Chapter 6 The X Ray Imaging System

This document discusses the key components and functioning of an x-ray imaging system, including the x-ray tube, high voltage generator, operating console, examination table, exposure timers and how they work together to produce x-rays and x-ray images.

Uploaded by

priradiology
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

lOMoARcPSD|35323155

Chapter 6 THE X-RAY Imaging System

Radiologic Technologist (University of Makati)

Studocu is not sponsored or endorsed by any college or university


Downloaded by Pri Radiology ([email protected])
lOMoARcPSD|35323155

CHAPTER 6
THE X-RAY IMAGING SYSTEM

Three Principal Parts of X-ray Imaging System Line Compensator


• X-ray Tube, Operating Console & High • It measures the voltage provided to the x-ray
Voltage Generator imaging system

Radiographic X-ray Tube AUTOTRANSFORMER


• It is attached to an overhead movable crane
assembly Autotransformer
• Location: examination room • It consists of only one winding of wire &
one core
Fluoroscopic X-ray Tube • Function: supplies a precise voltage to the
• It is located under the examination table filament circuit & high-voltage circuit
• It controls the kVp
Examination Table • Location: operating console
• It must be transparent to x-rays as possible • Step up voltage approximately twice the
• Composition: carbon fiber (strong & absorbs input voltage value
little radiation)
Primary Connection
Fluoroscopic Table Tilt • It conducts the input power to the
• Tilt: 90/30 autotransformer
• 90 degrees to the foot side
• 30 degrees to the head side Autotransformer Law
• It states that the voltage receive & provide
High Voltage Generator by the transformer is directly proportional to
• It is housed in an equipment cabinet the number of turns
positioned against a wall • Formula: Vs/Vp = Ns/Np
• It is always close to the x-ray tube
Adjustment of Kilovoltage Peak (kVp)
OPERATING CONSOLE • Major & minor kVp

Operating Console kVp


• Most familiar to the radiologic technologist • It determines the quality of the x-ray beam
• It is used to control the x-ray tube current &
voltage applied to the x-ray tube kVp Meter
• It provides control of line compensation, • It reads voltage (not kVp)
kVp, mA & exposure time • Location: output terminals of the
autotransformer
Radiation Quantity
• The number of x-rays Prereading kVp Meter
• The intensity of the x-ray beam • It allows the voltage to be monitored before
25

• Units: mR or mR/mAs an exposure


Page

Radiation Quality X-ray Tube Current


• The penetrability of the x-ray beam • It is measured in milliamperes (A)
• It is expressed in kVp or HVL (more
precisely)

STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO

Downloaded by Pri Radiology ([email protected])


lOMoARcPSD|35323155

CHAPTER 6
THE X-RAY IMAGING SYSTEM

Filament Temperature • It is connected at the center of the secondary


• It determines the number of electrons winding of the high-voltage step-up
emitted by the filament transformer
• Rationale: ensures electrical safety
Filament Current
• It controls the filament temperature Filament Transformer
• It is measured in amperes (A) • Full Name: Filament Heating Isolation Step-
• Range: 3-6 A down Transformer
• A step down transformer
Filament Circuit • Functions:
• It controls x-ray tube current o Received the voltage from the mA
selector switch
Space Charge Effect o Steps down the voltage to
• A phenomenon of the space charge that approximately to 12 V
makes it difficult for subsequent electrons to o Provides the current to heat the
be emitted by the filament because of filament
electrostatic repulsion • Primary windings:
o Thin copper
Space Charge o Current: 0.5-1 A
• Electron cloud near the filament o Voltage: 150 V
• Secondary windings:
Thermionic Emission o Thick
• The release of electrons from a heated o Current: 5-8 A
filament o Voltage: 12 V

Precision Resistor EXPOSURE TIMERS


• It is used to reduce the voltage to a value
that corresponds to the selected Guard Timer
milliamperage • It terminates an exposure after a prescribed
time (6 s)
Falling Load Generator
• Design in which exposure factors are Timer Circuit
adjusted automatically to the highest mA at • It consists of an electronic device
the shortest exposure time allowed by the o Make & break the high voltage
high-voltage generator across the x-ray tube
• Exposure begins at maximum mA & it drops o Always done on the primary side
as the anode heats ▪ Rationale: lower voltage
• Result: minimum exposure time
Four Types of Timing Circuits
• Synchronous Timer, Electronic Timer, mAs
26
mAs
• The product of x-ray tube current (mA) & Timer & Automatic Exposure Control
Page

exposure time (s)


• A measure of electrostatic charge (C) Synchronous Timer
• A precision device designed to drive a shaft
mA Meter at precisely 60 revolution per second
• It monitors the x-ray tube current • It cannot be used for serial exposures

STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO

Downloaded by Pri Radiology ([email protected])


lOMoARcPSD|35323155

CHAPTER 6
THE X-RAY IMAGING SYSTEM

Electronic Timer High Voltage Transformer


• Most sophisticated, most complicated & • A step-up transformer
most accurate (as small as 1 ms) • Secondary Voltage (kVp) > Primary Voltage
• It allows a wide range of time intervals to be (V)
selected • Secondary Current (mA) < Primary Current
• It is used for rapid serial exposures (A)
• Secondary Windings > Primary Windings
Most exposure timers are electronic & are • Voltage Waveform: sinusoidal
controlled by a microprocessor! • Amplitude: only difference in the primary &
secondary waveform
mAs Timer
• Functions: Turns Ratio
o Monitors the product of mA & • The ratio of the number of secondary
exposure time windings to the number of primary windings
o Terminates exposure when desired • Examples: 500:1 & 1000:1
mAs value is attained • Directly proportional to the voltage
o Provides the highest safe tube • Inversely proportional to the current
current for the shortest exposure for
any mAs selected Voltage Rectification
• Location: secondary side of the high-voltage • It ensures that electrons flow from cathode
transformer to anode only
• Applications:
o Falling-load Rectification
o Capacitor discharge imaging system • The process of converting alternating
current (AC) to direct current (DC)
Automatic Exposure Control (AEC)
• A device that measure the quantity of Rectifier
radiation that reaches the image receptor • An electronic device that allows current
• It automatically terminates the exposure flow in only one direction
when the image receptor has received the
required radiation intensity Diode
• An electronic device that contains two
Solid-state Detectors electrodes
• It is used to check timer accuracy (as short
as 1 ms) Valve Tube
• A vacuum tube (original rectifier)
HIGH VOLTAGE GENERATOR • It replaced by solid-state rectifier
o Composition: silicon
High Voltage Generator
• It increases the output voltage from the
27

Semiconductor
autotransformer to the kVp necessary for x- • Lies between insulators & conductors
Page

ray production
• 2 Types: p-type & n-type
Three Primary Parts
P-type Semiconductor
• High Voltage Transformer, Filament
• Have loosely bound electrons (free to move)
Transformer & Rectifiers
• Have spaces called holes (no electrons)
STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO

Downloaded by Pri Radiology ([email protected])


lOMoARcPSD|35323155

CHAPTER 6
THE X-RAY IMAGING SYSTEM

• Holes: as mobile as electrons Extinction Time


• Ending an exposure
Solid-state p-n Junction
• N-type material placed in contact with p- High Frequency Generator
type crystal • It produces a nearly constant potential
• It conducts electricity in only one direction voltage waveform
• Solid-State Diode: a rectifier • Advantages:
o Much smaller & less costly & more
Electron flow is used when medical imaging efficient
systems are described! o Improves image quality at lower
patient radiation dose
Half-Wave Rectification • It uses inverter circuits
• The voltage is not allowed to swing
negatively during the negative half of its Inverter Circuit
cycle • A high-speed switchers or choppers that
• Diodes: 0, 1 or 2 convert DC into a series of square pulses
• 60 pulses/second
• Disadvantages: Full-wave rectification or high-frequency
o It wastes half the supply of power voltage generation is used in almost all
o It requires twice the exposure time stationary x-ray!

Full-Wave Rectification Capacitor Discharge Generator


• The negative half-cycle corresponding to the • Tube voltage falls during exposure
inverse voltage is reverse • Approximately 1 kV/mAs
• Diodes: 4
• 120 pulses/second Grid-Controlled X-ray Tube
• Advantage: • An automatic lead beam stopper
o Exposure time reduced in half • It stops continues x-ray emission of
capacitor bank
Single-Phase Power • It is designed to be turned on & off very
• It results in a pulsating x-ray beam rapidly
• Disadvantage: • Applications:
o X-ray produced has a value near zero o Portable capacitor discharge imaging
systems
Three-Phase Power o Digital subtraction angiography
• The voltage impressed across the x-ray tube o Digital radiography
is nearly constant o Cineradiography
• 6 pulses/1/60 second • Grid: it refers to an element in the tube that
• Advantage: acts as a switch
28

o Voltage never drops to zero during


exposure Less Voltage Ripple
Page

• Disadvantages: • Greater radiation quantity


o Its size & cost o Higher efficiency of x-ray
production
Initiation Time
• Starting an exposure
STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO

Downloaded by Pri Radiology ([email protected])


lOMoARcPSD|35323155

CHAPTER 6
THE X-RAY IMAGING SYSTEM

• Greater radiation quality


o Fewer low-energy projectile
electrons pass from cathode to anode

CHARACTERISTICS OF HIGH FREQUENCY


X-RAY GENERATORS
Frequency Range Inverter Features
<1 kHz Thyristors
Large silicon-controlled
1-10 kHz
rectifier
Power field effect
10-100 kHz
transistors

Voltage Ripple
• Means of characterizing voltage waveforms

Waveform Ripple Voltage


Single Phase: Varies from zero to
Half-wave 100% maximum
Full-wave 100%
Three-Phase:
6-pulse 14% Never falls below
86% of maximum
value
12-pulse 4% Never falls below
96% of maximum
value

High Frequency < 1% Never falls below


99% of maximum
value

Power Rating (kW)


• Formula: for three-phase & high frequency
o (mA x kVp)/1000
• Formula: for single-phase
o (0.7)(mA x kVp)/1000

Three Main Section of X-ray Imaging System


• X-ray tube, Operating Console & High-
29

Voltage
Page

STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO