Begins Here.: Clear Vision
Begins Here.: Clear Vision
VISION
BEGINS
HERE.
EleVision™ HD 2 platform
installation guide
Clear vision is essential
in your OR.
The EleVision™ HD 2 platform
provides a customizable
solution for your visualization
needs — and it all starts with
a thorough installation.
CONTENTS
Trolley/Cart Installation
Camera Control Unit
Camera Head
TipVision™ Videoscope
LED Light Source
Trolley/Cart
CONTENTS
Trolley/Cart Installation
Camera Control Unit
3 | TROLLEY/CART INSTALLATION
Camera
CONTENTS
Trolley/Cart Installation
Camera Control Unit
†All settings are reset to the default settings. If necessary, only restore the default
settings of an individual specialty.
Navigation in the menu of the controller 3. Insert the endoscope including the eyepiece cup
using the camera head buttons and release the locking mechanisms.
1. Check that the glass surfaces of the camera head 4. Secure the endoscope against inadvertent
are completely clean. opening. Use the locking device on the endocoupler
to do so.
2. Check that the adjusting ring(s) is/are operational.
5. If applicable, draw the sterile drape over the camera
3. Check that the camera cable is undamaged. head and camera cable toward the controller until
the drape is completely unfolded. The endoscope is
4. Do not use the camera head if the camera cable is
now connected.
kinked, broken, or twisted.
6. Connect light cable to the endoscope.
5. Check that the contacts on the camera cable
connection plug are free from moisture and dirt.
Connecting the device to the mains power
6. Do not use the camera head if there is any moisture
in the plug or contacts are soiled. Do not connect the device to the mains until you have
completed all the previous activities. Observe the
7. Inspect the housing for external damage. safety notices provided at the start of this chapter.
8. Do not use the device if there is any external 1. Insert the power cord into the mains connection
damage to the housing. socket on the rear of the device.
Note: The user is responsible at all times for 2. Switch on the device and all neighboring devices.
maintaining the sterile barrier.
3. Connect the camera.
2. Press together the two locking mechanisms on the
endocoupler to spread apart the retaining clips. 4. Set the white balance.
Image Setting
∙ Contact our service department if you
require further assistance.
Optimized camera settings can be selected in
the configuration menu for selected endoscopic
specialties.
The following configurable parameters are available Choose a window that is similar in size or smaller
for each specialty: than the round endoscope image.
†This function is not available with the following camera heads: ELEVTV1000 TipVision™
videoscope 10 x 341 mm, 0 degree autoclavable, and ELEVTV1030 TipVision™
videoscope 10 x 341 mm, 30 degree autoclavable.
†As evaluated in a clinical study with the FICE platform (Fujinon Corporation) in the identification of gastroduodenal lesions.
‡Compared to white light image. 13 out of 19 surgeons surveyed agreed.
§Based on preclinical or animal studies. Results may not correlate to performance in humans.
Reset to Default Settings — Full Reset† If you wish to cancel the process, select cancel and
confirm. The dialog window closes and the setup
1. Press the menu button on the front side of the menu is shown.
device for approximately two seconds.
The main menu appears on the screen. If you wish to reset the device to the default settings,
confirm the selection.
2. Navigate to the options menu entry.
The device is reset to the default settings and
3. Press the menu button for approximately two restarts.
seconds. The setup menu appears on the monitor.
†All settings are reset to the default settings. If necessary, only restore the default
settings of an individual specialty.
†All settings are reset to the default settings. If necessary, only restore the default
settings of an individual specialty.
Note: Keep in mind that ambient light that deviates 2. Activate the release.
significantly from the color temperature of the light
source can affect the results of the white balance. 3. Depending on configuration and connected
accessories press the affected button on the
camera head, activate the affected foot switch.
The capture is saved.
12 | CAMERA HEAD
TipVision
CONTENTS
™ Trolley/Cart Installation
Camera Control Unit
Videoscope Buttons
General information and default button settings
balance can only be assigned to buttons 1 and 3. It’s 1. Hold the horizon holder (dark blue) in its horizontal
activated by holding down the button. orientation.
The default settings for 2. Turn the handpiece (light blue) to the left or right
the buttons are as follows: until you have found the desired viewing range.
13 | TIPVISION™ VIDEOSCOPE
LED Light
CONTENTS
Trolley/Cart Installation
Camera Control Unit
If this occurs, you should attach the light guide, attach 5. Check that the LED can go into standby mode and
it correctly, or switch the device off. be reactivated
2. The LED switches off, the device remains Activate automatic light control
switched on. Proceed as follows:
3. Pull out the light guide. 1. Using the MIS-Bus cable, connect the light
source to a compatible camera. Observe the
4. Detach the adapters used, and arrange for instructions for use supplied with the camera.
reprocessing the light guide. Observe the
instructions for use supplied with the light guide. 2. Activate auto light control via the camera
menu. When auto light control is activated, this
5. The light guide has been decoupled. message appears on the light source display.
Setting the light intensity manually Auto light control is now activated.
© 2020 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are
trademarks of Medtronic. All other brands are trademarks of a Medtronic company.