Grid
Grid
Radiographic Grids
Image-Forming X-Rays
Four X-ray paths
• a. X-rays interact with patient and
scatter away from the receptor
• b. X-rays interact and are absorbed
(photoelectric absorption) within
patient
• c. X-rays are transmitted through
patient without interaction and strike
receptor
• d. X-rays interact with patient
(Compton scatter) and scatter
towards
• C and D are referred to the image-
forming x-ray photons
Radiographic Grid
• Used to reduce scatter
radiation from reaching the
image receptor (IR) through
absorption
• Cleans up scatter radiation
• Inherent part of bucky,
placed between the patient
and IR • When primary x-rays interact
with the patient, x-rays are
• Table or upright bucky scattered from the patient in all
usage directions.
– >60 kVp, 10 cm tissue
1
Image Contrast
• Difference in optical
density between adjacent
structures
• High vs Low contrast High
– Skeletal anatomy
– Abdomen, Chest
Medium
Low
Grid Design
• Radiolucent interspace material
with alternating radiopaque strips
– Aluminum, plastic or carbon fiber
for interspace
– Lead, tungsten, platinum, gold
strips
• Transmits x-rays traveling in a
straight line, oblique x-rays
absorbed by strips
Surface area
of grid 6
2
Application
Grid Ratio
• Grid Ratio is the height of
the grid strip (h) divided
by the thickness of the
interspace material (D).
T = strip width.
– Grid Ratio = h/D
• Affected by changing
– Height of lead strips
– Thickness of strips
– Width of interspace
Grid Ratio
• High ratio grids absorb
more scatter yet
require higher mAs or
kVp
High-ratio grids are more effective than
– mAs is factor of choice low-ratio grids because the angle of
deviation is smaller.
• 5:1, 6:1, 8:1, 10:1,
12:1, 16:1 ratio
designs
3
Grid Ratio Application
10
Grid Frequency
• Number of strips or grid lines
per inch or cm
– 25 – 45 lines/cm, 60 – 110
lines/in
– 25 – 80 lines/cm, 60 – 200
lines/in
• Higher grid frequency
requires higher technique
– Less grid lines appear in image
– Often used in mammography
• 80 lines/cm, 200 lines/in
• Typically higher frequency grids
have thinner lead strips
11
12
4
Lead Content of Grid
• Lead content
– Most important factor in grid’s efficiency
– Measured in mass per unit area
• g/cm2
– High ratio grids tend to have highest lead content
– As lead content increases, removal of scatter
increases and therefore contrast increases
13
Application
14
Grid Performance
• Contrast Improvement Factor (k)
– Comparison of image contrast with a grid to image
contrast without a grid
– k is higher for higher ratio grids
• K = Radiographic contrast with grid
Radiographic contrast without grid
• Measured at 100 kVp using a step wedge
• Manufacturer Avg 1.5 – 2.5
– Use of a grid approximately doubles the contrast
15
5
Grid Performance
• Bucky Factor
– Higher technique required Approximate Bucky Factor Values
with grid usage
• mAs X bucky factor avg Grid 70 90 120 Avg
Ratio kVp kVp kVp
– Measurement of technical
Non- 1 1 1 1
factor and patient dose
Grid
increase based upon
penetration of primary & 5:1 2 2.5 3 2
scatter radiation through 8:1 3 3.5 4 4
the grid 12:1 3.5 4 5 5
– Bucky factor increases with 16:1 4 5 6 6
increased grid ratio and
increased kVp
16
Approximate Exposure
Factor Changes Necessary
for Standard Grids
Grid Ratio mAs Increase kVp Increase
Non-Grid 1X 0
5:1 2X 8 – 10
6:1 3X 11 – 12
8:1/10:1 4X 13 – 15
12:1 5X 20 – 25
16:1 6X 30 – 40
17
Grid Types
Linear Parallel
Linear Focused
• Linear/Parallel
– Vertical lead strips do not
coincide with the primary
beam
– Absorption of 10 beam (Grid
cut-off) occurs with:
• Short SID
• Large IR
18
6
Parallel Grid
Grid Types
• Linear/Focused
– Angled lead strips to
coincide with primary
beam divergence
– Focal distance set to SID
usage to minimize grid
cut-off
20
21
7
Grid Focus
Grid Types
• Crossed (Criss-Cross,
Cross-Hatched)
– 2 parallel grids
perpendicular
– Not common
– High Grid cut-off if off-
centered to CR
23
Grid Types
• Moving grid mechanism
– Upright or table bucky activation
– Eliminate grid lines from image
• High frequency = less grid lines
possible
• Low frequency = more grid lines
possible
– Reciprocating
• 2 cm movement transversely
• Motor drives grid back and forth
during exposure
– Oscillating
• Circular movement 2-3 cm
movement
• Electromagnet pulls grid to one side
• Releases it during exposure
24
8
Grid Types
• Grid Cap
– Permanently mounted grid, IR slides into device
– Used with a variety of IR sizes
• Grid Cassette
– Permanently mounted grid
– Specific grid sizes
• Wafer Grid
– Non-permanent grid mount, must be secured
– Specific grid size
25
Grid Types
• Air-Gap (Air Filtration)
– Common on dedicated
Chest X-ray units
– Part is @10-15 cm from IR
( 4” - 6”)
– Similar to 8:1ratio grid
• 10” air gap equivalent to 6”
15:1 ratio grid
– mAs increased 10% per cm
gap
– Magnification results
unless SID is increased
27
9
Grid Cutoff
• A decrease in the number of transmitted
photons that reach the image receptor
because of some misalignment of the grid
• Grid Errors
– Off-level grid
– Off-center grid
– Off-focus grid
– Upside-down focused grid
Grid Errors
• Off-level
– Parallel & focused
– Decreased density across
image
29
Grid Errors
• Off-Center (lateral decentering)
– Focused Grid
– Decreased density across
image
– Most common error
30
10
Grid Errors
• Off-focus
– Incorrect SID use
– Decreased density at edges of
image
• Direct relationship
– More critical with high ratio grid
• Upside-down
– Mobile radiography
– CR not directed to tube side
– Marked decreased density at
edges of image and points lateral
to CR
31
32
33
11
Grid Cutoff – Off Focus
34
35
Grid Cutoff
12
Grid Errors - Moire Effect
• Zebra pattern
• Caused by
– Similar grid frequency to laser scanning
frequency in CR processing
– Using a grid cassette in the bucky tray
• Correct by
– Selecting a high grid frequency
– Use a moving grid mechanism
– Do not use two grids
37
• Patient dose
Grid Selection
• Kvp usage
• Scatter absorption
– <90 kVp 8:1 satisfactory
– >90 kvp >8:1 grids used
39
13
Summary
40
14