sears Neutron scattering lengths and cross sections. Neutron News 3, 26–37 (1992).
41. L. Cser, G. Krexner, Markó, M., I. Sharkov, and Gy. Török, Direct observatotion of
local distortion of a crystal lattice at picometer accuracy using atomic resolution neutron
holography. Phys. Rev. Lett. 97, 255501(1)–255501(4) (2006).
42. A. Goffeau, Life with 482 genes. Science 270, 445–446 (1995).
43. J. Deisenhofer, O. Epp, K. Miki, R. Huber, and H. Michel, X-ray structure analysis of a
membrane protein complex: electron density map at 3 Å resolution and a model of the
chromophores of the photosynthetic reaction center from Rhodopseudomonas viridis.
J. Mol. Biol. 180, 385–398 (1984).
44. J. Deisenhofer, O. Epp, K. Miki, R. Huber and H. Michel, Structure of the protein
subunits in the photosynthetic reaction center of Rhodopseudomonas viridis at 3 Å resolu-
tion. Nature 318, 618–624 (1985).
45. M. Garavito and J.P. Rosenbusch, Isolation and crystallization of bacterial porin. Meth.
Enzymol. 125, 309–328 (1986).
46. S.W. Cowan, et al., The structure of OmpF porin in a tetragonal crystal form. Structure
3, 1041–1050 (1995).
47. R. Dutzler, et al., Crystal structure and functional characterization of OmpK36, the
osmoporin of Klebsiella pneumoniae. Structure 7, 425–434 (1999).
48. I.A. Vartanyants, I.K. Robinson, I. McNulty, C. David, P. Wochner, and Th.
Tschentscher, Coherent X-ray scattering and lensless imaging at the European XFEL
Facility. J. Synch. Rad. 14, 453–470 (2007).
49. H.N. Chapman, et al., Femtosecond diffractive imaging with a soft-X-ray free-electron
laser. Nat. Phys. 2, 839–843 (2006).
Chapter 10
Novel Imaging Techniques: Polarized Neutrons
and Neutron-Based Magnetic Resonance Imaging
Neutrons are sensitive to magnetic fields due to their magnetic moment, i.e., spin.
Therefore, similar to the conventional attenuation contrast image, the magnetic
field inside and around a sample can be visualized independently by detection of
N. Kardjilov (*)
Helmholtz Centre Berlin for Materials and Energy, Berlin, Germany
e-mail: [email protected]
Z
g
’ ¼ !L t ¼ L Bds (10:1)
v
path
Fig. 10.1 Schematic of the set-up used for imaging magnetic materials on CONRAD. The
neutron beam is first polarized, then precesses around a magnetic field is analyzed and finally
detected. Note how the intensity (represented by the height of the dark arrow) behind the
analyzer is smaller than the intensity behind the polarizer [1]
10 Novel Imaging Techniques 173
where I0(x,y) is the incident beam intensity, (x,y) is the linear attenuation
coefficient of the sample and (x,y) are the coordinates in the detector plane. The
cosine implies a periodic transmission function for the analyzed precession
angles and complicates a straightforward quantification with respect to the
traversed magnetic fields. By use of a calculation model based on the Biot-
Savart law, it was shown that quantitative analysis is possible [1].
Imaging experiments using the setup shown in Fig. 10.1 were carried out at
the neutron tomography facility CONRAD at HMI [1]. For this purpose the
instrument was equipped with solid-state polarizing benders [2] providing a
beam with a cross-section of 15 mm width and 45 mm height. For investigation
of larger samples, up to 20 cm width, a scanning arrangement was adapted. A
double crystal monochromator [3] was used to select a defined wavelength from
the cold neutron spectrum. The spatial resolution achieved in the radiography
images was around 500 mm for the given experimental geometry.
The potential of the method was demonstrated by visualization of the
magnetic field around a simple dipole magnet (Fig. 10.2). This shows the
Fig. 10.2 A radiograph showing the field lines surrounding a bar magnet. The magnetic field
decreases in strength with distance from the magnet, resulting in a series of maxima and
minima, where the beam polarization is sequentially parallel or antiparallel to the analyzer.
Very close to the magnets (where the field is strongest) the field lines are too close together to
be spatially resolved
174 N. Kardjilov et al.
decay of the magnetic field strength with increasing distance from the
magnet, resulting in an annular structure around the sample with an increas-
ing period due to the changing precession angles of the neutron spins on
their path through the strongly decaying field. The gray level scale used in
the images is related to the intensity variations induced by the sample and
the magnetic field (from black = minimum to white = maximum, repre-
senting the periodic 2p rotation of the neutron spin). The structure in the
center part cannot be seen due to the limited spatial resolution of
approximately 500 mm [4].
This imaging method can in some instances be extended into three dimen-
sions by a standard tomographic technique. Figure 10.3 shows the distribution
of a magnetic field trapped inside a polycrystalline lead cylinder, with a dia-
meter of 1 cm and length of 3 cm, that becomes superconducting when cooled
below the critical temperature, Tc = 7.2 K. The sample was cooled down to
6.8 K in a homogenous magnetic field of 10 mT. After this the magnetic field
was switched off, resulting in partially trapped magnetic fields in the
Fig. 10.3 A 3D reconstruction of trapped flux (solid regions left and right) inside a
polycrystalline cylinder of lead. When cooled to below its critical temperature (7.2 K) in the
presence of a weak magnetic field, some flux is present inside due to defects and grain
boundaries and this remains trapped even after the field is switched off [1]
10 Novel Imaging Techniques 175
superconductor due to grain boundaries and other defects [5, 6]. The tempera-
ture dependence of the residual field distribution inside the sample was visua-
lized by recording radiographic images during the heating process. The images
(Fig. 10.3) show an inhomogeneous residual field which decreases during heat-
ing and vanishes completely when the critical temperature Tc is reached. For the
weak trapped residual field at 7.0 K, a tomographic investigation was per-
formed by rotating the sample around its vertical axis. The beam attenuation
for each pixel can be related to absorption and magnetic contrast (Eq. 10.2)
assuming that the trapped magnetic field conserves its main orientation per-
pendicular to the beam polarization and is weak enough to cause spin rotations
smaller than p for all recorded projections. The volumetric data set was recon-
structed from 60 2D images, collected through rotation over 1808,
using a filtered back projection algorithm (Chapter 6). The results show the
three-dimensional (3D) representation of the flux trapped in the sample. Flux
concentrations could be found close to the end surfaces of the cylinder and at
the position where the sample was held with a screw.
This imaging method has many potential applications. The presence and
controlled application of magnetic fields are essential in many fields of science
and technology as well as in fundamental physics. For example, the flux
distribution and flux pinning in large superconducting samples, the skin effect
[7] in conductors, or magnetic domain distributions in bulk ferromagnets could
be visualized and studied in detail.
The Larmor precession described in the previous section forms the base of the
neutron spin-echo technique which is typically used to determine small energy or
momentum changes upon scattering [7], see also Neutron Scattering Instrumen-
tation [8]. In a typical spin-echo setup (Fig. 10.4), two identical regions of
magnetic field are applied before and after a ‘‘p-flipper.’’ A neutron with polar-
ization perpendicular to the field undergoes Larmor precession in the first field
region and its spin rotates by an angle of radians. Due to the energy spread of
the neutron beam, the spins of different energy neutrons acquire different pre-
cession angles and the neutron beam appears to be depolarized after the first field
region. The p-flipper between the two magnetic field regions rotates the neutron
spins about an axis perpendicular to the applied field by 1808. This has the
effect of reversing the angle of the neutron spin from to –. Since the two
magnetic field regions are identical, the neutron spin rotates an additional
angle of radians in the second field region. The spin precession in the two
field regions cancels and the beam ‘‘recovers’’ its polarization after going through
the field regions. This setup constitutes a neutron spin-echo instrument, and
the process through which the neutron beam recovers its polarization is called
‘‘spin echo.’’
176 N. Kardjilov et al.
Fig. 10.4 A basic neutron spin-echo instrument setup. The spin of a neutron undergoes
Larmor precession in the upstream magnetic field. A p-flipper between the magnetic field
regions inverted the spin rotation angle from to –. The neutron spin rotates an additional
angle equal to in the identical downstream field region and returns to its starting direction
Nbl2
n¼1 ; (10:3)
2p
where mn is the neutron mass, h is the Planck’s constant, and b is the coherent
scattering length. The quantity N(x)b(x) is more commonly known as the
neutron scattering length density. The magnitude of the neutron scattering
length density ranges from 10–7 Å–2 to 10–5 Å–2. Hence the spin angle change
measures the path integral of the scattering length density.
In a modification of the neutron spin-echo instrument discussed above,
placing a material in one arm of the pair of magnetic field regions results in a
10 Novel Imaging Techniques 177
net neutron spin angle that corresponds to the path integral of the scattering
length density. With a proper neutron beam arrangement, this effect can be
applied to measure the optical potential as a function of position in a material –
an imaging technique. A typical benchmark for an imaging technique is a 1 mm
spatial resolution. A common neutron polarization analyzer such as a polarized
3
He analyzer (Chapter 3 in this book) may resolve a spin angle of 108. If the
magnetic field strength is strong enough to rotate the neutron spin by 108 over
1 mm of its flight path, a 1 mm imaging resolution in the direction along the
beam path can be achieved. Using a typical scattering length density of
Nb=10–6 Å–2, Fig. 10.5 shows the magnetic field required to reach = 108
over a 1 mm flight path as a function of the wavelength. A potential imaging
device using this spin-echo technique would therefore use cold neutrons with
wavelengths ranging from 20 to 100Å and an applied field from several gauss
for 100Å neutrons to 1 Tesla for 30Å neutrons. The magnetic potential energy
U ¼ ~ ~ where
B, is the neutron magnetic moment, is equal to 0.06 meV in a 1
Tesla field. Hence, compared with a kinetic energy of 8.181 meV for 100Å
neutrons, the optical effect of the magnetic potential can be ignored.
One main difference between the imaging application and the neutron spin-
echo spectrometer is that in imaging, the additional change in the spin angle
occurs only when the neutron is inside the material, whereas the change accu-
mulates over the entire field region after the sample in the case of spin echo
spectrometer. This makes the spin angle change insensitive to the beam diver-
gence. The beam divergence, however, is still important in that neutrons
through different parts of a sample may end up striking the same detector
pixel. This blurring of the image is, however, a common concern for all imaging
techniques rather than being specific to the spin contrast imaging. In neutron
imaging, collimation is used to reduce the divergence and anti-grids can be used
between the sample and the detector system. The same setup can be applied in
the techniques discussed here.
To acquire a three-dimensional image of an object using imaging techniques
based on direct neutron absorption, measurements with the neutron beam
passing the object at different angles are first carried out and tomography
reconstruction is then applied to render the 3D image. The advantage of the
imaging technique discussed here is that it can be used with a time-of-flight
(TOF) technique to achieve 3D imaging without the need to measure the object
at many different angles and without using tomographic reconstruction.
The idea behind the technique is illustrated in Fig. 10.6. Initially the neutron
spins are aligned parallel to the applied field. At time t=0, a p/2 pulse is applied
over the length of the sample to rotate the neutron spins to a perpendicular
direction with respect to the field, which starts the precession of the neutron
spins. Along the neutron flight path, neutrons at different position along a flight
path go through a different path length through the material and therefore
acquire a different spin angle change . For instance, the spin change angle is 1
for neutrons starting at point 1 and 2 for neutrons starting at point 2, respec-
tively. The spin angle change is converted to intensity change by a polarization
analyzer before the detector. Since neutrons at point 1 at t=0 reach the detector
later than neutrons at point 2 at t=0, by recording the time-of-flight of each
neutron at the detector, the position of each neutron at t=0 can be determined.
Subtracting 2 from 2 gives the spin angle change that corresponds to the
scattering length density from point 1 to point 2. Applying this analysis to a
series of images taken at consecutive times-of-flight, a slice-by-slice 3D distri-
bution of the scattering length density of an object can be obtained. The speed
of 30Å neutrons is 131.87 m/s. It takes 7.6 ms to travel a distance of 1 mm. A
timing resolution of 1 ms would be sufficient to resolve the initial location of the
neutrons at the time the precession is started.
Fig. 10.6 Technique for direct 3D imaging. At time t=0, a p/2 flip starts the neutron
precession; neutrons at point 1 precess through a longer path in the sample than neutrons at
point 2. Subtracting the two gives the precession angle from point 1 to point 2
10 Novel Imaging Techniques 179
(a) (b)
(c) (d)
Fig. 10.7 Neutron intensities at the detector if spin precession starts (a) when the neutrons are
right before the sample, and (b) when the neutrons are passing the middle of the sample. The
color scales are relative; (c) The normalized integrated intensity at the detector along the
x-direction of the detector from y=+2 mm to y=+3 mm from the two detector images
shown in (a) and (b); (d) The spin angle change calculated from the two time-of-flight
intensities
both high magnetic field and very cold neutrons: From Eq. (10.1) above, 30Å
neutrons in a 1 T field precess 221 times every 1 mm flight path. If identical
upstream and downstream field regions are used, the location where the p-
flipping occurs will need to be accurate to 0.1 mm for a 108 accuracy. This
would be a formidable task using currently available technology. However,
only a small region that covers the sample will need the 1 T field. The rest of
the field regions can use a field that can be substantially lower. This is because the
spin precession angle depends upon both the field strength and the path length
along which the field is imposed. A smaller field over a long distance can result in
the same spin rotation as a larger field over a shorter distance. If a nominal 1 T
field is applied over a 1 cm length where the sample is located and the rest of the
setup is in a 20 gauss field, the second field region can extend 5 m and require an
10 Novel Imaging Techniques 181
accuracy of 0.06 mm in the p-flipping location for a 108 accuracy. The third
challenge results from the fact that the precession of the neutrons starts at different
locations along the beam. Hence either the downstream p/2 pulse or the p-flipping
will need to be applied to different locations along the flight path. This would
require a more sophisticated p-flipper setup than those currently used at neutron
beam lines, which typically apply p-flipping over a small distance along the beam.
Hence although we have shown that spin contrast tomography is theoreti-
cally feasible, there are some interesting technical challenges to be overcome
before the method can be applied in practice.
freeze liquid samples, and the microwaves suffer from the same penetration
depth problems as the lower-frequency RF. An alternative approach to mea-
suring the decay time must be found to make this technique accessible to more
types of samples. A potential solution is to use high magnetic fields.
With the availability of high magnetic fields, a simple extension of the
aforementioned techniques would be to use the field to raise the temperature
necessary for the saturation of the nuclear spins. Figure 10.8 examines this
possibility by looking at the nuclear polarization as a function of temperature
for different fields. As can be seen from the figure, T 10 mK is required for the
current state of the art field for neutron-scattering experiments of 15 T. If the
field was increased to 55 T, the required temperature is in the hundreds of mK
range. Such an increase makes the measurement simpler but does not relieve the
freezing problem. To use this technique at room temperature requires 105 T,
an unrealizable static field.
So the main challenge is to find another way to measure t. The approach
described below eliminates the need for microwaves and/or mK temperatures.
The general idea is to use spectroscopy to measure the Zeeman splitting of
relaxing nuclear spins at a high field, then quickly drop the field to a lower value
and watch the Zeeman splitting relax to the equilibrium value at the lower field.
A large static field is required to sufficiently separate the peak associated with
the Zeeman splitting from other incoherent processes. Protons are the nuclei of
choice for this study, as they are the most familiar NMR active nuclei. The
protons are assumed to be provided by a water sample at room temperature.
Therefore the nuclear relaxation time is assumed to be 200 ms. A static field of
1.0
0.8
H = 5x105 T
Polarization
H = 4.5T
H = 15T
H = 55T
0.6
0.4
0.2
0.0
Fig. 10.8 The normalized proton polarization as a function of temperature for different
values of magnetic field
10 Novel Imaging Techniques 183
Fig. 10.9 The resonance energy for a system of hydrogen nuclei (dashed curve) for a given
applied field (solid curve)
8.5
2000
8.0 1500
cnts
Resonance Energy (µeV)
7.5 1000
500
7.0
0
–30 –20 –10 0 10 20 30
6.5
E(µeV)
simulated data
6.0
fit
5.5
5.0
0.4 0.6 0.8 1.0 1.2 1.4 1.6
t(s)
Fig. 10.10 The resonance energy determined from the simulated spectra. The resonance
energy is fitted to an exponential decay to determine the relaxation time. The inset shows an
example simulated spectrum (circles) and a fit (solid)
on the decay curve should be counted for a total of 10 s. Therefore the measure-
ment would be repeated many times and the aggregated results analyzed. In
practice this means a 0.6 Hz pulsed magnet operated for 600 pulses. A full
relaxation time measurement would require 20 min.
Each of these resultant spectra are analyzed by fitting 3 Lorentzians of 1 meV
width, convoluted with the instrumental energy resolution of 3 meV, at each
time interval. The parameter to be tracked, as a function of time, to determine
the relaxation time is the splitting of the finite energy transfer peaks from the
center. The results from the simulated data for this parameter and its uncer-
tainties are plotted as points in Fig. 10.10. An exponential decay is then fit to the
points to determine the relaxation rate. The specific function used for this fit is
E ¼ Aet=t þ E0 with E0 the field splitting at the lower field value and A a
normalization constant. The results are as follows: t = 196 5 ms, A = 25
2 meV and Eo = 5.33 0.02 meV. The fitted relaxation or Zeeman splitting time,
t, is in excellent agreement with the input time of 200 ms. Therefore, simulations
show that neutron measurements of the relaxation time could be performed by
watching the timed decay of the Zeeman splitting.
10 Novel Imaging Techniques 185
To test the limits of this method, similar calculations were performed with a
10 T static field. However, for this configuration the Zeeman splitting was too
small at the longest times to be resolved from the central incoherent peak. This
resulted in an artificially large value of t from the fit. In principle the t
measurement will be influenced at longer times by the spins already flipped at
shorter times. However, less than 106 spins are flipped before the next field pulse
is applied, which is a negligible quantity when compared with the total number
of spins in the system.
In order to do this measurement, a high-resolution spectrometer such as
BASIS at the U.S. Spallation Neutron Source (SNS), a high (static) field
magnet in excess of 30 T such as that proposed for Zeemans at the SNS, and a
10 T pulsed magnetic field are required [18, 19].
Though the core technology behind neutron-probed MRI has been
described, other significant challenges remain. Position encoding is one of
them. In traditional MRI a field gradient is placed across the sample [20, 21].
Then the resonance condition is only met at one spot in the sample. The field
gradient is then varied to change this location. A field gradient would only
smear out the neutron-probed measurement as it measures all energies. Instead,
we would raster the sample through the beam so only a single-point is measured
at a time. We assume the state of the art pixel size of 1 1 mm to 0.1 0.1 mm,
as defined for the Vulcan instrument at SNS [22]. An additional restriction is
that the bore in the proposed magnet [19] is only 50 mm. Therefore rastering a
sample within this bore requires a small sample 1 1 cm and would provide
between 100 and 10000 unique points depending on the beam size. Extending
this technique to larger samples is a motivation for increasing the bore size in
the next generation of superconducting hybrid magnets. Another approach
would be to use small pulsed magnets that can achieve 30 T, but have a
repetition rate of 1 pulse per 7–8 minutes [23]. A sample sandwiched between
two of these magnets and rastered independently of the magnets provides a
way to increase the sample size. For this approach one would have to stop
measuring the relaxation after the peaks associated with the Zeeman splitting
coalesce with the main incoherent peak. The main restriction of this approach is
the long time between pulses. This time is controlled by internal heating of the
magnet. The time between pulses could be reduced to the time limit of the
capacitor bank power supply by rastering additional magnets while the first
pair cools.
For both styles of magnets, field gradients are important. Though for tradi-
tional MRI homogeneity on a few PPM [20] is expected, 100 ppm is accep-
table. The pulsed magnets are quite configurable so field gradients of this order
are reasonable. For the Zeemans style of magnet, a high-homogeneity (100 ppm)
resistive insert would have to be designed. This insert design would drop the peak
field to 28 T (M. D. Bird, private communications). Nevertheless, one could
envision running the magnet for a series of cycles in a high field mode and then
changing for another series of cycles to a high homogeneity mode.
186 N. Kardjilov et al.
Acknowledgments Garrett Granroth is grateful for useful discussions with J.-K. Zhao,
G. Greene, E. Iverson, M. W. Meisel, N. Sullivan, and B. Halperin. His work was performed
at Oak Ridge National Laboratory, managed for the U.S. Department of Energy by UT-
Battelle, LLC, under Contract No. DE-AC05-00OR22725.
References
1. N. Kardjilov, I. Manke, M. Strobl, A. Hilger, W. Treimer, M. Meissner, T. Krist, J.
Banhart, Three-dimensional imaging of magnetic fields with polarised neutrons, Nat.
Phys. 4, 399–403 (2008).
2. Th. Krist, S. J. Kennedy, T.J. Hick, F. Mezei, New compact neutron polarizer. Physica B
241–243, 82–85 (1998).
3. W. Treimer, M. Strobl, N. Kardjilov, A. Hilgar, I. Manke, Wavelength turable device for
neutron radiography and temography, Appl. Phys. Lett, 89, 203504 (2006).
4. A. Hilger, N. Kardjilov, M. Strobl, W. Treimer, J. Banhart, The new cold neutron
radiography and tomography instrument CONRAD at HMI Berlin, Physica B
385–386, 1213 (2006).
5. P. Gammel, D. Bishop, Fingerprinting vortices with smoke, Science 279, 410–411 (1998).
6. Ch. Jooss, J. Albrecht, H. Kuhn, S. Leonhardt, H. Kronmüller, Magneto-optical studies
of current distributions in high-Tc superconductors. Rep. Prog. Phys. 65, 651–788 (2002).
7. I. Manke, N. Kardjilov, M. Strobl, A. Hilger, J. Banhart, Investigation of the skin effect
in the bulk of electrical conductors with spin-polarized neutron radiography, J. Appl.
Phys, 104, 076109 (2008).
8. http://www.springer.com/series/8141, this series of books
9. A.I. Frank, I. Anderson, I.V. Bondarenko, A.V. Kozlov, P. Hoghoj, and G. Ehlers, Phys.
Atom. Nucl. 65, 2009 (2002).
10. T.A. Jyrkkiö, M.T. Huiku, K.N. Clausen, K. Siemensmeyer, K. Kakurai, and M. Steiner,
Calibration and applications of polarized neutron thermometry at milli- and microkelvin
temperatures, Z. Phys. B 71, 139 (1988).
11. J.B. Hayter, G.T. Jenkin, and J.W. White, Polarized-Neutron Diffration from Spin-
Polarized Protons: A Tool in Structure Determination? Phys. Rev. Lett. 33, 696 (1974).
12. M. Dobhal, NMR Data for Carbon-13. Part 4: Natural Products, Springer, Heidelberg
(2006).
10 Novel Imaging Techniques 187
13. http://www.lanl.gov/orgs/mpa/nhmfl/60TLP.shtml
14. http://www.lanl.gov/orgs/mpa/nhmfl/images/60TLPpulse.jpg
15. C.P. Slichter, Principles of Magnetic Resonance, Springer-Verlag, Heidelberg (1990).
16. MathCad, Parametric Technology Corporation.
17. G.E. Granroth, Prospects for Neutron probed NMR in The proceedings of ICANS–XVII,
J. Wei, S. Wang, W.L. Huang, J.S. Zhao, ed. http://www.icans-xviii.ac.cn/proceedings/
indexed/copyr/15. pdf.
18. http://www.sns.gov/instrument_systems/beamline_02_basis/index.shtml, (2007).
19. G.E. Granroth, C. Broholm, F. Klöse, G. Srajer, Z. Islam, Y. Lee, and J. Lang, Report
from Probing Matter at High Magnetic Fields with X-Rays and Neutrons, SNS Report
SNS107000000TR-005-R00, (2005).
20. J. Jin, Electromagnetic Analysis and Design in Magnetic Resonance Imaging, CRC Press,
Boca Raton (1999).
21. P. Sprawls, Magnetic Resonance Imaging, Medical Physics Publishing, Madison (2000).
22. X.-L. Wang, T.M. Holden, G.Q. Rennich, A.D. Stoica, P.K. Liaw, H. Choo, and C.R.
Hubbard, Physica B 385–386, 673 (2006).
23. K. Ohoyama, N. Katoh, H. Nojiri, Y.H. Matsuda, H. Hiraka, K. Ikeda, H.M. Shimizu,
Neutron diffraction under 30 T pulsed magnetic fields, J. Mag. Mag. Mat. 310, e974
(2007).
Section C
Chapter 11
Neutron Imaging for the Hydrogen Economy
Abstract The development of fuel cells and hydrogen storage materials will be one
of the highest global research and development priorities for the foreseeable future.
The particular abilities of neutrons to penetrate materials and to image hydrogen
will make neutron imaging a key technique, allowing the in-situ study of real
operational devices. This chapter describes the current state of the art in this field.
11.1 Introduction
Compared to most other forms of radiation, neutrons are highly efficient means of
probing complex structures because of their tremendous of penetrating capability
many common materials and their unique capability of distinguishing different
materials with similar chemical and physical properties. Consequently, neutron
imaging has always been regarded as one of the most potentially important tools
for many areas of industrial research. However, these promises have remained
largely unfulfilled, and neutron imaging techniques have seen limited applications,
primarily due to traditionally poor spatial resolution and slow processing time as
well as lack of identification of critical, high-impact areas of research. Most
industrial and academic research that can significantly benefit from this technique
involves dimensions that are a few tens of micrometers in length and interaction
times that last at longest a few seconds. The current practical limitations for
neutrons are about 100 mm in spatial resolution and tens of seconds in time
resolution. These limitations are mainly due to the relatively weak neutron sources
(nuclear reactors that are many orders of magnitude weaker than even table-top
photon sources), large mean free paths of neutrons and secondary radiation in
M. Arif (*)
National Institute of Standards and Technology, 100 Bureau Drive, MS 6100,
Gaithersburg, MD 20899-6100, USA
e-mail: [email protected]
materials (limits spatial resolution), and the inability to process data in real-time
while being statistically significant (limited by weak source).
However, in the past decade enough gains have been made to enable neutron
imaging to be applied in different fields of science and technology. One of the most
important prospects is the understanding of the interface physics, chemical kinetics,
and proper modeling of the processes that govern the operation of the future
generation of energy production and storage devices based on hydrogen. The
sensitivity of neutrons to hydrogen, together with other unique properties, makes
them an ideal probe for characterizing the structure, morphology, and hydrogen
dynamics that are key to the future use of these promising devices. The neutron
imaging technique, in particular, has emerged as one of the most important tools in
the development of robust and efficient fuel cells accompanying hydrogen storage
materials and devices. This article will focus on the application of neutron imaging
research in the development of low-temperature fuel cells that may power homes,
consumer electronics, and transport vehicles in the emerging hydrogen economy.
Fig. 11.1 Schematic of a fuel cell. The platinum (Pt) catalyst is generally applied to both sides
of the membrane to form the membrane electrode assembly. Formation of water at the
membrane can cause the cell to perform poorly
anode side of the MEA to the cathode side with the proton transport. In the
meantime, water is generated on the cathode side of the MEA. A water gradient
thus exists within the membrane. The PEM water content is extremely critical as
it impacts the ionic conductivity and mechanical properties such as strength and
gas permeability. Depending on the water concentration, the membrane per-
formance and lifetime can be significantly reduced. In order to maintain the
proton conductivity and preserve membrane life, PEM fuel cells typically
require humidified reactant gases. This vapor, together with the water gener-
ated during operation, poses significant challenges for water management. A
hydrophobic porous gas diffusion layer (GDL) is typically used to help manage
water distribution. Furthermore, the electrode composition may be hydropho-
bic or hydrophilic, which is another area of water management difficulty.
PEM fuel cells have gone through significant evolution in recent years, but
one of the remaining key challenges is efficient water management. The control
194 M. Arif et al.
of incoming humidity and the product water within the MEA and GDL is
generally acknowledged as one of the most crucial aspects of operating a PEM
fuel cell. Too much water within the MEA or GDL will result in flooding
conditions that impede gas diffusion; too little water will reduce the membrane
proton conductivity, thereby decreasing the cell performance; and cycling
between water levels will reduce membrane life. Thus, proper water manage-
ment is the key to a stable and long-life PEM fuel cell and must be achieved by
properly designing the flowfield, the GDL, the MEA, and their interfaces. This
requires a fundamental understanding of the in situ water distribution in an
operating fuel cell.
It is critically important for both optimal design and efficient operation of
the fuel cells to study and quantify these adverse effects in situ using non-
destructive means [2]. At present, neutron imaging is the only technology that
has the ability to look inside a standard, commercially viable PEM fuel cell
while the cell is operating and characterize the water content non-destructively.
Since neutrons are highly penetrating in most materials, they are ideal probes in
the investigation of internal structures of bulk objects like the fuel cell. Thermal
neutrons (low-energy neutrons) passing through matter experience the follow-
ing interactions: (a) absorption, (b) scattering (elastic, inelastic, small angle
scattering), (c) refraction, (d) change in the phase of the neutron wave,
(e) depolarization of an incident spin direction.
These interactions are used for signals in thermal neutron two-dimensional
(2D) (radiographic) and 3D (tomographic) imaging (Chapters 5 and 6, in this
volume). The most common processes are absorption and scattering. The ther-
mal cross-sections for absorption and scattering events are complex functions of
atomic mass, atomic number, nuclear isotope, and neutron energy. The contrast
obtainable between materials in radiographic images is a function of their differ-
ent cross-sections. For thermal neutrons, hydrogen-based compounds (such as
water) would be highly visible against aluminum, iron, or carbon, which are
commonly used in fuel cell construction (Fig. 6.1). This is due to the fact that the
so-called mass attenuation coefficient of water is nearly two orders of magnitude
larger than that of the background materials. This unique signature of hydrogen
allows for the study of water transport phenomena in fuel cells with sub-micro-
gram precision.
Vehicles powered by PEM fuel cells require, among other things, the presence of
onboard hydrogen storage vessels. The design of metal-hydride beds with
suitably low weight and volume to meet the demands of automotive applica-
tions is no trivial matter. The characterization of prototype technologies would
benefit greatly from an in situ imaging technique such as neutron radiography
to aid in understanding the hydrogen concentration gradients across these beds
11 Neutron Imaging for the Hydrogen Economy 195
Fig. 11.2 Neutron radiography study of hydrogen storage in LaNi4.78Sn0.22. (a) Photo of the
Sievert’s reactor. (b,c) Colorized (bottom) and raw greyscale (middle) images of the reactor and
the hydrogen content (top) in the uncharged and charged states, respectively
Fig. 11.3. Cross-sectional drawing of a prototype hydrogen storage bed. The porous alumi-
num is filled with LaNi4.78Sn0.22. With neutron tomography, the effects of hydrogen uptake,
such as swelling and temperature gradients, can be revealed
196 M. Arif et al.
Fig. 11.4 A sub-image of a neutron radiograph of a 50 cm2 active area fuel cell clearly showing
the water content (dark areas) of the separate channels. In this case the anode and cathode
channels were oriented perpendicularly to permit discrimination of anode vs. cathode channel
water.
Fig. 11.5 One of the first neutron radiographs of an operating fuel cell, showing the water
gradient across a 0.5 mm thick (wide) membrane, comprising several thermally bonded
Nafion 117 membranes. The darker region in the membrane center indicates more water
water retention [10]. Having a broad range of pore sizes can ensure that there
are sufficient pathways for reactant gas to reach the catalyst layer [11].
where I is the transmitted and I0 the incident neutron flux, N is the atom
number density, t is the sample thickness, and is the total scattering cross-
section responsible for the neutron loss in the forward direction passing through
the sample. Since the primary goal is to image and quantify the water content,
all measurements involve the basic steps of taking a dry and wet image of the
device. The wet image is normalized with respect to the dry image to determine
the profile of the water thickness. The detection system typically consists of a
scintillator viewed by either a charge-coupled device CCD or an amorphous
silicon flat panel. The spatial resolution, which is a convolution of detector
200 M. Arif et al.
where I0 is the incident fluence rate (cm–2 s–1), T is the acquisition time, and is
the detection efficiency [15]. Typical random uncertainties achieved at the
national institute for standards and technology (NIST) neutron imaging facility
range from 1 to 10 mm.
Both 2D and 3D neutron imaging are used to study water dynamics in fuel cells.
Neutron tomography can provide important information about an object’s
internal construction and about its constituent materials. The neutron tomo-
graphy technique allows one to reconstruct a 3D image of an object from the 2D
neutron radiographs taken for several rotations of the object about a rotation
axis (Chapter 6). Using the technique of filtered back projection (FBP), these
projection images are Fourier filtered and then back-projected across planes
orthogonal to the rotation axis. The result of this process is a stack of 2D
reconstructions of the planes of the object, orthogonal to the rotation axis. The
result of the FBP is a 3D map of the object’s neutron attenuation coefficient.
Subtracting the dry map from the wet map yields the 3D water profile of the fuel
cell. The spatial resolution of the map is determined by the resolution of the
detector system, on the order of 100 mm. Figure 11.6 shows the 3D water
content of an operating fuel cell at steady state.
Because of the time required to rotate the fuel cell by 1808, neutron tomo-
graphy can be used only as a steady state measurement. Often the desired
information is the average water gradient going from the anode to the cathode.
This ‘‘through-plane’’ water gradient must be imaged with higher resolution
than is typical of scintillator-based detectors, since the total thickness of the
PEM sandwich is only on the order of 0.5 mm. Various companies are devel-
oping new detector technologies aimed at achieving 10 mm or better resolution
in neutron detection in the near future [16]. These detectors are based on micro-
channel plates and have been demonstrated to have resolution better than
25 mm [17] (Chapter 3). With a smaller pixel size, image acquisition time must
11 Neutron Imaging for the Hydrogen Economy 201
Fig. 11.6 A small fuel cell (4 cm2) was used to demonstrate neutron tomography. The water
distribution in each slice plane from the anode to the cathode shows that anode flooding is
occurring in this fuel cell as a result of the thick (1 mm) gas diffusion layers. The data set was
acquired within 20 min
only are the bulk properties of the sample measured, but the elemental distribu-
tions of local inhomogenities are also observable. This technique is typically a
few hundred times more sensitive than normal radiography. It is ideally suited
for detecting laminate separation and minute thickness variations of polymer
electrolyte layers. NIST is developing a cold neutron imaging facility that will
enable research to measure the phase gradient, using a variety of such techni-
ques, including a grating method based on the Talbot effect [18].
Ultimately, these technologies can be used as a quality control tool for fuel
cell manufacturers. Fuel cells can be subjected to routine scans for the verifica-
tion of proper assembly and operational characteristics. Additional resolution
may be achieved using neutron point sources to magnify the neutron image of
the sample.
Fig. 11.7 (a) Photograph, and (b) schematic of the NIST neutron imaging facility. Also shown
is (c) the fuel cell test stand and (d) the sample area with connections to a typical fuel cell
sparger bottles, and heated inlet gas lines prevent condensation. The hydrogen is
supplied by a hydrogen generator so there is no stored inventory of hydrogen in
the confinement building. There is an extensive set of hydrogen safety features,
including several hydrogen sensors, a line pressure sensor, and flame detection in
the hydrogen vent that automatically shuts down all hydrogen systems in the
event of a detected emergency. Also, there is a freeze chamber, capable of
controlling the temperature between –408C and 508C for use in freeze/thaw
studies.
times under different conditions. The fuel cell was operated at 0.4 V with both
gas streams humidified to a relative humidity (RH) of 75 % for 2 h to ensure
that the GDL saturation was at equilibrium. After this precondition, the fuel
cell load was dropped. In the first case, the humidified gas flows were main-
tained, and in the second case the gas flows were stopped. By so doing, the
efficacy of the competing GDL water transport mechanisms of capillary action
versus evaporation and advection could be compared. Figure 11.8 shows the
first case in which the fuel cell load was stopped but the gas flow was main-
tained. The water content in the GDL is seen to decline rapidly to the comple-
tely dry state within about 15 min. Figure 11.9 shows the second case where
Fig. 11.8 Water transport in the GDL when the fuel cell load is stopped, but the under-
saturated flows are maintained. (A–D) Through-plane water profiles are shown from four of
the repeat measurements. The grayscale neutron images show the water profile change from a
time after the load was stopped
11 Neutron Imaging for the Hydrogen Economy 205
Fig. 11.9 Water transport in the GDL when the fuel cell load and the gas flows are stopped. (A–D)
Through-plane water profiles are shown from four of the repeat measurements. The grayscale
neutron images show the water profile change from a time after the load was stopped. Compared
with the evaporation/advection case, the change in water content of the GDL is negligible
both the fuel cell load and gas flows were stopped. The water content of the GDL
barely changes, even after 70 min, indicating that capillary action plays a smaller
role than evaporation and advection in the water transport through the GDL.
11.6 Conclusion
Today this technique is used by almost all automobile manufacturers and major
fuel cell developers. By all accounts, neutron imaging research has shortened
the development time of fuel cells and is providing reliable, real-time data
important for optimum design and performance that was almost impossible
to get only a few years ago. Though the primary focus of neutron imaging
research has been on fuel cells during the last few years, it has now started to
play an expanded role in the development of hydrogen storage materials and
devices. It is expected that this trend will continue in the coming years. In spite
of all the success that neutron imaging has enjoyed in recent years, even higher-
impact research is limited by the current temporal and spatial resolution of
neutron detection devices, the lack of very accurate analytical techniques, and
the lack of availability of wide infrastructures for hydrogen research at various
neutron facilities around the world. Fortunately, all of these issues are being
aggressively pursued by various government laboratories and industrial and
academic researchers. It is expected that in the very near future, both temporal
and spatial resolution will improve by an order of magnitude. This, coupled
with advanced methods such as phase contrast imaging, will provide much
needed enhanced capabilities and open new avenues of research that are not
possible today. Consequently, in the foreseeable future, neutron imaging will
continue to significantly contribute to enhanced efforts to develop and com-
mercialize pollution-free, hydrogen-powered energy devices that are the cen-
terpieces of the emerging hydrogen economy.
References
1. Certain trade names and company products are mentioned in the text or identified in an
illustration in order to adequately specify the experimental procedure and equipment used.
In no case does such identification imply recommendation or endorsement by NIST, nor
does it imply that the products are necessarily the best available for this purpose.
2. J. St-Pierre, J. Electrochem. Soc.154, B724 (2007).
3. P. Bhandari, M. Prina, M. Ahart, R. C. Bowman, and L. A. Wade, Sizing and Dynamic
Performance Prediction Tools for 20 K Hydrogen Sorption Cryocoolers in Cryocoolers 11,
Edited by R. G. Ross, Jr. Kluwer Academic/Plenum Press, New York, pp. 541–549 (2001).
4. M. Prina, P. Bhandari, R.C. Bowman, L. A. Wade, D. P. Pearson, and G. Morgante,
Performance Prediction of the Planck Sorption Cooler and initial Validation, Advances in
Cryogenic Engineering, Vol. 47, edited by S. Breon, et al., Am. Inst. Phys., New York,
pp. 1201–1208 (2002).
5. P. Bhandari, M. Prina, R.C. Bowman-Jr, C. Paine, D. Pearson, A. Nash, Sorption Coolers
using a Continuous Cycle to Produce 20 K for the Planck Flight Mission, Cryogenics 44,
395–401 (2004).
6. D. Pearson, R. Bowman, M. Prina, P. Wilson, The Planck Sorption Cooler: Using Metal
Hydrides to Produce 20 K, J. Alloys Compounds 446–447, 718–722 (2007).
7. J. P. Owejan., T. A. Trabold, D. L. Jacobson, M. Arif, and S. G. Kandlikar, Int. J.
Hydrogen Energy 32, 4489 (2007).
8. R. J. Bellows, M. Y. Lin, M. Arif, A. K. Thompson, and D. Jacobson, J. Electrochem. Soc.
146, 1099 (1999).
11 Neutron Imaging for the Hydrogen Economy 207
9. M. Mathias, J. Roth, J. Fleming, and W. Lehnert, ‘Diffusion Media for PEM Fuel Cells,’
in Handbook of Fuel Cells – Fundamentals, Technology and Applications. Fuel Cell
Technology and Applications, Vol. 3, W. Vielstich et al.(Eds.), John Wiley & Sons, Chapter
46 (2003).
10. R. Mukundan, J. R. Davey, T. Rockward, J. S. Spendelow, B. S. Pivovar, D. S. Hussey,
D. L. Jacobson, M. Arif, and R. L. Borup, ’Imaging of Water Profiles in PEM Fuel Cells
Using Neutron Radiography: Effect of Operating Conditions and GDL Composition’,
ECS Trans. 11(1), 411 (2007).
11. K. Yoshizawa, K. Ikezoe, Y. Tasaki, D. Kramer, E. H. Lehmann, and G. G. Scherer,
ECS Trans. 3, 397 (2006).
12. M. A. Hickner, N. P. Siegel, K. S. Chen, D. N. McBrayer, D. S. Hussey, D. L. Jacobson,
and M. Arif, J. Electrochem. Soc.153, A902 (2006).
13. M. A. Hickner, N. P. Siegel, K. S. Chen, D. N. McBrayer, D. S. Hussey, D. L. Jacobson,
and M. Arif, J. Electrochem. Soc. 155(4), B427–B434 (2008).
14. S. Kim, A.K. Heller, M.C. Hatzell, M.M. Mench, D.S. Hussey, D.L. Jacobson, High
Resolution Neutron Imaging of temperature-driven flow in polymer electrolyte fuel cells,
American Nuclear Society National Meeting, Anaheim, CA (2008).
15. D.S. Hussey, D.L. Jacobson, M. Arif, K.J. Coakley, and D.F. Vecchia, In situ fuel cell
water metrology at the NIST neutron imaging facility, Proceedings of the ASME Fuel Cell
Conference, New York, June 18–20, 2007.
16. O.H.W. Siegmund, J.V. Vallerga, A. Martin, B. Feller, M. Arif, D.S. Hussey, and D.L.
Jacobson, Nucl. Instrum. Meth. A 579, 188 (2007).
17. W.B. Feller, P.L. White, and P.B. White, Gamma Insensitive Highly Borated Micro-
channel Plates for Neutron Imaging, Proceedings of the 8th World Conference on Neutron
Radiography, 583–591 (2008).
18. F. Pfeiffer, C. Grünzweig, O. Bunk, G. Frei, E. Lehmann, and C. David, Phys. Rev. Lett.
96, 215505 (2006).
Chapter 12
Material Science and Engineering
with Neutron Imaging
D. Penumadu
Abstract This chapter summarizes some of the results related to the use of
neutron imaging (radiography and tomography) as applied to the broad area of
materials science and engineering research. These include multi-phase flow
visualization in metal casting techniques, energy-selective imaging of materials
and its use for texture and stress imaging in polycrystalline materials, charac-
terization of discrete particle systems, flow through porous media, and strobo-
scopic imaging. The importance of spatial resolution and neutron detector type
for given engineering applications is also addressed.
12.1 Introduction
Film and image plate-based thermal neutron imaging for nondestructive inspec-
tion has found many useful engineering applications in the past three decades, as
demonstrated by the establishment of an American Society for Testing and
Materials (ASTM) standard (E1316) for its routine implementation at various
research reactors (Chapter 5). However, in the past decade the use of thermal and
cold neutrons for material science and engineering applications has rapidly
evolved due to a dramatic improvement in the availability of large-area amor-
phous silicon flat panels, high-resolution (20482048) and large charge-coupled
device (CCD) chip-based digital cameras, sophisticated and inexpensive optical
lenses, and bright and thin (50–200 mm) neutron scintillation screens. This has
D. Penumadu (*)
University of Tennessee-Knoxville, Joint Institute for Advanced Materials (JIAM)
Chair of Excellence, Department of Civil and Environmental Engineering, Knoxville,
TN 37996-2010, USA
e-mail: [email protected]
(a) (b)
Fig. 12.1 Radiograph of a computer floppy disk using neutrons (a) shows the polymeric
components clearly while X-rays (b) are sensitive to the metallic components
Table 12.1 Material properties and the calculated thickness for 50% attenuation for thermal
neutrons
Molecular Macroscopic Thickness for
Chemical Density weight cross-section 50% attenuation
Material formula (g cm–3) (g mol–1) (cm–1) (cm)
Aluminum Al 2.70 27 0.0984 7.05
Styrene C8H8 0.9 106 3.12 0.22
Water H2O 1.0 18 2.62 0.26
Silica SiO2 2.65 60 0.259 2.68
(a) (b)
allows high-quality imaging. The neutron beams are collimated with an L/D ratio
of 50–400, resulting in good spatial resolution for digital still radiographs. The
radioscopic systems use a CF Thompson tube as the neutron camera, which
combined with a SIT (silicon intensified tube) camera allows image capture at a
rate of 30 frames/s. The radiography setup for LFC is shown in Fig. 12.3. The
polymer foams used in the study are EPS foams with a density of 22.4 kg/m3.
The casting metal used in the study was aluminum alloy. This was heated to
14008 F/7608C before pouring into the EPS patterns. Real-time neutron radio-
graphy was used to visualize the molten metal and pyrolysis foam fronts during
the whole LFC process. The casting samples were plates of size 100 150 mm
with varying thicknesses (4, 12, and 24 mm), and various gating orientations
(bottom, side, and top), as shown in Fig. 12.2.
(a) (b)
Fig. 12.3 Bottom-gating casting with pattern thickness of 4 mm. (a) Neutron radiograph
(at 19 s). (b) Optical picture of the final casting
Real-time digital videos were recorded during the entire process of metal entry,
foam decomposition, and casting solidification. Specific images of interest were
then extracted from the digital videos, further processed, and analyzed using
Image-Pro Plus V4.5 [7]. Figure 12.3 shows an example neutron radiograph of a
bottom-gated casting that did not fill completely at 19 s since the hot aluminum
metal was introduced to the down-sprue. There is a wealth of information in such
radiographs as a function of time that is valuable for understanding the pyrolysis
aspects of foam and the associated phase transitions of both polystyrene (from
foam to liquid to gaseous monomer) and metal (liquid to solid state).
Only neutrons provide this outstanding ability to visualize and model complex
multiphase and multiscale processes, penetrating metal and at the same time
clearly showing light elements such as the byproducts of polymer degradation.
The neutron images shown in Fig. 12.4 indicate the effect of foam thickness on
the degradation process during LFC. It is of great interest to note that for thinner
12 Material Science and Engineering with Neutron Imaging 215
Fig. 12.4 The effect of EPS foam pattern thickness on degradation; 4-mm thick plate (top) and
24-mm thick (bottom)
foam patterns, one does not see diffusion pathways, as was noted for thicker
foam, and often thicker castings have defects from incomplete pyrolysis of the
foam in certain regions. Data of this nature are of profound significance for
understanding the physics of the problem related to multiphase dynamic flow and
transport. These data can be used to develop suitable engineered solutions to
produce near-net-shaped castings with lower scrap rate. The spatial and temporal
information on a given phase during casting solidification is valuable for model-
ing and verification of the dynamic process using computational fluid dynamic
(CFD) codes such as Flow3D1 and Magma1.
relatively easy to identify individual grains; the particles shown are mostly
smaller than 800 mm. The sand core, obtained from a foundry, was produced
by a specialized technique which utilizes a vaporized or gaseous chemical that
operates as a catalyst or co-reactant in order to cure (polymerize) a resin binder
film. The resin is spread uniformly over the surface of the sand and compacted
into a mold (rectangular in the study illustrated). When the coating has fully
cured, each sand grain is firmly joined to the others in contact with it. An
understanding of the stability and durability of sand cores is important for
the foundry industry, but a realistic fundamental model of core behavior
from particle scale mechanics and numerical modeling is still a dream.
Neutron imaging offers the unique possibility to provide actual information
on particle contacts and voids with appropriate resolution. Using suitable
thresholding techniques with advanced image processing algorithms, it will
in the future be possible to describe the three-dimensional void and solid
particle fabric and its evolution as a function of externally applied mechan-
ical or thermal stress.
Many practical issues, such as the fate of contaminants in the sub-surface of
the earth, or of nutrients for plant systems (Chapter 17), can benefit from
information that is uniquely obtainable from neutron imaging. Using H2O
and D2O alternately, or in clever miscible modes, neutron imaging experiments
can be designed to extract information on porous media filled with fluid and gas
phases, and the arrangement of solid phases, with relatively good precision and
ease. They can also provide much needed information for the study of partially
saturated porous media and suction characteristic curves that are presently
unresolved.
The micromechanical analysis of granular assemblies using computer
simulations, through numerical methods such as the distinct element
method (DEM), can be evaluated by comparing the results for discrete
particulate model systems with carefully controlled neutron imaging
experiments that involve a mechanism to apply a known state of stress
or strain using a portable loading system. For simulations, the material
microproperties such as particle shear and tangential stiffness can be
evaluated from nanoindentation test results on individual particles. The
particle contact behavior, together with the nature of force chain structure
and other related micromechanics associated with one-dimensional com-
pression on granular materials having different shape and mechanical
properties, is then analyzed in conjunction with experimental data gener-
ated using both neutron diffraction and imaging. The goal of the DEM is
not to match the macroscopic response but to use the fundamental proper-
ties of the particles to predict the behavior of the assembly, and to
evaluate the basic mechanisms contributing to the strength and deforma-
tion behavior of particulate systems in terms of particle force-chain struc-
ture (Fig. 12.6).
218 D. Penumadu
Fig. 12.6 Example force distribution of a granular assembly under a punch obtained using a
particle flow code
There are some common engineering applications that require continuous fast
imaging with neutrons to obtain detailed information related to a fast process.
To solve such problems, major difficulties must be overcome related to the low
number of neutrons detected in one measurement time frame and the number of
pixels that can be read. To meet this challenge requires a high-flux neutron
source, low decay time neutron scintillation materials, and the use of strobo-
scopic imaging for fast but periodic processes. Using the stroboscopic
approach, the number of neutrons in one time window is still very low, but
many exposures of the same time window of a periodic process may be accu-
mulated on the detector before readout, thus increasing the available intensity
(though not improving the signal-to-noise ratio).
The stroboscopic technique can be illustrated by a study of fuel injection and oil
flux in a combustion engine. A commercial combustion engine was stripped of the
intake and exhaust system and coupled to a 2 kW electric motor with a transmis-
sion belt [13]. The spark plugs were also removed, so that the engine would not
generate compression in its cylinders. The electric motor was capable of turning the
engine at low rotation speeds and, due to the reduced heat production, the water
cooling could be omitted. A Hall sensor close to the transmission wheel on the
camshaft generated a pulse which was used to trigger the image intensifier of a
cooled CCD camera. The trigger signal could be delayed in order to shift the time
window within the cycle of the engine. Measurements were made at two facilities:
the Neutrograph facility at the Institut Laue-Langevin (ILL) reactor in Grenoble
with a high thermal neutron flux of 3 109 n cm–2 s–1and low L/D value of 170,
12 Material Science and Engineering with Neutron Imaging 219
and the ANTARES facility at the FRM-II reactor of the Technische Universitat
Munchen with a lower thermal neutron flux of 1108 n cm–2 s–1 but a high L/D
value of 400. At ILL, the detection system was a MCP-intensified CCD camera
with a Peltier-cooled chip (13001024 pixels) with 16 bit digitization.
The full cycle of the four-stroke engine running at 1000 rpm was split into 120
individual frames over two rotations; 150 individual images were recorded as an
on-chip accumulation of 200 ms exposure each. In this way, the total exposure
time for the full run was only of order 18 min. The field of view was 2424 cm.
At FRM-II the measurements were repeated with an Andor MCP intensified
CCD camera with 10241024 pixels. Because of the lower available flux, the
time window was extended to 1 ms and the rotation speed was reduced to
600 rpm. The images thus show more motional blurring but better definition
of the stationary parts. Figure 12.7 shows typical results that demonstrate the
differences in spatial and temporal resolution for two facilities related to strobo-
scopic imaging. As can be seen, the lubrication of a running engine can be
visualized; this can have important applications related to the design and
modeling of a future automotive powertrain.
Possible options for detection and triggering systems for short-time strobo-
scopic neutron imaging include LCD shutters, interline CCDs, CCDs with a
gated image intensifier, and electronically shuttered CCDs [14]. Stroboscopic
neutron radiography with a time resolution down to a few microseconds, as well
as dynamic tomography on a sub-minute time scale, can allow the visualization
of thick and strongly absorbing materials [15], thus providing the ability to also
investigate large samples with extremely low contrast as required in some fields
of engineering.
Fig. 12.7 Dynamic radiography of a running engine, (a) acquired at the Neutrograph facility
and (b) acquired at the ANTARES facility [13]
220 D. Penumadu
The requirements for dynamic imaging are clearly more severe since the
number of neutrons recorded in a single time frame is very low, even for the
highest flux sources, so a low background is extremely important as well as a
fast frame rate. A recently developed prototype detector for this type of appli-
cation is based on electron multiplying charge-coupled device (EMCCD) tech-
nology [16]. This CCD technology evolved from the original development of
the high-resolution -ray imaging detector by RMD [17], which uses a back-
illuminated, thermoelectrically cooled, 512 512 pixel EMCCD as a readout
sensor coupled to a LiF/ZnS:Ag scintillator. The scintillator screen is 225 mm in
thickness and provides bright emission for short exposure times (2 ms) with
an imaging area of 25 25 mm2 and an effective pixel size of 50 mm.
The camera permits a minimum of 2 ms data acquisition and can provide
30 frames per second (fps) imaging at full pixel resolution. In binned mode,
the frame rate can be increased to 225 fps. While the internal gain effectively
offsets read noise, reducing it to < 1 e– pixel–1 s–1, the dark noise is also
minimized to < 1 e– pixel–1 s–1 by cooling the EMCCD to –358C.
Using this detector, dynamic imaging in integrated mode was conducted at
the National Institute of Standards and Technology reactor facility using the
thermal neutron port BT2. Figure 12.8 shows neutron images of water droplets
from a plastic bottle with a small hole, imaged at 30 fps to compensate for the
low flux and relatively low detection efficiency of the LiF/ZnS:Ag screen. It is
important to consider the scattering effects for such images using tools such as
Fig. 12.8 Dynamic imaging (single frames) of a water droplet using the EMCCD detector
12 Material Science and Engineering with Neutron Imaging 221
the Monte Carlo method to extract contrast from linear attenuation asso-
ciated with absorption versus scattering.
The ability to use EM CCD technology now opens up many possibilities for
pseudo real-time neutron imaging applications that require the study of flow or
transport of fluids through porous media or channels. The following are examples:
optimization of the channel geometry for microfluidic-based devices for
thermal management of high-performance electronic chips and related
applications,
optimization of catalytic converters,
understanding of physico-thermal issues related to the conversion of low-
yield hydrocarbon-rich geo-materials for next-generation oil extraction,
study of the environmental degradation of polymeric composites due to
moisture and humidity exposure, and
development of novel drainage materials for hydraulic management and
filtration applications.
dc ðx; y; zÞ d0
"c ðx; y; zÞ ¼ ; (12:1)
d0
222 D. Penumadu
The stress tensor is then derived from the strain tensor using Hooke’s law.
A particular advantage of neutrons for residual stress imaging is the ability
to penetrate deep into large components, nondestructively, as illustrated in
Figs. 12.9 and 12.10. Hard (high energy) X rays can now offer similar penetra-
tion (depending on the material) but the relevant diffraction angles are then
very small, so the gauge volumes are ‘‘needle-like,’’ which is problematic for
Fig. 12.9 Wingspar from an Airbus A340 aircraft mounted for stress imaging on the Engine-X
diffractometer at the ISIS spallation neutron source
12 Material Science and Engineering with Neutron Imaging 223
Fig. 12.10 Image of the stress distribution in the head of a railway rail (see e.g., [19])
measuring one component of the stress. The two techniques are therefore highly
complementary and are increasingly used in combination.
Fig. 12.11. Image of the residual elastic strain distribution around a cold expanded hole in a
12-mm thick steel plate. The measured strains correspond to the through-thickness average of
the out-of-plane strains. The in-plane spatial resolution 22 mm2. [20]
In practice there can often be a combination of edge shift and height change,
averaged over the crystallites within the gauge volume, leading to significant
broadening and distortion of the sharp Bragg edges. Two samples of the same
material can therefore have quite different transmission spectra, which can be used
as a signature of the method used to create the samples, as illustrated in Fig. 12.12.
The Bragg edges also obviously differ between the different materials that
might make up a macroscopic system (Fig. 12.13). The ratio of images mea-
sured below and above a Bragg edge for one of the component materials gives a
good contrast for that material, while the other material effectively disappears
(provided that it does not also have a Bragg edge in the same wavelength
region).
In principle, full tomographic imaging can be carried out as a continuous
function of wavelength, with the possibility to, then, reconstruct images with
highlighted contrast to illustrate stress, texture, composition, or even complex
combinations. Currently, the available neutron fluxes and detector technolo-
gies mean that the resolution has to be relaxed in some respect to achieve
reasonable images; for example, either positional resolution is relaxed to
achieve high time resolution for high Bragg edge contrast, as in Fig. 12.13,
or wavelength resolution is relaxed to achieve higher spatial resolution, as in
Fig. 12.14. However, with the advent of higher-flux neutron sources, such as
12 Material Science and Engineering with Neutron Imaging 225
Fig. 12.13 Neutron transmission images for a copper cored iron slug, with Bragg edge
filtering used to enhance the contrast for either copper or iron [22]
226 D. Penumadu
the Spallation Neutron Source in the United States, and/or the design of
specialized instruments using advanced neutron optics and detectors, such
as the IMAT instrument proposed for the ISIS source in the United Kingdom,
these restrictions are likely to be overcome within the next 5 years. Then, this
will open the way for a much broader range of applications of neutron
imaging in materials science and engineering.
References
1. J. Campbell, The concept of net shape for castings. Mater. Design 21(4), 373–380 (2000).
2. X. Yao and S. Shivkumar Mould filling characteristics in lost foam casting process. Mater.
Sci. Technol. 13, 841–846 (1997).
3. R. E. Turner, Radiography: Major contribution to the metalcasting industry. Modern
Casting 61(1), 44–45 (1972).
4. B. Sirrell, M. Holliday, et al., The Benchmark Test 1995. 7th Conference on Modelling of
Casting. Welding and Advanced Solidification Processes, London, UK.
5. M. Cox, R. A. Harding, et al., Optimised running system design for bottom filled alumi-
nium alloy 2L99 investment castings. Mater. Sci. Technol. 19(5), 613–625 (2003).
6. X. Luo. Study on Infrastructure Materials by Using Neutron Radiography and Diffraction.
PhD dissertation, University of Tennessee, Knoxville, TN, USA (2007).
7. X. Luo and D. Penumadu, Visualization of the multiphase flow in the lost foam casting
process by neutron radiography and image processing. Automotive Light Metal Castings:
Technology and Applications, Materials Science and Technology (MS&T), Detroit, MI,
USA (2007).
8. D. Penumadu and N. Kardjilov, High resolution neutron tomography of sandcores used in
metal casting (2008). To be submitted.
12 Material Science and Engineering with Neutron Imaging 227
9. T.S. Majmudar and R.P. Behringer, Contact force measurements and stress-induced
anisotropy in granular materials. Nature 435(7045), 079–1082 (2005).
10. N.P. Kruyt, Contact forces in anisotropic frictional granular materials. Int. J. Solids
Struct. 40(13–14), 3537–3556 (2003).
11. M. Otto et al., Anisotropy in granular media: Classical elasticity and directed-force chain
network. Phys. Rev. E 67(3), 031302-1 to 24 (2003).
12. P. Lehmann, P. Wyss, et al., Tomographical imaging and mathematical description of
porous media used for the prediction of fluid distribution. Vadose Zone J. 5(1), 80–97
(2006).
13. B. Schillinger, J. Brunner, and E. Calzada, A study of oil lubrication in a rotating engine
using stroboscopic neutron imaging. Physica B 385–386, 921–923 (2006).
14. B. Schillinger, H. Abele J. Brunner, G. Frei, R. Gahler, A. Gildemeister, A. Hillenbach,
E. Lehmann, and P. Vontobel, Detection systems for short-time stroboscopic neutron
imaging and measurements on a rotating engine. Nucl. Instrum. Meth. Phys. Res. A 542,
142–147 (2005).
15. A. Hillenbach, M. Engelhardt, H. Abele, and R. Gahler, High flux neutron imaging for
high-speed radiography, dynamic tomography and strongly absorbing materials. Nucl.
Instrum. Meth. Phys. Res. A 542, 116–122 (2005).
16. V.V. Nagarkar, I. Shestakova, S. C. Thacker, S. R. Miller, D. Penumadu, J. F. Ankner,
H. Z. Bilheux, and C. E. Halbert, Time-resolved high resolution neutron imaging studies
at the ornl spallation neutron source. Session N14-4, Nuclear Science Symposium,
Honolulu, Hawaii, USA, Oct. 27-Nov. 3 (2007).
17. V.V. Nagarkar, I. Shestakova, V. Gaysinskiy, S.V. Tipnis, B. Singh, W. Barber,
B. Hasegawa, and G. Entine, CCD-Based Detector for SPECT. IEEE Trans. Nucl. Sci.
53(1), 54–58 (2006).
18. M.T. Hutchings, P. J. Withers, et al., Introduction to the Characterization of Residual
Stress by Neutron Diffraction. Boca Raton, CRC Press (2005).
19. P. J. Webster, X. Wang, G. Mills, G. A. Webster, Residual stress changes in railway rails.
Physica B 180–181, 1029 (1992).
20. J.R. Santisteban, L. Edwards, M.E. Fizpatrick, A. Steuwer, and P.J. Withers, Appl. Phys.
A 74, S1433–S1436 (2002).
21. J.R. Santisteban, S. Siano, and Daymond, M.R, Neutron strain scanning of archaeolo-
gical bronzes, Mater. Sci. Forum 524–525, 975–980 (2006).
22. W. Kockelmann, G. Frei, E.H. Lehmann, P. Vontobel, and J.R. Santisteban, Energy-
selective neutron transmission imaging at a pulsed source. Nucl. Instrum. Meth. Phys.
Res. A 578, 421–434 (2007).
Chapter 13
Novel Neutron Imaging Techniques for Cultural
Heritage Objects
Abstract The use of neutrons for cultural heritage (CH) research is illustrated
with special reference to neutron tomography (NT) methods, providing three-
dimensional (3D) images of neutron attenuation, and the analysis techniques
known as prompt gamma-ray activation analysis (PGAA) and neutron reso-
nance capture analysis (NRCA), providing the elemental composition of
an object. PGAA and NRCA are well-established nondestructive methods
for bulk analysis of CH objects, with sensitivities that can reach the
parts-per-million range. By improving the spatial resolution of PGAA and
NRCA it will be possible to measure the composition of small parts inside a
large object or even to provide a full 3D map of the elemental composition of
an artifact. The imaging techniques under development are called prompt
gamma-ray activation imaging (PGAI), neutron resonance capture imaging
(NRCI) and neutron resonance transmission (NRT) tomography. The
NRCA experience at the GELINA neutron source is the starting point for
the development of NRCI/NRT now taking place at the 100 times more power-
ful ISIS pulsed neutron source.
PGAI, NRCI, and NRT are unlikely to achieve the spatial resolution
obtained by conventional NT. It is the combined use of these new imaging
techniques with neutron and X-ray tomographies that is proposed here as the
way forward in neutron imaging. There are parallel developments in the use of
Bragg edges to improve the elemental sensitivity of NT (dichromatic tomogra-
phy). Another foreseen development is 3D phase imaging (neutron diffraction
imaging, NDI), which can essentially be undertaken using existing diffraction
instruments. The ongoing developments are promising and suggest that neu-
tron-based techniques will provide the basis for integrated analysis protocols
leading to important advances in the scientific characterization of the materials
that constitute a CH object.
C. Andreani (*)
University of Milano-Bicocca, Department of Materials Science, Via Roberto
Cozzi 53, 20125 Milano, Italy
The use of X-ray and neutron facilities for cultural heritage (CH) research is
rapidly growing [1, 2]. Detailed three-dimensional (3D) maps of neutron and
X-ray attenuation in CH objects are produced by the tomographic methods
described earlier in this book. X-ray and neutron facilities also provide a
number of nondestructive analytical techniques that are used in the CH
field. X-ray fluorescence (XRF) is a well-known example, in particular
because of the widespread use of portable XRF devices, a clear advantage
for analysis of valuable objects at (or close to) museums. More generally, the
application of nuclear-based techniques to CH research has expanded in recent
years as laboratories with small ion accelerators have been installed close to
museums, facilitating the study of CH objects by particle-induced X-ray and
gamma-ray emission (PIXE and PIGE) and Rutherford backscattering [3].
With XRF and PIXE, imaging is possible down to micrometer resolution; but
because both techniques are based on the detection of low-energy X-rays, only
information on the surface of an object is retrieved. Bulk analysis techniques are
alternatively available (see below) but do not offer a spatial resolution adequate
for measuring the composition of small parts inside a large object well. Ideally,
one would like to combine the spatial resolution of tomographic methods with the
power of nondestructive analytical techniques to provide a 3D map of the
elemental composition of a large object. Can neutrons be used for such a challen-
ging task? The question is addressed in this chapter by first reviewing some general
features of neutron (and X-ray) techniques as relevant for CH studies.
X-rays and neutrons conveniently complement each other in CH research. This
is because CH objects can be made from diverse materials sampling the Mende-
leev table at both low and high atomic numbers, making the attenuation of X-rays
and neutrons very different. X-rays have a lower penetration depth at high atomic
numbers. Thus X-ray radiography (XR) is suitable for probing the interior of
objects made from wood or bone, but it has problems with bulky metals such as
gold, silver, and lead, which are nearly opaque to X-rays. On the other hand,
neutron radiography (NR) can probe metal artifacts (e.g., copper, tin, iron,
bronze, lead) but does not penetrate thick layers of organic materials because of
strong neutron beam attenuation by hydrogen. An introduction to NR and its
applications can be found elsewhere in this book.
Neutron tomography (NT) is performed by taking a large set of radiographs
of an object at different orientations and reconstructing all data offline in a
virtual 3D image. For further details, see Chapters 6 and 7 in this book. Only a
few NT setups exist in the world (see [4] for a review of facilities in Europe)
because of the high neutron fluxes required to perform a full scan of an object in
13 Novel Neutron Imaging Techniques for Cultural Heritage Objects 231
Fig. 13.2 (a) Neutron radiograph of a spark plug at 6.9 Å; (b) result of the division of two
radiographs, one at 6.9 Å and the other at 3.2 Å, showing that the steel cladding (central part)
became transparent; (c) 3D result of dichromatic tomography. Detailed explanations can be
found in [5]
13 Novel Neutron Imaging Techniques for Cultural Heritage Objects 233
Bragg edges may be more useful for CH applications because of the texture
information embedded in their detailed shapes. Texture in the phases distorts
the Bragg edges, so it can be used as a fingerprinting technique with the
potential for texture mapping [8]. Appropriate transmission detectors need to
be developed before this technique can be used routinely at pulsed neutron
sources. On the other hand, phase-sensitive neutron diffraction (ND) tech-
niques are in regular use for phase and microstructure characterization of
ceramic and bronze artifacts, including texture analysis (see below). Texture
provides important clues to the deformation history and thus to historic pro-
duction steps, a prominent example being the neutron texture analysis of the
early Copper Age Iceman axe [9].
Three features of neutron interactions with nuclei are exploited to provide
analytical tools to determine the elemental compositions of CH objects:
1. radioactive nuclei produced by thermal/cold neutron capture,
2. prompt gamma rays from thermal/cold neutron capture, and
3. resonant neutron absorption at element-specific neutron energies in the
epithermal energy range.
The first feature has long been exploited by a method known as neutron
activation analysis (NAA). It started with the detection of radioactive isotopes
through their half-lives. Now, these isotopes are observed and quantified with
high-resolution gamma-ray detectors on the basis of gamma-ray energies as
well as gamma-ray half-lives. This technique, known as instrumental neutron
activation analysis (INAA), is now a valuable analytical tool for CH studies in
use at many facilities [10, 11]. INAA may be invasive because it may require the
234 C. Andreani et al.
Neutron resonance capture analysis makes use of the unique resonance absorp-
tion properties of epithermal neutrons for analyzing the bulk composition of
materials and objects. The absorption cross sections of most elements are well
known; they vary from one element to another and actually vary between
isotopes of the same element [16]. Neutron absorption is followed by the prompt
emission of a gamma-ray cascade with total cascade energies of up to about
8 MeV. The detailed energy spectrum of the gamma emission is also well known
in the case of thermal neutron capture, but it can vary among resonances. NRCA
is used at the GELINA pulsed neutron source for analysis of CH artifacts to
determine the concentration of neutron-sensitive elements [17–25].
So far NRCA has been applied to a large number of artifacts, mainly bronze
objects. A typical NRCA spectrum looks like the one in Fig. 13.4 [25] obtained
by exposing an Etruscan votive object to the GELINA neutron beam.
A number of resonance peaks of different elements (copper, lead) are clearly
visible, corresponding to neutron energies in the range from 3,250 to 3,650 eV.
The energies and intensities of the resonance peaks characterize the elemental
composition of the sample. Figure 13.4 shows the 3,357 eV resonance of
lead-206 and three copper resonances at 3,310, 3,503, and 3,588 eV. Because
1
Ancient Charm (Analysis by Neutron Resonant Capture Imaging and other Emerging
Neutron Techniques: Cultural Heritage and Archaeological Research Methods), http://
ancient-charm.neutron-eu.net/ach.
236 C. Andreani et al.
Fig. 13.4 Example of neutron resonance capture analysis NRCA spectrum featuring three
copper resonances and one lead resonance in the region of 3,250–3,650 eV used to determine
the Pb/Cu weight ratio of an Etruscan votive object [25]
these four resonances are weak, the Pb/Cu weight ratio can be obtained after
small self-shielding corrections. A set of Etruscan statuettes from a collection
originally owned by Earl Corazzi of Cortona (Tuscany), now at the National
Museum of Antiquities in Leiden (Netherlands), are among the objects inves-
tigated by NRCA. In this case, NRCA was used to distinguish suspected fakes
from genuine statuettes, solely on the basis of the elemental compositions of the
artifacts [20]. The conclusion was based on the observation that minor quan-
tities (up to several percent) of zinc occurred in some of these statuettes. Given
the melting techniques available to the Etruscan blacksmiths, no more than a
fraction of a percent of zinc would occur in a genuine artifact (Fig. 13.5).
Fig. 13.5 Photograph of two Etruscan statuettes, one genuine and one false [20]
Fig. 13.6 Nuclear processes involved in prompt gamma-ray activation analysis (PGAA) and
neutron activation analysis (NAA): neutron capture is immediately followed by emission of
prompt gamma rays. The compound nucleus decays by beta emission followed by the emis-
sion of delayed gamma rays
usual prompt gamma-ray measurements in that the gamma measured is not due to
the (n, ) reaction. Indeed, boron (like lithium-6) reacts with neutrons by emission
of -particles [i.e., via the reaction 10B(n, )7Li]. Most of the lithium-7 is formed in
an excited state and de-excited instantaneously (within 10–14 s) by the emission of
477 keV gamma rays.
The PGAA technique requires mainly a source of thermal or cold neutrons
and a high-resolution spectrometer for measurement of gamma rays with
energies over a range from about 100 keV to 11 MeV. The energies of the
prompt gamma rays identify the neutron-capturing elements, while the inten-
sities of the peaks at these energies are used to determine their concentrations
(Fig. 13.7). An advantage of the PGAA method is that nothing special is
required in the way of sample preparation: the sample can be in a solid, liquid,
or gaseous state. Depending on the amount of the investigated material,
Fig. 13.7 Example of prompt gamma-ray activation analysis (PGAA) spectrum from an
estuarine sediment
13 Novel Neutron Imaging Techniques for Cultural Heritage Objects 239
element, and target matrix, the detection limit for a PGAA installation can be as
low as 10 ppb (for B, Sm, Gd). The activity of the object after irradiation is low
or decreases rapidly in most cases; therefore, the object can be returned within a
few days of the measurement. Compared with NAA, where the sample is placed
close to the core of the reactor, the neutron beam intensity at the sample is six
orders of magnitude lower.
A typical setup for PGAA consists of a sample box, a shielding, and a
detection system [13, 27–30]. The object is exposed to an intense thermal or,
preferably, cold neutron beam on the order of 106–108 n/cm2s, originating from a
neutron guide. The neutron guide is preferably curved to avoid a direct view to
the moderator or cold neutron source, thus reducing the fast-neutron and
gamma-ray background. The huge number of peaks in a spectrum (up to about
a thousand) requires a high-purity germanium (HPGe) detector with high-energy
resolution (<2 keV at 1,332 keV). Moreover, many high-energy gamma rays will
not deposit their full energy inside the HPGe detector and thus build up large
Compton tails that will hide peaks at lower energies or generate (unwanted)
additional escape peaks. To reduce these effects, the HPGe detector is sur-
rounded by a fast gamma-ray detector, usually made of an annular bismuth
germanate oxide (BGO) or NaI(Tl) scintillator that serves to reject events simul-
taneously detected in the HPGe detector and the scintillator. This anti-Compton
shield acts as an active shield against unwanted, external high-energy gamma rays
that deposit part of their energy in the scintillator before scattering in the HPGe
detector. The whole detection system is then covered with lead bricks and addi-
tional neutron shielding to further improve the signal/background ratio. The
sample box is usually covered for the same reasons with lithium-6, except at the
entry of the neutron beam, to prevent scattered neutrons from reaching the
detector. The detection system is normally placed at 908 to the neutron beam.
Fig. 13.8 Main composition of 53 brooches studied by S. Baechler et al. [31] (left) and a
photograph of one of them (right)
240 C. Andreani et al.
The count rate of the acquisition system is limited to about 10 kHz by adjusting
the neutron beam size with lithium-6 collimators.
PGAAs have been performed on various samples, in a wide range of fields,
including material science, medicine, nuclear science, geology, and archeology.
In one of the first CH applications, S. Baechler and coworkers [31] studied 53 La
Tene and Roman brooches from western Switzerland and identified four dif-
ferent groups of copper-based alloys (Fig. 13.8).
Fig. 13.9 (a) X-ray radiograph at 87 keV of a piece from the Oklo natural reactor (Gabon);
(b) 3D distributions of lead and uranium resulting from tomographies taken at the high-
energy beam line (ID15) of the European Synchrotron Radiation Facility (Grenoble, France)
[7]; (c) spatial distributions of selected elements in the same piece, obtained by prompt
gamma-ray activation imaging at the Paul Scherrer Institut (Villigen, Switzerland) [32]
2
Forschungsreaktor München II (translation: Munich Research Reactor II), Munich
Technical University.
13 Novel Neutron Imaging Techniques for Cultural Heritage Objects 243
neutrons to a spot size of 4 7 mm2 at 10 cm from the end of the guide. The cold
neutron flux is expected to exceed 1010 n/cm2s at the focal point [35]. One of the
aims of the facility is to test different PGAI geometry options on CH objects.
The NRCA experience at the GELINA neutron source [17–25] is the starting
point for the development of an imaging method based on resonant neutron
capture. Imaging applications require high neutron fluxes, and the 100 times
more powerful ISIS pulsed neutron source (Chilton, UK) is now used for the
development of NRCI and NRT [15]. Figure 13.10 shows a sketch of the NRCI/
NRT setup that is under construction at ISIS. In the case of NRCI, the object is
scanned in front of a collimated neutron beam while a set of scintillator
detectors surrounding the object record the capture gamma rays. The composi-
tion of the object must be retrieved by a tomographic reconstruction corrected
for neutron self-absorption. In practice the spatial resolution is limited by the
need to collimate the beam of epithermal neutrons while keeping the signal/
background at an acceptable level. The best compromise at the moment seems
400 mm
Gamma
400 mm detectors
Collimator
300 mm
Fig 13.10 Sketch of a setup for performing neutron resonance capture imaging and neutron
resonance tomography. The sample is attached to a frame allowing translation and rotation
motion relative to the beam and detectors, which are fixed. The gamma-ray detectors for
NRCI are placed around the sample. A position-sensitive detector for NRT is placed in the
neutron beam behind the sample
244 C. Andreani et al.
In the past few years, neutron techniques have increasingly been used for the
study of CH materials, in particular to address their nature and authenticity,
their provenance and diffusion, the manufacturing techniques used in their
fabrication, and their state and conservation [37]. In the following discussion,
we show first results from the AC project [38–41] and examples of applications
13 Novel Neutron Imaging Techniques for Cultural Heritage Objects 245
of thermal and cold neutron tomographic techniques to the study of phase and
microstructure characterization of ceramic artifacts, ancient bronzes, and mar-
ble artifacts. A common feature in the selected examples is that the use of
neutrons was motivated either by the general aim of developing new imaging
techniques or by the need to perform a nondestructive analysis. Cast and
restored metal objects were inspected to elucidate the production technology
and the corrosion damage. Marble and ceramic artifacts were inspected to
probe the presence of adhesives, glued parts, epoxy filler materials, or epoxy
protective coatings.
For testing purposes, the AC archaeologists prepared ‘‘black boxes.’’ These
closed metal cubes contain objects of various shapes, forms, and elemental
composition representative of the composition of real archaeological objects.
The task was to recognize the elemental composition and the rest–strain of the
inner objects by PGAI, NRCI, and NDI [38–41]. Neutron and X-ray radio-
graphy and tomography were used to visualize the inner content of the black
boxes and to help with targeting the coordinates of the interesting spots for the
new 3D neutron techniques. NT was performed at the ANTARES3 facility [42] at
FRM-II, and the X-ray tomographies were performed at the Centre for X-ray
Tomography at the Ghent University [43]. Figure 13.11 shows a 3D visualization
(tomograph) of the reconstructed X-ray data (a) and the corresponding neutron
data (b) for one of the black boxes. Using solely the image content, it can be seen
that neither X-ray tomography nor NT alone allows the derivation of both types
of materials and the elemental composition of the scanned object. However, the
Fig. 13.11 X-ray tomograph (a) and neutron tomograph (b) of black box H-IX from the
Ancient Charm project [39]
3
Advanced Neutron Tomography and Radiography Experimental System.
246 C. Andreani et al.
combined use of the two techniques allows areas of differing materials and their
shapes to be distinguished and separated. At a later stage, to determine precisely
the elemental composition of these objects, PGAI analysis was performed
directly at the coordinates provided by the NT image. Figure 13.12 shows an
X-ray tomograph and the corresponding neutron tomographs for another black
box. Here three separate layers of materials can be visually distinguished. The
upper row shows corresponding horizontal slices of the object (looking ‘‘from
above’’); the bottom row shows vertical slices (looking ‘‘from the side’’). A
suitable isocontour of the X-ray slice helps in visualizing the initial misalignment
of the NT image and its correct alignment after the registration. This example
demonstrates in various ways the value added by the registration and the sub-
sequent point-to-point comparison.
The neutron tomograph in Fig. 13.13 refers to laminated bronze artifacts
from rich sepulchral complexes in the necropolis of Osteria-Poggio Mengarelli
and Cavalupo (eighth century B.C.) [44]. Quantitative phase analysis was per-
formed by neutron diffraction at the INES4 diffractometer installed at the ISIS
neutron source. This result was complemented by the analysis of the spatial
distribution of the components, obtained with an NT experiment (Fig. 13.13),
Fig. 13.12 (a) X-ray tomograph of black box D-VIII from the Ancient Charm project [39];
corresponding neutron tomograph in the initial position (b) and after automatic registration (c) [38]
4
Italian Neutron Experimental Station, a collaborative project between ISIS and the Italian
National Research Council [Consiglio Nazionale delle Ricerche].
13 Novel Neutron Imaging Techniques for Cultural Heritage Objects 247
α-spacing
(b)
006 3 006 18 006 22
Fig. 13.15 Example of neutron tomography of marble fragments from Villa Adriana [48]
Acknowledgment This work was performed with financial support from the European Com-
munity under contract HPRI-2001-50043 and within the CNR-Council for the Central
Laboratory of the Research Councils (UK) agreement. Carla Andreani and Giuseppe Gorini
acknowledge Consiglio Nazionale delle Ricerche (CNR), Italy, for financial support for the
experiments performed at ISIS. Thomas Materna acknowledges financial support from the
250 C. Andreani et al.
Swiss National fund, the University of Fribourg (Switzerland), and the University of Cologne
(Germany) for the experiments performed at PSI and the European Synchrotron Radiation
Facility. We also wish to thank Giulia Festa for critical revision of the text.
References
1. Portal (News and developments from the CCLRC). Looking forward to the past: science
and heritage. Special issue Autumn/winter (2006). http://www.scitech.ac.uk/Publications/
nl/Portal/.
2. ESRF Newsletter. Science and art. 44 (December 2006).
3. M. Menu et al., The dedicated accelerator-based IBA facility AGLAE at the Louvre,
Nucl. Instrum. Meth. Phys. Res. B 45, 610–614 (1990).
4. E.H. Lehmann, P. Vontobel, B. Schillinger, T. Bücherl, S. Baechler, J. Jolie, W. Treimer,
R. Rosa, G. Bayon, S. Legoupil, S. Körner, V. Micherov, M. Balasko, in Proceeding of
the 15th World Conf. on Non-Destructive Testing, Rome, Italy, October 15–21, (2000).
Available from http://www.ndt.net/article/wcndt00/papers/idn804/idn804.htm.
5. N. Kardjilov, S. Baechler, M. Bastürk, M. Dierick, J. Jolie, E. Lehmann, T. Materna, B.
Schillinger, P. Vontobel, Nucl. Instrum. Meth. Phys. Res. A 501, 536 (2003).
6. NEUTRA: Thermal neutron facility for imaging at the Paul Scherrer Institute, Villigen,
Switzerland http://neutra.web.psi.ch/facility/index.html.
7. T. Materna et al., Uranium-sensitive tomography with synchrotron radiation, J. Syn-
chrotron Rad. 6, 1059–1064 (1999).
8. W. Kockelmann, G. Frei, E.H. Lehmann, P. Vontobel and J.R. Santisteban, Nucl.
Instrum. Meth. Phys. Res. A 578, 42 (2007).
9. G. Artioli, M. Dugnani, I. Angelini, L. Lutterotti, A. Pedrotti, A. Fleckinger, Proc.
Archaeometallurgy Europe 2, 19–27, (2003); G. Artioli, M. Dugnani, T. Hansen, L.
Lutterotti, A. Pedrotti, G. Sperl, in: A. Fleckinger (ed.), La mummia dell’età del rame. 2.
Nuove ricerche sull’uomo venuto dal ghiaccio, Vol. 3, Folio Verlag, Bolzano, 9–22, (2003).
10. E. Ciliberto and G. Spoto (eds.) Modern Analytical Methods in Art and Archaeometry,
John Wiley & Sons, Inc. New York (2000).
11. D.R. Brothwell and A.M. Pollard (eds.) Handbook of Archaeological Sciences, John
Wiley & Sons, Inc., New York (2001).
12. C.O.Fischer et al, Neues zur Neutronen-Aktivierungs-Autoradiographie, Restauro 105,
426 (1999) (in German).
13. Zs. Révay and T. Belgya, Principles of the PGAA method, in G.L. Molnár ed., Hand-
book of Prompt Gamma Activation Analysis with Neutron Beams, Kluwer Academic
Publishers, pp. 1–305 (2004).
14. H. Postma, M. Blaauw, P. Bode, P. Mutti, F. Corvi and P. Siegler, Neutron-resonance
capture analysis of materials, J. Radioanal. Nucl. Chem. 248(1), 115–120 (2001).
15. G. Gorini for the Ancient Charm collaboration, Ancient Charm: A research project for
neutron-based investigation of cultural-heritage objects, Il Nuovo. Cimento. C, 30, 47–58
(2007).
16. S.F. Mughabghab et al., Neutron Cross-sections, vol.1, Academic Press, New York
(1981). Available online at http://atom.kaeri.re.kr/cgi-bin/endfplot.pl.
17. H. Postma, M. Blaauw, P. Schillebeeckx, G. Lobo, R. Halbertsma and A.J. Nijboer,
Non-destructive elemental analysis of copper-alloy artefacts with epithermal neutron-
resonance capture, Czech. J. Phys. 63(Suppl. A), A233–A240 (2003).
18. M. Blaauw, H. Postma, P. Mutti, An attempt to date an antique Benin bronze using
neutron resonance capture analysis, Appl. Radiat. Isot., 62, 429–433 (2005).
19. M. Blaauw, H. Postma and P. Mutti, Quantitative neutron capture resonance analysis
verified with instrumental neutron activation analysis, Nucl. Instrum. Meth. Phys. Res. A
505, 508–511 (2003).
13 Novel Neutron Imaging Techniques for Cultural Heritage Objects 251
20. H. Postma, P. Schillebeeckx, and R.B. Halbertsma, Neutron resonance capture analysis
of some genuine and fake etruscan copper-alloy statuettes, Archaeometry 46 635–646
(2004).
21. H. Postma and P. Schillebeeckx, Non-destructive analysis of objects using neutron resonance
capture, memorial issue Gabor Molnár, J Radioanal. Nucl. Chem. 265, 297–302 (2005).
22. R. C. Perego, H. Postma, M. Blaauw, P. Schillebeeckx, A. Borella, Neutron Resonance
Capture Analysis: improvements of the Technique for resonances above 3 keV and new
applications, J. Radioanal. Nucl. Chem. 271(1) 89–94 (2007).
23. H. Postma, R.C. Perego, P. Schillebeeckx, P. Siegler, and A. Borella, Neutron resonance
capture analysis and applications, J. Radioanal. Nucl. Chem. 271(1) 95–99 (2007).
24. H. Postma, D. Fontijn, P. Schillebeeckx, R.C. Perego, and J.J. Butler, Neutron Reso-
nance Capture Analysis of Geistingen axes, LUNULA, Archaeologia protohistorica XIII,
41–46, (2005).
25. H. Postma and P. Schillebeeckx, Neutron-resonance capture as a tool to analyse the
internal compositions of objects non-destructively, Notiziario Neutroni e Luce di Sincro-
trone 11(2), 14–18 (2006).
26. Z.B. Alfassi, C. Chung (Ed.), Prompt Gamma Neutron Activation Analysis, CRC Press,
Boca Raton, Florida, 1995.
27. M. Crittin et al. The new prompt gamma-ray activation facility at the Paul Scherrer
Institute, Switzerland, Nucl. Instr. Meth. Phys. Res. A 449, 221 (2000).
28. Zs. Revay et al., Cold neutron PGAA facility at Budapest, Nucl. Instrum. Meth. Phys.
Res. B 213, 385–388 (2004).
29. H-J Cho et al., Study on prompt gamma-ray spectrometer using Compton suppression
system, Nucl. Instrum. Meth. Phys. Res. B 229, 499–507 (2005).
30. Zs. Revay et al., Construction and characterization of the redesigned PGAA facility at the
University of Texas at Austin, Nucl. Instrum. Meth. Phys. Res. A 577, 611–618 (2007).
31. S. Baechler et al. The k0-method in cold-neutron prompt gamma-ray activation analysis,
J. Radioanal. Nucl. Chem. 256, 239–245 (2003).
32. S. Baechler et al., Non-destructive analysis of a bulky sample from a natural fossil reactor,
J. Radioanal. Nucl. Chem. 250, 39–45 (2001).
33. P. Kudejova et al., On the construction of a new instrument for cold-neutron prompt
gamma-ray activation analysis at the FRM II, J. Radioanal. Nucl. Chem. 265, 221–227
(2005).
34. Research reactor II (FRM-II), Technische Universität München, 2007 http://www.frm2.
tum.de/en/index.html.
35. P. Kudejova, Two new installation for non-destructive sample analysis: PIXE and PGAA,
Ph.D. Thesis, University of Cologne (2005).
36. E. Perelli Cippo et al, Meas. Sci. Technol. 19, 034027 (2008).
37. W. Kockelmann, L.C. Chapon, R. Engels, j. Schelten, C. Neelmejer, H.M. Walcha, G.
Artioli, S. Schalev, E. Perelli Cippo, M. Tardocchi, G. Gorini, and P. Radaelli, Neutrons
in cultural heritage research, J. Neutron Res. 14(1), 37–42 (2006).
38. P. Kudejova, J. Cizek, R. Schulze, J. Jolie, B. Schillinger, K. Lorenz, M. Mühlbauer, B.
Masschaele, M. Dierick, J. Vlassenbroeck, A marker-free 3D image registration for the
Ancient Charm project. Case study with neutron and X-ray tomography datasets, Noti-
ziario Neutroni e Luce di Sincrotrone 12(2), 6–13 (2007).
39. P. Kudejova for the Ancient Charm collaboration, Neutron and X-ray imaging of the
‘black boxes’ for the Ancient Charm project, Archeometriai Mũhely, 35–40 (2008/1).
40. G. Festa, W. Kockelmann, A. Kirfel and the Ancient Charm collaboration, Neutron
diffraction analysis of ‘black boxes’, Archeometriai Mũhely 2008/1 61–72.
41. Z. Kis et al., Prompt gamma activation imaging on ‘black boxes’ in the ‘Ancient Charm’
project, Archeometriai Mũhely. 41–60 (2008/1).
42. ANTARES Facility at the research reactor II (FRM-II), Technische Universität
München (2007). http://www.physik.tu-muenchen.de/antares/.
252 C. Andreani et al.
43. X-ray Radiography and Tomography Facility at the Ghent University (UGCT) (2007)
http://www.ugct.ugent.be.
44. G. Festa, P.A. Caroppi, A. Filabozzi, C. Andreani, M.L. Arancio, R. Triolo, F. Lo Celso,
V. Benfante, S. Imberti, Composition and corrosion phases of Etruscan Bronzes from
Villanovan Age, Measur. Sci. Technol. 19(3) 034004 (7 pp) (2008).
45. A. Filabozzi, A. Pietropaolo, C. Andreani, M. P. De Pascale, G. Gorini,.W. Kockelmann,
L.C. Chapon, Non invasive neutron diffraction analyses of marbles from the ‘‘Edificio
con Tre Esedre’’ in Villa Adriana, Il Nuovo Cimento C 29, 237–252 (2006).
46. A. Filabozzi, C. Andreani, M.P. De Pascale, G. Gorini, A. Pietropaolo, E. Perelli Cippo,
R. Senesi, M. Tardocchi and W. Kockelmann, Texture and structure studies on marbles
from Villa Adriana via neutron diffraction technique, J. Neutron Res. 14, 55–58 (2006).
47. S. Siegesmund and K. Ullemeyer, T. Weiss, and E.K. Tschegg, Int. J. Earth Sci. 89, 170
(2000).
48. R. Triolo, V. Bonfante, and F. Lo Celso, private communication (2007); see also N.
Kardijlov et al, Neutron tomography in modern archaeology, Notiziario Neutroni e Luce
di Sincrotrone 13(2), 6–9 (2008).
49. G. Kühne et al., CNR—the new beamline for cold neutron imaging at the Swiss spallation
neutron source SINQ, Nucl. Instrum. Meth. Phys. Res. A 542, 264–270 (2005); ICON:
Cold neutron facility for imaging at the Paul Scherrer Institute, Villigen, Switzerland,
described in http://neutra.web.psi.ch/facility/index.html.
50. G. Festa for the Ancient Charm collaboration, ANCIENT CHARM: A new project for
neutron-based 3D imaging with applications to CH research. Poster Section (Ist prize
winner): Looking Forward to the Past: Science and Heritage, Tate Modern London
(28 November 2006). Available online at http://ancient-charm.neutron-eu.net/ach/
conferences.
Chapter 14
Probing the Potential of Neutron Imaging
for Biomedical and Biological Applications
Abstract Neutron imaging of biological specimens began soon after the dis-
covery of the neutron by Chadwick in 1932. The first samples included tumors
in tissues, internal organs in rats, and bones. These studies mainly employed
thermal neutrons and were often compared with X-ray images of the same or
equivalent samples. Although neutron scattering is widely used in biological
studies, neutron imaging has yet to be exploited to its full capability in this area.
This chapter summarizes past and current research efforts to apply neutron
radiography to the study of biological specimens, in the expectation that clinical
and medical research, as well as forensic science, may benefit from it.
14.1 Introduction
Fig. 14.1 Neutron radiography of plant tissues demonstrates the sensitivity of neutrons
to hydrogenous materials [3]
Fig. 14.2 An early thermal neutron radiograph of a human chest was limited by multiple
scattering and poor detector spatial resolution [7]
Figure 14.3 shows images of thin tumor specimens: (A) adenocarcinoma, (B)
liposarcoma, and (C) leiomyoma [6]. Thermal neutron images are displayed in
the upper row. The control images for each tumor, 20-kV roentgenograms
(X-ray images), are shown in the lower row. The greater contrast seen in the
neutron images probably arises from differences in hydrogen density, since
many tumors appear to have greater hydrogen content than healthy tissues [6].
Fig. 14.3 Comparison of tumors imaged with thermal neutrons (top) and 20-kV X-rays
(bottom). The contrast in the neutron radiographs is more pronounced because of higher
hydrogen concentrations in tumorous tissue [6]
Fig. 14.4 Radiographs of a rat paw and tail using (A) thermal neutrons and (B) 42-kV X-rays.
Gd2O3 in oil was used as the contrast agent [6]
Fig. 14.5 Radiographs of a rat head using (A) epithermal (i.e., 0.5 to 2 eV) neutrons and (B)
42-kV X-rays. The rat nasopharynx is visible in (A), as indicated by the arrow. Although
In2O3 was used as a contrast agent, the epithermal neutron flux was not sufficient to
demonstrate its contrast capabilities [6]
1990s it was realized that epithermal beams should be more favorable for
BNCT, since body tissues act as a natural neutron moderator. Interest in
clinical applications reawakened, as did interest in thermal and epithermal
imaging of biological tissues.
258 K. Watkin et al.
As with any imaging method, there is a need to understand and evaluate the
benefits and risks of exposing a living subject to neutrons. In order to estimate
the potential radiological hazard that neutron imaging poses to soft biological
tissues, Monte Carlo simulations of neutron exposure of a simulated human
arm were conducted [9]. The conclusions were quite pessimistic. The simula-
tions predicted no effective contrast at given neutron energies of 1 keV, 120 keV,
fission spectrum, and 14 MeV. However, the study provided useful information
on dose deposition between tissue, bone cortex, and marrow. For example,
the calculated neutron dose at 120 keV energy to bone was twice the dose to
contiguous soft tissue.
While epithermal (1 eV–0.1 MeV) and fast (>0.1 MeV) neutrons are required
to penetrate thick tissues, they do not pose a direct radiological hazard. Rather,
hydrogenous tissue moderates these neutrons down to the thermal energy range
(<1 eV) at which they can be captured and thereby cause biological damage.
Fast neutrons also knock off protons from hydrogenous tissues, creating an
additional radiological hazard.
Recently, Monte Carlo simulations for neutron stimulated emission com-
puted tomography (NSECT), an in vivo tomographic spectroscopy using fast
neutrons (Chapter 15), demonstrated that the radiation dose deposited in a
human abdomen or breast using NSECT would be comparable to the dose from
an abdominal CT (computed tomography) exam or a mammogram, respec-
tively [10–12].
Clinical, i.e., small-animal, and pharmaceutical research may benefit the
most from epithermal neutron imaging. The exposure of the specimen is a lesser
concern in those fields, and imaging may aid medical diagnosis without the risk
of radiation exposure.
Recent data on the biological effects of low radiation exposures can be found
in Ref. [13]. Information on the biological effects of fast neutrons on biological
tissues can be found in Refs. [14, 15].
Very recently (Figs. 14.6 and 14.7), a study of rat femurs demonstrated the
relative strengths of X-ray and neutron imaging. Heavy elements (predomi-
nantly calcium and phosphorus in bone) determine X-ray contrast, so the X-ray
images highlight the hard bone framework, with the hydrogenous matter being
essentially invisible. Neutrons, on the other hand, highlight a relatively uniform
distribution of hydrogen throughout the interior and exterior regions of the
bone and do not resolve the filaments of trabecular hard bone running through
the marrow. Interestingly, the neutron images (Fig. 14.7) show an obvious bone
reduction in the decalcified samples – after inorganic compounds are dissolved
by soaking in a 15% ethylene diamine tetra acetic acid solution, as expected
Fig. 14.6 Images made by 40-keV X-ray radiography of (a) calcified bones (i.e., heated at
6008C for 5 h) and (b) natural bones. The distinction between the cortical (surface) and the
trabecular (inner) bone regions is evident [18]
Fig. 14.7 Neutron radiographs of rat bones from (a) decalcified and (b) natural bones [18]
260 K. Watkin et al.
from the egression of mineral material from inside the bone. Post-mortem bone
desiccation was also evaluated as a possible forensic tool [18].
Several recent neutron imaging experiments on biological samples have been
published, e.g., the remains of an ichthyosaur embryo imbedded in a shell [19],
arachnids [20], and insects [21]. Examples of radiographs of other biological
specimens can be found at http://neutra.web.psi.ch/gallery/biological.html.
Adult male porcine lung tissue (1.5 mm thick, 1 cm by 1 cm sample size) immersed
in formalin was successfully imaged using thermal neutrons. These slices were first
extracted from frozen samples from the anatomical region illustrated in Fig. 14.8.
Standard histological techniques were used to allow comparison of neutron
and optical images (Fig. 14.9). After neutron exposure, the samples were stained
using hematoxylin and eosin stain, which is known to provide good resolution
of lung tissue components such as muscle and cartilage.
Although the images in Fig. 14.9 look very similar, the distribution of their
pixel intensities can supply a true quantitative comparison. Recently developed
software [22] enables automatic integration of multiscale volumes from differ-
ent imaging modalities through a cyber-based portal imaging system. It was
found that 92% of the pixel intensities of these two images match, meaning the
pixel-by-pixel intensity between the histological and the neutron images is very
similar. These initial results for fixed porcine lung tissue provide significantly
more detail than any previously reported images taken with a thermal neutron
beam [7]. To compare thermal neutron imaging with X-ray imaging, the
specimen was also imaged using microCT, as shown in Fig. 14.10. The microCT
measurement was conducted before histological staining and subsequent opti-
cal imaging. The gross morphology in all the images is very similar. Cartilage
Fig. 14.8 Approximate location of the thin porcine lung specimen imaged using thermal
neutrons. Only the lung airways are illustrated
14 Probing the Potential of Neutron Imaging 261
Fig. 14.9 Histological and thermal neutron images of the same section of porcine lung, with a
92% pixel agreement (see text)
Fig. 14.10 Multimodal images of the same specimen of porcine lung tissue
supporting the airway is clearly visible in each modality, along with differentia-
tion of the alveolar regions, although the greatest definition is seen in the
histology images.
In a recent study, the anatomical changes in corn kernels affected by Asper-
gillus flavus fungus were analyzed quantitatively using neutron tomography
[23], demonstrating the effectiveness of the technique in identifying the degra-
dation of the biological tissues compared with tissues in uninfected control
kernels. Resistant kernels showed no sign of anatomical change. This interest-
ing research shows the sensitivity of neutron radiography to the effects of
diseases on biological tissues. The most recent and promising studies are illu-
strated by tomography of a rat lung [24].
262 K. Watkin et al.
There is a great and largely untapped potential for the use of neutrons in
biomedical imaging using isotopically labeled marker compounds. Medical
compounds incorporating neutron-absorbing isotopes can be targeted to spe-
cific organs, such as the heart, to enable neutron imaging of anatomical struc-
ture and organ function as well as to deliver radiation doses specifically to
tumor sites. Such a system could provide higher-resolution images than PET
and more precision in targeting disease. Neutrons may also be used with water-
stable SPGO (small particulate gadolinium oxide) nanoparticles [25–27] for
targeted or non-targeted contrast enhancement. Neutron imaging through
metal – e.g., stents and other metallic objects used for biological purposes – is
possible. Small-animal imaging studies, isotopic imaging, and targeted
contrast-enhanced imaging are all potential applications.
Recent experiments highlight the biomedical potential of the neutron-sensitive
SPGO contrast agent. Dextranized SPGO (10 nm) was injected into the portal
vessel of a porcine liver specimen. A thermal neutron radiograph of the thin
specimen (1.5 mm) with the dextranized SPGO highlighting the lumen of the
vessel is shown in Fig. 14.11. SPGO particles appear dark because of the high
thermal neutron capture cross section of gadolinium.
Fig. 14.11 Neutron images of SPGO-doped liver tissue. A: original liver specimen; B: color-
enhanced image of A; C: distribution of the SPGO nanoparticles created using simple thresh-
olding techniques
14 Probing the Potential of Neutron Imaging 263
14.6 Summary
Many recent advances in medical imaging are based upon technological advances
in computer visualization, refined image projection techniques, and advanced
charge-coupled device systems, among others. Although there have been several
recent attempts to create small neutron generating systems that may be useful at
the laboratory level, the use of larger reactor- and accelerator-based neutron
sources with dedicated imaging systems remains the most effective and efficient
way to use thermal neutrons for biological tissue imaging.
Neutron radiographs of biological tissues accompanied the first neutron
imaging experiments. Neutron imaging of biological samples flourished during
the 1960s and 1970s and then largely disappeared until the 1990s. The recent
emergence of new techniques and improved detection (e.g., for epithermal
neutrons – Chapter 4) are leading to a rejuvenation of the field of neutron
imaging.
Neutron imaging of living human subjects still provokes obvious radiologi-
cal concerns. However, its potential for applications such as clinical research,
small-animal imaging, and forensic science needs to be investigated further. The
high sensitivity of neutrons to hydrogen atoms should be of benefit for forensic
applications, where quantification of water changes in dead tissues is highly
important.
Acknowledgments This work is supported by the U.S. Department of Energy. The Oak Ridge
National Laboratory Spallation Neutron Source is managed by UT-Battelle, LLC, for the
Department of Energy under contract DE-AC05-00OR22725. We would like to thank Dr.
John Nemeth of Oak Ridge Associated Universities for financial support of part of this
research, under contract DE-AC05-06OR23100. We would also like to thank our many
colleagues, especially Daniel Hussey at the National Institute of Standards and Technology
Center for Neutron Research and Candice Halbert from Oak Ridge National Laboratory.
References
1. E. Lehmann, S. Hartmann, P. Wyer, Nucl. Instrum. Meth. A 542, 87 (2005).
2. H. Kallman, Research 1, 254 (1947).
3. J. Thewlis et al., Br. J. Appl. Phys. 7, 345 (1956).
4. H.L. Atkins, Biological application of neutron radiography, Mat. Eval. 23, 453–458
(1965).
5. Barton, J. P., Some possibilities of neutron radiography, Phys. Med. Biol. 9, 33–42 (1964).
6. M. Brown and P.B. Parks, Am. J. Roentg. 106, 472 (1969).
7. J. Anderson et al., Br. J. Radiol. 37, 957 (1964).
8. R.F. Barth, Coderre, M.G.H.Vicente and T.E.Blue, Clin. Can. Res. 11, 3987–4002 (2005).
9. T.F. Budinger et al., Phys. Med. Biol. 16(3), 439–450 (1971).
10. C.E. Floyd et al., Phys. Med. Biol. 51, 3375–3390 (2006).
11. J.E. Bender, A.J. Kapadia, A.C. Sharma, G.D. Tourassi, B.P. Harrawood, and C.E. Floyd,
Med. Phys. 34, 3866–3871 (2007).
12. A.J. Kapadia, A.C. Sharma, J.E. Bender, G.D. Tourassi, C.R. Howell, A.S. Crowell,
M.R. Kiser, B.P. Harrawood, R.S. Pedroni, and C.E. Floyd, IEEE Trans. Nucl. Sci.
55(1), 501–509 (2008).
264 K. Watkin et al.
A. J. Kapadia
Neutron spectroscopic techniques have been widely used in medical and biomedi-
cal research to detect the presence of elements in the body. Techniques, such as
instrumental neutron activation analysis (INAA), prompt-gamma neutron activa-
tion analysis (PGNAA), and delayed neutron activation analysis (DNAA), have
been used since the 1960s to identify elements in healthy and diseased tissue, and
techniques based on inelastic scattering of fast neutrons have been used frequently
in body composition studies to measure whole body carbon and oxygen content.
The use of NAA techniques for medical applications was first reported in
1964 by Anderson et al. [1] for measurement of sodium in the body. Between
1968 and 1972, Chamberlain reported the measurement of body calcium and
sodium in the body [2–4] and described techniques for whole-body NAA [5–7]
and pulsed NAA [8]. In 1971, Cohn and Dombrowski reported the measure-
ment of calcium, sodium chlorine, nitrogen, and phosphorus in the human
body through in vivo NAA [9]. Since then, NAA and PGNAA have been used
for a variety of applications, such as the measurement of nitrogen [10], carbon
and oxygen [11], cadmium [12], and manganese [13] in the body and in trace
element research to identify cancerous tissue [14–17]. The development of
NAA has been documented in several review articles [18–21], which the reader
may refer to.
Inelastic neutron scatter analysis (INSA) using fast neutrons was initially
reported by Kyere et al. in1982 [22], who used 14 MeV neutrons from a (d,T)
sealed-tube neutron generator to determine whole body carbon content as a
measure of energy expenditure in the body. Following this experiment,
Kehayias et al. reported methods to determine fat content in the body through
measurement of whole body carbon and oxygen through inelastic neutron
scatter [23–25], which they later combined with hydrogen and nitrogen mea-
surement through in vivo NAA [26]. Until recently, most experiments utilizing
fast-neutron inelastic scatter focused primarily on in vivo measurement of
body fat.
The use of nuclear resonance scattering (NRS) was reported by Wielopolski
et al. for detection of iron in the liver [27] and in the heart [28] using an indirect
method of nuclear excitation by gamma rays generated through neutron cap-
ture. Recently, these authors reported the use of 14 MeV neutrons for in vivo
measurement of liver iron through INSA and NRS [29].
In this chapter, we would discuss the development of a tomographic tech-
nique that uses inelastic scattering of fast neutrons and detection of the
resulting gamma rays to quantify spatial distributions of elements in a sample.
The technique, called neutron stimulated emission computed tomography
(NSECT) [30, 31], was pioneered at Duke University in 2003 by the late
Dr. Carey E. Floyd Jr. for the purpose of diagnostic medical imaging.
NSECT uses a beam of fast neutrons to excite stable isotopes of elements in
a sample to determine their concentration and spatial distribution within the
sample. Since it is sensitive to a wide variety of elements that naturally occur in
the human body, it has the potential to noninvasively diagnose several dis-
orders that are characterized by changes in element concentrations in the
diseased tissue [32–34].
Here we present an overview of the NSECT technique and discuss the
current status of the research. The reader may note that while medical imaging
of the human body is the ultimate goal of NSECT, the technique is still in
development stages, with most experiments being performed on phantoms and
excised tissue specimens. The methods and results presented in the chapter are
mainly for the purpose of demonstrating feasibility and proof-of-concept for
future medical applications.
15 Neutron Stimulated Emission Computed Tomography 267
15.1.1 Principle
NSECT has the potential for use in a variety of applications in both medical and
biological imaging research. It is sensitive to a wide variety of elements that naturally
occur in the human body and has the ability to provide a three-dimensional map
of the elemental distribution within the target object from a single scan. These
characteristics of NSECT have the potential to diagnose several disorders in the
human body that are characterized by changes in element concentrations in
the diseased tissue. Disorders such as iron and copper overload in the liver
[38–41] and several types of cancers including breast [14, 16, 17, 42–48], prostate
[49–51], and brain [52, 53] have been associated with changes in element concentra-
tions. NSECT has the potential to diagnose these disorders through a noninvasive
in vivo scan without the use of radioactive isotope tracers. Three potential medical
and biological applications of NSECT are discussed below.
testis, and female reproductive organs [54]. NSECT cancer diagnosis can be
performed by measuring the trace element composition of the tissue and analyzing
this information to identify combinations of elements that are markers of cancer.
Owing to NSECT’s tomographic ability, given adequate sensitivity, it can identify
both the location of the tumor as well as its cancer state from a single scan.
Fig. 15.1 Schematic of an NSECT system used in tomographic measurements. Major components
of the system are labeled as A – Neutron Collimator; B – Instantaneous Neutron Flux Monitor;
C – Sample Manipulation Tomographic Gantry; D – HPGe detector; E – Anticoincidence
Compton Shield; and F – Inelastically Scattered Neutron. The neutron beam (shown by the line
of dots) travels from the source to the sample (C) and scatters inelastically. Gammas emitted in the
inelastic scatter interaction (shown as line of dashes) are detected by the HPGe detector, while the
scattered neutron (F) continues along another trajectory (shown as a solid line).
(Figure modified from [94] pp. 2313–2326. # Institute of Physics and IOP Publishing
Limited 2008)
15 Neutron Stimulated Emission Computed Tomography 271
NSECT can be performed with a variety of neutron sources, some of which are
available commercially, while others are specialized sources available only at
state-of-the-art neutron facilities. An ideal source for an NSECT application
should produce a narrow collimated beam of fast neutrons with high neutron
flux and allow beam pulsing and energy tuning with a nearly monochromatic
energy profile. High neutron flux is required to achieve minimal scan durations,
while a collimated beam is required to attain spatial resolution in the tomo-
graphic acquisition. Beam pulsing or chopping is necessary to identify the time-
resolved inelastic scatter signal and separate it from other neutron and gamma
effects that occur later in time. Without beam pulsing (i.e., with a continuous-
beam source), the inelastic scatter signal can be obscured by radiation
detected from room and sample-related background effects that are observed
at different times, such as radioactive decay following neutron capture in the
sample. These effects can be reduced through time-of-flight (TOF) correction,
which needs beam pulsing. The TOF correction technique for NSECT has been
described in detail in a separate publication [93].
Energy tuning, while highly desirable, is not an essential requirement for a
neutron source. A requirement of inelastic scatter spectroscopy is that the energy
of the neutron beam must be higher than the maximum energy state to be excited.
However, unnecessarily high neutron energies will simply excite higher unwanted
states in both desired and undesired elements, leading to a busy spectrum
cluttered with noise. Further, as NSECT detectors are susceptible to Compton
noise from high gamma energies, exciting unwanted high energy states in the
sample will also produce higher Compton background noise signals in the
detectors, which can potentially obscure elements with low-lying characteristic
gamma lines. A monochromatic energy source producing neutrons with energy
slightly higher than the highest desired excited state is the most suitable. With a
tunable neutron source, a beam can be produced with energy that is sufficient to
excite the elements of interest and avoid other unwanted elements in the sample.
For example, to image carbon in a biological sample (at 4.439 MeV), a neutron
energy of 5.0 – 5.5 MeV is ideal as it excites the 4.439 MeV energy state in 12C but
avoids excitation of the 6.13 MeV excited state in 16O. A monochromatic source
also facilitates lower patient dose by restricting neutron illumination to the
required range of energy. Higher neutron energies that contribute to dose by
depositing more energy per interaction and lower thermal energies that contri-
bute to dose through neutron capture and subsequent radioactive emission are
eliminated in a true monochromatic source.
Table 15.1 lists some neutron sources that have been identified for use in
NSECT: Van-de-Graaff accelerators, deuterium-deuterium (DD) neutron
tubes, and deuterium-tritium (DT) neutron tubes. Van-de-Graaff accelerators
provide tunable collimated neutron beams that are suitable for tomography.
NSECT experiments performed thus far have used a Van-de-Graaff accelerator
272 A.J. Kapadia
Table 15.1 A comparison of three neutron sources suitable for NSECT applications
Characteristic Van-de-Graaff DD Neutron Tube DT Neutron Tube
Energy 5 MeV to 23.2 MeV 3.2 MeV 14 MeV
Flux 104 – 106 n/s.cm2 108 – 1011 n/s 1010 – 1013 n/s
Pulsing 2 ns @ 2.5 MHz 100 ms 100 ms
Collimation Available Available Available
Monochromatic Beam Yes Yes Yes
Energy Tuning Yes No No
Several gamma-ray detectors have been explored and evaluated for use in
NSECT, including single-crystal and clover HPGe detectors and the compara-
tively inexpensive options NaI-Tl and BGO. For multielement applications,
such as cancer diagnosis, only HPGe detectors are found to be suitable, provid-
ing the most desirable combination of energy resolution and efficiency. HPGe
detectors typically possess energy resolution of 0.1%, providing approximately
1 keV resolution at 1 MeV. BGO and NaI-Tl detectors provide typical energy
resolutions of approximately 7 and 12%, respectively, which is insufficient to
separate close-lying gamma lines in a multielement analysis. However, these
detectors are suitable for single-element evaluations where the element of interest
lies in a region, which is free from close-lying gamma lines from other elements.
NSECT experiments performed at Duke University have used two HPGe
segmented detectors in the two fold segmented clover detector configuration. In
this configuration, each detector consists of four coaxial n-type germanium
crystals mounted together in the shape of a 4-leaf clover. Each detector exhibits
minimum relative efficiency of 22% (relative efficiency compares the efficiency
of the detector at 1332 keV to that of a 3 inch cubic NaI detector) and full width
at half maximum less than or equal to 2.25 keV for 1.332 MeV gamma rays of
60
Co. Detectors are typically calibrated against a 22Na source and positioned at
135 degrees from the incident neutron beam. A majority of the elements of
interest to NSECT decay through electric quadrupole transitions, whose dis-
tribution has maximum intensity at 45 and 135 degrees. Placing the detectors at
the 135 degree orientation helps to maximize signal intensity and simulta-
neously prevent detector activation and damage from forward scattering neu-
trons. In Figure 15.1, the HPGe detector (labeled D) is visible surrounded by an
anticoincidence Compton shield (labeled E), which is used to reduce the effects
of Compton scattering in the detectors. These shields have a minimum peak-to-
Compton ratio of 41 for 60Co gamma rays at 1.332 MeV [32].
Fig. 15.2 Geometry of the phantom imaged in the tomography experiment. The vertical outer
bars represent copper while the diagonal inner (gray) bars represent iron. Each bar measures
0.6 cm by 6 cm by 2.5 cm.
(Figure from Floyd et al, ‘‘Neutron Stimulated Emission Computed Tomography of a Multi-
Element Phantom,’’ Phys Med Biol, vol. 53, pp. 2313–2326. # Institute of Physics and IOP
Publishing Limited 2008)
Fig. 15.3 Reconstructed image from the NSECT acquisition of the sample. The vertical outer
regions represent copper while the diagonal inner region represents iron. Each element was
reconstructed separately and then combined.
(Figure from Floyd et al, ‘‘Neutron Stimulated Emission Computed Tomography of a Multi-
Element Phantom,’’ Phys Med Biol, vol. 53, pp. 2313–2326. # Institute of Physics and IOP
Publishing Limited 2008)
Fig. 15.4 Gamma energy spectrum from the iron-copper phantom showing spectral lines
from six transitions in 56Fe and 63Cu:
1. 63Cu from 1st excited state to ground state; energy 660 keV
2. 56Fe from 1st excited state to ground state; energy 847 keV
3. 63Cu from 2nd excited state to ground state; energy 962 keV
4. 56Fe from 3rd to 2nd excited state; energy 1239 keV
5. 56Fe from 4th to 2nd excited state; energy 1811 keV
6. 63Cu from 6th to 1st excited state; energy 1864 keV
(Figure from [94] pp. 2313–2326. # Institute of Physics and IOP Publishing Limited 2008)
15 Neutron Stimulated Emission Computed Tomography 277
Fig. 15.5 A phantom of the human torso with two chambers – outer chamber corresponding
to an adult torso, and inner chamber corresponding to an adult human liver. Both chambers
can be filled separately with any desired material.
(Figure from Kapadia et al. [32], pp. 2633–2649. # Institute of Physics and IOP Publishing
Limited 2008)
to an adult human liver. The phantom was filled with bovine liver tissue with
artificially induced iron overload and was scanned using a 5 MeV neutron
beam. The resultant spectrum shown in Fig. 15.6 was analyzed to quantify
the concentration of iron in the liver. The gamma line at 847 keV was detected
for 4.18 g of 56Fe. Based on this spectrum, a clinically relevant projected
sensitivity of 6 mg/g was obtained for iron overload diagnosis. In addition,
gamma lines were identified for several other elements in the liver, including Cl,
Cu, K, Na, and Zn, which were confirmed through NAA, and for 12C from the
phantom. Gamma lines were also identified for 74Ge and 76Ge from the gamma-
ray detectors, 42K from the room background, and 1H (neutron capture) from
the phantom. The dose from the scan was calculated using a Monte-Carlo
simulation as 0.375 mSv [33], which is significantly lower than an abdominal
X-ray exam that typically delivers 2 mSv [99]. The experiment demonstrates
that NSECT has the potential to detect clinically relevant concentrations of
iron in the human body through a noninvasive scan with reasonable dose.
Although the scan time for this experiment was unreasonable at over 24 h, it
278 A.J. Kapadia
Fig. 15.6 Gamma energy spectrum from the uniform iron overload torso phantom, showing
peaks corresponding to elements detected in the liver. The peak at 847 keV corresponds to
56
Fe. Peaks are also seen for Ge from the detector and 12C and 1H from the tissue.
(Figure from Kapadia et al. [32], pp. 2633–2649. # Institute of Physics and IOP Publishing
Limited 2008)
can be brought down to a few seconds using high-flux neutron sources (such as
the aforementioned DD and DT tubes) and increasing the number of detectors.
Detailed analysis and a complete description of the experiment are available
in [33].
Fig. 15.7 NSECT spectrum of a benign breast sample showing elements identified through
NSECT spectroscopy.
(Figure from Kapadia et al. [32], pp. 501–509. # 2008 IEEE)
Fig. 15.8 NSECT spectrum of a malignant breast sample showing elements identified through
NSECT spectroscopy.
(Figure from Kapadia et al. [32], pp. 501–509. # 2008 IEEE)
280 A.J. Kapadia
Table 15.2 List of elements showing statistically significant differences between benign and
malignant spectra. Elements with negative differences showed a decrease in concentration in
the malignant sample. Statistical significance was calculated using a z-score test for difference
of means. (Table from Kapadia et al. [32], pp. 501–509. # 2008 IEEE)
Energy keV Element match Counts benign Counts malignant Diff p-val
79
219 Br 6 19 13 0.01
59
397 Co, 79Br 16 2 –14 0.01
81
1028 Br 13 29 16 0.05
39
1128 K, 68Zn 0 13 13 0.001
56
1306 Fe 10 0 –10 0.01
27
2299 Al 0 13 13 0.001
37
2469 Cl, 56Fe, 66Zn 5 15 10 0.05
35
3635 Cl 3 14 11 0.01
of a few hundred micrograms, which is below the sensitivity of the current system,
which is a few hundred milligrams [32]. Therefore, accurate quantification of these
microgram concentrations will require a significant improvement in sensitivity.
Detailed analysis and a complete description of the experiment are available
in [32].
Patient dose in NSECT is of significant concern due to the use of fast neutrons, which
have a dose quality factor (Q-factor) of 10. For the technique to translate successfully
to the clinical environment, NSECT dose must be comparable to the dose levels of
other ionizing imaging modalities. As the technique is still in early stages of develop-
ment, dose analysis in NSECT is currently performed using Monte-Carlo simulations.
The process, described in [96, 100], can be summarized in the following three steps:
(a) a Monte-Carlo simulation is used to estimate two parameters for an inci-
dent neutron beam – the number of neutrons that interact in the volume of
interest and the average energy deposited per interacting neutron,
(b) the resulting energy deposited in the volume is converted from MeV to J/kg
using the known mass of the volume to give the absorbed energy in Gray (Gy),
and
(c) the absorbed energy is multiplied by the quality factor for neutrons (10) and
the weighting factor for the organ of interest to give the effective dose
equivalent in Sieverts (Sv).
This technique has been used to calculate the patient dose for NSECT scans
of the abdomen [96], liver [33, 96, 100], and breast [100, 101]. Table 15.3
summarizes the results of these dose-analysis simulations.
In comparison, an abdominal X-ray scan typically delivers 2 mSv, a mam-
mogram delivers 0.7 mSv, while an abdominal CT exam delivers approximately
10 mSv [99]. The doses from NSECT scans appear comparable to, or even lower
than, the other techniques. This is largely due to the modest number of neutrons
15 Neutron Stimulated Emission Computed Tomography 281
Table 15.3 NSECT dose delivered to organs in the body from NSECT scans
Organ Spectroscopic scan Tomography scan
Abdomen 1–2 mSv 1–5 mSv
Liver 0.02–1 mSv 0.5–3 mSv
Breast 0.02–0.5 mSv 0.5–1 mSv
15.5 Summary
NSECT represents an exciting new imaging modality that has the potential for
application in both medical and biological research. Several human disorders
characterized by element changes can be diagnosed through noninvasive in vivo
scanning using this technique. NSECT has the ability to obtain tomographic
information about the spatial distribution of elements within a tissue or organ to
make a quantitative and spatial diagnosis. A prototype of the NSECT acquisition
system has been developed and built at Duke University using a Van-de-Graaff
accelerator and HPGe detectors. As demonstrated through experiments with the
prototype system, NSECT has the sensitivity to detect concentrations of iron that
represent a clinically relevant liver iron overload condition. Sensitivity evaluation
experiments indicate that concentrations as low as 3 mg/g may be quantifiable
through NSECT. Although MRI is able to quantify moderate concentrations of
iron overload, it suffers from a loss of accuracy for concentrations above 6 mg/g
wet weight (20 mg/g dry weight) due to a reduction in the signal intensity caused by
high concentrations of iron [89]. NSECT, on the other hand, shows an increase in
the signal with increasing iron concentration. This facilitates iron overload detec-
tion in patients with severe overload, where MRI begins to lose accuracy.
The image resolution observed with the prototype system is approximately
1 cm, which is passable for imaging large organs such as the liver. However, as
the resolution depends primarily on the width of the neutron beam, it is possible to
improve the resolution significantly by using a narrow collimated beam (at the cost
of additional scan time). Simulation experiments have demonstrated a resolution
of 5 mm when scanning was performed with a 5 mm rectangular beam [96]. While
the resolution of the NSECT system depends primarily on the size of the incident
neutron beam, the best resolution achievable in a hydrogen-rich biological sample
will also be limited by the noise component generated from neutron scatter by
hydrogen. Neutrons that are meant to illuminate a certain voxel in the sample may
282 A.J. Kapadia
scatter onto adjacent voxels containing an element of interest and induce inelastic
gamma emission from that voxel. Detection of the adjacent-voxel gammas may
lead to a spread in the tails of the resulting sinograms, which can lead to reduced
resolution. The contribution of this effect to the limits of resolution is currently
being investigated through Monte-Carlo simulations.
Patient dose for NSECT examinations of different organs have been found
to vary between 0.02 mSv for a breast spectroscopic exam [101] to under 1 mSv
[96] for a liver spectroscopic exam, evaluated through Monte-Carlo simula-
tions. Tomographic doses are slightly higher depending on the number of
projections used but are generally lower than the dose delivered from a corre-
sponding X-ray CT exam [96].
15.6 Future
Acknowledgment The author would like to acknowledge and thank all the members of the
Duke Advanced Imaging Laboratories (DAILabs) and Triangle Universities Nuclear
Laboratory (TUNL) who have been involved in the development of NSECT, especially
Georgia Tourassi, Amy Sharma, and Janelle Bender for their analytical contribution and
deep involvement in NSECT; Brian Harrawood for his unparalleled computing support;
Calvin Howell, Alexander Crowell, Matthew Kiser, and Robert Macri for their help and
guidance with NSECT acquisition experiments; and Anton Tonchev and Anthony Hutcheson
for their help with gamma detector setup and management. Finally, the author would like to
express his deep gratitude to Dr. Carey Floyd, the pioneer of NSECT, in whose memory and
name this research continues.
References
1. J. Anderson, S.B. Osborn, R.W. Tomlinson, D. Newton, J. Rundo, L. Salmon, and J.W.
Smith, Neutron-Activation Analysis in Man in Vivo. a New Technique in Medical Inves-
tigation, Lancet 2, 1201–1205, (Dec 5 1964).
2. M.J. Chamberlain, J.H. Fremlin, D.K. Peters, and H. Philip, Total body sodium by whole
body neutron activation in the living subject: further evidence for non-exchangeable
sodium pool, Br. Med. J. 2, 583–585 (Jun 8, 1968).
15 Neutron Stimulated Emission Computed Tomography 283
3. M.J. Chamberlain, J.H. Fremlin, D.K. Peters, and H. Philip, Total body calcium by
whole body neutron activation: new technique for study of bone disease, Br. Med. J. 2,
581–3, Jun 8 1968.
4. M.J. Chamberlain, J.H. Fremlin, D.K. Peters, and H. Philip, Measurement of whole body
calcium and sodium by neutron activation analysis in the living subject, Strahlentherapie
[Sonderb], 67, 178–85, 1968.
5. M.J. Chamberlain, Whole-body neutron activation analysis, Proc. R. Soc. Med. 62, 370–3
(Apr 1969).
6. M.J. Chamberlain, J.H. Fremlin, I. Holloway, and D.K. Peters, Use of the cyclotron for
whole body neutron activation analysis: theoretical and practical considerations, Int. J.
Appl. Radiat. Isot. 21, 725–34 (Dec 1970).
7. M.J. Chamberlain, J.H. Fremlin, D.K. Peters, and H. Philip, Applications of the whole-
body counter in total-body neutron activation analysis, Br. J. Radiol. 43, 287–8 (Apr
1970).
8. M.J. Chamberlain and J.H. Fremlin, Measurement of whole-body nitrogen by pulsed
neutron activation analysis, Strahlentherapie [Sonderb] 72, 88–93 (1972).
9. S.H. Cohn and C.S. Dombrowski, Measurement of total-body calcium, sodium, chlorine,
nitrogen, and phosphorus in man by in vivo neutron activation analysis, J. Nucl. Med. 12,
499–505 (Jul 1971).
10. D. Vartsky, K.J. Ellis, and S.H. Cohn, In vivo measurement of body nitrogen by analysis
of prompt gammas from neutron capture, J. Nucl. Med. 20, 1158–1155 (Nov 1979).
11. C.L. Hollas, L.E. Ussery, K.B. Butterfield, and R.E. Morgado, A method for in vivo
determination of carbon and oxygen using prompt gamma radiations induced by
14.7-MeV neutrons, Basic Life Sci. 55, 395–400 (1990).
12. H.C. Biggin, N.S. Chen, K.V. Ettinger, J.H. Fremlin, W.D. Morgan, R. Nowotny,
M.J. Chamberlain, and T.C. Harvey, Cadmium by in vivo neutron activation analy-
sis, in J. Radioanal. Chem, v. 19, no. 2, pp. 207–214; International colloquium on
activation analysis of very low amounts of elements; 2 Oct 1972; Saclay, France United
Kingdom (1974).
13. M.L. Arnold, F.E. McNeill, I.M. Stronach, A. Pejovic-Milic, D.R. Chettle, and A.
Waker, An accelerator based system for in vivo neutron activation analysis measurements
of manganese in human hand bones, Med. Phys. 29, 2718–24 (Nov 2002).
14. A. Garg, V. Singh, et al., An elemental correlation study in cancerous and normal breast
tissue with successive clinical stages by neutron activation analysis, Biol. Trace Element
Res. 46, 185–202 (1994).
15. A. Danielsen and E. Steinnes, A study of some selected trace elements in normal and
cancerous tissue by neutron activation analysis, J. Nucl. Med. 11, 260–4 (Jun 1970).
16. K.H. Ng, D.A. Bradley, and L.M. Looi, Elevated trace element concentrations in
malignant breast tissues, Br. J. Radiol. 70, 375–82 (Apr 1997).
17. K.H. Ng, D.A. Bradley, L.M. Looi, C.S. Mahmood, and A.K. Wood, Differentiation of
elemental composition of normal and malignant breast tissue by instrumental neutron
activation analysis, Appl. Radiat. Isot. 44, 511–6 (Mar 1993).
18. K.J. Ellis, Human body composition: in vivo methods, Physiol. Rev. 80, 649–80 (Apr
2000).
19. J.F. Sutcliffe, A review of in vivo experimental methods to determine the composition of
the human body, Phys. Med. Biol. 41, 791–833 (May 1996).
20. E. Witkowska, K. Szczepaniak, and M. Biziuk, Some applications of neutron activation
analysis: A review, J. Radioanal. Nucl. Chem. 265, 141–150 (2005).
21. S. Mattsson and B.J. Thomas, Development of methods for body composition studies,
Phys. Med. Biol. 51, R203–28 (Jul 7 2006).
22. K. Kyere, B. Oldroyd, C.B. Oxby, L. Burkinshaw, R.E. Ellis, and G.L. Hill, The
feasibility of measuring total body carbon by counting neutron inelastic scatter gamma
rays, Phys. Med. Biol. 27, 805–17 (Jun 1982).
284 A.J. Kapadia
23. J.J. Kehayias, S.B. Heymsfield, A.F. LoMonte, J. Wang, and R.N. Pierson, Jr., In vivo
determination of body fat by measuring total body carbon, Am. J. Clin. Nutr. 53, 1339–44
(Jun 1991).
24. J.J. Kehayias, Aging and body composition: possibilities for future studies, J. Nutr. 123,
454–8 (Feb 1993).
25. J.J. Kehayias, H. Zhuang, V. Hughes, and L. Dowling, Assessment of body fat and lean in
the elderly by measuring body carbon and oxygen: validation against hydrodensitometry,
Appl. Radiat. Isot. 49, 723–5 (May-Jun 1998).
26. J.J. Kehayias, S. Valtuena, A.B. Waitekus, C.A. Sheahan, and M. O’Neill, In vivo
elemental partition analysis using fast neutrons. A tool for testing the efficacy of new
clinical interventions, Ann. N Y. Acad. Sci. 904, 140–7 (May 2000).
27. L. Wielopolski, R.C. Ancona, R.T. Mossey, A.N. Vaswani, and S.H. Cohn, Nuclear reso-
nance scattering measurement of human iron stores, Med. Phys. 12, 401–4 (Jul-Aug 1985).
28. L. Wielopolski and E.C. Zaino, Noninvasive in-vivo measurement of hepatic and cardiac
iron, J. Nucl. Med. 33, 1278–82 (Jul 1992).
29. L. Wielopolski, Feasibility of measuring iron in vivo using fast 14 MeV neutrons, Cooley’s
Anemia Foundation Inc. BNL-7980–2005 (2005).
30. C.E. Floyd, C.R. Howell, B.P. Harrawood, A.S. Crowell, A.J. Kapadia, R. Macri, J.Q.
Xia, R. Pedroni, J. Bowsher, M.R. Kiser, G.D. Tourassi, W. Tornow, and R. Walter,
Neutron Stimulated Emission Computed Tomography of Stable Isotopes, in SPIE
Symposium on Medical Imaging, San Diego, CA, pp. 248–254 (2004).
31. C.E. Floyd, J.E. Bender, A. Sharma, A. Kapadia, J. Xia, B. Harrawood, G.D. Tourassi,
J. Lo, and C.R. Howell, Introduction to Neutron Stimulated Emission Computed Tomo-
graphy. Phys. Med. Biol. 50 (14), 3375–90 (2005).
32. A.J. Kapadia, A.C. Sharma, J.E. Bender, G.D. Tourassi, C.R. Howell, A.S. Crowell,
M.R. Kiser, B.P. Harrawood, and C.E. Floyd, Neutron Stimulated Emission Com-
puted Tomography for Diagnosis of Breast Cancer, IEEE Trans Nucl. Sci. 55, 501–509
(2008).
33. A.J. Kapadia, G.D. Tourassi, A.C. Sharma, A.S. Crowell, M.R. Kiser, and C.R. Howell,
Experimental detection of iron overload in liver through neutron stimulated emission
spectroscopy, Phys. Med. Biol. 53, 2633–2649 (2008).
34. C.E. Floyd, J.E. Bender, A.C. Sharma, A.J. Kapadia, J.Q. Xia, B.P. Harrawood, G.D.
Tourassi, J.Y. Lo, A.S. Crowell, and C.R. Howell, Introduction to neutron stimulated
emission computed tomography, Phys. Med. Biol. 51, 3375–3390 (2006).
35. M.N. Wernick and J.N. Aarsvold, Emission Tomography: The Fundamentals of PET and
SPECT: Academic Press, San Diego (2004).
36. A. Kak and M. Slaney, Principles of computerized tomographic imaging vol. 33. Philadel-
phia: society for industrial and applied mathematics (2001).
37. J.E. Bender, C.E. Floyd, B.P. Harrawood, A.J. Kapadia, A.C. Sharma, and J.L. Jesneck,
The effect of detector resolution for quantitative analysis of neutron stimulated emission
computed tomography, in SPIE Medical Imaging, pp. 1597–1605 (2006).
38. L.W. Powell, Diagnosis of hemochromatosis, Semin. Gastrointest Dis. 13, 80–8 (Apr 2002).
39. S. Joffe, Hemochromatosis, N. Lamki, B. Coombs, U. Schmiedl, R.M. Krasny, and J. Karani,
Eds., Medscape emedicine, emedicine.medscape.com/article/369012-overview, Mar 11, 2005.
40. L. Powell, Hemochromatosis, in D. Kasper, Fawci, AS, Longo, DL, Braunwald, E,
Hauser, SL, Jameson, JL, Ed.,Harrison’s Principles of Internal Medicine, 16 ed. vol. 2,
McGraw Hill, New York, NY, pp. 2298–2303 (2005).
41. G. Brewer, Wilson Disease, in D. Kasper, Fawci, AS, Longo, DL, Braunwald, E, Hauser,
SL, Jameson, JL, Ed., Harrison’s Principals of Internal Medicine, 16 ed. vol. 2, NY:
McGraw Hill, pp. 2313–2315 (2005).
42. H. Mussalo-Rauhamaa, S. Piepponen, J. Lehto, R. Kauppila, and O. Auvinen, Cu, Zn,
Se and Mg concentrations in breast fat of Finnish breast cancer patients and healthy
controls, Trace Elements Med. 10, 13–15 (1993).
15 Neutron Stimulated Emission Computed Tomography 285
43. K. Geraki, M.J. Farquharson, and D.A. Bradley, X-ray fluorescence and energy disper-
sive x-ray diffraction for the quantification of elemental concentrations in breast tissue,
Phys. Med. Biol. 49, 99–110 (Jan 7 2004).
44. U. Majewska, D. Banas, J. Braziewicz, S. Gozdz, A. Kubala-Kukus, and M. Kuchar-
zewski, Trace element concentration distributions in breast, lung and colon tissues, Phys.
Med. Biol. 52, 3895–911 (Jul 7 2007).
45. K. Geraki, M.J. Farquharson, and D.A. Bradley, Concentrations of Fe, Cu and Zn in
breast tissue: a synchrotron XRF study, Phys. Med. Biol. 47, 2327–39 (Jul 7 2002).
46. S.L. Rizk and H.H. Sky-Peck, Comparison between concentrations of trace elements in
normal and neoplastic human breast tissue, Cancer. Res. 44, 5390–4 (Nov 1984).
47. S. Rizk and H. Sky-Peck, Comparison between Concetrations of Trace Elements in
Normal and Neoplastic Human Breast Tissue, Cancer. Research. 44, 5390–539 (1984).
48. P.M. Santoliquido, H.W. Southwick, and J.H. Olwin, Trace metal levels in cancer of the
breast, Surg. Gynecol. Obstet. 142, 65–70 (Jan 1976).
49. M. Brys, A.D. Nawrocka, E. Miekos, C. Zydek, M. Foksinski, A. Barecki, and W.M.
Krajewska, Zinc and cadmium analysis in human prostate neoplasms, Biol. Trace. Elem.
Res. 59, 145–52 (Winter 1997).
50. A. Feustel, R. Wennrich, D. Steiniger, and P. Klauss, Zinc and cadmium concentration in
prostatic carcinoma of different histological grading in comparison to normal prostate
tissue and adenofibromyomatosis (BPH), Urol. Res. 10, 301–3 (1982).
51. V. Zaichick, T.V. Sviridova, and S.V. Zaichick, Zinc in the human prostate gland:
normal, hyperplastic and cancerous, Int. Urol. Nephrol. 29, 565–74 (1997).
52. E. Andrasi, M. Suhajda, I. Saray, L. Bezur, L. Ernyei, and A. Reffy, Concentration of
elements in human brain: glioblastoma multiforme, Sci. Total. Environ. 139/140, 399–402
(Nov 1 1993).
53. M. Persigehl, H. Schicha, K. Kasperek, and H.J. Klein, Trace element concentration in
human organs in dependence of age, Beitr. Path. 161, 209–220 (1977).
54. E.J. Margalioth, J.G. Schenker, and M. Chevion, Copper and zinc levels in normal and
malignant tissues, Cancer 52, 868–72 (Sep 1 1983).
55. P. Ghadirian, P. Maisonneuve, C. Perret, G. Kennedy, P. Boyle, D. Krewski, and A.
Lacroix, A case-control study of toenail selenium and cancer of the breast, colon, and
prostate, Cancer. Detect. Prev. 24, 305–13 (2000).
56. I. Kato, A.M. Dnistrian, M. Schwartz, P. Toniolo, K. Koenig, R.E. Shore, A. Zele-
niuch-Jacquotte, A. Akhmedkhanov, and E. Riboli, Iron intake, body iron stores and
colorectal cancer risk in women: a nested case-control study, Int. J. Cancer. 80, 693–8
(Mar 1 1999).
57. R.L. Nelson, F.G. Davis, E. Sutter, L.H. Sobin, J.W. Kikendall, and P. Bowen, Body iron
stores and risk of colonic neoplasia, J. Natl. Cancer. Inst. 86, 455–60 (Mar 16 1994).
58. K. Witkowski, A. Kozlowski, M. Pardela, J. Piecuch, and P. Walichiewicz, [Level of
copper in plasma and tissue of patients with esophageal and large bowel cancer], Wiad.
Lek. 46, 586–8 (Aug 1993).
59. K.Q. Xiao and W.J. Henderson, [Electron microscopy microanalysis and quantitative
detection of trace elements in carcinoma of the colon], Zhonghua Bing Li Xue Za Zhi 21,
142–5 (Jun 1992).
60. W.P. Banner, J.J. DeCosse, Q.H. Tan, and M.S. Zedeck, Selective distribution of sele-
nium in colon parallels its antitumor activity, Carcinogenesis 5, 1543–6 (Dec 1984).
61. G.C. Gregoriadis, N.S. Apostolidis, A.N. Romanos, and T.P. Paradellis, A comparative
study of trace elements in normal and cancerous colorectal tissues, Cancer 52, 508–19
(Aug 1 1983).
62. Z.M. Bataineh, I.H. Bani Hani, and J.R. Al-Alami, Zinc in normal and pathological
human prostate gland, Saudi. Med. J. 23, 218–20 (Feb 2002).
63. A. Feustel, R. Wennrich, and H. Dittrich, Investigations of trace elements in metastases
and primary carcinoma of the prostate, Urol. Res. 17, 107–9 (1989).
286 A.J. Kapadia
64. J.O. Ogunlewe and D.N. Osegbe, Zinc and cadmium concentrations in indigenous blacks
with normal, hypertrophic, and malignant prostate, Cancer 63, 1388–92 (Apr 1 1989).
65. I. Romics and L. Katchalova, Spectrographic determination of zinc in the tissues of
adenoma and carcinoma of the prostate, Int. Urol. Nephrol. 15, 171–6 (1983).
66. M. Yaman, D. Atici, S. Bakirdere, and I. Akdeniz, Comparison of trace metal concen-
trations in malign and benign human prostate, J. Med. Chem. 48, 630–634 (2005).
67. F.K. Habib, G.L. Hammond, I.R. Lee, J.B. Dawson, M.K. Mason, P.H. Smith, S.R.Y.K.
Stitch, A.G. Meade, E.P. Rack, and A.J. Blotcky, Metal-androgen interrelationships in
carcinoma and hyperplasia of the human prostate, J. Endocrinol., 61 (1), 133–41 (1976).
68. H. Kubo, S. Hashimoto, and A. Ishibashi, Simultaneous determinations of Fe, Cu, Zn,
and Br concentrations in human tissue sections, Med. Phys. 3, 204–9 (Jul-Aug 1976).
69. G.M. Brittenham and D.G. Badman, Noninvasive measurement of iron: report of an
NIDDK workshop, Blood 101, 15–9 (Jan 1 2003).
70. J.M. Alustiza, J. Artetxe, A. Castiella, C. Agirre, J.I. Emparanza, P. Otazua, M. Garcia-
Bengoechea, J. Barrio, F. Mujica, and J.A. Recondo, MR quantification of hepatic iron
concentration, Radiology 230, 479–84 (Feb 2004).
71. W.F. Avrin and S. Kumar, Noninvasive liver-iron measurements with a room-tempera-
ture susceptometer, Physiol. Meas. 28, 349–61 (Apr 2007).
72. H.L. Bonkovsky, R.B. Rubin, E.E. Cable, A. Davidoff, T.H. Rijcken, and D.D. Stark,
Hepatic iron concentration: noninvasive estimation by means of MR imaging techniques,
Radiology 212, 227–34 (Jul 1999).
73. G.M. Brittenham, S. Sheth, C.J. Allen, and D.E. Farrell, Noninvasive methods for
quantitative assessment of transfusional iron overload in sickle cell disease, Semin.
Hematol. 38, 37–56 (Jan 2001).
74. E. Cecchin, S. De Marchi, F. Querin, M.G. Marin, R. Fiorentino, and F. Tesio, Efficacy
of hepatic computed tomography to detect iron overload in chronic hemodialysis, Kidney.
Int. 37, 943–50 (Mar 1990).
75. R.W. Chapman, G. Williams, G. Bydder, R. Dick, S. Sherlock, and L. Kreel, Computed
tomography for determining liver iron content in primary haemochromatosis, Br. Med. J.
280, 440–2 (Feb 16 1980).
76. J.L. Chezmar, R.C. Nelson, J.A. Malko, and M.E. Bernardino, Hepatic iron overload:
diagnosis and quantification by noninvasive imaging, Gastrointest. Radiol. 15, 27–31
(Winter 1990).
77. R.M. Dixon, P. Styles, F.N. al-Refaie, G.J. Kemp, S.M. Donohue, B. Wonke, A.V.
Hoffbrand, G.K. Radda, and B. Rajagopalan, Assessment of hepatic iron overload in
thalassemic patients by magnetic resonance spectroscopy, Hepatology 19, 904–10 (Apr
1994).
78. Y. Gandon, D. Olivie, D. Guyader, C. Aube, F. Oberti, V. Sebille, and Y. Deugnier, Non-
invasive assessment of hepatic iron stores by MRI., Lancet 363, 357–362 (Jan 31, 2004 2004).
79. D. Guyader, Y. Gandon, J.Y. Robert, J.F. Heautot, H. Jouanolle, C. Jacquelinet, M.
Messner, Y. Deugnier, and P. Brissot, Magnetic resonance imaging and assessment of
liver iron content in genetic hemochromatosis, J Hepatol 15, 304–8 (Jul 1992).
80. J.M. Howard, C.N. Ghent, L.S. Carey, P.R. Flanagan, and L.S. Valberg, Diagnostic
efficacy of hepatic computed tomography in the detection of body iron overload, Gastro-
enterology 84, 209–15 (Feb 1983).
81. P. Liu, M. Henkelman, J. Joshi, P. Hardy, J. Butany, M. Iwanochko, M. Clauberg, M.
Dhar, D. Mai, S. Waien, and N. Olivieri, Quantification of cardiac and tissue iron by
nuclear magnetic resonance relaxometry in a nvel murine thalassemia-cardiac iron over-
load model, Can. J. Cardiol. 12, 155–64 (Feb 1996).
82. P. Nielsen, R. Engelhardt, M. Duerken, G.E. Janka, and R. Fischer, Using SQUID
biomagnetic liver susceptometry in the treatment of thalassemia and other iron loading
diseases, Transfus. Sci. 23, 257–8 (Dec 2000).
15 Neutron Stimulated Emission Computed Tomography 287
83. P. Nielsen, R. Engelhardt, J. Dullmann, and R. Fischer, Non-invasive liver iron quanti-
fication by SQUID-biosusceptometry and serum ferritin iron as new diagnostic para-
meters in hereditary hemochromatosis, Blood Cells Mol. Dis. 29, 451–8 (Nov-Dec 2002).
84. P. Nielsen, U. Kordes, R. Fischer, R. Engelhardt, and G.E. Janka, [SQUID-biosuscep-
tometry in iron overloaded patients with hematologic diseases], Klin Padiatr 214, 218–22
(Jul-Aug 2002).
85. H. Perrimond, C. Chagnon, I. Moulanier, G. Michel, H. Guidicelli, and P.J. Bernard, The
value of nuclear magnetic resonance in the study of iron overload in thalassemia patients,
Ann. Pediatr. (Paris) 38, 175–84 (Mar 1991).
86. S. Sheth, SQUID biosusceptometry in the measurement of hepatic iron, Pediatr. Radiol.
33, 373–7 (Jun 2003).
87. D. Vartsky, K.J. Ellis, D.M. Hull, and S.H. Cohn, Nuclear resonant scattering of gamma
rays – a new technique for in vivo measurement of body iron stores, Phys. Med. Biol. 24,
689–701 (Jul 1979).
88. D. Vartsky, L. Wielopolski, K.J. Ellis, and S.H. Cohn, The Use of Nuclear Resonant
Scattering of Gamma-Rays for Invivo Measurement of Iron, Nucl. Instrum. Meth. Phys.
Res. 193, 359–364 (1982).
89. E. Angelucci, A. Giovagnoni, G. Valeri, E. Paci, M. Ripalti, P. Muretto, C. McLaren, G.
M. Brittenham, and G. Lucarelli, Limitations of magnetic resonance imaging in measure-
ment of hepatic iron, Blood 90, 4736–42 (Dec 15 1997).
90. Z.J. Wang, J.C. Haselgrove, M.B. Martin, A.M. Hubbard, S. Li, K. Loomes, J.R. Moore,
H. Zhao, and A.R. Cohen, Evaluation of iron overload by single voxel MRS measure-
ment of liver T2, J. Magn. Reson. Imaging 15, 395–400 (Apr 2002).
91. A.J. Kapadia, A.C. Sharma, G.D. Tourassi, J.E. Bender, A.S. Crowell, M.R. Kiser, C.R.
Howell, and C.E. Floyd, Neutron Spectroscopy of Mouse Using Neutron Stimulated
Emission Computed Tomography (NSECT), in IEEE Nuclear Science Symposium, Med-
ical Imaging Conference, San Diego, CA, pp. 3546–3548 (2006).
92. A.C. Sharma, G.D. Tourassi, A.J. Kapadia, A.S. Crowell, M.R. Kiser, A. Hutcheson, B.
P. Harrawood, C.R. Howell, and C.E. Floyd, Elemental Spectrum of a Mouse Obtained
via Neutron Stimulation., in 2007 SPIE Symposium on Medical Imaging, San Diego, CA,
p. 65100 K (2007).
93. C.E. Floyd, A.C. Sharma, J.E. Bender, A.J. Kapadia, J.Q. Xia, B.P. Harrawood, G.D.
Tourassi, J.Y. Lo, M.R. Kiser, A.S. Crowell, R.S. Pedroni, R.A. Macri, S. Tajima, and C.
R. Howell, Neutron Stimulated Emission Computed Tomography: Background Correc-
tions, Nucl. Instrum. Meth. Phys. Res. Sect. B 254, 329–336 (2007).
94. C.E. Floyd, A.J. Kapadia, J.E. Bender, A.C. Sharma, J.Q. Xia, B.P. Harrawood, G.D.
Tourassi, J.Y. Lo, A.S. Crowell, M.R. Kiser, and C.R. Howell, Neutron Stimulated
Emission Computed Tomography of a Multi-Element Phantom, Phys. Med. Biol. 53,
2313–2326 (2008).
95. K. Lange and R. Carson, EM reconstruction Algorithms for Emission and Trans-
mission Tomography, Journal of Computer Assisted Tomography 8, 306–316 (1984).
96. A.J. Kapadia, Accuracy and Patient Dose in Neutron Stimulated Emission Computed
Tomography for Diagnosis of Liver Iron Overload: Simulations in GEANT4, in Biome-
dical Engineering. vol. PhD, Duke University, Durham, NC (2007).
97. W.A. Watson, III, J. Chen, G. Heyes, E. Jastrzembski, and D. Quarrie, CODA: a
scalable, distributed data acquisition system, in IEEE Transactions on Nuclear Science,
Vancouver, BC, Canada, pp. 61–68 (1994).
98. R. Fox, C. Bolen, K. Orji, and J. Venema, NSCLSpecTcl Meeting the Needs of Pre-
liminary Nuclear Physics Data Analysis in 11’th Annual Tcl/Tk Conference New Orleans,
Louisiana (2004).
99. RSNA, Radiation Exposure in X-ray Examinations, in American College of Radiology
(ACR) and the Radiological Society of North America (RSNA) (2007).
288 A.J. Kapadia
100. A.C. Sharma, B.P. Harrawood, J.E. Bender, G.D. Tourassi, and A.J. Kapadia, Neutron
stimulated emission computed tomography: a Monte Carlo simulation approach, Phys.
Med. Biol. 52, 6117–31 (Oct 21 2007).
101. J.E. Bender, A.J. Kapadia, A.C. Sharma, G.D. Tourassi, B.P. Harrawood, and C.
E. Floyd, Breast cancer detection using Neutron Stimulated Emission Computed
Tomography: prominent elements and dose requirements, Med. Phys. 34,
3866–3871 (2007).
102. A.J. Kapadia, C.E. Floyd, C.R. Howell, and B.P. Harrawood, Sampling Requirements
for Neutron Stimulated Emission Computed Tomography, in RSNA, Physics (Digital
Imaging, PACS) session Chicago, IL (2004).
Chapter 16
Visualizing Structures of Biological
Macromolecules Through Indirect Imaging
with Small-Angle Neutron Scattering
and Modeling
16.1 Introduction
16.1.1 Theory
Fig. 16.1 Schematic of small-angle scattering. Neutrons as plane waves are incident on a
sample particle and are scattered. The pattern observed results from the interference of
neutrons, scattered from different positions, within the particle
16 Visualizing Structures of Biological Macromolecules 291
ranging from a few hundred to a few thousand meters per second. For an
isotropic sample, such as for a protein tumbling freely in solution, the pattern
of scattered neutrons is symmetric about the beam and can be described
according to Eq. (16.1) [1, 2].
*Z 2 +
~R ~ 3
IðQÞ ¼ n ~
R s e iQ
d R ; (16:1)
V
where I(Q) is the scattered intensity, n is the number of particles per unit
volume, R ~ is the scattering length density of the particle at position R, ~
and s is the scattering length density of the solvent. The integral is taken over
the volume of the particle V [1, 2]. The integral is averaged over time, all
orientations, and the ensemble of structures present in the solution.
The fact that Eq. 16.1 describes the scattering signal relative to the background
solvent is incredibly important because it enables the use of SANS with contrast
variation [3]. The dramatic difference between the scattering lengths of hydro-
gen (–3.71 fm) and deuterium (6.67 fm) makes it possible to greatly alter the
scattering length density of a biological macromolecule or the aqueous back-
ground by substituting for hydrogenated material ones that are isotopically
labeled with deuterium. Contrast variation is a particularly powerful technique
for studying multisubunit complexes in which one subunit is labeled with
deuterium because it enables the separation of the scattering from the two
components within the complex. In a contrast variation experiment, a complex
containing a subunit with a different average scattering length density, such as a
deuterium-labeled protein or nucleic acid, is measured in a series of solutions
consisting of mixtures of H2O and D2O. Conceptually, the effect of changing the
H2O/D2O mixture in the solution is illustrated in Fig. 16.2. The intensity profiles
in the contrast variation series can be written in the following manner (Eq. 16.2):
IðQ; 1 ; 2 Þ ¼ 21 I1 ðQÞ þ 22 I2 ðQÞ þ 1 2 I12 ðQÞ (16:2)
Here, 1 and 2 are the differences between the hydrogenated and deu-
terated components of the complex relative to the solvent, respectively, and are
often called the contrasts. The three functions, I1(Q), I2(Q), and I12(Q) are
known as the basic scattering functions [3]. The basic scattering functions
are the scattering signals from the components within the complex having two
different scattering length densities and the cross term that provides informa-
tion on the relative disposition of the components. The measured contrast
variation series data define a set of linear equations that can be used to extract
the basic scattering functions for further analysis.
292 W.T. Heller and G.A. Baker
Fig. 16.2 Contrast variation. The scattering length density of the background solvent is varied
relative to the two subunits having different scattering length densities, thereby differentially
highlighting the structures of the subunits within the whole
There are two common, basic methods of SAS data analysis. For a dilute,
monodisperse solution of homogenous particles, the intensity I(Q) can be fit
for the radius of gyration (Rg) according to Guinier [2]:
2
QRg
ln½IðQÞ ¼ ln½Ið0Þ (16:3)
3
Here, I(0) is the scattered intensity at Q = 0, which is a shape-independent
function of the molecular weight of the particle and the average scattering
length density relative to the background solution. The radius of gyration, Rg,
is a shape-dependent function of the contrast-weighted distance from the center
of a particle to scattering elements within the particle. While Guinier analysis
may be used for any particle of unknown size or shape, there are instances when
it is not valid. To use Guinier analysis, the following three conditions must be
satisfied: (1) the system must be rotationally isotropic, (2) the particles must not
interact with each other, and (3) Q Rg must be less than 1.3, a condition
resulting from the approximation used to derive Eq. (16.3) for globular parti-
cles. Condition (3) must be further restricted for other shapes, with Q Rg less
than 1.0 being commonly applied.
Additional analysis can be performed to obtain more information on the
shape of the scattering particle. Equation (16.1) can be inverted to provide the
distance distribution function P(R) as a function of the experimental I(Q) using
the Fourier transform in Eq. (16.4).
Z1
1
PðRÞ ¼ 2 IðQÞQR sinðQRÞdQ : (16:4)
2p
0
16 Visualizing Structures of Biological Macromolecules 293
The actual approach uses inverse Fourier methods and includes the Glatter
algorithm [4, 5], the Moore algorithm [6], and the method implemented in the
software GNOM [7]. Fitting SAS data for P(R) provides a measure of dmax, the
maximum linear dimension of the particle.
An example SANS intensity profile calculated from the crystal structure of
hemoglobin (PDB ID: 4HHB [8]) using the software ORNL_SAS [9] is shown
in Fig. 16.3. As can be seen, the data have a relatively simple appearance that
suggests they do not contain a great deal of information, in contrast to high-
resolution crystallography data. Researchers have constructed arguments
based on the Shannon Sampling Theorem [10] to estimate the appropriate
number of degrees of freedom, nf, suitable for fitting SAS data [6, 11–13]. The
number is a function of the Q-range of the data collected and the size of the
scattering particle. If data are collected for a particle of size dmax over a range
that extends from Qmin to Qmax, then nf dmax(Qmax-Qmin)/ p.
Structural biology is a very visual science, but the basic analysis of SAS data
does not provide a structure that can be used to understand the interplay
between structure and function in complex macromolecular systems. Hence,
methods have been developed for constructing models from SAS data. Much
of the effort has focused on SAXS data, due to the greater availability of
instruments. Still, tools for building models from SANS data have been
developed that can be applied to contrast variation data of multisubunit
complexes. The kind of modeling possible is often a function of the amount
of existing structural information, particularly high-resolution structural
information.
294 W.T. Heller and G.A. Baker
In cases where no structural information exists for the protein or complex being
studied, several methods exist for building structural models that are referred
to as shape restoration methods. These techniques encompass a broad range of
approaches spanning simple shapes, spherical harmonics, and methods using
aggregates of shapes. Each approach has its strengths and weaknesses. A large
number of these methods are freely available to the scientific community,
making them attractive choices for building models from SAS data.
Some of the most established modeling methods for SAS data on proteins
and protein complexes use simple shapes. Model fitting using these approaches
employs a limited number of degrees of freedom, providing a degree of con-
fidence that the answer is uniquely determined by the data. Modeling using a
simple shape has the further advantage that the associated intensity profile is
quite often an analytical function [15], making it attractive for fitting data by
hand. More importantly, many of these geometrical shapes are implemented in
freely available software packages [16, 17]. For modeling using multiple shapes,
the only analytical function available is the Debye formula for aggregates of
spheres [18]. Other multishape structures lack analytical functions for the
scattering curves, although methods have been developed for building models
in this manner [19–27]. Such approaches are very well suited to modeling
contrast variation series of SANS data from biological macromolecular com-
plexes, where each subunit can be assigned to a single shape.
Shape reconstruction methods using spherical harmonics were among the
first ab initio methods developed for extracting structural information from
SAS data [11, 28–34]. In this approach, the structure is defined by a series
superposition of spherical harmonics, an infinite set of orthogonal functions. It
is trivial to limit the number of terms used in the reconstruction, which is
typically assigned to nf. The ability to readily restrict the number of free
parameters in the fit provides a measure of confidence that the structure
found is reasonable. Still, the resulting structures are restricted to those which
can be represented by spherical harmonics, and significantly more complicated
structures cannot be effectively modeled. The approach is readily applied to
SAXS data, but it is also applicable to contrast variation experiments. These
shape restoration methods can also be applied directly to basic scattering
functions extracted from such data sets.
Currently, the most popular methods for shape restoration from SAS data
are the aggregate ab initio techniques that use large sets of simple, relatively
small shapes to define the volume of the scattering particles. Spheres are the
most common base shape for aggregate techniques [35–38], but ellipsoids have
also been used [14]. The programs DAMMIN [39] and GASBOR [40] use
dummy atoms and dummy residues, respectively, for the shape reconstructions
and are the most popular software packages currently available for this pur-
pose. A related grid-based method has also been developed [41–43]. In general,
16 Visualizing Structures of Biological Macromolecules 295
any shape can be reproduced using these methods up to the limits of the
resolution of the data and the base shape used to build the structure, in principle
avoiding the limitations inherent in using spherical harmonics to define shapes.
Unfortunately, the large number of degrees of freedom inherent in such model-
ing makes degenerate solutions possible that fit the SAS data equally well. For
this reason, multiple independent structures are often generated and averaged
into a consensus shape [35–37, 44]. An implementation of DAMMIN exists that
is capable of directly modeling against contrast variation series of SANS data
from complexes with selective deuterium labeling [39]. The other techniques can
be applied to SAXS data, SAS data from structures that are not selectively
deuterated, or to basic scattering functions extracted from contrast variation
series data.
16.3 Examples
The skeletal and cardiac variants of troponin complexes have also been
studied by SANS with contrast variation. A variety of modeling approaches
have been used to visualize the structure of the complex, including mixing
high- and low-resolution structural information [80, 81], simple shapes [24],
ab initio modeling of the basic scattering functions followed by rigid body
modeling using the shapes [37, 58], as shown in Fig. 16.4, and investigating
the function of the complex from an available high-resolution structure [82].
Protein kinase A is another dynamic multisubunit complex for which SANS
plus contrast variation and modeling was used to obtain structural and function
information [26, 83], as shown in Fig. 16.5. Protein–nucleic acid complexes
benefit from the inherently different scattering length densities of hydrogenated
protein and nucleic acids, which was leveraged in a study of the assembly of the
methionine repressor protein, MetJ, on different lengths of DNA taken from
regulatory sequences of different genes within the met regulon [84], which is
shown in Fig. 16.6. By no means do these examples comprise an exhaustive
listing of the multisubunit complexes that have been studied to great benefit
by SANS with contrast variation, as other systems have also been studied
recently [26, 85–90], a trend certainly to continue as interest in the method
grows.
Fig. 16.4 Models of the ternary cardiac troponin complex (TnC–TnI–TnT(198–298) derived
from SANS with contrast variation using two different isotopic labeling patterns [91]. In the
left model, the TnC was deuterated (large dark spheres), while the TnI subunit was labeled
(small light spheres) for the contrast variation series used to derive the model on the right. The
density near the middle of the structure was inferred to be the TnT(198–298). The longest
dimension of the structure is 125 Å
298 W.T. Heller and G.A. Baker
Fig. 16.5 Model of the protein kinase A structure derived using SANS with contrast variation
showing the catalytic domains (gray), cyclic AMP binding domains (light gray), the dimeriza-
tion-docking domain (dark gray), and unstructured linker domains (black) [83]. The longest
dimension of the structure is 155 Å
Fig. 16.6 Models of MetJ–DNA complexes using different lengths of DNA. MetJ dimers
along the length of the DNA (black spheres) are colored light and dark gray in an alternating
pattern to show differences in the length-dependent packing of the protein dimers on the DNA
[84]. The radius from the axis of the DNA of the structure on the left is 40 Å
16.4 Conclusions
Acknowledgments This work was supported by the Oak Ridge Center for Structural
Molecular Biology (KP1102010) of the Office of Biological and Environmental Research of
the U. S. Department of Energy under contract No. DE-AC05-00OR22725 with Oak Ridge
National Laboratory, managed and operated by UT-Battelle, LLC. The submitted manu-
script has been authored by a contractor of the U.S. government under Contract DE-AC05-
00OR22725. Accordingly, the U.S. government retains a nonexclusive royalty-free license to
publish or reproduce the published form of this contribution, or allow others to do so, for U.S.
government purposes.
References
1. P. Debye and A. Bueche, Scattering by an Inhomogeneous Solid. J. Appl. Phys.
20, 518–525 (1949).
2. A. Guinier and G. Fournet, Small-angle Scattering of X-rays. New York, Wiley (1955).
3. K. Ibel and H.B. Stuhrmann, Comparison of Neutron and X-Ray-Scattering of Dilute
Myoglobin Solutions. J. Mol. Biol. 93, 255–265 (1975).
4. O. Glatter, New Method for Evaluation of Small-Angle Scattering Data. J. Appl. Crystallogr.
10, 415–421 (1977).
5. O. Glatter, Interpretation of Real-Space Information from Small-Angle Scattering
Experiments. J Appl Crystallogr 12: 166–175 (1979).
6. P.B. Moore, Small-Angle Scattering - Information-Content and Error Analysis. J. Appl.
Crystallogr. 13, 168–175 (1980).
7. D.I. Svergun, Determination of the Regularization Parameter in Indirect-Transform
Methods Using Perceptual Criteria. J. Appl. Crystallogr. 25, 495–503 (1992).
8. G. Fermi, M.F. Perutz, B. Shaanan, and R. Fourme, The Crystal-Structure of Human
Deoxyhemoglobin at 1.74a Resolution. J. Mol. Biol. 175, 159–174 (1984).
9. E. Tjioe and W.T. Heller, ORNL_SAS: software for calculation of small-angle scattering
intensities of proteins and protein complexes. J. Appl. Crystallogr. 40, 782–785 (2007).
10. C.E. Shannon, A Mathematical Theory of Communication. Bell Sys. Tech. J. 27,
379–423, 623–656 (1948).
11. D.I. Svergun, V.V. Volkov, M.B. Kozin, and H.B. Stuhrmann, New developments in
direct shape determination from small-angle scattering .2. Uniqueness. Acta. Crystallogr.
Sect. A 52, 419–426 (1996).
12. D. Taupin and V. Luzzati, Information-Content and Retrieval in Solution Scattering
Studies .1. Degrees of Freedom and Data Reduction. J. Appl. Crystallogr. 15, 289–300
(1982).
13. V.V. Volkov and D.I. Svergun, Uniqueness of ab initio shape determination in small-
angle scattering. J. Appl. Crystallogr. 36, 860–864 (2003).
300 W.T. Heller and G.A. Baker
14. D. Vigil, S.C. Gallagher, J. Trewhella, and A.E. Garcia, Functional dynamics of the
hydrophobic cleft in the N-domain of calmodulin. Biophys. J. 80, 2082–2092 (2001).
15. J.S. Pedersen, Analysis of small-angle scattering data from colloids and polymer solutions:
modeling and least-squares fitting. Adv. Colloid. Interface. Sci. 70, 171–210 (1997).
16. S.R. Kline, Reduction and analysis of SANS and USANS data using IGOR Pro. J. Appl.
Crystallogr. 39, 895–900 (2006).
17. M.V. Petoukhov, P.V. Konarev, A.G. Kikhney, and D.I. Svergun, ATSAS 2.1 – towards
automated and web-supported small-angle scattering data analysis. J. Appl. Crystallogr.
40, S223–S228 (2007).
18. P. Debye, Zerstreuung von Röntgenstrahlen. Annalen der Physik 351, 809–823 (1915).
19. S. Fujiwara, F.J. Kull, E.P. Sablin, D.B. Stone, and R.A. Mendelson, The Shapes of the
Motor Domains of 2 Oppositely Directed Microtubule Motors, Nod and Kinesin – a
Neutron-Scattering Study. Biophys. J. 69, 1563–1568 (1995).
20. S. Hansen Calculation of Small-Angle Scattering Profiles Using Monte-Carlo Simula-
tion. J Appl Crystallogr 23, 344–346 (1990).
21. S.P. Harris, W.T. Heller, M.L. Greaser, R.L. Moss, and J. Trewhella, Solution structure
of heavy meromyosin by small-angle scattering. J. Biol. Chem. 278, 6034–6040 (2003).
22. H. Hartmann, A. Bongers, and H. Decker, Small-angle neutron scattering reveals an
oxygen-dependent conformational change of the immunogen keyhole limpet hemocyanin
type 1 (KLH1). Eur. Biophys. J. Biophys. Lett. 30, 471–475 (2001).
23. W.T. Heller, ELLSTAT: shape modeling for solution small-angle scattering of proteins
and protein complexes with automated statistical characterization. J. Appl. Crystallogr.
39, 671–675 (2006).
24. S.J. Henderson, Monte Carlo modeling of small-angle scattering data from non-
interacting homogeneous and heterogeneous particles in solution. Biophys. J. 70,
1618–1627 (1996).
25. D.B. Stone, P.A. Timmins, D.K. Schneider, I. Krylova, C.H.I. Ramos, F.C. Reinach, and
R.A. Mendelson, The effect of regulatory Ca2+ on the in situ structures of troponin C
and troponin I: A neutron scattering study. J. Mol. Biol. 281, 689–704 (1998).
26. J.K. Zhao, E. Hoye, S. Boylan, D.A. Walsh, and J. Trewhella, Quaternary structures
of a catalytic subunit-regulatory subunit dimeric complex and the holoenzyme of the
cAMP-dependent protein kinase by neutron contrast variation. J. Biol. Chem. 273,
30448–30459 (1998).
27. J. Zhou, A. Deyhim, S. Krueger, and S.K. Gregurick, LORES: Low resolution shape
program for the calculation of small angle scattering profiles for biological macromole-
cules in solution. Comput. Phys. Commun. 170, 186–204 (2005).
28. J.G. Grossmann, Z.H.L. Abraham, E.T. Adman, M. Neu, R.R. Eady, B.E. Smith, and
S.S. Hasnain, X-Ray-Scattering Using Synchrotron-Radiation Shows Nitrite Reductase
from Achromobacter-Xylosoxidans to Be a Trimer in Solution. Biochemistry 32,
7360–7366 (1993).
29. F. Spinozzi, F. Carsughi, and P. Mariani, Particle shape reconstruction by small-angle
scattering: Integration of group theory and maximum entropy to multipole expansion
method. J. Chem. Phys. 109, 10148–10158 (1998).
30. H.B. Stuhrman, Interpretation of Small-Angle Scattering Functions of Dilute Solutions
and Gases – a Representation of Structures Related to a One-Particle-Scattering
Function. Acta Crystallogr. Sect. A-Crystal Phys. Diff. Theor. Gen. Crystallogr. A 26,
297–306 (1970).
31. D.I. Svergun, M.H.J. Koch, J.S. Pedersen, and I.N. Serdyuk, Structural Model of the 50-S
Subunit of Escherichia-Coli Ribosomes from Solution Scattering .2. Neutron-Scattering
Study. J. Mol. Biol. 240, 78–86 (1994).
32. D.I. Svergun, M.H.J. Koch, and I.N. Serdyuk, Structural Model of the 50-S Subunit of
Escherichia-Coli Ribosomes from Solution Scattering .1. X-Ray Synchrotron-Radiation
Study. J. Mol. Biol. 240, 66–77 (1994).
16 Visualizing Structures of Biological Macromolecules 301
33. D.I. Svergun and H.B. Stuhrmann, New Developments in Direct Shape Determination
from Small-Angle Scattering .1. Theory and Model-Calculations. Acta. Crystallogr. Sect.
A 47, 736–744 (1991).
34. D.I. Svergun, V.V. Volkov, M.B. Kozin, H.B. Stuhrmann, C. Barberato, and M.H.J.
Koch, Shape determination from solution scattering of biopolymers. J. Appl. Crystallogr.
30, 798–802 (1997).
35. P. Chacon, J.F. Diaz, F. Moran, and J.M. Andreu, Reconstruction of protein form with
X-ray solution scattering and a genetic algorithm. J. Mol. Biol. 299, 1289–1302 (2000).
36. P. Chacon, F. Moran, J.F. Diaz, E. Pantos, and J.M. Andreu, Low-resolution structures
of proteins in solution retrieved from X-ray scattering with a genetic algorithm. Biophys. J.
74: 2760–2775 (1998).
37. W.T. Heller, N.L. Finley, W.J. Dong, P. Timmins, H.C. Cheung, P.R. Rosevear, and
J. Trewhella, Small-angle neutron scattering with contrast variation reveals spatial
relationships between the three subunits in the ternary cardiac troponin complex and
the effects of troponin I phosphorylation. Biochemistry 42, 7790–7800 (2003).
38. D. Walther, F.E. Cohen, and S. Doniach, Reconstruction of low-resolution three-dimen-
sional density maps from one-dimensional small-angle X-ray solution scattering data for
biomolecules. J. Appl. Crystallogr. 33, 350–363 (2000).
39. D.I. Svergun, Restoring low resolution structure of biological macromolecules from
solution scattering using simulated annealing. Biophys. J. 76, 2879–2886 (1999).
40. D.I. Svergun, M.V. Petoukhov, and M.H.J. Koch, Determination of domain structure of
proteins from X-ray solution scattering. Biophys. J. 80, 2946–2953 (2001).
41. H. Hartmann, A. Bongers, and H. Decker, Small-angle X-ray scattering-based three-
dimensional reconstruction of the immunogen KLH1 reveals different oxygen-dependent
conformations. J. Biol. Chem. 279, 2841–2845 (2004).
42. H. Hartmann and H. Decker. Small-angle scattering techniques for analyzing conforma-
tional transitions in hemocyanins. Energ. Biol. Macromol. Pt D. 379, 81–106 (2004).
43. H. Hartmann, T. Muller, and H. Decker, Modeling techniques for analysing conforma-
tional transitions in hemocyanins by small-angle scattering of X-rays and neutrons.
Micron 35, 11–13 (2004).
44. M.B. Kozin and D.I. Svergun, Automated matching of high- and low-resolution
structural models. J. Appl. Crystallogr. 34, 33–41 (2001).
45. H.M. Berman, J. Westbrook, Z. Feng, G. Gilliland, T.N. Bhat, H. Weissig, I.N. Shindyalov,
and P.E. Bourne, The Protein Data Bank. Nucl. Acids. Res. 28, 235–242 (2000).
46. B.A. Fedorov, O.B. Ptitsyn, and L.A. Voronin, X-Ray Diffuse Scattering of Globular
Protein Solutions – Consideration of Solvent Influence. FEBS Lett. 28, 188–190 (1972).
47. S.C. Gallagher, A.J. Callaghan, J.K. Zhao, H. Dalton, and J. Trewhella, Global
conformational changes control the reactivity of methane monooxygenase. Biochemistry
38, 6752–6760 (1999).
48. R. Langridge, D.A. Marvin, W.E. Seeds, H.R. Wilson, C.W. Hooper, M.H.F. Wilkins,
and L.D. Hamilton, The Molecular Configuration of Deoxyriboucleic Acid: Molecular
Models and their Fourier Transforms. J. Mol. Biol. 2, 38–64 (1960).
49. F. Merzel and J.C. Smith, SASSIM: a method for calculating small-angle X-ray and
neutron scattering and the associated molecular envelope from explicit-atom models of
solvated proteins. Acta. Crystallogr. Sect. D-Biol. Crystallogr. 58, 242–249 (2002).
50. J. Ninio, M. Yaniv, and V. Luzzati, Comparative Small-Angle X-Ray Scattering Studies
on Unacylated, Acylated and Crosslinked Escherichia-Coli Transfer Rna Val(I). J. Mol.
Biol. 71, 217–229 (1972).
51. M.Y. Pavlov, M.A. Sinev, A.A. Timchenko, and O.B. Ptitsyn, A Study of Apo-Forms
and Holo-Forms of Horse Liver Alcohol-Dehydrogenase in Solution by Diffuse-X-Ray
Scattering. Biopolymers 25, 1385–1397 (1986).
52. C.A. Pickover and D.M. Engelman, On the Interpretation and Prediction of X-Ray-
Scattering Profiles of Biomolecules in Solution. Biopolymers 21, 817–831 (1982).
302 W.T. Heller and G.A. Baker
53. D. Svergun, C. Barberato, and M.H.J. Koch, CRYSOL – A program to evaluate x-ray
solution scattering of biological macromolecules from atomic coordinates. J. Appl.
Crystallogr. 28, 768–773 (1995).
54. D.I. Svergun, I. Aldag, T. Sieck, K. Altendorf, M.H.J. Koch, D.J. Kane, M.B. Kozin, and
G. Gruber, A model of the quaternary structure of the Escherichia coli F-1 ATPase from
X-ray solution scattering and evidence for structural changes in the delta subunit during
ATP hydrolysis. Biophys. J. 75, 2212–2219 (1998).
55. D.I. Svergun, Richard S, M.H.J. Koch, Z. Sayers, S. Kuprin, and G. Zaccai, Protein
hydration in solution: Experimental observation by x-ray and neutron scattering. Proc.
Natl. Acad. Sci. USA 95, 2267–2272 (1998).
56. J. Zhou, S. Krueger, and S.K. Gregurick, A coarse graining approach to determine
nucleic acid structures from small angle neutron scattering profiles in solution. Nucl.
Acids. Res. 33, 6361–6371 (2005).
57. A.W. Ashton, M.K. Boehm, J.R. Gallimore, M.B. Pepys, and S.J. Perkins, Pentameric
and decameric structures in solution of serum amyloid P component by X-ray and
neutron scattering and molecular modelling analyses. J. Mol. Biol. 272, 408–422 (1997).
58. W.T. Heller, E. Abusamhadneh, N. Finley, P.R. Rosevear, and J. Trewhella, The solution
structure of a cardiac troponin C-troponin I-troponin T complex shows a somewhat
compact troponin c interacting with an extended troponin I-Troponin T component.
Biochemistry 41, 15654–15663 (2002).
59. P.V. Konarev, M.V. Petoukhov, and D.I. Svergun. MASSHA – a graphics system
for rigid-body modelling of macromolecular complexes against solution scattering data.
J. Appl. Crystallogr. 34, 527–532 (2001).
60. M.B. Kozin and D.I. Svergun, A software system for rigid-body modelling of solution
scattering data. J. Appl. Crystallogr. 33, 775–777 (2000).
61. F. Gabel, B. Simon, and M. Sattler, A target function for quaternary structural refinement
from small angle scattering and NMR orientational restraints. Eur. Biophys. J. Biophys.
Lett. 35, 313–327 (2006).
62. A. Grishaev, J. Wu, J. Trewhella, and A. Bax, Refinement of multidomain protein
structures by combination of solution small-angle X-ray scattering and NMR data.
J. Am. Chem. Soc. 127, 16621–16628 (2005).
63. M. Nollmann, W.M. Stark, and O. Byron, A global multi-technique approach to study
low-resolution solution structures. J. Appl. Crystallogr. 38, 874–887 (2005).
64. M.V. Petoukhov and D.I. Svergun, Global rigid body modeling of macromolecular
complexes against small-angle scattering data. Biophys. J. 89, 1237–1250 (2005).
65. M.V. Petoukhov, N.A.J. Eady, K.A. Brown and D.I. Svergun, Addition of missing loops and
domains to protein models by X-ray solution scattering. Biophys. J. 83, 3113–3125 (2002).
66. M.H.J. Koch, P. Vachette, and D.I. Svergun, Small-angle scattering: a view on the
properties, structures and structural changes of biological macromolecules in solution.
Q Rev. Biophys. 36, 147–227 (2003).
67. D.I. Svergun and M.H.J. Koch, Small-angle scattering studies of biological macromole-
cules in solution. Rep. Prog. Phys. 66, 1735–1782 (2003).
68. J.A. Langer, D.M. Engelman, and P.B. Moore, Neutron-Scattering Studies of Ribosome
of Escherichia-Coli – a Provisional Map of Locations of Proteins S3, S4, S5, S7, S8 and S9
in 30-S Subunit. J. Mol. Biol. 119, 463–485 (1978).
69. V. Ramakrishnan, M. Capel, M. Kjeldgaard, D.M. Engelman, and P.B. Moore,
Positions of Protein-S14, Protein-S18 and Protein-S20 in the 30-S Ribosomal-Subunit
of Escherichia-Coli. J. Mol. Biol. 174, 265–284 (1984).
70. V.R. Ramakrishnan, S. Yabuki, I.Y. Sillers, D.G. Schindler, D.M. Engelman, and
P.B. Moore, Positions of Proteins S6, S11 and S15 in the 30-S Ribosomal-Subunit of
Escherichia-Coli. J. Mol. Biol. 153, 739–760 (1981).
71. M.S. Capel, D.M. Engelman, B.R. Freeborn, M. Kjeldgaard, J.A. Langer,
V. Ramakrishnan, D.G. Schindler, D.K. Schneider, B.P. Schoenborn, I.Y. Sillers,
16 Visualizing Structures of Biological Macromolecules 303
S. Yabuki, and P.B. Moore, A Complete Mapping of the Proteins in the Small
Ribosomal-Subunit of Escherichia-Coli. Science 238, 1403–1406 (1987).
72. D.I. Svergun, J.S. Pedersen, I.N. Serdyuk, and M.H.J. Koch, Solution Scattering from
50s Ribosomal-Subunit Resolves Inconsistency between Electron-Microscopic Models.
Proc. Natl. Acad. Sci. USA 91, 11826–11830 (1994).
73. D.I. Svergun, N. Burkhardt, J.S. Pedersen, M.H.J. Koch, V.V. Volkov, M.B. Kozin,
W. Meerwink, H.B. Stuhrmann, G. Diedrich, and K.H. Nierhaus, Solution scattering
structural analysis of the 70 S Escherichia coli ribosome by contrast variation .1. Invar-
iants and validation of electron microscopy models. J. Mol. Biol. 271, 588–601 (1997).
74. D.I. Svergun, N. Burkhardt, J.S. Pedersen, M.H.J. Koch, V.V. Volkov, M.B. Kozin,
W. Meerwink, H.B. Stuhrmann, G. Diedrich, and K.H. Nierhaus, Solution scattering
structural analysis of the 70 S Escherichia coli ribosome by contrast variation .2. A model
of the ribosome and its RNA at 3.5 nm resolution. J. Mol. Biol. 271, 602–618 (1997).
75. D.I. Svergun and K.H. Nierhaus, A map of protein-rRNA distribution in the 70 S
Escherichia coli ribosome. J. Biol. Chem. 275, 14432–14439 (2000).
76. J.K. Krueger, G.A. Olah, S.E. Rokop, G. Zhi, J.T. Stull, and J. Trewhella, Structures of
calmodulin and a functional myosin light chain kinase in the activated complex:
A neutron scattering study. Biochemistry 36, 6017–6023 (1997).
77. J.K. Krueger, G. Zhi, J.T. Stull, and J. Trewhella, Neutron-scattering studies reveal
further details of the Ca2+/calmodulin-dependent activation mechanism of myosin
light chain kinase. Biochemistry 37: 13997–14004 (1998).
78. J.K. Krueger, N.A. Bishop, D.K. Blumenthal, G. Zhi, K. Beckingham, J.T. Stull, and
J. Trewhella, Calmodulin binding to myosin light chain kinase begins at substoichio-
metric Ca2+ concentrations: A small-angle scattering study of binding and conforma-
tional transitions. Biochemistry 37, 17810–17817 (1998).
79. J.K. Krueger, S.C. Gallagher, G. Zhi, R. Geguchadze, A. Persechini, J.T. Stull, and
J. Trewhella, Activation of myosin light chain kinase requires translocation of bound
calmodulin. J. Biol. Chem. 276, 4535–4538 (2001).
80. G.A. Olah, S.E. Rokop, C.L. WangA, S.L. Blechner, and J. Trewhella, Troponin-I
Encompasses an Extended Troponin-C in the Ca2+-Bound Complex – a Small-Angle
X-Ray and Neutron-Scattering Study. Biochemistry 33, 8233–8239 (1994).
81. G.A. Olah and J. Trewhella, A Model Structure of the Muscle Protein Complex 4ca2+-
Center-Dot-Troponin-C-Center-Dot-Troponin-I Derived from Small-Angle Scattering
Data – Implications for Regulation. Biochemistry 33, 12800–12806 (1994).
82. W.A. King, D.B. Stone, P.A. Timmins, T. Narayanan, von A.A.M. Brasch,
R.A. Mendelson, and P.M.G. Curmi, Solution structure of the chicken skeletal
muscle troponin complex via small-angle neutron and x-ray scattering. J. Mol.
Biol. 345, 797–815 (2005).
83. W.T. Heller, D. Vigil, S. Brown, D.K. Blumenthal, S.S. Taylor, and J. Trewhella,
C Subunits binding to the protein kinase a RI alpha dimer induce a large conformational
change. J. Biol. Chem. 279, 19084–19090 (2004).
84. A.M. Augustus, P.N. Reardon, W.T. Heller, and L.D. Spicer, Structural basis for the
differential regulation of DNA by the methionine repressor protein MetJ. J. Biol. Chem.
281, 34269–34276 (2006).
85. P. Callow, A. Sukhodub, J.E. Taylor, and G.G. Kneale, Shape and subunit organisation
of the DNA methyltransferase M.AhdI by small-angle neutron scattering. J. Mol. Biol.
369, 177–185 (2007).
86. D. Comoletti, A. Grishaev, A.E. Whitten, I. Tsigelny, P. Taylor, and J. Trewhella,
Synaptic arrangement of the neuroligin/beta-neurexin complex revealed by X-ray and
neutron scattering. Structure 15, 693–705 (2007).
87. D.L. Ho, W.M. Byrnes, W.P. Ma, Y. Shi, D.J.E. Callaway, and Z.M. Bu, Structure-
specific DNA-induced conformational changes in Taq polymerase revealed by small
angle neutron scattering. J. Biol. Chem. 279, 39146–39154 (2004).
304 W.T. Heller and G.A. Baker
88. S. Krueger, S.K. Gregurick, J. Zondlo, and E. Eisenstein, Interaction of GroEL and
GroEL/GroES complexes with a nonnative subtilisin variant: a small-angle neutron
scattering study. J. Struct. Biol. 141, 240–258 (2003).
89. A.E. Whitten, D.A. Jacques, B. Hammouda, T. Hanley, G.F. King, J.M. Guss,
J. Trewhella, and D.B. Langley, The structure of the KinA-Sda complex suggests an
allosteric mechanism of histidine kinase inhibition. J. Mol. Biol. 368, 407–420 (2007).
90. J. Zimmer, D.A. Doyle, and J.G. Grossmann, Structural characterization and
pH-induced conformational transition of full-length KcsA. Biophys. J. 90, 1752–1766
(2006).
91. W.T. Heller, J.K. Krueger. and J. Trewhella, Further insights into calmodulin-myosin light
chain kinase interaction from solution scattering and shape restoration. Biochemistry 42,
10579–10588 (2003).
Chapter 17
Neutron Imaging Applied to Plant Physiology
T.M. Nakanishi
Abstract Until recently, the in situ study of roots and dynamic measurements
of water uptake in plants or trees has relied on physical, and therefore invasive
and often destructive, sampling or sensor installation in the surrounding soil.
These methods can, at the very least, alter the equilibrium of the rhizosphere of
the plant or require the removal of the tree or plant. The sensitivity of neutrons
to hydrogen atoms provides a noninvasive, nondestructive experimental
approach to understand real-time processes in plants (e.g., water uptake).
This chapter attempts to illustrate the effectiveness of neutron imaging techni-
ques in studying plant physiology, soils, and wood.
17.1 Introduction
specific element from other neighboring ones. Neutron imaging techniques are
naturally suited for materials containing hydrogen atoms and other low atomic
weight attenuating materials, permitting nondestructive in situ measurements
of the spatial and temporal distribution of water (and other fluids rich in
hydrogen) in plants and trees as well as in soils.
case of a chrysanthemum leaf [13]. Although cold neutrons are limited in their
use because of a low transmission, small veins in the leaf can be observed, and
the contrast is approximately 70% higher than that of a thermal neutron image.
In addition, the range of changes in water thickness observed in the leaf is
smaller than 30 mm [15].
Neutron beam is perfectly suitable for analysis of the tissue level of the
sample, such as water distribution within a cut flower [3, 16–20] or a wood
sample [21–24]. However the resolution afforded by the technique, typically tens
of microns, is insufficient to study the process taking place within a single cell
because much higher resolution is needed for observation within a cell.
To obtain 2D neutron transmission images of living plants and their root
system in soil, samples are often grown in thin aluminum containers (2–3 mm
thick). The total cross section (absorption and scattering) of aluminum is
approximately 1/50 that of hydrogen; therefore, aluminum is almost transpar-
ent to thermal neutrons, and images of samples are not much affected by it.
Fig. 17.1 A neutron radiograph of a soybean plant root embedded in soil after (A) 8 and
(B) 15 days of growth; (C) magnification of the root; and (D) conversion of (B) to the
3D image
Fig. 17.2 Root development pattern. A: Soybean root development in the presence of 50 mM
of vanadium (white aggregate). B: Radish root development. (Left) Low-pH soil containing
23 mM of aluminum; (right) Modified lignin mixed with the soil shown in the image on the left
to reduce inhibition of growth by the aluminum
the polymer, but the side roots grows larger to compensate. On the other hand,
when the polymer ‘‘doped’’ with vanadium is placed on one side of the root, the
lateral root facing the polymer side does not grow, but the root system on the
other side is longer. The position at which growth stops and the direction of root
growth are dependent on the vanadium concentration inside the polymer [29]. It
is also shown that the amount of water uptake at a certain height of the main root
increases before side root growth begins.
Since root development is an indicator of the soil condition, neutron imaging
can be applied to evaluate soil conditioning agents as well as the effect of a
chemically modified fertilizer. To study lignin derivatives used as a soil con-
ditioning agent, a radish plant is employed to observe the root profile in an
acidic soil [30]. The root growth pattern captured in the neutron image in
Fig. 17.2B clearly shows that the modified lignin is effective as a soil condition-
ing agent. Subsequently, the total root length is measured through image
analysis, and the soil conditioning effect of different chemicals is analyzed.
The root pattern shown in Fig. 17.2B is similar to patterns found in lightning
or in the production of cracks in a pane of glass. There is no reproducibility in
the pattern—a fundamental feature, especially for roots, for maintaining bio-
logical diversity. Because of this, experimental errors cannot be quantified
properly when analysis of root growth is measured using invasive techniques
such as the line intersection method, in which a lattice board is placed on the
root and then removed from the soil, allowing the number of intersections
between the root and the lattice to be counted. Neutron imaging, because it is
nondestructive, can use the same sample to trace the root activity over time,
which reduces the fundamental errors that arise from using different samples.
Recent laboratory experiments reproducing water infiltration in soil, plant,
and the atmosphere have reported water redistribution processes as a function
310 T.M. Nakanishi
of different conditions such as, for example, root structure, soil structure, and
moisture [31].
Fig. 17.3 In situ neutron tomography of the root system of a soybean during different days
Fig. 17.4 Quantitative measurements of (left) side root growth according to their position
along the main root and (right) water distribution around the main root days after watering
the specimen. Depth is in mm
Fig. 17.5 Images of a carnation flower. A photograph is on the left, a neutron image in the
center, and a magnification of the neutron image on the right. In the left and center images,
the two flowers on the inside are controls and the two on the outside were dried before the
neutron radiograph was taken [36]
312 T.M. Nakanishi
Increasing the viscosity of water (i.e., xenon is dissolved under high pressure
into water) can help prolong the life of a carnation flower after it has been
harvested, as illustrated in Fig. 17.6A [36]. This helps control the metabolism of
the flower, which slows down the deterioration process mediated by an enzy-
matic reaction. An example of a 3D neutron imaging study is shown in
Fig. 17.6B. The flower respiration rate while conducting neutron imaging
during the senescent (i.e., aging) stage has been evaluated [3, 18, 19]. Further-
more, the analysis of 3D images of a carnation flower demonstrates the impor-
tance of water inside the ovule to maintain the flowering stage longer. It is
Fig. 17.6 Neutron image of a carnation flower. (A) Neutron radiographs of a carnation
flower after being supplied with water containing xenon. (B) CT images at different heights
along the flower head [36]
17 Neutron Imaging Applied to Plant Physiology 313
shown that the ovary and pistil areas continue to increase in size after the
respiration rate stops increasing [36].
The study of physiological activity of a tree mainly relies on knowing the water
distribution inside the wood. The example below illustrates the use of neutron
imaging on wood disks extracted from a tree cut only a few hours before
experiments, which therefore permits the observation of the green moisture
image [23, 24]. A stem section, covered as a precaution to prevent water loss, is
further cut to obtain 1 cm thick wood disks just prior to being exposed to
neutrons. Figure 17.7 shows the neutron images of disks from a Japanese
cypress (Chamaecyparis obtusa; Fig. 17.7A) and a Sugi (Cryptomeria japonica;
Fig. 17.7B), whose ages are estimated to be about 19 and 24 years, respectively.
The upper images in the figure are photographs of wood disks, and the
Fig. 17.7 Images of wood disks, each 1 cm thick. Photographs of disks are shown in the top
row; neutron images in the center row; and reconstructions of the middle images, with high
points corresponding to water-rich areas, in the bottom row
314 T.M. Nakanishi
center images are the corresponding neutron images. The lower images are
reconstructed ones based on the images in the center row with the degree of
whiteness indicated as height.
In the Japanese cypress, the neutron image shows higher water content in the
outer rings, adjacent to the bark. This zone includes the living phloem tissue and
recently developed xylem tissue that transports water from the roots to the
leaves. Both tissues have high water content. Water content declines radially
from the bark into the older, less functional, air-filled xylem tissue. Water
distribution in some areas shows a wavy pattern that might indicate tissue
damage or buildup of specific chemicals within the rings. In the heartwood
region, rings appear to have much lower water content, likely indicating buildup
of decay-resistant extractive chemicals. Besides the large hydrogen-rich outer
part of the disk, many rings corresponding to the annual rings of the wood are
shown to contain a higher level of water than their immediate surroundings.
Sugi is a popular wood in Japan for building houses and furniture. Although
there is only one species of Sugi, neighboring trees of the same cultivar show
different water content at the heartwood. It is not known what causes the
differences in moisture content at the heartwood. When lumber is processed,
residual moisture causes serious problems. When the moisture content at the
heartwood is high in green lumber, it is difficult to remove water completely
during the drying process. However, moisture is gradually lost after the lumber
is used in houses or furniture, causing the shape to warp over the years.
When neutron radiographs of wood discs are taken during the drying pro-
cess, the mechanism by which moisture is lost can be directly analyzed [24]. In
the case of Sugi, another feature in the disk is observed in the heartwood. As
shown in Fig. 17.7B, there is always a water-deficient area between heartwood
and sapwood. Because of the lighter color of this zone, it is called the white
zone. As the tree grows, the area of heartwood increases. However, the water
content throughout the heartwood is maintained at a high level as the tree
grows. It is not known how the water in sapwood moves into heartwood across
the white zone.
A rapid decrease in water content can be observed when the plant tissue
is injured. For example, neutron imaging shows that when Sugi wood is
inoculated with a canker fungus, a water deficiency occurs and spreads
from the inoculated site [30]. The dynamic process of water absorption in
a wood sample can be evaluated with neutron imaging techniques [22].
Imaging of solvent absorption and water-loss processes in small wood samples
(4 4 2.7 cm) has been reported [21] when acryl resin solvents are applied
to protect the surface of the wood, which is of importance for the development
of wood-protective materials for preserving wood art with historical and
cultural value.
Several other neutron imaging studies of wood can be found in [37–41].
17 Neutron Imaging Applied to Plant Physiology 315
Besides the examples presented above, neutron imaging has been used in many
other agricultural applications. The water uptake of seeds during germination
provides useful information, for example, in investigating the method of
storing seeds because in some seeds, physical damage or the storage condition
may change the germination rate (Fig. 17.8) [42]. Even at an early stage of
germination, the germination rate of seeds can be estimated using neutron
images of water absorption.
Recently, a tomographic study of corn kernels has aimed at understanding
the effect of Aspergillus flavus (A-flavus fungus) to resistant species. Anatomic
changes are observed when resistant inoculated and uninoculated species are
compared to their nonresistant inoculated counterparts [43].
Fig. 17.8 Water absorption process in five kinds of seeds. Photographs of the seeds are shown
in the upper left corner (the black bar is 5 mm long). Seeds are soaked in water and neutron
images are taken after 0, 2, 4, and 6 h (top to bottom). For the broad bean, images taken after 0,
2, and 4 h are shown
316 T.M. Nakanishi
References
1. K.A. Garrett and H. Berger, Atom. Rev. 15, 125 (1977).
2. J.F.W. Markgraf and R. Matfield, Neutron beam, in J.C. Domanus, ed., Practical
Neutron Radiography, Kluwar Academic Publishers, Dordrecht, Boston, London,
pp. 26–50 (1992).
3. U. Matsushima, Y. Kawabata, C-M. Sim, K-Y. Nam, and T. Nishizawa, Proc. VIIIth IS
Postharvest Phys. Ornamentals, Eds. N. Marissen et al., Acta Hortic. 669, 111 ISHS
(2005).
4. J.M. Hainsworth and L.A.G. Alymore, Soil Sci. Soc. Am. J. 50, 841 (1986).
5. C.J. Moran, A. Pierret, and A.W. Stevenson, Plant Soil 223, 101 (2000).
6. P.T. Callaghan, Diffusion-limited resolution, in P.T. Callaghan ed., Principles of Nuclear
Magnetic Resonance Microscopy, Oxford Science, Oxford, pp. 203–206 (1991).
7. T.M. Nakanishi and M. Matsubayashi, J. Plant Phys. 151, 442 (1997).
8. S.T. Willatt et al., Agron. J. 70(4) 581–586 (1978).
9. S.T. Willatt et al., Ann. Bot. 43, 415–422 (1979).
10. P. Couchat et al., Agron. J. 72, 321–324 (1980).
11. A. Hilger, N. Kardjilov, M. Strobl, W. Treimer, and J. Banhart, Physica B 385(86), 1213
(2006).
12. Y. Kawabata, M. Hino, T. Nakano, H. Sunohara, U. Matsushima, and P. Geltenbort,
Nucl. Instrum. Meth. Phys. Res. A 542, 61 (2005).
13. Y. Kawabata, U. Matsushima, T. Horie, T. Nakano, and R. Maruyama, J. Radioanal.
Nucl. Chem. 264, 319 (2005).
14. G. Frei and E. Lehmann, Proceedings of 8th World Conference of Neutron Radiography.
Gaithersburg, USA, p. 21, October 16–19 (2006).
15. U. Matsushima, K. Kawabata, M. Hino, P. Geltenbort, and B. Nicolai, Nucl. Instrum.
Meth. Phys. Res. A 542, 76 (2005).
16. U. Matsushima, T. Ooshita, T.M. Nakanishi, M. Matsubayashi, Y. Seo, and Y. Kawagoe,
J. Jpn. Soc. Agr. Machin. 62, 70 (2000).
17 Neutron Imaging Applied to Plant Physiology 317
R.C. Lanza
18.1 Introduction
the limitations of X-ray techniques and the need for SNM detection, it is now
useful to reexamine neutron methodologies, particularly imaging. A significant
number of neutron-based techniques have been proposed and are under devel-
opment for security applications, especially SNM detection, but describing how
they work is beyond the scope of the chapter. Instead, one particular approach
to neutron imaging, neutron resonance radiography (NRR), is discussed in
detail as it illustrates many of the issues connected with imaging and detection.
Fig. 18.1 Influence of decision threshold on false positive rate relative to true positive rate
Fig. 18.2 Receiver–operator curve (ROC). The dashed line represents random performance
322 R.C. Lanza
18.1.3 Practicality
The SNMs are the basic ingredients for nuclear weapons, and their control
and interdiction are highly important. The current prevailing model for
addressing threats associated with cargo-container-borne weapons of mass
destruction, especially nuclear weapons, could be characterized as a cus-
toms-based approach. Under this scheme, new systems using either active or
passive techniques to detect nuclear signatures, from hidden fissile material
would be integrated into the existing customs infrastructure of ports. The
18 Homeland Security and Contraband Detection 323
18.1.5 Explosives
18.1.6 Others
One highly developed neutron approach that has been tested is pulsed fast
neutron analysis (PFNA) [6–9]. In this system, shown in Figs. 18.3 and 18.4,
fast neutrons are used to produce gamma rays from neutron reactions with
elements of interest; for explosives, this means particularly C, N, and O, but
other elements are also detectable. The approach in principle is also capable of
detecting SNM by detection of delayed gammas and neutrons between pulses,
but this has not been its primary design goal.
The operation is as follows: Neutrons in the range of 8 MeV are generated
by an electrostatic accelerator by the reaction D(d,n)He. These neutrons
are collimated into a beam and the beam is mechanically scanned across a
container. At the same time, the accelerator is pulsed to produce 1-ns pulses
of neutrons with a repetition rate of 1 MHz. Gamma rays from inelastic
interactions are detected by an array of scintillation detectors. The time differ-
ence between the accelerator pulse and gamma detection is used to determine
the position along the beam where the interaction took place. The pulsed
approach also reduces background and clutter for another reason: the number
of neutrons produced per pulse is low enough so that there is typically only one
Fig. 18.3 Pulsed fast neutron analysis system; from Brown [7]
326 R.C. Lanza
neutron interaction in the container during a given 1-ns pulse; the next potential
event comes during the next beam pulse, 1 microsecond later.
The beam line is mechanically scanned so as to move the beam through the
object, therefore the gammas can be correlated with the position of interaction
in three dimensions, and a three-dimensional image is formed. The gamma-ray
spectrum obtained for each voxel is used to determine relative elemental content
in the object under inspection. For 8-MeV neutrons, a time-of-flight measure-
ment of 1 ns yields an accuracy of 5 cm in space. The physics for this
particular system has been refined over a number of years. Note that the choice
of neutron energy, 8 MeV, means that scattered neutrons will generally lose
enough energy so that they fall below the threshold for gamma production and
thus do not contribute to spatial clutter. Another point to observe is that the
detectors, are conventional scintillators (NaI) rather than high-resolution
detectors, such as HPGe. The choice of detectors was made after both experi-
mental observation and cost analysis. NaI has considerably higher efficiency
and count rate capability than does HPGe, but it has lower energy resolution.
This disadvantage is overcome by the use of the entire spectrum rather than
single peaks to determine elemental composition. Rather than fit the data to
individual peaks, this approach fits the entire spectrum to a linear combination
of whole spectra from each of the elements, a method long used in areas such as
well-logging and online coal and mineral analysis.
18 Homeland Security and Contraband Detection 327
Fig. 18.5 The Lawrence Livermore National Laboratory (LLNL) nuclear ‘‘car wash’’
Fig. 18.6 Delayed gamma spectrum with special nuclear material (SNM) (black) and without
SNM (gray)
production and deployment as an air cargo inspection system [13, 14]. Its basic
approach is to image a container using a combination of 14-MeV neutrons
generated from a sealed DT neutron generator and gamma rays from a 60Co
source (1.17, 1.33 MeV). The ratio, R, of the neutron and gamma attenuation
coefficients can be determined on a pixel by pixel basis. This ratio shows a
significant correlation with material type and can thus be used to discriminate
among various materials. Figure 18.7 shows this ratio for various materials of
interest in cargo. The ratio is color-coded and displayed as an image for visual
inspection by the operator (Fig. 18.8).
Fig. 18.7 Ratio (R) of neutron to gamma attenuation coefficients for various materials
18 Homeland Security and Contraband Detection 329
Fig. 18.8 Optical picture (left) and R-ratio image of aircraft container (right)
Fig. 18.9 Total neutron cross section for H, C, N, and O in the MeV energy region. With the
exception of hydrogen, clear and unique resonance structures exist for each element
330 R.C. Lanza
Ix
ln ¼ Na a ðEÞxa ;
Iair
where lnðIx =Iair Þ represents the ratio of the neutron beam attenuated by
the object to the unattenuated beam, Na represents the atom density of
element a, and a(E) represents the total neutron cross section of material a
at energy E. All equations in this section apply to a single pixel. Aside from
the intrinsic energy variation in the neutron cross section, factors related to
the neutron production mechanism will introduce a smearing in the cross
section as a function of energy. The notation 1a is used to represent the
effective neutron cross section for material a at energy setting 1. Notation-
ally, superscripts will represent the energy setting while subscripts represent
the material. The measurement lnðIx =Iair Þat energy setting 1 is defined as
b1. Finally, Na xa is combined into a . In a two-dimensional projection
image, the attenuation image is the sum of elemental contents. For a
material comprising several different elements (a, b, c, d), the observed
attenuation becomes
18.2.5 Apparatus
Detector
source
Position: 60 Deg. rotated
Detector array
Neutron beam
RFQ accelerator
LD-3
Cont
ainer
Object
Collimators
Fig. 18.10 Neutron resonance radiography imaging system. The deuteron accelerator on the
rotating platform (in gray) rotated to the 608 position. Front and rear object collimators
are shown with the object being imaged (typically an LD-3 air cargo container) between
the collimators, resting on the object transport system. The detector array is behind the
rear object collimator
prevents stray radiation from the source from reaching the object, and the
rear object collimator prevents radiation scattered from the object from
reaching the detectors. The neutron beam is defined by aligning the
adjustable source collimator with the fixed object collimators and the detec-
tor array. The object to be imaged (typically a standard aircraft LD-3
container) is transported through the beam on a motorized trolley. The
system is housed in a warehouse with minimal climate control (heat only)
and no running water. The prototype occupies a radiation safety exclusion
zone of about 25 by 15 m.
The detector system is based on a combination of plastic scintillator and
photomultiplier tube in which the fast neutron is detected by proton recoil in the
plastic. Typically, gamma background signals are larger than those produced
by neutrons, and the electronics are based on a window discriminator for
gamma rejection.
The success of the NRR technique is largely determined by the accuracy of
the elemental cross sections. Initially, Monte Carlo methods were used to
18 Homeland Security and Contraband Detection 333
simulate neutron scattering on elements. The cross sections extracted from the
simulations were then used to study the possibility of creating elemental maps.
However, it was quickly realized that the ideal detector in the type of back-
ground-free environment assumed in the Monte Carlo studies does not reflect
actual experimental conditions. Hence, a precise experimental extraction of the
elemental cross sections was necessary as photon background can significantly
change the extracted neutron cross section. Overwhelming photon background
washes out the resonant features in the cross sections and makes NRR difficult;
reducing the photon-to-neutron ratio is essential to the success of the method.
Figure 18.11 shows experimentally determined carbon and nitrogen cross
sections. All elemental cross sections were lower than those calculated using
Monte Carlo simulations, largely as a result of gamma contamination. How-
ever, the resonant structure of these elemental cross sections is still observed,
allowing for a good elemental separation in compounds.
Once the elemental calibration had been completed, a set of angles was
selected for performing radiographic analysis. The best set of angles is the one
that allows the most precise determination of elemental composition in com-
pounds. Analytically, these angles would correspond to the most pronounced
features on elemental cross sections. For example, the oxygen cross section
exhibits the most unique behavior at high angles (100 through 1158), whereas
the nitrogen cross section has a distinct structure at lower angles (around 358).
The carbon cross section is most distinctive at around 908. Hydrogen and, to a
lesser extent, silicon cross sections exhibit no distinct resonance structure. A non-
trivial relationship among these five cross sections in determining elemental
composition is best studied quantitatively using a computer optimization code.
To find the best set of angles, compounds composed solely of the basis
materials (such as melamine, H6C3N6) were scanned at 58 increments in
accelerator angle. To resolve a set of five basis elements, scans of six or
more angles were required. Since a larger number of angles linearly increase
the scan time, it was found that seven angles represented a good balance
Fig. 18.11 Measurement of total neutron cross section for carbon (left) and for nitrogen
(right) as a function of accelerator angle (degrees)
334 R.C. Lanza
between quality and scan time. For each unique set of 7 angles picked from
a set of 24, the square of the difference between radiographically determined
and known elemental composition in each compound was calculated. The
sum of these differences was then minimized to produce the best set of
angles. Figure 18.12 (left) shows the results of the angle optimization
studies, comparing the theoretical atom fraction for melamine with the
one determined from the set of optimal angles. The resulting calibration
was also tested with various compounds not included in the optimization. A
scan of one such compound, ammonium nitrate (NH4NO3), in air is shown
in Fig. 18.12 (right). The relative amounts of hydrogen and nitrogen are in
good agreement with expected values. The amount of oxygen is somewhat
lower than the expectation, but still pronounced in comparison to carbon
and ‘‘other’’ represented by silicon.
Fig. 18.12 Left: Atom fraction of melamine sample as determined from the calibration
procedure. The basis elements are H, C, N, O, and Si. The black (right) bars are the
expected values and the gray (left) bars represent the best values obtained during
the optimal angle selection. The correctness of the melamine atom fraction was one of
the constraints of the angle selection. Right: Atom fraction of ammonium nitrate sample.
The basis elements are H, C, N, O, and Si. The black (right) bars are the expected values
and the gray (left) bars represent the best values obtained after the optimal angle
selection. The correctness of the ammonium nitrate atom fraction was not constrained
by the angle selection
The results in Fig. 18.12 are derived from blocks of pixels on isolated
objects rather than from a more realistic situation with multiple elements.
Figures 18.13 and 18.14 show images developed by the system together with
their elemental maps. Images presented are rotated by 908 from their actual
scan direction: top to bottom in the image is left to right in the scan; left to
right in the image is bottom to top in the scan. The leftmost part of
Fig. 18.13 represents the attenuation image at one of the 7 energies scanned
18 Homeland Security and Contraband Detection 335
with the presence of 12 different targets. Also shown are the hydrogen and
oxygen elemental maps and finally, on the right, the result of a simple
algorithm that correctly identifies the water sample by identifying the coin-
cident presence of hydrogen and oxygen. Similarly, Fig. 18.14 shows the
identification of the nitrogen-rich melamine.
In the next set of images, Figs. 18.15 and 18.16, a series of targets are
superimposed into typical air cargo contents. As an example of algorithm
development in this project, a preliminary algorithm was developed to detect
graphite. The carbon elemental image appears to be very stable and have low
noise. The carbon block is detected with high efficiency in an environment of
typical air cargo contents is seeded with other potential targets.
336 R.C. Lanza
Fig. 18.15 Automated detection of carbon object in an air cargo of cut flowers. The left pane is
the carbon image and the right pane is the sum of the elemental map for all five basis elements.
The carbon block alone is correctly identified by the algorithm among the clutter (100 cm
hydrangeas) and other seeded targets
Acknowledgments The work on NRR was a collaboration between the MIT Department of
Nuclear Science and Engineering, the MIT Laboratory for Nuclear Science, L3-Communica-
tions, and LLNL. The author especially thanks David Perticone (L-3), who was the principle
investigator on the NRR project and was responsible for much of the progress made on it, as
well as Vitalyi Ziskin (L-3 and MIT), Gongyin Chen (L-3 and Varian), Whitney Raas (MIT),
and Gordon Kohse (MIT) for their contributions throughout the project. Jim Hall (LLNL),
Brian Rusnak (LLNL), and John Watterson (University of the Witwatersrand) made invalu-
able contribution in the early parts of the project.
Others include the staff of the MIT-Bates Laboratory for their excellent contributions to
the construction and operation of the NRR prototype, especially Bob Fisk, Peter Binns, Jim
Kelsey, Peter Goodwin, Christoph Tschalaer, and Jan Van Der Laan; and the following
engineering staffs of L-3 Communications for the design of system components: Tim Hart,
Ken Prather, Jim Karon, Tony Antoniou, John Price, Keith McClelland, Brain Pearson,
Steve McDevitt, Jeff Stillson, and Sal Gargiulo.
References
1. D.L. Chichester and J.D. Simpson, Compact accelerator neutron generators, The Indus-
trial Physicist, Dec-Jan 2003/2004 9(6), 22–25 (2004).
2. ANSI 43.14, Radiation Safety Guidelines for Active Interrogation Systems for Security
Screening of Cargo. American National Standards Institute (2008).
3. J. Zukas and W. Walters, eds., Explosive Effects and Applications, Springer-Verlag, New
York, NY (1997).
4. A. Buffler, Contraband detection with fast neutrons, Rad. Phys. Chem. 71(3–4), 853–861
(2004).
5 R. Lanza, Neutron techniques for detection of explosives in Y. Yinon, ed., Counterterrorist
Detection Techniques of Explosives, Elsevier, Amsterdam, pp. 131–155 (2007).
338 R.C. Lanza
A E
Absorption, 5, 7, 10, 11, 22, 33, 36, 43, 47, 48, Epithermal neutrons, 48, 234, 235, 243–244,
49, 68, 70, 71, 75, 81, 82, 85, 92, 94, 97, 255, 256, 257, 258, 263
103–106, 110, 125–126, 140–142, 146,
164, 175, 178, 183, 194–195, 210, 221, 233,
234, 235, 243, 253, 258, 307, 314, 315 F
Accelerator, 14, 15–21, 22, 27, 28, 68–70, Fission, 4, 14, 15, 16, 21–22, 27–28, 48, 70,
168, 230, 263, 270–272, 281, 322, 325, 110, 157, 258, 327
331–333, 337 Flow, 69, 75, 78, 97–98, 196, 198, 202–203,
204–205, 211, 212, 215, 218, 221, 254, 319
Flower, 306–307, 311, 312–313, 336
B Flux, 5–6, 13–14, 15, 16, 17, 19–21, 22, 23, 24,
Biology, 122, 293, 316 25, 26, 27, 28, 29, 32, 34, 35, 37, 39, 49, 68,
Biomedical, 211, 253, 262, 265 70, 76–77, 82, 83, 106, 110, 114, 125–126,
Botany, 305–316 174, 175, 183–184, 186, 197, 199–200, 218,
219–220, 224, 243, 257, 270–271, 272, 278,
C 281, 282, 331
Carnation, 311–312 Fuel cell, 8, 75, 78–79, 97–98, 106, 109–110,
Casting, 209–226 192–199, 200–201, 202–205, 206, 210, 211
Charge-coupled device (CCD), 52–53, 76, 82, Functional, 254, 269–270, 314
84–85, 111–114, 118, 199, 209, 218, 219,
220, 221, 220, 306
G
Cold Neutrons, 47, 58, 94, 124, 177, 179–180,
Gas neutron detector, 56–58
209–210, 231, 236, 242, 254–255, 306–307
Granular, 212, 216, 217–218
Composites, 110, 221
Growth, 247, 307–311
Contrast agent, 74, 255, 257, 262
Cross section, 5–11, 18–19, 21–22, 24–25, 36,
42, 47, 48, 52, 70, 71, 73, 82, 94–95, 102, H
110, 111, 116, 117, 121, 122, 125, 130–131, Hydrogen storage, 181, 192–199, 202,
140, 143, 156, 157, 158, 164, 173–174, 206, 210
182–183, 186, 194, 195, 197, 199, 212, 213,
223, 231, 235, 244, 255, 262, 307, 329–330,
332–333 I
Imaging, 4, 8, 10, 11, 13, 26, 27, 29, 67–68, 69,
69, 72, 73, 76–78, 82, 85, 104, 109–110,
D 117–124, 129–130, 131, 136, 142, 149, 156,
2D imaging, 85, 86, 88 164, 168, 171–186, 194–195, 199–203,
3D imaging, 85, 86, 92, 130, 200, 230, 234, 211, 297–299, 306–313, 314, 315–316,
307–308, 312 320, 324, 329, 332, 337
339
340 Index
Imaging neutron detector, 47–60 69–70, 72, 73–74, 82, 83–84, 85, 86–87,
Inelastic scatter, 22, 36, 157–158, 176, 265, 93–94, 95, 97, 100, 101, 104, 106, 109–110,
266, 267–268, 271–272 111, 122, 124, 129, 130, 131, 132, 136–137,
In situ, 106, 168, 194–195, 205, 212, 221, 139, 145, 146, 149, 156–157, 158, 160,
306, 310 161–163, 164–165, 171–186, 191–192,
Instrument, 13–14, 16, 17, 20, 22, 23, 24, 27, 194, 209–211, 213, 214–215, 216, 218,
34, 35, 36, 38, 42, 44, 45, 47, 52, 53–54, 56, 219–220, 222, 230–240, 241, 242,
96, 102, 157, 168, 173–174, 175, 176–177, 243–244, 249, 253, 254–255, 256–257,
179, 183–184, 185, 216, 226, 229, 233, 258, 260, 262, 263, 265, 266, 270–272, 280,
234–235, 247, 248, 265, 293, 322 281, 289–290, 305–307, 310, 315–316,
325–326, 327, 332, 337
Neutron stimulated emission computed
M tomography (NSECT), 258, 266,
Magnetic fields, 4, 11, 87, 106, 171–172, 173, 267–268, 269–270, 272, 273–274, 275,
175, 182, 185 276, 277–278, 279, 280–282
Magnetic resonance imaging, 171–186, 254 Numerical, 73, 122, 130, 183, 195–196,
Materials, 4, 6, 8–10, 11, 14, 22, 24–25, 33, 217–218
40–41, 48, 52, 68, 70, 74–75, 78, 94, 96, 99,
109–110, 111, 115, 124, 125–126, 131, 139,
140, 143, 154, 156, 166, 168, 172, 176, 181, P
191–192, 194, 197, 199, 200, 206, 209–210, Plant physiology, 305–316
211, 212, 215–218, 219, 221, 223, 224–225, Polarized neutrons, 11, 82, 86, 106,
229, 230, 231, 234, 235, 244, 245–246, 249, 171–186
253, 254, 255, 289, 291, 305–306, 314, Portable source, 28
319–320, 322–323, 324, 328, 333–334 Proton, 3–4, 8, 14, 15–21, 26, 47, 78, 156, 182,
Medical imaging, 86, 123, 254, 262, 263, 265, 183, 192, 193, 194, 270, 332
266, 268–270 Proton Exchange Membranes (PEM),
Metals, 4, 42, 44, 75–76, 95, 96, 111, 142–143, 192–194, 196–197, 200–201, 203–205
144, 154, 194–195, 211–215, 223, 230, 245, Pulse, 19, 20, 21, 22, 25, 26, 35, 58, 178,
262, 311 179–180, 181, 183, 185, 186, 218, 325–326
Modeling, 106, 112–114, 192, 215, 217, 219, Pulse width, 23, 24, 26
289–299, 327 Pyrolysis, 212, 214, 215
Moderators, 4, 14, 15, 16, 20, 22, 23, 24, 25,
26, 32, 34, 70–71, 72, 239, 257
Multi-Phase, 214–215 R
Reactor source, 10, 14, 15, 17, 26–27, 149, 232
Reflectors, 15, 22, 24
Roots, 154–155, 306, 307–309, 310–311,
N 314, 316
Neutron converter, 47–49, 58
Neutron detector, 28, 47, 53, 56, 58, 59–60,
85–86, 179 S
Neutron energy, 4, 7, 10, 11, 18, 21, 22, 23, Sand, 211–212, 216–217, 307, 310
25, 26, 28, 111, 140, 149, 176, 194, 200, Scattering, 4, 6, 7–10, 11, 13–14, 15, 17,
226, 231–232, 271, 272, 326, 331–332 21–22, 26, 29, 34, 36, 39–40, 41–42, 48, 49,
Neutron imaging, 4, 5, 11, 14, 17, 20, 28, 47, 56, 68, 82, 94, 100–106, 110, 111, 122, 131,
50, 52, 72, 77, 78–79, 82–83, 106, 109–126, 143–145, 155, 156, 157–158, 159, 160, 161,
129, 179, 191–206, 209–226, 229–249, 162, 163, 164, 166, 168, 175, 176, 177, 183,
253–263, 289, 305–316, 319, 323, 324, 194, 239, 253–254, 255, 273, 289–290, 291,
329–330 293, 296, 297, 333
Neutron production, 15–21, 28, 330 Scintillation neutron detector, 52–53, 84,
Neutrons, 4, 5–6, 7, 8, 11, 13–14, 15–16, 18, 325–326
19, 21–22, 24–25, 26, 27–28, 31, 32, 33, Small particulate gadolinium oxide
35–36, 39, 40–41, 42, 43, 44, 47–60, 67–68, (SPGO), 262
Index 341
Soil, 306, 307–311, 316 Time-of-flight (TOF), 10, 14, 19, 21, 26, 27,
Solid state neutron detector, 58–60 36, 39, 47, 52, 53, 56, 59, 124, 178, 180,
Spallation, 4, 14, 16, 18, 110 244, 271, 326
Spallation source, 14, 19, 20, 21–25, 26, 27, Tomography, 11, 27, 53, 54, 77, 81–106, 111,
29, 70, 125–126, 157, 168, 221 115, 117, 130, 132–136, 140, 178, 181, 195,
Spectroscopy, 49, 166–167, 181–183, 210, 196, 200–201, 202–203, 210, 216, 219, 223,
258, 267–268, 269, 271, 274, 279, 306 230, 232, 233, 234, 240–242, 245, 247, 249,
Spin-echo, 175, 176–177 254, 258, 261, 265–282, 306, 310
Structural, 8, 122, 123, 137, 145, 166–167,
168, 210, 211, 254, 269–270, 289–290, 293,
294, 295, 297 W
Water dynamics, 196–197, 200
Water uptake, 306, 307, 309, 315
T Wood, 230, 254, 307, 313–316
Thermal Neutrons, 7, 8, 15, 21–22, 28, 29, 47,
51–52, 59, 68, 70, 78, 132, 149, 153–168,
194, 202–203, 209–210, 213, 218, 220, 232, X
253, 254–255, 256–257, 258, 260, 262, 307 X-ray imaging, 117, 260, 306, 324