YSIO 3DV Adjust Axrf-000.842.01.01.02
YSIO 3DV Adjust Axrf-000.842.01.01.02
h
a
F
s
n
m
ie
S
RAD/FLUORO
AX
Adjustment
System
07042034
07042018
© Siemens 2009
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise autho-
rized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with
or authorized by Siemens or one of its affiliates are directed to contact one of the local
offices of Siemens or one of its affiliates before attempting installation or service proce-
dures.
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Required Aids and Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Tolerance Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Torque Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
System Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Polydoros Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Service Radiation Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Product-specific Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Prerequisites. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Additionally Required Documents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
All tools, measuring equipment and aids, with the exception of the "standard installation
tool kit", are listed and specified in the Service Tools Catalogue.
Tolerance Data 0
Limit value for over 3 mm over 6 mm over 30 over 120 over 400 over 1000 over 2000
the nominal to 6 mm to 30 mm mm to 120 mm to 400 mm to mm to mm to
range mm mm 1000 mm 2000 mm 4000 mm
Adm. toler- ± 0,5 mm ± 1 mm ± 1,5 mm ± 2,5 mm ± 4 mm ± 6 mm ± 8 mm
ance
Torque Values 0
Notes 1.2
When performing the work steps and checks, the general safety information for medical
products must be observed. (TD00-000.860.01 / General Safety Notes)
WARNING A Gigalink (Class kx3A product) with a laser diode is used for
image transfer between the FD --- PIX Box --- image system.
The laser beam is dangerous for the human eye.
¹ Do not look into the laser beam. The laser beam is not
visible.
SS_Switch, D600.S1
See (Fig. 1 / p. 9)
ON = Position 2 Normal mode, the generator can be switched on via the system
switch-on circuit.
OFF = Position 1 The generator is switched off and cannot be switched on via the sys-
tem.
Depending on the version, the illustrations and drawings can show slight differences to the
shipped unit and should therefore be viewed only as examples.
Prerequisites 1.4
Abbreviations 1.5
If there are changes to the ceiling stand, e.g. caused by a replacement or removal of differ-
ent components, a check of the system and possibly a new adjustment is required.
Check
1. Position the ceiling stand over the table so that when the light localizer is switched on,
the front edge of the table is lined up with the center of the light cross.
2. Move the stand in the longitudinal direction over the complete length of the table.
3. If the table is aligned, it should move parallel to the center of the light cross. The toler-
ance over the entire length of the table is ± 2 mm.
Adjustment
1. Turn the table at one end until parallelism to the central beam is established over the
entire movement range.
Check
• Check the difference between the light and radiation fields. The position of the light field
can be readjusted at the collimator.
Adjustment
The adjustment can be found in the CB-Doc under Collimator > Adjustment
(AX74-020.842.01).
Check
1. Position the 3D in the center of the room, between the two transverse and longitudinal
rails.
2. Remove the cover panels on the rear of the support arm (Fig. 2 / p. 15) that cover the
column.
Fig. 3: Checking wobbling in the transverse Fig. 4: Checking wobbling in the longitudinal
direction of the support arm direction of the support arm
¹ If the result is not within the tolerance, wobbling of the support arm must be
adjusted.
Adjustment
Adjustment is made using the wobbling screws. The wobbling screws are comprised of a
securing screw (Allen screw) (1/Fig. 5 / p. 16) and an adjustment screw (hex screw)
(2/Fig. 5 / p. 16).
1. Loosen the two Allen screws (1/Fig. 6 / p. 16) and (1/Fig. 7 / p. 16) (approx. 4 turns).
Loosen the two hex screws (1/Fig. 6 / p. 16) and (1/Fig. 7 / p. 16) (approx. 2 turns).
Fig. 6: Wobbling screws Fully Automated Fig. 7: Wobbling screws Fully Synchronized
System System
2. Correction
In the longitudinal direction of the support arm:
- Loosen the two Allen screws (2/Fig. 6 / p. 16) or (2/Fig. 7 / p. 16) and
(3/Fig. 6 / p. 16) or (3/Fig. 7 / p. 16) (approx. 2 turns).
- Make a correction in the longitudinal direction with the hex screw (2/Fig. 6 / p. 16) or
(2/Fig. 7 / p. 16). With it, the support arm/tube assembly can be lowered or raised.
- Hand-tighten the Allen screws (2/Fig. 6 / p. 16) or (2/Fig. 7 / p. 16) and
(3/Fig. 6 / p. 16) or (3/Fig. 7 / p. 16) and check alignment in the horizontal, trans-
verse as well as in the longitudinal directions.
In the transverse direction of the support arm to the right or left:
- Loosen the two Allen screws (3/Fig. 6 / p. 16) and (3/Fig. 7 / p. 16) (approx. 2 turns).
- Make a correction in the transverse direction with the hex screw (3/Fig. 6 / p. 16) or
(3/Fig. 7 / p. 16). With it, the support arm/tube assembly can be tilted to the left or
right.
- Hand-tighten the Allen screws (3/Fig. 6 / p. 16) and (3/Fig. 7 / p. 16) and check align-
ment in the horizontal, transverse as well as in the longitudinal directions.
3. If the result of the check is okay, hand-tighten the two hex screws (1/Fig. 6 / p. 16) and
(1/Fig. 7 / p. 16).
4. Tighten all Allen screws to 10 Nm and use Loctite 243.
5. Reinstall the cover panels.
NOTE After pivoting the tube unit support arm, alignment of the
tube unit must be checked in the transverse direction.
Check
1. Position the 3D in the center of the room, between the two transverse and longitudinal
rails.
2. The check must be performed for the -90°, 0°, +90° and +180° support arm positions.
6. The measured drift in the movement range may not exceed +/-10 mm.
¹ If the result is not within the tolerance, tube unit drift must be adjusted in the trans-
verse direction.
Adjustment
4. Loosen the two screws on the support flange half-cowls (Fig. 13 / p. 20).
NOTE
After adjusting the tube unit in the transverse direction, the
RVA tube assembly drift (rotation around the vertical axis)
• must be checked for the Fully Synchronized system and,
if needed, must be mechanically adjusted.
• must be adjusted for the Fully Automated system in the
SSW system.
Check
on the Bucky wall stand
1. Align the tube unit to the BWS and move to the min. SID. Switch on the light localizer.
2. Mark the light field on the BWS (collimator centering cross).
3. Move the tube unit to the max. SID.
4. The measured drift to the left or right in the working range may not exceed +/-3 mm.
¹ If the drift is not within the specified tolerance, the drift must be readjusted.
Adjustment
With the Ysio Fully Automated Systems, perform this adjustment in the SSW in the menu
Tune Up > Stand > Home Position Adjustment, all axles per the Online Help.
Adjustment of the vertical tube rotation must be performed mechanically for the Ysio Fully
Synchronized System:
1. Remove the cover panels on the rear of the support arm (Fig. 15 / p. 21) that cover the
column.
2. To adjust the stop disk for the vertical axis, loosen the (Fig. 16 / p. 22) securing screw
(1/Fig. 17 / p. 22).
3. Release the brake by pressing the appropriate button for vertical tube rotation on the
TUI.
The brake must be released during the adjustment.
4. For the adjustment, turn the eccentric bolt (2/Fig. 17 / p. 22).
5. Retighten the securing screw (1/Fig. 17 / p. 22).
6. Recheck the drift.
7. Reinstall the cover panels.
Check
on the Bucky wall stand
1. Align the tube unit to the BWS and move to the min. SID. Switch on the light localizer.
2. Mark the light field on the BWS (collimator centering cross).
3. Move the tube unit to the max. SID.
4. The measured drift up or down in the working range may not exceed +/-3 mm.
on the table
1. Align the tube unit to the table and move to the min. SID. Switch on the light localizer.
2. Mark the light field on the table (collimator centering cross).
3. Move the tube unit to the max. SID.
4. The measured drift to the left or right in the working range may not exceed +/-3 mm.
¹ If the drift is not within the specified tolerance, the drift must be readjusted.
Adjustment
With the Ysio Fully Automated Systems, perform this adjustment in the SSW in the menu
Tune Up > Stand > Home Position Adjustment, all axles per the Online Help.
Adjustment of the horizontal tube rotation must be performed mechanically for the Ysio
Fully Synchronized System:
1. Remove the cover panels on the rear of the support arm (Fig. 18 / p. 23) that cover the
column.
3. Release the brake by pressing the appropriate button for horizontal tube rotation on the
TUI.
The brake must be released during the adjustment.
4. For the adjustment, turn the eccentric bolt (2/Fig. 20 / p. 23).
5. Retighten the securing screw (1/Fig. 20 / p. 23).
6. Recheck the drift.
7. Reinstall the cover panels.
Check
Adjustment
The adjustment can be found in the CB-DOC under Collimator > Adjustment
(AX74-020.842.01).
Check
Adjustment
Potentiometers 2.2
Adjustment of the Potentiometer for the transverse Axis (only Fully Synchronized
System) 0
Perform
• following replacement or removal of the potentiometer.
• if there are problems with the position display.
Adjust the potentiometer for position sensing of the transverse axis using the SSW.
1. Log in to the SSW.
Select the Tune Up > Stand > Position Check menu.
2. The check of the potentiometer range for the transverse axis is activated by selecting
Set. Position the 3D in the middle of the movement range of the transverse axis and
turn the potentiometer in the valid value range. When the value range has been
reached, the “Success“ audible signal sounds. If the 3D is outside of the valid value
range, the “Fail“ audible signal sounds (continuous signal).
3. The screen is exited by selecting Exit.
Brakes 2.3
Check
1. Remove the cover panels on the rear of the support arm (Fig. 23 / p. 28) that cover the
column.
Adjustment
1. During the adjustment, press the button on the TUI for the vertical movement direction.
The screws to adjust the brake gap (1/Fig. 26 / p. 29) for the vertical axis are located on
the bottom of the tube unit support arm.
Loosen the Allen screws (1/Fig. 27 / p. 29) and adjust the brake gap with the adjust-
ment screws (hex screws) (2/Fig. 27 / p. 29).
The brake gap must be 0,3 + 0,1 mm all around.
Check
1. Remove the cover panels on the rear of the support arm (Fig. 28 / p. 30) that cover the
column.
Adjustment
1. During the adjustment, press the button on the TUI for the movement direction.
The screws to adjust the brake gap for the horizontal axis (1/Fig. 31 / p. 31) and
(1/Fig. 32 / p. 31) are located on the back of the tube unit support arm.
Loosen the Allen screws (1/Fig. 33 / p. 31) and adjust the brake gap with the adjust-
ment screws (hex screws) (2/Fig. 33 / p. 31).
The brake gap must be 0,35 + 0,1 mm all around.
Transverse and Longitudinal Carriage Brake Gap (only Fully Synchronized Sys-
tem) 0
Check
If the brake gap is too large because the pads are worn, a noise (clattering) will develop
when the brakes are applied.
Check the brake gap with a feeler gauge ((Fig. 34 / p. 32) or (Fig. 35 / p. 32)).
Brake pad surface to the surface: 0,05 + 0,1 mm
Fig. 34: Checking the longitudinal direction brake Fig. 35: Checking the transverse direction brake
gap gap
Adjustment
NOTE The surface of the eccentric bolts must be flush with the
mounting bracket. Only turn the eccentric bolts, do NOT
press them!
Fig. 36: Brake assembly eccentric bolt Fig. 37: Brake assembly set screw
The detent forces for the horizontal and vertical axis rotation may have to be adjusted per
the customer's request.
1. Remove the cover panels on the rear of the support arm (Fig. 38 / p. 34) that cover the
column.
3. If only turning the screw for the vertical axis is not sufficient to adjust the detent force,
washers can be placed under the springs. Spare washers are located in the support
arm (Fig. 40 / p. 35).
If the sensor does not react, it is possible to adjust the sensor by moving it.
1. The spacing of S1 and S2 to the metal bracket for the stop assembly (1/Fig. 43 / p. 38)
is 3 ± 0.3 mm.
2. The spacing of S3 to the plate while switched on (stud is engaged, magnet is powered)
(2/Fig. 43 / p. 38) is 2 ± 0.3 mm.
3. The counter nut (3/Fig. 43 / p. 38) must be tightened! The plate can be moved!
If the drift of the central beam is adjusted, the adjustments in the SSW must then be per-
formed again.
Table 0
1. Select a suitable organ program and center the tube unit over the table.
2. Switch on the light localizer.
3. Position the ball phantom to the centering cross of the light field on the table
4. Release an exposure.
5. The ball in the image must appear inside the circle (Fig. 44 / p. 40).
4. Position the ball phantom to the centering cross of the light field on the front panel.
5. Release an exposure.
6. The ball in the image must appear inside the circle (Fig. 44 / p. 40).
IQAP 3.2
A change in the adjustment of the radiation field must be checked with IQAP and docu-
mented.
For this, start a Partial Test and perform the ColliCenter test point. If an adjustment of the
light/radiation field was performed or if the collimator was removed from the tube unit, per-
form the test point Light and XRay.