TMS01

This document provides technical information about the iTHERM MultiSens Flex TMS01 modular direct contact thermometer. It is designed for oil and gas and petrochemical applications to measure temperature using thermocouples or RTDs with a measuring range of -270 to 1150°C. It has a modular and flexible design that allows for infinite sensor layout configurations. The transmitter provides analog or digital communication outputs. The document describes the device's measurement principles, components, installation requirements, certifications, and ordering information.

Uploaded by

Bayu Firdaus
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
23 views

TMS01

This document provides technical information about the iTHERM MultiSens Flex TMS01 modular direct contact thermometer. It is designed for oil and gas and petrochemical applications to measure temperature using thermocouples or RTDs with a measuring range of -270 to 1150°C. It has a modular and flexible design that allows for infinite sensor layout configurations. The transmitter provides analog or digital communication outputs. The document describes the device's measurement principles, components, installation requirements, certifications, and ordering information.

Uploaded by

Bayu Firdaus
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 32

TI01256T/09/EN/04.

22-00 Products Solutions Services


71570972
2022-06-15

Technical Information
iTHERM
MultiSens Flex TMS01
Modular direct contact TC and RTD multipoint
thermometer for Oil & Gas and Petrochemical
applications

Application
• Easy-to-use device with modular and flexible design, ready to be installed either
for direct contact measurements, or in an existing thermowell
• Specifically designed for Oil & Gas and Petrochemical processing industries
• Measuring range:
• Resistance insert (RTD): –200 to 600 °C (–328 to 1 112 °F)
• Thermocouple (TC): –270 to 1 150 °C (–454 to 2 102 °F)
• Static pressure range: Up to 100 bar (1 450 psi). Specific maximum process
pressure achievable depending on the thermometer design and process
temperature
• Degree of protection: IP66/67

Head transmitter
All Endress+Hauser transmitters are available with enhanced accuracy and reliability
compared to directly wired sensors. Easy customizing by choosing one of the
following outputs and communication protocols:
• Analog output 4 to 20 mA
• HART®
• PROFIBUS® PA
• FOUNDATION Fieldbus™

Your benefits
• Infinite 3D sensors distribution layouts for any process monitoring configuration
• High density of measuring points thanks to the possibility to mount the ProfileSens
sensor technology
• High degree of customization thanks to a modular product design for easy
installation, process integration and maintenance
• Easy integration due to inserts according standard IEC 60584, ASTM E230 and IEC
60751
• Thanks to a proper junction box support frame design, overheating of electronic is
avoided resulting in a longer product lifetime
• Compliance to different types of protection for use in hazardous locations for a
wide and easy process integration
• Possibility of sensing elements replacement
iTHERM MultiSens Flex TMS01

Table of contents

Function and system design . . . . . . . . . . . . . . . . . . . 3 Test report and calibration . . . . . . . . . . . . . . . . . . . . . . 23


Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Material requirements . . . . . . . . . . . . . . . . . . . . . . . . . 23
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Welding requirements . . . . . . . . . . . . . . . . . . . . . . . . . 23
Equipment architecture . . . . . . . . . . . . . . . . . . . . . . . . . 4 Pressure equipment requirements . . . . . . . . . . . . . . . . . 23

Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Ordering information . . . . . . . . . . . . . . . . . . . . . . . 24


Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Device-specific accessories . . . . . . . . . . . . . . . . . . . . . . 27
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Service-specific accessories . . . . . . . . . . . . . . . . . . . . . . 28
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Family of temperature transmitters . . . . . . . . . . . . . . . . . 7 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Wiring diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Performance characteristics . . . . . . . . . . . . . . . . . . 12
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Response time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Shock and vibration resistance . . . . . . . . . . . . . . . . . . . . 13
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Mounting location . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . 15

Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Ambient temperature range . . . . . . . . . . . . . . . . . . . . . 15
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Climate class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . 16

Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Process temperature range . . . . . . . . . . . . . . . . . . . . . . 16
Process pressure range . . . . . . . . . . . . . . . . . . . . . . . . . 16

Mechanical construction . . . . . . . . . . . . . . . . . . . . 16
Design, dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Process connection . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Certificates and approvals . . . . . . . . . . . . . . . . . . . 23


CE Mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Hazardous area approvals . . . . . . . . . . . . . . . . . . . . . . . 23
Certification HART . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Certification FOUNDATION Fieldbus . . . . . . . . . . . . . . . . 23
Certification PROFIBUS® PA . . . . . . . . . . . . . . . . . . . . . 23
Other standards and guidelines . . . . . . . . . . . . . . . . . . . 23
Material certification . . . . . . . . . . . . . . . . . . . . . . . . . . 23

2 Endress+Hauser
iTHERM MultiSens Flex TMS01

Function and system design


Measuring principle Thermocouples (TC)
Thermocouples are comparatively simple, robust temperature sensors which use the Seebeck effect
for temperature measurement: if two electrical conductors made of different materials are connected
at a point, a weak electrical voltage can be measured between the two open conductor ends if the
conductors are subjected to a thermal gradient. This voltage is called thermoelectric voltage or
electromotive force (emf.). Its magnitude depends on the type of conducting materials and the
temperature difference between the "measuring point" (the junction of the two conductors) and the
"cold junction" (the open conductor ends). Accordingly, thermocouples primarily only measure
differences in temperature. The absolute temperature at the measuring point can be determined
from these if the associated temperature at the cold junction is known or is measured separately and
compensated for. The material combinations and associated thermoelectric voltage/temperature
characteristics of the most common types of thermocouple are standardized in the IEC 60584 and
ASTM E230/ANSI MC96.1 standards.

Resistance thermometer (RTD)


These resistance thermometers use a Pt100 temperature sensor according to IEC 60751. The
temperature sensor is a temperature-sensitive platinum resistor with a resistance of 100 Ω at
0 °C (32 °F) and a temperature coefficient α = 0.003851 °C-1.
There are generally two different kinds of platinum resistance thermometers:
• Wire wound (WW): Here, a double coil of fine, high-purity platinum wire is located in a ceramic
support. This is then sealed top and bottom with a ceramic protective layer. Such resistance
thermometers not only facilitate very reproducible measurements but also offer good long-term
stability of the resistance/temperature characteristic within temperature ranges up to
600 °C (1 112 °F). This type of sensor is relatively large in size and it is comparatively sensitive to
vibrations.
• Thin film platinum resistance thermometers (TF): A very thin, ultrapure platinum layer,
approx. 1 μm thick, is vaporized in a vacuum on a ceramic substrate and then structured
photolithographically. The platinum conductor paths formed in this way create the measuring
resistance. Additional covering and passivation layers are applied and reliably protect the thin
platinum layer from contamination and oxidation, even at high temperatures. The primary
advantages of thin film temperature sensors over wire wound versions are their smaller sizes and
better vibration resistance. A relatively low principle-based deviation of the resistance/
temperature characteristic from the standard characteristic of IEC 60751 can frequently be
observed among TF sensors at high temperatures. As a result, the tight limit values of tolerance
category A as per IEC 60751 can only be observed with TF sensors at temperatures up to approx.
300 °C (572 °F). For this reason, thin-film sensors are generally only used for temperature
measurements in ranges below 400 °C (752 °F).

Measuring system Endress+Hauser offers a complete portfolio of optimized components for the temperature measuring
point – everything you need for the seamless integration of the measuring point into the overall
facility.
This includes:
• Power supply unit/active barrier
• Configuration units
• Overvoltage protection
For more information, see the brochure 'System Components - Solutions for a Complete
Measuring Point' (FA00016K/09)

Endress+Hauser 3
iTHERM MultiSens Flex TMS01

3 4 5

7 7
6 6

5 5
1 1
+ +
2 2
- 4 - 4

3 3

7 7
6 6

5 5
1 1
+ +
2 2
- 4 - 4

3 3

+
-
+
-
+
-
+
-
+
-
+
-

3 OUT 4 3 OUT 4 3 OUT 4 3 OUT 4

1 2

HAW56x

HAW56x

HAW56x
HAW56x
1 IN 2 1 IN 2 1 IN 2 1 IN 2

A0028076

1 Application example in a reactor, mounted multipoint thermometer in a locally existing thermowell with
four measurement points and four built-in transmitters or terminal blocks.
1 Device configuration with application software FieldCare
2 Commubox
3 PLC
4 Active barrier RN221N (24 VDC, 30 mA) that has a galvanically isolated output for supplying voltage to loop-
powered transmitters. The universal power supply works with an input supply voltage of 20 to 250 V DC/AC;
50/60 Hz, which means that it can be used in all international power grids.
5 Mounted multipoint thermometer in a locally existing thermowell, optionally with built-in transmitters in the
junction box for 4 to 20 mA-, HART-, PROFIBUS® PA-, FOUNDATION Fieldbus™ communication or terminal
blocks for remote wiring.
6 Surge arrester modules HAW562 for protection of signal lines and components in hazardous areas, e.g.
4 to 20 mA- , PROFIBUS® PA, FOUNDATION Fieldbus™ signal lines. More information on this can be found
in the Technical Information →  29

Equipment architecture The multipoint thermometer belongs to a range of modular product configuration for multipoint
temperature detection with a design where subassemblies and components can be managed
individually for easy maintenance and spare part ordering.
It consists of the following main sub-assemblies:
• Single point insert: Done by a metal sheathed sensing measuring element (thermocouple or
resistance), extension cables and transition bushing. When applicable, each insert can be handled
as an individual spare part that may be replaceable by untightening its compression fitting
installed onto the process connection. They can be ordered via specific standard product order
codes (e.g. TSC310, TST310) or special codes. For the specific order code please contact the
Endress+Hauser service department.
• Multiple insert: Composed of a plurality of independent metal sheated thermocouple cables in
one probe, each of them provided with its seal pot and extension cables resulting in a dual seal
insert design (Endress+Hauser ProfileSens).
• Process connection: Represented by an ASME or EN flange, it might be provided with eyebolts for
lifting the device.
• Head: It is composed of a junction box provided with its components such as cable glands, draining
valves, earth screws, terminals, head transmitters, etc.
• Neck: It is designed to support the junction box by components such as supporting rods and plates
or tube extension.
• Additional accessories: Components that can be ordered independently from the selected product
configuration, such as clips, pads, tips, spacers and plates for tag sensors.
• Thermowells: They are directly welded on the process connection, designed to guarantee higher
degree of mechanical protection and corrosion resistance for each sensor.

4 Endress+Hauser
iTHERM MultiSens Flex TMS01

In general, the system measures the temperature profile inside the process environment by means of
many sensors, jointed to a suitable process connection which ensures the right tightness levels.
Externally, the extension cables are wired into the junction box, which can be directly mounted or
remote as option.

Design Description, available options and


materials

1: Head Hinged cover junction box for electrical


connections. It includes components such
as electrical terminals, transmitters and
cable glands.
• 316/316L
• Other materials on request

2a: Frame neck Modular frame support that is adjustable


for all available junction boxes.
316/316L

2b: Tube neck Modular tube support that is adjustable for


all available junction boxes and ensures
extension cable inspection.
316/316L

3: Compression fitting High performance compression fitting for a


proper tightness between process and
external environment, for a wide range of
process fluids concentration and severe
combination between temperature and
1 pressure.
• 316L
• 316H

4: Process connection Represented by a flange according to


international standards, or engineered to
satisfy specific process requirements.
→  21
2a
• 304/304L
• 316/316L
2b • 316Ti
3 • 321
7 • 347
4
• Other materials on request
5 5: Insert • Mineral insulated grounded and
6a ungrounded thermocouples or RTD
(Pt100 wire wound)
• Mineral insulated ungrounded
multipoint cable insert with
6b thermocouples (ProfileSens)
For details refer to the ordering
A0028078
information table

6a: Protecting thermowells The thermometer can be equipped:


6b: Thermowells closure tip • either with protecting thermowells for
increased mechanical strength and
corrosion resistance
• or open guiding tubes for installation in
an existing thermowell.
• 316/316L
• 321
• 347
• Alloy 600
• Other materials on request

7: Eyebolt Lifting device for easy handling during


installation phase.
316

Endress+Hauser 5
iTHERM MultiSens Flex TMS01

The modular multipoint thermometer is characterized by the following possible main configurations:
• Linear configuration
The different sensors are aligned along
the straight direction coinciding with the
longitudinal axis of the multipoint
assembly itself (linear multipoint
measurement). This configuration can be
used to install the multipoint either in an
1 existing thermowell as part of the reactor
or in direct contact with the process.
• 3D distribution configuration
For a high number of measuring points,
each multipoint cable sensor can be bent
and arranged in a three dimensional
configuration by fixing them through clips
or other equivalent accessories. This
configuration is commonly used to reach
several measurement points distributed at
different cross sections and levels. Specific
support frames can be provided and
installed on request if they are not already
available on site.

A0028362

2 Main configuration possibilities


1 Linear configuration
2 3D configuration

Input
Measured variable Temperature (temperature linear transmission behavior)

Measuring range RTD:


Input Designation Measuring range limits

RTD as per IEC 60751 Pt100 –200 to +600 °C (–328 to +1 112 °F)

Thermocouple:
Input Designation Measuring range limits

Thermocouples (TC) as per IEC Type J (Fe-CuNi) –210 to +720 °C (–346 to +1 328 °F)
60584, part 1 - using an Type K (NiCr-Ni) –270 to +1 150 °C (–454 to +2 102 °F)
Endress+Hauser - iTEMP Type N (NiCrSi-NiSi) –270 to +1 100 °C (–454 to +2 012 °F)
temperature head transmitter Type T (Cu-CuNi) –270 to +370 °C (–454 to +698 °F)

Internal cold junction (Pt100)


Cold junction accuracy: ± 1 K
Max. sensor resistance: 10 kΩ

6 Endress+Hauser
iTHERM MultiSens Flex TMS01

Input Designation Measuring range limits

Thermocouples (TC) - flying Type J (Fe-CuNi) –270 to +720 °C (–454 to +1 328 °F), typical
leads - as per IEC 60584 and sensitivity above 0 °C ≈ 55 μV/K
ASTM E230 Type K (NiCr-Ni) –270 to +1 150 °C (–454 to +2 102 °F) 1), typical
sensitivity above 0 °C ≈ 40 μV/K
Type N (NiCrSi-NiSi) –270 to +1 100 °C (–454 to +2 012 °F), typical
sensitivity above 0 °C ≈ 40 μV/K
Type T (Cu-CuNi) –270 to +370 °C (–454 to +698 °F), typical sensitivity
above 0 °C ≈ 43 μV/K

1) Limited by jacket material of insert

Output
Output signal Generally, the measured value can be transmitted in one of two ways:
• Directly-wired sensors - sensor measured values forwarded without a transmitter.
• Via all common protocols by selecting an appropriate Endress+Hauser iTEMP temperature
transmitter. All the transmitters listed below are mounted directly in the junction box and wired
with the sensory mechanism.

Family of temperature Thermometers fitted with iTEMP transmitters are an installation-ready complete solution to
transmitters improve temperature measurement by significantly increasing accuracy and reliability, when
compared to direct wired sensors, as well as reducing both wiring and maintenance costs.
PC programmable head transmitters
They offer a high degree of flexibility, thereby supporting universal application with low inventory
storage. The iTEMP transmitters can be configured quickly and easily at a PC. Endress+Hauser offers
free configuration software which can be downloaded from the Endress+Hauser Website. More
information can be found in the Technical Information.
HART® programmable head transmitters
The transmitter is a 2-wire device with one or two measuring inputs and one analog output. The
device not only transfers converted signals from resistance thermometers and thermocouples, it also
transfers resistance and voltage signals using HART® communication. It can be installed as an
intrinsically safe apparatus in Zone 1 hazardous areas and is used for instrumentation in the
terminal head (flat face) as per DIN EN 50446. Swift and easy operation, visualization and
maintenance using universal device configuration tools like FieldCare, DeviceCare or
FieldCommunicator 375/475. For more information, see the Technical Information.
PROFIBUS® PA head transmitters
Universally programmable head transmitter with PROFIBUS® PA communication. Conversion of
various input signals into digital output signals. High accuracy over the complete ambient
temperature range. The configuration of PROFIBUS PA functions and of device-specific parameters is
performed via fieldbus communication. For more information, see the Technical Information.
FOUNDATION Fieldbus™ head transmitters
Universally programmable head transmitter with FOUNDATION Fieldbus™ communication.
Conversion of various input signals into digital output signals. High accuracy over the complete
ambient temperature range. All transmitters are released for use in all important process control
systems. The integration tests are performed in Endress+Hauser's "System World". For more
information, see the Technical Information.
Advantages of the iTEMP transmitters:
• Dual or single sensor input (optionally for certain transmitters)
• Pluggable display (optionally for certain transmitters)
• Unsurpassed reliability, accuracy and long-term stability in critical processes
• Mathematical functions
• Monitoring of the thermometer drift, sensor backup functionality, sensor diagnostic functions
• Sensor-transmitter matching for dual sensor input transmitters, based on Callendar/Van Dusen
coefficients

Endress+Hauser 7
iTHERM MultiSens Flex TMS01

Power supply
• Electrical connecting cables must be smooth, corrosion resistant, easy to be cleaned and
inspected, robust against mechanical stresses, no-humidity sensitivity.
• Grounding or shielding connections are possible via ground terminals on the junction box.

Wiring diagrams Type of sensor connection RTD

1 2 3
R3 RD
R1 RD R2 R1 RD B5 BK
R1 R3
R1 W3 R1 RD
R3 R1
R3 RD R2 RD B6 BK

W5 B6 G4 W6 W4 W4 WH

W5 WH B5
W3 WH G4 YE W6 WH
4
A0045453

3 Terminal block mounted


1 3-wire, single
2 2 x 3-wire, single
3 4-wire, single
4 Outside screw

3 2
6
RD
5
RD
1
1 mA
4

WH
3 2
WH

A0045600

4 Head mounted transmitter TMT18x (single sensor input)


1 Power supply, head transmitter and analog output 4 to 20 mA or fieldbus connection
2 RTD, 3-wire
3 RTD, 4-wire

Only available with screw terminals

1 2

RD 6
+
RD 5
1+
+

4 -2-
WH -
WH 3

3
A0045464

5 Head mounted transmitter TMT7x or TMT31 (single sensor input)


1 Sensor input, RTD and Ω: 4-, 3- and 2-wire
2 Power supply or fieldbus connection
3 Display connection/CDI interface

8 Endress+Hauser
iTHERM MultiSens Flex TMS01

2 1 3
BK

BK RD 7
6
RD
5
1
+
2
WH 4
-

YE WH 3

4
A0045466

6 Head mounted transmitter TMT8x (dual sensor input)


1 Sensor input 1, RTD: 4- and 3-wire
2 Sensor input 2, RTD: 3-wire
3 Power supply or fieldbus connection
4 Display connection

Mounted field transmitter: Fitted with screw terminals

WH WH RD RD

2 3
2
4
GN
1
5 4 3+ -
3
3 S1 +
BK S2 +
6
!-
+

-
-
BK
A0045732

7 TMT162 (dual sensor input)


1 Sensor input 1, RTD: 3- and 4-wire
2 Sensor input 2, RTD: 3-wire
3 Power supply, field transmitter and analog output 4 to 20 mA or fieldbus connection

Endress+Hauser 9
iTHERM MultiSens Flex TMS01

+
-
5 4 3

1
RD

+
-
2

!
3+ -
RD

S1
3

4
4
WH

WH

1
A0045733

8 TMT142B (single sensor input)


1 Sensor input RTD
2 Power supply, field transmitter and analog output 4 to 20 mA, HART® signal
3 2-wire
4 3-wire
5 4-wire

Type of sensor connection thermocouple (TC)

1 x TC 2 x TC
-
+
-

+ +

A0012700

9 Terminal block mounted

10 Endress+Hauser
iTHERM MultiSens Flex TMS01

Head mounted transmitter TMT18x (single sensor input) 1) Head mounted transmitter TMT8x (dual sensor input) 2)

1 2 1
3
TC TC
6

7
6

1+ 5
4 1+
2-

2-
4

3
A0045467

1 Power supply, head transmitter and analog output 4 to 20 mA


or fieldbus communication 4
A0045474

1 Sensor input 1
2 Sensor input 2
3 Fieldbus communication and power supply
4 Display connection

Head mounted transmitter TMT7x (single sensor input) 2) Mounted field transmitter TMT162 or TMT142B 1)

1 1
6
3
4 3 2
1+
+
5 5 -
1
+ 2 4 3+

S1
3 + +
6 S2
!
+

-
- -
4 -
-2-
3
1 2
3 2 TC TC
A0045353 2 1 5 6

1 Sensor input TC, mV


2 Power supply, bus connection
3 Display connection/CDI interface A0045636

1 Sensor input 1
2 Sensor input 2 (not TMT142B)
3 Supply voltage for field transmitter and analog output 4 to 20 mA or fieldbus
communication

1) Fitted with screw terminals


2) Fitted with spring terminals if screw terminals are not explicitly selected or a dual sensor is installed.

Thermocouple wire colors


As per IEC 60584 As per ASTM E230

• Type J: black (+), white (-) • Type J: white (+), red (-)
• Type K: green (+), white (-) • Type K: yellow (+), red (-)
• Type N: pink (+), white (-) • Type N: orange (+), red (-)
• Type T: Brown (+), white (-) • Type T: Blue (+), red (-)

Endress+Hauser 11
iTHERM MultiSens Flex TMS01

Performance characteristics
Accuracy RTD resistance thermometer as per IEC 60751

Class Max. tolerances (°C) Characteristics


1)
Cl. AA, former 1/3 ± (0.1 + 0.0017 · |t| )
Max. deviation (°C)
Cl. B 3.0

Cl. A ± (0.15 + 0.002 · |t| 1)) 2.5

Cl. B ± (0.3 + 0.005 · |t| 1))


2.0
Temperature ranges for compliance with the
B
tolerance classes 1.5

Wire wound Cl. A Cl. AA


sensor (WW): 1.0
– –50 to +250 °C A
100 to +450 °C 0.5 AA

Thin-film version Cl. A Cl. AA


(TF):
-200 -100 0 100 200 300 400 500 600°C
Standard –30 to +300 °C 0 to +150 °C
AA
- 0.5

A
- 1.0

- 1.5 B

- 2.0

- 2.5

- 3.0 Max. deviation (°C)

A0045598

1) |t| = absolute value °C

In order to obtain the maximum tolerances in °F, the results in °C must be multiplied by a factor
of 1.8.
Permissible deviation limits of thermoelectric voltages from the standard characteristic for
thermocouples as per IEC 60584 or ASTM E230/ANSI MC96.1:

Standard Type Standard tolerance Special tolerance

IEC 60584 Class Deviation Class Deviation

J (Fe-CuNi) 2 ±2.5 °C (–40 to 333 °C) 1 ±1.5 °C (–40 to 375 °C)


±0.0075 |t| 1) (333 to 750 °C) ±0.004 |t| 1) (375 to 750 °C)

K (NiCr-NiAl) 2 ±2.5 °C (–40 to 333 °C) 1 ±1.5 °C (–40 to 375 °C)


±0.0075 |t| 1) (333 to 1 200 °C) ±0.004 |t| 1) (375 to 1 000 °C)

1) |t| = absolute value °C

Standard Type Standard tolerance Special tolerance

ASTM E230/ANSI Deviation, the larger respective value applies


MC96.1
J (Fe-CuNi) ±2.2 K or ±0.0075 |t| 1) (0 to 760 °C) ±1.1 K or ±0.004 |t| 1)
(0 to 760 °C)

K (NiCr- ±2.2 K or ±0.02 |t| 1) (–200 to 0 °C) ±1.1 K or ±0.004 |t| 1)


NiAl) ±2.2 K or ±0.0075 |t| 1) (0 to 1 260 °C) (0 to 1 260 °C)

1) |t| = absolute value °C

12 Endress+Hauser
iTHERM MultiSens Flex TMS01

Response time Response time for the sensor assembly without transmitter. It refers to inserts in direct contact
with process. When thermowells are selected specific evaluation should be done.

RTD
Calculated at an ambient temperature of approx. 23 °C by immersing the insert in running water (0.4
m/s flow rate, 10 K excess temperature):

Insert diameter Response time

Mineral-insulated cable, 3 mm (0.12 in) t50 2s

t90 5s

RTD insert StrongSens, 6 mm (¹⁄₄ in) t50 < 3.5 s

t90 < 10 s

Thermocouple (TC)
Calculated at an ambient temperature of approx. 23 °C by immersing the insert in running water (0.4
m/s flow rate, 10 K excess temperature):

Insert diameter Response time

Grounded thermocouple: t50 0.8 s


3 mm (0.12 in), 2 mm (0.08 in)
t90 2s

Ungrounded thermocouple: t50 1s


3 mm (0.12 in), 2 mm (0.08 in)
t90 2.5 s

Grounded thermocouple t50 2s


6 mm (¹⁄₄ in)
t90 5s

Ungrounded thermocouple t50 2.5 s


6 mm (¹⁄₄ in)
t90 7s

Cable probe diameter (ProfileSens) Response time

8 mm (0.31 in) t50 2.4 s

t90 6.2 s

9.5 mm (0.37 in) t50 2.8 s

t90 7.5 s

12.7 mm (¹⁄₂ in) t50 3.8 s

t90 10.6 s

Shock and vibration • RTD: 3G / 10 to 500 Hz according to IEC 60751


resistance • RTD iTHERM StrongSens Pt100 (TF, vibration resistant): Up to 60G
• TC: 4G / 2 to 150 Hz according to IEC 60068-2-6

Calibration Calibration is a service that can be performed on each individual insert, either in order phase, or after
multipoint installation.
When calibration shall be performed once the multipoint is installed, please contact the
Endress+Hauser service to get full support. Together with the Endress +Hauser service any
further activity can be organised to achieve the calibration of the target sensor. In any case it is
forbidden to unscrew any threaded component on the process connection under operating
conditions = running process.
Calibration involves comparing the measured values of the sensing elements of the multipoint
inserts (DUT device under test) with those of a more precise calibration standard using a defined and

Endress+Hauser 13
iTHERM MultiSens Flex TMS01

reproducible measurement method. The aim is to determine the deviation of the DUT measured
values from the true value of the measured variable.
In case of multipoint cable sensor, temperature controlled calibration baths from
–80 to 550 °C (–112 to 1 022 °F) can be used only for the last measuring point (when NL-LMPx <
100 mm (3.94 in)) for either factory calibration or accredited calibration. Special bore through
calibration furnaces with homogeneous distribution of temperature are used for thermometer
factory calibration along the length: 200 to 550 °C (392 to 1 022 °F)
Two different methods are used for the inserts:
• Calibration at fixed-point temperatures, e.g. at the freezing point of water at 0 °C (32 °F).
• Calibration compared against a precise reference thermometer.
Evaluation of inserts
If a calibration with an acceptable uncertainty of measurement and transferable measurement
results is not possible, Endress+Hauser offers an insert evaluation measurement service, if
technically feasible.

Installation
Mounting location The installation location must meet the requirements listed in this documentation, such as ambient
temperature, protection classification, climatic class, etc.. Care should be taken when checking the
sizes of possible existing support frames or brackets welded on the reactor’s wall (usually not
included in the scope of delivery) or of any other existing frame in the installation area.

Orientation No restrictions. The multipoint thermometer can be installed either in horizontal, oblique or in
vertical configuration, related to the reactor or vessel vertical axis.

A0028440

 10 Installation examples - no restrictions to the installation orientation


1 Vertical installation with linear configuration
2 Oblique installation with 3D distribution configuration
3 Horizontal installation with 3D distriburtion configuration

14 Endress+Hauser
iTHERM MultiSens Flex TMS01

Installation instructions The modular multipoint thermometer is designed to be installed with a flanged process connection
into a vessel, reactor, tank or similar environment. All parts and components have to be handled
with care. During the installation phase, lifting and introduction of the equipment through the
preset nozzle, the following must be avoided:
• Misalignment with the nozzle axis.
• Any load on the welded or threaded parts due to the action of the weight of the device.
• Deformation or crushing of the threaded components, bolts, nuts, cable glands and compression
fittings.
• Bending radius of the thermowells smaller than 20 times the diameter of the thermowell.
• Friction between the temperature probes and the internals of the reactor.
• Fixing the temperature probes to the reactor's infrastructures without allowing axial displacements
or movements.
• A bending radius of the sheathed cable (inserts) smaller than 5 times the outer diameter of the
sheathed cable.

A0028441

 11 Multipoint thermometer installation in a reactor nozzle via flange process connection.

During installation the whole thermometer must only be lifted and moved by using ropes
properly mounted on the eyebolt of the flange (1).

Environment
Ambient temperature range Junction box Non-hazardous area Hazardous area

Without mounted –40 to +85 °C (–40 to +185 °F) –40 to +60 °C (–40 to +140 °F)
transmitter

With mounted head –40 to +85 °C (–40 to +185 °F) Depends on the respective hazardous area
transmitter approval. Details see Ex documentation.

Storage temperature Junction box

With head transmitter –40 to +95 °C (–40 to +203 °F)

With DIN rail transmitter –40 to +95 °C (–40 to +203 °F)

Endress+Hauser 15
iTHERM MultiSens Flex TMS01

Humidity Condensation according to IEC 60068-2-14:


• Head transmitter: Permitted
• DIN rail transmitter: Not permitted
Maximum relative humidity: 95% according to IEC 60068-2-30

Climate class Determined when the following components are installed into the junction box:
• Head transmitter: Class C1 according to EN 60654-1
• Multi-channel transmitter: Tested as per IEC 60068-2-30, meets the requirements regarding class
C1-C3 in accordance with IEC 60721-4-3
• Terminal blocks: Class B2 according to EN 60654-1

Degree of protection • Specification for conduit: IP68


• Specification for the junction box: IP66/67

Electromagnetic Depending on the transmitter used. For detailed information see the related Technical Information,
compatibility (EMC) listed at the end of this document.

Process
The process temperature and process pressure are the minimum input parameters for the selection
of the right product configuration. If special product features are requested, additional data such as
process fluid type, phases, concentration, viscosity, stream and turbulences, corrosion rate have to be
considered as mandatory for the whole product definition.

Process temperature range Up to +1 150 °C (+2 102 °F).

Process pressure range 0 to 100 bar (0 to 1 450 psi)


Anyhow, the maximum required process pressure has to be combined with the maximum
design process temperature. Process connections like compression fittings, flanges with their
specific ratings, thermowells, selected according to the plant requirements, define the
maximum process conditions at which the device has to operate. Endress+Hauser experts can
support the customer on any related questions.
Process applications:
• Olefins
• Ethylene
• Propylene
• Aromatics
• Benzene
• N-based inorganics
• Ammonia
• Urea
• NGTL
• Distillation units and hydrogenation

Mechanical construction
Design, dimensions The overall multipoint assembly is composed of different sub-assemblies. Both linear and 3D
configurations have the same features, dimensions and materials. Different inserts are available,
based upon specific process conditions, in order to have the highest accuracy and an extended
lifetime. In addition, protecting thermowells can be selected to further increase mechanical
performances and corrosion resistance, and to allow insert replacement. Associated shielded
extension cables are provided with high resistance sheath materials to withstand different
environmental conditions and to ensure steady and noiseless signals. The transition between the
inserts and the extension cable is obtained by the usage of specially sealed bushings, ensuring the
declared IP degree protection.

16 Endress+Hauser
iTHERM MultiSens Flex TMS01

I I

H
H
A
B

L
F

MPx
L MPx

A0028080

 12 Design of the modular multipoint thermometer, with frame neck on the left side or with frame neck and
covers on the right side. All dimensions in mm (in)
A, B, Dimensions of the junction box, see following figure
C
MPx Numbers and distribution of measuring points: MP1, MP2, MP3 etc.
LMPx Different immersion length of sensing elements or thermowells
I, H Encumbrance of the junction box and support system
F Extension neck length
L Overall device length

Extension neck F in mm (in)

Standard 250 (9.84)


Specifically customized extension necks are available on request.

Endress+Hauser 17
iTHERM MultiSens Flex TMS01

Immersion lengths MPx of sensing elements/thermowells:

Based on customer requirements

Junction box

A
C

B
1
2

A0028118

1 Cable gland
2 Junction box
3 Frame

The junction box is suited for chemical agents environments. Sea water corrosion resistance and
extreme temperature variation stability is guaranteed. Ex e/Ex i terminals can be installed.
The multipoint thermometer can be provided of both ground terminals or shield terminals.
Please follow your plant guidelines for a proper cables connection.

Possible junction box dimensions (A x B x C) in mm (in):


A B C

Stainless Steel Min. 170 (6.7) 170 (6.7) 130 (5.1)

Max. 500 (19.7) 500 (19.7) 240 (9.5)

Aluminium Min. 100 (3.9) 150 (5.9) 80 (3.2)

Max. 330 (13) 500 (19.7) 180 (7.1)

Type of specification Junction box Cable glands

Material AISI 316 NiCr Plated brass


AISI 316 / 316L

Ingress protection (IP) IP66/67 IP66

Ambient temperature range –55 to +110 °C (–67 to +230 °F)


(ATEX)

Approvals ATEX, IECEx, UL, CSA, EAC approval for use in hazardous area

18 Endress+Hauser
iTHERM MultiSens Flex TMS01

Type of specification Junction box Cable glands

Marking • ATEX II 2GD Ex e IIC According to the junction box approval


T6/T5/T4 Gb
Ex ia IIC T6/T5/T4 Ga
Ex tb IIIC T85°C/T100°C/
T135°C Db IP66
• IECEx Ex e IIC T6/T5/T4
Gb/
Ex ia IIC T6/T5/T4 Ga
Ex tb IIIC T85°C/T100°C/
T135°C Db IP66
• UL913 Class I, Zone 1, AEx
e IIC; Zone 21, AEx tb IIIC
IP66
• CSA C22.2 No.157 Class I,
Zone 1 Ex e IIC; Class II,
Groups E, F and G

Cover Hinged -

Maximum sealing diameter - 6 to 12 mm (0.24 to 0.47 in)

Neck extension
The neck extension ensures the connection between the flange and the junction box. The design has
been developed to ensure several mounting layouts to deal with possible obstacles and constraints
that can be met in any plant such as the reactor’s infrastructure (step ways, loading structures,
supporting skirts, stairs, etc.) and reactor thermal insulation. The neck extension design ensures easy
access for monitoring and maintaining inserts and extension cables. It guarantees a high stiffness
connection for the junction box and vibration loads. No closed volumes are present in the neck
extension. This avoids the accumulation of waste and potentially dangerous fluids coming from the
environment that can damage the instrumentation allowing continuous ventilation.

Insert and thermowells


Different insert and thermowell types are available. For any different requirement that is not
described here, please contact the Endress+Hauser sales department.
In case of multipoint cable insert (ProfileSens), see Technical Information TI01346T

Thermocouple
Diameter in mm (in) Type Standard Hot junction type Sheath material

1x type K
2x type K
6 (0.24)
1x type J
4.5 (0.18)
2x type J IEC 60584 /ASTM Alloy600 / AISI 316L /
3 (0.12) Grounded/Ungrounded
1x type N E230 Pyrosil
2 (0.08)
2x type N
1.5 (0.06)
1x type T
2x type T

RTD
Diameter in mm (in) Type Standard Sheath material

1x Pt100 WW
3 (0.12) 2x Pt100 WW
IEC 60751 AISI 316L
6 (¹⁄₄) 1x Pt100 TF
2x Pt100 TF

Endress+Hauser 19
iTHERM MultiSens Flex TMS01

Thermowells
External diameter in mm (in) Sheath material Type Thickness in mm (in)

AISI 316/316L closed or open


AISI 316Ti
1 (0.04) or
6 (0.24) AISI 321
1.5 (0.06)
AISI 347
Alloy 600

AISI 316/316L closed or open


AISI 316Ti 1 (0.04) or
8 (0.32) AISI 321 1.5 (0.06) or
AISI 347 2 (0.08)
Alloy 600

AISI 316/316L closed or open


AISI 316Ti
10.2 (¹⁄₈) AISI 321 1.73 (0.068)
AISI 347
Alloy 600

Weight The weight can vary depending on the configuration: Dimension and content of the junction box,
neck length, dimensions of process connection and the number of inserts. The approximate weight
of a typically configured multipoint thermometer (number of inserts = 12, flange size = 3", medium
size junction box) = 40 kg (88 lb)

Materials It refers to insert sheath, neck extension, junction box and all wetted parts.
The temperatures for continuous operation specified in the following table are only intended as
reference values for use of the various materials in air and without any significant compressive load.
The maximum operation temperatures are reduced considerably in some cases where abnormal
conditions such as high mechanical load occur or in aggressive media.

Material Short form Recommended Properties


name max.
temperature for
continuous use
in air

AISI X5CrNiMo 17-12-2 650 °C (1 202 °F) • Austenitic, stainless steel


316/1.4401 • High corrosion resistance in general
• Particularly high corrosion resistance in chlorine-
based and acidic, non-oxidizing atmospheres
through the addition of molybdenum (e.g.
phosphoric and sulfuric acids, acetic and tartaric
acids with a low concentration)

AISI 316L/ X2CrNiMo17-12-2 650 °C (1 202 °F) • Austenitic, stainless steel


1.4404 X2CrNiMo18-14-3 • High corrosion resistance in general
1.4435 • Particularly high corrosion resistance in chlorine-
based and acidic, non-oxidizing atmospheres
through the addition of molybdenum (e.g.
phosphoric and sulfuric acids, acetic and tartaric
acids with a low concentration)
• Increased resistance to intergranular corrosion
and pitting
• Compared to 1.4404, 1.4435 has even higher
corrosion resistance and a lower delta ferrite
content

Alloy600/ NiCr15Fe 1 100 °C • A nickel/chromium alloy with very good


2.4816 (2 012 °F) resistance to aggressive, oxidizing and reducing
atmospheres, even at high temperatures
• Resistance to corrosion caused by chlorine gases
and chlorinated media as well as many oxidizing
mineral and organic acids, sea water etc.
• Corrosion from ultrapure water
• Not to be used in sulfur-containing atmospheres

20 Endress+Hauser
iTHERM MultiSens Flex TMS01

Material Short form Recommended Properties


name max.
temperature for
continuous use
in air

AISI X5CrNi18-10 850 °C (1 562 °F) • Austenitic, stainless steel


304/1.4301 • Well usable in water and lowly pollute waste
water
• Only at relatively low temperatures resistant to
organic acids, saline solutions, sulphates,
alkaline solutions, etc.

AISI 304L/ X2CrNi18-9 850 °C (1 562 °F) • Good welding properties


1.4307 • Impervious to intergranular corrosion
• High ductility, excellent drawing, forming, and
spinning properties

AISI 316Ti/ X6CrNiMoTi17-12-2 700 °C (1 292 °F) • Addition of titanium means increased resistance
1.4571 to intergranular corrosion even after welding
• Broad range of uses in the chemical,
petrochemical and oil industries as well as in coal
chemistry
• Can only be polished to a limited extent, titanium
streaks can form

AISI X6CrNiTi18-10 815 °C (1 499 °F) • Austenitic stainless steel


321/1.4541 • High resistance to intergranular corrosion even
after welding
• Good welding characteristics, suitable to all
standard welding methods
• It is used in many sectors of the chemical
industry, petrochemical, and pressurized vessels

AISI X6CrNiNb10-10 800 °C (1 472 °F) • Austenitic stainless steel


347/1.4550 • Good resistance to a wide variety of
environments in the chemical, textile, oil-
refining, dairy and food industries
• Added niobium makes this steel impervious to
intergranular corrosion
• Good weldability
• Main applications are furnace fire walls, pressure
vessels, welded structures, turbine blades

Process connection

3
A0028122

 13 Flange as process connection


1 Flange
2 Compression fittings
3 Eyebolt

Endress+Hauser 21
iTHERM MultiSens Flex TMS01

Standard process connection flanges are designed according to the following standards:

Standard 1) Size Rating Material

ASME 1½”, 2”, 3”, 4”, 6”, 8” 150#, 300#, 400#, 600# AISI 316, 316L, 304, 304L, 316Ti,
321, 347
EN DN40, DN50, DN80, DN100, PN10,PN16, PN25, PN40,
DN150, DN200 PN63, PN100

1) Flanges according to GOST standard are available on request.

Compression fittings
The compression fittings are welded or threaded into the flange to ensure tightness to the process
connection. Dimensions are coherent with the insert dimensions. Compression fittings comply with
the highest standards of reliability in terms of materials and performances required.

Material AISI 316/316H

Operability
For details of operability, see the Technical Information of the Endress+Hauser temperature
transmitters or the manuals of the related operating software. →  29

22 Endress+Hauser
iTHERM MultiSens Flex TMS01

Certificates and approvals


CE Mark The complete assembly is provided with individual components CE marked, to ensure safe use in
hazardous areas and pressurized environments.

Hazardous area approvals The Ex approval applies to individual components like junction box, cable glands, terminals. For
further details on the available Ex versions (ATEX, UL, CSA, IECEx, NEPSI, EAC Ex), please contact
your nearest Endress+Hauser sales organization. All relevant data for hazardous areas can be found
in separate Ex documentation.
ATEX Ex ia inserts are available only for diameters ≥ 1.5 mm (0.6 in). For further details contact an
Endress+Hauser technician.

Certification HART The HART® temperature transmitter is registered by the FieldComm Group. The device meets the
requirements of the HART® Communication Protocol Specifications.

Certification FOUNDATION The FOUNDATION Fieldbus™ temperature transmitter has successfully passed all test procedures
Fieldbus and is certified and registered by the Fieldbus Foundation. The device thus meets all the
requirements of the following specification:
• Certified according to FOUNDATION Fieldbus™ specification
• FOUNDATION Fieldbus™ H1
• Interoperability Test Kit (ITK), up to date revision status (device certification no. available on
request): the device can also be operated with certified devices of other manufacturers
• Physical layer conformance test of the FOUNDATION Fieldbus™

Certification PROFIBUS® PA The PROFIBUS® PA temperature transmitter is certified and registered by the PNO (PROFIBUS®
Nutzerorganisation e. V.), PROFIBUS user organization. The device meets all the requirements of the
following specifications:
• Certified according to FOUNDATION Fieldbus™ specification
• Certified in accordance with PROFIBUS® PA Profile (the up to date profile version is available on
request)
• The device can also be operated with certified devices of other manufacturers (interoperability)

Other standards and • EN 60079: ATEX certification for hazardous areas


guidelines • IEC 60079: IECEx certification for hazardous areas
• IEC 60529: Degree of protection of housing (IP code)
• IEC 60584 and ASTM E230/ANSI MC96.1: Thermocouples

Material certification The material certificate 3.1 (according to EN 10204) can be requested separately. The certificate
includes a declaration related to the materials used to produce the thermometer. It guarantees the
traceability of the materials through the identification number of the multipoint thermometer.

Test report and calibration The "Factory calibration" is carried out according to an internal procedure in a laboratory of Endress
+Hauser accredited by the European Accreditation Organization (EA) to ISO/IEC 17025. A
calibration which is performed according to EA guidelines (LAT/Accredia) or (DKD/DAkkS) may be
requested separately. The calibration is performed on the inserts of the multipoint.

Material requirements Endress+Hauser can supply components according to AD 2000 W2 And W10 standards.

Welding requirements Endress+Hauser has been audited according to DIN EN ISO 3834-2:2005.

Pressure equipment Endress+Hauser can supply devices according to 2014/68/EU.


requirements

Endress+Hauser 23
iTHERM MultiSens Flex TMS01

Ordering information
Overview of the scope of delivery see the configuration table below.
Detailed ordering information is available from your Endress+Hauser Sales Center:
www.addresses.endress.com

Process connection: Flange

Standard • ASME B16.5 


• EN 1092-1 
Others on request

Material 316 + 316L, 316Ti, 304, 304L, 321, 347 ________


Others on request

Face • RF 
• RTJ 
Others on request

Size • 1½", 2", 3", 4", 6", 8" ________


• DN40, DN50, DN80, DN100, DN150, DN200 ________
Others on request

The values reported in the table below are indicative, based on calculations for nozzles with
standard dimensions. So the maximum number of measurement points can differ from the
maximum number of the configuration table. It depends on the dimensions of the nozzle used
on location.

Flange size Maximum number of thermowells Maximum number of inserts


(considering a with insert-⌀:1.5 mm (0.06 in) or
schedule 40 2 mm (0.08 in)
nozzle)
Thermowell diameter Inserts diameter

10.24 mm 6 mm 8 mm 3 mm 4.5 mm 4.8 mm 6 mm ProfileSens


(¹⁄₈ in) (0.24 in) (0.32 in) (0.12 in) (0.18 in) (0.19 in) (0.24 in) 8 mm (0.31 in),
9.5 mm (0.37 in) or
12.7 mm (¹⁄₂ in)

1½" 3 3 1

2" 5 5 1

3" 8 8 2

4" 16 16 4

6" 30 30 11

8" 48 48 20

Insert, sensor

Measuring principle • Thermocouple (TC) 


• Resistance Temperature Detection (RTD) 
• Multipoint cable sensor ProfileSens (TC) 

Type TC: J, K, N, T ________


RTD: Pt100

Design • TC: Single, duplex 


• RTD: 3-wire, 4-wire, 2x3-wire 

Execution • TC: Grounded, Ungrounded 


• RTD: Wire wound (WW); Thin film (TF) 

Sheath material 316L, Alloy 600, Pyrosil® ________

Approvals • Intrinsic safety


• Non hazardous ________

24 Endress+Hauser
iTHERM MultiSens Flex TMS01

Insert, sensor

Insert diameter • 1.5 mm (0.06 in) 


• 2 mm (0.08 in) 
• 3 mm (0.12 in) 
• 4.5 mm (0.18 in) 
• 4.8 mm (0.19 in) 
• 6 mm (0.24 in) 
• ProfileSens 8 mm (0.31 in) 
• ProfileSens 9.5 mm (0.37 in) 
• ProfileSens 12.7 mm (¹⁄₂ in) 
Others on request

Standard/Class IEC/Class 1 for TC


ASTM/Class special for TC
IEC/Class A for RTD ________
IEC/Class AA for RTD
Others on request

Measurement point distribution

Positioning • Equi spaced 


• Customized 

Number 2, 4, 6, 8, 10, 12 ... 48 1) ________

Insertion length 2) TAG (description) (LMPx) in mm (in)

MP1 _________________________________________ __________

MP2 _________________________________________ __________

MP3 _________________________________________ __________

MP4 _________________________________________ __________

MP5 _________________________________________ __________

MP6 _________________________________________ __________

MPx _________________________________________ __________

1) Different numbers/configurations are available on request


2) If the multipoint cable insert (ProfileSens) will be used, refer to TI01346T

Junction box (head)

Material • Stainless steel (standard) 


• Aluminum (to be specified) 
Others on request

Electrical connection Terminal block wiring:


• Terminal block - standard/number  / _____
• Terminal block - compensated/number  / _____
• Terminal block - spare/number  / _____
Transmitter wiring:
• HART protocol, e. g.: TMT182, TMT82

• PROFIBUS PA protocol, e. g.: TMT84

• FOUNDATION Fieldbus protocol, e. g.: TMT85,

TMT125 (multi-channel transmitter)
_______
• Quantity

Approvals Ex e / Ex ia / Ex d _______
Others on request

Cable entries (process side) Single or multiple, _______ / _______


type: M20, NPT ½", Quantity _______ / _______
Others on request

Cable entries (user side) Single or multiple, _______ / _______


type: M20, M25, NPT ½", NPT 1" / Quantity _______ / _______
Others on request

Endress+Hauser 25
iTHERM MultiSens Flex TMS01

Extension neck

Length F in mm (in) 250 mm (9.84 in) 


Or as specified ________

TAG

Device information Refer to customer specification 


As specified  (table)

Measuring point Refer to customer specification 


information Location, as specified:

• Tagging (TAG), on extension wires insert
If the multipoint cable 
• Tagging (TAG), RFID
sensor (ProfileSens) will be 
• Tagging (TAG), on tip
used, multiple tagging 
• Tagging (TAG), on insert bushing
along the probe will be 
• Tagging (TAG), on device
provided. 
• Tagging (TAG), by customer

• Tagging (TAG), on transmitter
Special version, to be specified

Additional requests

Extension wire length, only


for remote head Specification in mm: ________

Extension wires sheath • PVC 


material • HYFLON 
Others on request

On-site existing Yes 


thermowell No 

Test, Certificate, Declaration

Inspection certificate 3.1, EN10204 (material certificate wetted parts) 1) 

Inspection certificate 3.1, short form, EN10204, (material certificate wetted parts) 

Internal pressure test according to Endress+Hauser procedure, test report (in case of 
thermowells)

Internal helium leak test according to Endress+Hauser procedure, test report (in case 
of thermowells) 1)

PMI test, Endress+Hauser procedure, (wetted parts), test report 

Final assembly functional test, test report 1) 

Final inspection report 1) 

External pressure test according to Endress+Hauser procedure, test report (max length 
10 m)

Routing design including 3D drawing 1) 

2D dimensional drawing 

Welding book (including welding map) 

Radiographic inspection certificate for thermowell welds 


1)
Radiographic inspection certificate on hot junctions/tips for sensors 

Manufacturer declaration 

Dye penetrant test, thermowell welding, test report 

Inspection test report (Sensor/TMT), inspection certificate 1) 

Quality control plan 

1) (recommended)

26 Endress+Hauser
iTHERM MultiSens Flex TMS01

Accessories
Various accessories, which can be ordered with the device or subsequently from Endress+Hauser, are
available for the device. Detailed information on the order code is available from your local Endress
+Hauser sales center.

Device-specific accessories Accessories Description

Tip end Terminal closure welded at the tip end of the probe in
order to protect the insert (or the thermowell) from
aggressive process conditions and to facilitate its fixing
by means of metallic tie wraps.

A0028427

Thermal contact system

Insert and spacers • Used on straight configurations and in case of an


existing thermowell for axial centering of the insert
bundle
• Prevent the twisting of the inserts
• Give flexural stiffness to the sensor bundle

2
A0033485

1 Insert
2 Spacer

Guiding tubes and spacers • Used on straight configurations and in case of an


existing thermowell for axial centering of the insert
1 2 bundle
• Give flexural stiffness to the sensor bundle
• Allow sensor replacement
• Guarantee thermal contact between the sensor tip and
the existing thermowell
• Modular design 1)
3

A0028783

1 Spacer
2 Guiding tube
3 Insert

Endress+Hauser 27
iTHERM MultiSens Flex TMS01

Accessories Description

Thermowells and spacers Used on straight configurations and existing thermowells


Avoid twisting of sensor cables
Give flexural stiffness to the sensor bundle
Allows sensor replacement

1
2

A0028434

1 Thermowell
2 Spacer

Bimetallic stripes • Used on straight configurations and inside existing


thermowells
2 • Guarantee thermal contact between the sensor tip and
the thermowell due to bimetallic stripes activated by
3 temperature difference
• No friction during installation even with already
installed sensors

1
A0028435

 14 Bimetal strips with or without guiding tubes


1 Spacer
2 Guiding tube
3 Bimetallic stripe

1) Can be mounted in-house or on-site

Service-specific accessories Accessories Description

Applicator Software for selecting and sizing Endress+Hauser measuring devices:


• Calculation of all the necessary data for identifying the optimum measuring
device: e.g. pressure loss, accuracy or process connections.
• Graphic illustration of the calculation results
Administration, documentation and access to all project-related data and
parameters over the entire life cycle of a project.
Applicator is available:
Via the Internet: https://portal.endress.com/webapp/applicator

Accessories Description

Configurator Product Configurator - the tool for individual product configuration


• Up-to-the-minute configuration data
• Depending on the device: Direct input of measuring point-specific information
such as measuring range or operating language
• Automatic verification of exclusion criteria
• Automatic creation of the order code and its breakdown in PDF or Excel output
format
• Ability to order directly in the Endress+Hauser Online Shop
The Configurator is available on the Endress+Hauser website at: www.endress.com
-> Click "Corporate" -> Select your country -> Click "Products" -> Select the product
using the filters and search field -> Open product page -> The "Configure" button to
the right of the product image opens the Product Configurator.

28 Endress+Hauser
iTHERM MultiSens Flex TMS01

FieldCare SFE500 FDT-based plant asset management tool from Endress+Hauser.


It can configure all smart field units in your system and helps you manage them. By
using the status information, it is also a simple but effective way of checking their
status and condition.

 For details, see Operating Instructions BA00027S and BA00065S


DeviceCare SFE100 Configuration tool for devices via fieldbus protocols and Endress+Hauser service
protocols.
DeviceCare is the tool developed by Endress+Hauser for the configuration of
Endress+Hauser devices. All smart devices in a plant can be configured via a point-
to-point or point-to-bus connection. The user-friendly menus enable transparent
and intuitive access to the field devices.

 For details, see Operating Instructions BA00027S


Accessories Description

W@M Life cycle management for your plant


W@M offers assistance with a wide range of software applications over the entire
process: from planning and procurement to the installation, commissioning and
operation of the measuring devices. All the relevant information is available for
every measuring device over the entire life cycle, such as the device status, device-
specific documentation, spare parts etc.
The application already contains the data of your Endress+Hauser device.
Endress+Hauser also takes care of maintaining and updating the data records.
W@M is available:
Via the Internet: www.endress.com/lifecyclemanagement

Documentation
• Operating manuals iTEMP temperature transmitters:
• TMT180, PC-programmable, single-channel, Pt100 (KA00118R)
• HART® TMT82, two-channel, RTD, TC, Ω, mV (BA01028T)
• PROFIBUS® PA TMT84, two-channel, RTD, TC, Ω, mV (BA00257R)
• FOUNDATION FieldbusTM TMT85, two-channel, RTD, TC, Ω, mV (BA00251R)
• Supplementary ATEX documentation:
ATEX/IECEx (Ex ia IIC): XA01647T
• Technical Information of inserts:
• Resistance thermometer insert Omnigrad T TST310 (TI00085T)
• Thermocouple insert Omnigrad T TSC310 (TI00255T)
• Multipoint temperature cable probe iTHERM ProfileSens TS901 (TI01346T)
• Technical Information application example:
HAW562 surge arresters, (TI01012K)

Endress+Hauser 29
*71570972*
71570972

www.addresses.endress.com