NOTES ON FUNCTION
GENERATOR
ÖNEM
2020/2021
FUNCTION GENERATOR
1. Introduction
The AATech AWG 1020A is a voltage waveform generator. It can be used to apply
periodic voltage waveforms to a circuit (e.g., sinusoids, square waves, triangle waves, and
pulses). You will use the function generators to supply time-varying input to circuits for
transient or frequency analysis (e.g., to characterize a filter or an amplifier).
2. Quick Notes
• Make sure the “Output” button is lit (in the lower-right corner).
• Set parameters in Ampl/Offset mode, not in HiLevel/LoLevel mode. If
HiLevel/LoLevel mode is selected, push the blue buttons below HiLevel/LoLevel until
Ampl/Offset mode is selected instead.
• The Ampl parameter sets the peak-to-peak amplitude, not the peak amplitude. See
Figure 1.
• Delivered voltage varies with output load impedance. With an infinite load impedance
at the output, the voltage amplitude delivered to the load will be double the panel
setting. See Section 3.
• Choose parameters to adjust via the blue buttons that are below the screen. Change the
parameter values using the top-right knob as well as the arrow keys; an alternative is
to use the numbered keys to input the desired value. If you do use the numbered keys,
note that after punching in the numbers, you must select the proper units and order of
magnitude using the blue buttons below the screen.
• Offset is the voltage offset for the waveform. You can use it to add a constant voltage
(DC component) to the wave (e.g. if you want to have a cosine vary around 2 V).
• The function generator’s ground is earth-grounded, i.e. the black lead from the BNC
cable is not a floating ground. Since the oscilloscope is earth-grounded as well, the
function generator and oscilloscope grounds are connected.
When displaying voltage value settings (e.g. the Ampl parameter displaying Vpp), the
function generator panel displays the voltage that will be delivered to the load at the output
under the assumption that the total load impedance is 50 Ω. The function generator also has
an internal impedance of 50 Ω, and so the load impedance forms a voltage divider with the
internal impedance of the function generator. The equivalent voltage divider is shown in
Figure 2.
Under the assumption that the load impedance is 50 Ω, the voltage delivered to the load
(the effective voltage output) is half the generated voltage of vs. Loading the output with an
impedance much greater than 50 Ω (such as connecting the output directly to an oscilloscope)
will drop the majority of the voltage across the load, and thus the measured output voltage can
be up to double the panel setting. Loading the function generator with RL < 50 Ω will cause
the voltage delivered to be less than the panel setting.
The function generator is often used to create an input signal to amplifiers. Voltage
amplifiers often have a high input impedance (e.g. an effectively infinite input impedance for
a MOS amplifier), and so the function generator will often be loaded with an infinite RL.
Thus the voltage delivered to the amplifier input will often be double the panel setting.
4. Interface Details
To begin, you should know where the power button and output port are located on the
function generator. The power button is in the lower-left corner of the front panel. The output
ports are in the lower-right corner and require a special cable. Notice that after turning on the
function generator, the display shows the value of the currently selected parameter as well as
the units and order of magnitude (e.g. mV V).
The notes below correspond to the button groups shown in Figure 3. They are generally in
the order that you would use them to set up an output waveform.
Upper-Right
Bottom-
Right
Left-
Middle
Middle
4.1.Bottom-Right Buttons
The Output buttons turn on the output voltages, which should be off when the function
generator is initially turned on. So, press the button once to turn on the output voltage.
4.2.Middle Buttons
First grey button selects the channel (CH1 or CH2) and the other five buttons allow
you to select parameters of the output waveform. After selecting a parameter, the parameter
value can be adjusted with the numbered keys, arrow buttons, and the knob located in the
upper-right corner of the front panel. Depending on the middle button selected, not all of these
five left-middle buttons will be active (i.e. not every waveform will require the use of all six
buttons).
Each left-middle button has two labels above it, with only one of the two labels lit at
any given time. The highlighted label corresponds to the parameter that is currently selected,
and its value will be shown on the display. You will only need to adjust the Freq, Ampl,
Offset, Width, and/or Duty Cycle parameters. If HiLevel and LoLevel are lit, press the button
below LoLevel or HiLevel until Ampl and Offset are selected instead.
• Freq: Frequency in repetitions per second. In the case of a pulse, this parameter
corresponds to the frequency at which repeated pulses are sent.
• Ampl: Amplitude (peak-to-peak), defined as the distance between the highest and
lowest voltage points of a waveform.
• Offset: DC voltage offset. The time-varying waveform is offseted (“raised”) by the
constant offset component (e.g. if you want a sinusoid of amplitude 1 V centered at 3
V, the offset should be 3 V).
• Width: Pulse width (for the pulse waveform).
• Duty Cycle: duration of a pulse. This control on the function generator changes the
ratio of high voltage to low voltage time in a square wave signal, i.e. changing the
waveform from a square wave with a 1:1 duty cycle to a pulse waveform, or a
triangular waveform with equal rise and fall times to a saw-tooth.
4.4.Upper-Right Buttons
In the upper-right corner of the front panel, there is a knob, two arrow buttons, and a set of
numbered buttons. Each of these offers a different way of adjusting the displayed parameter
value:
• Knob: This can be turned clockwise and counterclockwise to adjust the displayed
parameter value. Note that the knob only changes the digit (or number) selected by the
blinking cursor.
• Arrow Buttons: Press the left and right arrow buttons to manually move the blinking
cursor.
• Numbered Buttons: An alternative to the knob and arrow buttons is the set of
numbered keys. Simply punch in the desired value using the numbered key pad, and
when prompted, select the proper order of magnitude. Note that the polarity of the
output can be changed with the +/− button.