WTW Inolab PH Level2
WTW Inolab PH Level2
inoLab
pH Level 2
Laboratory pH meter
ba12218e04 10/2002 71
Accuracy when The use of advanced technology and the high quality stan-
going to press dard of our instruments are the result of continuous devel-
opment. This may result in differences between this
operating manual and your instrument.
We cannot guarantee that there are absolutely no errors in
this manual. We are sure you will understand that we cannot
accept any legal claims resulting from the data, figures or
descriptions.
72
List of contents
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
1.1 Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
1.3 Sockets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . .80
2.2 General safety instructions . . . . . . . . . . . . . . . . . .80
3 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . 83
3.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . .83
3.2 Initial commissioning . . . . . . . . . . . . . . . . . . . . . . .83
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
4.1 Switch on the instrument . . . . . . . . . . . . . . . . . . . .85
4.2 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86
4.2.1 Measuring the pH value . . . . . . . . . . . . . . .88
4.2.2 Measuring the Redox voltage . . . . . . . . . . .89
4.2.3 Transmitting measured values . . . . . . . . .90
4.3 Calibrating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
4.3.1 Calibration interval (Int 3) . . . . . . . . . . . . . .94
4.3.2 AutoCal TEC . . . . . . . . . . . . . . . . . . . . . . .95
4.3.3 AutoCal DIN . . . . . . . . . . . . . . . . . . . . . . . .98
4.3.4 ConCal . . . . . . . . . . . . . . . . . . . . . . . . . . .101
4.4 Storing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .104
4.4.1 Manual storage . . . . . . . . . . . . . . . . . . . . .104
4.4.2 Switching on AutoStore (Int 1) . . . . . . . . .106
4.4.3 Outputting the data storage . . . . . . . . . . .108
4.4.4 Clearing the storage . . . . . . . . . . . . . . . . .111
4.5 Data transmission . . . . . . . . . . . . . . . . . . . . . . . .112
4.5.1 Data transmission interval (Int 2) . . . . . . .112
4.5.2 Recorder (analog output) . . . . . . . . . . . . .114
4.5.3 PC/external printer (RS232 interface) . . .115
4.5.4 Remote control . . . . . . . . . . . . . . . . . . . . .116
4.6 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . .117
4.7 Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .121
73
List of contents
8 Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
9 Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
74
Overview
1 Overview
The compact inoLab pH Level 2 precision pH meter lets you
perform pH measurements rapidly and reliably. The inoLab
pH Level 2 provides the highest degree of operating com-
fort, reliability and measuring safety for all applications.
1 Keypad
2 Display
3 Sockets
Note
The measuring instrument can also be delivered as part of
a set.
Information on this and other accessories is available in the
WTW catalog LABORATORY AND FIELD INSTRUMENTATION or
via the Internet.
75
Overview
1.1 Keyboard
2 4
1 5
76
Overview
1.2 Display
Status line
Function and
temperature display
1.3 Sockets
4
3
1 2
Connectors:
1 pH electrode
2 Temperature probe
3 Reference electrode
4 RS 232 interface / analog output
5 Plug-in power supply
Caution
Only connect probes to the instrument that cannot feed ex-
cessive voltages or currents (> SELV and > circuit with cur-
rent limiter).
77
Overview
78
Safety
2 Safety
This operating manual contains basic instructions that you
must follow during the commissioning, operation and main-
tenance of the instrument. Consequently, all responsible
personnel must read this operating manual before working
with the instrument.
The operating manual must always be available within the
vicinity of the instrument.
Target group This measuring instrument was developed for use in the lab-
oratory.
Thus, we assume that, as a result of their professional train-
ing and experience, the operators will know the necessary
safety precautions to take when handling chemicals.
Symbols used
Caution
indicates instructions that have to be followed to prevent
damage to your instrument.
Warning
indicates instructions that have to be followed to protect
yourself and the instrument from dangerous electrical volt-
age.
Note
Indicates notes that draw your attention to special features.
Note
Indicates cross-references to other documents, e.g. appli-
cation reports, operating manuals of combination elec-
trodes, etc.
79
Safety
Function and opera- The smooth functioning and operational safety of the instru-
tional safety ment can only be guaranteed if the generally applicable
safety measures and the specific safety instructions in this
operating manual are followed.
Caution
The instrument is only allowed to be opened by personnel
authorized by WTW.
80
Safety
Obligations of the The operator of this measuring instrument must ensure that
operator the following laws and guidelines are observed when using
dangerous substances:
l EEC directives for protective labor legislation
l National protective labor legislation
l Safety regulations
l Safety datasheets of the chemical manufacturer
81
Safety
82
Commissioning
3 Commissioning
3.1 Scope of delivery
l Laboratory measuring instrument, inoLab pH Level 2
l Plug-in power supply
l Operating manual and short manual in the cover
l 4 x type AA Mignon 1.5 V batteries
83
Commissioning
Connecting the The plug-in power supply supplies the pH meter with low
plug-in power voltage (7.5 V ... 12 V DC).
supply
Warning
The line voltage on site must lie within the input voltage
range of the original plug-in power supply unit (see chapter
7 TECHNICAL DATA).
Caution
Use original plug-in power supplies only
(see chapter 7 TECHNICAL DATA).
1 Insert the plug (1) into the socket (2) of the pH meter.
2 Connect the original WTW plug-in power supply (3) to
an easily accessible mains socket.
Note
You can also perform measurements without a plug-in pow-
er supply.
84
Operation
4 Operation
4.1 Switch on the instrument
Note
The instrument has an energy saving feature to avoid
unnecessary battery depletion.
The energy saving feature switches the instrument off if no
key has been pressed for an hour.
The energy saving feature is not active:
85
Operation
4.2 Measuring
Preparatory Perform the following preparatory activities when you want
activities to measure:
Note
Incorrect calibration of the pH electrode will result in incor-
rect measured values. Therefore, regularly perform calibra-
tion before measuring.
Caution
When connecting an earthed PC/printer, measurements
cannot be performed in earthed media as incorrect values
would result.
The RS232 interface is not galvanically isolated.
86
Operation
Note
The pH meter automatically recognizes the type of the tem-
perature probe used. As a result, you can connect elec-
trodes with the NTC30 or Pt1000.
Note
When calibrating without a temperature probe, set up the
current temperature of the respective buffer solution manu-
ally by pressing the u d keys.
87
Operation
88
Operation
Note
The current AutoRead measurement (with acceptance of
the current value) can be terminated at any time by pressing
g.
Note
Redox electrodes are not calibrated. However, you can
check Redox electrodes using a test solution.
89
Operation
Note
If you connect a recorder (analog output), the output to the
digital output is switched off.
90
Operation
4.3 Calibrating
Why calibrate? pH electrodes age. This changes the asymmetry (zero
point) and slope of the pH electrode. As a result, an inexact
measured value is displayed. Calibration determines the
current values of the asymmetry and slope of the electrode
and they are stored in the instrument.
Thus, you should calibrate at regular intervals.
91
Operation
Note
You can automatically print out a calibration protocol after
the calibration. To do so, connect a printer according to sec-
tion 4.5.3 before the calibration. After a valid calibration, the
protocol is printed.
Sample printout:
CALIBRATION PROTOCOL
02.03.99 14:19
Device No.: 12345678
CALIBRATION pH
Cal Time: 01.03.99 / 15:20
Cal Interval: 7d
AutoCal DIN Tauto
Buffer 1 1.679
Buffer 2 4.008 *
Buffer 3 6.865
Buffer 4 9.180 *
C1 174.1mV 25.0°C
C2 -133.3mV 25.0°C
S1 -59.4 mV/pH
ASY1 - 4 mV
Probe: +++
92
Operation
93
Operation
Note
To ensure the high measuring precision of the measuring
system, perform a calibration after the calibration interval
expires.
Setting the The calibration interval (Int 3) is set to 7 days in the factory.
calibration interval The interval can be changed (1 ... 999 days):
94
Operation
Note
Steps 2, 6 and 13 are not required if you use a temperature
probe.
95
Operation
96
Operation
Note
You can also prematurely terminate the three-point calibra-
tion by pressing m. The values of the two-point calibration
for the slope and asymmetry are then stored.
97
Operation
Note
Steps 2, 6 and 13 are not required if you use a temperature
probe.
98
Operation
99
Operation
Note
You can also prematurely terminate the three-point calibra-
tion by pressing m. The values of the two-point calibration
for slope and asymmetry are retained.
100
Operation
4.3.4 ConCal
Two-point Use two buffer solutions for this procedure:
calibration
l pH 7.0 ± 0.5
l any other buffer solution
Note
Steps 2 and 9 are not required if you use a temperature
probe.
101
Operation
102
Operation
Single-point Use a buffer solution in the range pH = 7.0 ± 0.5 for this pro-
calibration cedure.
Note
Only the electrode asymmetry is determined in single-point
calibration. The slope of the last two-point calibration is
retained.
Note
Step 2 is not required if you use a temperature probe. The
TP message indicates an active temperature measurement.
103
Operation
4.4 Storing
The pH meter has an internal data storage device. Up to 800
data records can be stored in it.
A complete data record consists of:
l Memory location
l Date/time
l Measured value
l Temperature
l Temperature measurement procedure
l I.D. number
l Manual storage
l Switching on the AutoStore function (Int 1).
104
Operation
2 Confirm with g.
The display changes to the input of the I.D. number.
StoFull message This message appears if all 800 memory locations are full.
105
Operation
Setting the storage The storage interval (Int 1) is set to OFF in the factory.
interval Thus, the AutoStore function is switched off.
To switch the function on, set up a time interval (5 s, 10 s,
30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min):
106
Operation
Note
The AutoStore function is interrupted if you perform other
functions, e.g. output data storage.
After completing the other function, the AutoStore function
continues. However, as a result, gaps can occur in the
recording of the measured values.
107
Operation
l display
l interface
Outputting to the 1 Press the r key repeatedly until Sto disp appears
display on the display.
2 Press the g key.
A measured value appears on the display.
The memory location of the data store appears for
approx. 2 s. This is followed by the corresponding
temperature.
108
Operation
Note
If you want to find a specific parameter (e.g. date), proceed
as follows:
Outputting to the 1 Press the r key repeatedly until Sto SEr appears
interface on the display.
Note
You can cancel the transmission by pressing m or g.
109
Operation
Sample printout:
CALIBRATION PROTOCOL
02.03.99 14:19
Device No.: 12345678
CALIBRATION pH
Cal Time: 01.03.99 / 15:20
Cal Interval: 7d
AutoCal DIN Tauto
Buffer 1 1.679
Buffer 2 4.008 *
Buffer 3 6.865
Buffer 4 9.180 *
C1 174.1mV 25.0°C
C2 -133.3mV 25.0°C
S1 -59.4 mV/pH
ASY1 - 4 mV
Probe: +++
No. 1:
09.03.99 17:10
pH 10.01 25 °C
Tman AR
Ident : 1
No. 2:
09.03.99 17:11
pH 10.01 24,7 °C
Tauto AR
Ident : 1
No. 3:
09.03.99 17:12
305 mV
Tauto
Ident : 13
l calibration protocol
l storage contents
110
Operation
Note
The Clear store function only appears if data records have
already been stored in the storage. Otherwise, the pH meter
automatically changes to the measuring mode.
Note
The calibration data remain in the storage and can be called
up via the calibration protocol.
111
Operation
Note
The setting of the interval (Int 2) only has an effect when the
storage interval (AutoStore function) is switched off.
112
Operation
Note
If the AutoStore function is active, the data transmission is
performed according to the setting of the storage interval
(Int1). Set the storage interval (Int1) to OFF to activate the
data transmission interval (Int2).
Note
You can also set the data transmission interval (Int2) in the
Configuration menu (see page 117).
113
Operation
1 Free
Socket assignment 2 Plug coding
1 2
3 Ground
4 3 4 Analog output
RS 232
low impedance electrode signal
REC corresponds to 50 ... 62 mV/pH
(internal resistance < 5 Ohm)
Note
Activate the analog output by connecting 2 and 3.
114
Operation
Note
The RS232 interface is not galvanically isolated.
If it is connected to an earthed PC/printer, measurements
cannot be made in earthed media as this would give incor-
rect results!
1 CTS
Socket assignment
2 RxD
1 2
3 Ground
4 3 4 TxD
RS 232
REC
115
Operation
Note
A more detailed description is provided within the scope of
delivery of the communication package.
116
Operation
4.6 Configuration
You can adapt the pH meter to your individual requirements.
To do this, the following parameters can be changed (the
status on delivery is marked in bold):
Note
You can leave the configuration menu at any time. Parame-
ters that have already been changed are stored.
To do this, press the m key.
Baud rate
117
Operation
Data transmission
interval
Calibration
interval
118
Operation
Temperature unit
119
Operation
Adjusting the
1 Press and hold down the g key.
resolution
2 Press the m key.
The measured values appear with the higher resolu-
tion on the display, e.g. pH = 4.012.
3 Press the g key and m key again.
The measured values with the lower
resolution appear on the display, e.g. pH = 4.01.
120
Operation
4.7 Reset
You can reset (initialize) measuring and configuration
parameters separately from one another.
Measuring The following measuring parameters (pH InI) are reset to the
parameters values they had on delivery:
Measuring mode pH
Asymmetry 0 mV
Slope -59.16 mV/pH
Calibration procedure AutoCal TEC
Temperature, manual 25°C
Resolution of pH display 0.01
Note
When the measuring parameters are reset, the calibration
data are lost. After the parameters have been reset, cali-
brate!
121
Operation
122
Maintenance, cleaning, disposal
Caution
Make sure that the poles of the batteries are the right way
round. The ± signs in the battery compartment must corre-
spond to the ± signs on the batteries.
Only use leakproof alkaline manganese batteries.
Note
See the relevant operating manual of the electrode for in-
structions on maintenance.
123
Maintenance, cleaning, disposal
5.2 Cleaning
Occasionally wipe the outside of the measuring instrument
with a damp, lint-free cloth. Disinfect the housing with iso-
propanol as required.
Caution
The housing is made of synthetic material (ABS). Thus,
avoid contact with acetone or similar detergents that contain
solvents. Remove any splashes immediately.
5.3 Disposal
Packing The measuring instrument is sent out in a protective trans-
port packing.
We recommend: Keep the packing material. It protects the
instrument against damage during transport.
Batteries This note refers to the battery regulation that applies in the
Federal Republic of Germany. We would ask end-consum-
ers in other countries to follow their local statutory provi-
sions.
Note
In compliance with §14 of the BATTERY REGULATION, we
would like to point out that this instrument contains batteries.
Batteries that have been removed must only be disposed of
at the recycling facility set up for this purpose or via the retail
outlet.
It is illegal to dispose of them in household refuse.
124
What to do if...
6 What to do if...
Measuring instrument:
– Incorrect calibration – Select correct
procedure procedure
– Incorrect solution – Set up correct
temperature (without temperature
temperature probe)
– Socket damp – Dry socket
125
What to do if...
Buffer solutions:
– Incorrect buffer solutions – Change calibration
procedure
– Buffer solutions too old – Only use once.
Note the shelf life
– Buffer solutions depleted – Change solutions
Sample:
– pH value not stable – Measure with air
excluded if necessary
– Temperature not stable – Adjust temperature if
necessary
Electrode + sample:
– Conductivity too low – Use suitable electrode
– Temperature too high – Use suitable electrode
– Organic liquids – Use suitable electrode
126
What to do if...
Cause Remedy
LoBat display
– Batteries almost – Replace batteries
depleted (see section
5.1 MAINTENANCE)
Cause Remedy
to display
– Timeout of the interface – Checkout connected
instrument
127
What to do if...
Cause Remedy
StoFull message – All 800 memory locations – Output data store and
are full clear data store
128
Technical Data
7 Technical Data
Accuracy pH ± 0.005
(± 1 digit) (in the measuring range (at operating temperature
of 2 pH units around the + 15°C ... + 35 °C)
calibration point)
± 0.01
U [mV] ± 0.3 (at + 15 °C ... + 35 °C)
±1
T [°C] NTC 30: ± 0.1
PT 1000:
± 0.5 at 0 °C ... 15 °C
± 0.1 at 15 °C ... 35 °C
± 1 at 35 °C ... 55 °C
T [°F] NTC 30: ± 0.2
PT 1000:
± 0.9 at 32 °F ... 59 °F
± 0.2 at 59 °F ... 95 °F
± 1.8 at 95 °F ... 131 °F
129
Technical Data
130
Technical Data
131
Technical Data
132
Lists
8 Lists
This chapter provides additional information and orientation
aids.
Specialist terms The glossary briefly explains the meaning of the specialist
terms. However, terms that should already be familiar to the
target group are not described here.
Index The index helps you find the topics that you are looking for.
133
Lists
Abbreviations
AR AutoRead (drift control)
ARng Automatic range switching
Measuring instrument measures with high-
est resolution
ASY Asymmetry
AutoCal DIN Automatic calibration with DIN buffer
solutions
AutoCal TEC Automatic calibration with WTW technical
buffer solutions
Cal Calibration
Cd... Calibration with DIN buffer solutions
(acc. to DIN 19 266)
ConCal Conventional one/two point calibration
Ct... Calibration with WTW technical buffer
solutions
disp Display
Displays the data storage on the screen
E3 Error message (see WHAT TO DO IF ...)
InI Initialization
Resets individual basic functions to
the status they had on delivery
LoBat Low Battery
Batteries are almost empty
mV Voltage unit
mV/pH Unit of the electrode slope
OFL Overflow
Display range exceeded
pH pH value
RCL Recall
memory dump
134
Lists
S Slope
SELV Safety Extra Low Voltage
SEr Serial interface
Output of the data storage on the RS 232
or on the internal printer
SLO Slope
Slope setting on calibration
Sto Store
Memory
TP Temperature probe
Temperature measurement active
UASY Asymmetry potential
°C Temperature unit, °Celsius
°F Temperature unit, Fahrenheit
135
Lists
Glossary
AutoRead Monitors the electrode drift and releases the measured val-
ue only after the stability criterion has been reached. In this
way, this procedure ensures the highest degree of precision
and reproducibility.
136
Lists
Index
A interval
analog output 114 calibration 94
asymmetry 91 data transmission 112
authorized use 80 storing 106
AutoCal DIN 91, 98
AutoCal TEC 91, 95 K
AutoRead 88 keys 76
B L
battery compartment 123 LoBat 127
baud rate
setting 117 M
measuring precision 94
C
calibrating 91 O
calibration operational safety 80
evaluation 93
P
interval 94
place of the instrument 85
procedures 91
plug-in power supply
protocol 92
connecting 84
ConCal 91, 101
printing
condition on delivery 121
calibration protocol 92
D measured values 90
data record 104 printing measured values 90
data transmission (interval) 113
R
date
Redox electrode 89
setting 83, 119
Redox voltage 89
display 77
remote control 116
drift control 88
replacing the batteries 123
E reset 121
energy saving feature 85 RS232 interface 115
error messages 125
S
I safety 79
initial commissioning 83 safety precautions 79
initialize 121 scope of delivery 83
137
Lists
single-point calibration 91
ConCal 103
slope 91
sockets 77
T
temperature probe 87
three-point calibration 91
AutoCal DIN 100
AutoCal TEC 97
time
setting 83, 119
two-point calibration 91
AutoCal DIN 99
AutoCal TEC 96
ConCal 101
138
Appendix
9 Appendix
139
Appendix
Note
If you need further informationen or application notes you
can ask for:
l Application reports
l Primers
l Safety data sheets.
140