TLM8 Manual EN
TLM8 Manual EN
TLM8
2014/30/UE
SYSTEM IDENTIFICATION
KEY TO SYMBOLS
Below are the symbols used in the manual to draw the reader's attention:
Further information.
GUARANTEE
24 months from the delivery document date. The guarantee covers only defected parts and includes the replacement parts and
labour. All shipping and packing costs are paid by the customer. It is possible to have the repair in guarantee on condition that the
returned product has not been transformed, damaged or repaired without authorization. No guarantee is applicable on returned
products without the original label and/or serial number. No guarantee against misuse.
Batteries: Laumas provides 1 year guarantee from the date of delivery note, against material defects or battery manufacturing faults.
The terminals indicated on the instrument’s wiring diagram to be connected to earth must
have the same potential as the weighed structure (same earthing pit or earthing system). If
you are unable to ensure this condition, connect with an earthing wire the terminals of the
instrument (including the terminal – SUPPLY) to the weighed structure.
The cell cable must be individually led to its panel input and not share a conduit with other cables;
connect it directly to the instrument terminal strip without breaking its route with support terminal strips.
Use “RC” filters on the instrument-driven solenoid valve and remote control switch coils.
Avoid inverters in the instrument panel; if inevitable, use special filters for the inverters and separate
them with sheet metal partitions.
The panel installer must provide electric protections for the instruments (fuses, door lock switch
etc.).
It is advisable to leave the equipment always switched on to prevent the formation of condensation.
INSTALLING LOAD CELLS: The load cells must be placed on rigid, stable in-line structures; it
is important to use the mounting modules for load cells to compensate for misalignment of the
support surfaces.
PROTECTION OF THE CELL CABLE: Use water-proof sheaths and joints in order to protect the
cables of the cells.
-1-
MECHANICAL RESTRAINTS (pipes, etc.): When pipes are present, we recommend the use of
hoses and flexible couplings with open mouthpieces with rubber protection; in case of hard pipes,
place the pipe support or anchor bracket as far as possible from the weighed structure (at a distance
at least 40 times the diameter of the pipe).
WELDING: Avoid welding with the load cells already installed. If this cannot be avoided, place the
welder ground clamp close to the required welding point to prevent sending current through the load
cell body.
EARTHING THE WEIGHED STRUCTURE: By means of a copper wire with suitable cross-section,
connect the cell upper support plate with the lower support plate, then connect all the lower plates to
a single earthing system. Electrostatic charges accumulated because of the product rubbing against
the pipes and the weighed container walls are discharged to the ground without going through or
damaging the load cells. Failure to implement a proper earthing system might not affect the
operation of the weighing system; this, however, does not rule out the possibility that the cells and
connected instrument may become damaged in the future. It is forbidden to ensure earthing system
continuity by using metal parts contained in the weighed structure.
OK OK NO
NO NO OK
-2-
LOAD CELL INPUT TEST (QUICK ACCESS)
3s
From the weight display, press for 3 seconds: the top of the
display shows the gross weight; the bottom shows the response
signal of each load cell expressed in mV with three decimals.
Example: a load cell with 2.000 mV/V sensitivity provides a
response signal between 0 and 10 mV.
IF ONE OF THE ABOVE CONDITIONS IS NOT MET, PLEASE CONTACT THE TECHNICAL
ASSISTANCE SERVICE.
-3-
MAIN SPECIFICATIONS OF THE INSTRUMENT
Weight transmitter with 6-wire load cells inputs suitable for assembly on back panel fitted
Omega/DIN rail. Dimensions: 148x92x60 mm. Backlit LCD graphic display, 128x64 pixel resolution,
60x32 mm viewing area. 5-key keypad.
16 bit analog output (tension or current).
RS485 serial port for connection to: PC/PLC up to 32 instruments (max 99 with line repeaters) by
ASCII Laumas or ModBus R.T.U. protocol, remote display, printer. Optional: integrated Profibus DP,
DeviceNet, CANopen, Profinet IO, Ethernet/IP, Ethernet TCP/IP, Modbus TCP, RS232/RS485,
SERCOS III, PowerLink, CC-Link output.
-4-
TECHNICAL SPECIFICATIONS
-5-
ELECTRICAL CONNECTIONS
BASIC INFORMATION
WIRING DIAGRAM
SH
SH
SH
-SIG
+SIG
-SIG
+SIG
-SIG
+SIG
-SIG
+SIG
-REF
+REF
-REF
+REF
-REF
+REF
-REF
+REF
+EX
+EX
+EX
+EX
-EX
-EX
-EX
-EX
1 2 3 4 5 6 7 8 9 10 11 12 13
14 15 16 17 18 19 20 21 22 23 24 25 26
+REF
+REF
+REF
+REF
-REF
-REF
-REF
-REF
+SIG
+SIG
+SIG
+SIG
-SIG
-SIG
-SIG
-SIG
+EX
+EX
+EX
+EX
-EX
-EX
-EX
-EX
SH
SH
SH
SH
In case of difficulty connecting all the reference wires of the installed load cells, simply connect
those of the load cell located at the average distance from the instrument. The reference wires not
used must be individually isolated.
USE OF 4-WIRE LOAD CELLS: if all the load cells used are 4-wire, make a jumper between EX-
(10) and REF- (7) and between EX+ (17) and REF+ (20).
-6-
27
OUT 1 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42
IN 1
OUT COMMON
IN COMMON
OUT 2
+
IN 2
OUT 4
OUT 3
IN 3
OUT 5
+ 0÷10 VDC
+SUPPLY
-SUPPLY
+ 0÷20 - 4÷20 mA
-COMMON
+
-
OUTPUTS INPUTS ANALOG RS485 12/24 VDC
max 115 VAC 150 mA supply 5÷24 VDC OUTPUT SUPPLY
TERMINALS LEGEND
-7-
KEYS AND SYMBOLS FUNCTIONS
Long press
KEY Short press Into menus
(3 s)
Cancel or return to previous
Semi-automatic zero Tare resetting
menu
Select figure to be modified
Gross Net Net Gross
or go to previous menu item.
Modify selected figure or go
Weight print mV load cell test
to next menu item
Confirm or enter in
Setting setpoint and hysteresis
submenu
Setting general parameters (press
+
immediately followed by )
Setting preset tare (press
+
immediately followed by )
Load distribution display
SYMBOL Function
net weight (semi-automatic tare or preset tare)
zero (deviation from zero not more than ±0.25 divisions)
stability
unit of measure: kg
unit of measure: g
not used
The symbols are activated in sequence within the menus to indicate that the display is
not showing a weight.
-8-
MENU MAP
Into menus changes are applied right after pressing the ENTER key (no further confirmation is
required).
SETPOINT
SYSTEM PARAMETERS
-9-
3s
+
- 10 -
INSTRUMENT COMMISSIONING
Check that the display shows the weight and that when loading the load cells there is an increase in
weight. If there is not check and verify the connections and correct positioning of the load cells.
- If the instrument has already been theoretical CALIBRATED (plant system identification tag
present on the instrument and on the cover: load cell’s rated data already entered):
▫ If the system uses load cells with different sensitivity perform a real or theoretical equalization
(see section EQUALIZATION).
▫ Reset to zero (see section TARE WEIGHT ZERO SETTING).
▫ Check the calibration with sample weights and correct the indicated weight if necessary (see
section REAL CALIBRATION (WITH SAMPLE WEIGHTS)).
- If the instrument HAS NOT BEEN CALIBRATED (missing plant system identification tag)
proceed with calibration:
▫ If the system uses load cells with different sensitivity perform a real or theoretical equalization
(see section EQUALIZATION).
▫ If load cells data are unknown, follow the procedure in section REAL CALIBRATION (WITH
SAMPLE WEIGHTS).
▫ Enter the rated data of load cells following the procedure given in section THEORETICAL
CALIBRATION.
▫ Reset to zero (see section TARE WEIGHT ZERO SETTING).
▫ Check the calibration with sample weights and correct the indicated weight if necessary (see
section REAL CALIBRATION (WITH SAMPLE WEIGHTS)).
- If you use the analog output, set the desired analog output type and the full scale value (see
section ANALOG OUTPUT).
- If you use serial communication, set the related parameters (see section SERIAL
COMMUNICATION SETTING).
- If setpoint are used, set the required weight values and the relevant parameters (see sections
SETPOINT PROGRAMMING and OUTPUTS AND INPUTS CONFIGURATION).
- 11 -
PROGRAMMING OF SYSTEM PARAMETERS
From the weight display, press simultaneously keys MENU and ESC to access the parameter setting.
THEORETICAL CALIBRATION
- By modifying the theoretical full scale, the sensitivity or the division, the real calibration
is cancelled and the theoretical calibration only is considered valid.
- If the theoretical full scale and the recalculated full scale in real calibration (see section
REAL CALIBRATION (WITH SAMPLE WEIGHTS)) are equal, this means that the
calibration currently in use is theoretical; if they are different, the calibration in use is the
real calibration based on sample weights.
- By modifying the theoretical full scale or the divisions, the system’s parameters
containing a weight value will be set to default values (setpoint, hysteresis, etc.).
- 12 -
MAXIMUM CAPACITY
: maximum displayable weight (from 0 to max full scale; default: 0). When the weight exceeds
this value by 9 divisions, the display shows . To disable this function, set 0.
Perform this procedure after having set the THEORETICAL CALIBRATION data.
Use this function to set to zero the weight of the empty system after commissioning and then later
on to compensate zero variations due to the presence of product residues.
Procedure:
- Confirm the message by pressing ENTER.
- The weight value to be set to zero is displayed. In this phase all of the symbols are flashing.
- Confirming once again, the weight is set to zero (the value is stored to the permanent memory).
- Press to display the value of the total weight reset by the instrument, given by the sum of all
of the previous zero settings.
DIAGNOSTICS ON ZERO: if diagnostics on zero has been enabled, the display shows the current
load distribution and the weight value; by pressing and , the stored load distribution and the
weight set to zero are displayed (if the weight is zero and all channels are , it means that no
zero-setting has yet been performed). Press ENTER and select or to store or not
the current distribution and the zero mV values (see section TEST). Confirm by pressing ENTER to
perform a new zero-setting.
WARNING: diagnostics on zero is performed only if the load distribution has been stored
at least once.
- 13 -
REAL CALIBRATION (WITH SAMPLE WEIGHTS)
Load onto the weighing system a sample weight, which must be at least 50% of the maximum
quantity to be weighed.
By confirming the message the flashing value of the weight currently on the system is
displayed. In this phase all of the symbols are off. Adjust the value on display by using the arrow
keys if necessary. After confirming, the new set weight will appear with all the symbols flashing.
After an additional confirmation, the message will be restored and by repeatedly pressing
the key ESC the weight will once again be displayed.
Example: for a system of maximum capacity 1000 kg and 1 kg division, two sample weights are
available, one of 500 kg and the other one of 300 kg. Load both weights onto the system and correct
the indicated weight to 800. Now remove the 300 kg weight, the system must show 500; remove the
500 kg weight too; the system must read zero. If this does not happen, it means that there is a
mechanical problem affecting the system linearity.
WARNING: identify and correct any mechanical problems before repeating the procedure.
- If theoretical full scale and recalculated full scale in real calibration are equal, it means
that the theoretical calibration is currently in use; otherwise, the real calibration based on
sample weights is in use.
- If the correction made changes the previous full scale for more than 20%, all the
parameters with settable weight values are reset to default values.
- 14 -
CONFIRMATION AND CHANGE OF ACTIVE CHANNELS
After performing the calibration and verifying that the system works properly, you can confirm the
channels automatically detected by the instrument; in this way, in case of accidental interruption of
the cable of one or more load cells, the instrument displays the alarm.
After editing, you must repeat equalization, zero setting and calibration using a sample
weight.
- 15 -
EQUALIZATION
At the end of the equalization you must perform the TARE WEIGHT ZERO SETTING
and, if necessary, the REAL CALIBRATION.
REAL EQUALIZATION
Use a sample weight equal to at least 50% of the single load cell capacity.
: unload the scale, wait for stability and confirm by pressing ENTER.
: place the sample weight on load cell 1, wait for stability and confirm by pressing ENTER.
: place the sample weight on load cell 2, wait for stability and confirm by pressing ENTER.
Repeat the operation for each connected load cell.
If equalization is successfully completed, the display shows , confirm by pressing ENTER to
exit; if an error occurs, the display shows , confirm by pressing ENTER and repeat the
procedure.
THEORETICAL EQUALIZATION
÷ : set the sensitivity for each load cell, leaving it at 0 for non-active channels.
EQUALIZATION COEFFICIENTS
÷ : it displays the equalization coefficients calculated for each active channel.
EQUALIZATION DELETION
Confirmation is requested (), press ENTER to reset the equalization or press ESC to cancel
the command.
- 16 -
FILTER ON THE WEIGHT
The filter enables to stabilise a weight as long as its variations are smaller than the corresponding
“response time”. It is necessary to set this filter according to the type of application and to the full
scale value set.
The "A" filter can only be set if the instrument is connected to one load cell only.
ANTI PEAK
When the weight is stable, the anti peak filter removes any sudden disturbances with a maximum duration
of 1 second. Confirm the filter on the weight with ENTER and select one of the following options:
- : anti peak filter enabled (default);
- : anti peak filter disabled.
- 17 -
ZERO PARAMETERS
(from 0 to max full scale; default: 300; considered decimals: 300 – 30.0 – 3.00 – 0.300): this
parameter indicates the maximum weight value resettable by external contact, keypad or serial
protocol.
(from 0 to max 20% of full scale; default: 0): If at switch-on the weight value is lower than
the value set in this parameter and does not exceed the value, the weight is reset. To
disable this function, set 0.
ZERO TRACKING
(from 1 to 5, default: ): When the weight value is stable and, after a second, it
deviates from zero by a figure in divisions smaller or equal to the figure in divisions set in this
parameter, the weight is set to zero. To disable this function, set .
Example: if the parameter is set to 5 and is set to 2, the weight will be
automatically set to zero for variations smaller than or equal to 10 ( x ).
- 18 -
SETTING UNITS OF MEASURE
: kilograms
: grams
: tons
: pounds*
: newtons*
: litres*
: bars*
: atmospheres*
: pieces*
: newton metres*
: kilogram metres*
: other generic units of measure not included in the list*
If the print function is enabled, the symbol corresponding to the selected unit of measure will be
printed after the measured value.
For the units marked with * it’s possible to set also the display coefficient (parameter
, see the related section). To use is necessary to enable it, closing the
input (see section OUTPUTS AND INPUTS CONFIGURATION).
DISPLAY COEFFICIENT
If one of the inputs is set to mode (see section OUTPUTS AND INPUTS
CONFIGURATION) when the input is closed the value will be displayed modified according to the
coefficient; when the input is opened the standard weight display will be restored.
: (max settable value: 99.9999; default: 1.0000) will have different meanings according to the
value set in , i.e. the selected unit of measure. (see section SETTING UNITS OF MEASURE).
- 19 -
If the unit of measure chosen is:
: pounds, the value set in will be multiplied by the weight value currently displayed;
: newton, the value set in will be multiplied by the weight value currently displayed;
: litres, in set the specific weight in kg/l, assuming that the system is calibrated in kg;
: bar, the value set in will be multiplied by the weight value currently displayed;
: atmosphere, the value set in will be multiplied by the weight value currently displayed;
: pieces, in set the weight of one piece;
: newton metres, the value set in will be multiplied by the weight value currently
displayed;
: kilogram metres, the value set in will be multiplied by the weight value currently
displayed;
: generic unit of measure not included in the list, the value set in will be multiplied by
the weight value currently displayed.
WARNING: All other settings (setpoint, hysteresis, calibration ...) are expressed in weight
value. If you want to convert them to the new unit of measurement, perform one of the
following procedures for changing the system calibration.
The parameter must remain set to 1.0000.
- 20 -
OUTPUTS AND INPUTS CONFIGURATION
OUTPUTS
The outputs are set by default as follows: / / / / .
Possible operation modes:
- (normally open): the relay is de-energised and the contact is open when the weight is
lower than the programmed setpoint value; it closes when the weight is higher than or equal to
the programmed setpoint value.
- (normally closed): the relay is energised and the contact is closed when the weight is
lower than the programmed setpoint value; it opens when the weight is higher than or equal to
the programmed setpoint value.
- : the contact will switch on the basis of weight, according to setpoint (see section SETPOINT
PROGRAMMING).
- : the contact will not switch on the basis of weight, but is controlled by remote protocol
commands.
- : relay switching occurs when the weight is stable.
- : relay switching occurs when one of the following alarms is triggered: , ,
, , , , ; the operation mode is forced to (normally
closed).
If the operation mode is selected, the following options are also active:
- : the contact will switch on the basis of gross weight.
- : the contact will switch on the basis of net weight (If the net function is not active, the
contact will switch on the basis of gross weight).
- : relay switching occurs for both positive and negative weight values.
- : relay switching occurs for positive weight values only.
- : relay switching occurs for negative weight values only.
By confirming with ENTER the setpoint operation can be set to the value 0:
- : relay switching will not occur if the setpoint value is 0.
- :
- setpoint = 0 and switching = : relay switching occurs when the weight is 0; the
relay will switch again when the weight is different from zero, taking hysteresis into
account (both for positive and for negative weights).
- setpoint = 0 and switching = : relay switching occurs for a weight higher than or equal
to 0, the relay will switch again for values below 0, taking hysteresis into account.
- setpoint = 0 and switching = : relay switching occurs for a weight lower than or equal
to 0, the relay will switch again for values above 0, taking hysteresis into account.
- 21 -
INPUTS
Default: input 1 = input 2 = input 3 =
To perform a net operation (SEMI-AUTOMATIC TARE), close the NET/GROSS input or press the
TARE key for less than 3 seconds. The instrument displays the net weight (just set to zero) and the
NET symbol lights up. To display the gross weight again, keep the NET/GROSS input closed or
press TARE for 3 seconds.
This operation can be repeated many times by the operator to allow the loading of several products.
Example:
Put the box on the scale, the display shows the box weight; press TARE, the display shows the net
weight to zero; introduce the product in the box, the display shows the product weight. This
operation can be repeated several times.
While the net weight is displayed, keep pressed to display gross weight. When
the key is released the net weight will be displayed again.
The semi-automatic tare operation is not allowed if the gross weight is zero (the display
shows ).
- 22 -
PRESET TARE (SUBTRACTIVE TARE DEVICE)
+
It is possible to manually set a preset tare value to be subtracted from the display
value provided that the ≤ max capacity condition is verified.
By default the instrument shows the last programmed preset tare value: to apply it press and
then ENTER.
After setting the tare value, going back to the weight display, the display shows the net weight
(subtracting the preset tare value) and the NET symbol lights up to show that a tare has been
entered. To delete a preset tare and return to gross weight display, hold down TARE for about 3
seconds or keep the NET/GROSS input (if any) closed for the same length of time (3 seconds). The
preset tare value is set to zero. The NET symbol is turned off when the gross weight is displayed
once again.
While the net weight is displayed, keep pressed to display the gross weight.
When the key is released the net weight will be displayed again.
ALL THE SEMI-AUTOMATIC TARE (NET) AND PRESET TARE FUNCTIONS WILL BE
LOST WHEN THE INSTRUMENT IS TURNED OFF.
By closing the SEMI-AUTOMATIC ZERO input, the weight is set to zero; alternatively, by pressing
the 0 key for less than 3 seconds, the message is displayed for 3 seconds, by
pressing ENTER the weight is set to zero.
This function is only allowed if the weight is lower than the value (see section
RESETTABLE WEIGHT SETTING FOR SMALL WEIGHT CHANGES), otherwise the alarm
appears and the weight is not set to zero.
- 23 -
PEAK
By keeping the PEAK input closed the maximum weight value reached remains displayed. By
opening the input the current weight is displayed.
If you wish to use this input to view a sudden variation peak, set the FILTER ON THE
WEIGHT to 0.
ANALOG OUTPUT
- : it selects the analog output type (4÷20 mA, 0÷20 mA, 0÷10 V, 0÷5 V; default: 4÷20 mA).
- : choice of a weight followed by the analog output: gross () or net (). If the net
function is not active, the analog output varies according to gross weight.
- : set the weight value for which you wish to obtain the minimum analog output value.
Only set a value different from zero if you wish to limit the analog output range; for
instance: for a full scale value of 10000 kg you require an 4 mA signal at 5000 kg and
20 mA at 10000 kg, in this case, instead of zero, set 5000 kg.
- : set the weight value for which you wish to obtain the maximum analog output value; it
must correspond to the value set in the PLC program (default: calibration full scale). E.g.: if I am
using a 4÷20 mA output and in the PLC program I wish to have 20 mA = 8000 kg, I will set the
parameter to 8000.
- : analog output correction to zero: if necessary adjust the analog output, allowing the PLC
to indicate 0. The sign “-“ can be set for the last digit on the left. E.g.: if I use a 4÷20 mA output
and, with the minimum analog setting, the PLC or tester read 4.1 mA, I must set the parameter to
3.9 to obtain 4.0 on the PLC or tester.
- : correction of analog output to full scale: if necessary permit modification of the analog
output by allowing PLC to indicate the value set in the parameter . E.g. if I am using a
4÷20 mA output with the analog set to full scale and the PLC or tester reads 19.9 mA, I must set
the parameter to 20.1 to get 20.0 on the PLC or tester.
Minimum and maximum values which can be set for zero and full scale corrections:
ANALOG OUTPUT TYPE Minimum Maximum
0÷10 V -0.150 10.200
0÷5 V -0.150 5.500
0÷20 mA -0.200 22.000
4÷20 mA -0.200 22.000
- 24 -
NOTE: the analog output may also be used in the opposite manner, i.e. the weight setting that
corresponds to the analog zero ( ) may be greater than the weight set for the analog full scale
( ). The analog output will increase towards full scale as the weight decreases; the analog
output will decrease as the weight increases.
For example:
= 10000 = 0 analog output 0÷10 V
All analog outputs of the instrument are ACTIVE and SINGLE ENDED type, therefore
they can be connected only to PASSIVE receiver devices. The minimum load allowed for
voltage outputs is 2 kohm, the maximum load allowed for current outputs is 400 ohm.
- 25 -
- : continuous weight transmission protocol to RIP6100, RIP675, RIP6125C series remote
displays (set: = , = , = ).
When the remote display is set to gross weight:
- if the instrument displays the gross weight, the remote display shows the gross weight.
- if the instrument shows the net weight, the remote display shows the net weight alternated
with the message .
- : printer.
- : communication protocol with inclinometer (see Inclinometer manual).
- : transmission speed (2400, 4800, 9600, 19200, 38400, 115200; default: 9600).
- : instrument address (from 1 to 99; default: 1).
- : maximum transmission frequency (10 – 20 – 30 – 40 – 50 – 60 – 70 – 80 – 100 –
200 – 300; default: 10); to be set when the transmission protocol is selected.
Maximum setting frequency ():
- 20 Hz with minimum baud rate 2400 baud.
- 40 Hz with minimum baud rate 4800 baud.
- 80 Hz with minimum baud rate 9600 baud.
- 100 Hz with minimum baud rate 19200 baud.
- 200 Hz with minimum baud rate 38400 baud.
- 300 Hz with minimum baud rate 38400 baud.
- : delay in milliseconds which elapses before the instrument replies (from 0 to 200
ms; default: 0).
- :
- : no parity (default).
- : even parity.
- : odd parity.
- : stop bit (1 – 2; default: 1).
- : stability character ( – ; default: ); to be set when the
transmission protocol is selected in mode (see section CONTINUOUS FAST
WEIGHT TRANSMISSION PROTOCOL in Protocols manual).
- : number of copies(1 – 9; default: 1).
- : number of blank lines between one printout and the next.
- : printing of custom heading from PC ( – ; default: ).
- : connected printer type:
-
-
-
For more information about protocols and methods of communication, request the
proper manual to technical assistance.
- 26 -
RS485 SERIAL COMMUNICATION
GND
GND
GND
RS485 +
RS485 +
RS485 +
RS485 -
RS485 -
RS485 -
max 500 m
24 VDC
-
+
PC RS232
Since a two-wire RS485 output may be used directly on the RS-232 input of a PC or remote display
or printer, it is possible to implement instrument connection to an RS-232 port in the following
manner:
INSTRUMENT RS232
RS485 – → RXD
RS485 + → GND
- 27 -
AUTOMATIC DIAGNOSTICS OF LOAD DISTRIBUTION
Only use this function in systems where load distribution can be repeated with each
change of weight (for example: liquid weighing).
Load diagnostics: the instrument, with stable weight, calculates and stores the load percentage on
each channel. If under normal operation the load percentage error is higher than the value set in
parameter , the display shows the alarm alternated with the weight; the alarm
remains active also upon instrument power-off, press ENTER to cancel.
Diagnostics on zero: when a zero-setting is performed from the menu, the instrument
calculates the load percentage on each channel; diagnostics on zero is performed only if the load
distribution has been stored at least once (see section TARE WEIGHT ZERO SETTING).
If with unloaded system the load percentage error is higher than the value set in parameter
, the display shows the alarm alternated with the weight; the alarm remains active
also upon instrument power-off, press ENTER to cancel.
- 28 -
TEST
- 29 -
EVENTS LOG
- : it displays the last 50 events stored, starting from the most recent one:
: zero-setting from the calibration menu, press ENTER to display the value set to zero.
: theoretical calibration, press ENTER to display the full scale set.
: real calibration, press ENTER to display the sample weight used.
: tare setting via the keypad, press ENTER to display the set value.
: load distribution error, press ENTER to display the weight value that triggered the
alarm, press ENTER again to display the difference between the load percentage and the
stored value.
: weight alarm, press ENTER to display the alarm type.
: load distribution deletion.
: equalization.
: modification or deletion of the manual selection of active channels.
- : it deletes stored events; confirmation is requested (),press ENTER to proceed
or press ESC to cancel.
- : it prints all events.
INFO MENU
- 30 -
SETPOINT PROGRAMMING
From the weight display, press MENU to access the setpoint setting.
- (from 0 to max full scale; default: 0): setpoint; relay switching occurs when the weight
exceed the value set in this parameter. The type of switching is settable (see section OUTPUTS
AND INPUTS CONFIGURATION).
- (from 0 to max full scale; default: 0): hysteresis, value to be subtracted from the setpoint
to obtain contact switching for decreasing weight. For example with a setpoint at 100 and
hysteresis at 10, the switching occurs at 90 for decreasing weight.
These values are set to zero if the calibration is changed significantly (see sections
THEORETICAL CALIBRATION and REAL CALIBRATION (WITH SAMPLE WEIGHTS)).
- 31 -
USE WITH W SERIES INSTRUMENTS
When the TLM8 is used in combination with a W serie instrument, the load cells are connected to
the multi-channel weight transmitter, which transmits the weight to the indicator; all the operations of
calibration, zeroing, equalization and channel selection can be performed remotely through the
indicator.
- 32 -
ADDITIONAL MENU MAP
WARNING: the map shows only the additional menu items that are enabled on the W
series instruments connected to the TLM8.
Into menus changes are applied right after pressing the ENTER key (no further confirmation is
required).
- 33 -
TLM8 REMOTE CONTROL
The following functions of TLM8 can be performed remotely through the weight indicator:
INSTRUMENT COMMISSIONING
TARE WEIGHT ZERO SETTING
CONFIRMATION AND CHANGE OF ACTIVE CHANNELS
EQUALIZATION
FILTER ON THE WEIGHT
ANTI-PEAK
AUTOMATIC DIAGNOSTICS OF LOAD DISTRIBUTION
LOAD DIAGNOSTICS
DIAGNOSTICS ON ZERO
TEST
EVENTS LOG
If the configuration is performed on the TLM8, the W series instrument connected must be
restarted to allow synchronization.
ADDITIONAL ALARMS
WARNING: the list shows only the additional alarms that are enabled on the W series
instruments connected to the TLM8.
: TLM8 is not responding, check connections and serial ports settings.
: TLM8 responds incorrectly, check connections and serial ports settings.
: issues with the serial communication device.
: the instrument is synchronizing with TLM8, wait for the end of the operation.
- 34 -
ALARMS
- 35 -
Serial protocol alarms:
MODE
Bit LSB 76543210 76543210 76543210 76543210 76543210 The response to the
xxxxxxx1 xxxx1xxx xxxxxx1x xxxxx1xx On gross: zero command is a
Status 'value not valid' error
xxx1xxxx
Register (error code 3)
On net:
MODBUS RTU
xx1xxxxx
ASCII __O-F_ __O-L_ __O-F_ __O-L_ __O-F_ &aa#CR
RIP * __O-F_ __O-L_ __O-F_ __O-L_ __O-F_ __O-F_
HDRIP-N _ERCEL _ER_OL _ER_AD ###### _ER_OF O__SET
CONTIN _ERCEL _ER_OL _ER_AD ^^^^^^ _ER_OF O__SET
* For RIP remote displays, if the message exceeds 5 digits the display reads .
With an alarm the relays open and the analog outputs go to the lowest possible value
according to the following table:
RANGE 0÷20 mA 4÷20 mA 0÷5 V 0÷10 V
Output value -0.2 mA 3.5 mA -0.5 V -0.5 V
- 36 -
PRINTING EXAMPLES
If the printer has been set (see section SERIAL COMMUNICATION SETTINGS), from the weight
display press the PRINT key:
BASIC PRINTOUT
-----------------------
TLM8 BASE Addr:01
GROSS 878 kg
NET 589 kg
TARE 289 kg
GROSS 1204 kg
NET 831 kg
TARE 373 kg
PEAK 2103 kg
UNIT kg | bar
G 1195 | 1792
N 1195 | 1792
T 0 | 0
- 37 -
LOAD DISTRIBUTION PRINT
Current distribution: press the PRINT key from the menu.
Current and stored distribution: from the and menus, keep pressed the PRINT key
for more than 3 seconds while the weight is displayed.
CURRENT DISTRIBUTION
-----------------------
TLM8 BASE Addr:01
CURRENT (STATUS)
GROSS 2014 kg
CH1: 23.5 %
CH2: 24.1 %
CH3: 15.5 %
CH4: 16.7 %
CH5: 9.0 %
CH6: 10.2 %
CH7: ERR (load cell connected but in error)
CH8: OFF (load cell not connected)
- 38 -
RESERVED FOR THE INSTALLER
MENU LOCKING
Through this procedure, it’s possible to block the access to any menu on the instrument.
Select the menu that you wish to lock:
press ESC and simultaneously for 3 seconds, the display shows
(the left point on the text indicates that this menu item is now locked). If the operator tries
to enter this menu, the access is denied and the display shows .
MENU UNLOCKING
press ENTER and simultaneously for 3 seconds, the display shows
(the left point on the text is off to indicate that this menu item is unlocked).
press and simultaneously for 3 seconds: it is now possible to enter
and modify all menus including those which are locked. By returning to weight display, the menu
lock is restored.
Upon instrument power-on hold down the ESC key until the display shows , then proceed as
follows:
CONSTANTS RESTORE (does not erase the calibration): confirm , use arrow keys to select
, set code 6935 and confirm.
- 39 -
PROGRAM SELECTION: confirm and use the arrow keys to select the desired program:
: basic program, setpoint management only.
: to be used when the loaded weighing system correspond to not loaded cells and vice versa
(product increases while weight on load cells actually decreases).
: weight remote display program with setpoint.
After confirming the choice of the program (except and ), the user must choose its
approval status among the following possible choices:
*) Contact technical assistance to request the proper manual and the correct procedures for
approval, indicating mandatory hardware code and serial number (see section
INSTRUMENT COMMISSIONING).
When the TLM8 is used in combination with a W serie instrument, the approval status set
on both devices must be the same.
If you do not have a specific manual for the newly set program, you can request it
to technical assistance.
Press ESC immediately followed by hold them down for about 5 seconds (this operation is also
possible via the MODBUS and ASCII protocols):
- : no lock.
- : keypad lock: if active, when a key is pressed the message is displayed for 3
seconds.
- : keypad and display lock: if active, the keypad is locked and the display shows the
instrument model (weight is not displayed); by pressing a key the display shows for 3
seconds.
- 40 -
DECLARATION OF CONFORMITY - EU
Dichiarazione di Dichiariamo che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di
I
conformità seguito citate.
Declaration of We hereby declare that the product to which this declaration refers conforms with the following
GB
conformity standards.
Declaración de Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con
E
conformidad las siguientes normas
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden
D Konformitäts-erklärung
Normen übereinstimmt.
Déclaration de Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est
F
conformité conforme aux normes citées ci-après.
Tímto prohlašujeme, že výrobek, kterého se toto prohlášení týká, je v souladu s níže uvedenými
CZ Prohlášení o shode
normami.
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna
NL Conformiteit-verklaring
vermelde normen overeenstemt.
Declaração de Declaramos por meio da presente que o produto no qual se refere esta declaração, corresponde às
P
conformidade normas seguintes.
Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy, jest zgodny z
PL Deklaracja zgodności
poniższymi normami.
Заявление о Мы заявляем, что продукт, к которому относится данная декларация, соответствует
RUS
соответствии перечисленным ниже нормам.
Models: TLM8
M 2014/31/EU
NAWI Directive
EN 45501:2015
OIML R76-1:2006
(only if “M” mark is applied)
- 41 -