Edited Clean Room Checklist (ISO 14644)
Edited Clean Room Checklist (ISO 14644)
Document
Para. Req.
REQUIREMENTS
2.1.1
The Contamination Control Services Laboratory shall Request to see the filter replacement
maintain the certification, High Efficiency Particulate Air log or database and verifty the status
Filter (HEPA)/Ultra Low Penetration Air Filter (ULPA), on frequency replacement of filters.
and pre-filter replacement status for all of these sites in
order to comply with the documentation requirements of
ISO.
2.2
The Classification of the Clean Areas shall be Ensure that the log or database
documented per the ISO 14644-1 standard (available identifies the classification of the
through http://standards.nasa.gov/). Clean Areas.
2.3
The ISO 14644-1 tests shall include but are not limited to Select random samples and verify
airborne particulate counting and sizing, air velocity that tests are performed per ISO
testing, and differential pressure testing. standard above.
2.5
The Controlled Environment areas (> class 100,000) shall Verify records that particulate
require only particulate testing unless otherwise requested testing has been performed on Clean
by the customer. Rooms Class 100,000 or greater.
square footage of each Clean Room or Controlled that the square footage has been
Environment to establish the number of samples to be measured and recorded.
taken in the area.
3.2
The Contamination Control personnel shall take the Select a ramdom sample of test
proper number of samples at each location as defined by results and verify that the
the ISO 14644-1 Annex B Section B.4.1.1 using a (DPC) appropriate number of samples were
Calibrated particle counter (Calibrated within 180 days taken per the ISO standard.
designated to count the required particle sizes for the area
which will determine the classification of the clean area.
3.3
Contamination Control personnel shall perform a velocity Ensure that the velocity test was
test in the clean area. performed.
3.4
The number of tests performed shall be based on the Verify that the number of tests
square footage of the area. coincide with square footage of the
area.
3.5
Testing information shall be recorded on the certification Review the certification document
data or memo to the customer. and verify that testing information is
included.
3.6
Contamination Control personnel shall determine if the Seek evidence that
flow is adequate for the specified area and make recommendations have been made in
recommendations to the customer in writing where the writing by reviewing test records.
flow is deemed to be low or out of specification.
3.7
Differential pressure readings shall be performed in all Ensure that differential pressure test
clean areas to determine the existence of positive was performed and the resulting data
pressure. was provided to the cusotmer.
3.10
The Contamination Control personnel shall perform a Ensure that a visual for cleanliness
visual cleanliness survey of the Clean Area to be tested has been performed per specification
under the ISO standard. and recorded and filed.
Note :
RESULTS CODES; M = MAJOR NONCONFORMANCE C = COMMENDATION
N = MINOR NONCONFORMANCE S = VERIFIED SATISFACTORY
O = OBSERVATION N/A = NOT AUDITED
IL Contract ATTRIBUTE METHOD OF VERIFICATION Supplier Results REMARKS
Document
Para. Req.
Note :
The following testing shall be optional.
Table 1 (below) shall be used as the minimum Review the specification and verify
requirements for clean rooms, controlled environments, that the requirements listed in Table
and workstations (as specified in ASTM E2352-04, 1 have been met as a minimum.
available through http://standards.nasa.gov/)
5.3
The contractor shall supply to the JPL Contamination Ensure that the contractor satisfies
RESULTS CODES; M = MAJOR NONCONFORMANCE C = COMMENDATION
N = MINOR NONCONFORMANCE S = VERIFIED SATISFACTORY
O = OBSERVATION N/A = NOT AUDITED
IL Contract ATTRIBUTE METHOD OF VERIFICATION Supplier Results REMARKS
Document
Para. Req.
Control Laboratory a documentation package that this requirement and verify that it is
certifies the work complies with all the ISO standard per ISO.
requirements, and specifies the type of work performed,
and any optional tests performed.
5.4
The contractor shall build the new Clean Room facility in Review the documents supplied to
accordance to the ISO 14644 documents and supply the the JPL Contamination Control
JPL Contamination Control Laboratory documentation Laboratory documentation and
which shall include all tests in accordance with that verify that requirements in ISO have
document, and sizes, part number of manufacturer been met.
number, and statement that the filters were not tested
with Dioctylphthalate (DOP), plasticizer test materials or
other non-approved JPL test methods with a CNC for the
HEPA or ULPA filter installation.
Note:
6.2
The Contamination Control personnel shall determine by Ensure that Contamination Control
the tests performed in the certification process when the certification includes tests
filter require replacement based on the airflow or the performed to determine when filter
differential pressure drop of the filter. replacement is required.
6.4
All HEPA filters older than 7 years old shall be replaced. Review filter log/records and verify
that these fitlers have been replaced
Notes: at the 7th year of use or before.
Waivers for filters older than 7 years shall be reviewed
on an individual basis.
7.7
Re-certification tests shall be performed on the clean Select a sample of records and verify
rooms at specific intervals to demonstrate compliance that re-certification tests were
with ISO 14644-1 according to the classification of the performed at specific intervals.
installation.
Note:
10.4
An Inter-Office Memo shall be initiated with the Request a copy of the memo and
classification, differential pressure and velocity data, verify all the pertinent information
customer’s name, charge number, location of the work has been included.
performed and the calibration date of the particle counter.