The document is an instruction manual for a compressor nebulizer model NE-C29. It contains sections on intended use, important safety notes, how to operate the device, and care/maintenance instructions. The safety notes section warns users to carefully follow instructions, clean parts properly before and after use, avoid fluid contact with the compressor, and only use approved accessories to prevent injuries or device malfunctions.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
44 views48 pages

Omron NE C29 CompAir Pro Compressor Nebulizer

The document is an instruction manual for a compressor nebulizer model NE-C29. It contains sections on intended use, important safety notes, how to operate the device, and care/maintenance instructions. The safety notes section warns users to carefully follow instructions, clean parts properly before and after use, avoid fluid contact with the compressor, and only use approved accessories to prevent injuries or device malfunctions.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 48

NE-C29-C1_D_M.qxd 11.11.

24 4:16 PM Page 1
NE-C29-C1_D_M02_111125.pdf

INSTRUCTION MANUAL

TM

Compressor Nebulizer
Model NE-C29

5
3

10/2011
1649823-2D
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 1

CONTENTS
Before Using the Unit
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Important Safety Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Know Your Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Guide to Using the NE-C29 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8

Operating Instructions
How to Prepare the Nebulizer for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
How to Take Medication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13

Care and Maintenance


Cleaning and Daily Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Maintenance and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
How to Change the Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Troubleshooting Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Optional Parts and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25

1
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 2

Before Using the Unit

INTENDED USE
Medical Purpose This product is intended to be used for inhaling medication
for respiratory disorders.
Intended User • Legally certified medical experts, such as doctor, nurse and
therapist, or healthcare personnel or patient under the
guidance of qualified medical experts.
• The user should also be capable of understanding general
machine operation and the content of instruction manual.
Intended Patients This product should not be used by patients, who are
unconscious or are not breathing spontaneously.
Environment This product is intended for use in a medical facility, such as
Hospital, clinic and doctor’s office, and in a room of general
household.
Durable period Durable periods are as follows, provided the product is used
to nebulize saline 10 times a day for 10 minutes each time at
room temperature (23°C).
Durable period may vary depending on usage environment.
Main unit 3 years
Nebulizer Kit 100 days
(Cap, Inhalation Top,
Vaporiser Head, Medication Cup)
Mouthpiece 100 days
Air Tube 100 days
Air Filter 60 days
Adult Mask 100 days
Child Mask 100 days
Precautions for use Warnings and cautions described in the instruction manual
should be observed.

2
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 3

IMPORTANT SAFETY NOTES


To assure the correct use of the product basic safety measures should always be followed
including the precautions listed below:

WARNING:
Indicates a potentially hazardous situation which, if not avoided, could result in death or
serious injury.
(Usage)
For type, dose, and regime of medication follow the instructions of your doctor or
respiratory therapist.
Clean and disinfect the nebulizer kit, mouthpiece and optional nosepiece or masks
before using them for the first time after purchase, if the device has not been used for
a long period of time.
If the device is used by more than one person, be sure to clean and disinfect parts
before they are used by each person.
Always dispose of any remaining medication after use, and use fresh medication each
time.
Do not use only water in the nebulizer, for nebulizing purposes.
Do not use the device where it may be exposed to flammable gas.
Do not block the ventilation slots. Never place the device where the ventilation slots
may be obstructed during operation. Do not cover the compressor with a blanket or
towel etc., during use.
(Risk of electrical shock)
Never unplug the power cord with wet hands.
The compressor and power cord are not waterproof. Do not spill water, or other
liquids, on these parts. If liquid does spill on these parts, immediately unplug the
power cord and wipe off the liquid with gauze or other absorbent material.
Do not use or store the device in humid locations, such as a bathroom.
Do not immerse the main unit in water or other liquid.
Although this device fulfils the provisions of the EMC (Electromagnetic Compatibility)
directive, the use of it should be avoided in direct vicinity of other electric devices.
(Maintenance and Storage)
Store the device and accessories in a clean location.
Always rinse the nebulizing parts with clean hot tap water after disinfecting them.
Never leave the cleaning solution in the nebulizing parts.
Be sure to wash the parts after use, and dry the parts immediately after washing them.
Keep the device out of the reach of unsupervised infants and children. The device may
contain small pieces that can be swallowed.
Do not store the air tube while there is moisture or medication remaining inside it.
3
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 4

IMPORTANT SAFETY NOTES


CAUTION:
Indicates a potentially hazardous situation which if not avoided, may result in minor or
moderate injury, or physical damage.
(Usage)
Provide close supervision when this device is used by, on, or near children or invalids.
Make sure that the vaporiser head, air filter, and other parts are attached correctly.
Make sure that the air filter is clean. If the air filter has changed colour, or has been
used on average for more than 60 days, replace it with a new one.
Do not tilt the nebulizer kit so that it is at an angle of greater than 45º, or shake it
while in use.
Do not use or store the device while the air tube is bent.
Approved for human use only.
Use only original OMRON parts and accessories. Use of unsupported parts may
damage the device.
Do not add more than 7 ml of medication to the medication cup.
Do not drop the compressor or nebulizer kit, or subject them to strong shocks.
Do not distort the vaporiser head, or jab the nozzle of the medication cup with a pin
etc.
Do not insert fingers or objects inside the compressor.
Do not block the air filter cover.
Do not disassemble or attempt to repair the compressor or power cord.
To avoid injury to the nose mucosa, do not squeeze the nosepiece into the back of the
nose.
Do not block the slit between the cap and the inhalation air inlet.
Do not use a microwave oven or hair dryer to dry the nebulizing parts.
Changes or modifications not approved by OMRON HEALTHCARE will void the
user warranty.
(Risk of electrical shock)
Plug the device into the appropriate voltage outlet. Do not overload power outlets or
use extension cords.
Always unplug the device from the power outlet after use.
Always disconnect the power cord before cleaning.
(Maintenance and Storage)
Do not carry or leave the nebulizer while the medication cup contains medication.
When sterilizing parts by boiling, make sure that the container does not boil dry.
4
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 5

IMPORTANT SAFETY NOTES


Never place, or attempt to dry the compressor, the power cord or nebulizing parts in a
microwave oven.

General Safety Precautions:


Inspect the device and parts before using the them each time, and check that there are
no problems. In particular, be sure to check the following:
- That the vaporiser head, nozzle or air tube are not damaged.
- That the nozzle is not blocked.
- That the compressor operates normally.
When using this device, there will be some noise and vibration caused by the pump in
the compressor. There will also be some noise caused by the emission of compressed
air from the nebulizer kit. This is normal and does not indicate a malfunction, or
reduce the working life of the device.
Operate the device only as intended. Do not use the device for any other purpose.
Do not use the device at temperatures greater than +40ºC.
Dispose of the device and any used accessories or optional parts according to
applicable local regulations. Unlawful disposal may cause environmental pollution.

5
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 6

KNOW YOUR UNIT


Compressor (Main Unit)

Nebulizing Parts
Storage Compartment
Cover
Inhalation Attachments
Power
Switch

Mouthpiece
5
3

Ventilation Adult Mask


Slots
Nebulizer Kit
Holder
Air Plug Air Filter Cover (*Air Filter Inside)
Air Connector Child Mask

Air Plug

Air Tube Nosepiece


(Option)

Rear View Nebulizer Kit

Cap

Inhalation Air Inlet


Inhalation Top

Ventilation
Slots
Vaporiser Head
Power Plug
Power Cord
NOTE:
Above shows the power plug Nozzle Medication Cup
for 230V 50Hz model. 5
3

Air Tube Connector

6
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 7

KNOW YOUR UNIT


Accessories:

Spare Air Filters x 5 Instruction Manual

7
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 8

GUIDE TO USING THE NE-C29

HOW TO PREPARE THE NEBULIZER FOR USE (Page 9)


NOTE:
Clean and disinfect the nebulizer kit and applicable inhalation attachments
before using the device for the first time, or if it has not been used for a
while.

HOW TO TAKE MEDICATION (Page 13)


NOTES:
• Do not tilt the nebulizer kit so that it is at an angle of
greater than 45°.
• Do not bend or fold the air tube.
max
5
3

HOW TO CLEAN PARTS AFTER USE (Page 16)


NOTE:
Failure to clean the nebulizing parts after use may lead
to infection.

HOW THE NEBULIZER KIT WORKS Baffle


The medication that is pumped up through the Aerosol Aerosol
medication channel is mixed with compressed
air which is generated by a compressor pump. Medication
Nozzle
The compressed air mixed with medication is channel
turned into fine particles and sprayed by being
Medication Medication
impacted to the baffle.

Compressed Air

8
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 9

Operating Instructions

HOW TO PREPARE THE NEBULIZER FOR USE


WARNING:
Clean and disinfect the nebulizer kit, mouthpiece and optional nosepiece or masks
before using them for the first time after purchase, if the device has not been used
for a long period of time.
For directions on cleaning and disinfecting the parts of the device refer to page 16 in this
instruction manual.
NOTE:
Make sure that the air filter is changed regularly. (See “How to Change the Air Filter”
on page 19.)

1. Make sure that the power switch is turned off.


Check that the power switch is in the position.

2. Plug the power plug into a power outlet.

3. Remove the mouthpiece and cap from the nebulizer kit.

4. Remove the inhalation top from the medication cup. 2


1) Rotate the inhalation top counter clockwise.
2) Lift the inhalation top out of the medication cup. 1
5
3

9
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 10

HOW TO PREPARE THE NEBULIZER FOR USE


5. Remove the vaporiser head from the medication cup.

6. Add the correct amount of prescribed medication to the


medication cup.

CAUTION:
The capacity of the medication cup is
2 to 7 ml. max
5
3

7. Reattach the vaporiser head to the medication cup.

8. Put the inhalation top back inside the medication cup.


1) Align the protrusion on the inhalation top with the
indent on the medication cup, as shown. 1
2) Rotate the inhalation top clockwise and align with
the medication cup until they click in place.
2
max
5
3

10
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 11

HOW TO PREPARE THE NEBULIZER FOR USE


9. Attach the desired inhalation attachment, as described below.
Attach the cap when using the mouthpiece or optional nosepiece. Using the cap with
these attachments will reduce the amount of medication that is discharged into the air.

Attaching the Mouthpiece 2


1) Attach the cap to the inhalation air inlet.
2) Attach the mouthpiece to the inhalation top. 1

Attaching the Adult Mask


Attach the mask to the inhalation top.
NOTE:
Do not attach the cap.

Attaching the Child Mask


Attach the mask to the inhalation top.
NOTE:
Do not attach the cap.

Attaching the Nosepiece (Option)


2
1) Attach the cap to the inhalation air inlet.
2) Attach the nosepiece to the inhalation top. 1

11
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 12

HOW TO PREPARE THE NEBULIZER FOR USE


10. Attach the air tube.
1) Attach the air tube to the air connector on the compressor.
2) Attach the other end of the air tube to the air tube connector on the medication cup.

NOTES:
• Make sure that the air tube is securely connected. Otherwise, air may leak from
the air tube during use.
• Take care not to spill the medication when attaching the air tube. Keep the
nebulizer kit upright.
• Use the nebulizer kit holder as a temporary holder for the nebulizer kit.

12
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 13

HOW TO TAKE MEDICATION


1. Hold the nebulizer kit as indicated on the right.

max
5
3

CAUTION:
Do not tilt the nebulizer kit so that it
is at an angle of greater than 45°.
Medication may flow into the
mouth. 45°

x
ma
5
3

2. Press the power switch to the on ( ) position.


The compressor starts and nebulization starts.
NOTE:
Perform functional test, observe whether
aerosol is generated.

To stop nebulization
Press the power switch to the off ( ) position.

WARNING:
Do not block the ventilation slots on the compressor. This could result in the
compressor overheating or malfunctioning.
CAUTION:
Do not block the air filter cover.

13
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 14

HOW TO TAKE MEDICATION


3. Inhale medication as according to your doctor’s and/or health care adviser’s
instructions.

Using the Mouthpiece


Insert the mouthpiece into your mouth and inhale the
medication breathing normally. Exhale as normal through the
mouthpiece.

max
5
3

Using the Adult Mask


Attach the mask so that it covers your nose and mouth and
inhale the medication. Exhale as normal through the mask.
NOTE:
If the amount of aerosol is too great, attach the cap.
max
5
3

Using the Child Mask


Attach the mask so that it covers your nose and mouth and
inhale the medication. Exhale as normal through the mask.
NOTE:
If the amount of aerosol is too great, attach the cap.
max
5
3

Using the Nosepiece (Option)


Inhale the medication through the nosepiece, as shown and
exhale through the mouth.

max
5
3

14
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 15

Ca

HOW TO TAKE MEDICATION


4. Complete your treatment.

1) When you have finished inhaling your medication,


turn the compressor off.
Press the power switch to the off position.

2) Unplug the air tube from the nebulizer kit.


NOTE:
Always hold the air plug when disconnecting the air tube.

3) Check that there is no condensation or moisture remaining in the air tube.

If there is moisture or liquid remaining in the air tube, be sure to follow


the procedure below to remove any moisture from within the air tube.
i) Make sure that the air tube is still connected to the air connector on the
compressor.
ii) Turn on the compressor and pump air through the air tube to expel the moisture.

4) Unplug the air tube from the compressor.


WARNING:
Condensation may occur in the air tube. Do not store the air tube while it
contains condensation or moisture. This could result in infection via
bacteria.

5) Unplug the device from the power outlet.

15
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 16

Care and Maintenance

CLEANING AND DAILY DISINFECTING


Following the cleaning instructions after each use will prevent any remaining
medication in the bottle from drying resulting in the device not nebulizing effectively
and also for preventing infections.
WARNING:
Be sure to wash the parts after use, and dry the parts immediately after
washing them.
1. Remove the nebulizer kit from the air tube and disassemble the nebulizer kit.
Separate the inhalation attachment and the nebulizer kit, and disassemble the nebulizer
kit. (See page 9)
2. Discard any remaining medication in the medication cup.

3
5
max
3. Wash all the parts of the nebulizing parts in warm water and mild detergent, or wash in
a dishwasher.
For instructions on how to disinfect the parts, see the following section.
4. After cleaning and disinfecting the parts, rinse them
thoroughly with clean water and allow the parts to
air dry in a clean environment. 5
3

Disinfecting the Parts of the Nebulizing Parts


Always disinfect the medication cup, mouthpiece and other parts of the nebulizing parts
after the last treatment of the day. To disinfect the parts, use one of the methods
described below:
A. Use a commercially available disinfectant. Follow the instructions provided by the
disinfectant manufacturer.
B. Parts may be boiled for more than 15 minutes (except for mask).
1. Submerge the parts in the cleansing solution for the specified period.
2. Remove the parts and discard the solution.
3. Rinse the parts with clean hot tap water.

16
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 17

MAINTENANCE AND STORAGE


1. Assemble the nebulizing parts.
2. If the compressor is dirty, wipe it clean with a soft
cloth moistened with water or mild detergent.

Mild Detergent
WARNING:
The compressor is not waterproof.
3. If the outside of the air tube is dirty, wipe it clean with
a soft cloth moistened with water or mild detergent.
4. Store the parts in the storage compartment in the
compressor.
NOTE:
Make sure that the parts are clean and dry. Do not place parts in the storage
compartment if they are still wet.
5. Store the unit in a clean location.
NOTE:
Nebulizer kit should be replaced after using for 6 months.

Handling the Vaporiser Head


The vaporiser head is an important part that is used to vaporise medication.
Be sure to observe the following precautions when handling it.

CAUTION:
• Always wash the vaporiser head after each use. Baffle
• Do not use a brush or pin etc., to clean the parts.
• When disinfecting the parts by boiling, be sure to boil
them in plenty of water.
• Do not boil vaporiser head together with other
objects, other than applicable nebulizer accessories.
• Do not use a microwave oven, dish dryer or hair dryer
to dry parts.

Vaporiser head

17
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 18

MAINTENANCE AND STORAGE


To keep your device in the best condition and protect the device from damage follow
these directions:

CAUTION:
• Do not drop the compressor or nebulizer kit, or subject them to strong shocks.
• Do not carry or leave the nebulizer while the medication cup contains
medication.

NOTES:
• Do not store the device in extreme hot or cold temperatures, high humidity or under
direct sunlight.
• Dispose of the device and any used accessories or optional parts according to
applicable local regulations.
• Never clean with benzene, thinner or a flammable chemical.

Benzene

Thinner

18
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 19

HOW TO CHANGE THE AIR FILTER


If the air filter becomes dirty, or has water or medication spilled on it, replace the air
filter with a new one. On average, the air filter should be changed every 60 days even if
it does not appear dirty.

1. Pull the air filter cover off the compressor.

2. Change the air filter.


Remove the old air filter with a sharp object, such as a
toothpick, and insert the new air filter.

3. Put the air filter cover back in place.


NOTES:
• Use only OMRON air filters designed for this unit.
Do not operate without a filter.
• There is no front or back orientation for the air
filters.
• Check that air filters are clean and free of dust before inserting them.
• Do not attempt to wash or clean the filter. If the air filter becomes wet, replace it.
Damp air filters can cause blockages.
• To prevent the mesh of the air filter cover becoming blocked, wash the cover
regularly. Do not boil the air filter cover.
Remove the air filter cover, wash it, and make sure it is dry and reattached
correctly.
• Please contact your OMRON dealer for additional air filters, or if the parts
supplied with the device are damaged or missing.

19
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 20

TROUBLESHOOTING TIPS
Check the following if your unit should fail during operation. You can also refer to the
pages of this manual for complete instructions.
PROBLEM CAUSE REMEDY
Check that the plug is
Nothing happens when Is the power plug inserted in an electrical
the power switch is plugged into an electrical outlet. Unplug then reinsert
pressed. outlet? the plug if necessary.
Is there medication in Fill the medication cup.
medication cup?
Is there too much/little Add the correct amount of
medication in the medication to the medication
medication cup? cup.
Is the vaporiser head Attach the vaporiser head
missing and not correctly.
assembled correctly?
Is the nebulizer kit Assemble the nebulizer kit
assembled correctly? correctly.
Make sure that the nozzle is
Is the nozzle blocked?
No nebulization or low free of blockages.
nebulization rate, when
the power is on. Make sure that the nebulizer
Is the nebulizer kit tilted kit is not tilted at an angle of
at a sharp angle? more than 45 degrees.
Make sure that the air tube is
Is the air tube connected correctly connected to the
correctly? compressor and nebulizer kit.
Is the air tube folded or Make sure that the air tube
damaged? does not contain kinks.
Make sure that the air tube is
Is the air tube blocked? free of blockages.

Is the air filter dirty? Replace the air filter with a


new one.
The device is abnormally Is the air filter cover Attach the air filter cover
loud. attached correctly? correctly.

The device is very hot. Are the ventilation slots Do not block the ventilation
blocked? slots.
NOTES:
• If the suggested remedy does not solve the problem, do not try to repair the device - no
parts of the unit are user serviceable. Return the unit to an authorized OMRON dealer.
• The device should be serviced once a year by an approved OMRON service centre.
20
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 21

SPECIFICATIONS
Model: NE-C29
Type: Compressor Nebulizer
Electrical: 230V, 50Hz model ; 230V, 60Hz model
Power Consumption: 138VA
Operating Temperature/ +10°C to +40°C, 30% to 85% RH
Humidity:
Storage Temperature/ –20°C to +60°C, 10% to 95% RH
Humidity:
Weight: Approx. 2.3kg (compressor only)
Dimensions: Approx. 186mm (w) × 180mm (h) × 216mm (d)
Contents: Compressor, Nebulizer Kit, Air Tube, 5 pcs Replacement
Filters, Mouthpiece, Adult Mask, Child Mask,
Instruction Manual.

NOTES:
• Subject to technical modification without prior notice.
• Please note that specifications may vary with medication type used.
• Do not use the device where it may be exposed to flammable gas.
• This unit fulfils the provisions of the EC directive 93/42/EEC (Medical Device
Directive) and the European Standard EN13544-1:2007, Respiratory therapy
equipment - Part1: Nebulizing systems and their components.
• This OMRON product is manufactured under the strict quality system of OMRON
HEALTHCARE Co., Ltd., Japan.

= Class II =Type B

= Read the instruction manual carefully

21
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 22

SPECIFICATIONS
Technical data for the OMRON Comp Air Pro compressor with the OMRON V.V.T.
(Virtual Valve Technology) Nebulizer kit:

Particle Size: *MMAD **approximately 3µm


Medication Tank Capacity: 7 ml maximum
Appropriate Medication Quantities: 2 ml minimum - 7 ml maximum
Sound: Noise level (at 1 m distance) 60 db
Nebulization Rate: 0.4 ml/min (by weight loss)
Aerosol Output: **0.4 ml (2ml, 1%NaF)
Aerosol Output Rate: **0.06 ml/min (2ml, 1%NaF)

NOTES:
• Performance may vary with drugs such as suspensions or high viscosity. See drug
supplier’s data sheet for further details.
• The data above is referenced by OMRON HEALTHCARE.
• See web site of OMRON HEALTHCARE EUROPE to update technical information.

Result of cascade impactor **measurements for particle size with OMRON Comp Air
Pro compressor and OMRON V.V.T. Nebulizer kit

* MMAD = Mass Median Aerodynamic Diameter


** Independently measured at SolAero Ltd., Canada, Dr. John Dennis, according to EN
13544-1

22
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 23

SPECIFICATIONS
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile (cellular)
telephones, medical devices in use may be susceptible to electromagnetic interference
from other devices. Electromagnetic interference may result in incorrect operation of
the medical device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.

In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with
the aim to prevent unsafe product situations, the EN60601-1-2:2007 standard has been
implemented. This standard defines the levels of immunity to electromagnetic
interferences as well as maximum levels of electromagnetic emissions for medical
devices.

This medical device manufactured by OMRON HEALTHCARE conforms to this


EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate strong
electrical or electromagnetic fields, near the medical device. This may result in
incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct operation of
the device in case the distance is shorter.

Further documentation in accordance with EN60601-1-2:2007 is available at OMRON


HEALTHCARE EUROPE at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.

23
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 24

SPECIFICATIONS
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)

This marking shown on the product or its literature, indicates that it should not be
disposed of, with other household wastes at the end of its working life. To prevent
possible harm to the environment or human health from uncontrolled waste disposal,
please separate this from other types of wastes and recycle it responsibly to promote
the sustainable reuse of material resources.

Household users should contact either the retailer where they purchased this product,
or their local government office, for details of where and how they can take this item
for environmentally safe recycling.

Business users should contact their supplier and check the terms and conditions of the
purchase contract. This product should not be mixed with other commercial wastes for
disposal.

This product does not contain any hazardous substances.

24
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 25

OPTIONAL PARTS AND ACCESSORIES


Nebulizer Kit Set Nosepiece
Model No. C28NEBKIT Model No. C28NP
Nebulizer Kit

Mouthpiece

Air Tube Mouthpiece Filter Cover


Model No. C30TUBE Model No. C28MP Model No. C28FLCVR

Adult Mask Child Mask Air Filters (x 5)


Model No. C28AM Model No. C28CM Model No. C28FL

25
Compressor
Nebulizer System

Model
NE-C29

5
3


 !"
 !"# KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK O
 ! KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK Q
kbJ`OV= ! KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK S

 !
 !"#$%&' KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK T
 !"# KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK NN

 !"
 !"#$ KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK NQ
 !" KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK NR
 !"#$%& KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK NS
 !"#$% KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK NT
 KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK NU
 !"#! KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK NV

1
 !"#
 !"#$%&'()*+,-./01&234567018

 !"#$%&'()*+,-./%&01'23456789:;<=
 !
 !"#$%&'()*+,-./0123456789:4;
 !"#$%&'#()*+,-./0123456789:;<=>?@6ABC
 !"#$%!&'()*+,-./0
 !"#$%&'()*+,%&'-.(/0 !123456789:
 !"#$%&'()*+,-./0123!"4+,5
 !"#$%&'(!")*+,-./
 !"#$%&'()*+,-./012
 !"#$%&'()*+,-"#$./0 !123%45(45678
 !"#$%&
 !"#
 !"#$%&'(
 !"#$%&'()*+,(-./0123456789):;0123<456
 !"#$%&'()!*+,-./0123456789:;<
 !"#$%&'()*+,-./012
 !"#$%&'()*+,-
 !"#$=bj` !"# !"#$%&'()*+,-./01234()
 !"
 !"#$
 !"#$%&'()*+,
 !"#$%&'()*+,-./0123456789:;<=>?@A BCD
 !
 !"#$%&'()*+,-./01
 !"#$%&%'()*+,-./0123456789,:;<=/
 !"#$%&'()*+,-./012

2
 !"#

 !"#$%&'()*+,-./%&01'234567897:;'8<=>?:

 !
 !"#$%&'#(!"#$%&'#)*!"#$%&+,-.'/01

 !"#$%&'()*+,-./01
 !"#$%&'() !"#$*+,-.*/0123"=SM= !"#$%&
 !"
 !"#$%&'()*=QR  !"#$%&'(
l

 !"#$%&'()*+,
 !"#$
 !=ljolk= !"#!$%&'()*+ !,-./0$
 !"#$%&'(=T= !"#
 !"#$%&'()%*+,-./012
 !"#$%&' ()*+,-./01-2345
 !"#$%&'()*+,
 !"#$%&'
 !"#$%&'()"*+,-
 !"#$%&$'()*+,-.$/012
 !"#$%&'()*+,*-
 !"#$%&'()*+,-./
=ljolk=eÉ~äíÜÅ~êÉ= !"#$%&'()*+$,-.
 !"#
 !"#$"%&'()*+,-./012#3+,4562789:/
 !"#$%&'()*+,-.
 !"#$%&'(
 !"#$
 !"#$%&'()*+,- .
 !"#$%&'()*+,-./!01
 !"#$%&'()*+,-./012(345./01 6789:
 !"#$%&
 !"#$%&'()*+,-./0123456789:;<=>?@ABCD
J  !"#$%&'()*
J  !"#
J  !"#$%
J  !"#$%
 !"#$%&'()*+,-./01234567#$89:(%&;*<=>,
 !"#$%&'()*+,-./*01234567"
 !"#$%&'#()*+#,-.#()
 !=HQM `= !"#$%&
l

 !"#$%&'&()*+,-"./0123/45#%&6789:;<=>4
3

C29_01_19 3 5/17/06, 10:15 AM


 !
  !

 !
 !

 !
 !

 !
5
3

  !

 !"
  !"G  !"#$%
 !
 !



  !

 ! 

 !

 !





 !


 

 !"#$%&'( 5
3

OPMs=RMeò 
 !

4
 !



R= !"#$%  !"

5
kbJ`OV= !

 !"#$%&' 7

 !"#$%&'("&)*+,-./0"123456789:

 !"# 11

•   !"#$%&=QRç 
•  !"#$%&'
max
5
3

 !"#$%& 14

 !"#$%&'()*+,-./


 !"#$
 
 !"#!"$%&'()*+,-
 !"#$%&'()*+,-./
 
 !"#$%&'()*+,-./
 !"#$%&'()*+,-./  
 !"#$

 !

6
 !"#$%&'


 !"#$%&'("&)*+,-./012345678$9:;<=>?@

 !"#$%&'()*+,-./0123)4=NQ=

 !"#$%&'() !"=NS= !"#$%&'(

NK  !"#$#%& !"#'(= =

OK  !"!# !$%

PK  !"#$%&'() !*+

2
QK  !"#$%&'(
NF  !"#$%&
OF  !"#$%&'( 1
5
3

7
 !"#$%&'
RK  !"#$%&'(

SK  !"#$%&'()*+,
=
 !"#=O==T=
max
5
3

TK  !"#$%&'()*+

UK  !"#$%&'(
NF  !"#$%&'()*+,-./01
1
 !"
OF  !"#$%&'()*+,-./01
 ! !" max
2
5
3

8
 !"#$%&'
VK  !"# $%&'()*+,-
 !"#$%&'()*+,-./0123./014567#8"39:;<
 !"#$%&
2
 !"#
NF  !"#$%&'()*
1
OF  !"#$%&'()

 !"#
 !"#$%&'

 !"#$%&

 !"#
 !"#$%&'

 !"#$%&

 ! ! 2
NF  !"#$%&'()*
1
OF  !"#$%&'

9
 !"#$%&'
NMK  !"
NF  !"#$%&'( ")*
OF  !"#$%&'()*+! &,-

=====
1


•  !"#$%&'()*+ ,-. !/0%
•  !"#$%&'()*+,-./0)
•  !"#$% &'!"(

10
 !"#
NK  !"#$%&'()

max
5
3


  !"#$%& 45o 
 !"#$%&

x
ma
5
3

OK  !"#$%&'(")=E F=
 !"#$%&'(

 !"#$%&'()*+,-

 !"
 !"#$%&'(")=E= =F=


 !"#$%&'()*+,-./0"#$1234567*

 !"#$%&'

11
 !"#
PK  ! L  !"#$%&'()*

 !"#
 !"#$%&'()*+,-.+$/012
 !"#$%&'

max
5
3

 !"#
 !"#!$%&'()"*+,-./012
 !"#

 !"#$%&'() *+,
max
5
3

 !"#
 !"#!$%&'()"*+,-./012
 !"#

 !"#$%&'() *+,
max
5
3

 ! !
 !"#$%&'()*"#$+,-./

max
5
3

12
 !"#
QK  !"

NF  !"#$%&'()*+,
 !"#$%&'(#)= =

OF  !"#$%&

 !"#$%&' ()

PF  !"#$%&'()*+,-./

 !"#$%&'()*+,-./0123*45678 !%&9
áF  !"#$%&'( !)*+,-
ááF  !"#$%&'()*(+,(#-./01

QF  !"#$%&

 !"#$%&'( )*+%&',-./0123456(7!"89:
 !"#$

RF  !"#$%&'(

13
 !"#$
 !"#$%&'()&'*+,-./0123456789:;<,=>?@ABC
 !"#$%&'()*+

 !"#$%&'()*+#$",-./0

NK  !"#$%&'()*+,-./0123$%&'(+,$%&- !=T


OK  !"#$%& '(

3
5
max
PK  !"#$%&'(&)*+,-./0,12345"#)67&)8
 !"#$%&'()*+,-.$/0

QK  !"#$%&'()*+,-.#/
 !"#$%&'()*+,
5
3

 !"# $%&'()
 !"#$%&'($)*+,-./01234567894:;<=>?@+:;<
 !"#$%&'()*+
^K  !"#$%&'()*+,-.&'()*/01#$%23+
_K  ! !"# !"#$=NR= !"
NK  !"#$%&'()*+",-./012
OK  !"#$%&'()*+
PK  !"#$%&'(!)*+

14
 !"

NK  !"#$

OK  !"#$%&'()*+,-./012

Mild Detergent
 !"#$%&

 !"#$%&'(

PK  !"#$%&'()*+,-./0123456789%&:

QK  !"#$%&'!"()*

 !"#$%&'() !*+,-./01231245'

RK = !"#$%&'()

=S= !"#$%&'(

 !"#$%&'()*+,-$./+01234567


•  !"#$%&'()%*+,-./012
•  !"#$%&'()*+,- .


•  !"#$%&'()*+)*(,-./01234567
•  !"#$%&'&()*+,-"./0123/4
Gasoline

•  !"#$%&'()*+,-./012345
Benzene

Thinner

15
 !"#$%&

 !"#$%&'()*+,-./0)1234(5678 !"#$9:;<=>?
=SM= !"#$%&'()*+,-./0123

NK  !"#$%&'()*

OK  !"#$%
 !"# !" !"#$%&'
 !"#$%&'

PK  !"#$%&'

•  !"#$%!&=ljolk= !"#$%
 !"#$%&'()*
•  !"#$%&'()
•  !"#$%&'()!*+,-./0123
•  !"#$%&'()*+$%&,-./01234'56-.7*+$%&8
 !"#
•  !"#$%&'()*+,-./0#$1%2345678!"#$%29:
 !"#$%&'(#)*+,-./
•  !"#$%&'()*+,-.!/012345)678(9:;<1
ljolk= !

16

C29_01_19 16 4/10/06, 1:37 PM


 !"#$%

 !"#$%&'()*+,-./0123456/789:;&<=>?+@ABC


  
 !"#$%& '()
 !"#$%&  !"#$% &
 !"#$%&'()$*
 !"  !
 !"!#$
 !"#$%&  !"#$%&
 !"#$%&'
 !"#$%&'()
 !
 !"#$%&'(
 !"#$%

 !"#$%&'  !"#$%
 !"#$%&
 !"#$  !"#$%
 !"#$
 !"#$%&'(  !"#$%&'(=QRç 
 !"#$%&'()*
 !"#$%&
 !
 !"#$%&'(  !"#$%
 !"#$  !"#$%
 !"#$%&'  !"#$%&'(
 !"#$%&  !"#$%&'
 !"#$%&'
 !"# 
 !"#$%&
 !"#$%  !"#$%
 !


•  !"#$%&'(&)*+,-./012345-67458(9:;<2"=
 !"#$%&'()=ljolk= !
•  !"#$%&'=ljolk= !"#$%&

17
NE-C29

230V, 50Hz ; 230V, 60Hz


138VA
0.4 ml/min
5 µm (MMD)* EN13544-1
7 ml (cc)
2 - 7 ml (cc)
+10˚C +40˚C, 30% 85% RH
-20˚C +60˚C, 10% 95% RH
2.3kg
186 x 180 x 216 mm
5

*MMD =




• EMC IEC60601-1-2:2001

• 这 OMRON 产品是在 OMRON HEALTHCARE Co., Ltd, 日本,严格的质量体系下生产的。

請仔細閱讀本說明書

18

C29_01_19 18 4/10/06, 12:41 PM


 !"#!
 !"#$
 `OUkb_hfq

  !

  !   !  !
 `PMqr_b  `OUjm  `OUkm  `OU^j  `OU`j

 !ñR  !"
 `OUci  `OUci`so

19
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 26

NOTES
NE-C29-C1_D_M.qxd 11.12.15 4:41 PM Page 27

OMRON HEALTHCARE Co., Ltd.


53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp
THE NETHERLANDS

OMRON HEALTHCARE SINGAPORE PTE LTD.


Asia Pacific HQ 438A Alexandra Road, #05-05/08
Alexandra Technopark, Singapore 119967
OMRON (DALIAN) CO., LTD.
Production facility
Dalian, CHINA

You might also like