Detective DX Sdministrator Manual
Detective DX Sdministrator Manual
Detective-EX/DX Series ®
Administrator’s Manual
WARRANTY
ORTEC* warrants that the items will be delivered free from defects in material or workmanship. ORTEC makes no other
warranties, express or implied, and specifically NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE.
ORTEC’s exclusive liability is limited to repairing or replacing at ORTEC’s option, items found by ORTEC to be defective
in workmanship or materials within one year from the date of delivery. ORTEC’s liability on any claim of any kind, including
negligence, loss, or damages arising out of, connected with, or from the performance or breach thereof, or from the
manufacture, sale, delivery, resale, repair, or use of any item or services covered by this agreement or purchase order, shall in
no case exceed the price allocable to the item or service furnished or any part thereof that gives rise to the claim. In the event
ORTEC fails to manufacture or deliver items called for in this agreement or purchase order, ORTEC’s exclusive liability and
buyer’s exclusive remedy shall be release of the buyer from the obligation to pay the purchase price. In no event shall ORTEC
be liable for special or consequential damages.
Quality Control
Before being approved for shipment, each ORTEC instrument must pass a stringent set of quality control tests designed to
expose any flaws in materials or workmanship. Permanent records of these tests are maintained for use in warranty repair and
as a source of statistical information for design improvements.
Repair Service
If it becomes necessary to return this instrument for repair, it is essential that Customer Services be contacted in advance of
its return so that a Return Authorization Number can be assigned to the unit. Also, ORTEC must be informed, either in writing,
by telephone (865.482.4411) or by facsimile transmission (865.483.2133), of the nature of the fault of the instrument being
returned and of the model, serial, and revision (“Rev” on rear panel) numbers. Failure to do so may cause unnecessary delays
in getting the unit repaired. The ORTEC standard procedure requires that instruments returned for repair pass the same quality
control tests that are used for new-production instruments. Instruments that are returned should be packed so that they will
withstand normal transit handling and must be shipped PREPAID via Air Parcel Post or United Parcel Service to the designated
ORTEC repair center. The address label and the package should include the Return Authorization Number assigned.
Instruments being returned that are damaged in transit due to inadequate packing will be repaired at the sender’s expense, and
it will be the sender’s responsibility to make claim with the shipper. Instruments not in warranty should follow the same
procedure and ORTEC will provide a quotation.
Damage in Transit
Shipments should be examined immediately upon receipt for evidence of external or concealed damage. The carrier making
delivery should be notified immediately of any such damage, since the carrier is normally liable for damage in shipment.
Packing materials, waybills, and other such documentation should be preserved in order to establish claims. After such
notification to the carrier, please notify ORTEC of the circumstances so that assistance can be provided in making damage
claims and in providing replacement equipment, if necessary.
*ORTEC® is a registered trademark of Advanced Measurement Technology, Inc. All other trademarks used herein are the
property of their respective owners.
ADDITIONAL WARRANTY STATEMENT
Please note that the personal digital assistant (PDA) that controls the
Detective-EX/DX is intended exclusively for the tasks detailed in the
Detective-EX/DX Series Operator’s Manual and Administrator’s Manual.
Using this PDA for any other purpose could void your warranty.
If you have any questions about the use or maintenance of this instrument,
please contact your ORTEC representative or our Global Service Center
first.
iii
iv
TABLE OF CONTENTS
WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
Cleaning Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix
1. GENERAL INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1. How the Detective-EX Stores Spectra and Search Files . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1.1. Live-Spectrum Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1.2. Stored Spectrum and Search Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1.3. File Formats and Filenaming Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2. A Note About the DX and DX-100 User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
v
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
2.5.1.4. Lin/Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.5.2. Save . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.5.3. Pause/Continue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.6. Advanced Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.6.1. Calibrate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.6.1.1. Start the Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.6.1.2. Apply New Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
2.6.1.3. Restore Default . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2.6.1.4. Back . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2.6.2. Cal. Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2.6.3. Audio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
2.6.4. Turn Cooler Off/On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
2.6.5. Spectra . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.6.5.1. Ask for Sample Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.6.5.2. Default Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.6.5.3. Data Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
2.6.5.4. File Save Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
2.6.5.5. Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
2.6.5.6. Delete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
2.6.6. Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
2.6.7. Search . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
2.6.8. Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
2.6.8.1. Set or Change Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
2.6.8.2. Allow Display of Spectrum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
2.6.8.3. Buttons Beep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
2.6.8.4. About . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
2.6.9. SNM Search . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
2.6.10. Exit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
2.6.11. More >> (Advanced Setup, Page 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
2.6.12. Dose Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
2.6.13. Identify . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
2.6.13.1. Real Time Preset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
2.6.13.2. ID Display Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
2.6.14. List Nuclides (View Analysis Library) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
2.7. Identification Messages — “Classify” ID Display Mode . . . . . . . . . . . . . . . . . . . . . . . 33
2.7.1. SNM Search Mode Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
2.7.2. Identification Mode Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
2.7.2.1. Messages on the Main ID Mode Screen . . . . . . . . . . . . . . . . . . . . . . . 33
2.7.2.2. Messages on the Found Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
2.7.2.3. Messages on the Suspect Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
vi
TABLE OF CONTENTS
4. SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
4.1. Electrical and Mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
4.2. Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4.3. Docking Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4.4. PAC-II Power Adaptor/Charger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4.5. PC Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
4.6. Shipping the Detective-EX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
vii
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
5. TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
5.1. “Error Saving .SPC File” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
5.2. If the Detective-EX Program Stops . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
5.3. If You Drop the Detective-EX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
5.4. If the Detective-EX Will Not Turn On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
5.4.1. Starting from an Exhausted Battery (Reinitialization) . . . . . . . . . . . . . . . . . . . 56
5.5. If the Docking Station Spontaneously Shuts Down and Restarts . . . . . . . . . . . . . . . . . 58
5.6. If the Display is Lost or Posts a “Display Disabled” Message . . . . . . . . . . . . . . . . . . . 58
5.7. If the “USB Host Connected” Message Persists After Disconnection . . . . . . . . . . . . . 58
5.8. “Batteries Hot or Overheated” Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.9. If You Forgot the Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.10. Calibration Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.10.1. Calibration Takes Very Long to Complete or Does Not Finish at All . . . . . . 59
5.10.2. The Centroid, FWHM, or Calibration Adjustment Value Has Changed
Dramatically Since the Last Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.11. ActiveSync: “Couldn’t Find Any Mobile Device” . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.12. Troubleshooting MAESTRO-Related Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5.12.1. MAESTRO-32 Does Not Connect with the Detective-EX . . . . . . . . . . . . . . . 60
5.13. Other Problems with the Detective-EX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
INDEX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
viii
Safety Instructions and Symbols
This manual contains up to three levels of safety instructions that must be observed in order to
avoid personal injury and/or damage to equipment or other property. These are:
DANGER Indicates a hazard that could result in death or serious bodily harm if the safety
instruction is not observed.
WARNING Indicates a hazard that could result in bodily harm if the safety instruction is not
observed.
CAUTION Indicates a hazard that could result in property damage if the safety instruction is
not observed.
DANGER–High Voltage
ATTENTION–Refer to Manual
Please read all safety instructions carefully and make sure you understand them fully before
attempting to use this product.
Cleaning Instructions
To clean the instrument exterior:
! Disconnect the instrument from the power source.
! Remove loose dust on the outside of the instrument with a lint-free cloth.
! Remove remaining dirt with a lint-free cloth dampened in a general-purpose detergent and
water solution. Do not use abrasive cleaners.
CAUTION To prevent moisture inside of the instrument during external cleaning, use
only enough liquid to dampen the cloth or applicator.
! Allow the instrument to dry completely before reconnecting it to the power source.
ix
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
x
1. GENERAL INFORMATION
The Detective-EX/DX® Administrator’s Manual is a supplement to the Detective-EX/DX
Operator’s Manual (Part No. 931040), and contains technical information that is generally not
needed on a daily basis.
The ORTEC Detective-EX is a portable nuclide identifier using both a high-purity germanium
(HPGe) detector and a moderated neutron detector for detection and identification of
radionuclides. The Detective-DX is identical to the -EX but has no neutron detector. In both
manuals, the term Detective-EX refers to all members of this instrument family, exceptions
noted.
The user interface is a high-resolution color display coupled with a touch panel. This
combination gives an easy-to-use instrument both for reading the display and control. The result
is a lightweight and rugged instrument that delivers unmatched identification and detection for
the needs of nuclear security. The unit is shipped calibrated and ready to use.
The Detective-EX uses an advanced method1 for identifying and classifying radionuclides
according to the requirements of ANSI N42.34 and other standards, as follows:
! Industrial, such as 57Co, 60Co, 133Ba, 137Cs, 192Ir, 241Am, and 75Se
! Medical, such as 18F, 67Ga, 99mTc, 111In, 123I, 131I, 133Xe, and 201Tl
! NORM,2 such as 40K, 226Ra, 232Th, and 238U
! Nuclear, such as 233U, 235U, 237Np, 239Pu, and 252Cf
! Special Nuclear Material (SNM), such as 235U and 239Pu
Other lists of nuclides can be supplied on special order. This method is able to determine the
nuclides present very quickly based on the gamma-ray peaks in the spectrum and on other
aspects of the total spectrum. It does not operate in the traditional mode of spectrum collection,
then analysis. In the mode of radionuclide identification in inspections, the counting geometry
and shielding are not known, so the quantity of a nuclide cannot be determined.
The Detective-EX/DX’s SNM Search Mode3 lets you quickly sense the possible presence of
special nuclear materials 235U and 239Pu, even if other radioactive materials are present. It is
designed for rapid area searches as well as fast scanning of containers or vehicles. It perfectly
complements the Detective-EX’s radionuclide Identification Mode by allowing you to
expeditiously locate possible SNM and then test more rigorously using Identification Mode.
1
UCRL-MI-151849, March 2003, RadScout Hand-Held Radioactive Material Identifier, M. Rowland, et. al.
2
NORM — Naturally occurring radioactive material. NORM-Th is naturally occurring thorium and decay
products.
3
Patent pending.
1
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
SNM Search Mode also takes advantage of the fact that the comparatively innocuous 133Ba can
be used as a plutonium surrogate. Advanced users can choose to enable a “search for 133Ba”
feature within SNM Search Mode both for training purposes and for developing a site’s concept
of operations.
Before using SNM Search Mode in the field, you must develop your concept of operations for
most effective use. The Detective-EX allows you to adjust the length of the SNM data sampling
(“dwell”) time. ORTEC specialists are available to help you optimize the instrument settings;
contact your ORTEC representative or our Global Service Center.
The operating system is Microsoft® Windows® Mobile™, with the Detective-EX as the only
program running. The operation of the software is controlled by soft buttons in much the same
way as any other Windows program. A soft keyboard is automatically displayed when any
alphanumeric input is needed.
The Detective-EX can store a nearly unlimited number of Identification Mode spectra and
Search Mode count-rate “strip charts” in ORTEC file formats, ANSI N42.42 formats, or both.
See Section 1.1 for more information on file types and filenaming conventions. The files can be
stored on the internal flash ROM disk or on user-supplied, removable SecureDigital (SD)
memory cards.4 Files stored on the internal or removable memory can be downloaded via the
DATA port to any PC running under Microsoft Windows 2000 Professional SP4, XP®
Professional SP2 or higher, or Vista® Ultimate. See the “File Transfer to PC” chapter in the
accompanying Operator’s Manual. Alternatively, if using memory cards, you can copy the
spectrum files directly to your PC with a memory card reader and Windows Explorer.
The Detective-EX comes with either a docking station or the more compact PAC-II power
adaptor/ charger (collectively referred to as charger bases). Both types of charger base contain
the charger for the internal battery. The docking station’s fans both cool the charger unit and
dissipate heat from the Detective-EX when the unit is docked. The docking station can also hold
a weak (~37 kBq) 137Cs source used to verify the energy calibration.
Checking the Detective-EX’s calibration is easy, whether you have a docking station with its
built-in 137Cs source; a PAC-II charger, which does not include a calibration source; or are
working in the field under battery power. The Detective-EX allows you to choose the calibration
source that best suits your needs — even the 40K naturally present in concrete or low-sodium salt
substitute.
On PCs running under Windows 2000, XP, or Vista, you can use MAESTRO®-32 MCA
Emulator Software (A65-B32) v6.05 or later, communicating via CONNECTIONS-32 v6.09.06 or
4
Some earlier models also support CompactFlash™ (CF) memory cards.
2
1. GENERAL INFORMATION
later,5 to view the search and identification spectrum files from the Detective-EX. In addition,
you can use MAESTRO-32 to control the data collection in the MCA and store the resulting
spectra to disk as ORTEC spectrum files. For more information on these operations, see
Chapter 3.
IMPORTANT
The Detective-EX is designed to function with the Regional Settings in
Windows Mobile set to the English (United States) region and all of its
default settings, as discussed in Section 5.13. Using other than the
factory default settings can cause operational problems.
To save the current gamma-ray spectrum or search strip chart in stored-spectrum memory, press
the Save button.
The Spectra dialog (Section 2.6.5.3) allows you to select the spectrum file formats and storage
location; and view the contents of the stored-spectrum memory (Fig. 29 on page 26).
5
To determine the CONNECTIONS-32 version being used by MAESTRO-32, open File/About MAESTRO....
3
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
The .SPC file contains three parameters in addition to the contents specified in the ORTEC
Software File Structure Manual for DOS and Windows Systems (Part No. 753800; hereinafter
called the File Structure Manual):
YYYY_MM_DD_hh_mm_sss
Search Mode “Strip Charts” — Search Mode data are “strip charts” of count rate vs. elapsed
time. These can be stored in the ORTEC .CHN format; or in ANSI N42.42 format, formatted for
an InstrumentType of PersonalRadiationDetector.
The .CHN format is simpler than the .SPC format, and is described in the File Structure Manual.
The .N42 file includes the sample description and a count rate log for the gamma strip chart; the
count rate log for the neutron strip chart (if the unit is so equipped); and GPS location log (if so
equipped).
YYYY_MM_DD_HH_MM_SSStype
When saving in the ORTEC .CHN format, two files are saved for each search, one for the gamma
count and one for the neutron count; type is either G for the gamma-ray chart or N for the
neutron chart.
4
1. GENERAL INFORMATION
When saving in ANSI .N42 format, only one file is generated; type is B indicating that it
contains both the gamma-ray chart and (for units so equipped) the neutron chart.
5
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
6
2. OPERATION DIALOGS AND RADIONUCLIDE
IDENTIFICATIONS
2.1. Survey Mode
The Survey Mode dialog (Fig. 1) displays the gamma count,
neutron count and gamma dose rates; the Detective-EX
overall Status; the battery status, and the current power
source (e.g., internal battery, external power source, dock-
ing station or PAC-II power adaptor/charger; see Fig. 2).
Date/Time, Location, Battery Status, Storage Space — The top line in the upper right corner
alternates between the date/time and the latitude/longitude. The lower line switches between
from the battery status (charging vs. time to full discharge) and the number of spectra that can be
stored in the currently selected storage location.
Instrument Status Area — The center of the Survey Mode screen posts messages about the
Detective-EX’s hardware status. The Status is either READY (Fig. 1) or Detector is warm. In
addition, warnings such as the high gamma dose rate message are posted in this part of the
screen.
7
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
2.1.1.1. Auto-Calibration
The Detective-EX can be configured to automatically check its own calibration. When the
auto-calibration routine is running, the screen appears as shown in Fig. 3. The Time remaining
readout in the center of the screen shows the estimated time remaining to complete the auto-
calibration sequence. For more detailed information, see Section 2.6.2.
8
2. OPERATION DIALOGS AND RADIONUCLIDE IDENTIFICATIONS
2.2.2. Save
To save the displayed search results, tap Save. If the Ask
for Sample Description option is enabled on the Spectra Fig. 6. The Search Mode Strip
setup screen (Section 2.6.5.1), a soft keyboard will open so Chart with Fixed Y-Axis Scaling.
9
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
you can enter a description for this data acquisition; tap Set to complete the save. If saving in
ORTEC .CHN format, two count-rate “strip chart” files are saved for each search, one for the
gamma count (labeled with a G suffix) and one for the neutron count (with an N suffix). For
ANSI .N42 files, one file with both gamma and neutron count rate (with a B suffix) is created.
The filenames are automatically generated (Section 1.1.2) and saved according to the settings on
the Spectra screen (Section 2.6.5).
2.2.3. Pause/Continue
To pause a search acquisition, tap Pause. The Pause button
will change to Continue, as shown in Fig. 7. Tap Continue
to resume the search. The display will be continuous, that is,
it will not show a gap for the paused time.
Fig. 7. The Search Mode, Paused.
2.2.4. Identify
To go directly to the Identification Mode, tap Identify. If you have not already saved the current
Search data, they will be lost when you change modes.
10
2. OPERATION DIALOGS AND RADIONUCLIDE IDENTIFICATIONS
high-probability reading means certainty, while an unstable reading during a fast scan is
analogous to a survey meter reading. Higher is more significant than lower.
First you scan the area looking for the site with the highest stable readings. Upon locating a
potential hotspot, you can then run Identification Mode to verify the presence of a suspect
radionuclide with very little risk of an incorrect identification (i.e., a false positive). This takes
longer than the SNM search operation but yields a final determination as to whether or not SNM
is present in detectable quantities.
Data acquisition dwell time and alarm threshold parameters are set on the SNM Search dialog
under Advanced Settings; see Section 2.6.9.
For a walkthrough of a successful SNM search, see the SNM Search Mode discussion in
Chapter 3 of the Operator’s Manual.
The Detective-EX nuclide identification scheme uses a “peak quality factor” to quantify the
quality of a gamma-ray peak in the spectral data. This quality factor, Q, is defined as:
where:
G = the gross counts in a region of the spectrum centered at the gamma-ray peak energy
with a width based on the FWHM.
B = the number of background counts in the same region
σN = the uncertainty associated with the value G!B. The numerator of this equation
represents the signal or net counts in the peak. The denominator is the statistical error
expected in the signal.
In the standard radionuclide Identification Mode, one of the tests applied to the peaks in the
spectrum is “does the value of σN exceed a threshold value which is internal to the software and
can be different for different nuclides?” This is a “necessary but not sufficient” criterion for
identification.
11
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
In SNM Search Mode, there is no threshold value as such. The value of σN for 235U and 239Pu
(and 133Ba) is computed and re-computed in real time, and plotted on the display as the square of
the computed value.
If no SNM (or 133Ba) is present, the display bar graph “bounces” around statistically. The degree
to which it bounces depend on the background count rate and the dwell-time setting. When a
source is encountered, the reading stabilizes and the closer the instrument is to the source, the
higher is the value of the index displayed, just as for a traditional rate meter but this time
specific to 235U or 239Pu (or 133Ba).
Once the high and stable reading has been verified, Identification Mode can then be used, while
keeping the Detective-EX in a fixed position. A confirmatory identification means that all of the
required conditions have been met, and not just the Q value used to locate suspicious material in
SNM Search Mode.
NOTE The existing spectrum is erased before the new data are collected. If you wish to
save the data from the most recent identification or search session, you must do
so before starting the new identification session.
12
2. OPERATION DIALOGS AND RADIONUCLIDE IDENTIFICATIONS
During data acquisition, a simulated green LED flashes at the top left of the screen. On units
with GPS capability, the top right of the screen alternates between the elapsed acquisition time
for the current spectrum and the GPS coordinates. In addition, the neutron count rate and the
gamma dose rate continually update.
Tap Save to end this identification acquisition and save the data to a .SPC file (see
Section 2.4.2). To end without creating a .SPC, tap Back. This immediately returns you to the
Survey Mode. The current spectrum data is retained in memory until the next identification or
search. You can display the spectrum (and decide whether or not to save it) from the Survey
Mode screen.
2.4.1.1. No Preset
To pause an identification acquisition, tap the Pause button. In pause mode, the Pause button
changes to Continue and the simulated LED is hidden (see Fig. 7). While paused, you can
display and/or save the spectrum.
13
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
To resume the identification process, tap Continue. When acquisition resumes, the LED
indicator reappears.
NOTE Tapping Pause or Continue does not clear the spectrum. Pausing simply suspends data
collection. When you continue, subsequent data are added to the existing spectrum
until you stop acquisition by tapping Back to return to the Survey Mode screen. The
spectrum is retained in memory until you start the next identification or search
acquisition. If you have not already viewed and/or saved the spectrum, you can do so
from the Survey Mode screen.
During the preset counting interval, Pause/Continue works the same as when counting without
a preset.
After the preset time has elapsed and counting has ended, the Pause button changes to
Continue. At this point, you may either Save the spectrum, go Back to the Survey Mode screen,
or extend acquisition for another period equal to the preset. (You may extend the count time
even if you have already saved the spectrum.) d
Example: If we set a preset of 15 seconds and tap Identify, the Elapsed Time readout at the top
right of the screen counts up from 0 of 15 to 15 of 15, data acquisition stops, and the Pause button
changes to Continue. If we then tap Continue, the readout updates to 15 of 30 and counts up to
30 of 30, at which point data acquisition again stops.
NOTE Tapping Pause or Continue does not clear the spectrum. Pausing simply suspends data
collection. When you continue, subsequent data are added to the existing spectrum
until the next preset is reached or you tap Back to return to the Survey Mode screen.
The spectrum is retained in memory until you start the next identification or search
acquisition. If you have not already viewed and/or saved the spectrum, you can do so
from the Survey Mode screen.
2.4.2. Save
Tap Save to save the current identification spectrum and results in the selected data storage
location and file format (see Sections 2.6.5.3 and 2.6.5.4, respectively). The results are saved in
the spectrum file and can be viewed both on the Detective-EX (Section 2.6.5.5) and in
14
2. OPERATION DIALOGS AND RADIONUCLIDE IDENTIFICATIONS
MAESTRO-32. An identifying text description can also be entered if the Ask for Sample
Description feature is enabled on the Spectra dialog (Section 2.6.5.1).
Tap Save to save the current spectrum and identification results in the selected data storage
location.
2.4.4. Search
To go directly to the Search Mode, tap Search. If you do not Save the current spectrum and
identification results data before going to Search Mode, these data will be lost.
2.4.5. Viewing the Found and Suspect Nuclides — “Classify” ID Display Mode
In the Classify ID Display Mode, to see the list of specific nuclides found, tap any Found
or Suspect message (see the examples in Fig. 9). Suspects are nuclides for which the
characteristic spectral lines are present, but not enough counts have been accumulated for
definitive identification.
15
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
Found nuclides (high-confidence IDs) and suspect IDs are displayed as shown in Figs. 13
and Fig. 14, respectively. Section 2.7.2 lists the Identification Mode messages for the
Classify mode. Section 2.8 shows the list of radionuclide IDs in the current library.
Tap Save to save the current spectrum and identification results to disk. To select the data
storage location and format(s), see Section 2.6.5.
Fig. 13. View Found Nuclides. Fig. 14. View Suspect Nuclides.
16
2. OPERATION DIALOGS AND RADIONUCLIDE IDENTIFICATIONS
You can also view identification spectra stored on the Detective-EX’s flash ROM
disk, My Documents folder, or removable storage card, as described in Section 2.6.5.5.
2.5.1.1. Full/Zoom
In the expanded view, the Full button restores the display to 8000 channels. In the full view, the
Full button changes to Zoom and zooms the display to the last horizontal expansion before Full.
In this way, you can easily look at the peaks in the spectrum in detail (Zoom in) and move
quickly from energy to energy in Full mode.
2.5.1.2. <>
The <> button (Zoom in) expands the display by halving the number of channels, so the peaks
appear wider. The center channel of the new display will be the marker position.
17
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
2.5.1.3. ><
The > < button (Zoom out) contracts the display by doubling the number of channels, so the
peaks appear narrower. The center channel of the new display will be the marker position or
channel 0 is the leftmost corner of the display or channel 8192 is the rightmost corner of the
display.
2.5.1.4. Lin/Log
The Lin button changes the vertical display to linear mode and automatic vertical limits. The
Lin button changes to Log in the linear mode. Tap Log to change the vertical scale to log mode.
2.5.2. Save
Tap Save to save this spectrum and identification results to the selected data storage location
(see Section 2.6.5.3).
2.5.3. Pause/Continue
If the Detective-EX is acquiring data, you can pause the data collection and resume it later,
according to the instructions in Section .
If a password has been set, the screen shown in Fig. 16 will be displayed. Tap in the password
text box to show the soft keyboard. Enter the password and tap OK. If you entered the correct
password, the first Advanced Setup dialog will be displayed (Fig. 17).
If the password was incorrect, the Password invalid message will be displayed. Try to enter the
password again, or tap Back to return to the Survey Mode dialog. If no password was set,
the Detective-EX will go directly to the first Advanced Setup dialog (Fig. 17).
18
2. OPERATION DIALOGS AND RADIONUCLIDE IDENTIFICATIONS
2.6.1. Calibrate
The Detective-EX offers both manual and automatic calibration features. Calibrate is the
manual calibration feature, which you can use at any time to check or change the calibration.
The calibration source to be used must be pre-set within the Cal. Settings function, as discussed
in Section 2.6.2. The Detective-EX has two default source entries, 137Cs or 40K. You can
optionally define up to four sources.
You can abort the calibration process at any time; no changes will be made to the calibration
coefficients (i.e., the existing calibration will still be in effect).
At the end of the calibration period, a message will indicate that the calibration (1) is OK, (2) is
usable but can be improved, or (3) has failed. If the calibration succeeded, the display will
show the energy of the calibration source peak. Ideally, the Centroid value should be:
! 137
Cs 661.62 keV
! 40
K 1460.75 keV
19
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
Fig. 18. Start Manual Calibration Fig. 19. Waiting for Calibration to
Using Source Peak Energy. Complete.
or the user-defined peak energy entered on the Cal. Settings setup screen under Advanced
Settings.
In addition, the screen will display the full width at half-maximum (FWHM), an indicator of
peak shape; and the calibration adjustment (Cal. Adjust). The calibration adjustment is the
difference between the spectrum position of the calibration source’s peak in the new calibration
and the position of the same peak in the current calibration, expressed in energy, as a percentage
of the actual peak energy.
If the calibration fails, an Unable to Check Calibration! message will be displayed. For
more detailed calibration troubleshooting, see Section 5.9.
When the calibration is satisfactory, tap Back to return to the Advanced Setup dialog.
20
2. OPERATION DIALOGS AND RADIONUCLIDE IDENTIFICATIONS
2.6.1.4. Back
If you do not want to apply the new calibration, tap Back
to return to the Advanced Setup dialog.
To define up to four sources, tap the Setup... button to open the screen shown in Fig. 22. Tap in
the Source Name field and use the soft keyboard to enter the name, tap in Energy Peak fields
21
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
and enter the peak energy in keV, then tap Back to return to the calibration settings
screen. The newly created entry will now be available on the source list.
To set the auto-calibration interval, tap the Schedule lists and select the desired time of day as
well as one of the following intervals (Fig. 23):
! Daily.
! Weekly; choose the Day of the week.
! Monthly; tap in the Day of the month field and use the soft keyboard to enter the day.
During auto-calibration, the screen displays the time remaining to completion of the calibration,
as shown in Fig. 3. This number is the real time for the calibration spectrum acquisition, based
on the peak count rate.
Tap Back to interrupt the auto-calibration and return to the Survey Mode. The Detective-EX will
continue trying to calibrate until the calibration peak passes the quality test.
If you disconnect it from the charger during auto-calibration, the routine will attempt to run to
completion and must be aborted. When the unit is reconnected to the power adaptor, auto-
calibration will start at the next scheduled interval.
22
2. OPERATION DIALOGS AND RADIONUCLIDE IDENTIFICATIONS
2.6.3. Audio
This feature (Fig. 24) lets you adjust the number of counts per click for the clicker signals and
the volume. As soon as you make your selection. The neutron sensitivity is lower than the
gamma sensitivity because the neutron count rate is lower.
Fig. 24. Adjust the Audio Fig. 25. Confirm Cooler Off.
Sensitivity.
NOTE The Detective-EX internal battery does not have enough power to cool down
the detector from room temperature, therefore, external power is required. In addition,
an exhausted internal battery should be charged for 2 hours before the Detective-EX is
cooled using external power. If you are using a standard 12 V auto-mobile battery
as the external power source, the detector cooling can significantly discharge the
battery. Cooler startup without external power cause operational anomalies; these are
discussed in troubleshooting Sections 5.5 and 5.6.
23
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
NOTE We recommend that the internal battery be recharged before it is exhausted. There is a
chance that operating until the battery goes dead could result in the loss of data stored
on the internal flash ROM disk (data on SD or CF card will not be affected), and the
instrument could require the reinitialization procedure discussed in Section 5.4.1.
You can turn the cooler off for 10–20 minutes without significantly warming the detector. As
soon as the cooler is restarted, the Detective-EX typically shows a Ready status within a few
seconds.
The cooler can be restarted at any time, without harming the detector crystal, regardless of
whether the detector is warm, partially cooled, or cooled to the operating temperature range.
2.6.5. Spectra
Figure 26 shows the Spectra dialog. You can select
the spectrum description options and storage location,
choose spectrum file formats, view the list of stored
spectrum files, display a stored spectrum and its identi-
fication results, or delete a spectrum file.
Tap in the dialog to display the soft keyboard and tap Set to save your entry.
When the Ask dialog is shown, the default description is in the dialog. You can add to it or
change it completely. This is a good way to enter the main portion of the description ahead of
time, then add a unique identifier, such as a sequence number, to each spectrum as it is acquired.
24
2. OPERATION DIALOGS AND RADIONUCLIDE IDENTIFICATIONS
Tap the down arrow to open a list of available locations. If SD or CF cards are not installed, they
will not be on the list. Tap the location you want.
Note that spectrum files stored on the built-in flash ROM disk and user-supplied memory cards
are non-volatile, that is, they will be retained even if the Detective-EX internal batteries are
exhausted. However, in some earlier Detective-EXs, the contents of the My Documents folder
could be volatile, meaning any files stored there might be lost if the batteries are exhausted.
The File Save Format feature (Fig. 28) lets you choose to generate these files as ORTEC file
formats, ANSI N42.42 format files (compatible with ANSI N42.42-2006), or both. For more
detailed information on file formats and filenaming conventions, see Section 1.1.3.
Fig. 27. Select the Data Location. Fig. 28. Select the Data Format(s).
25
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
2.6.5.5. Display
Tap Display to display a stored spectrum. This will open
the Choose File dialog (Fig. 29), displaying the Data Loca-
tion and the list of spectra in that location. The display
controls are discussed in Section 2.5. Use the scrollbar to
browse up and down through the list of spectra.
To view the spectrum, tap the name and then tap Choose.
2.6.5.6. Delete
To delete a stored spectrum, tap Delete. This opens the Choose File dialog (Fig. 29), displaying
the Data Location and the list of spectra in that location.
Use the scrollbar to browse up and down through the list of spectra.
2.6.6. Status
The Status dialog (Fig. 30) allows you to monitor the
hardware diagnostic and data acquisition parameters.
The top three fields show the Live Time, Real Time,
and Dead Time for the most recent Identification Mode
acquisition. The Real Time is the time elapsed since
the data acquisition was started. The Live Time is the
amount of time that the Detector is available to accept
Fig. 30. Status Dialog.
26
2. OPERATION DIALOGS AND RADIONUCLIDE IDENTIFICATIONS
another pulse (i.e., is not busy), and is equal to the real time minus the Dead Time (the time the
Detector is not available).
The remaining nine indicators are live (i.e., they continually update to reflect current hardware
conditions). For the first few seconds after you open this screen, the nine live indi-
cators normally read Waiting, after which they acquire their actual current values or status
messages (e.g., OK, Error, Waiting, Off).
On startup and until the detector has cooled and the HV has automatically turned on, Detector
Temp, Cold Tip Temp, and Bias Voltage read Error. Once cooldown is complete and the HV
automatically turns on, these indicators change to OK. When the cooler is switched off, the Cold
Tip Temp and Cooler Drive indicators read OFF.
The OK indicator means the parameter is within the allowed limits. The Error indicator means the
parameter is outside the limits and must be corrected before the Detective-EX can be used. If the
Waiting indicator persists beyond a few seconds or if you receive an Error reading for any
of these parameters after the nominal cooldown time has elapsed, contact your ORTEC
representative or our Global Service Center.
When all parameters are within allowed limits, the Survey Mode screen shows the Ready status
message.
2.6.7. Search
This dialog (Fig. 31) allows you to set the data Collection
Time for Search Mode. Search Mode tracks the total counts
vs a specified period of time and presents the results in a
“strip chart” emulation. The total counts in each time period
are represented on the vertical scale, and the time period
always occupies one pixel width on the display (that is,
changing the dwell time does not change the width of the
lines on the strip chart). As soon as you choose a new
dwell setting, the change will go into effect.
2.6.8. Settings
The Settings (Fig. 32) dialog allows you to change the
Administrator Mode password, lock access to Display
Mode so users cannot view the spectrum, and turn on the
beep when a button is tapped. Fig. 31. Search Time per
Horizontal Unit Selection.
27
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
NOTE There is no master password so be sure to record your password, including uppercase
and lowercase letters. If you lose this password, contact your ORTEC representative or
our Global Service Center for assistance.
28
2. OPERATION DIALOGS AND RADIONUCLIDE IDENTIFICATIONS
NOTE If spectra are saved (whether or not the Display Mode is locked), a properly configured
PC and MAESTRO-32 can be used to view the saved spectra. In such cases,
the Detective-EX, and any memory cards containing spectra, must be secured from
unauthorized access.
2.6.8.4. About
This dialog (Fig. 34) shows the instrument type, firmware and application software version
numbers, and serial number. Tap Back to return to the Settings dialog.
Fig. 34. The About Dialog. Fig. 35. Configure the SNM
Search.
29
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
To set the Q Alarm Threshold for each gamma emitter, tap the radioisotope name, tap in the
threshold field to open the soft keyboard, enter the desired value, and tap Set. Figure 35 shows
the setup dialog for Pu-239.
Figure 36 shows the setup dialog for Neutron count rate, in counts per second. Tap the
radioisotope name, tap in the threshold field to open the soft keyboard, enter the desired
value, and tap Set.
To enable the Ba-133 search (Fig. 37), mark the Include in Search checkbox and enter the
desired alarm threshold value, then tap Set. When this feature is disabled, the Ba-133 bar graph
is not displayed on the SNM Search Mode screen.
Fig. 36. Enter the Neutron Count Fig. 37. Enable the Ba-133 Search
Rate Threshold. and Alarm Threshold.
2.6.10. Exit
The Exit button is used to close the Detective-EX software application and go to the PDA
operating system. This is not normally used, but could be needed for special applications. A
confirmation dialog is displayed before the program exits. To return to the Detective-EX
program without exiting, tap Cancel.
30
2. OPERATION DIALOGS AND RADIONUCLIDE IDENTIFICATIONS
To restart the Detective-EX application, tap Start, then locate and tap on the DetectiveEX
program shortcut. (The DetectiveEx executable file is located in the My Device\Program Files\
DetectiveEX folder.)
Fig. 38. Advanced Setup, Second Fig. 39. Set the Dose Rate Units
Page. and Alarm Thresholds.
To set either Threshold, tap in the field to display the soft keyboard. Enter the desired value
then tap Set.
To change units, tap either the µSv/h or mrem/h radio button to mark it.
31
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
2.6.13. Identify
This dialog (Fig. 40) lets you set a counting preset;
and choose to display radionuclides by category (e.g.,
Industrial, Medical, NORM) or by radioisotope name.
After the preset time has elapsed and counting has ended, the Pause button changes to
Continue. At this point, you may either Save the spectrum, go Back to the Survey Mode screen,
or extend acquisition for another period equal to the preset. (You may extend the count time
even if you have already saved the spectrum.) For more information and an example, see
Section 2.4.1.2.
NOTE Tapping Pause or Continue does not clear the spectrum. Pausing simply suspends data
collection. When you continue, subsequent data are added to the existing spectrum
until the next preset is reached or you tap Back to return to the Survey Mode screen.
32
2. OPERATION DIALOGS AND RADIONUCLIDE IDENTIFICATIONS
33
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
If the count rate is within certain limits, but the criteria for positive identification have not yet
been met, you might also see:
If the count rate is >30,000 counts per second, you will see:
If uranium has been found you will see one of the following (these are mutually exclusive):
34
2. OPERATION DIALOGS AND RADIONUCLIDE IDENTIFICATIONS
! Pu
! Pu, including Am 241
! Pu - Can’t determine type
If the count rate is >30,000 counts per second, you will see:
! possibly Cd109
! Am241, Count for > 3 minutes
In addition, any individual nuclide found will be listed by name (e.g., Am241, K40, Cf252-249).
35
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
36
2. OPERATION DIALOGS AND RADIONUCLIDE IDENTIFICATIONS
2.7.3.14. Count rate is high. May require longer count or more distance
This message is displayed when the Detective-EX spectrometer is extremely “busy” processing
pulses from a high-activity source (count rate >30,000 counts per second). In such situations,
positive identifications can take much longer than expected. If possible, increase the detector-to-
source distance until this message is no longer displayed.
37
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
38
3. USING THE DETECTIVE-EX WITH A PC AND
ORTEC SPECTRUM ANALYSIS PROGRAMS
The Detective-EX is intended for completely independent operation, with no PC required.
However, when you connect the Detective-EX CONTROL port to a PC running an ORTEC
CONNECTIONS-32 MCA emulation program, you can use the Detective-EX as a high-
performance HPGe spectrometer/digital MCA. Note that only limited hardware adjustments are
allowed. The Detective-EX will not allow you to modify hardware settings, in the MCA
emulation program, in ways that might hamper the unit’s performance as an identifier. For
example, you cannot change amplifier gain, conversion gain, and other similar settings.
Installing the accompanying MAESTRO-32 MCA Emulator, v6.05 or later, will make it possible
for you to use the Detective-EX with both current and earlier releases of other CONNECTIONS-32
programs such as our GammaVision®-32 Gamma-Ray Analysis and MCA Emulator or
ISOTOPIC-32 Program for Radioactive Waste Characterization. Using CONNECTIONS-32 driver
v6.09.06 or later will allow you to view the stored spectrum reports described in Section 3.3.6.
The MAESTRO-32 Software User’s Manual (Part No. 777800) contains complete instructions
on software installation, network protocol setup, and configuration of the MCBs attached to your
PC. When installing MAESTRO-32, choose the USB-Based Instruments selection to install the
correct driver for the Detective-EX.
The DATA port connects the internal computer (running Windows Mobile) to another Windows
PC so you can copy spectra from the Detective to the PC. In this mode, you do not have access
to the MCA hardware functions, only the files stored on the instrument. For Windows XP and
2000 users, the DATA mode requires Microsoft ActiveSync, which must be downloaded and
installed before use. Windows Vista has a built-in mobile-to-PC communications capability; no
additional software is required. See the “File Transfer to PC” chapter in the Operator’s Manual
for instructions. When connected via the DATA port, the Detective-EX is listed on your PC as a
mobile device.
The CONTROL port connects the Detective-EX’s MCA to the PC and disables the other
features of the instrument. This allows you to use the unit as a high-performance HPGe MCA
controlled by MAESTRO-32 and other ORTEC CONNECTIONS-32 spectroscopy applications.
Within approximately 10 seconds after connecting via the CONTROL port, the Detective-EX
screen changes to the display shown in Fig. 42.
39
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
Note, however, that the Detective-EX is not designed to store efficiency information in the .SPC
files it creates. Therefore, for quantitative determinations, special attention is needed to prepare
the calibration (.CLB) file so that GammaVision-32 or ISOTOPIC-32 will add the efficiency
information during processing.
6
If, for some reason, the driver cannot be auto-located, it is installed in c:\Program Files\Common Files\ORTEC
Shared\UMCBI.
40
3. USING THE DETECTIVE-EX WITH A PC AND ORTEC SPECTRUM ANALYSIS PROGRAMS
To prepare this calibration file, load a Detective-EX calibration spectrum into GammaVision-32
or ISOTOPIC-32. Use the software application’s Calibration Wizard to obtain the additional
efficiency calibration, keeping the existing energy calibration. After the efficiency calibration is
computed, save the complete set of calibrations as a .CLB file.
To add the efficiency information to an existing .SPC file (typically from a standalone
Detective-EX):
! Transfer the .SPC file from the Detective-EX to the PC (see the “File Transfer to PC” chapter
in the Operator’s Manual).
! Load the .CLB file containing the proper efficiency information. This can be done either with
the application’s Recall Calibration command; or by specifying the .CLB file as the
calibration override in the GammaVision-32 analysis options (.SDF) file or the ISOTOPIC-32
configuration.
NOTE If the .CLB file is specified in the GammaVision .SDF file or ISOTOPIC configuration,
the .An1 file generated during analysis will contain the proper calibration information
but the original .SPC file (without the efficiency information) will not change.
To load the efficiency information before acquiring a new spectrum (with the Detective-EX
attached to a PC):
! Load the .CLB file either with the Recall Calibration command; or by specifying the .CLB
file as the calibration override in the GammaVision .SDF file or ISOTOPIC configuration.
41
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
Note that as you enter characters in the data-entry fields, the characters will be underlined until
you move to another field or until 5 seconds have lapsed since a character was last entered.
During the time the entry is underlined, no other program or PC on the network can modify this
value.
NOTE The changes you make on most property tabs take place immediately. There is no
cancel or undo for the dialog.
3.3.1. Amplifier
Figure 43 shows the Amplifier tab. This
tab displays the settings for the Coarse
and Fine Gain, which are factory preset
and cannot be changed. The fine gain
changes when the unit is recalibrated.
3.3.2. Amplifier 2
Figure 44 shows the Amplifier 2 tab,
which displays the fixed Rise Time
and Flattop Width settings. You can
also use the InSight Virtual Oscillo-
scope™ mode to view the digital signal
processor’s actual sampled waveform
on a reference graticule; however, the
shaping settings cannot be adjusted.
42
3. USING THE DETECTIVE-EX WITH A PC AND ORTEC SPECTRUM ANALYSIS PROGRAMS
3.3.4. ADC
This tab (Fig. 46) allows you to monitor
the current Real Time, Live Time, and
Count Rate. In addition, you can view
the fixed Conversion Gain and the
Lower and Upper Level Discrim-
inator settings.
7
Patent pending.
43
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
3.3.6. Reports
This tab (Fig. 48) displays the reports
available for the current live spectrum
or spectrum file, including the ID,
Found Nuclides, Suspect Nuclides,
Top Lines, and (for instruments so
equipped) GPS reports. You must be
running MAESTRO-32 or other
ORTEC programs that use CON-
NECTIONS-32 v6.09.06 or later.
For more information on the contents of these reports, see the section entitled “Displaying
Stored Spectra” in the Operator’s Manual.
8
A reference to our SMART-1® detector technology. For more information on SMART-1, see the ORTEC catalog
or visit www.ortec-online.com.
44
3. USING THE DETECTIVE-EX WITH A PC AND ORTEC SPECTRUM ANALYSIS PROGRAMS
3.3.7. About
This tab (Fig. 49) displays hardware
and firmware information about the
Detective-EX as well as the data
Acquisition Start Time (Sample
description is not used). In addition,
the Access field shows whether the
Detector is currently locked with a
password by MAESTRO-32. Read/
Write indicates that the Detector is
unlocked and Read Only means it is
locked. See the note concerning
passwords on page 40, at the end of
Section 3.1.
Fig. 49. Detective-EX About Tab.
Use the Firmware Revision field to
determine both the type of Detective-EX (regular or -100 model, with or without neutron
detector) and firmware version. This parameter is formatted DETX-nnn, where X is C for the
regular Detective-EX, D for the Detective-EX-100, F for the regular Detective-DX, and G for the
Detective-DX-100; and nnn is the firmware version. In the accompanying illustration, DETF-304
indicates a regular Detective-DX running firmware V3.04. You can also use the About tab to
view this information for .SPC-format spectra downloaded from the Detective-EX.
3.3.8. Status
Figure 50 shows the Status tab. Ten
parameters are continuously monitored
in real time. Satisfactory status is re-
ported as OK or as a numerical va2lue. A
failure is reported as ERR or a descrip-
tive message. Use the listboxes to select
any six parameters to be displayed
simultaneously on the Status tab. You
can change the selected parameters at
any time.
45
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
! Gamma Dose Rate — reported in µSv/h, and indicating whether this reading was made
with the detector (Ge for low dose rates) or the Geiger-Müller tube (GM for high dose rates).
See Gamma Dose-Rate Determination, page 50, for further discussion of the two dose-rate
meters.
! HV Bias — in volts.
To resolve status problems, refer to the troubleshooting list in Chapter 5. For further assistance,
contact your ORTEC representative or our Global Service Center.
3.3.9. Presets
Figure 51 shows the Presets tab. The
presets can only be set when the
Detective-EX is not acquiring data
(during acquisition the preset field
backgrounds are gray indicating that
they are inactive). You can use any
or all of the presets at one time. To
disable a preset, enter a value of zero.
If you disable all of the presets, data
acquisition will continue until man-
ually stopped.
46
3. USING THE DETECTIVE-EX WITH A PC AND ORTEC SPECTRUM ANALYSIS PROGRAMS
(disconnected from the PC). When operating as a standalone identifier all presets are
cleared. When the Identify button is pressed, the Detective-EX will always continue to
take data until the Stop button is pressed.
When more than one preset is enabled (set to a non-zero value), the first condition met during
the acquisition causes the Detector to stop. This can be useful when you are analyzing samples
of widely varying activity and do not know the general activity before counting. For example,
the Live Time preset can be set so that sufficient counts can be obtained for proper calculation
of the activity in the sample with the least activity. But if the sample contains a large amount of
this or another nuclide, the dead time could be high, resulting in a long counting time for the
sample. If you set the ROI Peak preset in addition to the Live Time preset, the low-level
samples will be counted to the desired fixed live time while the very active samples will be
counted for the ROI peak count. In this circumstance, the ROI Peak preset can be viewed as a
“safety valve.”
The values of all presets for the currently selected Detector are shown on the Status Sidebar.
These values do not change as new values are entered on the Presets tab; the changes take place
only when you Close the Properties dialog.
Enter the Real Time and Live Time presets in units of seconds and fractions of a second. These
values are stored internally with a resolution of 20 milliseconds (ms) since the Detector clock
increments by 20 ms. Real time means elapsed time or clock time. Live time refers to the amount
of time that the Detector is available to accept another pulse (i.e., is not busy), and is equal to the
real time minus the dead time (the time the Detector is not available).
Enter the ROI Peak count preset value in counts. With this preset condition, the Detector stops
counting when any ROI channel reaches this value unless there are no ROIs marked in the
Detector, in which case that Detector continues counting until the count is manually stopped.
Enter the ROI Integral preset value in counts. With this preset condition, the Detector stops
counting when the sum of all counts in all channels for this Detector marked with an ROI
reaches this value. This has no function if no ROIs are marked in the Detector.
Marking the Overflow checkbox terminates acquisition when data in any channel exceeds 231!1
(over 2×109) counts.
47
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
48
4. SPECIFICATIONS9
System Conversion Gain The Detective-EX/DX operates at a conversion gain of 8192
channels.
Shaping-Time Constants
! Rise time: 3.4 µs, fixed.
! Flattop: 0.8 µs width, fixed.
Linearity
! Integral Nonlinearity: <±0.025% over top 99.5% of spectrum, measured with a mixed source
(55Fe at 5.9 keV to 88Y at 1836 keV).
! Differential Nonlinearity: <±1% (measured with a BNC pulser and ramp generator).
Overload Recovery At maximum gain, recovers to within 2% of rated output from ×1000
overload in 2.5 non-overloaded pulse widths (measured using our InSight™ Virtual
Oscilloscope).
Pulse Pileup Rejector Automatically set threshold. Pulse-pair resolution typically 500 ns.
Low Frequency Rejector (LFR) Filter12 The digital signal processor incorporates ORTEC’s
exclusive LFR filter, designed to reduce microphonic and low-frequency periodic noise (if
present) from germanium detector output signals, thereby producing significant improvement in
spectral resolution. The LFR is specifically beneficial in systems cooled by mechanical coolers,
9
Subject to change without notice.
10
Ron Jenkins, R. W. Gould, and Dale Gedcke, Quantitative X-Ray Spectrometry (New York: Marcel Dekker,
Inc.), 1981, pp. 266–267.
11
The CF Option model also supports CompactFlash memory cards.
12
Patent pending.
49
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
which are known to often contribute microphonic noise to the spectrum (thereby degrading
spectral resolution). The LFR is also capable of reducing any periodic noise signal resulting
from surrounding electronics and equipment or ground loops.
Neutron Detector Internal, 4 each, 3He tubes, 1.3 cm × 10 cm active length, 20 atm, stainless
steel body, RS-P4-0404-214, inside 2.5 cm × 18 cm × 14 cm moderator.
Internal Compensated Geiger-Müller Tube LND 71322 Geiger Müller tube. GM tube
specifications are available on request.
The Detective-EX uses two detectors to determine the gamma dose rate over a wide range from
<0.05 µSv/h to >10000 µSv/h, a dose-rate range of around six decades. For low dose rates,
below ~20 µSv/h, the dose rate is determined from the Ge detector spectrum. For dose rates
above this value, the internal compensated GM tube is used. The Detective-EX switches
between the two automatically.
For radiation fields from extended sources (e.g., from a container full of NORM) or far-field
point sources (>1 ft [0.3 m] from the instrument), the precise location of the detector in use is
not important. When point sources are close to the instrument, the detector readings might not be
in agreement because they are in different places within the instrument, and the dose rate is
varying rapidly with position. It will be different at these two detector locations.
NOTE The dose rate function in the Detective-EX should not be considered as adequate for
accurate dosimetry purposes. Its primary purpose is to protect the user of the Detective-
EX from high radiation fields.
If you must test the dose-rate measurement capability with a point source, especially at close
range, it is important to know the reference points of the instrument detectors, since test
procedures such as ANSI or IAEA prescribe use of point sources at specified distances from the
50
4. SPECIFICATIONS
reference point of the instrument. For the Ge detector, the reference point is the geometric center
of the Ge detector endcap disk (see Fig. 52). For the GM and neutron detectors, the reference
point is shown in Fig. 52 and is marked on the case of the instrument.
Fig. 52. Reference Points for the HPGe and Neutron Detectors and the Geiger-Müller Tube (all
measurements in inches).
Presets Multiple presets can be set within MAESTRO-32 for any or all of the following:
! Real Time Stops data collection when the real time reaches this value (in increments
of 1 s).
! Live Time Stops data collection when the live time reaches this value (in increments
of 1 s).
! ROI Integral Integral count stops data collection when the sum of all ROI channels
reaches this value.
! ROI Peak Peak count stops data acquisition when any ROI channel reaches this value
(maximum value of 232!1 counts [4 ×109]).
51
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
! Data Overflow Terminates acquisition when data in any channel exceeds 231!1 (2 ×109)
counts.
Detective-EX 100: 39.4 cm L × 18.3 cm W × 34.9 cm H (15.5 in. × 7.2 in. × 13.1 in.)
Detective-DX 100: 39.4 cm L × 16.3 cm W × 34.9 cm H (15.5 in. × 6.55 in. × 13.75 in.)
! Weight
Detective-EX: 11.75 kg (25.9 lb). Detective-EX 100: 12 kg (26.3 lb).
DX versions are 0.9 kg (1.9 lb) lighter.
Maximum Shock Tolerance 20 g. An accelerometer inside the instrument tracks the maximum
shock to which the Detective-EX has been subjected, and a shock in excess of 20 g will void the
warranty.
Internal Battery Rechargeable, nominal 14.4-V lithium-ion battery pack, ORTEC Part
No. 798780. Charging circuitry and battery management circuitry internal to the docking station
or PAC-II. Battery lifetime 3–5 years. Access door in right side panel of case. The compartment
is semi-sealed and protects against penetration of dirt and moisture. See Chapter 6 for
instructions on replacing the battery.
Battery Life >3 hours at 25°C when HPGe detector is cold. Battery life can be extended
indefinitely by the use of optional, external battery belt. The unit is expected to be kept running
once cold.
Note that the internal battery charge is not sufficient to start the Detective-EX cooler or cool a
warm detector; external power is required. In addition, an exhausted internal battery should be
charged for 2 hours before the Detective-EX is cooled using external power. Cooler startup
without external power and/or with an exhausted internal battery can cause operational
anomalies. If the internal battery is completely exhausted (but can still hold a charge), see
52
4. SPECIFICATIONS
troubleshooting Section 5.4.1. For other startup issues, see Sections 5.5 and 5.6. Also note that
using an automobile battery to cool the detector could significantly discharge the auto battery.
4.2. Connectors
DATA Rear-panel USB connector, with dust cover, for PC communications.
CONTROL Rear-panel USB connector, with dust cover, for MCA communications.
INPUT POWER Rear-panel external power input, with dust cover, 12–17 V dc, 60 W, or from
battery or auto-sensing ac-powered battery charger.
Headphones ( ) Rear-panel female connector, with dust cover, accepts standard male
headphone jack. Headphone volume controllable from the Detective-EX buttonpad. Note that
the audible high-dose-rate alarm sounds whether or not headphones are connected to the
Detective-EX. This alarm is not routed through the headphones and its volume is independent of
the headphone volume setting.
53
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
Charger Status LEDs Front-panel red and green LEDs. When the Detective-EX’s internal
battery is charging, the red indicator is continuously lit. The red indicator flashes if you
disconnect the Detector before its battery is fully charged, or if the Detective-EX is connected
but the internal battery is faulty. When the battery is fully charged, the green indicator turns on
and the red indicator turns off.
4.5. PC Prerequisites
In addition to completely independent, standalone operation, the Detective-EX can also be
operated as a benchtop MCA, in conjunction with ORTEC CONNECTIONS-32 software such as
MAESTRO-32, by connecting it to any PC running Microsoft Windows 2000 Professional, XP
Professional SP2 or higher, or Vista Ultimate, and equipped with a USB port.
54
5. TROUBLESHOOTING
5.1. “Error Saving .SPC File”
This message is generated if the spectrum storage location is set to a removable card drive and
no card is in the specified drive. Insert the card and tap Save again. If you do not have a card for
the specified drive, you must stop the current identification session, enter Advanced Setup, go to
the Spectra dialog, and select a valid location.
! Turn off the MCA electronics (not the cooler) off by disconnecting from any external power
source, disconnecting any USB connection, and pressing the On/Off button. Wait 5 seconds,
then turn the unit back on. This should reset any mechanical relays that might have been
deactivated by the drop.
55
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
! Listen to the cooler: If you hear any unusual noises (rattling or pinging) emanating from the
cooler, contact your ORTEC representative or our Global Service Center.
! Monitor the unit for the next few hours to ensure that it remains Ready (indicating that the
cooler and bias voltage are functioning properly), calibrates successfully, and correctly
identifies the specified calibration source. If, on recalibration, a large adjustment is proposed
(>0.3% or more), this could indicate a problem.
If the unit fails in one or more of these respects or if one or more hardware diagnostic settings
indicate a failure (see Section 5.2.4 of the Operator’s Manual), contact your ORTEC
representative or our Global Service Center for further assistance. Remember that there are no
user-serviceable parts inside the Detective-EX and opening the case will void the warranty.
! If the internal batteries have become exhausted but can still hold a charge, use the
reinitialization procedure in Section 5.4.1.
! If the battery does not hold a charge, ensure that it is properly installed. If properly installed,
the battery must be replaced according to the instructions in Section 6. If restarting from an
exhausted battery, reinitialization might be necessary; see Section 5.4.1.
56
5. TROUBLESHOOTING
! Reinitialize.
1. If the Detective-EX is off, press the front-panel On button. After initial processing, the
display will read Microsoft Windows Powered Pocket PC (seeing this Powered Pocket
PC display is the major indicator that the internal memory has been lost and that the
instrument must be reinitialized.)
3. Follow the onscreen instructions to align the screen. Use a stylus for this as it calibrates
the screen, which should be as accurate as possible.
4. Tap the Next button and continue following the onscreen recovery instructions.
5. After successfully performing the cut-and-paste sequence, tap the Next button.
2. Change the Home time and date to the current time and date, then tap the OK at the
upper right hand side.
3. Tap the Install text; this will reinstall the Detective-EX software.
4. Finally, reset the system. Using File Explorer, navigate back to the My Device folder,
choose Program Files and the DetectiveEX folder, then tap on the DetectiveEX file to start
the software. The initial screen will be displayed; see the Operator’s Manual for more
information.
57
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
! Disconnect the Detective-EX from the charger base or external power source and from any
USB connections.
! Turn the unit off by pressing the ON button under the protective dust cover. Wait 15–20
seconds; this will return the MCA board to a normal off state.
! Press the ON button to power on the Detective-EX display. It might take a few seconds for
the display to appear, so wait at least 1 minute before trying the procedure again.
For further assistance, contact your ORTEC representative or our Global Service Center.
58
5. TROUBLESHOOTING
! Use a higher-activity calibration source or position the source on the detector endcap.
! If the calibration routine does not complete (continues running beyond one minute), tap
STOP to abort it, correct the situation, and restart the calibration.
! If the energy calibration appears highly variable, contact our Global Service Center or your
ORTEC representative. Routine changes of the order of 0.05% are to be expected. This
might simply be due to counting statistics in the spectrum. A change much larger than this,
say 0.5%, bears investigation. Apparent changes in the reported FWHM calibration are most
likely to be statistical variations. If the (gain) calibration adjustment value is stable, the unit
is functioning properly.
59
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
necessary, create an ActiveSync exception. If the PC is running under Windows XP, Service
Pack 2 or higher must be installed. Also, be sure the PC is up-to-date on all critical operating
system updates. If using an older version of ActiveSync, check the Microsoft website to see if a
more recent version is available; if all of the above conditions have already been met, installing a
newer version of ActiveSync will very likely resolve the firewall conflict.
! Make sure the Detective-EX is turned on and connected to the PC via the CONTROL port.
! Check the Master Instrument List on the PC to ensure that the Detective-EX is on the list. If
it is not listed, manually run the MCB Configuration program by clicking on Start,
Programs, MAESTRO 32, MCB Configuration. See the MAESTRO-32 User’s Manual
for detailed instructions on this operation.
! On the Advanced Setup screen (see Fig. 17, page 19), tap Exit, then confirm that you wish to
end the Detective-EX program.
60
5. TROUBLESHOOTING
! On the Windows desktop, tap the Start square in the upper left corner of the screen, then tap
Settings.
! In the Settings dialog, tap the System tab, then select Regional Settings.
! On the Region tab, select English (United States) from the listbox.
! On the Number tab, choose the period ( . ) for the Decimal symbol and the comma ( , ) for
the Digit grouping symbol.
! On the Time tab, select h:mm:ss tt as the Time style and the full colon ( : ) as the Time
separator.
! On the Date tab, use M/d/yy as the Short date, the forward slash ( / ) as the Date separator,
and the Long date format dddd, MMMM dd, yyyy.
! Tap the upper right OK circle, then the upper right X circle to return to the Windows
desktop.
For further assistance, contact your ORTEC representative or our Global Service Center.
61
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
62
6. CHANGING THE INTERNAL BATTERY
The typical service life of the Detective-EX internal battery is 3–5 years. When the internal
battery no longer retains a charge, it must be replaced with ORTEC Part No. 798780. This is a
straightforward operation that takes just a few minutes, so the detector does not significantly
warm up during the procedure. When connected to external power such as the docking station,
the new battery will typically reach full charge within 2–4 hours. If your Detective-EX was
purchased after July 1, 2005, use the following instructions. For older units, see Section 6.1.
For more information, contact your ORTEC representative or our Global Service Center.
1. Turn off the cooler and the Detective-EX electronics by reversing the instructions in
Sections 3.1 and 3.2 of the Operator’s Manual.
2. Figure 55 shows the battery hatch on the side panel. Remove the five (5) screws from the
hatch with a Torx® T10 screwdriver (there is no sixth screw under the black rubber “foot”).
Gently pry up one end of the hatch, then slip it from behind the rubber foot and expose the
battery (see Fig. 62).
NOTE If the hatch is not easily freed, it might have a sixth screw hidden under the rubber
foot. Do not force the hatch open! Go to the instructions in Section 6.1.
63
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
1. Turn off the cooler and the Detective-EX electronics by reversing the instructions in
Sections 3.1 and 3.2 of the Operator’s Manual.
2. Using a Torx T10 screwdriver, remove the 11 screws from the neutron detector housing and
take off the hatch (Fig. 57). Next, remove the two screws that fasten the detector housing to
the Detective-EX side panel.
3. Gently pull the neutron detector from the housing, being sure not to stress the two wires
that pass into the body of the Detective-EX. Figure 58 shows a closeup of the neutron
detector housing, wires, and neutron detector.
4. To separate the gold connector, grasp each half of it by the knurled areas only (not the
wires!) and pull gently; see Fig. 59.
64
6. CHANGING THE INTERNAL BATTERY
Fig. 57. Remove the Housing Cover and the Two Side-Panel Screws.
65
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
5. To separate the white connector: Grasp the male half of the connector (originating in the
neutron detector housing) with two fingers of one hand. Grasp the base of the female half
with the thumb and middle finger of your other hand. Use your index fingernail (not a
screwdriver) to gently lift and hold the “tee-shaped” tang a few millimeters above its detent.
While holding up the tang, gently separate the connector halves. See Fig. 60.
6. Place the Detective-EX on its side and use a Phillips screwdriver to remove the three screws
that hold the black rubber “foot” piece to the bottom of the Detective-EX body, as shown in
Fig. 61. Set the instrument upright.
7. With a Torx T15 screwdriver, remove the 6 screws that hold the neutron detector housing to
the front of the Detective-EX. Pull the housing and rubber foot forward 1–2 inches, taking
care not to stress the wires (Fig. 61). Typically, there is no need to pull the wires through
the neutron detector housing.
Fig. 59. Grasp the Fig. 60. Lift the “Tee-Shaped” Tang then
Knurling to Separate the Two Halves of the White
Disconnect the Connector. (Do not pull on the wires!)
Gold Connector.
66
6. CHANGING THE INTERNAL BATTERY
Fig. 61. Pull the Rubber “Foot” and Neutron Detector Housing
1–2” Away from the Detective-EX. (The dashed black arrow shows
the path of the wires.)
8. Figure 62 shows the battery hatch on the side panel; remove it using a Torx T10
screwdriver.
67
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
9. Open the tie that holds the battery in place. Note the orientation of the battery and connector
as you disconnect and remove the old battery. The connecting wire for the new
battery (refer to Fig. 56) should feed from the front of the battery compartment, across the
top edge of the battery pack (the edge closest to the top of the instrument), then toward the
back of the compartment. Close the tie around the new battery.
10. Reverse Steps 2–8 to replace the hatch and reconnect/reassemble the neutron detector
housing assembly. When reinstalling the neutron detector housing cover (see Step 2),
use a torque driver to apply 4 in-lb (.452 Nm) to the 11 screws. The cover will break if too
much torque is applied.
11. Connect the Detective-EX to external power. Remember that the cooler cannot be restarted
from the internal battery.
12. Restart the Detective-EX, turn on the cooler, and wait for the instrument status to change to
Ready.
68
7. CALIBRATION SOURCE STATEMENT
The quantity of radioactive material contained in this product is extremely small and presents no
known radiation hazard. However, it is always good practice to minimize exposure by following
the basic radiation principles of time, distance, and shielding. The solid sealed source in this
instrument requires no special handling.
Exempt quantity licensed products containing radioactive material should be used only as
intended by the manufacturer and in accordance with the instructions provided. All radioactive
materials should be securely stored when not in use.
This product may be disposed of in regular waste without regard to the radioactive content
providing the customer is not a specific licensee and all radiation symbols have been removed or
defaced. If the customer receiving the exempt quantity is a specific licensee, then the customer is
subject to the requirements of 10 CFR Part 20 in areas where 10 CFR 30.18 is silent (e.g., waste
disposal).
This product conforms to the conditions and limitations specified in 49 CFR 173.424 for
radioactive material, excepted package-instruments or articles, UN2911.
69
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
70
INDEX
Administrator Mode (advanced setup) Identification Mode
advanced setup . . . . . . . . . . . . . . . . . . . . . . 18 identification (.SPC and .N42) files . . . . . 4, 14
hiding the spectrum display . . . . . . . . . . . . . 28 nuclide identification messages . . . . . . . . . . 33
password . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 MAESTRO-32
Battery MCB Properties setup . . . . . . . . . . . . . . . . . 41
cannot cool detector . . . . . . . . . . . . . . . . . . . 23 Monitor detector state of health . . . . . . . . . . 45
exhausted, startup from . . . . . . . . . . . . . . . . 56 Monitor hardware status . . . . . . . . . . . . . . . . 45
power source indicator . . . . . . . . . . . . . . . . . . 7 troubleshooting . . . . . . . . . . . . . . . . . . . . . . 60
replacing . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 Neutron count rate . . . . . . . . . . . . . . . . . . . . . . . 31
Calibration check . . . . . . . . . . . . . . . . . . . . . . . 19 Nuclide identification messages . . . . . . . . . . . . . 33
enable auto-calibration . . . . . . . . . . . . . . . . . 21 Nuclide library . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Power
sources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 exhausted battery, startup from . . . . . . . . . . . 56
troubleshooting . . . . . . . . . . . . . . . . . . . . . . 59 power source indicator . . . . . . . . . . . . . . . . . . 7
user-defined . . . . . . . . . . . . . . . . . . . . . . . . . 21 Quantitative analysis . . . . . . . . . . . . . . . . . . . . . 40
Connecting to a PC . . . . . . . . . . . . . . . . . . . . . . 39 Reference point . . . . . . . . . . . . . . . . . . . . . . . . . 51
using MAESTRO-32 . . . . . . . . . . . . . . . . . . 41 Reports
Cooler found nuclides . . . . . . . . . . . . . . . . . . . . . . . 15
detector warmup . . . . . . . . . . . . . . . . . . . . . 54 suspect nuclides . . . . . . . . . . . . . . . . . . . . . . 15
shipping . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 top lines (most intense) . . . . . . . . . . . . . . . . . 15
shutdown for brief periods . . . . . . . . . . . . . . 24 viewing in MAESTRO-32 . . . . . . . . . . . . . . 44
startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 viewing stored spectra and reports . . . . . . . . 26
Detective-EX Search Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
ADC setup . . . . . . . . . . . . . . . . . . . . . . . . . . 43 "strip chart" (.CHN and .N42) files . . . . . . . 4, 9
amplifier settings . . . . . . . . . . . . . . . . . . . . . 42 dwell time per channel . . . . . . . . . . . . . . . . . 27
docking station . . . . . . . . . . . . . . . . . . . . . . . 53 Shipping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
exhausted battery, startup from . . . . . . . . . . 56 SNM Search Mode . . . . . . . . . . . . . . . . . . . . . . 10
high-voltage setup . . . . . . . . . . . . . . . . . . . . 44 messages . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
power adaptor/charger . . . . . . . . . . . . . . . . . 53 setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
presets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Spectrum
reinitialization (complete restart) . . . . . . . . . 56 Display Mode . . . . . . . . . . . . . . . . . . . . . . . . 16
software and firmware versions . . . . . . . . . . 29 enter a description . . . . . . . . . . . . . . . . . . . . 24
stored-spectrum memory . . . . . . . . . . . . . . . . 3 file storage location . . . . . . . . . . . . . . . . . . . 25
use as conventional MCA . . . . . . . . . . . . . . 60 filename formats . . . . . . . . . . . . . . . . . . . 3, 25
Display Disabled message . . . . . . . . . . . . . . . . . 58 hiding the spectrum display . . . . . . . . . . . . . 28
Display Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 16 identification (.SPC and .N42) files . . . . . 4, 14
hiding the spectrum display . . . . . . . . . . . . . 28 quantitative analysis . . . . . . . . . . . . . . . . . . . 40
Dose-rate detector viewing stored spectra and reports . . . . . . . . 26
reference points . . . . . . . . . . . . . . . . . . . . . . 51 zoom, marker, scaling . . . . . . . . . . . . . . . . . 17
Gamma dose rate Spectrum memory
setup and defaults . . . . . . . . . . . . . . . . . . . . 31 stored spectra . . . . . . . . . . . . . . . . . . . . . . . . . 3
WARNING!!! High Gamma Dose Rate! . . . . 7 stored-spectrum reports . . . . . . . . . . . . . . . . 26
Hardware status . . . . . . . . . . . . . . . . . . . . . . . . . 26 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . 55
71
Detective-EX/DX® Series HPGe-Based Portable Nuclide Identifiers — Administrator’s Manual
72