Validation of Analysis Method For Determining Content of Chlorpheniramine Malate (CTM) in Tablet Preparation by Uv Spectrophotometry
Validation of Analysis Method For Determining Content of Chlorpheniramine Malate (CTM) in Tablet Preparation by Uv Spectrophotometry
INTRODUCTION
149 | Science and Community Pharmacy Journal, Volume. 2,Number. 2, December 2023
METHODS
1. Materials
The materials used in this study include CTM tablets, namely CTM standards, 3 types of CTM samples from
different production factories (Orphen, PIM, and Generic) and 0.1 N sulfuric acid solvent
2. Equipment
Equipment used in this study includes The materials that will be used in determining the levels include CTM
tablets, namely CTM standards, 3 types of CTM samples from different production factories (Orphen, PIM,
and Generic) and 0.1 N sulfuric acid solvent
3. Procedure
Samples Standard chlorpheniramine maleate (ctm) powder was weighed as much as 25 mg and
dissolved with 0.1 N sulfuric acid in a beaker. The solution was put into a 25 ml measuring flask until it
reached the mark and was homogeneous
A standard solution of chlorpheniramine maleate (ctm) with a concentration of 15 ppm was made by
pipetting 1.5 ml of the CTM stock solution and then placing it in a 10 ml volumetric flask. The solution was
diluted with 0.1 N sulfuric acid to the limit mark, then shaken until homogeneous. The maximum wavelength
of the solution was measured in the wavelength range between 200-400 nm.
From the 1000 ppm stock solution a standard solution was made with a concentration series of 6; 10 ;
1.5 ; 2 and 2.5 ppm as much as 10 ml. The series solutions that have been prepared are then measured for
the absorption of each concentration at the maximum wavelength obtained. The absorbance data obtained
is then calculated by calculating the standard curve equation to obtain the line equation y = a + bx.
Five test sample tablets that met weight uniformity were then crushed until smooth and homogeneous. The
powder sample was weighed as much as 5 mg and dissolved with 0.1N sulfuric acid and placed in a 25 ml
measuring flask up to the mark. 1 ml of the solution was pipetted, then diluted with sulfuric acid to a
concentration of 20 ppm in 10 ml. Measure the absorption of the solution at the maximum wavelength. If the
absorption of the test sample solution is still outside the absorption range of the standard solution, then the
solution is diluted until the absorption falls within the range. Determination of levels was carried out by
repeating 3 times.
150 | Science and Community Pharmacy Journal, Volume. 2,Number. 2, December 2023
RESULT AND DISCUSSION
Determination of the Maximum Wavelength
Determination of the maximum wavelength is carried out because the wavelength of a compound can be
different when determined under different conditions and tools. The maximum wavelength (λmax) is the
wavelength at which electronic excitation occurs which provides maximum absorbance. The aim of
measuring at the maximum wavelength is that the change in absorbance for each concentration unit is
greatest at the maximum wavelength, so that maximum analytical sensitivity will be obtained (Gandjar and
Rohman, 2007).
Figure 1 shows the results of the maximum CTM wavelength measurement obtained which is 263 nm. This
maximum wavelength indicates that the CTM absorption is in the UV region because it is in the wavelength
range of 200–400 nm. Theoretically the maximum absorption for CTM is 263 nm (Anne, 2018)
Determination Linearity
Linearity shows the ability of an analytical method to obtain test results that correspond to the analyte
concentration in the sample in a certain concentration range (Ermer and Miller, 2005). This can be done by
making a calibration curve from a standard solution whose concentration is known. The calibration curve is
a standard method that can be used to determine the concentration of an analyte based on the Lambert-
Beer law. Determination of the calibration curve is carried out by analyzing the concentration of CTM
standard solutions, including 6; 10 ; 15 ; 20 and 25 ppm. Solutions with a series of concentrations were
measured for their respective absorbance at the maximum CTM wavelength, namely 263 nm. Measuring
the absorbance of CTM standard solutions at the maximum wavelength is because in this area the largest
absorption point will be obtained for each CTM standard solution.
The measurement results show that the greater the concentration of the CTM standard solution measured,
the greater the absorbance obtained. This is because at higher concentrations, the concentration level of
the CTM compound is also higher. In addition, the Lambert-Beer law shows that changing the concentration
of a particular sample will change the absorbance at each wavelength by a constant factor (Skoog and
West, 1971). Preparation of a standard calibration curve was carried out by plotting the CTM standard
solution (axis x) and absorbance (y axis), then the points are connected with a straight line.
151 | Science and Community Pharmacy Journal, Volume. 2,Number. 2, December 2023
Figure 2. CTM standard curve
Based on the results of absorbance measurements of CTM solutions with various concentrations, it gives a
linear equation y = 0.01904x + 0.03139 with a correlation coefficient (R) value of 0.995 and a coefficient of
determination (R2) value obtained of 0.9947. The correlation coefficient value obtained is the relationship
between CTM concentration and absorbance, that is, it meets the linear criteria (parameters), because a
good correlation coefficient close to the value 1. The linear range values obtained indicate that the Lambert-
Beer law applies in the calibration curve, so that the line equation can be used to determine the validation
of the method for determining CTM levels using an ultraviolet spectrophotometer.
Determination of Precision
Accuracy or precision is a measure that indicates the degree of agreement between individual test results,
measured through the spread of individual results from the average if the procedure is applied repeatedly
to samples taken from a homogeneous mixture. Accuracy can be expressed as repeatability or
reproducibility. Repeatability is the accuracy of a method if it is carried out repeatedly by the same analyst
under the same conditions and within a short time interval. Reproducibility is the accuracy of the method if
carried out under different conditions. Precision values are calculated using standard deviation (SD) to
produce Relative Standard Deviation (RSD) or Coeficient Variation (CV). Accurate criteria are given if the
method provides a %RSD value ≤2%. The smaller the standard deviation value obtained, the smaller the
coefficient of variation value (Riyadi, 2009).
Based on the data in Table 1, it shows that the standard deviation value obtained by sample A was 0.4483
and the % relative standard deviation (%RSD) value was 14.6%, sample B was 0.2020 and the value %
relative standard deviation (%RSD) is 6.4%, sample C is 0.0542 and the % relative standard deviation
(%RSD) value is 1.3%. The results obtained indicate that the test method used in validating the method for
determining CTM levels in drug C samples using UV Visible spectrophotometry meets the acceptable
%RSD value requirements. But samples A and B do not meet the value requirements %RSD received. The
152 | Science and Community Pharmacy Journal, Volume. 2,Number. 2, December 2023
precision value can provide information that this method cannot be used as a permanent method used in
the laboratory.
Limit of Detection (LOD) is the smallest amount or concentration of analyte in a sample that can be detected,
but does not need to be measured according to the actual value. Limit of Quantitation (LOQ) or limit of
quantitation is the smallest amount of analyte in a sample that can be determined quantitatively with a good
level of precision and precision. Quantitation limits are quantitative test parameters for low analyte
concentrations in complex matrices and are used to determine the presence of impurities or product
degradation.
Based on Table 2, it can be seen that the LOQ values obtained for samples A, B and C are not accepted
as quantitation limits because they do not meet precision (CV < 20%) and accuracy (bias < ± 20%).
Meanwhile, if the LOD value at this concentration is measured, it cannot provide accurate analysis.
Determination of Accuracy
The accuracy of an analytical method is the closeness of the test result value obtained by the
procedure to the actual value. Accuracy is a measure of the precision of analytical procedures
(Rohman, 2007). Accuracy is expressed as the percent recovery (%Recovery) of the added
analyte.
Table 3 shows that the average % recovery results obtained for samples A, B, and C were 100%. According
to Harmita (2004), the acceptable range of percent (%) analyte recovery values is 90-110%. This range is
flexible depending on the condition of the analyte being examined based on the number of samples and
laboratory conditions. The %recovery obtained is within the acceptable range, namely 90-110%, so it can
be said that this method has good accuracy.
Determination of CTM levels is carried out by measuring the test sample solution which is suspected to
contain CTM at a maximum wavelength of 263 nm with repetition 3 times/sample. Determination of these
levels aims to guarantee the quality and safety of a medicinal product.
153 | Science and Community Pharmacy Journal, Volume. 2,Number. 2, December 2023
Table 4. Data on CTM Level Ranges
The data in Table 4 shows that the recovery value of mg/tablet in tablet drug samples with 3 different
trademarks is 6.3 mg, 5.8 mg, and 4.03 mg. According to Werner, et al (2010) the CTM level in tablet
preparations is 4 mg. The average level of paracetamol obtained was more than the level that should be
present in the tablet preparation of the drug. Meanwhile, the range of CTM levels obtained was 159.307%,
145.025% and 126.195%. According to the requirements of the Indonesian Pharmacopoeia (FI) Edition VI
2020, the level of CTM active substance in a tablet is not less than 98.5% and not more than 101.10%. The
results obtained from this test show a discrepancy between the test results and the established standards.
Factors that influence inconsistency in test results are when taking/pipetting the solution, inaccuracy in
weighing which causes less accurate results and the lack of a shaking factor before the sample solution to
be measured also affects the results obtained in the test, because the solution must be truly homogeneous
in order to obtain good results. maximum in testing.
CONCLUSION
Based on the results of the research that has been carried out, it can be concluded as the CTM levels
obtained in drug tablet samples did not comply with established standards, the UV spectrophotometry
method used in the research does not meet the parameters set in the validation test except for determining
accuracy, so this method can be applied for analysis of determining chlorpheniramine maleate levels in a
laboratory. More observations, furthermore, related to the development of methods for analysis drugs.
REFERENCES
Anne, Hafiezah & Tursino. 2018. Penetapan Kadar Pseudofedrin Hcl Dan Klorfeniramin Maleat Dengan
Metode Spektrofotometri Derivatif Dalam Sediaan Sirup. Jurnal Farmagazine. Tangerang.
Ermer, J., dan Miller, J.H.McB, 2005, Method Validation in Pharmaceutical Analysis, A Giude to Best
Practice, Weinheim: Wiley-VchVerlag GmbH dan Co. KGaA. Halaman 253.
Farmakope Indonesia edisi VI. 2020. Departemen Kesehatan. Republik Indonesia. Jakarta: Kementrian
Kesehatan Republik Indonesia.
Gandjar, G.H., dan Rohman, A., 2007. Kimia Farmasi Analisis. Yogyakarta: Pustaka Pelajar.
Harmita, 2004, Petunjuk Pelaksanaan Validasi Metode dan Cara Perhitungannya, didalam: Majalah Ilmu
Kefarmasian, Desember. Vol. 1, No.3, pp. 117 –135. Jakarta: Departemen Farmasi FMIPA-UI.
International Conference on Harmonization (ICH), 2005, Validation of Analytical Procedures: Text and
Methodology Q2(R1), in http://www.ich.org, access at 12 April 2022.
Ribka, K. M., Julius, P., Paulina, V. Y., Yamlean., 2018. Analisis Kandugan Merkuri (hg) pada Beberapa
krim Pemutih wajah Tanpa Ijin BPOM yang Beredar di Pasar 45 Manado. Jurnal Ilmiah Farmasi –
UNSRAT vol.7 No.3 agustus 2018 ISSM : 2302-2493.
Rohman, A. 2007, Kimia Farmasi, Yogyakarta: Pustaka Pelajar.
Skoog, D., dan West, D., 1971, Principles of Instrumental Analysis, New York: Holt, Rinehart and Winston,
Inc.
Werner, D., Thuman, C., Maxwell, J., 2010, Apa yang Anda Kerjakan Bila Tidak Ada Dokter (Where There
Is No Doctor). Yogyakarta: C.V Andi Offset.
154 | Science and Community Pharmacy Journal, Volume. 2,Number. 2, December 2023