Current Probes: User Manual
Current Probes: User Manual
User Manual
ETS-Lindgren L.P. reserves the right to make changes to any product described herein in order to improve function, design, or for any other reason. Nothing contained herein shall constitute ETS-Lindgren L.P. assuming any liability whatsoever arising out of the application or use of any product or circuit described herein. ETS-Lindgren L.P. does not convey any license under its patent rights or the rights of others. Copyright 2008 by ETS-Lindgren L.P. All Rights Reserved. No part of this document may be copied by any means without written permission from ETS-Lindgren L.P. Trademarks used in this document: The ETS-Lindgren logo is a trademark of ETS-Lindgren L.P. Revision Record MANUAL DESCRIPTION | Part #399297, Rev. A Revision A Description Initial Release Date December, 2008
ii
Table of Contents
3.0 Specifications..................................................................... 13
Model 91197 Series Current Probe ............................................................. 13 Model 91197 Series Physical Specifications ....................................... 13 Model 91197 Series Electrical Specifications ...................................... 14 Model 91550 Series Current Probe ............................................................. 15 Model 91550 Series Physical Specifications ....................................... 15 Model 91550 Series Electrical Specifications ...................................... 15 Model 91550 Series Pulse Power Limits ............................................. 17 Model 93511 Series Current Probe ............................................................. 18 Model 93511 Series Physical Specifications ....................................... 18 Model 93511 Series Electrical Specifications ...................................... 18 Model 93686 Series Current Probe ............................................................. 19 Model 93686 Series Physical Specifications ....................................... 19 Model 93686 Series Electrical Specifications ...................................... 20 Model 94106 Series Current Probe ............................................................. 21 Model 94106 Series Physical Specifications ....................................... 21 Model 94106 Series Electrical Specifications ...................................... 21 Model 94106 Series Pulse Power Limits ............................................. 22 Model 94111 Series Current Probe ............................................................. 23 Model 94111 Series Physical Specifications ....................................... 23 Model 94111 Series Electrical Specifications ...................................... 23 Model 94111 Series Pulse Power Limits ............................................. 24 Model 94430 Series Current Probe ............................................................. 25 Model 94430 Series Physical Specifications ....................................... 25 Model 94430 Series Electrical Specifications ...................................... 25 Model 94430 Series Pulse Power Limits ............................................. 26 | iii
Model 94606 Series Current Probe ............................................................. 27 Model 94606 Series Physical Specifications ....................................... 27 Model 94606 Series Electrical Specifications ...................................... 27
iv
Model 91550-5 Transfer Impedance ................................................... 49 Model 93511 Series Current Probe ............................................................. 50 Model 93511-1 Transfer Impedance ................................................... 50 Model 93511-1L Transfer Impedance ................................................. 50 Model 93686 Series Current Probe ............................................................. 51 Model 93686-1 Transfer Impedance ................................................... 51 Model 93686-8 Transfer Impedance ................................................... 51 Model 94106 Series Current Probe ............................................................. 52 Model 94106-1 Transfer Impedance ................................................... 52 Model 94111 Series Current Probe ............................................................. 53 Model 94111-1 Transfer Impedance ................................................... 53 Model 94111-2 Transfer Impedance ................................................... 53 Model 94430 Series Current Probe ............................................................. 54 Model 94430-1 Transfer Impedance ................................................... 54 Model 94606 Series Current Probe ............................................................. 55 Model 94606-1 Transfer Impedance ................................................... 55
vi
Caution: Denotes a hazard. Failure to follow instructions could result in minor personal injury and/or property damage. Included text gives proper procedures.
Warning: Denotes a hazard. Failure to follow instructions could result in SEVERE personal injury and/or property damage. Included text gives proper procedures.
See the ETS-Lindgren Product Information Bulletin for safety, regulatory, and other product marking information.
vii
viii
1.0 Introduction
The ETS-Lindgren Current Probe is a clamp-on RF current transformer that determines the intensity of RF current present in an electrical conductor or group of conductors. The current probe is designed for use with electromagnetic interference (EMI) test receivers or spectrum analyzers, or with any similar instrument having a 50-ohm input impedance.
A current probe provides a way to accurately measure net (common mode) radio frequency current flowing on a wire or bundle of wires without requiring a direct connection to the conductor(s) of interest. The probe clamps around the test conductor which becomes a one turn primary winding; the probe forms the core and secondary winding of an RF transformer. Measurements can be made on single-conductor and multi-conductor cables, grounding and bonding straps, outer conductors of shielding conduits and coaxial cables, and so on.
Introduction
91197-1L
93511-1L
10
Introduction
2.0 Maintenance
Before performing any maintenance, follow the safety information in the ETS-Lindgren Product Information Bulletin included with your shipment.
WARRANTY
Maintenance is limited to external components such as cables or connectors. If you have any questions concerning maintenance, contact ETS-Lindgren Customer Service.
Annual Calibration
See the Product Information Bulletin included with your shipment for information on ETS-Lindgren calibration services.
Service Procedures
For the steps to return a system or system component to ETS-Lindgren for service, see the Product Information Bulletin included with your shipment.
Maintenance
11
12
Maintenance
3.0 Specifications
At the lower frequencies, the signal current IP level can be as great as allowed for maximum power current. When both signal and power currents are high, their sum should not exceed the given limits.
Model 91197 Series Current Probe MODEL 91197 SERIES PHYSICAL SPECIFICATIONS
Window Diameter: 3.18 cm 1.25 in Outside Diameter: 8.26 cm 3.25 in Width: 3.56 cm 1.4 in Output Connector: Weight: Type N 0.6 kg 1 lb 5 oz Impedance: 50
Specifications
13
14
Specifications
Model 91550 Series Current Probe MODEL 91550 SERIES PHYSICAL SPECIFICATIONS
Window Diameter: 3.18 cm 1.25 in Outside Diameter: 8.89 cm 3.5 in Width: 7.29 cm 2.87 in Output Connector: Weight: Type N 0.6 kg 1.31 lb Impedance: 50
100 amps
100 amps
100 amps
Specifications
15
91550-1 Maximum Power Current (400 Hz): Maximum Power Voltage: Internal Loading: Rated Ouptut Load Impedance: Sensitivity Under Rated Load: 350 amps, 50 Hz1500 Hz
No limitation; subject to adequate conductor insulation No 50 Yes 50 No 50 1.0 microampere with 1 microvolt sensitivity receiver and 1 ohm transfer impedance, or 10 mV across 50 ohm load for 0.01 amp signal
0.17 microampere 1.0 microampere with 1 microvolt sensitivity receiver and 6 ohm transfer impedance with 1 microvolt sensitivity receiver and 1 ohm transfer impedance, or 10 mV across 50 ohm load for 0.01 amp signal
16
Specifications
Specifications
17
Model 93511 Series Current Probe MODEL 93511 SERIES PHYSICAL SPECIFICATIONS
Window Diameter: 3.18 cm 1.25 in Outside Diameter: 8.26 cm 3.25 in Width: 3.56 cm 1.4 in Output Connector: Weight: Type N 0.6 kg 1 lb 5 oz Impedance: 50
200 amps
200 amps
18
Specifications
93511-1 Maximum Power Current (400 Hz): Maximum Power Voltage: Sensitivity Under Rated Load: 350 amps
93511-1L
800 amps
No limitation; subject to adequate conductor insulation 1.0 microampere with 1 microvolt sensitivity receiver and 1 ohm transfer impedance 1.0 microampere with 1 microvolt sensitivity receiver and 1 ohm transfer impedance
Model 93686 Series Current Probe MODEL 93686 SERIES PHYSICAL SPECIFICATIONS
Window Diameter: 6.65 cm 2.62 in Outside Diameter: 13.97 cm 5.5 in Width: 5.38 cm 2.12 in Output Connector: Weight: Type N 2.27 kg 5 lb Impedance: 50
Specifications
19
200 amps
62 amps
20
Specifications
Model 94106 Series Current Probe MODEL 94106 SERIES PHYSICAL SPECIFICATIONS
Window Diameter: 3.18 cm 1.25 in Outside Diameter: 8.26 cm 3.25 in Width: 3.56 cm 1.4 in Output Connector: Weight: Type N 0.43 kg 15 oz Impedance: 50
RF Current Range (RF CW): RF Current Range (Pulse): Maximum Power Current (DC60 Hz):
50 amps
200 amps
Specifications
21
94106-1 Maximum Power Current (400 Hz): Maximum Power Voltage: Rated Output Load Impedance: Sensitivity Under Rated Load: No limitation; subject to adequate conductor insulation 200 amps
50 0.1 microampere with 1 microvolt sensitivity receiver and 10 ohm transfer impedance
22
Specifications
Model 94111 Series Current Probe MODEL 94111 SERIES PHYSICAL SPECIFICATIONS
Window Diameter: 3.18 cm 1.25 in Outside Diameter: 8.89 cm 3.5 in Width: 3.56 cm 1.4 in Output Connector: Weight: Type N 0.42 kg 15 oz Impedance: 50
50 amps
200 amps
Specifications
23
94111-1 / 94111-1L Maximum Power Current (DC400 Hz): Maximum Power Voltage: Sensitivity Under Rated Load: 200 amps
94111-2
200 amps
No limitation; subject to adequate conductor insulation 0.2 microampere with 1 microvolt sensitivity receiver and 5 ohms transfer impedance 1 microampere with 1 microvolt sensitivity receiver and 1 ohm transfer impedance
24
Specifications
Model 94430 Series Current Probe MODEL 94430 SERIES PHYSICAL SPECIFICATIONS
Window Diameter: 1.91 cm 0.75 in Outside Diameter: 5.72 cm 2.25 in Width: 2.54 cm 1.0 in Output Connector: Weight: Type BNC 0.18 kg 6.5 oz Impedance: 50
70 amps
Specifications
25
94430-1 / 94430-1L Maximum Power Voltage: Sensitivity Under Rated Load: No limitation; subject to adequate conductor insulation 0.125 microampere with 1 microvolt sensitivity receiver and 8 ohm transfer impedance
26
Specifications
Model 94606 Series Current Probe MODEL 94606 SERIES PHYSICAL SPECIFICATIONS
Window Diameter: 12.7 cm 5.0 in Outside Diameter: 19.81 cm 7.8 in Width: 5.38 cm 2.12 in Output Connector: Weight: Type N 3.18 kg 7.0 lb Impedance: 50
0 amps100 amps
0 amps300 amps
0 amps100 amps
0 amps400 amps
Specifications
27
94606-1 Maximum Power Voltage: Rated Output Load Impedance: Sensitivity Under Rated Load: 50 0.2 microampere with 1 microvolt sensitivity receiver and 5 ohms transfer impedance
94606-6 No limitation; subject to adequate conductor insulation 50 1 microampere with 1 microvolt sensitivity receiver and 1 ohm transfer impedance
28
Specifications
The current probe is an inserted-primary type of radio frequency current transformer. When the probe is clamped over the conductor or cable in which current is to be measured, the conductor forms the primary winding. The clamp-on feature of this probe enables easy placement around any conductor or cable.
Circuit
The circuit is that of a radio frequency transformer, as illustrated on page 30. Because the current probe is intended for clamp-on operation, the primary shown on page 30 is the electrical conductor in which interference currents are to be measured. This primary is considered as one turn since it is assumed that the noise currents flow through the conductor and return to the source by way of a ground conductor such as a frame, common ground plane, or earth. On some current probe models the secondary output terminals are resistively loaded internally to provide substantially constant transfer impedance over a wide frequency range.
Principles of Operation
29
BASIC RF TRANSFORMER
30
Principles of Operation
Sensitivity
Probe sensitivity in microamperes depends on the sensitivity in microvolts of the receiving equipment with which it is used. The following tables show the relationship of receiving sensitivity in microvolts to the overall sensitivity of the probe and receiver in microamperes. This data is based on the transfer impedance of each model.
Principles of Operation
31
32
Principles of Operation
Principles of Operation
33
The influence of intermodulation on the current probe output as measured with the EMI test equipment is negligible for primary conductor power frequency currents under 300 amperes. For primary power currents above 300 amperes, measurements taken by the EMI test equipment generally will not be affected by intermodulation due to the averaging characteristics for the quasi-peak and peak functions; the readings will increase with current.
Transfer Impedance
The RF current (IP) in microamps in the conductor under test is determined from the reading of the current probe output in microvolts (ES) divided by the current probe transfer impedance (ZT).
IP
Or, in dB:
ES ZT
IP dBA
ES dBV
ZT dB
34
Principles of Operation
The typical transfer impedance of the current probe throughout the frequency range is shown in Typical Data on page 47. It is determined by passing a known RF current (IP) through the primary test conductor and noting the voltage (ES), developed across a 50-ohm load.
ZT
ES IP
Principles of Operation
35
36
Principles of Operation
37
Installation Instructions
The window (aperture) of the probe will accommodate cables up to the following maximum outside diameters. Current Probe Model 91197 Series Model 91550 Series Model 93511 Series Model 94106 Series Model 94111 Series Model 94430 Series Model 93686 Series Model 94606 Series Maximum Outside Diameter 1.25 inches 1.25 inches 1.25 inches 1.25 inches 1.25 inches 0.75 inches 2.62 inches 5.0 inches
For greatest accuracy, the conductor under measurement should be centered in the window of the current probe. Place the probe around the conductor(s) to be measured and then carefully lock the probe jaws. Otherwise, inadequate shielding or incorrect air gap will result and the measurement will not be accurate. The connecting cable used between the current probe and the EMI test equipment must have 50-ohm characteristic impedance and matching cable connectors. The current probe is calibrated for use only with a 50-ohm load. Therefore, the EMI test equipment must have a 50-ohm input impedance. Observe precautions regarding minimum bending radius when installing and using the cable. For long cables and at high frequencies, cable loss may also be a factor. Use low loss cables and perform cable loss corrections if necessary. | Assembly and Installation
38
The probe rejection of any external pickup from conductors not passing through the window is better than 60 dB. The presence of very strong magnetic fields will likely have an effect on probe sensitivity. Do not place the unit close to permanent magnets or the magnetic field structures of motors or generators.
39
40
6.0 Operation
Before connecting any components, follow the safety information in the ETS-Lindgren Product Information Bulletin included with your shipment.
If measuring uninsulated conductors: Use extreme care when installing the current probe and taking measurements. If possible, de-energize the test sample during assembly and disassembly of the setup. Also, arrange to center the test conductor in the current probe window for additional voltage breakdown protection. Do not permit the uninsulated current probe connector and cable connectors to come in contact with the ground plane or other nearby conductors. This will prevent possible measurement error due to ground loops, and will avoid danger from high voltages. Ensure that the 50-ohm load is capable of safely dissipating the incurred power. Should the load become disconnected, the developed voltage will be come much greater and may be very dangerous.
The RF current probe is a broadband RF transformer for use with EMI test equipment. Radio frequency currents can be measured in cables without physically disturbing the circuit.
Operation
41
42
Operation
Example: Frequency is 10.0 kHz; step 1 measurement is 52 dB above one microvolt. For example, suppose the transfer impedance of the current probe used in the example was 8.0 dB below one ohm at 10.0 kHz. Then, as outlined in step 2: 52 dB + 8.0 dB = 60 dB above one microampere at meter input.
IN TERMS OF dB ABOVE ONE MICROAMPERE PER MEGAHERTZ AT METER INPUT (BROADBAND INTERFERENCE MEASUREMENT)
1. Adjust the EMI test equipment for standard gain and make a peak measurement of the broadband interference (voltage output from the current probe) in terms of dB above one microvolt per megahertz. Use procedures outlined in the EMI test equipment instruction manual. 2. Subtract the transfer impedance of the current probe in dB at the test frequency from the dB measurement of the previous step. The result is the value of the broadband interference in terms of dB above one microamp per megahertz at meter input. At meter input as used in the MIL-I-26600 and MIL-I-6181D specifications refers to the current in the test sample lead. Example: Frequency is 100 kHz; step 1 measurement is 41 dB above one microvolt per megahertz. For example, suppose the transfer impedance of the current probe was 8.0 dB below one ohm at 100 kHz. Then, as outlined in step 2: 41 dB + 8.0 dB = 49 dB above one microamp per megahertz at meter input. This result is beyond the limit of 46.2 dB above one microamp per megahertz.
Operation
43
IN TERMS OF MICROAMPERE PER MEGAHERTZ IN TEST SAMPLE LEAD (BROADBAND INTERFERENCE MEASUREMENT)
1. Adjust the EMI test equipment for standard gain and make a measurement of the broadband interference (voltage output from current probe), in terms of microvolts per megahertz at meter input. Use procedures outlined in the EMI test equipment instruction manual. 2. Divide the microvolt per megahertz measurement of the previous step by the transfer impedance in ohms at the test frequency. The result is the value of conducted broadband interference in terms of microamps per megahertz in the test sample lead. Example: Frequency is 10.0 kHz; step 1 measurement is 8000 microvolts per megahertz. For example, suppose the transfer impedance of the current probe was 0.39 ohms. Then, as outlined in step 2, 8000/0.39 = 20513 microamps per megahertz in test sample lead.
44
Operation
Signal Injection
Current probes may be used to inject RF currents into test conductors when performing susceptibility tests. Injection is best accomplished with current probes that do not have internal loading. Internal loading will absorb part (or most) of the driving power and can seriously limit the maximum levels of voltage and current that the current probe can handle as an injection device. The current probe does not have an internal load and therefore is suited for signal injection. However, it will be limited by connector voltage rating (500 V) and by the thermal limit of the coil windings. A maximum continuous injection current into the current probe coils of 4.2 amps may be used; two or three times this level may be used for short periods of time. When injecting pulse signals, the average current should be held within the above limits, and the peak voltage held below 500 V. Because of variable circuit impedances, there is no easy way to compute the RF current that may be injected into the test conductor. The practical way to determine the injected current is to measure it with a second current probe on the test conductor. The second current probe can be any model that covers the frequency range of interest.
Operation
45
46
Operation
Typical Data
47
48
Typical Data
Typical Data
49
50
Typical Data
Typical Data
51
52
Typical Data
Typical Data
53
54
Typical Data
Typical Data
55
56
Typical Data
Appendix A: Warranty
See the Product Information Bulletin included with your shipment for the complete ETS-Lindgren warranty.
DURATION OF WARRANTIES
All product warranties, except the warranty of title, and all remedies for warranty failures are limited to two years.
Product Warranted Model 91197 Series Current Probe Model 91550 Series Current Probe Model 93511 Series Current Probe Model 93686 Series Current Probe Model 94106 Series Current Probe Model 94111 Series Current Probe Model 94430 Series Current Probe Model 94606 Series Current Probe
Duration of Warranty Period 2 Years 2 Years 2 Years 2 Years 2 Years 2 Years 2 Years 2 Years
Warranty
57