9 Quality Control
9 Quality Control
WARNING
Select the Infinia station with the desired QC reports and click [OK].
The QC image folder opens.
Usage
The Logs panel allows the user to view, print, and export reports to HTML
for diagnostic purposes.
Service
Opens Service Reports Folder
Reports
NM Daily QC
9.5 QC Overview
Table 9–1:
Periodically or to
Image QC Periodical NM Calibration & QC Section 9.6.5
diagnose problems
X-ray to NM
Registration Test Section 9.11
Registration
There are two Daily QC procedures, each relevant for a specific scan type,
as follows:
QC Tests Description
Perform the relevant daily quality control procedure in the order specified
above.
The time required to perform these procedures is about one half hour.
WARNING
Failing to perform the relevant daily quality control procedure before starting the
corresponding scans may cause non-identification of serious deterioration in
image quality, resulting in misinterpretation of the acquired images.
Table 9-2 shows the radionuclide sources, source activity, and collimator that
should be used, depending on the QC procedure to be performed.
1 mCi=37 MBq
or
27µCi=1 MBq
Rec.
Source
Count Usage
Activity
Rate
Req.
Source
Collim. HE
NM HE
NM Period. COR
mCi MBq kc/sec. Daily Period.
Daily QC Test
QC QC
QC
Tc99m LEHR or
Flood Source ~ 10 to 20 370-740 <45 9 9 9 9
LEGP
(re-fillable)
Source
Source Collimator Start Position
Placement
a. If a distance of 3.0 m is not possible, the point source can be placed closer, but not less
than 2.5m away from the active detector.
Note The minimal size of the flood source should be 440 mm x 580 mm in order to provide
adequate exposure of the entire FOV and to simplify the positioning of the flood
source between the detectors.
Usage:
9.6.3.1 Overview
The Infinia system is supplied with Default QC parameters, which should not
be modified in most cases. In some cases, it may be necessary to modify
parameters based on the type of isotope or the source used and therefore, the
QC Gantry Mode. In this case, the modes can be modified. These modes are
referred to as “Home Positions”.
The table below shows details of a few of the user home positions for QC
supplied by the factory.
Table 9-4 Predefined QC User Home Positions/Modes
Start Positions
Gantry
Usage
Mode
Motion
QC reports can be reviewed any time from the Logs. For more information
on the Logs, refer to Appendix B – Logs.
QC Procedure Description
Background Rate Performed without a radioisotope in the area. Checks whether there is any
Measurement test radiation in the area that could interfere with image acquisition.
Energy Peak Position Checks whether peaks are in the correct position and that energy
and Resolution test resolution is acceptable.
Uniformity Test This test checks Flood Field Uniformity Stability (not NEMA).
When this pop-up appears, the user has the option of stopping Daily QC and
starting NM Periodic QC in order to diagnose (and possibly fix) the problem.
The pop-up also shows the estimated time for NM Periodic QC to finish. If
ignored, the pop-up remains and NM Daily QC continues in the background.
The popup does not close until either [OK] or [Cancel] are pressed. Any time
[OK] is pressed, the self-healing action begins (even after NM Daily QC is
completed).
z To perform NM Periodic QC, press [OK].
NM Daily QC continues.
To rebuild the uniformity map (for Tc99m only), perform the following:
1. Press [OK].
The Key Parameters screen for Uniformity Maps appears. Verify that the
value is 120,000 kcts for Stop on counts.
Note The Self-healing procedure only builds a Uniformity Map for Tc99m.
2. Press [Apply].
3. Press [Start].
Usage
Note The Uniformity test done during NM Daily QC uses a ROI of 1.55 x 1.55 cm2
for the uniformity calculation. The uniformity value received after this procedure
should be used to monitor the uniformity stability of the system.
To run the Uniformity Test according to NEMA Standards Publication
NU 1-2001, refer to Performance (Acceptance) Tests on page 9-66.
During execution of the procedure, look for instructions in the prompt area
(at the top) of the Daily QC screen.
The NM Daily QC Test consists of two tests which are part of the NM Daily
QC flow:
QC Procedure Description
Setup
For Co57
Source
Go to Gantry
Go to Gantry
Mount the collimators to be used (either
LEHR or LEGP, see Table 9-2).
Go to Gantry
Double-click
Image QC
Background Test
[Apply]
Go to Gantry
Image
Quality Test The NM Daily QC input dialog opens (see
on left side of Figure 9-8).
screen
[Apply]
Go to Gantry
Notes:
QC Report
[Close NM
Daily QC] or To be pressed when NM Daily QC is
[Close completed.
Session]
Detector
Usage Gantry Mode
Configuration
D1
Used for both detectors with Flood Source H QC D1D2
D2
Maps
Acceptable
Detector Status
Results
Note In case of high background level use a source with a rate 10 times higher than the
background level but not higher than 45 kc/s.
1. Review all the test results. Results that appear in black have passed the
acceptance criteria, i.e. are within the limits indicated in the right column
of the results table. Results that appear in red have not passed the
acceptance criteria. The NM Daily QC procedure is successful only if all
of the tests have passed.
2. If any of the of the tests failed, refer to NM Daily QC Test Failure on
page 9-29.
3. The name that appears in the name field at the top of the screen is the
user login name. To change the name (e.g. if a default logon was used),
type a new name in this field. Click [Save] when done.
The test results can be accessed at any time from the Logs panel. Refer to
Reviewing QC & System Setup Reports on page 9-6.
Note If Daily NM QC was performed for each detector separately, a separate report is
created for each detector.
If the NM Daily QC test fails, check which test failed, use the recovery
suggested below and repeat the QC procedure.
If the procedure still fails, review the Logs (located in bottom right area of
QC Report Results). If significant error messages appear, contact qualified
GE Service Personnel for assistance.
QC Test Recovery
If the reason for test failure is not found, perform Periodical NM Calibration
& QC on page 9-32.
9.6.4 HE NM Daily QC
Usage:
z Periodically used to re-calibrate the photomultiplier gain and to verify
the image uniformity.
z Should be performed only if NM Daily QC is close to the failure limit
or when recommended by the Self-healing mechanism during Daily
QC (see Self-healing Procedure on page 9-16).
9.6.5.1 Description
This procedure can be run with a Tc99m source or with a Co57 source. However,
Important
it is strongly recommended to always use the same isotopes (Tc99m or Co57) and
not to switch between the isotopes.
Setup
For Co57
Source
Go to Gantry
Go to Gantry
Double-click
Image QC
Background Test
[Apply]
Go to Gantry
Source
Positioning
The system displays the Source Position
Acq.
Acq. Parameters Dialog.
on left side of
screen
[Apply]
Go to Gantry
[Next] at
bottom of Ignore the results of this report.
screen
Calibration &
Image
The system displays the Calibration &
Quality Test
Image Quality Test Parameters Dialog
on left side of
screen
[Next] at
bottom of The [Start] button is enabled.
screen
[Close NM
Daily QC] or To be pressed when Periodical NM
[Close Calibration is completed.
Session]
Review the QC report that appears at the end of the test. The test report that
appears has the same format as the NM Daily QC Test.
Note The tolerance for the Peak Position is different from NM Daily QC tolerance.
9.6.7.1 Overview
The Hawkeye QC phantom is used to measure the X-ray image quality. The
phantom is a plastic container in the form of a 20 cm diameter cylinder
containing water and a set of plastic test objects. For additional information
on the Hawkeye QC phantom, refer to QC Phantom on page 9-101.
Note X-ray Daily QC verifies slice uniformity only. The X-ray QC Tests under Performance
Tests (see X-ray QC Tests on page 9-98) uses all of the sections of the phantom.
Step Description
XRAY Warm-up Warm up of the X-Ray Tube by acquiring a few slices in low Voltage
(120 KV). Lasts approximately 3 minutes.
Air Table Acquisition of a single full rotation slice, using the nominal X-ray values.
Creation This slice is used to create the Air Table.
Water Used to check the quality of the X-ray image by acquiring the water
Uniformity section of the QC Phantom and analyzing image uniformity.
9.6.7.2 Operation
To access the X-ray Daily QC procedure, double click on the X-ray Daily
QC sub-category in the Image QC folder. For additional information refer to
Accessing the Quality Control Utilities on page 9-8.
Action Messages/Events
2. Press [Apply].
Action Messages/Events
10.Adjust the ROI over the phantom image, so that The Gantry moves to position. The X-ray
the ROI is positioned exactly on the phantom confirmation dialog for Water Uniformity
upper edge that points toward the gantry, as appears (the pop-up will disappear after five
shown in Figure 9-12 and press [Next]. minutes if not responded to).
ROI
Position on Phantom
Color
Green Top
Blue Bottom
Red In center
Attach to end
of Pallet
Position
rectangular
support bracket
over this
Alignment
Rectangular Circumferential
Support Bracket Top Alignment Line Line
To Gantry
Lateral
Alignment
Table Line
ROI positioned
exactly on the edge
of the phantom
Plastic Wall
Water
The X-ray Daily QC report appears (see Figure 9-14). Review the report.
Results that are outside the acceptance criteria are displayed in red.
Usage
This utility allows the operator to check if the system is performing according
to imaging specifications. Data is analyzed for energy shift, energy resolution
and spatial resolution.
During execution of the procedure, look for instructions in the prompt area
(at the top) of the CoDe Daily QC screen.
If a point source is used, the source must be aligned to the center of the
detector.
If a line source is used, the source must be aligned to the center of the detector
situated on the Y axis.
Setup
Double-click
Image QC
[Close CoDe
Daily QC] or To be pressed when CoDe Daily QC is
[Close completed.
Session]
Usage:
This test is used to check the Center of Rotation correction of the gantry and
the detectors.
The purpose of the COR Test is to check the consistency of the COR
calibration parameters measured during the calibration phase and is provided
to the customer to test the stability of the COR in both H and L modes.
During the test, a 360° Tomo acquisition is performed with COR correction
applied. The acquisition radius (or Lateral) should be in the middle of the
radius (lateral) range. The Sinus fit is implemented in order to exclude the
point source position effect on resulting data.
The test is performed once for H-mode and once for L-mode, with Collimators
mounted on the detectors.
Go to Gantry
COR Test
Acquisition The Acquisition Key Parameters screen
(left side of appears (see Figure 9-19).
screen)
[Apply]
Note The Source Positioning Acquisition stage can be skipped if the operator is
experienced with the appropriate positioning of the source for achieving a
favorable calibration.
Go to Gantry
Source
Positioning
Acq. under
Verify that the correct parameters are
COR Test
selected.
L-mode (left
side of
screen)
COR Test
Acquisition
under COR The Acquisition Key Parameters screen
Test L-mode appears (see Figure 9-19).
(left side of
screen)
[Apply]
COR Test
The C.O.R. Test Processing screen
Processing
appears. Select the appropriate image. The
(left side of
default is the latest acquired image.
screen)
The graphs of x and y deviations are displayed on the Graphs tab (see
Figure 9-21 and Figure 9-23). The yellow lines represent a visual reference
range. The graphs of Detector 1 appear in blue while the graphs of Detector
2 appear in red. The graphs show COR X/Y Offset Residual for each
projection where Cor X/Y Offset Residual reflects the inaccuracy of the
applied COR correction. It represents imperfection of the measured data
caused by the following factors:
z Acquisition Statistics–the weaker the source and the shorter the
acquisition time, the graphs are noisier;
z Gantry Mechanics– changes in detector sagging during rotation and
collimator attachment instability
The horizontal yellow lines represent maximal allowed spikes and jumps
caused by deficiency in acquisition statistics or/and gantry mechanics. The
vertical axis has a scale multiplied by 0.1 mm, so the limits are +/- 3mm in
X and +/- 1mm in Y.
The numbers in the table are within the acceptance limits. All numbers should
appear in black (not red). See Figure 9-23.
The following results are acceptable for both H-mode and L-mode:
The tests described in this section are performed according to NEMA Standards
Publication NU 1-2001.
The following radioactive sources are required for various Performance Tests.
1 mCi=37 MBq
or
27µCi=1 MBq
9.8.1.2 Phantoms
z Two Standard Line Slit Phantoms, one in X and one in Y direction
(see Figure 9-26) as described below.
The Infinia Line Slit X and Y Phantom Kit consists of two cases:
Case 1 – PAS000153 (see Figure 9-24) contains:
X Phantom
Foam
Insert
Y Phantom
Foam
Insert
30mm
CFOV
UFOV
Middle slit
over center
of UFOV
Details of the Low Energy QC Performance Tests and the acceptable test
results are listed in Table 9-11.
The test are done according to the NEMA Standard NU1-2001, for UTEST
only in case the frame size is 512.
Table 9-11: Low Energy Tests and Acceptable Test Results (Continued)
9.8.3.1 Description
This test checks the response of the camera to a uniform flux of radiation.
Uniformity is characterized by the count density variation caused by a
uniform input gamma flux over a defined large area of the detector.
Note The results will be according to NEMA NU1-2001 only for frame size 512 after
applying all of the corrections to the image.
< 45 ± 5 kc/s
H-WB LEGP or LEHR Centered between the Detectors
per detector
To access the Uniformity Test, double click on the Performance folder. For
additional information refer to Accessing the Quality Control Utilities on
page 9-8.
Action Messages/Events
Reached on:
Zoom 1 1
Table 9–15: Uniformity Test Acceptance Criteria For Uniformity Corrected Images
3/8" 1"
Parameter Name Acceptance Acceptance
Criteria Criteria
9.8.4.1 Description
This test checks the system's ability to accurately resolve spatially separated
radioactive sources.
Step Description
Source Activity
Radioactive Source Collimator
mCi MBq
9.8.4.3 Phantoms
z Standard Line Slit Phantom in X-direction
z Standard Line Slit Phantom in Y-direction
Ceiling
Distance
Angle
Dete 2
ctor ctor
1 Dete
Separate procedures are described below for performing the test in L-Cardiac
mode and Maps mode.
Action Messages/Events
5. Double click on the Resolution and Linearity Opens First Acquisition Key Parameters (see
Test. Figure 9-30).
14.Press [Back]
Action Messages/Events
18.Press [Apply]
19.Press [Start]
27.Select Resolution Test on the left side of This opens the Resolution panel.
the screen.
Action Messages/Events
Maps Mode
Action Messages/Events
6. Double click on the Resolution and Linearity Opens First Acquisition Key Parameters (see
Test. Figure 9-30).
Action Messages/Events
20.Select Resolution Test on the left side of This opens the Resolution panel.
the screen button.
Zoom 1
All other
0
parameters
3/8" 1"
Parameter Name Acceptance Acceptance
Criteria Criteria
3/8" 1"
Parameter Name Acceptance Acceptance
Criteria Criteria
9.8.5.1 Description
This test checks the system ability to distinguish between photons of different
energies.
Source Activity
Radioactive Source Collimator
mCi MBq
Performance
Tests
Tc99m Acquisition
Energy Test
Etest Report appears. Compare report to
(left side of
Section 9.8.5.5
screen)
Navigation:
Navigation:
This test refers to the phantom with six segments of rods, with diameter of
4.8, 6.4, 7.9, 11.1 and 12.7 mm, and six spheres with diameters of 9.5, 12.7,
15.9, 19.1, 25.4 and 31.8 mm; for example, Data Spectrum deluxe SPECT
phantom # 043-750.
Uniformity
Contrast
Resolution
9.9.3.1 Prerequisites
A high statistics sensitivity map acquired for both detectors with LEHR
Collimators. The statistics suggested for the maps is at least 120 Mc per Head.
The map should be acquired either with a Co57 flood source or Tc99m refillable
flood source with the “right” amount of water inside, so that neither hot nor
cold center is observed.
9.9.3.2 Preparations
1. Fill the Jaszczak phantom with distilled water, leaving a volume of about
10 ml of air.
2. With the table retracted to maximum OUT position, place the phantom
holder on the table frame and secure with a single screw.
3. Put the phantom in the holder with the off-centered filling screw facing
up.
Then make the following adjustments with reference to the main
symmetry axis:
a. Set the pitch angle of the phantom to 0o, using a small spirit level on
the phantom. The adjustment should be made with the screw in the
holder below the securing screw.
b. Set the yaw angle of the phantom to 0o. This adjustment is best
performed by setting the Gantry to 90o in H-Shape Mode, and
position the Heads close to the phantom (RADIAL IN). Make sure
that the sides of the phantom are parallel to the Collimators.
You may put a ruler between the phantom and the Heads to improve
the accuracy of our estimation of parallel lines.
The change in the phantom yaw angle is made with the two screws
on the sides of the security screw.
c. Check again the pitch angle to ensure that the yaw adjustment did
not affect it.Readjust if necessary.
4. Set the minimum possible rotation radius (RADIAL IN) for both
detectors, as follows:
a. Start with Gantry set to H-WB Mode, rotation = 90o, and the two
detectors positioned a few centimeters from the phantom.
Move Detector 1 IN until it is a few millimeters away from the
phantom. Rotate the Gantry slowly, ensuring that the Collimator
never touches the phantom. If necessary, change the Up/Down
position of the Table. Note the location of the critical point and
adjust the clearance between the collimator and the phantom with the
holder to 1 - 2 mm.
b. Measure the distance between Detector 1 and the phantom at two
rotation angles: 0o and 180o. The two distances should be equal. If
not, change the Table’s height.
c. After Detector 1 is properly positioned, adjust the position of
Detector 2.
The quickest way is to set Detector 2 at the critical point and then
move it IN until the Collimator is 1 - 2 mm away from the phantom
and its holder.
6. The phantom is now ready for filling with Tc99m. To calculate the activity
needed, bear in mind that, due to the phantom absorption, the output is
about 40% of the inserted Tc99m activity.
The suggested rate at the beginning of acquisition is about 30 kc/s (15 kc
per detector).
15 kc/s 1 2250
A = ---------------------- * --- = -------- [mCi]
S*1/60 (kc/s)/mCi 0.4 S
9.9.3.3 Acquisition
1. Open the To Do List.
2. Select the [New Study...] button to schedule a new Tomo study (see
Figure 9-33).
3. Type details of test (for assistance, refer to Figure 9-33. Press [Select
Study Protocol...]. Select Tomo protocol from the Factory\Basic
folder.
4. In the opened Tomo Key Parameters panel, select/verify the following
parameters:
Table 9–21: Tomo Key Parameters
Rotation: Total
360 (Arc per Detector: 180)
Angular Range:
Zoom: 2
frame)
Isotope: Tc99m
Collimator LEHR
9.9.3.4 Processing
1. Select the Jaszczak Planar Study from the Xeleris Data management
panel.
2. Select General SPECT protocol residing within General SPECT
Applications container on the Xeleris Data Management panel and press
[Start].
3. Select the SPECT options.
4. Select the correction reconstruction filter as shown below and click
[Proceed].
5. The expected result after the filter changes is the following (compare to
the previous results).
There are five test procedures performed in the X-ray QC dialog. It is possible
to select all of the procedures to be performed or just some of them.
z Uniformity
z Low Contrast
z Density
z High Contrast
z Slice Thickness
The table below shows the various X-ray QC tests and when they should be
performed.
After
Replacement of
After System Image Quality
During PM Source or
Installation Deterioration
Detector
Assembly
Uniformity 9 9 9 9
Density Scale 9 – 9 9
Slice Thickness 9 – 9 9
9.10.1.1 Uniformity
This test acquires scans of the water section in the QC phantom. The mean
CT number and the pixel noise (measured by the standard deviation of the
CT numbers in a ROI) are measured on the images. This is the most sensitive
test of system performance, since in a well-calibrated system, the mean CT
number of water should be uniform and the noise should be uniform random
noise inversely proportional to the square root of the number of X-ray photons
detected. Any loss of Tube output or Detector efficiency will be detected as
an increase in noise. A change in the Tube High Voltage will change the CT
number of water. Poor calibration of the Detector will cause ring artifacts,
which increase the measured image noise.
The detectability of small low contrast objects depends both on the noise and
the resolution of the system. It is assessed by imaging a pattern of holes in a
thin sheet of plastic.
9.10.1.3 Density
This test measures the densities of different materials such as water, air and
perspex. It can be used to confirm that the effective energy of X-ray scanning
is unchanged and is therefore important in quantitative uses of the CT images.
Additional Usage:
Perform this test after measuring the water CT levels in the Noise and
Uniformity Test.
This test measures the densities of different materials such as water, air and
perspex. It can be used to confirm that the effective energy of X-ray scanning
is unchanged and is therefore important in quantitative uses of the CT images.
It should be performed after measuring the water CT levels in the Noise and
Uniformity Test.
During this test five different groups of bars are scanned. The bar distance
are different from group to group. The
The image’s slice thickness is measured using four inclined hole ramps.
9.10.2 QC Phantom
A phantom is supplied with the Hawkeye option for the purpose of measuring
the X-ray image quality. The phantom is a plastic container in the form of a
20-cm diameter cylinder containing water and a set of plastic test objects.
Distance from
Section Function
Alignment Section
Rectangular
Support Bracket
Water
Section
Section1 30 mm 30 mm
Slice Alignment
Slice Thickness
Resolution
Section 2
Low Contrast
Resolution
Section 3
Water
The CT scale is measured using plastic and water in the phantom and air
outside the phantom.
The Slice Sensitivity profile is measured using the Slice Width gauge,
consisting of a set of four inclined rows of holes drilled at 63.4 degrees to the
horizontal.
Line Pairs/cm
2.0
Center Slice
Marker
2.5
3.0
The ability of the Hawkeye to detect small objects with small differences in
density between the surrounding tissue depends on both spatial resolution
and the noise of the system. A direct measurement of the Low Contrast
Resolution can be made with the Low Contrast section of the Hawkeye QC
phantom.
The Low Contrast section contains a thin plastic plate with an image density
2.5% greater than water when imaged with a 10 mm slice thickness. The plate
has several holes of different diameter which are filled with water. The hole
location and sizes are shown in Figure 9-37.
12 6 4 15
2 3
8 5
The phantom holder is placed on the stretcher (without the mattress) with the
small tips attached to the end of the stretcher. The phantom is placed on the
holder with the rectangular support bracket inserted over the right-angled
protrusion on the holder, facing out of the Gantry and the small alignment
line at the top. See Figure 9-38. Make sure that the phantom is not tilted or
skewed. After placing the phantom on the holder, make sure that the tips at
the edge of the holder come in contact with the end of the stretcher.
Attach to
End of Stretcher
To Gantry
Circumferential
Alignment
Line
Lateral
Alignment
Line
Rectangular
Support Bracket
Place on Holder
Table
Figure 9-39 QC Phantom Placed on Stretcher for Imaging (Rectangular Bracket Faces Out of Gantry)
X-ray Image
Quality Test
3
2
4
5
7
6
8
Acquire First
5 Range 5000 mm
Use Study
7 Select Study
8 Select Series
When selecting a slice, it is possible that there are more than three slices in the
Important
scan (if it was not obtained as a QC scan). In this case choose the appropriate
slice for every test before pressing [Apply].
9.10.2.7 Prerequisites
Before running the X-ray QC tests, verify that X-ray Daily QC (see X-ray
Daily QC (Hawkeye Option) on page 9-39) was performed at some time
during the day.
Usage:
Instructions:
Action Messages/Events
1. Select the Start Position and Range. Use arrows to increase or decrease number
4. Press <GO/PAUSE> on the RCU to begin. The Start X-ray Confirmation Dialog appears.
6. Position the locators exactly at the top and The Start X-ray Confirmation Dialog appears
bottom of the phantom. Click [Next]. again.
Usage:
Instructions:
Action Messages/Events
1. Select Study
2. Select Series
3. Select Slice
1. The first test that is selected in the X-ray QC Test dialog (see
Figure 9-40) is the first test that is run after [Apply] is pressed.
2. A test is run only if it is “checked” in the X-ray QC Test dialog.
3. Tests are run in sequential order and automatically continue to the next
selected test.
4. Once all of the selected tests are completed, a series of reports appear.
5. The individual reports for each test appear in the same sequence that the
tests were run.
6. To zoom in on an image, click on the image with the right mouse button.
7. To move an image, click on the image with the left mouse button.
The procedure described below is the flow if all X-ray tests are checked.
Action Messages/Events
6. Press <Start> on the RCU to begin. The X-ray Confirmation Dialog appears.
Uniformity Test
When the test starts, slice 1 of the phantom
appears with six round ROIs (see
Figure 9-42).
11.Press [Next].
Action Messages/Events
13.Press [Next]
Density Test
Slice 3 of the phantom is displayed on the
screen with three round ROIs of different
colors (see Figure 9-44).
15.Press [Next].
17.Press [Next].
Action Messages/Events
Green Top
Blue Bottom
Red In center
9.11.1 Overview
This procedure is used to measure the X-ray (CT) and Emission (NM)
images registration.
The registration test uses the Hawkeye alignment phantom, which is built
from low attenuation material with holes for placing 6 syringes filled with
radiation sources (Tc99m or other – according to the collimator used).
These sources serve as landmarks for calculating the distance between the
CT and NM images.
The processing and calculations for this test are done using a dedicated
protocol on the Xeleris workstation.
The processing protocol measures the distance between the CT and NM
images in the X, Y and Z-axes, and displays the Mean (average) of the
absolute values measured on the landmarks together with the acceptance
limits.
These values are the main outcome of this test, as they reflect the current
system’s performance, i.e.: if these values are within the acceptance limits,
then the CT – NM registration is acceptable. Otherwise, the system must be
calibrated using the X-ray to NM calibration procedure.
For a definition of the system’s axis and angles, see Figure 9-48.
In order to view and archive the results of the test, the results of the
processing must be fetched back to the Infinia acquisition station.
The test must be executed separately for every collimator set that is used on
the system (similar to C.O.R. test).
9.11.2 Pre-requisites
z If X-ray QC has not been performed within the last 24 hours, perform
X-ray QC (see X-ray QC Tests in Section 9.10).
z If C.O.R. QC Test has not been performed in the last 30 days, perform
the C.O.R. QC test (see C.O.R. QC Test in Section 9.7).
z Verify that the table is properly calibrated in relation to stretcher
support (See Section 6.2 – Table Height Calibration on page 6-21
of this manual
z Verify that stretcher motion when on the support and loaded with
weight is smooth and straight. See Installing the Table on page 4-67
of the Infinia Installation Manual
z Verify that the Infinia acquisition station and the Xeleris workstations
are mutually configured as remote DICOM stations
(see Section 4.12.3 of this manual
z Verify that the protocol Hawkeye registration QC appears under
Miscellaneous in the Xeleris Workstation.
3. If the Hawkeye Registration is OFF, select Yes to turn the protocol ON.
A confirmation message appears.
Note Once this procedure is completed, the Hawkeye Registration QC protocol is added
to the Miscellaneous group on the Xeleris workstation.
Six 10 ml syringes with needles (supplied with the kit) see Figure 9-52. If
these syringes are not available, any similar 10ml syringe may be used,
provided that the inner radius of the syringes is at least 15mm in diameter,
and that they fit securely in their grooves in the phantom.
1. Fill each syringe with 10 ml of liquid. Verify that all the syringes contain
the same volume and activity.
• When filling the syringe with activity it is important to avoid bubbles in the
Important
syringe – see Figure 9-52 and Figure 9-53.
• When filling the syringes, avoid differences in activity between syringes. In
order to do so, fill all the syringes from the same pool, and verify that all the
syringes contain the same volume (and no less than 6ml in each syringe).
2. Cover the needle with the plastic cover so that the liquid does not leak
out from the syringe.
3. Insert the syringes into the numbered holes of the phantom (three
syringes on each side). See Figure 9-54 and Figure 9-55.
Important
Locate the phantom in linear position 70-90 (according to stretcher
side ruler).
Action Messages/Events
1. Determining ‘time per projection’ - for All collimators Note: the count rate registered in the detectors
except Coincidence (SEPTA, FILTER, …) (an estimation of the average is sufficient,
a. Position the gantry in H mode, rotation angle 90. high accuracy is not required)
Move the stretcher in, so that the phantom is located See Figure 9-56.
between the detectors. Use Table 9-28 to determine the time per
b. Apply the Count Rate Measurement. projection needed during the acquisition
3. Determining ‘time per projection’ - for Coincidence Make sure that the Coincidence rate is not
collimators (SEPTA, FILTER, …): grater than 14Kcts/sec. and the singles rate are
a. Position the gantry in H mode, rotation angle 90. in the green range and not greater than
Move the stretcher in, so that the phantom is located 500Kcts/sec. (the average of the two singles
between the detectors. rate- high accuracy is not required). If it is
greater than the mentioned rates then the
b. Schedule a CoDe scan and apply it.
activity in the syringes should be decreased.
The Emission time should be 4 Min.
3. Move to Registration Test Acquisition The Registration test (Tomo) acquisition page
opens.
6. Press <Set> on the RCU Gantry moves to start position and then asks to
press <Go>
7. Press <Go> on the RCU The X-ray confirmation pop-up appears (on
host).
8. Press [Start] to confirm. This part (scout acquisition) Data is acquired. The acquisition starts with a
should take 2-3 minutes. Scout scan, and when completed, the results of
the scout scan are displayed.
9. Adjust the level of the window in order to see Note: It is only possible to set a Full range
registration phantom and syringes. Move locators such FOV (40 slices) – the locators do not enable
that the syringes are located symmetrically between the setting a different FOV and the tolerance
locators (see Figure 9-57) and press [Accept]. provided is only relevant to the location of the
Important: Avoid positioning the locators near the FOV.
syringes and ensure that there are no The Hawkeye sequence continues, NM part first
objects between the locators (approximately 5-10 minutes, depending on the
selected time/projection), followed by CT Slices
part (approximately 10 minutes).
11.Select the relevant study to process. Note: Studies are named according to the date
and time of creation. The study most recently
acquired should be automatically selected in the
list.
If you wish to process an old study, check the
check box ‘Show all Studies’, and select the
required study from the drop-down list.
12.Select the Xeleris processing station, where the study The application sends the selected study to the
is to be processed and verify that the workstation selected processing station.
configuration was completed. Press [Process]. Note: The [Fetch Results] button becomes
enabled after successfully sending the study to
the processing station. But fetching the results
should be done after processing the data on the
processing station.
13.Verify that the data has been successfully sent to the The study is marked as successfully sent.
selected processing station: 'data management’ Æ
transfer log tab
14.Go to the Xeleris workstation. In the The protocol is launched. The result screen shows the
patient selector screen, select the following:
registration study. Find the Hawkeye A zoomed image of one of the syringes: an NM image (top
Registration QC application that appears right), a CT image (top left), and a fused image, which shows
under 'Miscellaneous', and double click on the NM image contour super-imposed over the CT image
it to launch. (bottom left).
Note: For Xeleris version 2.* the protocol An MIP image of the syringes (bottom right), with numbers
should customized to run with Fourier series next to the identified syringes.
and not IRNC in CoDe calibration and test.
The CT - NM registration mean absolute values [mm] (left
side of screen).
See Figure 9-59.
15.Compare the MIP image to Figure 9-59 The MIP image should contain 6 syringes, labeled with
numbers, exactly as seen in. Figure 9-59:
2 horizontal lines (syringes 2 & 5)
2 vertical lines (syringes 3 & 6)
2 'points' (syringes 1 & 4)
16.Review the results. Check whether any of Pressing on each button displays both an NM and a CT
the values exceed the acceptance limits. image of a single syringe.
It is possible to view zoomed images of If any of the mean absolute values exceed the acceptance
the 6 syringes by clicking on their number limits, they should be Bold (this implies that the system must
on the left side of the screen. be calibrated).
Notes: The CT slice thickness is 1cm. This leads to the
following effects: The position of the syringe in the axial
direction might appear slightly shifted compared to the
position in the NM image. This visual effect is amplified by the
zoom x10 of the display.
The shape of syringes # 1, 2, 4 & 5 might not appear to be
circular
Both of these effects are normal and do not reflect a problem
in the QC procedure.
Return to the Infinia application and click The Infinia application fetches the results and displays the
the [Fetch results] button. processed information. See Figure 9-60.
Table displays the mean of the absolute CT to NM
differences, with the acceptance limits. The values in this
table are identical to the values in the table displayed on the
Xeleris workstation.
Interpreting the results: Note:
If all values in the table (absolute values) The acceptance limits for the CT to NM average absolute
are within the acceptance limit, then the difference (1) are 3mm for X and Y, and 5mm for Z. If any of
system’s performance is within the values exceed their limits, it should be marked red.
specifications. However, it is The maximum correction value allowed (limits of cumulative
recommended to accept the values. calibration values) (2.3) are: for X +/-8mm, for Y +/-5mm, for Z
This may improve the camera’s image +/-10mm, for Rotation +/-1 degree. If any of the values
quality. exceed their limits, it should be marked
If one or more values in the 1st table red.
exceed the acceptance limit (marked
red), then the system’s performance is
below specifications. CT to NM
calibration must be executed.
Press [Next] to display the final report. The application displays the final report.