BM-7AC Manual ENG Rev2
BM-7AC Manual ENG Rev2
LUMINANCE COLORIMETER
Rev.2
Introduction
Thank you for purchasing the TOPCON TECHNOHOUSE Luminance Colorimeter BM-7AC.
This instrument is a Luminance Colorimeter for measuring light with high precision from a light emitting
element such as a CRT, LCD, LCD backlight, LED or the like, or reflected light from a painted surface or
printed material.
Read this manual carefully before using the BM-7AC in work applications.
1
Display for safe use
In order to prevent unforeseeable injury to users and others, prevent damage to property or the like, and in
order that the instrument is used safely, important items are indicated on the instrument and in the
instruction manual.
After gaining a good understanding of the following displays and notation, carefully read the section 'Safety
Precautions', and obey all precautions.
Display Meaning of Display
* 1: This indicates injury such as a burn, electric shock or the like that does not require
hospitalization or long-term medical attention.
* 2: Physical damage includes damage to the building, property, pets or the like.
2
Safety Precautions
Warning
Do not use the instrument in areas near flammable or combustible vapors (gasoline, etc.).
Doing so could be the cause of fire.
Prohibited
In the case of any strange noise, smell or smoke from the instrument, immediately turn
OFF the power to the device and unplug the AC adapter from the power outlet.
Continuing to use the device as is could cause fire.
Mandatory
Contact the dealer or TOPCON TECHNOHOUSE CORPORATION.
3
Caution
Do not look directly into the sun or at the light bulb filament.
Doing so could result in damage to the eyes.
Prohibited
Use only specified screws when using the tripod screw and screw holes for jig
attachment.
Mandatory Do not tighten the screws any more than necessary. Doing so might cause internal
breakage.
Escape Clause
・ This company bears no liability for damage to the instrument resulting fire, earthquake, behavior
of a third party, other accidents occurring due to negligence of the user, unauthorized use, or
other abnormal conditions
・ This company does not bear any liability due to collateral damage (loss in business profit,
interruption of work, etc.) from the use of or inoperability of the instrument.
・ This company bears no liability for damage that occurs due to use other than that explained in
this instruction manual.
・ This company bears no liability for damage that occurs due to malfunctioning or the like
resulting from combined use with connected equipment.
User Maintenance
In order to maintain safety and performance, any maintenance work other than that instructed in
this manual should only be performed by a qualified service representative. However, the
following maintenance can be performed by the user.
4
Table of Contents
Introduction................................................................................................................................................................................1
Display for safe use ................................................................................................................................................................2
Safety Precautions .................................................................................................................................................................3
Notation Rules Used in This Manual ................................................................................................................................8
5
3.1.2 Data/Setting Item Display ........................................................................................................................................................................43
3.1.3 Numerical Value Setting ............................................................................................................................................................................44
6
4.2.17 FGR Command ...............................................................................................................................................................................................67
4.2.18 RG K Command ...........................................................................................................................................................................................68
4.2.19 RG L Command .............................................................................................................................................................................................68
4.2.20 WG K Command ............................................................................................................................................................................................68
4.2.21 WG L Command.............................................................................................................................................................................................69
4.2.22 WHO/VER/SRL Command.......................................................................................................................................................................69
4.2.23 UT Command ..................................................................................................................................................................................................69
4.2.24 CT Command ..................................................................................................................................................................................................70
4.2.25 AM/SM Command ........................................................................................................................................................................................70
Appendix ....................................................................................................................................................................77
Specifications......................................................................................................................................................................... 77
Appended Tables .................................................................................................................................................................. 82
System Diagram .................................................................................................................................................................... 83
External Dimensions ............................................................................................................................................................ 84
Internal Calculation Processing ...................................................................................................................................... 85
Terminology ............................................................................................................................................................................ 87
Warranty.....................................................................................................................................................................88
7
Notation Rules Used in This Manual
The following notation rules are used in this manual.
Notation Explanation
[MODE], [UP] Indicates buttons, or title of the screen displayed on the LCD display.
☞「」 Indicates a reference location in the manual.
☞『』 Indicates another manual as a reference.
Indicates items to keep in mind when performing an operation, or items
that require attention.
Request
Indicates items to consider when performing an operation, or information
Memo that is useful.
8
1. Before Use
1. Before Use
1.1 Checking the Unit and Supplied Parts
Make sure that the main unit and all of the following supplied parts are in proper order.
If there are any missing parts, contact the dealer or TOPCON TECHNOHOUSE CORPORATION.
・ Main unit 1
・ Instruction manual(Quick manual) 1
・ AC adapter 1
・ Inspection report 1
・ CD-ROM 1
(Instruction manual / Colorimetry program CS-900A)
・ Carrying case (option) 1
・ Analog output plug 3 * For analogue output model only
9
1. Before Use
Objective lens
Eyepiece
LCD
Panel switches
USB connector
RS-232C connector
Analog output connector (X, Y, Z) DC input connector
* Analog output model only Power switch
Name Function
Inner switches/buttons Buttons for performing various settings in the
function mode, and dip switches for performing
various settings for the remote mode, change in
object color etc.
☞「■ Inner Switches/Buttons」
Diopter adjustment ring Used when adjusting the focus of the finder on the
reticle mark.
LCD LCD with backlight for displaying various kinds of
information such as measurement results and
measurement conditions.
Panel switches Switches for starting/stopping measurement, and
switches for changing the response speed of the
photo-receiver circuits and display mode.
☞「■ Panel Switches」
Power switch Power switch to the instrument.
DC input connector Connector for plugging in the output plug from AC
adapter.
USB connector Connector for connecting a PC when performing
measurement in the remote mode.
PC connection ☞「1.3.2 Connecting a PC」
RS-232C connector Connector for connecting a PC when performing
measurement in the remote mode.
PC connection ☞「1.3.2 Connecting a PC」
10
1. Before Use
Name Function
Analog output connector Connected when using analog output.
* Analog output is a function only There are three connectors corresponding to
used for analog output model. tristimulus value X, Y, Z.
Memo
・ Use the manual range when using this output.
・ Do not insert or remove connectors when the power of the instrument is ON.
Request ・ Only uses the authorized AC adapter.
11
1. Before Use
Focus ring
Name Function
Focus ring Used when focusing on the measured object during measurement.
Focus ring set screw hole When the instrument is used in a stationary position and it is not
necessary to adjust the focus, a screw is screwed into this hole and the
ring is fix.
Screws that can be used are a metric M3 screw, round-tipped, 5mm long
or more slotted set screw, or hexagonal socket head set screw.
Tool attachment screws Used when assembling a system, and are screws for attaching the
BM-7AC.
M4×0.7 (Dia: 4 ㎜, Pitch: 0.7 ㎜) screws
Details ☞「Appendix: External Dimensions」
Tripod screw Screw used when attaching the instrument to a tripod.
Nominal designation: 1/4-20UNC No, of threads: 20
Pitch: 1.270 ㎜ Depth: 6 ㎜
Measuring field selector switch Used when selecting the measurement angle.
The size of the measurement range and measurement areas differ
depending on measuring angle.
Details ☞「Appendix: Specifications」
・ Only use a specified screw in the Focus ring set screw hole. Do not tighten the screw
Request any more than necessary. It may cause internal breakage.
・ Use only specified screws when using the tripod screw and screw holes for jig
attachment. Do not tighten the screws any more than necessary. Doing so might cause
Request internal breakage.
12
1. Before Use
■Panel Switches
LCD
FAST/SLOW switch
LIGHT switch
AUTO/MANU Switch
This switche is used to select AUTO or MANUAL measurement ranges.
The range is set to AUTO when the power is ON.
AUTO range: Automatically sets the measurement range according to the brightness of the
object being measured.
MANUAL range: Performs measurement in the measurement range set in the function mode.
Used when using analog output.
Factory default setting is range 3.
FAST/SLOW Switch
This switch is uded to select the response speed of the photo-receiver circuit. The response
speed is the time required for the analog output of the instrument to reach 90% of the peak value
when measuring an LED driven by a square wave from a function generator.
SLOW is set when the power is ON, and SLOW is used for normal measurement.
FAST : The response speed of the photo-receiver circuit is increased.
This is set when measuring fixed light such as the sun or halogen lamp, and a light
source having high frequency lamp such as a LCD.
Set when measuring a flashing light source and analog output is connected to an
oscilloscope. (Analog output is only used for analog output model.)
The response speed differs depending on the measurement range (30 milliseconds
to 1 millisecond).
SLOW : The response speed of the photo-receiver circuit is decreased.
This is used for normal measurement such as measuring fixed light, and flashing
light such as high-frequency flashing light, fluorescent light, and a CRT.
Set when recording waveforms of analog output using a pen recorder.
13
1. Before Use
RUN/HOLD Switch
This switch is used to select free-run status or hold status for display.
This switch is set to RUN when the power is ON.
It is also used as a switch for starting measurement of a standard sample, or white board.
CAL Switch
This switch is used to perform zero-adjustment.
Internal shutter of BM-7AC make a shaded state and perform measurement in the shaded state,
and saves the value.
The saved value in the shaded state is subtracted from the measurement value and the result is
displayed.
Zero-adjustment should be performed with the cap to the Eyepiece attached.
MODE Switch
By pressing this switch in the HOLD state (when [HOLD] of the RUN/HOLD switch is blinking),
the display mode of the measurement data can be changed.
Each time the switch is pressed, the display mode changes as shown below.
xy/L → u'v'/L → XYZ → Tc/duv/L → xy/L ……
This switch is set to xy/L mode when the power is turned ON.
LIGHT Switch
This switch is used to switches ON/OFF the lighting (backlight) to the LCD.
The switch is set ON when the power is turned ON.
■Inner Switches/Buttons
Dip switches Buttons
ON
1 2 3 4 5 6 7 8
FUNCTION ENTER CHANGE ROTATION SHIFT
ON
LOCAL USB SOURCE ABS FACTOR A PROTECT
REMOTE RS-232C OBJECT DIF ON ONLY ON ONLY WRITE
FACTOR B
1 2 3 4 5 6 7 8
― Buttons ―
FUNCTION
This button is used to enter/return from the function mode.
Press this button once to enter function mode, and press it again to return form function mode.
ENTER
This button is used when changing the display page and when storing the data in the memory
after entering value.
CHANGE
This button is used to change setting.
14
1. Before Use
ROTATION
This button is used to select a parameter to be inputted.
Each time the button is pressed, the selection candidate is displayed in order.
The displayed contents differ according to the parameter type.
Example : ・ → 0 → 1→ 2 → ……… → 8 → 9 → + → - → E → ・ → ………
2400 → 4800 → 9600 → 19200 → 38400 → 2400 → ………
EVEN → ODD → NONE → EVEN → ………
SHIFT
This button is used to move among digits when entering a multi-digit numerical value.
It moves from the highest order digit to the lowest order digit.
― Dip Switches ―
1 LOCAL/REMOTE
This switch is used to select the control state of the instrument.
LOCAL : Sets the LOCAL mode.
REMOTE : Sets the REMOTE mode.
・ The connection method is read into the instrument when the power is turned
ON. To change the setting、turn the power OFF and ON again.
Request
2 USB/RS-232C
This switch is used to select the connection method between the instrument and PC.
USB : Connection with PC through USB.
RS-232C : Connection with PC through RS-232C.
・ The connection method is read into the instrument when the power is turned
ON. To change the setting、turn the power OFF and ON again.
Request
3 SOURCE/OBJECT
SOURCE : Set when measuring the light source.
OBJECT : Set when measuring the object.
・ The connection method is read into the instrument when the power is turned
ON. To change the setting、turn the power OFF and ON again.
Request
4 ABS/DIF
This switch is used to select the display between an absolute value and difference.
ABS : Sets the absolute value display.
DIF : Sets the difference display.
Memo
・ This switch can be changed during power ON.
15
1. Before Use
5 FACTOR A/B
FACTOR A : Select when setting a single correction factor.
The correction is set in the function mode.
FACTOR B : Select when setting the direct correction function, using other instrument
BM-7AC, SC-777 or SR-3. The correction is set in the function mode.
Function Mode ☞「3.1 Function Mode」
Direct Correction Function
☞「2.1.4 Reducing Error Among a Plurality of
Measurement Instruments and Performing
Measurement」
Memo
・ This switch can be changed during power ON.
6 NC
This is an empty switch and should always be set to the ON side.
7 NC
This is an empty switch and should always be set to the ON side.
8 PROTECT/WRITE
PROTECT : Prevents writing to internal memory.
This switch should always be set to the PROTECT side during measurement.
WRITE : Allows writing to internal memory.
This switch should be set to the WRITE side when changing memory contents.
Memo
・ This switch can be changed during power ON.
16
1. Before Use
■ LCD
Measurement results are displayed on the LCD as follows.
Correction factor
00 : No correction factor
01 – 10 : Correction factor number
17
1. Before Use
1.3 Preparation
1.3.1 Connecting the AC Adapter
The procedure for connecting the AC adapter to the instrument is explained below.
2 Insert the connector on the output side of the AC adapter to the DC input connector of
the instrument.
18
1. Before Use
3 Plug in the AC adapter plug into the power outlet.
19
1. Before Use
1.3.2 Connecting a PC
When connecting a PC, use a RS-232C cable or USB cable to connect the BM-7AC and PC.
When using a RS-232C cable, use a straight cable supporting for a DOS/V PC. It is possible to
connect up to a maximum of 16 instruments to one PC by USB, also parallel use with an RS-
232C cable is possible. However, the USB and RS-232C cannot be used at the same time for
one BM-7AC.
Memo
・ The connection cables are not supplied with the unit. If cables are to be used, they should
be purchased separately.
・ For details regarding the connection on the PC side, refer to the manual for the PC being
used.
USB connection
- BM-7AC side - - PC side -
RS-232C connection
- BM-7AC side - - PC side -
20
1. Before Use
RS-232C specification for this unit is as shown below.
・ Cable Straight cable supporting for a DOS/V PC
・ Communication method Full duplex
・ Synchronous system Start stop synchronization
・ Baud rate 2400/4800/9600/19200/38400 BPS (Bits Per Second)
・ Bit configuration Data length 7 bits/8 bits
Parity EVEN/ODD/NONE
Stop bits 1 bit/2 bits
・ Communication format ASCII format
・ Delimiter "CR+LF" or "CR" added to the end of communication data strings.
When the power is turned ON in this state, the instrument will warm up for 10 minutes, then the
instrument enter the remote mode and the following screen will be displayed.
Turning on the Power ☞「1.3.4 Turning ON/OFF the Power」
Remote Mode
Warm-up..
. Ver.1.00
21
1. Before Use
Do not look directly at a bright light such as the sun or a bulb filament.
Doing so could damage the eyes.
Prohibited
Use only specified screws when using the tripod screw and screw holes for jig
Request attachment. Do not tighten the screws any more than necessary. Doing so might
cause internal breakage.
Reticle Mark
4 Collimate with the measured object, and turn the focus ring of the objective lens and adjust
the focus on the measured object.
5 Change the measurement angle according to the size and brightness of the measured
object.
Memo
・ When changing the measurement angle, use the Measuring field selector switch.
・ The current measurement angle will be displayed on the LCD.
Measuring field selector switch, LCD Display
☞「1.2 Part Names and Function」
・ The measurable luminance range changes according to the measurement angle.
☞「Appendix: Specifications Luminance Measurement Range」
22
1. Before Use
After the power has been turned ON, the instrument will warm up for 10 minutes, then the zero-
adjustment starts and then measurement starts automatically.
Calibrating
To finish measurement, press the RUN/HOLD switch. The final measurement results of
continuous measurement will be displayed.
Memo
・ If it has been 1 year or more since calibration has been performed by TOPCON
TECHNOHOUSE factory, a screen recommending maintenance will be displayed before zero-
adjustment.
Maintenance Recommended Screen Display ☞「1.4 Maintenance Recommended Display」
・ In the REMOTE mode (when dip switch 1 is on the REMOTE side), the screen that is
displayed after the power is turned ON will differ.
☞「1.3.2 Connecting a PC」「Dip Switch Settings」
When the power is turned ON and dip switch 8 is on the WRITE side, the following screen is
displayed.
Attention !!
Please set
the PROTECT/WRITE SW
to the PROTECT side.
After this screen is displayed, set dip switch 8 to the PROTECT side.
23
1. Before Use
Canceling the warm up
To cancel the warm up, press the CAL switch during warm up.
The following screen is displayed, and when the CAL switch is pressed again, zero-adjustment
begins.
DARK MEASURE
Push CAL Key.
・ When turning the power ON again after a short time has elapsed after the power
Request is turned OFF, do not cancel the warm up. Doing so could result in not being
able to obtain accurate measurement results.
To turn the power OFF, flip the power switch to the left side.
Measurement Method
FAST Make sure the [FAST] LED is blinking, then press the RUN/HOLD switch and
start measurement.
SLOW Make sure the [SLOW] LED is blinking, then press the RUN/HOLD switch
and start measurement.
Memo
・ When observing a blinking light source using analog output, set the speed to the FAST
mode. (Analogue output is a function only used for analog output model)
・ To obtain stable measurement result, set the speed to SLOW.
24
1. Before Use
Attention!!
A maintenance is
required.
Hit Inner SW
This display is displayed and a buzzer sounds until any inner switch is pressed, then the display
advances to the next screen.
To automatically pass this process, set the setting for the maintenance recommendation screen
to OFF.
Maintenance Recommended Display Setting ☞「3.13 Maintenance Recommended Display」
25
1. Before Use
26
2. Measurement Operation
Displayed when the calibration cannot be completed. Check that the cap is
attached to the eyepiece and the measurement condition is in operating
Calibration ERROR conditions.
If an error is still displayed even after retrying, repair may be required.
Contact the dealer or TOPCON TECHNOHOUSE office.
RS232C ERROR Displayed when the communication with PC is failed. Check the connection of
RS-232C or USB cable and communication condition.
27
2. Measurement Operation
2. Measurement Operation
2.1 Example of use
An example of measurement will be given here. Use it as a reference when using this instrument.
28
2. Measurement Operation
■ Using CS-900A
The reference light source sample should be the same kind of light source as the light source
actually being measured, and should have stable light output.
The procedure for performing correction between instruments using CS-900A, refer to the [CS-
900A INSTRUCTION MANUAL].
・ The reference light source sample should have stable light output and be the
same type as the light source to be measured.
Request
・ To use the Direct Correction function, the RS-232C parameters of the two
luminance colorimeters must be the same.
When using the Direct Correction function, first, measure the light source using the reference
instrument, and write that measurement data to the BM-7AC to be corrected.
Source
RS-232C
Data
BM-7AC SR-3
29
2. Measurement Operation
Next, place the BM-7AC to be corrected in the location where the reference instrument placed,
and measure the same light.
Source
Source
BM-7AC
A correction factor is calculated for the BM-7AC to be corrected based on the data of the
reference instrument, and the data of the BM-7AC to be corrected.
The procedure of measurement for correction between instruments using the Direct Correction
function is as explained below.
1 Set beforehand the data output method and communication method for the reference
instrument in reference to the table below. Then connect it with the BM-7AC to be
corrected using a specified cable. Also, set the RS-232C parameter settings of the BM-
7AC to be corrected are the same as those of the reference instrument.
2 Set dip switch 5 of the BM-7AC to be corrected to the FACTOR B side, and turn ON the
power.
3 Turn on the power to the reference luminance colorimeter.
4 The BM-7AC to be corrected changes to the function mode, and the [*Direct-Correction
Factor*] screen is displayed.
Moving to the Function Mode ☞「3.1.1 Moving/Returning to the Function Mode」
Display Method ☞「3.1.2 Data/Setting Items Display」
*Direct-Conn Factor*
Measure Reference
Measure Factor
All Factor Clear
30
2. Measurement Operation
6 Press the [ROTATION] button, and select "Measure Reference" ("*" is displayed) and
press the [ENTER] button.
Measure Reference : Perform measurement using the reference instrument, and write
the measurement data in the BM-7AC to be corrected.
Measure Factor : Perform measurement using the BM-7AC to be corrected, then
a correction factor is calculated based on data found in the
"Measure Reference".
All Factor Clear : Clear the correction factor and measurement data found above.
*Direct-Conn Factor*
*Measure Reference
Measure Factor
All Factor Clear
REF11
NO DATA
R=NEXT S=PAGE
Memo
・ It is possible to store up to 5 items of correction factor data.
31
2. Measurement Operation
9 Press the RUN/HOLD switch.
The message "Measuring" is displayed on the LCD, and measurement by the reference
luminance colorimeter starts. After measurement end, the following screen is displayed.
Attention !!
Please set
the PROTECT/WRITE SW
to the PROTECT side.
・ In the case that each of the various communication settings is not correct,
Request data communication cannot be performed. Check whether setting has
been performed properly according to the Error display.
Meaning of Error Displays ☞「1.5 Error Display」
10 Switch dip switch 8 of the BM-7AC to be corrected to the WRITE side to write data.
11 Place the BM-7AC to be corrected at the location where the reference luminance
colorimeter is placed, then press the [ENTER] button to display the following screen.
Select "Measure Factor" and press the [ENTER] button.
*Direct-Conn Factor*
*Measure Factor
Exit
12 Specify a number to write the data for the BM-7AC to be corrected. Select the number
where the data for the reference luminance colorimeter is written, then press the
RUN/HOLD switch to start measurement.
K11
NO DATA
R=NEXT S=PAGE
13 The message "Measuring" is displayed on the LCD, and the BM-7AC to be corrected
begins measurement. After finish of measurement, the following screen will be displayed
for about 3 seconds.
SMP11 x= 0.4475
y= 0.4320
L= 2.071E+01
32
2. Measurement Operation
14 The correction factor is calculated and the following screen is displayed.
15 When "SET" is displayed on the following screen by pressing the [CHANGE] button,
indicates that the correction factor has been set.
16 After returning to the function mode, the correction factor becomes valid.
Entering to the Function Mode
☞「3.1.1 Entering /Returning from the Function Mode」
33
2. Measurement Operation
34
2. Measurement Operation
Attention !!
Please set
the Std sample,then
push RUN/HOLD key.
・ Reference data is stored internally until the power is turned OFF. After
Request returning to ABS after once setting DIF, if DIF is set again, the reference sample
is not measured. Difference measurement will be performed using the stored
reference data. To change the reference data, turn OFF the power to the
instrument and turn it ON again, then perform the procedure again from step 1.
Memo
・ To return to absolute value measurement, set dip switch 4 to the ABS side.
・ The stored reference data can be check in the function mode.
Checking Reference Data ☞「3.2 Standard sample」
35
2. Measurement Operation
1 Make sure the power to the instrument has been turned OFF. If the power is ON, turn it
OFF.
2 Set dip switch 3 to the SOURCE side.
・ If the dip switch is changed while the power is ON, the change will not take
Request effect.
3 Turn the power to the instrument ON. The instrument will be in the free run state and
measurement will start.
36
2. Measurement Operation
1 Make sure the power to the device has been turned OFF. If the power is ON, turn it OFF.
2 Set dip switch 3 to the OBJECT side.
・ If the dip switch is changed while the power is ON, the change will not take
Request effect.
Attention !!
Please set
the WS (100.0%),then
push RUN/HOLD key.
4 Make sure that the displayed luminance factor for the white board matches the luminance
factor for the white board used. If it does not match, set the proper value.
Setting the Luminance Factor of the White Board
☞「3.4 White Board Luminance Factor」
Memo
・ The factory setting for the white board luminance factor is 100.0%.
After measurement, the measurement result will be displayed for about 3 seconds, then
measurement of the object color will start automatically in continuous measurement.
Memo
・ The white board measurement value is store internally until the power is turned OFF.
However, in the following cases, the white board data will be lost.
・ When the white board luminance factor is changed
・ When the white board correction factor is changed
Checking the White Board Measurement Value ☞「3.3 White Board」
・ To measure the white board again, turn the power OFF and ON again, then perform the
operation from step 1.
37
2. Measurement Operation
38
2. Measurement Operation
― Object Color ―
Chromaticity xy Luminance Lv
CIELAB
CIELUV
39
2. Measurement Operation
Memo
・ The analog output connector is a function that can only be used for analog output
model.
Memo
・ In regards to the analog output for X and Z, theoretically it do not have relationship between
the analog output voltage and luminance (even though the luminance is the same, the analog
output voltage differs when the spectral distribution of the light source is not same). Details
of relationship between analog output for X / Z and luminance not be given here.
40
2. Measurement Operation
2.7.3 Usage
The procedure for analog output measurement is as described below.
1 Solder and attach the shielded wire of the supplied analog output plug, and connect it to
the device.
・ When an absolute value is necessary for the tristimulus value, take the
Request correlation with the display into consideration.
・ When used with the output voltage in the saturated state, proper
measurement cannot be performed. Be sure to set the proper
measurement range.
・ The output impedance is about 100Ω. Use a recorder having an input
impedance of 10kΩ or more.
・ The zero point may move due to temperature change, but this does not means the
instrument is not broken.
41
3. Various Settings
3. Various Settings
3.1 Function Mode
The function mode is a mode for checking or changing various data or settings that are store in
the internal memory of the instrument.
The procedure for returning from the function mode is as described below.
When dip switch 8 is on the WRITE side when ending the function mode, the following screen is
displayed.
Attention !!
Please set
the PROTECT/WRITE SW
to the PROTECT side.
Select dip switch 8 to the PROTECT side, the setting changes become valid, and the
measurement mode screen is displayed.
42
3. Various Settings
Memo
・ The Standard sample and white board are not displayed when measurement of them are not
performed.
When measuring light source color - Dip switch 3 is on the SOURCE side -
・ Standard sample ☞「3.2 Standard sample」
・ Correction Factor ☞「3.6 Correction Factor」
・ Area Correction Factor ☞「3.7 Area Correction Factor」
・ Measurement Range ☞「3.8 Measurement Range」
・ RS-232C Parameter ☞「3.9 RS-232C Parameter」
・ Buzzer Sound ☞「3.10 Buzzer Sound」
・ Communication Format ☞「3.11 Communication Format」
・ Average Measure ☞「3.12 Average Measure」
・ Maintenance Recommended Display ☞「3.13 Maintenance Recommended Display」
When measuring object color - Dip switch 3 is on the OBJECT side -
・ Standard sample ☞「3.2 Standard sample」
・ White board ☞「3.3 White Board」
・ White Board Luminance Factor ☞「3.4 White Board Luminance Factor」
・ White Board Correction Factor ☞「3.5 White Board Correction Factor」
・ Correction Factor ☞「3.6 Correction Factor」
・ Measurement Range ☞「3.8 Measurement Range」
・ RS-232C Parameter ☞「3.9 RS-232C Parameter」
・ Buzzer Sound ☞「3.10 Buzzer Sound」
・ Communication Format ☞「3.11 Communication Format」
・ Average Measure ☞「3.12 Average Measure」
・ Maintenance Recommended Display ☞「3.13 Maintenance Recommended Display」
43
3. Various Settings
* White board *
Luminance factor
100.0 %
Attention !!
Please set
the PROTECT/WRITE SW
to the WRITE side.
Memo
・ When dip switch 8 is on the WRITE side, this screen is not displayed. In that case,
advance to step 3.
* White board *
Luminance factor
.
* White board *
Luminance factor
1
Memo
・ When entering a numerical value having more than one digit, press the [SHIFT] button
to move to the next digit, and repeat step 3 to select the value to be input.
44
3. Various Settings
4 After setting has been completed, press the [ENTER] button.
Memo
・ If a mistake is made during input of the numerical value, complete the procedure to step 4
then start over again from step 1.
* Standard sample *
x= 0.2279
y= 0.2615
L= 8.100E+02 cd/m^2
* White board *
X= 1.020E+03
Y= 1.000E+03 cd/m^2
Z= 1.143E+03
Memo
・ Not displayed when there are no data.
・ Displayed by XYZ regardless of the display mode.
・ Cannot be changed from just the display.
45
3. Various Settings
* White board *
Luminance factor
100.0 %
W KX=2.481E+02
KY=3.745E+02
KZ=1.364E+02
S=PAGE
The input/change method is the same as the input/change method for the correction factor.
☞「3.6.2 Inputting the Correction Factor」
Memo
・ It is possible to set just one white board correction factor.
46
3. Various Settings
* FACTOR NUMBER *
SET FACTOR 01
The correction factor is a factor for correcting the measurement value by using it to multiply the
measurement value.
The procedure for changing the factor is as described below.
FACTOR
DISP&SET OFF
INPUT *BACK
CLEAR
2 Press the [ROTATION] button, and select the process to be performed (indicated by "*").
DISP&SET : Selected when checking the value of the registered correction factors
(K01 to K10), or when setting the number of the correction factor to be
used.
☞「3.6.1 Checking/Setting the Correction Factor」
INPUT : Selected when entering or changing the correction factor.
☞「3.6.2 Inputting the Correction Factor」
CLEAR : Selected when clearing the correction factor.
☞「3.6.3 Clearing the Correction Factor」
OFF : Selected when not using a correction factor.
BACK : Selected when returning to the previous screen.
3 Press the [ENTER] button.
The screen corresponding to the selected process is displayed.
Memo
・ The correction factor is not valid when dip switch 5 is set to the FACTOR B side.
・ The factory settings for KX, KY, and KZ are all 1.
47
3. Various Settings
[CHANGE] : The displayed correction factor takes effects. When it is set, "SET" will be
displayed underneath the number of the correction factor.
When the value of the correction factor is not set, the input screen for
inputting the correction factor will be displayed.
Correction Factor Input Screen ☞「3.6.2 Inputting the Correction Factor」
[ROTATION] : Display the correction factor of next number.
[SHIFT] : Each time the button is pressed, the display changes as shown below.
Reference data → Sample data → Correction factor data → …
- Reference Data -
The reference data from when the displayed correction factor was calculated
is displayed.
REF01 x= 0.4476
y= 0.4074
L= 1.000E+02
C=SET R=NEXT S=PAGE
- Sample Data -
The sample data from when the displayed correction factor was calculated is
displayed.
SMP01 x= 0.4476
y= 0.4074
L= 1.000E+02
C=SET R=NEXT S=PAGE
Memo
・ Numbers such as "REF01" and "SMP01" that are displayed at the top left of the reference
data screen and sample data screen correspond to the number of the correction factor. In
other words, when the correction factor for K01 is calculated, the reference data for K01 is
REF01 and the sample data for K01 is SMP01.
48
3. Various Settings
1 Press the [ROTATION] button to display the data for the number of the correction factor
to be input
[ROTATION] : The correction factor for the next number is displayed.
[SHIFT] : Each time the button is pressed, the display changes as shown below.
Reference data → Sample data → Correction factor data → …
[ENTER] : Returns to the previous screen.
2 Press the [CHANGE] button.
The following screen is displayed.
FACTOR INPUT
DIRECT
REF. & MEASURE
*BACK
3 Press the [ROTATION] button, and select the process to be performed (indicated by "*").
DIRECT : Selected when inputting the correction factor directly.
☞「Direct Input」
REF. & MEASURE
: Selected when inputting already known luminance and chromaticity data,
performing sample measurement and calculating the correction factor.
☞「Calculation from Measurement」
BACK : Selected when returning to the previous screen.
4 Press the [ENTER] button.
49
3. Various Settings
Direct Input
The correction factor input screen is displayed.
The procedure for inputting the correction factor directly is as described below.
The procedure for calculating the correction factor from measurement is as described below.
Attention !!
Please set
the Ref.sample,then
push RUN/HOLD key.
50
3. Various Settings
3 Set the sample data and press the RUN/HOLD switch.
After measurement, the measurement result is displayed for about 10 seconds, then, the
correction factor is calculated using the input reference data and measured sample data is
displayed.
51
3. Various Settings
FACTOR CLEAR
ALL CLEAR
SELECT CLEAR
*BACK
Press the [ROTATION] button, and select the process to perform (indicated by "*").
ALL CLEAR : Selected when clearing all correction factors.
SELECT CLEAR : Selected when clearing individual correction factors.
・ When the reference data is cleared, correction factor data and sample data of
Request the same number are also cleared.
52
3. Various Settings
* FACTOR GROUP *
OFF
The area correction factor corrects measurement values in an arbitrarily specified area of a xy
chromaticity diagram or u'v' chromaticity diagram.
Memo
・ To register an area correction factor, it is necessary to use software CS-900A, or create
communication software that references a command list.
Command List ☞「4. Communication with a PC」
Area correction factors can be registered for a total of 10 groups to the BM-7AC. Groups 1 to
5 are assigned to area correction factors for the xy chromaticity diagram, and groups 6 to 10 are
assigned to area corrections factors for the u'v' chromaticity diagram. Also, it is possible to set
5 chromaticity areas in 1 group. In the chromaticity area setting, the correction factors are
specified by KX, KY and KZ, and the minimum and maximum areas of application of correction are
specified by x, y, or u' ,v', and minimum Luminance are specified by L.
G1 Color area 1
Specified color area
G2 Color area 2
G5 Color area 5
G6
G7
u',v'
G10
53
3. Various Settings
The area correction function enable measurement data to activate the correction factor
automatically at preset area in chromaticity diagram when measured chromaticity value enter the
preset area. The area can be specified up to 5 areas at once. For example, when you intend to
measure the luminance and chromaticity of R (red), G (green), B (blue), and W (white) color with
each correction factor applied, once you preset correction factor and area data for each of colors,
it is not necessary to change the setting of correction factor at each color measurement.
W
B R
It is possible to specify up to a maximum of 5 areas. Also, the minimum luminance value is set.
・ The chromaticity area should be specified within the limits of 0.03 between
Request minimum value and maximum value.
54
3. Various Settings
GROUP SET
GROUP *OFF
GROUP SET
GROUP *OFF
2 Press the [ROTATION] button, and select the area correction factor group for which the
chromaticity area is to be displayed.
3 Press the [CHANGE] button.
The data of the set chromaticity area is displayed.
G1-1 x0.2000-0.2300
SET y0.3000-0.3300
L1.000E+02
C=SET R=NEXT S=PAGE
55
3. Various Settings
* Measuring range *
X :Range1
Y :Range3
Z :Range3
・ The measurement range is from 1 to 5, with range 1 having the maximum sensitivity.
Measurement Range Details ☞「Appendix: Specifications」
・ When the AUTO/MANU switch is set to AUTO, and dip switch 1 is set to the remote side,
the measurement range setting is invalid.
・ The factory settings for X, Y and Z are all range 3.
・ After selecting Z, press the [ENTER] button. The next setting item (screen) is displayed.
56
3. Various Settings
*RS-232C Parameters*
Baud rate=38400
Length=7 Parity=ODD
Stop bit= 1
57
3. Various Settings
* SET BUZZER *
BUZZER OFF
* DATA FORMAT *
BM-7A Series
58
3. Various Settings
* Average Measure *
Single
When measuring varying-luminance light source, use this function. Measurement result in
Average measure is the average value of 5 times measurement and, measurement interval
is about 1 second.
Maintenance Display
*ON
ON : A buzzer will sound and the display will not advance to the next screen until any
button is pressed.
OFF : The message is displayed but the buzzer does not sound and it is possible to
advance to the next screen.
The procedure for making changes is as described below.
59
4. Communication With a PC
4. Communication With a PC
4.1 Communication Commands
This instrument can communicate with a PC. The communication format is either the BM-7A
Series mode and the former BM-7fast mode.
Changing the Communication Format ☞「3.11 Communication Format」
This chapter described the commands for creating communication program with the BM-7AC on
user own.
BM-7A Series Mode Communication Commands
☞「4.2 BM-7A Series Mode Communication Commands」
BM-7fast Mode Communication Commands ☞「4.4 BM-7fast Mode Communication Commands」
60
4. Communication With a PC
FK[n] Switches between normal correction/direct correction.
n: 1 or 2; 1: Normal correction, 2: Direct correction
FKR Reads the current correction type.
1: Normal correction 2: Direct correction
FAG[n] Enables area correction.
n: Number of the area correction group (1 to 10)
FO Disables area correction.
CGL[n] Clears the chromaticity area in the area correction factor per
group units.
n: Number of the area correction group (1 to 10)
FGR Reads the number of the area correction group currently
being used.
RG[m]K[n] Reads the area correction factor.
m: Group number (1 to 10) n: Area number (1 to 5)
RG[m]L[n] Reads the chromaticity area in the area correction factor.
m: Group number (1 to 10) n: Area number (1 to 5)
WG[m]K[n]‗####‗####‗#### Writes the area correction factor.
m: Group number (1 to 10) n: Area number (1 to 5)
WG[m]L[n]‗####‗####‗#### Writes the chromaticity area in the area correction factor.
m: Group number (1 to 10) n: Area number (1 to 5)
WHO Reads the product name (“BM-7AC” is output).
VER Reads the software version.
SRL Reads the manufacturing serial number.
UT Reads the measurement unit.
C:cd/㎡
CT Reads the amount of time elapsed since calibration.
"_" indicates a space. "####" indicates a number.
When a communication command is sent from the PC, the BM-7AC returns "OK" as a reception
check command. When the BM-7AC received an improper command, it returns "NO".
61
4. Communication With a PC
4.2.1 ST Command
When this command is sent to the BM-7AC, measurement starts. After measurement is finished,
the measurement data is sent from the BM-7AC.
PC Command BM-7AC
ST+(Cr・Lf)
OK+(Cr・Lf)
Measurement Start
・ ・ ・ ・ ・
Measurement end
Data communication
Data 1 + (Cr・Lf)
Data 2 + (Cr・Lf)
・
・
・
Data 21 + (Cr・Lf)
END + (Cr・Lf)
62
4. Communication With a PC
4.2.2 CA Command
The Zero-adjustment is performed.
PC Command BM-7AC
CA + (Cr・Lf)
OK + (Cr・Lf)
END + (Cr・Lf)
4.2.3 TF Command
The response speed of the photo-receiver circuit of the instrument is changed to the FAST mode.
PC Command BM-7AC
TF + (Cr・Lf)
OK + (Cr・Lf)
4.2.4 TS Command
The response speed of the photo-receiver circuit of the instrument is changed to the SLOW
mode.
PC Command BM-7AC
TS + (Cr・Lf)
OK + (Cr・Lf)
4.2.5 MA Command
The measurement range of the instrument is changed to the AUTO range.
PC Command BM-7AC
MA + (Cr・Lf)
OK + (Cr・Lf)
63
4. Communication With a PC
4.2.6 MM Command
The measurement range of the instrument is changed to the MANUAL mode.
Sets the X,Y and Z setting ranges.
MM‗X[l]‗Y[m]‗Z[n] : l is the measurement range for X 1 to 5
m is the measurement range for Y 1 to 5
n is the measurement range for Z 1 to 5
"‗" indicates a space.
PC Command BM-7AC
MM‗X3‗Y3‗Z3 + (Cr・Lf)
OK + (Cr・Lf)
4.2.7 FR Command
The correction factor currently being used by the instrument are read out.
PC Command BM-7AC
FR + (Cr・Lf)
OK + (Cr・Lf)
Correction factor number
END + (Cr・Lf)
4.2.8 F Command
The correction factor currently to be applied to the instrument are stored the instrument.
F [n] : n is the number of the correction factor to be currently being used. 0 to 15
PC Command BM-7AC
F15 + (Cr・Lf)
OK + (Cr・Lf)
64
4. Communication With a PC
4.2.9 R Command
The Correction factor in the instrument is read out.
R [n] : n is the number of the correction factor to be read. 1 to 15
PC Command BM-7AC
R15 + (Cr・Lf)
OK + (Cr・Lf)
Correction Factor
END + (Cr・Lf)
4.2.10 W Command
The correction factor is stored in the instrument.
W [n] : n is the number of the correction factor to be stored. 1 to 15
PC Command BM-7AC
W10 + (Cr・Lf)
OK + (Cr・Lf)
4.2.11 CF Command
The correction factor in the instrument is cleared.
CF [n] : n is the number of the correction factor to be cleared. 1 to 15
PC Command BM-7AC
CF10 + (Cr・Lf)
OK + (Cr・Lf)
65
4. Communication With a PC
4.2.12 FK Command
Normal correction/direct correction can be selected by this command.
FK[n]:n is 1 or 2. 1: Normal correction 2: Direct correction
PC Command BM-7AC
FK1 + (Cr・Lf)
OK + (Cr・Lf)
FKR + (Cr・Lf)
OK + (Cr・Lf)
Correction type
END + (Cr・Lf)
PC Command BM-7AC
FAG2 + (Cr・Lf)
OK + (Cr・Lf)
66
4. Communication With a PC
4.2.15 FO Command
The area correction can be deactivated by this command.
PC Command BM-7AC
FO + (Cr・Lf)
OK + (Cr・Lf)
PC Command BM-7AC
CGL1 + (Cr・Lf)
OK + (Cr・Lf)
PC Command BM-7AC
FGR + (Cr・Lf)
OK + (Cr・Lf)
Group Number
END + (Cr・Lf)
67
4. Communication With a PC
4.2.18 RG K Command
The area correction factor data can be read out by this command.
RG[m]K[n] : m is the area correction factor group number 1 to 10
n is the area number 1 to 5
PC Command BM-7AC
RG1K3 + (Cr・Lf)
OK + (Cr・Lf)
Area correction factor + (Cr・Lf)
END + (Cr・Lf)
4.2.19 RG L Command
The chromaticity area data to be applied area correction factor are read out by this command.
RG[m]L[n] : m is the number of the area correction factor group. 1 to 10
n is the area number. 1 to 5
PC Command BM-7AC
RG1L3 + (Cr・Lf)
OK + (Cr・Lf)
chromaticity area + (Cr・Lf)
END + (Cr・Lf)
4.2.20 WG K Command
The area correction factor can be stored in the instrument.
WG[m]K[n] : m is the number of the area correction factor group. 1 to 10
n is the area number. 1 to 5
PC Command BM-7AC
WG1K3‗####‗####‗#### + (Cr・Lf)
OK + (Cr・Lf)
####‗####‗#### + (Cr・Lf)
END + (Cr・Lf)
The values for the correction factors KX, KY and KZ are input in order in ####_####_####.
68
4. Communication With a PC
4.2.21 WG L Command
The chromaticity area to be applied the area correction factor can be store in the instrument by
this command.
WG[m]L[n] : m is the number of the area correction factor group. 1 to 10
n is the area number. 1 to 5
PC Command BM-7AC
WG1L3‗####‗####‗####‗####‗#### + (Cr・Lf)
OK + (Cr・Lf)
####‗####‗####‗####‗#### + (Cr・Lf)
END + (Cr・Lf)
The minimum value of the chromaticity x, the minimum value of the chromaticity y, the maximum
value of the chromaticity x, the maximum value of the chromaticity y and the luminance are input
in order in ####‗####‗####‗####‗####.
PC Command BM-7AC
WHO/VER/SRL + (Cr・Lf)
OK + (Cr・Lf)
Model name/serial number/version
+ (Cr・Lf)
END + (Cr・Lf)
4.2.23 UT Command
The unit of luminance of the instrument can be read out by this command.
C:cd/㎡
PC Command BM-7AC
UT + (Cr・Lf)
OK + (Cr・Lf)
Unit + (Cr・Lf)
END + (Cr・Lf)
69
4. Communication With a PC
4.2.24 CT Command
The amount of elapsed time since the last calibration of the instrument can be read out by this
command.
PC Command BM-7AC
CT + (Cr・Lf)
OK + (Cr・Lf)
Time [minutes] + (Cr・Lf)
END + (Cr・Lf)
PC Command BM-7AC
AM/SM + (Cr・Lf)
OK + (Cr・Lf)
70
4. Communication With a PC
71
4. Communication With a PC
■ BM-7fast Mode
Output data are three kinds of data; measurement conditions, color coordinates/correlated color
temperature and tristimulus values.
-M0 Command-
T*R*XnYnZnUCFn_
x=_#.####_y=_#.####_
X=_#.###E±##_Y=_#.###E±##_Z=_#.###E±##CR
-M1 Command -
T*R*XnYnZnUCFn_
u’=_#.####_v’=_#.####_
X=_#.###E±##_Y=_#.###E±##_Z=_#.###E±##CR
-M2 Command -
T*R*XnYnZnUCFn_
Tc=_#.####_duv=_#.###_
X=_#.###E±##_Y=_#.###E±##_Z=_#.###E±##CR
* A space is indicated by _.
* A numerical value is indicated by #.
* A carriage return is indicated by CR.
72
4. Communication With a PC
Communication Function
Command
ST Sets the instrument in the measurement status.
After measurement, returns measurement data from the device.
CA Performs calibration.
TF Switches the response speed of the light-receiver circuit to the FAST
mode.
TS Switches the response speed of the light-receiver circuit to the SLOW
mode.
RA Switches the measurement range to the AUTO range.
RM Switches the measurement range to the MANUAL range.
X[n] Sets the MANUAL range for X.
n: Measurement range (1 to 5)
Y[n] Sets the MANUAL range for Y.
n: Measurement range (1 to 5)
Z[n] Sets the MANUAL range for Z.
n: Measurement range (1 to 5)
M0 Switches the display mode (xyL mode).
M1 Switches the display mode (u'v'L mode).
M2 Switches the display mode (TC, duv, L mode).
FR Reads out the number of the correction factor currently being used.
F[n] Writes the number of the correction factor to be used.
n: Number of the correction factor to be used (0 to 15)
R[n] Reads out the correction factor.
n: Number of the correction factor to be read out (1 to 15)
W[n] Writes the correction factor.
n: Number of the correction factor to be write (1 to 10)
CF[n] Clears the correction factor.
n: Number of the correction factor to be cleared (1 to 15)
WHO Reads out the product name.
VER Reads out the software version.
SRL Reads out the manufacturing serial number.
73
4. Communication With a PC
74
5. USB Driver
5. USB Driver
The procedure for installing the USB Driver in the PC is as follows.
1 Place the CS-900A CD-ROM in the CD-ROM drive.
2 Select and double-click the “dpinst.exe” file in [\USB_Driver\{os name}\ {x86} or {x64}]
folder in the CD-ROM via Explorer.
OS Windows® 7 Ultimate(32bit/64bit)
Windows® 7 Professional (32bit/64bit)
Windows® 8.1 Pro (32bit/64bit)
Windows® 10 Pro (32bit/64bit)
75
5. USB Driver
5 In the case of Windows7, 8, the Windows Security dialog appears. Click the [Install].
The following dialog will be displayed, start installing the USB Driver.
76
Appendix
Appendix
Specifications
Measurement angle
2°/1°/0.2°/0.1°Switch type
Optical system
Finder visual field : 5°
Objective lens : f = 80 ㎜ F2.5
Photo-detector
Silicon photodiode 3-element type (X, Y, Z)
Measurement diameter
(Unit: ㎜φ)
Measurement Measurement distance (m)
angle 0.35 0.5 1 5 10
2° 10.0 15.4 32.8 169 341
1° 5.0 7.7 16.4 85 170
0.2° 1.0 1.5 3.3 17 34
0.1° 0.5 0.8 1.6 8 17
Measurement Functions
The following measurement can be performed by using display mode.
x, y, L (x, y: Chromaticity coordinates, L: Luminance) ±⊿
u', v', L (u', v': Chromaticity coordinates, L: Luminance) ±⊿
X, Y, Z (X, Y, Z: Tristimulus values) ±⊿
Tc, duv, L (Tc: Color temperature, duv: Deviation) ±⊿
CIE 1976 L*a*b* ⊿Eab* ±⊿
CIE 1976 L*u*v* ⊿Euv* ±⊿
* ±⊿ indicates the difference from the reference value.
For example, in the case of x, y and L, it means ⊿x, ⊿y and ⊿L.
77
Appendix
Response Speed
Display interval: About 0.5 seconds for both FAST and SLOW.
Analog Output (Unit: ms)
Range1 Range2 Range3 Range4 Range5
X 30 30 30 0.3 0.3
FAST Y 30 30 30 0.3 0.3
Z 30 30 30 0.3 0.3
The response speed in the table above means the time that it takes analog output from the
instrument to reach 90% of the peak value, when measuring an LED driven by a square wave from
a function generator.
Luminance Unit
Candela per square meter (cd/㎡)
Luminance accuracy
For a measurement angle of 2°:
1 to 5cd/㎡ : ±4% or less (auto range, standard sources A)
5cd/㎡ or greater: ±2% or less (auto range, standard sources A)
Guaranteed Accuracy Range
Measurement
Range and Specification
Angle
1 to 5cd/㎡ : ±4% or less
2°
5cd/㎡ or more : ±2% or less
4 to 20cd/㎡ : ±4% or less
1°
20cd/㎡ or more : ±2% or less
100 to 500cd/㎡ : ±4% or less
0.2°
500cd/㎡ or more : ±2% or less
400 to 2000cd/㎡ : ±4% or less
0.1°
2000cd/㎡ or more : ±2% or less
78
Appendix
Chromaticity accuracy
Guaranteed accuracy range
Measurement
Luminance (cd/㎡)
Angle
2° 10 ~ 30,000
1° 40 ~ 120,000
0.2° 1,000 ~ 3,000,000
0.1° 4,000 ~ 12,000,000
⊿x, ⊿y : ±0.002 or less(auto range, standard sources A)
* Chromaticity is displayed even at a luminance less than that given above.
Repeatability
Luminance
For 2°
1 to 5cd/㎡ : 1% or less (2σ, SLOW Mode, AUTO range, standard sources A)
5cd/㎡ or greater : 0.5% or less (2σ, SLOW mode, AUTO range, standard sources A)
Measurement Range and Specification
Angle
1 to 5cd/㎡ : 1% or less
2°
5cd/㎡ or more : 0.5% or less
4 to 20cd/㎡ : 1% or less
1°
20cd/㎡ or more : 0.5% or less
100 to 500cd/㎡ : 1% or less
0.2°
500cd/㎡ or more : 0.5% or less
400 to 2000cd/㎡ : 1% or less
0.1°
2000cd/㎡ or more : 0.5% or less
Chromaticity
For 2°
1 to 5cd/㎡ : Chromaticity xy 0.005 or less (SLOW mode, AUTO range, standard
sources A)
5cd/㎡ or more : Chromaticity xy 0.002 or less (SLOW mode, AUTO range, standard
sources A)
Measurement Range and Specification
Angle
1 to 5cd/㎡ : Chromaticity xy 0.005 or less
2°
5cd/㎡ or more : Chromaticity xy 0.002 or less
4 to 20cd/㎡ : Chromaticity xy 0.005 or less
1°
20cd/㎡ or more : Chromaticity xy 0.002 or less
100 to 500cd/㎡ : Chromaticity xy 0.005 or less
0.2°
500cd/㎡ or more : Chromaticity xy 0.002 or less
400 to 2000cd/㎡ : Chromaticity xy 0.005 or less
0.1°
2000cd/㎡ or more : Chromaticity xy 0.002 or less
79
Appendix
Temperature Characteristics
±3% or less (23℃ is regarded as standard in the range of 0℃ to 40℃)
Humidity Characteristics
±3% or less (60%RH is regarded as standard in the range of 60%RH or less)
Calibration Reference
TOPCON TECHNOHOUSE calibration reference (Standard sources A, 23℃±3℃)
User calibration reference Switchable
Display
Dot-matrix LCD 20 characters x 4 rows with backlight function
Interface
USB/RS-232C
Power Supply
AC adapter (standard supplied part)
Power Consumption
Approximately 2.5VA when using an AC adapter
Operating Conditions
Temperature : 0℃ to 40℃
Humidity : 85%RH or less
Storage Conditions
Temperature : -20℃ to 60℃
Humidity : 85%RH or less(No condensation)
External Dimensions
Approximately 325 ㎜ (Length) × 120 ㎜ (Width) × 162 ㎜ (Height)
Weight
Approximately 3kg
* The precision value (|uminance ancl chromaticity) is the specification value by our standard light
source and measurement condition.
Measurement error that is out of specification value may occur by the difference of the light source,
measurement condition and measurement environment.
80
Appendix
81
Appendix
Appended Tables
Foot-Lamberts
cd/㎡ (nt) cd/cm2 (sb) cd/ft2 rlx (asb) Lamberts
(ft-L)
1 10-4 0.0929 3.1416 0.000314 0.2919
3.426 0.0003426 0.3183 10.764 0.001076 1
Chromaticity accuracy of the combination of standard light source A and our colored glass
filter. (Reference)
⊿x, ⊿y : ±0.03 or less (auto range, standard sources A)
82
Appendix
System Diagram
Block Diagram
Eyepiece
Mirror
Relay lens
Reticle
Analog output
Relay lens
Buffer connector
X X
Y Y
Z
Infrared cut filter Z
Correction system
LCD Control switch
Spectral Sensitivity Curves
2.0
1.5
1.0
0.5
0
400nm 500nm 600nm 700nm
83
Appendix
External Dimensions
Use only specified screws when using the tripod screw and screw holes for jig
Request attachment. Do not tighten the screws any more than necessary. Doing so might
cause internal breakage.
Tripod seat
(For tooling) reference surface
Rubber leg
Installation surface
Chromaticity Coordinates
xy chromaticity coordinates of XYZ color coordinate system
X Y
x y
XYZ XYZ
4X 9Y
u´ v´
X 15Y 3Z X 15Y 3Z
Lightness Index
1
Y 3 Y
0.008856
L 116 16
*
Y Yn
n
Y : Tristimulus value Y in XYZ color coordinate system
Yn : Y value according to standard light on perfect reflecting diffuser
However, when Y is less than 0.008856, the following formula must be applied.
Yn
Y
* Y ≦ 0.008856
L 903.29 Yn
Yn
Chromaticness Index
CIE 1976 L*a*b* color coordinate system
1 1
* X 3 Y 3 X
a 500 >0.008856
Xn Yn Xn
1 1
* Y 3 Z 3 Y
b 200 >0.008856
Yn Zn Yn
Z
>0.008856
Zn
X, Y, Z : Tristimulus values in the XYZ color coordinate system
Xn, Yn, Zn : Tristimulus values in the XYZ color coordinate system on a perfect reflecting
diffusier.
X Y Z
However, if X , Y or Z have a value less than 0.008856, substitute the corresponding curbe
n n n
roots in the above formula with the values as shown below;
85
Appendix
X 16
7.787
Xn 116
Y 16
7.787
Yn 116
Z 16
7.787
Zn 116
CIE 1976 L*u*v* color coordinate system
* *
u 13L (u'un ' )
* *
v 13L (v'vn ' )
u' , v' : u'v' color coordinates
un', vn' : u' and v' values according to standard light on perfect reflecting diffuser.
Color Difference
CIE 1976 L*a*b* color difference 1
2
* *2
⊿E ab ⊿L ⊿a
* 2
⊿b
* 2
⊿L*, ⊿a*, ⊿b*: Difference in lightness indexes L*, and chromatic indexes a*, b* between two
non-luminous object color.
CIE 1976 L*u*v* color difference 1
2
* *2
⊿E uv ⊿L
⊿u
* 2
⊿v
* 2
⊿L*, ⊿u*, ⊿v*: Difference between the lightness indexes L *, and chromatic indexes u*, v*
between two non-luminous object color.
86
Appendix
Terminology
Correction Factor :
This means three factors, KX, KY and KZ to correct tristimulus values.
Standard sample :
This is a sample used as standard for color matching.
White Board :
This is a white board with good diffusion, which is made of barium sulfate, etc.
Local Mode :
This means a normal measuring status.
The instrument is operated by the panel switches, allowing no control from outside.
Remote Mode :
The instrument is controlled by an external personal computer via interface.
It can also be controlled by its panel switches.
Function Mode :
This is a mode to display and change the data stored in the memory of this instrument.
87
Warranty
Warranty Period
One year from the date of shipment
Maintenance Period
The repair parts (*1) are retained by us for eight years (*2) after purchase.
The repairable period is this period that parts are kept in stock.
Even after the storage period has elapsed, there are cases in which repair may be possible, so contact
the dealer or TOPCON TECHNOHOUSE CORPORATION.
(*1) Maintenance and repair parts are parts that are necessary to maintain the function of the
product.
(*2) We make our most effort to keep maintenance and repair parts in stock for the complete
storage period, however, due to some unexpected occurrence, the storage period may have to
be shortened.
Make inquiries to the address indicated on the base of the instruction manual.
Luminance
Colorimeter
Contact information: